K Number
K022782

Validate with FDA (Live)

Date Cleared
2002-09-10

(19 days)

Product Code
Regulation Number
878.3300
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For the reconstruction of hernias and soft tissue deficiencies and for the temporary bridging of fascial defects.

Device Description

Biocompatible, expanded polytetrafluoroethylene (ePTFE) with a removable, tinted silicone component (DualMesh® EMERGE Biomaterial). Biocompatible, ePTFE loaded with antimicrobial preservative agents chlorhexidine diacetate and silver carbonate with a removable, tinted silicone component (DualMesh® EMERGE PLUS Biomaterial). Both devices have one open microstructure surface and one closed microstructure surface. The open microstructure surface is textured with a "ridges and valleys" pattern to aid in surface identification and proper surface orientation.

AI/ML Overview

The provided text is a 510(k) Premarket Notification summary for the DualMesh® EMERGE Biomaterial and DualMesh® EMERGE PLUS Biomaterial. This submission describes a modification to an existing, currently marketed device and claims substantial equivalence to a predicate device.

For such a submission, the "acceptance criteria" and "device performance" are typically framed in terms of demonstrating that the modified device performs equivalently to the predicate device and that the modification does not raise new questions of safety or effectiveness. The studies performed are primarily designed to establish this equivalence, not to establish novel performance metrics against pre-defined clinical acceptance criteria like those seen for entirely new devices or AI software.

Therefore, the table below reflects the equivalence claims as the "acceptance criteria" and the "reported device performance" as confirmation of that equivalence through various tests.

1. Table of Acceptance Criteria and the Reported Device Performance

Acceptance Criteria (Equivalence Claims)Reported Device Performance
Mechanical Strength & Material Characterization: Applicant devices are substantially equivalent to predicate devices.Bench test data reveal the applicant devices have mechanical strength and material characterization values that are substantially equivalent to the predicate devices.
Antimicrobial Activity (PLUS version): Applicant device functions safely and effectively to inhibit bacterial colonization for up to ten days post-implantation, equivalent to the predicate.In-vitro antimicrobial activity test data demonstrate that the antimicrobial version of the applicant device functions both safely and effectively to inhibit bacterial colonization of the device for up to ten days post-implantation. (Also stated: "perform an equivalent preservative function").
Tissue Response: Tissue response for applicant devices is equivalent to histological controls for the predicate devices.In-vivo animal test data document that the tissue response for the applicant devices is equivalent to histological controls for the predicate devices.
Biocompatibility: No change from predicate device.The biocompatibility for the applicant devices has not changed from those for the predicate device.
Packaging & Sterilization: No change from predicate device.The packaging and sterilization process for the applicant devices have not changed from those for the predicate device.
Intended Use & Indications: No change from predicate device.These changes do not change the device's intended use or indications.
Overall Safety & Effectiveness: Modification does not raise new questions of safety or effectiveness.The modification described in this Premarket Notification does not raise questions of safety or effectiveness that have not been previously addressed. Both the applicant devices and the predicate devices perform their equivalent clinical functions by incorporating biocompatible materials to permanently or transiently bridge or support a tissue defect.

2. Sample size used for the test set and the data provenance

The document does not specify exact sample sizes for the bench tests, in-vitro antimicrobial activity tests, or in-vivo animal tests. It only states that these tests were performed and their results. The data provenance is not explicitly mentioned (e.g., country of origin), but it can be inferred that these studies were conducted to support a US FDA 510(k) submission. These studies would be considered prospective in nature, as they were conducted specifically for this submission to evaluate the modified device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This type of information is not applicable to this submission. "Ground truth" in the context of expert consensus, as typically discussed for AI devices in diagnostics, is not relevant here. The studies involve laboratory measurements (mechanical properties, antimicrobial activity) and animal histological assessments, where "ground truth" is derived from established scientific methods and benchmarked against the predicate device. The experts involved would be those conducting these specialized tests (e.g., mechanical engineers, microbiologists, veterinary pathologists), but their number and specific qualifications are not detailed.

4. Adjudication method for the test set

This information is not applicable and not provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving human interpretation (e.g., radiology reads) to resolve discrepancies. The bench and animal studies described here do not involve such a process.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done

No. An MRMC comparative effectiveness study is not applicable and was not done for this device. MRMC studies are typically used for diagnostic devices, especially those incorporating AI, to assess the impact of the device on human reader performance. This submission concerns a surgical mesh biomaterial, a physical implant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable and was not done. This device is a biomaterial, not an algorithm or AI software for which standalone performance would be assessed.

7. The type of ground truth used

The "ground truth" for the various tests can be categorized as:

  • Bench Test Data: Derived from standardized testing protocols and measurements of physical and material properties, often compared directly to specifications or predicate device values.
  • In-vitro Antimicrobial Activity: Established through laboratory microbiological assays demonstrating inhibition of bacterial growth.
  • In-vivo Animal Test Data: Established through histological examination of animal tissues, assessed by veterinary pathologists, and compared to the tissue response observed with the predicate device.

8. The sample size for the training set

This information is not applicable and not provided. The concept of a "training set" is relevant to machine learning and AI, which are not involved in the development or evaluation of this biomaterial device.

9. How the ground truth for the training set was established

This information is not applicable. As no "training set" in the AI/ML sense was used, no ground truth needed to be established for it.

{0}------------------------------------------------

K022782

SEP 1 0 2002

Attachment IV Special 510(k) Premarket Notification DualMesh® EMERGE Biomaterial and DualMesh® EMERGE PLUS Biomaterial

Premarket Notification 510(k) Summary

  • A. Submitter W.L. Gore and Associates, Inc. 3750 W. Kiltie Lane P.O. Box 900 Flagstaff, AZ 86002-0900
    Contact : R. Larry Pratt

Date Submitted: August 21, 2002

  • B. Applicant Device
    Trade Name: DualMesh® EMERGE Biomaterial and DualMesh® EMERGE PLUS Biomaterial

Classification Name: Surgical Mesh

  • Applicant Device Description C.
    Biocompatible, expanded polytetrafluoroethylene (ePTFE) with a removable, tinted silicone component (DualMesh® EMERGE Biomaterial). Biocompatible, ePTFE loaded with antimicrobial preservative agents chlorhexidine diacetate and silver carbonate with a removable, tinted silicone component (DualMesh® EMERGE PLUS Biomaterial). Both devices have one open microstructure surface and one closed microstructure surface. The open microstructure surface is textured with a "ridges and valleys" pattern to aid in surface identification and proper surface orientation.

  • D. Applicant Device Indications For Use
    DualMesh® EMERGE Biomaterial and DualMesh® EMERGE PLUS Biomaterial are indicated for use in the reconstruction of hernias and soft tissue deficiencies and for the temporary bridging of fascial defects. The antimicrobial preservative agents act as preservatives, thereby inhibiting bacterial colonization of the device for up to ten days post-implantation. The removable, tinted silicone component facilitates introduction and fixation of the biomaterial. Following fixation of the biomaterial, the silicone component is easily removed.

  • E . Predicate Device
    The previously cleared GORE-TEX® DualMesh® Biomaterial and GORE-TEX® DualMesh® PLUS Biomaterial are cited as the predicate devices.

{1}------------------------------------------------

Technological Characteristics F.

This Premarket Notification submission is for a modification to an existing, currently marketed device. The modification is to attach a silicone component to the closed microstructure surface of the ePTFE biomaterial. The silicone component is removed following fixation of the ePTFE component and is not implanted.

These changes do not change the device's intended use or indications. Similarly, the biocompatibility, packaging and sterilization process for the applicant devices have not changed from those for the predicate device.

Bench test data reveal the applicant devices have mechanical strength and material characterization values that are substantially equivalent to the predicate devices.

In-vitro antimicrobial activity test data demonstrate that the antimicrobial version of the applicant device functions both safely and effectively to inhibit bacterial colonization of the device for up to ten days post-implantation.

In-vivo animal test data document that the tissue response for the applicant devices is equivalent to histological controls for the predicate devices.

Design control and verification testing have been performed for this device modification.

Safety and Effectiveness Conclusions G.

This Premarket Notification submission concerns a modification to existing, currently marketed devices. The modification is to attach a silicone component to the closed microstructure surface of the ePTFE biomaterial. The silicone component is removed following fixation of the ePTFE component and is not implanted. The implanted portion of the applicant devices is identical to the implanted portion of the predicate devices.

Bench testing, in-vivo animal testing and in-vitro antimicrobial testing demonstrate the applicant devices perform equivalent to the predicate devices.

The modification described in this Premarket Notification does not raise questions of safety or effectiveness that have not been previously addressed. Both the applicant devices and the predicate devices perform their equivalent clinical functions by incorporating biocompatible materials to permanently or transiently bridge or support a tissue defect. The antimicrobial agents loaded on both the applicant DualMesh® EMERGE PLUS device and the predicate GORE-TEX® DualMesh® PLUS device perform an equivalent preservative function by inhibiting bacterial colonization for up to ten days post-implantation.

The applicant devices are substantially equivalent to the previously cleared predicate devices.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

ED 1 0 2002

W. L. Gore and Associates, Inc. R. Larry Pratt Regulatory Affairs 3750 West Kiltie Lane P.O. Box 900 Flagstaff, Arizona 86002-0900

Re: K022782

Trade/Device Name: DualMesh® EMERGE and EMERGE PLUS Biomaterial Regulation Number: 878.3300 Regulation Name: Surgical mesh, polymeric Regulatory Class: Class II Product Code: FTL Dated: August 21, 2002 Received: August 22, 2002

Dear Mr. Pratt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 – Mr. R. Larry Pratt

This letter will allow you to begin marketing your device as described in your Section 510(k). premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours

Celia M. Witten, Ph.D., M.D.

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of I

020 510(k) Number (if known): DualMesHOEMERGE Biomaterial Device Name: DualMesh EMERGE PLUS Biomaterial.

Indications For Use:

For the reconstruction of hernias and soft tissue deficiencies and for the temporary bridging of fascial defects.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office/of Device Evaluation (ODE)

(Division Sign-Off)

Division of General, Restorative
and Neurological Devices

510(k) NumberK022782
------------------------
Prescription UseOROver-The-Counter Use
(Per 21 CFR 801.109)(Optional Format 1-2-)

(Optional Format 1-2-96)

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.