(132 days)
The indications for use of the Paros CRP are for aiding in diagnosis and monitoring of inflammatory diseases. The primary utility is for screening for the presence of inflammatory disease, by measuring CRP on anti-coagulated whole blood samples, and thus eliminating the requirement for sample centrifugation. CRP measurement on serum samples is also possible.
A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein, aids in evaluation of the amount of inflammatory injury to body tissues.
The Paros CRP is a benchtop C-Reactive Protein Immunological Test System. It is a single parameter instrument (CRP only), with the ability to measure CRP on Whole Blood and Serum samples in-vitro. It employs the same measurement principles as the CRP measurement module of the ABX / Horiba ™MICROS CRP (K002646, October 2000). The Paros CRP does not have a cell counting module.
The CRP levels are measured in patients by reacting anti-CRP antibody coated latex particles with lyzed blood, and determining the rate of the turbidimetric reaction in the near infrared spectrum.
Here's an analysis of the Paros CRP device's acceptance criteria and study, based on the provided text:
Acceptance Criteria and Device Performance for Paros CRP
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated as numerical targets in the provided document. Instead, the document describes performance metrics that demonstrate "good," "excellent," or "no significant" issues. Based on the "Discussion of Performance Data Summary" section, the following can be inferred and presented:
| Performance Metric | Acceptance Criteria (Inferred from text) | Reported Device Performance |
|---|---|---|
| Precision | "Good precision" | Total Imprecision ranged from 0 to 2.6 CV%. |
| Accuracy / Bias | "No evidence of significant bias" | No significant bias reported (assessed by NCCLS EP 9-A). |
| Correlation (Whole Blood) | "Good correlation" with predicate device (Beckman Immage) | R²=0.99 (between Paros CRP and Beckman Immage). |
| Correlation (Serum) | "Excellent correlation" with predicate device (Beckman Immage) | R²=0.98 (between Paros CRP and Beckman Immage). |
| Linearity (Whole Blood) | Supports linearity claim from 0.2 to 10 mg/dL | Demonstrated linearity across tested range (Paros CRP result = 1.05 X Expected Target CRP value, R²=0.99) for 0.2 to 10 mg/dL. |
| Linearity (Serum) | Supports linearity claim from 0.2 to 7 mg/dL | Demonstrated linearity across tested range (Paros CRP result = 0.9 X Expected Target CRP value, R²=0.98) for 0.2 to 7 mg/dL. |
| Sample Stability | Reproducibility of results over 72 hours, with acceptable deviation | Reproducibility over 72 hours. At 0.2 mg/dL, % difference attributed to single decimal point and CRP level. All other results <10% deviation from baseline. |
| Carry-over | Acceptable level of carry-over | Zero % Carry-over. |
| Interfering Substances | Performance acceptable in presence of common interfering conditions and pathologies | Investigated samples with hyperbilirubinemia, hemolysis, hypergammaglobulinemia, hyperlipemia, Multiple Myeloma, Rheumatoid Arthritis, Giant Cell Arteritis, Polymyalgia Rheumatica, Polymyositis, non-specific arthropathies, and somatic carcinomas. (No specific performance values given for these, but implicitly deemed acceptable for bias assessment). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for the test set in the precision, linearity, stability, carry-over, or correlation studies.
Regarding data provenance:
- Country of Origin: Not specified in the provided text.
- Retrospective or Prospective: Not specified in the provided text.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
The concept of "experts" and "ground truth" in the context of diagnostic device validation for CRP measurement typically refers to the reference method or predicate device used for comparison, and samples with known CRP concentrations (e.g., from calibrators or controls). The document refers to:
- Predicate Devices: ABX / Horiba™ MICROS CRP (K002646) and Beckman Coulter Immage IMMUNOCHEMISTRY SYSTEM (K981638). These devices essentially served as the "ground truth" for comparison in the correlation studies.
- Qualified Personnel: The study was "Prepared By: Dr Ian Giles ABX Diagnostics." His qualifications are not explicitly detailed beyond being "ABX Scientific Affairs Manager."
- There is no mention of a panel of human experts (e.g., radiologists) establishing ground truth, as this is an in-vitro diagnostic device for C-reactive protein, not imaging.
4. Adjudication Method for the Test Set
The provided text does not describe an adjudication method (like 2+1 or 3+1) for the test set. Such methods are typically employed in studies where human interpretation of results is involved and discrepancies need to be resolved. For an IVD device measuring a biomarker, the comparison is usually against a reference method or predicate device, not human expert consensus needing adjudication.
5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done or is applicable. This study is for an in-vitro diagnostic device (Paros CRP) that automates the measurement of C-reactive protein. It does not involve human readers interpreting images or data, nor does it involve AI assistance for human interpretation. Therefore, the concept of human readers improving with AI assistance is not relevant to this device's validation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, a standalone performance study was done. The entire evaluation described for the Paros CRP (precision, linearity, stability, carry-over, and correlation) focuses on the performance of the device itself (the "algorithm only," if you consider the instrument's measurement principles as such) without any human interpretation in the measurement process. The device directly measures CRP levels in samples.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth used for performance evaluation primarily consisted of:
- Comparison to a Predicate Device/Reference Method: For accuracy/bias and correlation studies, the Beckman Immage system served as the reference for comparison for whole blood and serum samples. The ABX / Horiba™ MICROS CRP is also listed as substantially equivalent, implying its use as a reference for methodology.
- Expected Target Values: For linearity studies, expected target CRP values were used (likely from calibrators or serially diluted samples with known concentrations).
- Internal Controls and Calibrators: The document mentions "CRP Calibrator: CRPCAL" and "CRP Control: CRPTROL (L, H)" for quality control, suggesting these are part of establishing and verifying ground truth during daily operation and potentially in some validation stages.
8. The Sample Size for the Training Set
The provided text does not mention a "training set" in the context of machine learning or AI. The Paros CRP device is described as an "Immunological Test System" that employs turbidimetric reactions. It is a traditional IVD device, not an AI-based system that requires a training set in the typical sense.
9. How the Ground Truth for the Training Set Was Established
As there is no mention of a training set for an AI/ML algorithm, this question is not applicable to the Paros CRP device as described in the document. The device operates based on established chemical and optical principles, not on learned patterns from a training dataset.
{0}------------------------------------------------
DEC 2 3 2002
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: KO 32690
- ABX Diagnostics Company: 34 Bunsen Drive Irvine CA92618 USA Phone: (949)-453-0500 Fax: (949)-453-0600 Contact Person: Ian Giles Date Prepared: May 20, 2002
Paros CRP Trade Name:
Common Name: In Vitro Diagnostic C-Reactive Protein Immunological Test System
Classification Name: C-Reactive Protein Immunological Test System
Device Classification: Class II
Regulation Number: 21 CFR (866.5270)
Substantial Equivalence:
The Paros CRP is substantially equivalent to the following devices:
- ABX / Horiba ™ MICROS CRP (K002646 October 2000) .
- Beckman Coulter Immage IMMUNOCHEMISTRY SYSTEM . (K981638 June 1998).
Description:
The Paros CRP is a benchtop C-Reactive Protein Immunological Test System. It is a single parameter instrument (CRP only), with the ability to measure CRP on Whole Blood and Serum samples in-vitro. It employs the same measurement principles as the CRP measurement module of the ABX / Horiba ™MICROS CRP (K002646, October 2000). The Paros CRP does not have a cell counting module.
The CRP levels are measured in patients by reacting anti-CRP antibody coated latex particles with lyzed blood, and determining the rate of the turbidimetric reaction in the near infrared spectrum.
{1}------------------------------------------------
Indications for Use:
The indications for use of the Paros CRP are for aiding in diagnosis and monitoring of inflammatory diseases. The primary utility is for screening for the presence of inflammatory disease, by measuring CRP on anti-coagulated whole blood samples, and thus eliminating the requirement for sample centrifugation. CRP measurement on serum samples is also possible.
Discussion of Performance Data Summary:
The determination of substantial equivalence is based on precision, linearity, stability and carry-over studies as well as an inter-procedural correlation study.
The data presented in this 510K Pre-market Notification demonstrate good precision as assessed by NCCLS EP5-A. Total Imprecision ranged from between 0 and 2.6 CV%.
Accuracy / bias assessment (NCCLS EP 9-A) showed no evidence of significant bias. Good correlation was demonstrated between the Paros CRP and the Beckman Immage for Whole Blood (R2=0.99). Similarly, the correlation for Serum samples between the Paros CRP and the Beckman Immage was excellent (R =0.98). For Whole blood samples, the linear regression formula of the trend line was Y-0.92X + 0.03. A comparison of the serum results showed linear regression trend line formula of: Y=0.95X.
Linearity assessment data supports a Whole Blood CRP linearity claim from 0.2 to 10 mg/d1; and for Serum samples, a linearity range between 0.2 and 7 mg / dl is supported.
The data shows linearity across the tested range for Whole Blood (Paros CRP result = 1.05 X Expected Target CRP value) R2=0.99; and serum: (Paros CRP result = 0.9X Expected Target CRP value) Re = 0.98.
The Sample Stability Study showed reproducibility of results over the entire 72 hour assessment period. At a level of 0.2 mg / dl, the % difference in results can be attributed to the fact that there is a single decimal point on the CRP result, and to the level of CRP testing. All other results showed <10% deviation from the baseline measurement.
There was Zero % Carry-over in this study.
Potential Interfering Clinical Conditions were investigated. Samples with hyperbilirubinemia, hemolysis, hypergammaglobulinemia, and hyperlipemia were included in the bias assessment study. Pathological diagnoses in the bias assessment study included: Multiple Myeloma, Rheumatoid Arthritis, Giant Cell Arteritis, Polymyalgia Rheumatica, Polymyositis, non-specific arthropathies, and somatic carcinomas.
Prepared By: Dr Ian Giles ABX Diagnostics
{2}------------------------------------------------
The principles of immuno-turbidimetry on the MICROS CRP and PAROS CRP are summarized below:
Image /page/2/Figure/1 description: This image shows a diagram of a three-phase process. In phase 1, whole blood cells and antigens are mixed with R1, resulting in hemolysis. In phase 2, the product of phase 1 is mixed with R2, resulting in zero interference. In phase 3, the product of phase 2 is mixed with R3, which consists of latex beads coated with antibodies, and the resulting mixture is analyzed by a sensor using a light source LED at 850nm.
Principles of CRP Measurement on Whole Blood.
:
- Phase 1 : Hemolysis, using reagent R1 ; R1 contains saponin as active principle.
- Phase 2 : Incubation of the hemolysate with R2, which inhibits potential interference during the incubation with the specific R3 reagent.
- Phase 3 : Incubation with latex beads coated with anti-CRP antibodies. The recognition of CRP by anti-CRP antibodies induces bead aggregation. This increases the turbidity (opacity) of the incubation medium. The rate of change in turbidity is measured by using a light source (850nm LED) and an optical sensor.
{3}------------------------------------------------
1. MICROS CRP SPECIFICATIONS
MICROS CRP technical specifications are summarized in Table 3.
MICROS CRP SPECIFICATIONS
]
| Parameters | |
|---|---|
| CBC/DIFF Mode | 18 parameters (16 for the US) with graphics forRBC, PLT and WBC populationsRBC, HGB, HCT, MCH, MCHC, RDWPLT, MPV, PDW*, THT*LYM, MON, NEU (% and #) |
| CRP Mode | 19 parameters (17 for USA) with graphics forRBC, PLT and WBC populationsRBC, HGB, HCT, MCH, MCHC, RDWPLT, MPV, PDW*, THT*LYM, MON, NEU (% and #)CRP |
| Reagents | |
| CBC/DIFF Mode | Three reagents :ABX MINIDIL LMG, ABX ALPHALYSE, ABXMINICLEAN |
| CRP Mode | Six reagents :ABX MINIDIL LMG, ABX ALPHALYSE, ABXMINICLEANCRP-100 (CRP-R1, CRP-R2, CRP-R3) |
| Quality Control | Calibrator (Hematology) : MinocalControl Blood : Minotrol 16 (L, H, M)CRP Calibrator : CRPCALCRP Control : CRPTROL (L, H) |
| Principles of Measurement | Spectrophotometry (HGB)Impedance (CBC/DIFF) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes extending from its back, representing the department's commitment to health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the logo.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administrati 2098 Gaither Road Rockville MD 20850
Horiba, Ltd. c/o Dr. Ian Giles ABX Scientific Affairs Manager ABX Diagnostics 34 Bunsen Drive Irvine, CA 92618
Re: K022690
Trade/Device Name: Paros CRP Regulation Number: 21 CFR 866.5270 Regulation Name: C-reactive protein immunological test system Regulatory Class: Class II Product Code: DCN Dated: November 20, 2002 Received: November 22, 2002
Dear Dr. Giles:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
INDICATIONS FOR USE STATEMENT
S10(k) Number (if known): K O22690
Device Name: Paros CRP
Indications For Use: Indications for Use:
The indications for use of the Paros CRP are for aiding in diagnosis and monitoring of inflammatory diseases. The primary utility is for screening for the presence of inflammatory disease, by measuring CRP on anti-coagulated whole blood samples, and thus eliminating the requirement for sample centrifugation. CRP measurement on serum samples is also possible.
A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein, aids in evaluation of the amount of inflammatory injury to body tissues.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | ✓ |
|---|---|
| (Per 21 CFR 801.109) |
OR
| Over-The-Counter Use | |
|---|---|
| (Optional Format 1-2-96) |
| (Division Sign-Off) | J. Reeve for J. Bautista |
|---|---|
| Division of Clinical Laboratory Devices |
| 510(k) Number | KC022690 |
|---|---|
| --------------- | ---------------------------------------------------------- |
§ 866.5270 C-reactive protein immunological test system.
(a)
Identification. A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.(b)
Classification. Class II (performance standards).