K Number
K022493
Device Name
EMS 400
Date Cleared
2002-10-11

(74 days)

Product Code
Regulation Number
890.5850
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Body Clock EMS 400 is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area. It is indicated for the following: - Relaxation of muscle spasms; - Prevention or retardation of disuse atrophy; - Increasing local blood circulation; - Muscle re-education; - Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis; and - Maintaining or increasing range of motion.
Device Description
The EMS 400 a dual channel EMS (Powered Muscle Stimulator) unit. It is identical to the Altoona EMS 400 (K913272). It has two intensity (amplitude) dials for each channel, placed on the front of the unit. Directly below the left-hand-side intensity dial is a pulse rate dial, which sets the pulse rate for both channels. Below the right-hand dial is a timer dial, which controls the time of each cycle when cycle or reciprocal mode is used. If the dial is set to 0, the unit provides a constant mode.
More Information

No
The description details a standard electrical muscle stimulator with manual controls for intensity, pulse rate, and timer. There is no mention of AI, ML, or any adaptive or learning capabilities. The device is described as identical to a predicate device from 1991 (K913272), which predates widespread medical AI/ML applications.

Yes

The "Intended Use / Indications for Use" section explicitly states that the device is "intended for medical purposes" and lists several therapeutic indications such as "Relaxation of muscle spasms" and "Prevention or retardation of disuse atrophy".

No

The device description and intended use indicate that the Body Clock EMS 400 is an electrically powered muscle stimulator used for therapeutic purposes such as muscle relaxation, prevention of atrophy, and increasing blood circulation, not for diagnosing conditions.

No

The device description explicitly states it is an "electrically powered device" and a "dual channel EMS (Powered Muscle Stimulator) unit" with physical dials, indicating it is a hardware device.

No, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to examine specimens taken from the human body (like blood, urine, tissue) to provide information about a person's health. This testing is done outside of the body ("in vitro").
  • Body Clock EMS 400 Function: The description clearly states that the Body Clock EMS 400 is an electrically powered device that contracts muscles by passing electrical currents through electrodes contacting the affected body area. This is a physical interaction with the body, not the analysis of a biological specimen.
  • Intended Use: The listed indications for use (muscle relaxation, preventing atrophy, increasing circulation, etc.) are all related to physical therapy and rehabilitation, not diagnostic testing of biological samples.

The Body Clock EMS 400 is a therapeutic device, specifically a powered muscle stimulator (EMS).

N/A

Intended Use / Indications for Use

The Body Clock EMS 400 is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area. It is indicated for the following:

  • Relaxation of muscle spasms; 1.
  • Prevention or retardation of disuse atrophy; 2.
    1. Increasing local blood circulation;
    1. Muscle re-education;
  • ನ : Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis; and
    1. Maintaining or increasing range of motion.

Product codes

IPF

Device Description

The EMS 400 a dual channel EMS (Powered Muscle Stimulator) unit. It is identical to the Altoona EMS 400 (K913272). It has two intensity (amplitude) dials for each channel, placed on the front of the unit. Directly below the left-hand-side intensity dial is a pulse rate dial, which sets the pulse rate for both channels. Below the right-hand dial is a timer dial, which controls the time of each cycle when cycle or reciprocal mode is used. If the dial is set to 0, the unit provides a constant mode.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Not Applicable as there are no new or innovative aspects that have been introduced.

Key Metrics

Not Found

Predicate Device(s)

K913272

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).

0

10 510(k) Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K022493

10.1 Submitter's Identification:

OCT 11 2002

Body Clock Health Care Ltd 108 George Lane South Woodford London E18 1AD United Kingdom +44 (0)20 8532 9551 Tel: Fax: +44 (0)20 8532 9551

Contact: Jonathan Bash Date Prepared: July 23rd 2002

10.2 Name of Device:

Proprietary Name: EMS 400

Common or Usual Name: Powered Muscle Stimulator.

Classification Name: Powered Muscle Stimulator.

10.3 Predicate Device Information:

The EMS 400 is substantially equivalent to the EMS 400 (K913272)

10.4 Device Description:

The EMS 400 a dual channel EMS (Powered Muscle Stimulator) unit. It is identical to the Altoona EMS 400 (K913272). It has two intensity (amplitude) dials for each channel, placed on the front of the unit. Directly below the left-hand-side intensity dial is a pulse rate dial, which sets the pulse rate for both channels. Below the right-hand dial is a timer dial, which controls the time of each cycle when cycle or reciprocal mode is used. If the dial is set to 0, the unit provides a constant mode.

1

10.5 Intended Use:

  • Relax muscle spasms ●
  • Prevent or retard disuse atrophy .
  • Maintain or increase range of motion .
  • Increase local blood circulation .
  • Re-educate muscle .
  • Immediate post-surgical stimulation of calf muscle to prevent venous thrombosis .

10.6 Technological Comparison to Predicate Devices:

The EMS 400 has basic technological characteristics that are substantially equivalent to the predicate devices.

This unit uses dials to change the settings. It also uses shrouded patient cable connectors to comply with the FDA's Final Rule "Medical Devices: Establishment of Performance Standards for Electrode Lead Wires and Patient Cables."

10.7 Non-clinical Testing:

All Units are fully CE marked i.e. compliant with EEC Directive 93/42/EEC Annex V, classified as "Internally powered Equipment Type BF. They are intended for continuous operation." 150 9002 ISO 13488 EN 46002 EN 60601-1-2:1993 (EEC Directive 89/336/EEC)

10.8 Clinical Testing:

Not Applicable as there are no new or innovative aspects that have been introduced.

10.9 Conclusions:

The EMS 400 has the same intended use and same technical characteristics as the EMS 400 (K913272)

The information supplied in this 510(k) illustrate that the device does not pose any new questions of safety and effectiveness. EMS 400 is substantially equivalent to the predicate device.

2

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three lines representing its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The seal is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD. 20850

Mrs. Rashelle Preston Body Clock Health Care Limited 108 George Lane South Woodford London E18 1AD United Kingdom

Re: K022493

Trade/Device Name: EMS 400 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: IPF Dated: July 23, 2002 Received: July 29, 2002 Amended: July 29, August 16, and September 4, 2002

Dear Mrs. Preston:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic

3

Page 2 - Mrs. Preston

product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k)

premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

A. Htpt. Rwelu
Colin M. Wittes, Ph.D., M.D.

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

Statement of Indications For Use 8

510(k) Number (if known):

022493

Device Name: Body Clock EMS 400

Indications for Use: The Body Clock EMS 400 is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area. It is indicated for the following:

  • Relaxation of muscle spasms; 1.
  • Prevention or retardation of disuse atrophy; 2.
    1. Increasing local blood circulation;
    1. Muscle re-education;
  • ನ : Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis; and
    1. Maintaining or increasing range of motion.

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hyt Plurda

(Division Sign-Off) Division of General, Restorative and Neurological Devices Ko22493 510(k) Number

X Prescription Use (Per 21 CFR 801.109)