K Number
K022084
Manufacturer
Date Cleared
2002-09-26

(91 days)

Product Code
Regulation Number
870.2700
Panel
AN
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The indications for use for the Atlas Monitor, model numbers 200, 210 and 220 include the monitoring o the following human physiological vital signs:

  • Blood Oxygenation (SpO2) measurement .
  • ECG waveform derived from 3 or 5 lead measurement .
  • Respiration rate/waveform derived from ECG or CO2 e
  • Temperature measurement via YSI 400 series probes .
  • Non-Invasive Blood Pressure (NIBP) measurement .
  • CO2 End-Tidal side stream/waveform .
  • Heart Rate derived from selected source (ECG, SpO2) .
    The target populations are adult and pediatic populations. The monitor is intended for use within the healthcare facility setting.
Device Description

The Atlas Monitor is a multi-parameter device used to monitor human physiological vital signs. It combines a CRT to display ECG and CO2 waveforms and LED's for other numeric values. There are (3) different models available.

FEATURESMODEL 200MODEL 210MODEL 220
ECGYesYesYes
SpO2 Nellcor MP506YesYesNo
SpO2 NoninNoNoYes
NIBPYesYesYes
TemperatureNoYesYes
Impedance RespirationNoYesOnly if ETCO₂ is not running
ETCO₂NoNoYes
PrinterOptionOptionStandard
Battery BackupNot AvailableStandardStandard
RS423 I/ONot AvailableStandardStandard

The ECG front end will meet all applicable AAMI and harmonized EU standards (see safety reference section for particular standards). The Atlas Monitor will provide a 5 wire front end and will be compatible with both a 3 wire and 5 wire cable. Diagnostic (0.05Hz to 100Hz) and monitor (0.5Hz to 40Hz) bandwidth will be provided. The corresponding ECG waveform is displayed on the CRT. User can select which lead the monitor is displaying. If optional printer is installed that waveform can be printed.
The SpO2 value is obtained by the measurement of the red and infrared light absorbed by the patient's tissue. A probe, consisting of a detector and emitter, is placed on a patient at a point where profusion of a body part is close to the skin surface, like a hand digit. The changes in absorption are used to determine oxygen saturation and heart rate. The pulse signal graph LED bar indicates the relative strength of the pulses detected by the SpO2 module. The control board requires a single voltage input and generates all the necessary internal voltages. Communication to the Atlas Main board is via an internal serial communication interface. The SpO2 capability in Atlas is obtained by the utilization of OEM modules.
The NIBP portion of the Atlas monitor utilizes the oscillometric method of blood pressure determination. In this method, the patient's arm is compressed and blood flow occluded through the use of a cuff and bladder combination. Each time the patient's heartbeats, a slight variation of pressure occurs in the cuff. The cuff pressure is decreased in a step fashion. The device measures and catalogs the pressure pulses at each step in cuff pressure. The patient's systolic and diastolic pressures are determined through the examination of these pulses.
The temperature portion of the Atlas monitor uses well established precision YSI 400 probes to monitor surface temperature of a patient. The YSI probes work on the principle of interchangeable thermistors (NTC Type) that exhibits a steep drop in resistance as temperature changes, providing high sensitivity to temperature changes.
The impedance respiration feature utilizes the ECG lead set and patient surface ECG electrodes, together with additional electronics processing. The respiration rate in breaths/minute is displayed in numerical format on the CRT.
The end tidal CO2 (ETCO>) measurement is performed by utilizing an OEM PCB from Prvon Corporation. The Pryon side stream CO2 control board (LC101) is designed to acquire CO2 data utilizing an on board pump to aspirate a patient gas sample. Employing a proprietary side stream sensor, the C-cap bench. Patient waveform is obtained. The sensor is based on single beam single frequency and dual thermopile detector. The control board requires a single voltage input and generates all the necessary internal voltages. The control board outputs the CO2 waveform and performs all calculations for CO2 data and respiration rate, Communication to the Atlas main board is via an internal serial communication interface.

AI/ML Overview

The provided text is a 510(k) Summary of Safety and Effectiveness for the Welch Allyn Atlas Monitor. It describes the device, its indications for use, and its substantial equivalence to a predicate device. However, it does not contain the detailed performance testing results, acceptance criteria, or study design information typically found in a clinical study report or a more comprehensive validation document.

Therefore, I cannot populate most of the requested fields because the information is not present in the provided text.

Here's what can be extracted:

Acceptance Criteria and Device Performance (Limited Information)

Acceptance CriterionReported Device Performance
Overall Safety and Effectiveness"The Atlas Monitor and associated accessories have been tested and found to comply with the recognized national and international performance, safety and electromagnetic compatibility standards for medical devices and product specifications listed in the Atlas labeling."
"A risk analysis, identifying potential hazards and documenting mitigation of the hazards, has been developed and verified/validated as part of Welch Allyn, Inc. product development procedures. Welch Allyn, Inc. Quality System conforms to 21 CFR 820 and is certified to ISO 9001 and EN46001."
"Welch Allyn, Inc. conclusion is that the Atlas Monitor is safe, effective,... and equivalent to the Atlas Monitor currently on the market."
ECG Performance"The ECG front end will meet all applicable AAMI and harmonized EU standards..."
SpO2 PerformanceThe improved SpO2 channel is stated to be "substantially equivalent to the SpO2 channel in the Nellcor model N-395 pulse oximeter." (This implies it meets the performance characteristics of the predicate, but specific criteria for the Atlas Monitor itself are not detailed.)

Missing Information (Not found in the provided text):

  1. Sample size used for the test set and the data provenance: Not specified.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/specified. This type of device does not typically rely on expert human interpretation for its outputs like an AI diagnostic tool would. Its performance is validated against established measurement standards for vital signs.
  3. Adjudication method: Not applicable/specified.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a vital signs monitor, not an AI-assisted diagnostic tool for human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The monitor operates standalone in terms of its vital signs acquisition and display. The summary indicates it was "tested and found to comply with recognized national and international performance, safety and electromagnetic compatibility standards," which implies standalone performance evaluation against these standards. However, specific details of these tests are not provided.
  6. The type of ground truth used: For vital signs monitors, ground truth is typically established using calibrated reference devices or validated physiological models according to recognized standards (e.g., AAMI, ISO standards). The document states compliance with "recognized national and international performance, safety and electromagnetic compatibility standards," which would implicitly define the ground truth methods used, but doesn't explicitly state them.
  7. The sample size for the training set: Not applicable. This product is a vital signs monitor, not an AI/machine learning device that requires a "training set" in the conventional sense.
  8. How the ground truth for the training set was established: Not applicable.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).