K Number
K021541

Validate with FDA (Live)

Date Cleared
2002-10-28

(171 days)

Product Code
Regulation Number
862.1555
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NeoGram PKU Tandem Mass Spectrometry Kit is intended for screening for Phenvlketonuria (PKU). It involves the quantitative determination of phenvlalanine, tyrosine and the phenylalanine to tyrosine ratio (Phe/Tyr) in blood specimens dried on filter paper. This will aid in identifying newborns with PKU. It is intended for use by trained, qualified laboratory personnel.

Device Description

The assay involves the extraction of dried blood spots with a solution containing stable, isotopically labeled internal standards. The response of each amino acid relative to the isotopically labeled standards in the kit is proportional to their actual concentration. The analysis of the material is performed on a tandem mass spectrometer. Control material contained in the kit allows for verification of performance for the test run.

AI/ML Overview

Here's the information about the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a quantitative format for the NeoGram PKU Tandem Mass Spectrometry Kit. However, it implicitly demonstrates performance through comparison with a legally marketed predicate device (Isolab Phenylalanine Test Kit) and provides descriptive statistics. The implicit acceptance criterion is likely substantial equivalence to the predicate device, demonstrated by similar performance characteristics.

Performance MetricPredicate Device (Fluorometric method)NeoGram PKU Tandem Mass MethodImplied Acceptance Criteria (relative to predicate)Device Performance (NeoGram PKU)
Mean Phenylalanine Concentration (mg/dl)1.11.1Similar mean concentration1.1 mg/dl
Standard Deviation (mg/dl)0.30.2Similar or improved precision0.2 mg/dl
Linearity (approx.)21 mg/dl (fluorometric)35 mg/dl (TMS)Equal or wider linearity range35 mg/dl (TMS)

2. Sample size used for the test set and the data provenance:

  • Sample Size: 3679 samples
  • Data Provenance: Not explicitly stated whether retrospective or prospective, or country of origin.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. The "ground truth" for the test set appears to be the results obtained by the legally marketed predicate device (Isolab Phenylalanine Test Kit), which served as the comparator.

4. Adjudication method for the test set:

This information is not provided in the document. The comparison is made against the results from the predicate device, implying a direct comparison rather than an adjudication process involving multiple human readers for discrepancies.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a diagnostic kit for quantitative determination of analytes, not an AI-based image analysis or decision support system that involves human readers in the capacity described by MRMC studies. The "AI" concept isn't relevant to this type of device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, in essence. The performance presented is for the NeoGram PKU Tandem Mass Spectrometry Kit itself, as a standalone diagnostic system, performing the quantitative determination of phenylalanine, tyrosine, and Phe/Tyr ratio. While trained laboratory personnel operate the machine, the performance metrics (mean and standard deviation) are those of the machine's output.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" used for comparison appears to be the results obtained from a legally marketed predicate device (Isolab Phenylalanine Test Kit), specifically the "Fluorometric method." This provides a benchmark for expected performance in a clinical setting.

8. The sample size for the training set:

This information is not provided in the document. The document refers to a "calibration curve" for the predicate device and "internal standard comparison" for the new device, but details on the data used to establish these are not given.

9. How the ground truth for the training set was established:

This information is not provided in the document. For analytical devices, the "training set" would typically involve calibrators and control materials with known concentrations, but the specifics of how these known concentrations were established (e.g., against a reference standard) are not detailed.

{0}------------------------------------------------

OCT 2 8 2002

510(k) Summary

This summary of 510(k) safety and effectiveness information is being in accordance with the requirements of the SMDA of 1990 and 21 CFR 807.92.

The assigned 510(k) number is:________________________________________________-

Date: 3/28/01

Submitted by: PerkinElmer Life Sciences 3985 Eastern Rd. Norton, OH 44203

Telephone: 330-825-4525 Fax: 330-825-8520

Contact person: Hank Juske, Director, Regulatory Affairs

Trade name: NeoGram PKU Tandem Mass Spectrometry Kit (MS-7960) Common name: Phenylalanine screening test kit Classification name: Phenylalanine test system

Legally marketed predicate device: Isolab Phenylalanine Test Kit

Device description: The assay involves the extraction of dried blood spots with a solution containing stable, isotopically labeled internal standards. The response of each amino acid relative to the isotopically labeled standards in the kit is proportional to their actual concentration. The analysis of the material is performed on a tandem mass spectrometer. Control material contained in the kit allows for verification of performance for the test run.

Intended use: The NeoGram PKU Tandem Mass Spectrometry Kit is intended for screening for Phenvlketonuria (PKU). It involves the quantitative determination of phenvlalanine, tyrosine and the phenylalanine to tyrosine ratio (Phe/Tyr) in blood specimens dried on filter paper. This will aid in identifying newborns with PKU. It is intended for use by trained, qualified laboratory personnel.

Similarities/ Differences with the predicate device:

Similarities:
-Intended use is the same. Both kits are for screening of newborns.
-Both kits are intended to screen for the defect PKU.
-Both measure the analyte Phenylalanine
-Both use dried blood spots on filter paper for sample collection
-Both use standards and controls dried onto filter paper
-Both assays are quantitative
-Sample size (3mm punched spot) is the same
-Both employ a 96 well microtiter plate

{1}------------------------------------------------

-Fluorometric based chemistry vs. spectrometric assay Differences: -Linearity of approximately 21mg/dl fluorometric vs. 35 mg/dl TMS -Measurement by calibration curve versus internal standard comparison -Zinc sulfate/ethanol extraction solution vs. methanol / water solution

Mean value/ standard deviation of 3679 samples run by:

Fluorometric method: X= 1.1 mg/dl S.D.= 0.3

Tandem Mass Method: X= 1.1 mg/dl S.D = 0.2

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal are three stylized human profiles facing to the right, with flowing lines representing hair or movement.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

OCT 2 8 2002

Ms. Carroll Martin Regulatory Affairs Manager PerkinElmer Life Sciences, Inc. 3985 Eastern Road Norton, Ohio 44203

Re: K021541

Trade/Device Name: NeoGram PKU Tandem Mass Spectrometry Kit Regulation Number: 21 CFR 862.1555 Regulation Name: Phenylalanine test system Regulatory Class: Class II Product Code: JNB; CDR; JJE Dated: August 21, 2002 Received: August 22, 2002

Dear Ms. Martin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for PerkinElmer life sciences. The logo consists of a stylized "P" shape, with a thick vertical bar and an arrow-like shape pointing to the right. Below the symbol is the text "PerkinElmer" in a bold, sans-serif font. Underneath "PerkinElmer" is the text "life sciences." in a smaller, lighter font.

PerkinElmer Life Sciences 3985 Eastern Rd. Norton, OH 44321 Phone: 330-825-4525 Fax330-825-8520 http://www.perkinelmer.com

INDICATIONS FOR USE STATEMENT

510(K) Number K021541

Device name: NeoGram PKU Tandem Mass Spectrometry Kit

Indications for use: The kit is intended for screening for phenylketonuria (PKU). It involves the quantitative determination of phenylalanine, tyrosine and the phenylalanine to tyrosine ratio (Phe/Tyr) in blood specimens dried on filter paper. This will aid in identifying newborns with PKU. It is intended for use by trained, qualified laboratory personnel.

San Cooper

(Division Sign-Off)

Division of Clinical Laboratory Devices K 8 21 54 510(k) Number

Concurrence of CDRH, Office of Device Evaluation (ODE)

or

Prescription Use (per 21 CFR 801.109)

Over-the-Counter Use

(Division Sign-Off)
Division of Clinical Laboratory Devices

Division of Clinical Laboratory Devices
510(k) Number K061541

§ 862.1555 Phenylalanine test system.

(a)
Identification. A phenylalanine test system is a device intended to measure free phenylalanine (an amino acid) in serum, plasma, and urine. Measurements of phenylalanine are used in the diagnosis and treatment of congenital phenylketonuria which, if untreated, may cause mental retardation.(b)
Classification. Class II.