K Number
K020945
Date Cleared
2002-05-31

(67 days)

Product Code
Regulation Number
862.1545
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AxSYM® Intact PTH is a Microparticle Enzyme Immunoassay (MEIA) for the in vitro quantitative determination of intact human parathyroid hormone (I PTH) in human serum or plasma on the AxSYM® system. The AxSYM® I PTH assay is intended for use as an aid in the differential diagnosis of hypercalcemia and hypocalcemia.

Device Description

AxSYM® I PTH is based on the Microparticle Enzyme Immunoassay (MEIA) technology. The AxSYM® I PTH assay uses paired monoclonal and polyclonal antibodies, each reactive with epitopes in the N-terminal portion of intact human PTH (1-84). The polyclonal antibody is biotin labeled. I PTH in the sample serves to bridge the two antibodies.

AI/ML Overview

The provided text does not contain a study that proves the acceptance criteria of the device. Instead, it details the substantial equivalence of the "AxSYM Intact PTH" device to a predicate device, the "DSL-8000 ACTIVE™ Intact PTH IRMA," through method correlation and comparison of various performance characteristics.

Therefore, the following points address the questions based on the information available regarding the substantial equivalence study, rather than a study specifically proving acceptance criteria.


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria in this context are not explicitly stated numerical targets for device performance, but rather the demonstration of "substantial equivalence" to a predicate device. This is primarily shown through method correlation and comparable performance characteristics.

FeatureAcceptance Criteria (Implied by Predicate Performance / Substantial Equivalence Goal)Reported AxSYM® I PTH Performance
Method CorrelationVery high correlation (r value close to 1) between AxSYM® I PTH and predicate device.r = 0.97 (p < 0.0001)
Regression SlopeSlope close to 1.0 (95% CI covering 1.0)0.99 (95% CI: 0.95, 1.05)
Regression InterceptIntercept not statistically significantIntercept not statistically significant (p < 0.35)
Precision (Within Run)Comparable to predicate device.Level I: 5.9 % CV; Level II: 6.2 % CV; Level III: 6.5 % CV
Precision (Inter)Comparable to predicate device.Level I: 9.4 % CV; Level II: 8.1 % CV; Level III: 8.8 % CV
LinearityComparable to predicate device.93.8 ± 8.9 %
RecoveryComparable to predicate device.95.3 ± 8.9 %
Analytical SensitivityEqual to or better than predicate device.≤ 2.0 pg/mL I PTH
SpecificityNo detection of human PTH fragments 53-84, 44-68 and 39-84.No detection of human PTH fragments 53-84, 44-68 and 39-84
InterferenceAcceptable interference levels across various substances.Bilirubin (20 mg/dL): < 10 %; Triglycerides (1500 mg/dL): < 10 %; Total Protein (3 mg/dL): < 10 %; Hemoglobin (1000 mg/dL): < 10 %; Red Blood Cells (0.2 % v/v): < 10 %
Dynamic Range0 - 2000 pg/mL0 - 2000 pg/mL
Expected Values8 - 57.8 pg/mL or comparable to predicate.8 - 57.8 pg/mL

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 120 samples were used for the method correlation study.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). However, given that it's a comparison for regulatory submission, it is likely data collected specifically for this purpose. The samples were "serum and plasma samples."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of information is not relevant or provided for a chemical assay. The "ground truth" for a quantitative diagnostic assay like this is typically established by measurements from a reference method or a predicate device, as seen here. There are no "experts" in the sense of human adjudicators for establishing ground truth values of PTH in samples.

4. Adjudication Method for the Test Set

Not applicable. As noted above, this involves a comparison to a predicate device, not subjective interpretation requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This is a quantitative diagnostic immunoassay, not an AI-driven imaging or diagnostic system that involves human readers or AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Yes, the performance characteristics and method correlation described are for the "AxSYM® Intact PTH" device operating in a standalone capacity (i.e., algorithm/device only). The device provides quantitative results directly from the sample. It is stated that "The I PTH assay is not used as a stand-alone test. Clinical interpretation requires simultaneous measurement of serum calcium levels." This refers to clinical utility, not device performance. The device itself performs its measurement independently.

7. The Type of Ground Truth Used

The "ground truth" for the performance evaluation, in this context of a substantial equivalence claim, is the performance and measurements obtained from the predicate device, DSL-8000 ACTIVE™ Intact PTH IRMA. The study compared the AxSYM I PTH results to those of the predicate device.

8. The Sample Size for the Training Set

Not applicable. This is a chemical immunoassay, not an AI/machine learning device that typically requires a "training set" in the computational sense. The device's calibration and internal parameters are set through manufacturing processes and control procedures, not through training on a dataset of clinical cases.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" in the context of this device technology.

{0}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS 4.0

Diagnostic Systems Laboratories, Inc. Submitter: 445 Medical Center Boulevard Webster, Texas 77598 Phone: 281-332-9678 FDA Registration # 1628193

Contact: Carroll Potts Manager of Regulatory Affairs Office: (281) 332-9678 Fax: (281) 338-1895 cpotts@dslabs.com

Date of Summary:May 22, 2002
Device Trade Name:AxSYM Intact PTH
Regulation Cite:21CFR862.1545
Classification Name:Parathyroid Hormone
Analyte Code and Name:(CEW) Parathyroid Hormone Test System
Regulatory Class:II

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 to provide sufficient detail to understand the basis for a determination of substantial equivalence.

8020945 The assigned 510(k) number is:________________________________________________________________________________________________________________________________________________

Test Description: 4.1

Parathyroid hormone (PTH, parathyrin) is synthesized as a pre-pro hormone 4.1.1 by the chief cells of the parathyroid glands [1,2]. In normal physiology, PTH levels increase in response to decreased serum calcium and decrease in response to increased serum calcium resulting in a close inverse relationship between PTH and calcium concentrations [1,6]. Abnormal PTH levels can thus be detected by examining this interrelationship. Concurrent increases in PTH and calcium occur in the various forms of hyperparathyroidism, including multiple endocrine neoplasia (MEN) types I and II, parathyroid turnors, and idiopathic hyperparathyroidism [6,7]. Concurrent declines in calcium and PTH occur in the various forms of primary hypoparathyroidism and in disorders affecting PTH secretion, such as hypo- or hyper-magnesemia [8]. Abnormally low calcium levels in the presence of extremely elevated PTH levels can occur in a group of genetic disorders called pseudohypoparathyroidism [9]. Abnormally high calcium levels in the presence of low or normal PTH levels are characteristic of malignancy-associated hypercalcemia secondary to PTH-related peptide (PTHrP) production [10]. Abnormal concentrations and/or actions of PTH with essentially normal serum calcium levels may also occur, e.g. in renal osteodystrophy [11].

NOTE: The bibliography notations used make reference to the Bibliography Section at the end of the AxSYM I PTH Package Insert.

{1}------------------------------------------------

Immunoassays for I-PTH and its two major metabolic fragments have been 4.1.2 developed. Clinical investigations indicate that the diagnostic sensitivities of the Nterminal and C-terminal assays are limited, possibly due in part to variable concentrations of minor fragments and the performance characteristics of the assays [1,4]. Selective measurement of mainly I PTH has theoretical advantages since it may be more reflective of biologically active PTH production and action.

The AxSYM® I PTH assay uses paired monoclonal and polyclonal 4.1.3 antibodies, each reactive with epitopes in the N-terminal portion of intact human PTH (1-84). The polyclonal antibody is biotin labeled. I PTH in the sample serves to bridge the two antibodies.

4.2 Principle Of The Test:

AxSYM I PTH is based on the Microparticle Enzyme Immunoassay 4.2.1 (MEIA) technology.

The AxSYM I PTH reagents and sample are pipetted in the following sequence:

4.2.2 SAMPLING

  • Sample and all AxSYM® I PTH Reagents required for one test are pipetted by . the sampling probe into various wells of a Reaction Vessel (RV).
    The RV is immediately transferred into the Processing enter. Further pipetting is done in the Processing Center by the Processing Probe.

PROCESSING 4.2.3

  • An aliquot of the reaction mixture containing the antibody-antigen complex . bound to the microparticles is transferred to the matrix cell. The microparticles bind irreversibly to the glass fiber matrix..
  • The matrix cell is washed with the TRIS Wash Buffer. ●
  • The Anti-PTH: Biotin Conjugate is dispensed onto the matrix cell and binds . with the antibody-antigen complex.
  • The matrix cell is washed to remove unbound materials. .
  • The Anti-Biotin: Alkaline Phosphatase Conjugate is then dispensed onto the . matrix cell and binds with the biotinylated antibody-antigen complex.
  • The matrix cell is washed to remove unbound materials. .
  • The substrate, 4-Methylumbelliferyl Phosphate, is added to the matrix cell and . the fluorescent product is measured by the MEIA optical assembly.

{2}------------------------------------------------

4.3 Intended Use:

AxSYM® Intact PTH is a Microparticle Enzyme Immunoassay (MEIA) for the in vitro quantitative determination of intact human parathyroid hormone (I PTH) in human serum or plasma on the AxSYM® system. The AxSYM® I PTH assay is intended for use as an aid in the differential diagnosis of hypercalcemia and hypocalcemia.

4.4 Substantial Equivalence:

The AxSYM® Intact PTH assay is substantially equivalent to another 4.4.1 device marketed in the United States. We claim equivalence to the DSL-8000 ACTIVE™ Intact PTH IRMA (K896294).

  • To demonstrate substantial equivalence between the two assays, serum and 4.4.2 plasma samples were assayed using both methods. Method correlation between the AxSYM® I PTH and the predicate device was examined using 120 samples. The resulting regression equation was determined to be:
    [AxSYM® I PTH] = 0.99 [DSL ACTIVE™ I PTH IRMA] + 10.0 r = 0.97 p < 0.0001

The results of least squares regression analysis indicated very high correlation between the two methods (r = 0.97). The 95% confidence interval about the slope is [95% CI (0.95, 1.05)]. This correlation between the two methods is highly significant (p<0.0001). The 95% CI about the intercept is [95% CI (-11.72, 31.82)]. The intercept is not statistically significant (p<0.35).

Specimen values ranged from 4 - 1961 pg/mL with the AxSYM® I PTH and 6 – 1836 pg/mL with the predicate device, DSL ACTIVE™ Intact PTH IRMA.

{3}------------------------------------------------

Similarities: The following table compares the similarities between the AxSYM® I PTH assay with the predicate device:

FeatureAXSYM® I PTHPredicate Device(K896294)
Assay PrincipleSandwich AssaySandwich Assay
Intended Usefor the in vitro quantitativedetermination of intact humanparathyroid hormone (I PTH)in human serum or plasma onthe AxSYM® systemfor the quantitativemeasurement of Intact PTH (I-PTH) in serum or plasma.
Indication for Usean aid in the differentialdiagnosis of hypercalcemia andhypocalcemiaNone stated in package insert
Sample typeHuman serum or plasmaHuman serum or plasma
Sample required185 µL200 µL
Dynamic Range0 - 2000 pg/mL0 - 2000 pg/mL
Expected Values8 - 57.8 pg/mL9 - 55 pg/mL
TraceabilityTraceable to DSL I PTHIRMATraceable to Nichols I PTHAssay

The following table compares the differences between the AxSYM I PTH Differences: assay and the predicate device:

FeatureAxSYM I PTHPredicate Device (K896294)
Assay TechnologyMicroparticle EnzymeImmunoassay (MEIA)employing the "sandwich"principleImmunoradiometric Assay(IRMA) employing the"sandwich" principle
InstrumentAxSYM ImmunoassayAnalyzerGamma Counter

{4}------------------------------------------------

Performance Characteristics:

The following table compares the assay performance characteristics between the

CharacteristicsAxSYM® Intact PTHPredicate Device (K896294)
Precision (WithinRun)Level I (35.3 pg/mL): 5.9 % CVLevel I (13.0 pg/mL): 3.1 % CV
Level II (202.9 pg/mL): 6.2 % CVLevel II (23.0 pg/mL): 3.5 % CV
Level III (649.7 pg/mL): 6.5 % CVLevel III (476 pg/mL): 1.9 % CV
Precision (Inter)Level I (35.3 pg/mL): 9.4 % CVLevel I (13.0 pg/mL): 3.1 % CV
Level II (202.9 pg/mL): 8.1 % CVLevel II (23.0 pg/mL): 5.2 % CV
Level III (649.7 pg/mL): 8.8 % CVLevel III (472.0 pg/mL): 2.6 % CV
Linearity (Mean ± SD)93.8 ± 8.9 %94.2 ± 10.5 %
Recovery (Mean ± SD)95.3 ± 8.9 %97.7 ± 5.2 %
Analytical Sensitivity≤ 2.0 pg/mL I PTH6.0 pg/mL I PTH
SpecificityNo detection of human PTH fragments53-84, 44-68 and 39-84No detection of human PTHfragments 53-84, 44-68 and 39-84
InterferenceBilirubin (20 mg/dL): < 10 %ND
Triglycerides (1500 mg/dL): < 10 %ND
Total Protein (3 mg/dL): < 10 %ND
Hemoglobin (1000 mg/dL): < 10 %ND
Red Blood Cells (0.2 % v/v): < 10 %ND
LimitationsNo high dose effect up to 100K pg/mLNo high dose effect up to 300K pg/mL
On-board Stability14 days @ 31°CN/A

AxSYM® I PTH and the predicate device:

N/A = Not Applicable ND = Not Determined

Safety & Effectiveness: 4.5

  • As explained in Section 4.1 of this submission, the I PTH levels are used in 4.5.1 conjunction with calcium levels as a clinical diagnostic tool to support the diagnosis of a disorder, as explained in the Summary and Explanation of the Test in the beginning of the AxSYM® I PTH Package Insert in Section 6.0. The I PTH assay is not used as a stand-alone test. Clinical interpretation requires simultaneous measurement of serum calcium levels. The levels of I PTH and calcium are repeated over a period of time for a diagnosis and during any type of treatment for effectiveness.
  • 4.5.2 The safety of use for both I PTH assays is substantially equivalent, but the AxSYM® I PTH assay has less risk associated with it due to the absence of radioactive materials.
  • Therefore, we conclude the safety and effectiveness of both assays are 4.5.3 substantially equivalent and favor the AxSYM® I PTH assay.

{5}------------------------------------------------

Substantial Equivalence Conclusion 4.6

We conclude that the data and information supplied in this summary provide enough detail and support for our claim of substantial equivalence between the AxSYM® Intact PTH assay and the predicate device, DSL-8000 ACTIVE™ Intact PTH IRMA.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines forming its body and wings. The bird is positioned within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAY 3 1 2002

Mr. Carroll Potts, M.S. Manager of Regulatory Affairs Diagnostic Systems Laboratories, Inc. 445 Medical Center Blvd. Webster, TX 77598

K020945 Re: Trade/Device Name: AxSYM® Intact PTH Regulation Number: 21 CFR 862.1545; 21 CFR 862.1150; 21 CFR 862.1660 Regulation Name: Parathyroid hormone test system; Calibrator; Quality control material (assayed and unassayed) Regulatory Class: Class II Product Code: CEW; JIT; JIX Dated: March 22, 2002 Received: March 25, 2002

Dear Mr. Potts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

INDICATIONS FOR USE 1.0

510 (k) Number (if known): Kozo945

Device Name: AxSYM® Intact PTH

Indications for Use:

AxSYM® Intact PTH is a Microparticle Enzyme Immunoassay (MEIA) for the in vitro quantitative determination of intact human parathyroid hormone (I PTH) in vitro quantiturre determination oxSYM® system. The AxSYM® I PTH assay is intended for use as an aid in the differential diagnosis of hypercalcemia and hypocalcemia.

Jean Cooper

(Division Sign-OB)
Division of Clinical Laboratory Devices
510(k) Number K020945

(Please Do Not Write Below This Line. Continue On Another Page If Needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ن Prescription Use (Per 21 CFR 801.109) OR

Over-The Counter Use _________________________________________________________________________________________________________________________________________________________

§ 862.1545 Parathyroid hormone test system.

(a)
Identification. A parathyroid hormone test system is a device intended to measure the levels of parathyroid hormone in serum and plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.(b)
Classification. Class II.