K Number
K020797

Validate with FDA (Live)

Date Cleared
2002-05-03

(52 days)

Product Code
Regulation Number
868.6250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pulmo-Mist compressor is designed for use by adult or pediatric patients requiring aerosolized medications.
The Pulmo-Mist Nebulizer Compressor is intended to be used to supply a continuous positive pressure to a nebulizer device for the purpose of generating an aerosol medicine.

Device Description

The Pulmo-Mist compressor is an electrically powered piston type compressor that provides approximately 6 I/min of air flow against a back pressure of 10 psig. It is designed to supply air to a small volume nebulizer. The package optionally contains a legally marketed small volume disposable nebulizer kit, specifically the Westmed "VIXONE®", manufactured by Westmed, Inc. and private labeled for distribution by Nidek Medical Products, in the original manufacturer's unopened package, Each such package includes a mouthpiece, a nebulizer, a connector tube and flexible air tubing and labeling. The nebulization particle size varies slightly depending on the nebulizer used and on the viscosity of the material being nebulized. When used with the VIXONE, the MMAD is approximately 1 µm with a saline solution. Other manufacturers of small volume nebulizers report similar values for similar flow rates.
The Pulmo-Mist compressor is a portable unit. It weighs approximately 3.2 pounds and has dimensions of 7.5 in w x 4.5 in h x 7.5 in d.

AI/ML Overview

The provided text describes the 510(k) submission for the Nidek Medical Products, Inc. Pulmo-Mist Compressor (K020797). This device is a physical medical device (a compressor for nebulizers) and not an AI/ML powered device. Therefore, many of the requested categories related to AI/ML device studies (such as sample size for test/training sets, ground truth establishment, expert adjudication, MRMC studies, and standalone algorithm performance) are not applicable to this submission.

However, I can extract the acceptance criteria and the study performed related to its regulatory approval as a traditional medical device.

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
Electrical PerformanceHome Use Respiratory Devices (March 1992 Reviewer Guidance)Met all required performance criteria
Mechanical PerformanceHome Use Respiratory Devices (March 1992 Reviewer Guidance)Met all required performance criteria
Environmental PerformanceHome Use Respiratory Devices (March 1992 Reviewer Guidance)Met all required performance criteria
Safety StandardsUL544 (Medical and Dental Equipment)Designed according to UL standard
UL1431 (Personal Hygiene and Healthcare Appliances)Designed according to UL standard
FunctionalityFunctioned as intended (e.g., providing ~6 L/min airflow against 10 psig back pressure for nebulization)Functioned as intended
Technical EquivalenceComponent similarity to predicate (a.c. motor driven compressor, ABS thermoplastic enclosure, wiring, switch, interconnecting tubing)Technically equivalent to referenced legally marketed device
Electromagnetic CompatibilityNo microprocessors or other electronic components that emit or are susceptible to electromagnetic interferenceDevice contains no such components, implying inherent compatibility

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified as this is a physical device. The testing was likely conducted on a limited number of units of the Pulmo-Mist Compressor.
  • Data Provenance: The device was submitted by Nidek Medical Products, Inc. in Birmingham, AL, USA. The testing was conducted in the context of US regulatory approval (FDA 510(k) process). It's a prospective evaluation of the manufactured device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. This is a physical device for which performance is measured against engineering specifications and industry standards, not against expert interpretation of data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. Performance testing for this type of device does not involve adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is a physical device, not an AI/ML system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a physical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for this device's performance is established by adherence to engineering specifications, electrical and mechanical safety standards (UL544, UL1431), and industry performance guidelines (FDA Reviewer Guidance for Home Use Respiratory Devices). Its functionality is assessed based on its ability to generate the specified airflow and pressure.

8. The sample size for the training set

  • Not applicable. This is a physical device, not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

  • Not applicable.

Study Description:

The study proving the device meets the acceptance criteria was a performance testing study conducted by Nidek Medical Products, Inc. (or a designated testing facility) on the Pulmo-Mist Compressor.

  • Methodology: The device was assessed against specific benchmarks and standards.

    • Regulatory Guidance: The testing followed the Anesthesiology and Respiratory Devices Branch's "Reviewer Guidance for Home Use Respiratory Devices" (March 1992) and "Reviewer Guidance for Premarket Notification Submissions" (November 1993).
    • Safety Standards: Compliance with Underwriter's Laboratory (UL) standards UL544 (Medical and Dental Equipment) and UL 1431 (Personal Hygiene and Healthcare Appliances) was also a critical component.
    • Technical Equivalence: The submission also highlighted the technical equivalence of the Pulmo-Mist compressor to the legally marketed predicate device (Invacare Model IRC 1001 Env) by identifying common components and design principles (same motor supplier, similar enclosure, wiring, etc.).
    • Functionality: Performance aspects like airflow rate (approximately 6 L/min) and pressure (10 psig back pressure) were likely measured and verified. The nebulization particle size reference (approx. 1 µm MMAD with VIXONE and saline) is a characteristic of the nebulizer when used with the compressor, not directly the compressor itself, but implies the compressor provides adequate power for this effect.
  • Results: The "Performance Data" section explicitly states: "The Pulmo-Mist compressor met all the required performance criteria and functioned as intended." It also notes: "Summary results for the legally marketed device show equivalent performance," reinforcing the substantial equivalence claim.

In summary, for K020797, the "study" was a series of engineering, electrical, mechanical, and safety tests conducted in accordance with recognized standards and FDA guidance documents to demonstrate that the physical device functions as intended and is substantially equivalent to a predicate device.

{0}------------------------------------------------

K020797

1300/ HADS

MAY 0 3 2002

H. 510(k) Summary

Submitter:Nidek Medical Products, Inc.3949 Valley East Industrial DriveBirmingham, AL 35217Tel: 205-856-7200Fax: 205-856-0533
-------------------------------------------------------------------------------------------------------------------------------------------------

Contact Person: L. G. Suelter Tel: 205-856-7200 Ext 222 Email: Isuelter@nidekmedical.com

  • Date Submitted: 31 December 2001
    Trade Name of Device: Pulmo-Mist Compressor

Common Name: Nebulizer

Classification Name: Nebulizer (Direct Patient Interface) 21 CFR 868.5630

Equivalent Legally Marketed Device: Invacare Model IRC 1001 Env

Description: The Pulmo-Mist compressor is an electrically powered piston type compressor that provides approximately 6 I/min of air flow against a back pressure of 10 psig. It is designed to supply air to a small volume nebulizer. The package optionally contains a legally marketed small volume disposable nebulizer kit, specifically the Westmed "VIXONE®", manufactured by Westmed, Inc. and private labeled for distribution by Nidek Medical Products, in the original manufacturer's unopened package, Each such package includes a mouthpiece, a nebulizer, a connector tube and flexible air tubing and labeling. The nebulization particle size varies slightly depending on the nebulizer used and on the viscosity of the material being nebulized. When used with the VIXONE, the MMAD is approximately 1 µm with a saline solution. Other manufacturers of small volume nebulizers report similar values for similar flow rates.

The Pulmo-Mist compressor is a portable unit. It weighs approximately 3.2 pounds and has dimensions of 7.5 in w x 4.5 in h x 7.5 in d.

Image /page/0/Picture/12 description: The image shows the logo for NIDEK Medical. The word "NIDEK" is written in large, bold, black letters. Below the word "NIDEK" is the word "Medical" written in a smaller, cursive font. The word "Medical" is also black, but it is not as dark as the word "NIDEK".

{1}------------------------------------------------

Intended Use: The intended use of the Pulmo-Mist compressor is to supply compressed air to a nebulizer to produce aerosolized medications as prescribed by a physician. The Pulmo-Mist compressor is designed for use by adult or pediatric patients requiring aerosolized medications.

Technological

Characteristics: The Pulmo-Mist compressor is technically equivalent to the referenced legally marketed device. Both devices comprise three basic components: a) an a.c. motor driven compressor which is from the same component supplier and the same model compressor, b) an ABS thermoplastic enclosure, and c) wiring, switch and interconnecting tubing. Both devices supply air to small volume nebulizers. The device contains no microprocessors or other electronic components that emit or are susceptible to electromagnetic interference.

Performance Data:

The Pulmo-Mist compressor was tested in accordance with the electrical, mechanical and environmental performance requirements for home use respiratory devices described in the Anesthesiology and Respiratory Devices Branch's March 1992 document entitled "Reviewer Guidance for Home Use Respiratory Devices" and the November 1993 document entitled "Reviewer Guidance for Premarket Notification Submissions". The Pulmo-Mist compressor met all the required performance criteria and functioned as intended. Additionally, the Pulmo-Mist compressor was designed according to Underwriter's Laboratory (UL) UL544 Standard for Medical and Dental Equipment and UL 1431 Standard for Personal Hygiene and Healthcare Appliances. Summary results for the legally marketed device show equivalent performance.

Conclusion:

Based on information contained in this 510(k) submission. Nidek Medical Products, Inc. concludes that the Pulmo-Mist compressor is substantially equivalent to the referenced legally marketed device.

Image /page/1/Picture/7 description: The image shows the logo for NIDEK Medical. The logo is in black and white, with the word "NIDEK" in large, bold letters. Below the word "NIDEK" is the word "Medical" in a smaller, cursive font. The logo is simple and professional, and it is likely used to represent a medical company.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing right, with a swooping line above them.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 7 2002

Nidek Medical Products, Inc. c/o Pamela K. Gwynn Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle, NC 27709

Re: K020797

Pulmo-Mist Compressor Model #4323 Regulation Number: 868.6250, and 868.5630 Regulation Name: Portable Air Compressor, and Nebulizer Regulatory Class: II (two) Product Code: 73 BTI, and 73 CAF Dated: April 18, 2002 Received: April 22, 2002

Dear Ms. Gwynn:

This letter corrects our substantially equivalent letter of May 3, 2002 regarding the indications for use of your device. Our letter incorrectly limited your device to use in military environments.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your

{3}------------------------------------------------

Page 2 - Ms. Pamela K. Gwynn

device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Indications for Use Statement

510(k) Number K020747

Device Name:

Pulmo-Mist Nebulizer Compressor

Indications for Use:

The Pulmo-Mist Nebulizer Compressor is intended to be used to supply a continuous positive pressure to a nebulizer device for the purpose of generating an aerosol medicine.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

Division of Cardiovascular & Respiratory Devices
510(k) Number K020797

Image /page/4/Picture/10 description: The image contains the logo for NIDEK Medical. The logo is in black and white, with the word "NIDEK" in large, bold letters. Below the word "NIDEK" is the word "Medical" in a smaller, cursive font. The logo is simple and professional, and it is likely used to represent the company on its products and marketing materials.

§ 868.6250 Portable air compressor.

(a)
Identification. A portable air compressor is a device intended to provide compressed air for medical purposes, e.g., to drive ventilators and other respiratory devices.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9.