K Number
K020724

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-05-09

(64 days)

Product Code
Regulation Number
862.1475
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dade Behring Dimension® Automated LDL Cholesterol Flex® reagent cartridge method is an in vitro diagnostic test intended for the quantitative determination of lowdensity lipoprotein cholesterol in human serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and various liver and renal diseases.

Device Description

The Dade Behring Dimension® Automated LDL Cholesterol Flex® reagent cartridge method is an in vitro diagnostic test that consists of prepackaged reagents in a flexible plastic cartridge for use on the Dimension® clinical chemistry system. The Dimension® Automated LDL Cholesterol Flex® reagent cartridge assay is a homogeneous method for directly measuring low-density lipoprotein cholesterol in human serum or plasma without the need for off-line pretreatment or centrifugation steps. The method is in a two reagent format and depends on the properties of detergent 1 which solubilizes only non- LDL particles. The cholesterol released is consumed by cholesterol esterase and cholesterol oxidase in a non-color forming reaction. Detergent 2 solubilizes the remaining LDL particles. The soluble LDL-C is then oxidized by the action of cholesterol esterase and cholesterol oxidase forming cholestenone and hydrogen peroxide (H2O2). The enzymatic action of peroxidase on H2O2 produces color in the presence of disodium salt (DSBmT) and 4-aminoantipyrine (4-AA) that is measured as a bichromatic endpoint technique at 540 nm and 700 nm. The color development is proportional to the amount of LDL-C present in the sample.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Dade Behring Dimension® Automated LDL Cholesterol Flex® reagent cartridge method:

General Statement: The document provided is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study protocol and acceptance criteria in the manner one might find in an academic publication or a PMA submission. As such, the "acceptance criteria" are implicitly defined by the comparison to the predicate devices and the "study" is the comparative performance evaluation.


1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of a 510(k) submission, explicit, pre-defined "acceptance criteria" are not stated as clear thresholds (e.g., "slope must be between 0.9 and 1.1"). Instead, the performance is demonstrated by comparison to legally marketed predicate devices, with the underlying "acceptance" being FDA's determination of substantial equivalence based on the presented data.

Acceptance Criteria (Implicit)Reported Device Performance (vs. Predicate)
Substantial Equivalence in Design and Performance to Legally Marketed Predicate Devices. This is demonstrated by:- Intended Use: "in vitro use" (Matches predicate) - Sample Size: "3 uL" (Matches predicate) - Measurement: "Direct LDL determination; Bichromatic endpoint; 540 nm and 700nm" (Similar to predicate's "Direct LDL determination; Bichromatic endpoint; 546 nm and 660 nm") - Reagents: "Two-detergent Genzyme N-geneous™ LDL Cholesterol reagent set" (Matches predicate)
Strong Correlation and Agreement with Predicate Homogeneous LDL-C Assay.Comparison to Genzyme N-geneous™ LDL Cholesterol assay: - Slope: 0.95 - Intercept: 4.7 mg/dL [0.12 mmol/L] - Correlation Coefficient: 0.997
Strong Correlation and Agreement with Beta-Quantification Reference Method (CRMLN), considered the gold standard for LDL-C.Comparison to Beta-Quantification method (CRMLN): - Slope: 1.01 - Intercept: 3.3 mg/dL [0.08 mmol/L] - Correlation Coefficient: 0.982

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set:
    • Versus Genzyme N-geneous™ LDL Cholesterol assay: 122 samples (n=122)
    • Versus Beta-Quantification method (CRMLN): 49 samples (n=49)
  • Data Provenance: Not explicitly stated in the document. It's common in such submissions to use samples from a clinical laboratory setting, potentially encompassing a range of patient demographics, but specific details like country of origin or whether the data was retrospective or prospective are not provided. Given the context of a comparative performance study for a new analyzer, it is highly likely these were prospective samples run concurrently on both the new device and the predicate/reference method.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The concept of "experts establishing ground truth" in the context of this device (an LDL cholesterol assay) is different from image-based diagnostics.

  • For the Genzyme N-geneous™ LDL Cholesterol assay comparison: The "ground truth" was established by the predicate device itself. No human experts were involved in determining the "truth" for this comparison; it was an instrument-to-instrument comparison.
  • For the Beta-Quantification method (CRMLN) comparison: The "ground truth" was established by the Cholesterol Reference Method Laboratory Network (CRMLN) Beta-Quantification reference method. This method is considered a highly accurate and standardized laboratory procedure for LDL-C determination. While this method is performed by highly qualified laboratory personnel, the document does not specify the "number of experts" or their individual qualifications. The "expertise" lies in the method's standardization and rigorous execution within a reference laboratory network.

4. Adjudication Method for the Test Set

Adjudication methods (like 2+1, 3+1) are typically used in studies where human readers interpret results (e.g., radiology images) and disagreement needs to be resolved. This is not applicable here as the "test set" involves quantitative measurements from an automated clinical chemistry system. The "adjudication" is inherent in the comparison against established analytical methods, particularly the CRMLN Beta-Quantification reference method, which is itself a highly controlled and validated process.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.
MRMC studies are relevant for diagnostic devices that involve human interpretation of outputs (e.g., radiologists reading images with or without AI assistance). This device is an automated in vitro diagnostic test that produces a quantitative numerical result for LDL-C, requiring no human interpretation for its primary output. Therefore, the concept of "how much human readers improve with AI vs without AI assistance" does not apply.


6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, a standalone performance study was done. The entire premise of the submission is to demonstrate the performance of the device algorithm (the automated Flex® reagent cartridge method) by itself, producing an LDL-C value. The comparative studies describe its performance without human intervention in the analyte measurement process. The results (slope, intercept, correlation coefficient) directly reflect the algorithm's standalone analytical performance against other established methods.


7. The Type of Ground Truth Used

  • For the comparison to the Genzyme N-geneous™ LDL Cholesterol assay: The ground truth was effectively the measurements from the predicate device. This is a common practice in 510(k) submissions to show equivalence.
  • For the comparison to the Beta-Quantification method: The ground truth was the CRMLN Beta-Quantification reference method, which is considered a highly accurate and standardized analytical reference method. This is a gold standard for LDL-C measurement.

8. The Sample Size for the Training Set

The document does not provide information on a training set sample size. This is typical for a 510(k) summary for an in vitro diagnostic device, especially one involving chemical reagents and an automated system. Such devices are developed and optimized through internal R&D processes, but a distinct "training set" (as understood in machine learning) is not typically detailed or reported in regulatory submissions unless the device heavily relies on a data-driven model or AI component. The methods described are chemical reactions, not a machine learning algorithm that needs explicit training data.


9. How the Ground Truth for the Training Set Was Established

As no training set is explicitly mentioned or detailed in the document, the method for establishing its ground truth is not provided. The development of the assay itself would have involved extensive analytical chemistry, experimentation, and optimization using various known samples and reference materials, but these would be part of the product development process rather than a formalized "training set" as defined in studies for data-driven algorithms.

{0}------------------------------------------------

K020724

Summary of Safety and Effectiveness Information

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter's Name:Richard M. VaughtDade Behring Inc.P.O. Box 6101Newark, DE 19714-6101
Date of Preparation:March 5, 2002
Name of Product:Dade Behring Dimension® Automated LDL Cholesterol

FDA Classification Name: Lipoprotein test system

Predicate Device: The Genzyme N-geneous™ LDL-Cholesterol assay (K971573) and the Beta-Quantification reference method utilized by the Cholesterol Reference Method Laboratory Network (CRMLN).

Flex® reagent cartridge method

Device Description: The Dade Behring Dimension® Automated LDL Cholesterol Flex® reagent cartridge method is an in vitro diagnostic test that consists of prepackaged reagents in a flexible plastic cartridge for use on the Dimension® clinical chemistry system. The Dimension® Automated LDL Cholesterol Flex® reagent cartridge assay is a homogeneous method for directly measuring low-density lipoprotein cholesterol in human serum or plasma without the need for off-line pretreatment or centrifugation steps.

The method is in a two reagent format and depends on the properties of detergent 1 which solubilizes only non- LDL particles. The cholesterol released is consumed by cholesterol esterase and cholesterol oxidase in a non-color forming reaction. Detergent 2 solubilizes the remaining LDL particles. The soluble LDL-C is then oxidized by the action of cholesterol esterase and cholesterol oxidase forming cholestenone and hydrogen peroxide

{1}------------------------------------------------

(H2O2). The enzymatic action of peroxidase on H2O2 produces color in the presence of (11507). The Chilymatio action of Pe, disodium salt (DSBmT) and 4-aminoantipyrine (4-N,N-015(++Suffooury) in continer, around (540 nm, 700 nm) endpoint technique. The AA) that is measured as as a cremoral to the amount of LDL-C present in the sample.

Image /page/1/Figure/1 description: This image shows a series of chemical reactions. In the first reaction, nonsoluble LDL-C, VLDL-C, HDL-C, and Chylomicrons are converted to soluble Non-LDL-C, (VLDL-C, HDL-C, Chylomicrons) using Detergent 1. Soluble Non-LDL-C is converted to a non-color forming product using Cholesterol esterase and Cholesterol oxidase. In another reaction, nonsoluble LDL-C is converted to soluble LDL-C using Detergent 2, and soluble LDL-C + O2 is converted to Cholestenone + H2O2 using Cholesterol esterase and Cholesterol oxidase. Finally, H2O2 + DSBmT + 4-AA is converted to a color development product using Peroxidase.

Intended Use: The Dimension® Automated LDL Cholesterol Flex® reagent cartridge method is an in vitro diagnostic test intended for the quantitative determination of lowdensity lipoprotein cholesterol in human serum and plasma.

Comparison to Predicate Device: A summary of features for the Dimension® Automated LDL Cholesterol Flex® reagent cartridge method and the predicate device, the Genzyme N-geneous™ LDL Cholesterol reagent set (K971573) as utilized on automated analyzers are provided in the following chart:

Dade Behring Dimension® AutomatedLDL Cholesterol Flex® reagent methodGenzyme N-geneous™ LDL Cholesterolassay (for use on automated analyzers)
Intended Usein vitro usein vitro use
Sample size3 uL3 uL
MeasurementDirect LDL determination;Bichromatic endpoint;540 nm and 700nmDirect LDL determination;Bichromatic endpoint;546 nm and 660 nm
ReagentsTwo-detergent Genzyme N-geneous™LDL Cholesterol reagent setTwo-detergent Genzyme N-geneous™LDL Cholesterol reagent set

{2}------------------------------------------------

Additionally, comparative performance studies were conducted between the Dimension® Automated LDL Cholesterol Flex® reagent cartridge method and both the Genzyme Ngeneous™ LDL Cholesterol assay (on the Beckman CX-9 automated analyzer) and the CRMLN Beta-Quantification reference method. The results are summarized below:

ComparativeMethodSlopeIntercept(mg/dL) [mmol/L]CorrelationCoefficientn
Genzyme N-geneous™ LDLCholesterol assay0.954.7 [0.12]0.997122
Beta-Quantification method1.013.3 [0.08]0.98249

Comments on Substantial Equivalence: The Dimension® Automated LDL Cholesterol Flex® reagent cartridge method has the same intended use, employs the same design, and utililzes the same reagent set as the predicate device, the Genzyme N-geneous™ LDL Cholesterol reagent set (K971573).

Conclusion: The Dimension® Automated LDL Cholesterol Flex® reagent cartridge method is substantially equivalent in design and performance to both the Genzyme Ngeneous LDL Cholesterol assay and the CRMLN Beta-Quantification reference method based on the comparison studies as described above.

Richard M. Hodes

Richard M. Vaught Regulatory Affairs and Compliance Manager March 5, 2002

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" surrounding it.

MAY 0 9 2002

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Richard M. Vaught Regulatory Affairs and Compliance Manager Dade Behring Inc. Chemistry/Immunochemistry Glasgow Business Community Bldg. 500, M. S. 514 P.O. Box 6101 Newark, DE 19714-6101

K020724 Re:

Trade/Device Name: Dimension® Automated LDL Cholesterol Flex® Reagent Cartridge Method Regulation Number: 21 CFR 862.1475 Regulation Name: Lipoprotein test system Regulatory Class: Class I, reserved Product Code: MRR Dated: March 5, 2002 Received: March 6, 2002

Dear Mr. Vaught:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket I his leation. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 II you dostre specific an in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (1 sase contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications For Use Statement

K020724

Device Name:

Dimension® Automated LDL Cholesterol Flex® reagent cartridge method

Indications for Use:

The Dade Behring Dimension® Automated LDL Cholesterol Flex® reagent The Dado Don'ting Diniensitro diagnostic test intended for the quantitative cartings method is an in This Catagratein cholesterol in human serum and plasma. determination of 10% cents are used in the diagnosis and treatment of lipid Lipoprotem measurements are uses a and various liver and various liver and renal diseases.

Rem Vaught
Richard M. Vaught

Regulatory Affairs and Compliance Manager

March 5, 2002

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Cooper
(Division Sign-Off)
Division of Ciucal Laboratory Devices
510(k) Numbe: K020774

Prescription Use (Per 21 CFR 801.109)

OR

Over-the-counter Use

(Optional format 1-2-96)

§ 862.1475 Lipoprotein test system.

(a)
Identification. A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.