(119 days)
PMA890023/S4,S6,S7
No
The summary describes a standard contact lens with UV blocking properties and provides clinical study data related to its safety and efficacy for vision correction. There is no mention of AI, ML, image processing, or any computational analysis of data that would suggest the use of these technologies. The performance studies focus on clinical outcomes like visual acuity, wearing time, and lens deposits, not on the performance of an AI/ML algorithm.
No.
The device is indicated for the correction of visual acuity in persons with non-diseased eyes, not for treating any disease or condition.
No
Explanation: The device is a contact lens indicated for the correction of visual acuity; it does not diagnose medical conditions.
No
The device description clearly states it is a physical contact lens made of ocufilcon B material, with specific physical dimensions and properties. It is not a software-only device.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The provided text describes contact lenses. Contact lenses are medical devices used to correct vision by being placed directly on the surface of the eye. They do not analyze samples taken from the body.
- Intended Use: The intended use clearly states the lenses are for "correction of visual acuity" in persons with non-diseased eyes. This is a therapeutic/corrective function, not a diagnostic one.
- Device Description: The description focuses on the physical properties and manufacturing of the contact lenses.
- Performance Studies: The performance studies evaluate the safety and efficacy of the lenses in terms of visual acuity, wearing time, and lens deposits, all related to their function as a vision correction device.
Therefore, based on the provided information, the BIOMEDICS® 52 1-Day contact lenses are not an In Vitro Diagnostic device. They are a medical device used for vision correction.
N/A
Intended Use / Indications for Use
Spherical:
BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or lyperopic (farsighted) and may exhibit refractive astigmatism of 2.00 diopters that does not interfere with visual acuity.
Toric:
BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic)UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of up to 10.00 diopters.
The lens may be prescribed for Daily Wear in not- aphakic persons. The eyecare practitioner may prescribe the contact lens for wither single use disposable wear.
The BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
Product codes (comma separated list FDA assigned to the subject device)
LPL; MVN
Device Description
The BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses are available with ultraviolet absorbing additive (benzophenone based):
- in the power range of -10.00 to +6.00 diopters for sphere
- in the cylinder power range pl to -6.00D cylinder .
- with center thickness from 0.025mm to 0.27mm
- with base curves of 8.00mm to 9.20mm .
- . with diameter of 12.00mm to 18.00mm.
This lens material, design, cast molding manufacturing and sterilization process is equivalent to BIOMEDICS® 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses described in submission 52 PMA890023/S4,S6 and S7, 510(K) K003136.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Eyes (Cornea)
Indicated Patient Age Range
Not Found
Intended User / Care Setting
eyecare practitioner
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Clinical studies were performed to establish the safety and efficacy of the Biomedics® 52 1-Dav lenses with an in-process manufacturing change. The purpose of the clinical investigation was to verify that the change made in the manufacturing process to an unextracted process, substantiated by safety toxicity testing data results, do not raise additional questions of safety or effectiveness.
The study was designed as an open-label, prospective, concurrent cohort control, randomized clinical trial. Thirty-six (36) subjects (72 eyes) were enrolled into the study at three (3) investigational sites. All of the 36 subjects were determined to be eligible to enter the study and were dispensed either the Test or the Control lenses based upon a randomization schedule provided to the site. Scheduled follow-up visits were planned for the 1-Week, 2-Week and 4-Week visit intervals. No subjects were discontinued prior to completion. No adverse events were reported for either the Test or Control cohort through the study period.
Safety Results:
Safety is demonstrated if the test lens complications are substantially equivalent in frequency and severity to the control lens complications.
- Permanent decreases in BCVA: There were no reports of 2-line decreases in best corrected Snellen visual acuity for any Test or Control cohort eyes.
- Persistent corneal staining of Grade 3 or 4 at two or more follow-up visits: No reports of Grade 3 or 4 staining for either Test or Control cohort eyes.
- Neovascularization more than 1.5mm and/or more than 1.0mm in 3 or more quadrants: No neovascularization reported of greater than 1.5mm in any quadrant. Incidence remained stable.
- Persistent hyperemia of Grade 3 or 4 at two or more follow-up visits: No Grade 3 or 4 limbal or bulbar hyperemia reported.
- Peripheral or central ulcerative keratitis: No peripheral or central ulcerative keratitis reported.
- Infiltrates of Grade 2 or worse: No infiltrates reported.
- Subjective symptoms, problems or complaints: Discomfort was reported more frequently and with greater severity for the Test cohort eyes (12.5% of follow-up examinations, average severity 2.1 out of 4.0) compared to Control (7.3% of follow-up examinations, average severity 1.1 out of 4.0). Blurred vision was reported more frequently for the Control cohort (6.3%, average severity 1.7 out of 4.0) compared to Test (1.6%, average severity 1.0 out of 4.0). The findings for symptoms, problems and complaints do not indicate any safety issues with the Test lenses when compared to the Control lenses, and discomfort was not related to objective findings.
Both the Test and the Control cohort eyes had similar outcomes in the evaluation of the safety variables. Very few complications were reported during the study and most were mild and transient. The Test lens demonstrates equivalent safety when compared to the Control lens.
Visual Acuity Results:
The comparison of the beginning and ending lens visual acuties within and between the Test and the Control cohort eyes provides evidence that the Test lens is safe and effective with respect to lens visual acuity. Both the Test and Control lens visual acuities are essentially the same at the final examination.
- Proportion of Lens Visual Acuities 20/30 or Better at Final Visit: 100.0% for Completed Test Eyes, 100.0% for Completed Control Eyes.
- Proportion of Lens Visual Acuities 20/20 or Better at Final Visit: 84.8% for Completed Test Eyes, 83.3% for Completed Control Eyes.
Average Wearing Time:
Test cohort average wearing time was 12.8-13.2 hours per day. Control cohort average wearing time was 13.1-13.3 hours per day. Both cohorts achieved acceptable and similar average daily wearing times.
Lens Deposits over course of the study:
Test lenses had a rate of 12.5% with deposits, Control lenses had a rate of 11.1%. Test eyes reported 1.4% medium deposits, Control eyes only light deposits.
Efficacy Summary:
- All subjects in both Test and Control groups successfully completed the 1-month study period.
- The percentage of Completed eyes reporting visual acuities of 20/20 or better at the final study visit was 84.8% for the Test cohort eyes and 83.3% for the Control eyes. The percentage of Completed eyes reporting visual acuities of 20/30 or better at the final study visit was 100% for both Test and Control cohort eyes.
- Overall, the visual performance of the Test lens was similar to the performance of the Control lens.
Conclusions:
The results of the slit lamp examinations indicate that the Test lens performs at least as well as the Control lens in terms of safety. Visual acuities and ocular health were all maintained during the study. Except for discomfort, most of the symptoms, problems and complaints were reported at similar rates between the Test and the Control cohorts. The completion of all of the subjects recruited for the study and the excellent visual acuity results demonstrate the efficacy of the Test lens during the study. The findings of this study demonstrate that the Test lens performs equivalently in terms of safety and efficacy when compared to the Control lens cohort.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
PMA890023/S4,S6,S7
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 886.5925 Soft (hydrophilic) contact lens.
(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.
0
JUN 0 4 2002
K 0 20389
: SUMMARY OF SAFETY AND SUBSTANTIAL EQUIVALENCE
1. Applicant's Name and Address
Ocular Sciences Inc. 1855 Gateway Blvd. Suite 700 Concord, CA 94520 USA
Contact Person
Richard E. Lippman, OD FAAO Senior Consultant R.P. Chiacchierini & Associates, LLC. 17003 Horn Point Drive Gaithersburg, Maryland 20878 Telephone: (301) 330-4660 Fax: (310) 593-1759
Identification of device 2.
Soft Contact Lens Common Name: BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Trade Name: Contact lenses Classification: Daily Wear Soft (hydrophilic) Contact Lens Device classification: Class II (21 CFR 886.5925 (b) (1))
Description of device 3.
The BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses are available with ultraviolet absorbing additive (benzophenone based):
- in the power range of -10.00 to +6.00 diopters for sphere ●
- in the cylinder power range pl to -6.00D cylinder .
- with center thickness from 0.025mm to 0.27mm ●
- with base curves of 8.00mm to 9.20mm .
- . with diameter of 12.00mm to 18.00mm.
This lens material, design, cast molding manufacturing and sterilization process is equivalent to BIOMEDICS® 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses described in submission 52 PMA890023/S4,S6 and S7, 510(K) K003136.
4. Intended use
Spherical:
BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or lyperopic (farsighted) and may exhibit refractive astigmatism of 2.00 diopters that does not interfere with visual acuity.
1
510(K) PREMARKET NOTIFICATION
SUMMARY OF SAFETY AND SUBSTANTIAL EQUIVALENCE
Reference OSL52 : 3 Section Version : 1 Page :218
Toric:
BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic)UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of up to 10.00 diopters.
The lens may be prescribed for Daily Wear in not- aphakic persons. The eyecare practitioner may prescribe the contact lens for wither single use disposable wear.
The BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
5. Predicate devices
The predicate lenses were selected to address: material (FDA Group IV: high water, ionic polymer), intended use (daily wear) and lens designs (sphere, toric).
Lens material, spherical and toric lens design and intended use:
BIOMEDICS® 52 1-Day (ocufilcon B) Soft (hydrophilic) UV Blocking Contact lenses, FDA Group IV, high water content, ionic soft contact lenses for daily wear marketed internationally by OCULAR SCIENCES Inc. under K003136.
6. Characteristics
The characteristics of the BIOMEDICS® 52 1-Day (ocufilcon B) Soft (hydrophilic) UV Blocking Contact lenses are compared to the characteristics of the predicate device BIOMEDICS® 52 in the following table.
Material and Parameter comparison | ||
---|---|---|
Predicate device | ||
BIOMEDICS® 52 | Subject device | |
BIOMEDICS® 52 1-Day | ||
PRODUCTION METHOD | Cast molded process | Cast molded process |
INTENDED USE | Daily wear | |
Correction of ametropia | Daily wear | |
Correction of ametropia | ||
MATERIAL | ocufilcon B | ocufilcon B |
Type | Group IV | Group IV |
Color additive | Vat Blue 6 Dye | |
21 CFR 73.3119 | ||
CAS #130-20-1 | Vat Blue 6 Dye | |
21 CFR 73.3119 | ||
CAS #130-20-1 | ||
UV additive | Yes | Yes |
Characteristics comparison | Labeled | Labeled |
Water Content % @ 20°C | 52 | 52 |
Refractive Index @ 20°C | 1.41 | 1.41 |
Dk permeability, ISO 9913-1 | ||
Polarimetric method with edge | ||
correction @ 35°C | ||
x 10-11 (cm²/sec) (ml O₂/ml x mm Hg) | 16.8 | 16.8 |
Light transmittance | Tv, % @ 20°C | |
Illuminant A with a 2º observer (between | ||
380 and 780 nm) | ||
97.7% average | Tv, % @ 20°C | |
Illuminant A with a 2º observer (between 380 and | ||
780 nm) | ||
97.7% average |
TABLE 1
2
510(K) PREMARKET NOTIFICATION
SUMMARY OF SAFETY AND SUBSTANTIAL EQUIVALENCE
Reference OSL52 Section : 3 Version : 1 Page :318
| Characteristics comparison, -300 D
40 lenses | Labeled | Labeled |
---|---|---|
Base Curve, mm | 8.70 | 8.70 |
Diameter, mm | 14.20 | 14.20 |
Power, D | -3.00 | -3.00 |
7. Non clinical studies
Ocular Sciences (ocufilcon B) soft (hydrophilic)
contact lens
Non-clinical studies are summarized below:
Chemistry and leachability
- Material property data were generated on the BIOMEDICS® 52 and the BIOMEDICS® 52 1-Day lenses ● processed with the manufacturing process changes. The material properties were substantially equivalent.
- The lens care product manufacturers have previously shown compatibility of Group IV lenses with their . products.
- The shelf life stability for BIOMEDICS® 52 1-Day lenses is based upon stability protocols included with this . notification.
- Studies were conducted to determine the residual monomers on the subject device. .
Toxicology, lenses materials
In accordance with the May 1994 Guidance Document for Daily Wear contact lenses, toxicology studies have been conducted on the BIOMEDICS® 52 1-Day containing UV blocker. The results are summarized below:
Cytotoxicity Test: .
Cytotoxicity Tests have been conducted on the subject device according to ISO 10993: Biological Evaluation of Medical Devices, Part 5: Tests for Cytotoxicity: In vitro Methods guidelines, was conducted on the test articles, to determine the potential for Cytotoxicity.
The negative controls and the positive controls performed as anticipated. Under the conditions of the study, the test articles were not cytotoxic.
- Acute Systemic Injection Test in the mouse: ●
An evaluation of the test articles for systemic toxicity in mice after a single intravenous administration or a single intraperitoneal administration has been conding to the ISO 10993; Biological Evaluation of Medical Devices, Part 11: Tests for Systemic Toxicity.
No evidence of systematic toxicity was observed from the test article extracts. Each test article met the test requirements.
- . Ocular Eye Irritation Test in the rabbit:
An evaluation of the ocular irritation of 0.9% NaCl of the subject article after a single instillation in the rabbit has been conducted according to ISO 10993: Biological Evaluation of Medical devices, Part 10: Tests for the Irritation and Sensitization.
No evidence of ocular irritation was observed in the rabbits. The test are not considered irritants to the ocular tissue of rabbits.
. 22-Day Ocular Eye Irritation Study in the rabbit
For the evaluation of the safety of the subject device with regard to ocular irritation was conducted in rabbits in accordance with ISO 9394:1997(E). Lens wear was a minimum of 7 hours for 21 consecutive days. On lens wear day 22, lenses were worn for a least 4 hours.
3
510(K) PREMARKET NOTIFICATION
SUMMARY OF SAFETY AND SUBSTANTIAL EQUIVALENCE
Laboratory observations during the study included daily health observations, macroscopic eye examinations (during the last hour of lens wear), weekly biomicroscopic slit-lamp examinations and body weight observations.
Under the conditions of this study, eyes treated with test lenses were similar to the untreated control eyes.
Solution Compatibility
Microbiology: The lens care product manufacturers have established a reasonable assurance of disinfection efficacy of their care products with lens groups for which they are approved.
Non-clinical studies and manufacturing information provided by reference PMA890023 S4/S7 and K003136
- Cast molded manufacturing process .
- Toxicology packaging materials .
- Microbiology and sterilization ●
8. Clinical data
Clinical studies were performed to establish the safety and efficacy of the Biomedics® 52 1-Dav lenses with an in-process manufacturing change. The purpose of the clinical investigation was to verify that the change made in the manufacturing process to an unextracted process, substantiated by safety toxicity testing data results, do not raise additional questions of safety or effectiveness.
The study was designed as an open-label, prospective, concurrent cohort control, randomized clinical trial. Subjects were recruited based upon specified inclusion criteria to assure that the appropriate subject population samples were included and inappropriate subjects were excluded. Scheduled follow-up visits were planned for the 1-Week, 2-Week and 4-Week visit intervals based upon the date the subject was dispensed the Test or Control lenses. No subjects were discontinued prior to completion in this investigation. No adverse events were reported for either the Test or Control cohort through the study period.
Thirty-six (36) subjects (72 eyes) were enrolled into the study at three (3) investigational sites. All of the 36 subjects were determined to be eligible to enter the study and were dispensed either the Test or the Control lenses based upon a randomization schedule provided to the site.
The distribution of age and gender of the subjects enrolled into the study by cohort is shown below.
Distribution of Age and Gender | |||
---|---|---|---|
Test | Control | Totals | |
Female Subjects | 19 | 10 | 29 |
Male Subjects | 5 | 2 | 7 |
Ratio F/M | 3.8:1 | 5:1 | 4.1:1 |
Average Age (years) | 36.0 | 38.8 | 36.9 |
Safety Results:
Safety is demonstrated if the test lens complications are substantially equivalent in frequency and severity to the control lens complications.
-
- Permanent decreases in BCVA: There were no reports of 2-line decreases in best corrected Snellen visual acuity for any of the Test or Control cohort eyes.
4
510(K) PREMARKET NOTIFICATION
SUMMARY OF SAFETY AND SUBSTANTIAL EQUIVALENCE
OSL52 Reference Section : 3 Version : I : 5 / 8 Page
- Persistent corneal staining of Grade 3 or 4 at two or more follow-up visits: There were no reports 2. of Grade 3 or 4 staining for either the Test or the Control cohort eyes.
- Neovascularization more than 1.5mm and/or more than 1.0mm in 3 or more quadrants: There was 3. no neovascularization reported of greater than 1.5mm in any quadrant. The incidence of neovascularization reported for the Completed Test and Control cohort eyes remained stable throughout the period of the study.
- Persistent hyperemia of Grade 3 or 4 at two or more follow-up visits: No Grade 3 or 4 limbal or 4. bulbar hyperemia was reported during the period of the study for any Test or Control cohort eyes.
- Peripheral or central ulcerative keratitis: No peripheral or central ulcerative keratitis was reported 5. during the period of the study for any Test or Control cohort eyes.
- Infiltrates of Grade 2 or worse: No infiltrates were reported during the period of the study for any 6. Test or Control cohort eyes.
-
- Subrective symptoms, problems or complaints: Subjective symptoms were reported for both the Test and the Control eves over the course of the study. The areas where the Test and Control cohorts presented with differences were in discomfort and blurred vision.
Discomfort was reported more frequently and with greater severity for the Test cohort eyes with 12.5% of the follow-up examinations for Test eyes resulting in a report of discomfort with an average severity of 2.1 out of 4.0. This is in contrast to the Control cohort eyes where discomfort was reported for 7.3% of the follow-up examinations with an average severity of 1.1 out of 4 0.
Blurred vision was reported more frequently for the Control cohort eves with 6.3% of the follow-up examinations for Control eves resulting in a report of blurred vision with an average severity of 1.7 out of 4.0. This is in contrast to the Test cohort eyes where blurred vision was reported for 1.6% of the follow-up examinations with an average severity of 1.0 out of 4.0.
The findings for symptoms, problems and complaints do not indicate any safety issues with the Test lenses when compared to the Control lenses. The symptom of discomfort was not related to objective (slit lamp) findings for the Test cohort eyes and does not by itself indicate any safety concerns.
Both the Test and the Control cohort eyes had similar outcomes in the evaluation of the safety variables. Very few complications were reported during the study and most were mild and transient. Based upon the safety criteria established in the study protocol and reviewed above, the Test lens demonstrates equivalent safety when compared to the Control lens.
Visual Acuity Results:
The comparison of the beginning and ending lens visual acuties within and between the Test and the Control cohort eyes provides evidence that the Test lens is safe and effective with respect to lens visual acuity. Both the Test and Control lens visual acuities are essentially the same at the final examination.
5
510(K) PREMARKET NOTIFICATION
SUMMARY OF SAFETY AND SUBSTANTIAL EQUIVALENCE
| | Proportion of Lens Visual Acuities
Dispensing Versus Final Examination | |
|----------------------------------------------------|---------------------------------------------------------------------------|-------------|
| Visit Sample Size | Dispensing | Final Visit |
| Completed Test Eyes | 48 | 46 |
| Completed Control Eyes | 24 | 24 |
| Proportion of Lens Visual Acuities 20/30 or Better | | |
| Eyes with VA of 20/30 or better | Dispensing | Final Visit |
| Completed Test Eyes | 100.0% | 100.0% |
| Completed Control Eyes | 95.8% | 100.0% |
| Proportion of Lens Visual Acuities 20/20 or Better | | |
| Eyes with VA of 20/20 or better | Dispensing | Final Visit |
| Completed Test Eyes | 83.3% | 84.8% |
| Completed Control Eyes | 54.2% | 83.3% |
Average Wearing Time:
Based upon the average lens wearing times reported by the Test and Control subjects, both cohorts were achieving acceptable and similar average daily wearing times. For the Test cohort the was 12.8 hours per day to 13.2 hours per day over the period of the study. For the Control cohort the average wearing time was 13.1 hours per day to 13.3 hours per day over the period of the study.
Lens Deposits over course of the study:
Over the period of the study the Test and the Control cohort lenses were reported to have similar rates, Test 12.5% and Control 11.1%, with the Test eves reporting 1.4% of the lenses as having medium deposits and the Control eyes only reporting light deposits.
Efficacy Summary:
-
- The proportion of subjects in each group (Test or Control) successfully completing the protocol specified study duration: All of the Test and Control cohort subjects completed the 1 month study period. No subjects were discontinued for any reason.
-
- The percentage of lens visual acuties of 20/20 and the percentage of lens visual acuittes of 20/30 or Better: The percentage of Completed eyes reporting visual acuities of 20/20 or better at the final study visit was 84.8% for the Test cohort eyes and 83.3% for the Control eves. Both of these proportions are greater than those measured at the study. The percentage of Completed eves reporting visual acuities of 20/30 or better at the final study visit was 100% for the Test cohort eves and 100% for the Control cohort eyes. The percentage of Test eyes reporting VAs of 20/30 at the final visit met the expectations set at the inception of the study.
-
- Overall, the visual performance of the Test lens was similar to the performance of the Control lens.
Conclusions:
These conclusion are based upon the review of the data reported during the 1 month evaluation of the safety and efficacy of the Biomedics® 52 1-Day (ocufilcon B) soft (hydrophilic) contact lenses manufactured under a process change used following a daily disposable contact lens regimen.
The results of the slit lamp examinations indicate that the Test lens performs at least as well as the Control lens in terms of safety. Visual acuities and ocular health were all maintained during the study. Except for
6
510(K) PREMARKET NOTIFICATION
SUMMARY OF SAFETY AND SUBSTANTIAL EQUIVALENCE
Reference OSL52 Section : 3 Version : I Page : 7 / 8
discomfort, most of the symptoms, problems and complaints were reported at similar rates between the Test and the Control cohorts. The completion of all of the subjects recruited for the study and the excellent visual acuity results demonstrate the efficacy of the Test lens during the study.
The difference in the reports and severity of the symptom discomfort also suggest that the Test lens is associated with excessive discomfort for some of the Test subjects. The findings of this study demonstrate that the Test lens performs equivalently in terms of safety and efficacy when compared to the Control lens cohort.
9. Conclusions drawn from studies
Validity of Scientific Data:
A contract laboratory using Good Laboratory Practices conducted the Toxicology studies. Chemistry leachables studies were conducted by Ocular Sciences, UK Ltd.
Substantial Equivalence:
Information provided in this 510(k) establishes that the Biomedics® 52 1-Day lenses are equivalent in optical, chemical and physical properties of the predicate devices any questions of safety and effectiveness. Therefore, the device is substantially equivalent to the predicate devices® 52 1-Day (ocufilcon B) soft (Hydrophilic) UV blocking contact lenses under K003136.
Risk and Benefits:
The risks of the subject device are the same as those normally attributed to the wearing of soft (hydrophilic) contact lenses on a daily wear base. The patient are the same as those for other soft (hydrophilic) contact lenses.
10. Route chosen in the Flow Chart for 510 (k) Daily Wear Contact Lens Materials Submission
FIGURE 1 [NEXT PAGE]
BIOMEDICS® 52 1-Dav
7
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/7/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird-like shape.
Food and Drug Administration 9200 Corporate Boulevard . Rockville MD 20850 JUN 0 4 2002
Ocular Sciences, Inc. c/o Richard E. Lippman, O.D., F.A.A.O. Senior Consultant Official Correspondent to Ocular Sciences, Inc. R.P. Chiacchierini & Associates, LLC 17003 Horn Point Drive Gaithersburg, MD 20878
Re: K020389
Trade/Device Name: BIOMEDICS® 52 1-Day (ocufilcon B) Soft (hydrophilic) UV Blocking Contact Lens for Daily Wear (Spherical and Toric)
Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL; MVN Dated: May 2, 2002 Received: May 3, 2002
Dear Dr. Lippman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
8
Page 2 - Richard E. Lippman, O:D., F.A.A.O.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address . http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
9
INDICATION FOR USE STATEMENT
SECTION 2 : INDICATIONS FOR USE STATEMENT
510(k) Number (if known)
Device Name:
Indications for Use:
Spherical:
BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of 2.00 diopters that does not interfere with visual acuity.
Toric:
BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic)UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of 10.00 diopters.
The lens may be prescribed for Daily Wear in not- aphakic persons. The eyecare practitioner may prescribe the contact lens for single use disposable wear.
The BIOMEDICS® 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
CDRII Office of Device Evoluction (QDF)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use | |
---|---|
------------------ | --------------- |
OR
Over-the-counter-use | |
---|---|
---------------------- | --------------- |
(Division Sign-off) | |
---|---|
Division of Ophthalmic Devices |
510(k) Number | K020389 |
---|---|
--------------- | --------- |