K Number
K020150
Device Name
CRC-15BT
Manufacturer
Date Cleared
2002-08-02

(198 days)

Product Code
Regulation Number
892.1360
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The CRC 15BT dose calibrator is intended to be used by trained medical physicists, or qualified technologists to measure the activity or output of radioactive brachytherapy sources, including HDR, LDR and IVBT sources.
Device Description
The CRC 15BT dose calibrator is a well type ionization chamber specifically designed to accurately and rapidly measure all types of brachytherapy sources with appropriate calibration and source positioning device. The ionization chamber is connected to a readout, which converts measured current into a displayed value in units of either Ci (activity) or U (air kerma strength). The unit is sealed and pressurized with UPC Argon. which improves sensitivity and eliminates the need for temperature and pressure corrections.
More Information

90 KPT

CRC 15R

No
The document explicitly states "Mentions AI, DNN, or ML: Not Found" and the device description focuses on standard ionization chamber technology.

No
The device measures the activity of radioactive sources used in brachytherapy; it does not directly treat a patient or perform a therapeutic function itself.

No

The device measures the activity or output of radioactive brachytherapy sources. This is a measurement of the source itself, not of a patient's physiological state or disease, which would be characteristic of a diagnostic device. It aids in quality control and calibration of medical devices, rather than direct diagnosis.

No

The device description explicitly states it is a "well type ionization chamber" and describes physical components like the chamber, readout, and pressurized gas, indicating it is a hardware device.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to "measure the activity or output of radioactive brachytherapy sources." This is a measurement of a physical property of a radioactive source, not a test performed on a biological sample (in vitro).
  • Device Description: The device is described as a "well type ionization chamber" that measures current and converts it to activity or air kerma strength. This is a physical measurement device, not a device designed to analyze biological samples.
  • Lack of IVD Characteristics: The description does not mention any of the typical characteristics of an IVD, such as:
    • Testing biological samples (blood, urine, tissue, etc.)
    • Detecting or measuring substances in biological samples
    • Diagnosing, monitoring, or predicting disease based on biological sample analysis

The device is clearly intended for use in a clinical setting to measure the output of radioactive sources used in brachytherapy, which is a form of radiation therapy. This falls under the category of medical devices used in treatment, not in vitro diagnostics.

N/A

Intended Use / Indications for Use

The CRC 15BT dose calibrator is intended to be used by trained medical physicists, or qualified technologists to measure the activity or output of radioactive brachytherapy sources, including HDR, LDR and IVBT sources.

Product codes

90 KPT

Device Description

The CRC 15BT dose calibrator is a well type ionization chamber specifically designed to accurately and rapidly measure all types of brachytherapy sources with appropriate calibration and source positioning device. The ionization chamber is connected to a readout, which converts measured current into a displayed value in units of either Ci (activity) or U (air kerma strength). The unit is sealed and pressurized with UPC Argon. which improves sensitivity and eliminates the need for temperature and pressure corrections.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

trained medical physicists, or qualified technologists

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

The CRC 15BT has been evaluated by the University of Wisconsin ADCL, and found to be a suitable instrument for all clinically available gamma and high-energy beta brachytherapy sources, including HDR, LDR, and IVBT in any number of configurations including, seeds, ribbons, and source trains with appropriate source holders.

The CRC 15BT design is a minor modification to the Capintec CRC 15R dose calibrator to increase the activity range to accommodate HDR sources. The CRC 15R has been commercially available for over 10 years and has an excellent reputation for long term stability, reliability, safety and effectiveness. The modifications to the 15BT are limited to reduced fill gas pressure and in increase bias voltage to accommodate a higher activity range required for HDR sources. The basic chamber design and readout are the same as the CRC 15R. Consequently, the CRC 15BT is expected to maintain the same excellent quality as the CRC 15R. with the same level of safety and effectiveness and long term reliability.

The CRC 15BT has also been tested and approved to the following safety standards for medical equipment:

Medical Electrical Equipment, Part 1 General Requirements - EN60601-1: 1990+A1+A2 Medical Electrical Equipment, Part 1 General Reguirements - IEC60601-1: 1998+A1+A2 Medical Electrical Equipment, Part 1 General Requirements Section 1.2 Collateral standard: Electromagnetic Compatibility-EN60601-1-2 Medical Electrical Equipment, Part 1 General Requirements for Safety Section 1.4 Collateral Standard: Programmable Electrical Medical Systems-IEC60601-1-4:1996 Medical Electrical Equipment, Part 1 General Requirements-Can/CSA C22.2 No.606101-M90 Medical Electrical Systems-IEC 606601-1-1:1993+A1

Key Metrics

Not Found

Predicate Device(s)

Not Found

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.1360 Radionuclide dose calibrator.

(a)
Identification. A radionuclide dose calibrator is a radiation detection device intended to assay radionuclides before their administration to patients.(b)
Classification. Class II.

0

AUG 0 2 2002

Image /page/0/Picture/1 description: The image shows the logo for CAPINTEC, INC. The logo consists of the letters "CII" stacked on top of each other inside of a square. Below the square is the company name, CAPINTEC, INC. To the left of the logo is the number 2002.

January 10, 2001

RE: Summary of Safety and Effectiveness Information for the Capintec CRC 15BT calibrator.

The CRC 15BT dose calibrator is a well type ionization chamber specifically designed to accurately and rapidly measure all types of brachytherapy sources with appropriate calibration and source positioning device. The ionization chamber is connected to a readout, which converts measured current into a displayed value in units of either Ci (activity) or U (air kerma strength). The unit is sealed and pressurized with UPC Argon. which improves sensitivity and eliminates the need for temperature and pressure corrections.

The CRC 15BT has been evaluated by the University of Wisconsin ADCL, and found to be a suitable instrument for all clinically available gamma and high-energy beta brachytherapy sources, including HDR, LDR, and IVBT in any number of configurations including, seeds, ribbons, and source trains with appropriate source holders.

The CRC 15BT design is a minor modification to the Capintec CRC 15R dose calibrator to increase the activity range to accommodate HDR sources. The CRC 15R has been commercially available for over 10 years and has an excellent reputation for long term stability, reliability, safety and effectiveness. The modifications to the 15BT are limited to reduced fill gas pressure and in increase bias voltage to accommodate a higher activity range required for HDR sources. The basic chamber design and readout are the same as the CRC 15R. Consequently, the CRC 15BT is expected to maintain the same excellent quality as the CRC 15R. with the same level of safety and effectiveness and long term reliability.

The CRC 15BT has also been tested and approved to the following safety standards for medical equipment:

Medical Electrical Equipment, Part 1 General Requirements - EN60601-1: 1990+A1+A2 Medical Electrical Equipment, Part 1 General Reguirements - IEC60601-1: 1998+A1+A2 Medical Electrical Equipment, Part 1 General Requirements Section 1.2 Collateral standard: Electromagnetic Compatibility-EN60601-1-2 Medical Electrical Equipment, Part 1 General Requirements for Safety Section 1.4 Collateral Standard: Programmable Electrical Medical Systems-IEC60601-1-4:1996 Medical Electrical Equipment, Part 1 General Requirements-Can/CSA C22.2 No.606101-M90 Medical Electrical Systems-IEC 606601-1-1:1993+A1

1

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 0 2 2002

Ms. Mary Anne Dell CAPINTEC, Inc. 540 Alpha Drive PITTSBURGH PA 15238 USA Re: K020150

Trade/Device Name: CRC 15BT Calibrator Regulation Number: 21 CFR 892.1360 Regulation Name: Radionuclide dose calibrator Regulatory Class: II Product Code: 90 KPT Dated: May 13, 2002 Received: May 14, 2002

Dear Ms. Dell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

2

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-459
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-461
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-461
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-465
Other(301) 594-469

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Henry C. Snogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

3

Appendix 2

Indications for Use Form

Page

510(k) Number (if known):

Device Name: CRC 15BT

Indications For Use: The CRC 15BT dose calibrator is intended to be used by trained medical physicists, or qualified technologists to measure the activity or output of radioactive brachytherapy sources, including HDR, LDR and IVBT sources.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

Nancy Brogdon

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K020150