K Number
K014044
Manufacturer
Date Cleared
2002-03-04

(87 days)

Product Code
Regulation Number
886.1330
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Notal Vision Macular Computerized Psychophysical Test (MCPT) is indicated for The early detection of central and paracentral irregularities (vision abnormalities) in the visual field, most commonly associated with macular degeneration.

Device Description

The device is a computerized interactive software device that provides the user with a series of images oriented in vertical and horizontal planes on a imaging screen that is designed to identify irregularities in the central and paracentral macular field of the human visual system. The device analyzes the image presentation in the process of identifying visual field abnormalities and stores the analysis in a computer server. The device is housed in a PC computer.

AI/ML Overview

The Notal Vision Macular Computerized Psychophysical Test (MCPT) is indicated for the early detection of central and paracentral irregularities (vision abnormalities) in the visual field, most commonly associated with macular degeneration.

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state quantitative "acceptance criteria" for the MCPT in terms of specific sensitivity or specificity targets. Instead, the study's primary objective was to demonstrate that MCPT is equal to or better than the Amsler grid in detecting AMD-related retinal abnormalities. The reported performance compared the MCPT directly against the Amsler Grid.

MetricMCPT PerformanceAmsler Grid PerformanceP-Value (MCPT vs Amsler)Conclusion (based on P-Value)
Overal Sensitivity68.4%25.6%0.0000001MCPT significantly better
Overall Specificity81.8%100%(Implicitly worse)Amsler grid better

Note: While the overall P-value for sensitivity strongly favors MCPT, the P-value for Category 1 (healthy subjects) shows MCPT having significantly lower specificity compared to the Amsler Grid.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: The document does not explicitly state a total sample size for the test set. However, a breakdown of categories is provided:
    • Category 1 (Healthy): 33 subjects
    • Category 2 (AMD): 51 subjects
    • Category 3 (AMD): 20 subjects
    • Category 4 (AMD): 27 subjects
    • Category 5 (AMD): 19 subjects
    • Total subjects across all categories = 150 subjects.
  • Data Provenance: The study was a "prospective single blinded randomized clinical investigation conducted at multiple investigational sites." The document does not specify the country of origin of the data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

The document does not provide information on the number of experts used to establish ground truth or their qualifications. The categories are described as "individuals with defined age-related macular degeneration (AMD) at different stages, and the control subjects were those with normal healthy eyes," implying a clinical diagnosis as ground truth, but not the specific experts or their process.

4. Adjudication Method for the Test Set:

The document does not describe any specific adjudication method for the test set. The categories (AMD at different stages and healthy controls) were "defined," suggesting a pre-established clinical diagnosis, but the process of reaching that definition is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

Yes, a comparative effectiveness study was done. The study compared the MCPT to the Amsler Grid.

  • Effect Size of Human Readers Improve with AI vs. without AI assistance: This study compares an automated device (MCPT) against a traditional manual test (Amsler grid), not the improvement of human readers with AI assistance versus without AI assistance. Therefore, an effect size of human improvement cannot be determined from this document as it's not a reader study with AI assistance. The study focuses on the standalone performance of the MCPT and the Amsler grid themselves.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

Yes, a standalone study was performed. The "MCPT" (Macular Computerized Psychophysical Test) is described as an "interactive software device" that "analyzes the image presentation in the process of identifying visual field abnormalities." The results presented for MCPT are its direct performance, acting as an algorithm-only assessment, which is then compared against the Amsler grid.

7. Type of Ground Truth Used:

The ground truth used was clinical diagnosis/classification of "defined age-related macular degeneration (AMD) at different stages" and "normal healthy eyes." This falls under clinical diagnosis rather than pathology or outcomes data specifically.

8. Sample Size for the Training Set:

The document does not mention a separate training set or its sample size. The clinical investigation described appears to be for validation/testing.

9. How the Ground Truth for the Training Set Was Established:

Since no training set is described, information on how its ground truth was established is not available in the provided document.

{0}------------------------------------------------

K014044

sion Inc.

1 VOLUT - 972-3-6293, ISRAEL. Tel: 972-3-6293763 Fax: 972-3-6293764

MAR 4 2002

510(k) Summary Statement

Submitter:Company Name:Notal Vision, Inc.
Address:5 Droyanov StreetTel Aviv, 63143 Israel
Contact Person:Dr. Zeev Even Chen
Phone:(972) 3 6293763 Ext. 103
Fax:(972) 3 6293764
Registration :Pending

Manufacturer Information:

Company Name:Notal Vision, Inc.
Address:5 Droyanov StreetTel Aviv, 63143 Israel
Contact Person:Dr. Zeev Even Chen
Phone:(972) 3 6293763 Ext. 103

Registration Number: Pending

Official Correspondent:Richard E. Lippman, O.D., F.A.A.O.
Address:C.L. McIntosh, Inc.12300 Twinbrook Parkway Suite 230Rockville, Maryland 20852
Phone :301-770-9590
Fax:301-770-9584

DEVICE IDENTIFICATION:

Trade Name:Macular Computerized Psychophysical Test (MCPT)
Common Name:Amsler Grid
Classification Name:Amsler Grid

CLASSIFICATION Name and Reference:

21 CFR 886.1330 Amsler Grid

REGULATORY CLASS Class I Amsler Grid COMMON NAME: ESTABLISHMENT REGISTRATION: Pending

{1}------------------------------------------------

INDICATION for USE:

The Notal Vision Macular Computerized Psychophysical Test (MCPT) is indicated for The early detection of central and paracentral irregularities (vision abnormalities) in the visual field, most commonly associated with macular degeneration.

SUBSTANTIALLY EQUIVALENT TO:

The Notal Vision Macular Computerized Psychophysical Test is substantially equivalent to an Amsler Grid, a Class I Exempt medical device (21 CFR 886.1330) that is a chart with grids that is held at 30 centimeters distance from the patient and intended to rapidly detect central and paracentral irregularities in the visual field.

DEVICE DESCRIPTION:

The device is a computerized interactive software device that provides the user with a series of images oriented in vertical and horizontal planes on a imaging screen that is designed to identify irregularities in the central and paracentral macular field of the human visual system. The device analyzes the image presentation in the process of identifying visual field abnormalities and stores the analysis in a computer server. The device is housed in a PC computer.

CLINICAL INVESTIGATION

A prospective single blinded randomized clinical investigation was conducted at multiple investigational sites to evaluate the presence of central and paracentral visual field defects in the diagnosis of macular degeneration. The test device was the interactive software driven Macular Computerized Psychophysical Test (MCPT). The Amsler Grid chart was used as a comparison tool in two separate sets of subjects. Test subjects were those individuals with defined age-related macular degeneration (AMD) at different stages, and the control subjects were those with normal healthy eyes. Only one eye was tested in either the test group or the control group.

The primary objective of the study was to evaluate the ability of each technique to distinguish between patients with AMD at different stages and healthy controls without retinal abnormality.

The statistical tests, which was used for analyzing the results of this study, include comparative analysis to demonstrate the hypothesis that the MCPT is equal to or better than the Amsler grid in the detection of AMD-related retinal abnormalities, i.e., to

{2}------------------------------------------------

reject the null hypothesis which states: "The Amsler grid is better then the MCPT in detecting abnormalities in the macular area".

CLINICAL OUTCOMES AND CONCLUSIONS

Efficacy:

The MCPT indicated the presence of AMD significantly more often than the Amsler grid. The overall sensitivity of MCPT was 68.4%, compared with 25.6% of the Amsler grid. This difference is statistically significant. However, MCPT was also inclined to diagnose more Category 1 subjects (healthy subjects-false positives) as having AMD. The specificity of MCPT was 81.8%, compared to 100% for the Amsler grid.

CategoryMCPTAmsler GridP-Value¹
-+SensSpec-+SensSpec
12760.8183301.00.0312500
230210.4124740.0780.0000763
36140.701640.200.0019531
41260.96315120.4440.0001221
50191.09100.5260.0039063
TOTAL²37800.68487300.2560.0000001

MCPT and Amsler Grid Results per Category

= = Positive results; - = Negative results; Sens. = Sensitivity' Spec. = Specificity

1 Calculated according to mcNemar Test for two related dichotomous variables.

2 Categories 2 to 5.

The MCPT sensitivity analysis was calculated as follows: Sensitivity = (21+14 +26 + 19)[(21 +14 +26 +19) + 30 +6 +1 +0)] = 0.684 Specificity = 27/(27 +6) = 0.818

The Amsler Grid sensitivity analvsis was calculated as follows: Sensitivity = (4 +4 +12 + 10)[(4 +4 +12 +10) + (47 +16 +15 +9)] = 0.256 Specificity = 33/(33) = 1

Safety:

No adverse events were observed or reported by the subjects throughout the study. It is concluded that both the MCPT and the Amsler grid are safe.

LABELING

The Notal Vision Macular Computerized Psychophysical Test (MCPT) is provided with a User Manual for the Practitioner. The information is available from the company:

Notal Vision, Inc. 5 Droyanov Street Tel Aviv 63143. Israel (972) 3 6293763

0000312

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized human figure with three faces in profile, one behind the other. The figure is positioned to the right of the text, which reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement around the figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

4 2002 MAR

Notal Vision, Inc. c/o Richard E. Lippman, O.D., F.A.A.O. C.L. McIntosh, Inc. 12300 Twinbrook Parkway Suite 230 Rockville, Maryland 20852

Re: K014044

Trade Name: Macular Computerized Psychophysical Test (MCPT) Regulation Number: CFR 886.1330 Regulation Name: Amsler Grid Regulatory Class: Class I Product Code: HOQ Dated: December 1, 2001 Received: December 7, 2001

Dear Dr. Lippman:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bocaon of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enorosale) to 10gails and the Medical Device Amendments, or to Conimeres proc to may 20, 1976, are excordance with the provisions of the Federal Food, DDIg. de noos that have been that do not require approval of a premarket approval application (PMA). and Cosmette Act (110t) that do not requent to the general controls provisions of the Act. The r ou may , are every missas of the Act include requirements for annual registration, listing of general volusion profiteruring practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I load or how love a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Richard E. Lippman, O.D., F.A.A.O.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

A. Ralph Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

Copy

Indications Statement

510(k) Number (if known)

The Notal Vision Macular Computerized Psychophysical Test (MCPT) Device Name:

Indications for Use:

The Notal Vision Macular Computerized Psychophysical Test (MCPT) is indicated for the eatly dection of central The Notal Vision Macqual Computerized in the visual field, most commonly associated with macular degeneration.

Additional Claims:

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Presription Use

OR

Dennis L. McCarthy

(Division Sign-Off)

Division of Ophthalmic Ear,

Nose and Ti

510(k) Number

Over -The-Counter Use _

(Oprtional Format 1-2-96)

000015

§ 886.1330 Amsler grid.

(a)
Identification. An Amsler grid is a device that is a series of charts with grids of different sizes that are held at 30 centimeters distance from the patient and intended to rapidly detect central and paracentral irregularities in the visual field.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality system regulation in part 820 of this chapter, with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.