(84 days)
The NeuroSensor™ System is intended for use by a qualified neurosurgeon in monitoring cerebral blood flow in patients at risk of cerebral ischemia, and for the direct monitoring of intracranial pressure in both sub-dural and intraparenchymal applications.
The NeuroSensor™ system consists of a single-use combined 2mm diameter parenchymal probe for the real-time measurement of cerebral blood flow (CBF) and intracranial pressure (ICP) and a monitor for the display and storage of these measured variables and the computation and display of derived variables. CBF is measured using Laser Doppler flowmetry and provides real-time measurements of local blood flow in the brain. In the NeuroSensor™ system, low power laser light at 780nm is transmitted down a central fiber to the tip of the probe and illuminates the cerebral tissue. The laser light is scattered by the moving red blood cells and is collected by an array of collecting fibers at the tip of the probe. The reflected light is measured and the resulting signal processed to produce a measure of the perfusion or flux of blood in the local tissue sample volume. The flow measurement is converted into absolute units of ml/min/100g using an algorithm determined by a comparison between measurements made using the combined Laser Doppler probe and the reference method of Quantitative Autoradiography. ICP is monitored directly by a solid state sensor mounted on the side of the NeuroSensor™ probe close to its tip. The sensor is precalibrated in the factory with probe identification and calibration values stored within each probe and there is no requirement for the user to calibrate the probe before use. The NeuroSensor™ monitor uses a color LCD display to show the two measured variables CBF and ICP continuously in real time, both in digital form and as a realtime trace. The monitor can accept measured arterial pressure from an external patient monitor, and can use this data and the measured CBF and ICP to derive cerebral perfusion pressure (CPP) and cerebrovascular resistance (CVR). Data is stored by the monitor and can be displayed as a trend graph over a period of 15 minutes, or 1,2,8 or 24 hours. Completing the system is a single-use cranial access port featuring a titanium alloy bolt, and a convenience procedure kit for cranial access.
The provided text is a 510(k) summary for the NeuroSensor™ system, which measures cerebral blood flow (CBF) and intracranial pressure (ICP). This document does not contain a study that proves the device meets specific acceptance criteria with reported device performance metrics. Instead, it states that "In vitro testing shows that the device meets similar performance specifications as those for the predicate devices."
Therefore, I cannot provide a table of acceptance criteria and reported device performance from the provided text, nor can I answer questions about sample sizes, ground truth establishment, or specific study designs (like MRMC) as this information is not present.
However, I can extract information related to the device and its claimed equivalence:
1. Table of Acceptance Criteria and Reported Device Performance:
- Acceptance Criteria: The document implies that the acceptance criteria are "similar performance specifications as those for the predicate devices." However, the exact quantitative specifications for either the predicate devices or the NeuroSensor™ system are not provided within this text.
- Reported Device Performance: The text states, "In vitro testing shows that the device meets similar performance specifications as those for the predicate devices." No specific performance values (e.g., accuracy, precision, bias) are reported for either CBF or ICP measurements.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- The document mentions "In vitro testing," but it does not provide details on the sample size used, whether the data was retrospective or prospective, or the country of origin of the data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- This information is not provided in the document.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- This information is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- The device is a monitoring system (hardware and software for physiological measurements), not an AI-assisted diagnostic tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable to this device and no such study is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The document implies that the device's performance was evaluated in "in vitro testing." This would typically be a standalone performance evaluation of the device's measurement capabilities. The text states:
- "CBF is measured using Laser Doppler flowmetry... The flow measurement is converted into absolute units of ml/min/100g using an algorithm determined by a comparison between measurements made using the combined Laser Doppler probe and the reference method of Quantitative Autoradiography."
- "ICP is monitored directly by a solid state sensor... The sensor is precalibrated in the factory with probe identification and calibration values stored within each probe..."
- These descriptions suggest a standalone evaluation was performed for the measurement algorithms and sensors, but specific study details are not provided.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For Cerebral Blood Flow (CBF) measurement, the ground truth mentioned for algorithm determination is the "reference method of Quantitative Autoradiography."
- For Intracranial Pressure (ICP) measurement, the ground truth is established through "precalibrated" sensors in the factory.
8. The sample size for the training set:
- This information is not provided. The document mentions an "algorithm determined by a comparison" for CBF, implying some form of training or calibration data was used, but the sample size is not stated.
9. How the ground truth for the training set was established:
- For the CBF measurement algorithm, the ground truth was established by "a comparison between measurements made using the combined Laser Doppler probe and the reference method of Quantitative Autoradiography."
- For ICP, the sensors are "precalibrated in the factory," indicating that the factory calibration process itself establishes the ground truth for the sensor's accuracy.
§ 882.1620 Intracranial pressure monitoring device.
(a)
Identification. An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.(b)
Classification. Class II (performance standards).