K Number
K013713
Date Cleared
2002-07-15

(249 days)

Product Code
Regulation Number
876.5820
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Amuchina Automatic Reprocessing Machine (ARM) is a device designed for both (a) reprocessing hemodialyzers for reuse, and (b) preprocessing hemodialyzers prior to first use. Reprocessed hemodialyzer will be reused on the same patient on who originally used the hemodialyzer. Both the reprocessing and preprocessing procedures use peracetic acid/hydrogen peroxide based disinfectant.

Device Description

The Amuchina ARM Automatic [Dialyzer] Reprocessing Machine is a stand alone device designed for the automated reprocessing of hemodialyzers for reuse and for the pre-processing of hemodialyzers prior to first use. The ARM Unit has 4 stations which can sequentially process up to 4 dialyzers at one time. The ARM Unit has no direct or indirect patient contact. The ARM Unit uses a peracetic acid/hydrogen peroxide based disinfectant as both a cleaning solution and a disinfectant. The disinfectant concentrate is diluted to the inuse strength with AAMI quality water. When reprocessing dialyzers, the ARM Unit uses the following cycles: Rinse, Cleaning, Flush, Volume & Leak Test, and Disinfection. When pre-processing dialvzers, the following cycles are used: Flush, Volume & Leak Test (only if instructed for by the user), and Disinfection. For regularly scheduled maintenance, the ARM Unit has the following system cycles: System Rinse, System Disinfect, and System Self Test. Other cycles which are included in the ARM Unit include: Prime Pump for priming the chemical pump with the disinfecting agent, Line Volume Calibration for use in determining the total cell volume of the dialyzer, and System Void to purge fluids from the circuits prior to moving the machine. The ARM Unit incorporates the feature of including a patient photograph on the dialyzer label, thus reducing the possibility of reused dialyzers being used on the wrong patient.

AI/ML Overview

Here's an analysis of the provided 510(k) summary regarding the Amuchina ARM Automatic Reprocessing Machine:

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary for the Amuchina ARM Automatic Reprocessing Machine (K013713) does not explicitly present a table of "acceptance criteria" in the typical sense of quantitative performance metrics with associated thresholds (e.g., sensitivity > 90%, specificity > 85%). Instead, the "acceptance criteria" are implied by the comparative features and the results of the non-clinical performance and in-vitro testing.

The primary acceptance criterion appears to be functional equivalence to the predicate device (Seratronic DRS-4 Dialyzer Reprocessing System, K860674) for the intended uses of reprocessing and pre-processing dialyzers. The device's performance is demonstrated by functionality tests and in-vitro testing confirming proper dilution of the disinfectant.

FeatureAcceptance Criteria (Implied by Predicate & Intended Use)Reported Device Performance (Amuchina ARM)
Intended UseReprocessing hemodialyzers for reuseReprocessing hemodialyzers for reuse on the same patient.
Pre-processing hemodialyzers prior to first usePre-processes hemodialyzers prior to first use.
Number of Stations44
Cleaning SolutionsCapable of using Peracetic Acid/Hydrogen PeroxideUses Peracetic Acid/Hydrogen Peroxide.
DisinfectantsCapable of using Peracetic Acid/Hydrogen PeroxideUses Peracetic Acid/Hydrogen Peroxide.
CyclesExpected cycles for reprocessing (Clean, Disinfect, Test)Rinse, Cleaning, Flush, Volume & Leak Test, Disinfection (reprocessing). Flush, Volume & Leak Test (optional), Disinfection (pre-processing). System Rinse, System Disinfect, System Self Test, Prime Pump, Line Volume Calibration, System Void (maintenance/utility).
Test Cycle ContentsPressure Leak Test, Total Blood Cell VolumePressure Leak Test, Total Blood Cell Volume.
Dialyzer LabelingComprehensive patient/dialyzer information for reuse safetyBar Code, Patient Name, Patient Photograph (digital), Dialyzer Lot Number, Dialyzer Serial Number, Number times reprocessed, Number times reused, Test Results, Dialyzer Status.
Microprocessor ControlYesYes
Interactive Touch ScreenYesYes
Operator Defined ParametersYesYes
Patient Photograph Feature(Not applicable to predicate, but an added feature for safety)Yes (reduces possibility of wrong patient use).
Water QualityAAMI QualityAAMI Quality
FunctionalityPerform all intended/programmed functions as designed."Each individual function of the ARM Unit was tested to see if they performed as intended/programmed. No errors or failures either-were detected."
Dilution AccuracyProperly dilute cleaner/disinfectant concentrate."In-vitro testing was performed to assure the ARM Unit properly diluted the cleaner/disinfectant concentrate to the in-use concentrations of active ingredients. The results from these tests show that the ARM Unit performed as expected."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a sample size for a "test set" in the context of a clinical study or a large-scale data analysis. The testing described is "Non-Clinical Performance Data" and "In-vitro testing."

  • Test Set Sample Size: Not specified in terms of number of dialyzers or patient cases. The description implies a series of functional tests on the device itself and in-vitro tests on the dilution capabilities.
  • Data Provenance: The testing is described as non-clinical and in-vitro, meaning it was likely conducted in a controlled laboratory environment by the manufacturer (Amuchina International, Inc.). There is no mention of country of origin for data specifically, beyond the company's address in Gaithersburg, MD, USA. It is retrospective in the sense that the testing was completed before the 510(k) submission. No prospective clinical data is mentioned.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the summary. For non-clinical and in-vitro functional testing of a device like this, ground truth is typically established by engineering specifications, chemical analysis, and adherence to established standards (e.g., AAMI quality water). There's no mention of human experts interpreting results for a test set in the way a radiologist might for medical images.

4. Adjudication Method for the Test Set

This is not applicable as there's no mention of human interpretation of results requiring adjudication (e.g., by multiple experts). The tests described are objective functional and chemical tests.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done, nor is it applicable. This device is an automated reprocessing machine; it does not involve human "readers" or AI assistance for diagnostic interpretation. Its function is to mechanically and chemically prepare dialyzers for reuse.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a form of standalone performance was demonstrated through the non-clinical and in-vitro testing. The device's "individual functions" and its ability to "properly dilute the cleaner/disinfectant concentrate" were tested as an automated system without human intervention influencing the core performance metrics. The machine itself is designed to operate without a "human-in-the-loop" for its primary reprocessing functions, beyond operator setup and maintenance.

7. The Type of Ground Truth Used

The ground truth used for this device's evaluation is primarily based on:

  • Engineering Specifications and Design Requirements: Ensuring each function performs "as intended/programmed."
  • Chemical Analysis: Verifying "in-use concentrations of active ingredients" following dilution.
  • Comparison to Predicate Device's Performance Characteristics: Demonstrating substantial equivalence to the Seratronic DRS-4.
  • Adherence to Standards: Implicitly, AAMI quality water standard.

There is no mention of pathology, outcomes data, or expert consensus in a clinical diagnostic sense.

8. The Sample Size for the Training Set

There is no mention of a training set in this 510(k) summary. This type of device does not utilize machine learning or AI models that require a "training set." Its functionality is based on programmed logic and mechanical/chemical processes.

9. How the Ground Truth for the Training Set Was Established

As there is no training set mentioned or applicable for this device, this question is not applicable.

{0}------------------------------------------------

JUL 15 2002

K013713

510(k) Summary

as required by section 807.92(c).

Name:Ludovico GlavottoPresident
Address:Amuchina International, Inc.8-8 Metropolitan CourtGaithersburg, MD 20878
Telephone:FAX:301-330-7597301-330-6432
Contact Person:Ludovico Glavotto
Date SummaryPrepared:01 November 2001
Trade Name:Common Name:Classification Name:ARM Automatic Reprocess Machinedialyzer reprocessing systemdialyzer reprocessing system, panel 78, procode LIF, unclassified

Legally Marketed Device claiming Substantial Equivalence to, §807.92(a)(3):

Seratronic DRS-4 Dialyzer Reprocessing System, K860674

Description of Device, §807.92(a)(4):

The Amuchina ARM Automatic [Dialyzer] Reprocessing Machine is a stand alone device designed for the automated reprocessing of hemodialyzers for reuse and for the pre-processing of hemodialyzers prior to first use. The ARM Unit has 4 stations which can sequentially process up to 4 dialyzers at one time. The ARM Unit has no direct or indirect patient contact.

The ARM Unit uses a peracetic acid/hydrogen peroxide based disinfectant as both a cleaning solution and a disinfectant. The disinfectant concentrate is diluted to the inuse strength with AAMI quality water.

When reprocessing dialyzers, the ARM Unit uses the following cycles: Rinse, Cleaning, Flush, Volume & Leak Test, and Disinfection. When pre-processing dialvzers, the following cycles are used: Flush, Volume & Leak Test (only if instructed for by the user), and Disinfection.

For regularly scheduled maintenance, the ARM Unit has the following system cycles: System Rinse, System Disinfect, and System Self Test. Other cycles which are included in the ARM Unit include: Prime Pump for priming the chemical pump with the disinfecting agent, Line Volume Calibration for use in determining the total cell volume of the dialyzer, and System Void to purge fluids from the circuits prior to moving the machine.

{1}------------------------------------------------

KO13713-

The ARM Unit incorporates the feature of including a patient photograph on the dialyzer label, thus reducing the possibility of reused dialyzers being used on the wrong patient.

Intended Use of Device, §807.92(a)(5):

The ARM Automatic [Dialyzer] Reprocessing Machine is a device to reprocess hemodialyzers for reuse on the same patient. The patient population are those patients who have End-Stage Renal Disease and are on chronic hemodialysis. The ARM Unit is designed to be either a stand-alone device or act as either a server or a slave Unit in a network of several ARM Units.

The ARM Unit also can pre-processes hemodialyzers prior to first use, where preprocessing is the medical policy of the health care facility.

Summary of Technological Characteristics of ARM Automatic Reprocessing Machine and the DRS-4 Dialyzer Reprocessing Systems, §807.92(a)(6):

FreatureAmuchina ARMSeratronics DRS-4
Indications for UseReprocessing hemodialyzersfor reuse and Pre-processinghemodialyzers prior to firstuseReprocessing Hemodialyzersfor multiple use andPrepacking hemodialyzersfor specific patients
Number of ProcessingStations44
Cleaning SolutionsPeracetic Acid/HydrogenPeroxidePeracetic Acid/ HydrogenPeroxideBleach/Sodium Hypochlorite
DisinfectantsPeracetic Acid/HydrogenPeroxidePeracetic Acid/HydrogenPeroxideFormaldehyde
CyclesRinse, Clean, Flush, Test,DisinfectClean, Disinfect, Test
Test Cycle IncludesPressure Leak TestTotal Blood Cell VolumePressure Leak TestTotal Blood Cell VolumeUltrafiltration Rate
Dialyzer LabelsProcessing for ReuseIncludes:Bar CodePatient NamePatient Photograph (digital)Dialyzer Lot NumberDialyzer Serial NumberNumber times reprocessedNumber times reusedTest ResultsDialyzer StatusIncludes:Bar CodePatient ID Number & NameSocial Security NumberDialyzer Type (code #)Dialyzer CodeNumber times reusedTest ResultsDialyzer Status

{2}------------------------------------------------

SL 313

Dialyzer LabelWhen Added to List for(No label printed. If pre-
Pre-processingPre-processing:processing a dialyzer prior to
Bar Codefirst use, the DRS-4 system
Status: Do Not Userecords the TCV, and the
Dialyzer ModelKUF values foe each
Dialyzer Lot Numberdialyzer, and enters the
Dialyzer Serial Numberresults in the database.)
After Pre-processing:
Bar Code
Dialyzer Lot Number
Dialyzer Serial Number
Number times reprocessed
Number times reused
Microprocessor controlledYesYes
Interactive Touch ScreenYesYes
Operator can define someYesYes
parameters
Camera for PatientYesNo
Photograph
Water Requirements:
Pressure30-115 psig25-80 psig
QualityAAMI QualityAAMI Quality
Temperature15 - 25 degrees C24-35 degrees C
Flowup to 1500 ml/min at 2 bar1800 ml/min @ 1.75 kg/cm
Drain Heightup to 16 inchesup to 16 inches max

Non-Clinical Performance Data, §807.92(b)(1):

Each individual function of the ARM Unit was tested to see if they performed as intended/programmed. No errors or failures either-were detected and the performance characteristics of the down-stream processing procedures were not affected by the preceding test(s).

In-vitro testing was performed to assure the ARM Unit properly diluted the cleaner/ disinfectant concentrate to the in-use concentrations of active ingredients. The results from these tests show that the ARM Unit performed as expected.

Conclusions Drawn from Non-Clinical Performance Data, §807.92(b)(3):

The functionality tests on the ARM Unit demonstrate that the ARM Automatic Reprocessing Machine will perform as labeled for the reprocessing of hemodialyzers. The results of these tests demonstrate that the ARM Automatic Reprocessing Machine is substantially equivalent to the DRS-4 Dialyzer Reprocessing System. which is commercially distributed, for the same intended use that is the reprocessing of hemodialyzers for reuse.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of an eagle with three human profiles embedded within the eagle's body.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 5 2002

Mr. Gary Mishkin Vice President of Research and Development Alcavis International, Inc. 8-8 Metropolitan Court GAITHERSBURG MD 20878 Re: K013713 Trade/Device Name: ARM Automatic Reprocessing Machine Regulation Number: None Regulatory Class: Unclassified Product Code: 78 LIF

Dated: April 12, 2002 Received: April 16, 2002

Dear Mr. Mishkin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known):

Device Name: ARM Dialyzer Reprocess Unit

Indications for Use:

The Amuchina Automatic Reprocessing Machine (ARM) is a device designed for both (a) reprocessing hemodialyzers for reuse, and (b) preprocessing hemodialyzers prior to first use. Reprocessed hemodialyzer will be reused on the same patient on who originally used the hemodialyzer. Both the reprocessing and preprocessing procedures use peracetic acid/hydrogen peroxide based disinfectant.

8013713

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancyc Brogdon

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K013713

Prescription Use
(Per 21 CFR 801.109)

(Optional Format 3-10-98)

§ 876.5820 Hemodialysis system and accessories.

(a)
Identification. A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient.
(2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860).
(3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860).
(4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
(b)
Classification. (1) Class II (performance standards) for hemodialysis systems and all accessories directly associated with the extracorporeal blood system and the dialysate delivery system.(2) Class I for other accessories of the hemodialysis system remote from the extracorporeal blood system and the dialysate delivery system, such as the unpowered dialysis chair, hemodialysis start/stop tray, dialyzer holder set, and dialysis tie gun and ties. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.