K Number
K013232
Date Cleared
2001-11-20

(54 days)

Product Code
Regulation Number
886.5928
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CyberCases by Bausch & Lomb are indicated for storage of hard, rigid gas permeable (fluro silicone acrylate and silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection.

Device Description

CyberCases by Bausch & Lomb consist of a polypropylene cap and polypropylene body for storage of hard, rigid gas permeable (fluro silicone acrylate and silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection.

AI/ML Overview

Acceptance Criteria and Study for CyberCases™ by BAUSCH & LOMB

This submission is for a contact lens case, which is a relatively low-risk device. The regulatory pathway chosen is a 510(k) premarket notification, which relies on demonstrating substantial equivalence to a legally marketed predicate device. Therefore, a comprehensive clinical study with specific performance metrics such as sensitivity, specificity, or AUC is not typically required or performed for this type of device. The "acceptance criteria" are primarily established through a series of in-vitro studies to demonstrate safety and effectiveness equivalent to the predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied from Submission)Reported Device Performance
Biocompatibility/Toxicity: No significant toxicological effects on contact lenses or ocular tissues from materials in the contact lens case."A series of in-vitro toxicological... studies were performed to assess the safety... All studies were conducted in compliance with the Good Laboratory Practice Regulation for Nonclinical Laboratory Studies. Results of the testing demonstrate that the safety... of the CyberCases by Bausch & Lomb is equivalent to the predicate device, Bausch & Lomb Sight Savers Contact Lens Case."
Chemical Compatibility/Leachables: Materials in the contact lens case do not leach harmful substances into contact lens solutions or affect lens integrity."A series of in-vitro... chemical studies were performed to assess the... effectiveness of the CyberCases by Bausch & Lomb in accordance with the guidelines set forth in FDA's May 1, 1997 Guidance for Industry - Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products." Results support equivalence to the predicate device.
Material Durability/Functionality: The device should maintain its structural integrity and function (e.g., proper sealing) during its intended use.Implied by "effectiveness" studies demonstrating equivalence to the predicate. No specific performance metrics (e.g., seal strength, number of cycles) are provided in the summary.
Indications for Use: The device can safely and effectively store hard, rigid gas permeable, and soft contact lenses during chemical disinfection."CyberCases by Bausch & Lomb are substantially equivalent in terms of indications for use, safety and effectiveness to the predicate device, Bausch & Lomb Sight Savers Contact Lens Case..."
Equivalence to Predicate Device: Demonstration of substantial equivalence to the Bausch & Lomb® Sight Savers® Contact Lens Case (K852384) in terms of safety and effectiveness."Results of the testing demonstrate that the safety and effectiveness of the CyberCases by Bausch & Lomb is equivalent to the predicate device, Bausch & Lomb Sight Savers Contact Lens Case."

2. Sample Size Used for the Test Set and Data Provenance

The provided 510(k) summary does not specify the sample sizes used for the in-vitro toxicological and chemical studies. It generically states "A series of in-vitro toxicological and chemical studies were performed." Without explicit mention, it is assumed these were laboratory-based studies, not involving human subjects or patient data. Therefore, "data provenance" in terms of country of origin or retrospective/prospective does not apply in the typical sense for clinical data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

N/A. Given that the studies were "in-vitro toxicological and chemical studies" conducted in a lab environment, there would not be "experts" establishing a "ground truth" in the clinical sense (e.g., radiologists interpreting images). The "ground truth" would be objective scientific measurements and observations from the laboratory tests (e.g., chemical assays, cell viability tests), interpreted by qualified scientists or toxicologists. The submission does not specify the number or qualifications of the personnel who interpreted these lab results.

4. Adjudication Method for the Test Set

N/A. This is not applicable to in-vitro toxicological and chemical studies. Adjudication methods like 2+1 or 3+1 are used for clinical trials or image interpretation where human expert agreement is needed to establish ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically conducted for medical imaging devices or diagnostic aids where human interpretation is involved and the impact of an AI algorithm on reader performance is being evaluated. This submission relates to a contact lens case, not a diagnostic imaging device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

N/A. This device is a physical product (contact lens case), not an algorithm or AI-powered system. Therefore, the concept of "standalone performance" for an algorithm is not applicable. The device's "performance" is demonstrated through its physical and chemical properties and interaction with contact lenses and solutions.

7. Type of Ground Truth Used

The "ground truth" for the performance evaluation of the CyberCases™ was established through objective laboratory measurements and scientific observations from in-vitro toxicological and chemical studies. This would include direct measurements of material properties, chemical leachables, and effects on cell cultures or contact lens materials.

8. Sample Size for the Training Set

N/A. This device is not an AI/ML algorithm or software that requires a "training set." The submission describes a physical product, and its evaluation relies on established in-vitro testing methodologies rather than machine learning.

9. How the Ground Truth for the Training Set Was Established

N/A. As there is no training set for this physical device, there is no ground truth to establish for it.

{0}------------------------------------------------

K013232

NOV 2 0 2001

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

CyberCases™ by BAUSCH & LOMB

1. Submitter Information:

Bausch & Lomb Incorporated 1 400 North Goodman Street P.O. Box 30450 Rochester, New York 14603-0450

Contact Person:Kim S. DeVitto
Manager, Regulatory Affairs
Telephone No.:716-338-6401
Fax No.:716-338-0702

2. Device Name:

Classification Name: Contact Lens Cases

Proprietary Name: CyberCases™ by Bausch & Lomb

Predicate Device: 3.

The Bausch & Lomb® Sight Savers® Contact Lens Case was selected as the predicate device for the CyberCases by Bausch & Lomb.

4. Description of the Device

CyberCases by Bausch & Lomb consist of a polypropylene cap and polypropylene body for storage of hard, rigid gas permeable (fluro silicone acrylate and silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection.

{1}------------------------------------------------

ડ. Indications for Use:

CyberCases by Bausch & Lomb are indicated for storage of hard, rigid gas permeable (fluro silicone acrylate and silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection.

Description of Safety and Substantial Equivalence ર્.

A series of in-vitro toxicological and chemical studies were performed to assess the safety and effectiveness of the CyberCases by Bausch & Lomb in accordance with the guidelines set forth in FDA's May 1, 1997 Guidance for Industry - Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products. All studies were conducted in compliance with the Good Laboratory Practice Regulation for Nonclinical Laboratory Studies.

Results of the testing demonstrate that the safety and effectiveness of the CyberCases by Bausch & Lomb is equivalent to the predicate device, Bausch & Lomb Sight Savers Contact Lens Case.

7. Substantial Equivalence

CyberCases by Bausch & Lomb are substantially equivalent in terms of indications for use, safety and effectiveness to the predicate device, Bausch & Lomb Sight Savers Contact Lens Case, cleared for marketing under 510(k) Notification #K852384 on 7/31/1985. The addition of different colorants to the lens case does not effect the use of this product.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it and a pair of wings at the top.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 2 0 2001

Bausch & Lomb, Inc. c/o Kim S. DeVitto Manager, Regulatory Affairs 1400 N Goodman Street P.O. Box 30450 Rochester, NY 14603

Re: K013232

Trade/Device Name: CyberCases™ by Bausch & Lomb® Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) Contact Lens Care Products (contact lens cases) Regulatory Class: Class II Product Code: LRX Dated: September 26, 2001 Received: September 27, 2001

Dear De Vitto:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

A. Ralph Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

K013232

NOV 2 0 2001

Bausch & Lomb Incorporated 1400 North Goodman Street P.O. Box 30450 Rochester, NY 14603-0450

INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

N/A

Device Name:

CyberCases™ by Bausch & Lomb

Indications for Use:

CyberCases by Bausch & Lomb are indicated for use in the storage of hard, rigid gas permeable (fluro silicone acrylate and silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
------------------------------------------------------------
Prescription Use Use (Per 21 CFR 801.109)
---------------------------------------------
OROver-The-Counter
(Division Sign-Off) Division of Ophthalmic Devices
510(k) NumberK013232
------------------------

§ 886.5928 Soft (hydrophilic) contact lens care products.

(a)
Identification. A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.(b)
Classification. Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”