K Number
K012953
Device Name
GIOTTO IMAGE
Manufacturer
Date Cleared
2001-11-30

(87 days)

Product Code
Regulation Number
892.1710
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Giotto Image X-ray Mammographic System is intended to provide filmscreen X-ray imaging of the breast that can be imaged on a 18 x 24cm or 24 x 30cm film. With optional accessories the Giotto Image can also be used as the imaging device for stereotactic needle biopsy.
Device Description
The Giotto Image is a conventional film-screen x-ray mammographic device. This device is designed to generate a high resolution image of the breast on film using an x-ray source designed specifically for mammography. It incorporates a state-of-the-art microprocessor controlled high frequency single phase x-ray generator and automatic exposure control, molybdenum rotating anode x-ray tube with 0.3 and 0.1 mm focal spots, gantry, compression device and film holder. The unique open design of the Giotto Image allows positioning of the patient with the operator facing the patient. This face-to-face positioning allows the operator to carefully center the breast and compress it quickly and efficiently. Optional accessories allow both prone and seated breast needle biopsy.
More Information

Not Found

No
The description focuses on conventional film-screen X-ray technology and hardware components, with no mention of AI or ML.

No
The device is described as an X-ray imaging system for mammography and stereotactic needle biopsy, which are diagnostic and interventional procedures, not therapeutic.

Yes
The device is described as an "X-ray Mammographic System" intended to "provide filmscreen X-ray imaging of the breast" and can also be used as an "imaging device for stereotactic needle biopsy." These functions involve creating images of internal body structures to aid in the diagnosis of diseases or conditions, which is the definition of a diagnostic device.

No

The device description explicitly details hardware components such as an x-ray generator, x-ray tube, gantry, compression device, and film holder, indicating it is a physical medical device, not software-only.

Based on the provided information, the Giotto Image X-ray Mammographic System is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device Function: The Giotto Image is an X-ray imaging system. It uses X-rays to create images of the breast on film. This is an in vivo process (performed on a living organism), not an in vitro process (performed outside the body).
  • Intended Use: The intended use is to provide X-ray imaging of the breast and to be used as an imaging device for stereotactic needle biopsy. Both of these are imaging procedures performed directly on the patient.

Therefore, the Giotto Image falls under the category of medical imaging devices, not in vitro diagnostics.

N/A

Intended Use / Indications for Use

The Giotto Image X-ray Mammographic System is intended to provide filmscreen X-ray imaging of the breast that can be imaged on a 18 x 24cm or 24 x 30cm film. With optional accessories the Giotto Image can also be used as the imaging device for stereotactic needle biopsy.

Product codes (comma separated list FDA assigned to the subject device)

90IZH

Device Description

The Giotto Image is a conventional film-screen x-ray mammographic device. This device is designed to generate a high resolution image of the breast on film using an x-ray source designed specifically for mammography. It incorporates a state-of-the-art microprocessor controlled high frequency single phase x-ray generator and automatic exposure control, molybdenum rotating anode x-ray tube with 0.3 and 0.1 mm focal spots, gantry, compression device and film holder. The unique open design of the Giotto Image allows positioning of the patient with the operator facing the patient. This face-to-face positioning allows the operator to carefully center the breast and compress it quickly and efficiently. Optional accessories allow both prone and seated breast needle biopsy.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

X-ray

Anatomical Site

Breast

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K973856

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 892.1710 Mammographic x-ray system.

(a)
Identification. A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.

0

NOV 30 2001

510(k) Summary

K012953
page 1 of 2

Submitter Information 1.

  • Submitter: 1.1. MEDX Incorporated 3456 N. Ridge Ave., #100 Arlington Heights IL 60004 Phone: (847) 463-2020 Fax. (847) 463-2019
  • Manufacturing Facility: 1.2. Internazionale Medico Scientifica S.r.l. Via Sagittario, 5 - 40044 Pontecchio Marconi Bologna, Italy
  • 1.3. Contact: Floyd Rowan
  • Date: August 30, 2001 1.4.

Device Name 2.

| 2.1. | Classification Name:
Classification Number: | System Mammographic
90IZH |
|------|------------------------------------------------|------------------------------|
| 2.2. | Trade/Proprietary Name: | Giotto Image |
| 2.3. | Predicate Device: | Giotto HT (DC K973856) |

Device Description 3.

3.1. Function

The Giotto Image is a conventional film-screen x-ray mammographic device. This device is designed to generate a high resolution image of the breast on film using an x-ray source designed specifically for mammography. It incorporates a state-of-the-art microprocessor controlled high frequency single phase x-ray generator and automatic exposure control, molybdenum rotating anode x-ray tube with 0.3 and 0.1 mm focal spots, gantry, compression device and film holder. The unique open design of the Giotto Image allows positioning of the patient with the operator facing the patient. This face-to-face positioning allows the operator to carefully center the breast and compress it quickly and efficiently. Optional accessories allow both prone and seated breast needle biopsy.

1

510(k) Summary

K012953
Page 2 of 2

Scientific Concepts: 3.2.

X-ray imaging began in the late nineteenth century with the discovery of xrays by William Conrad Roentgen. X-ray imaging of the human anatomy, including the breast, began immediately after this discovery. The danger of x-ray exposure, tissue damage caused by ionizing radiation, was quickly x ray expectand today is known to be a major limitation of x-ray imaging. It was not until the 1980's that the unique requirements of x-ray mammography for the detection of non-palpable lesions were realized, manimography the development of specialized x-ray mammographic units. These devices incorporate special x-ray tubes and low ripple x-ray generators producing much lower energy x-rays than conventional x-ray systems. These low energy x-rays allow penetration of the breast tissue while at the same time producing sufficient contrast on the film to detect non-palpable breast lesions. Film screen systems for mammography have been optimized to produce the high resolution images required by mammography while at the same time minimizing dose to the patient.

Physical and Performance Characteristics: 3.3.

Mammography has been demonstrated to be the best imaging choice for screening of women for breast cancer by many studies and is currently recommended as a routine procedure for women over 50 years of age. The FDA has introduced the MQSA program to ensure consistent quality among mammography providers. The MQSA has adopted the accreditation program administered by the American College of Radiology (ACR). This program sets forth requirements for mammography equipment including image resolution, contrast resolution, dose, kV accuracy, etc. The Giotto Image has been designed to meet or exceed all the ACR requirements.

Device Intended Use 4.

  • The intended uses of the Giotto Image are identical to the intended uses of 4.1. the Giotto HT predicate device (Premarket notification K973856).

Device Technological Characteristics 5.

  • The characteristics of the Giotto Image system compare substantially with 5.1. the Giotto HT, in both materials used, technology applied, and functional methodology. Differences of note do not affect safety and effectiveness of the device, intended use, or application methods. The device operates in a manner substantially equivalent to other cleared devices in this category, and performs as well as the predicate Giotto HT.
  • The components of the Giotto Image that come in direct contact with the 5.2. patient (paddles, supports, holders, Bucky) are of the same materials as the Giotto HT predicate device (Premarket notification K973856).

2

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 3 0 2001

Mr. Floyd R. Rowan Executive Vice President Medx, Inc. 3456 N. Ridge Ave., #100 ARLINGTON HEIGHTS IL 60004 Re: K012953 Trade/Device Name: Giotto Image Mammographic x-ray system Regulation Number: 21 CFR 892.1710 Regulation Name: Mammographic x-ray system Regulatory Class: II Product Code: 90 IZH Dated: August 30, 2001 Received: September 4, 2001

Dear Mr. Rowan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the exercised of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

3

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Yancy C. Snigdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

Page 1 of 1

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: Giotto Image X-ray Mammographic System

Indications For Use:

The Giotto Image X-ray Mammographic System is intended to provide filmscreen X-ray imaging of the breast that can be imaged on a 18 x 24cm or 24 x 30cm film. With optional accessories the Giotto Image can also be used as the imaging device for stereotactic needle biopsy.

(Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off Division of Reproduct ·············································································································································································· ാ 10(k) Number

(Optional Format 3-10-98)

Prescription Use