K Number
K012631

Validate with FDA (Live)

Date Cleared
2001-12-06

(115 days)

Product Code
Regulation Number
878.4400
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Reprocessed soft tissue ablators are to be used in patients requiring arthroscopic surgery of the knee, shoulder, ankle, hip, wrist and elbow to resect, ablate and excise soft tissue.

Device Description

Soft tissue ablators are radiofrequency (RF) surgical tools designed for removal and dissection of tissue in arthroscopic surgeries. A generator or controller serves as the power unit. Soft tissue ablators are available in a wide range of sizes and angles; some models include thermal or suction functions.

AI/ML Overview

This document describes the reprocessing of soft tissue ablators, not an AI device. Therefore, the specific questions regarding AI device performance metrics, such as ground truth establishment with experts, multi-reader multi-case studies, and training set sample sizes, are not applicable.

The acceptance criteria and device performance for the Reprocessed Soft Tissue Ablators are based on the principle of substantial equivalence to predicate devices, ensuring safety and effectiveness are maintained after reprocessing.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
BiocompatibilityValidation of reprocessing procedures confirms biocompatibility of the reprocessed devices.
FunctionalityPerformance testing demonstrates that Reprocessed Soft Tissue Ablators perform as originally intended by the manufacturer.
Intended UseThe reprocessed devices maintain the same intended use: to remove soft tissue and control bleeding using high-frequency electrical current in arthroscopic surgeries.
MaterialsThe reprocessed devices are identical in materials to the predicate devices.
DesignThe reprocessed devices are identical in design (standard mechanical design, shapes, and sizes) to the predicate devices.
Performance SpecificationsThe reprocessed devices maintain the same performance specifications as the predicate devices.
Clinical ApplicationsNo changes to clinical applications compared to predicate devices.
Patient PopulationNo changes to the patient population compared to predicate devices.
Method of OperationNo changes to the method of operation compared to predicate devices.

Study Proving Device Meets Acceptance Criteria:

A study involving bench and laboratory testing was conducted to demonstrate the performance (safety and effectiveness) of the Reprocessed Soft Tissue Ablators.

2. Sample Size for Test Set and Data Provenance

The document does not specify a distinct "test set" in the context of AI development. Instead, the performance was validated through bench and laboratory testing. The sample size for these tests is not explicitly stated in the provided text.
Data provenance: Not applicable in the context of AI. The tests were conducted on reprocessed physical devices.

3. Number of Experts and Qualifications for Ground Truth

Not applicable. The "ground truth" for this device relates to its physical and functional integrity after reprocessing, not diagnostic accuracy requiring expert consensus on medical images or patient data. The standard for ground truth is equivalence to the original manufacturer's specifications.

4. Adjudication Method for Test Set

Not applicable. Adjudication methods are typically relevant for human interpretation of data, particularly in clinical effectiveness studies or for establishing ground truth in AI. The performance validation here involves objective engineering and biological tests.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This device is not an AI diagnostic or assistive tool for human readers. No MRMC study was conducted.

6. Standalone (Algorithm Only) Performance Study

Not applicable. This is a reprocessed physical medical device, not an algorithm or software.

7. Type of Ground Truth Used

The ground truth used for this study is implicitly the original manufacturer's specifications and performance standards for the single-use devices. The reprocessed devices must perform "as originally intended" and be "substantially equivalent" to these new devices. This includes biocompatibility, functional performance, and maintaining the original design, materials, and intended use.

8. Sample Size for Training Set

Not applicable. There is no training set as this is not an AI device.

9. How Ground Truth for Training Set Was Established

Not applicable. As this is not an AI device, there is no training set or ground truth establishment method for a training data set.

{0}------------------------------------------------

DEC 0 6 2001

K 012631

510(k) Summary of Safety and Effectiveness

Submitter:Alliance Medical Corporation10232 South 51st StreetPhoenix, Arizona 85044
Contact:Don SelveyVice President, Regulatory Affairs and Quality Assurance(480) 763-5300
Date of9 August 2001
preparation:Name of device:Reprocessed Soft Tissue Ablators
Common Name:Soft Tissue Ablators
ClassificationName:Electrosurgical Cutting and Coagulation Device and Accessories
Reprocessed device(s):
ManufacturerDescriptionModel Number
AthroCare®Arthro Wand® Right AngleA 1325-01
AthroCare®Arthro Wand® Right AngleA 1330-01
AthroCare®Arthro Wand® Right AngleA 1335-01
AthroCare®Arthro Wand® Right AngleA 1336-01
AthroCare®Arthro Wand® Right AngleA 1345-01
AthroCare®Arthro Wand® DomeA 3625-01
AthroCare®Arthro Wand® DomeA 3525-01
AthroCare®Arthro Wand® DomeA 3630-01
AthroCare®Arthro Wand® DomeA 3430-01
AthroCare®Arthro Wand® DomeA 3530-01
AthroCare®Arthro Wand® BevelA 2630-01
AthroCare®Arthro Wand® BevelA 2430-01
AthroCare®Arthro Wand® BevelA 2530-01
AthroCare®Arthro Wand® Small JointA 2723-01
AthroCare®Arthro Wand® Small JointA 2823-01
AthroCare®Arthro Wand® Small JointA 1115-01
AthroCare®Arthro Wand® BisectorA 4330-01
AthroCare®Arthro Wand® StraightA 1125-01
AthroCare®Arthro Wand® StraightA 1130-01
AthroCare®Arthro Wand® AngleA 1225-01
AthroCare®Arthro Wand® AngleA 1230-01

{1}------------------------------------------------

Mitek®VAPR™ Thermal Side Effect Electrode (3.5mm)225101
Mitek®VAPR™ Thermal Reversed Angled Side Effect Electrode(3.5mm)225112
Mitek®VAPR™ Thermal Angled End Effect Electrode (3.5mm)225104
Mitek®VAPR™ Side Effect Electrode (3.5mm)225301
Mitek®VAPR™ Angled Side Effect Electrode (21°, 3.5mm)225302
Mitek®VAPR™ End Effect Electrode (3.5mm)225303
Mitek®VAPR™ Angled End Effect Electrode (21°, 3.5mm)225304
Mitek®VAPR™ 90° Hook Electrode (3.5mm)225305
Mitek®VAPR™ Flexible Side Effect Electrode (3.5mm)225312

Predicate device(s):

Image /page/1/Picture/2 description: The image contains a check mark. The check mark is black and is on a white background. The check mark is oriented from the bottom left to the top right. The check mark is used to indicate that something is correct or has been completed.

K000036ArthroCare, ArthroCare® ENTec™ ReFlex™ Wand
K000044ArthroCare, ArthroCare® Orthopedic ElectrosurgerySystem
K000074ArthroCare, ArthroCare® Electrosurgery System
K000228ArthroCare® ENTec™ Surgery System
K000511ArthroCare® Orthopedic Electrosurgery System
K000936Mitek, VAPR™ 2.3mm Wedge Electrode
K001302ArthroCare, ArthroCare® Electrosurgery System
K001936ArthroCare ENTec® Surgery system, ArthroCare®Orthopedic Surgery System, and ArthroCare®Electrosurgery System (Electrosurgery Systems)
K943450ArthroCare, ArthroCare® Arthroscopic ElectrosurgerySystem 970
K955531ArthroCare® Bipolar Loop Electrosurgery System
K960169ArthroCare® Urologic Multi-electrode ElectrosurgerySystem
K962445ArthroCare® Dental Wand Electrosugery System
K963123Arthrocare, ArthroCare® Arthroscopic ElectrosurgerySystem 980 (Model 2000)
K963783Mitek, VAPR™ Electrosurgical System
K964849ArthroCare® Dermatology Electrosurgery System
K971532ArthroCare® Electrosurgery System
K973478ArthroCare®, AccENT Head and Neck ElectrosurgerySystem
K974022Mitek, VAPR™ T Thermal Electrode
K981870ArthroCare® Visage Cosmetic Surgery System
K992180ArthroCare® Visage™ Cosmetic Surgery System
K992406Mitek, VAPR™ 2.3mm End Effect Electrode
K992581ArthroCare, ArthroCare® Orthopedic ElectrosurgerySystem
K992972ArthroCare, ArthroCare® Electrosurgery System

Device description:

Soft tissue ablators are radiofrequency (RF) surgical tools designed for removal and dissection of tissue in arthroscopic surgeries. A generator or controller serves as the power unit. Soft tissue ablators are available in a wide range of sizes and angles; some models include thermal or suction functions.

{2}------------------------------------------------

  • Reprocessed Soft Tissue Ablators are intended to remove soft tissue and Intended use: control bleeding by providing tissue ablation and coagulaiton utilizing high-frequency electrical current in patients requiring general or arthroscopic surgery. Reprocessed soft tissue ablators are to be used in patients requiring Indications arthroscopic surgery of the knee, shoulder, ankle, hip, wrist and elbow statement: to resect, ablate and excise soft tissue. The design, materials, and intended use of the Reprocessed Soft Tissue Technological Ablators are identical to the predicate devices. The mechanism of action characteristics: of the Reprocessed Soft Tissue Ablator is identical to the predicate devices in that the same standard mechanical design, materials, shapes and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. Bench and laboratory testing was conducted to demonstrate Performance data: performance (safety and effectiveness) of the Reprocessed Soft Tissue Ablators. Biocompatibility ● . Validation of reprocessing Function test(s) . Performance testing demonstrates that Reprocessed Soft Tissue Ablators perform as originally intended.
    Conclusion:

In accordance with the Federal Food, Drug and Cosmetic Act 21 CFR Part 807 and based on the information provided in this premarket notification, Alliance Medical Corporation concludes that the modified device (the Reprocessed Soft Tissue Ablator) is safe, effective and substantially equivalent to the predicate devices as described herein.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 0 6 2001

Alliance Medical Corporation Mr. Don Selvey Vice President Regulatory Affairs and Quality Assurance 10232 South 51ª Street Phoenix, Arizona 85044

Re: K012631

Trade Name: Reprocessed Soft Tissue Ablators Regulation Number: 888.1100, 878.4400 Regulation Name: Arthroscope, Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: HRX, GEI

Dated: November 13, 2001 Received: November 20, 2001

Dear Mr. Selvey:

We have reviewed your Section 510(k) premarket notification of intent to market the device we have it viewed your boouse evermined the device is substantially equivalent (for the iteleficed above and we have done) to devices marketed in interstate commerce prior to may 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have May 26, 1970, the chartence with the provisions of the Federal Food, Drug, and Cosmetic Act
been reclassified in accordance with the provisions of the Federal Food, Drug, and occh reclassified in accessarproval of a premarket approval application (PMA). You may, (Act) that do not require upper to the general controls provisions of the Act. The general thorefore, maniot the act include requirements for annual registration, listing of devices, Controls provisions or the rev labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can thay be subject to sacherital resillations, Title 21, Parts 800 to 898. In addition, FDA may ou found in the Ood ucements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease oc devisou that I Dr Unesan that your device complies with other requirements of the Act that I Dri has intace a and regulations administered by other Federal agencies. You must or any 1 odolar batates and registements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 & Mr. Don Selvey

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you've begal finding of substantial equivalence of your device to a legally premated notification: "The classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you desire specific acrison. I for in vitro diagnostic devices), please contact the Office of additionally 21 31 31 594-4659. Additionally, for questions on the promotion and advertising of Compliance at (301) 591-1007 Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Outler general manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Walker, MD

Image /page/4/Picture/5 description: The image contains a handwritten symbol or character. It appears to be a cursive letter or a stylized mark, possibly part of a signature or logo. The character is formed with a loop at the top and a downward stroke, giving it a flowing appearance.

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number (if-known):

.. . . . . . . . . . . . . . . .

KO12631

Device Name: Alliance Medical Corporation Reprocessed Soft Tissue Ablators

Indications for Use: Reprocessed soft tissue ablators are to be used in patients requiring Indications for Ose. "Reprocessed both assubtions as to to a common to resect, ablate and excise soft tissue.

ManufacturerDescriptionModel NumberManufacturerDescriptionModel Number
AthroCare®Arthro Wand® Right AngleA 1325-01AthroCare®Arthro Wand® Small JointA 1115-01
AthroCare®Arthro Wand® Right AngleA 1330-01AthroCare®Arthro Wand® BisectorA 4330-01
AthroCare®Arthro Wand® Right AngleA 1335-01AthroCare®Arthro Wand® StraightA 1125-01
AthroCare®Arthro Wand® Right AngleA 1336-01AthroCare®Arthro Wand® StraightA 1130-01
AthroCare®Arthro Wand® Right AngleA 1345-01AthroCare®Arthro Wand® AngleA 1225-01
AthroCare®Arthro Wand® DomeA 3625-01AthroCare®Arthro Wand® AngleA 1230-01
AthroCare®Arthro Wand® DomeA 3525-01Mitek®VAPR™ Thermal Side Effect Electrode (3.5mm)225101
AthroCare®Arthro Wand® DomeA 3630-01Mitek®VAPR™ Thermal Reversed Angled Side Effect Electrode (3.5mm)225112
AthroCare®Arthro Wand® DomeA 3430-01Mitek®VAPR™ Thermal Angled End Effect Electrode (3.5mm)225104
AthroCare®Arthro Wand® DomeA 3530-01Mitek®VAPR™ Side Effect Electrode (3.5mm)225301
AthroCare®Arthro Wand® BevelA 2630-01Mitek®VAPR™ Angled Side Effect Electrode (21°, 3.5mm)225302
AthroCare®Arthro Wand® BevelA 2430-01Mitek®VAPR™ End Effect Electrode (3.5mm)225303
AthroCare®Arthro Wand® BevelA 2530-01Mitek®VAPR™ Angled End Effect Electrode (21°, 3.5mm)225304
AthroCare®Arthro Wand® Small JointA 2723-01Mitek®VAPR™ 90° Hook Electrode (3.5mm)225305
AthroCare®Arthro Wand® Small JointA 2823-01Mitek®VAPR™ Flexible Side Effect Electrode (3.5mm)225312

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Walker.

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K012631

Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________

Prescription Use (per 21 CFR 801.109)

CONFIDENTIAL

Alliance Medical Corporation Reprocessed Soft Tissue Ablators Traditional 510(k)

or

65

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.