K Number
K012590

Validate with FDA (Live)

Manufacturer
Date Cleared
2001-09-07

(28 days)

Product Code
Regulation Number
876.5010
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cordis PALMAZ® GENESIS™ Transhepatic Biliary Stent on OPTA™ PRO .035" Delivery System is indicated for use in the palliation of malignant neoplasms in the biliary tree.

Device Description

The PALMAZ GENESIS Transhepatic Biliary Stent is a balloon-expandable, laser cut stent made from 316L Stainless Steel tubing. The stent is sold mounted on a Cordis percutaneous transluminal angioplasty (PTA) balloon catheter. The stent is provided in nominal, unexpanded stent lengths from 19 to 79 mm. The stent is designed for expansion to diameters from 8 to 10 mm, depending on the diameter of the associated balloon upon which it is mounted.

The Cordis OPTA PRO PTA balloon catheter, formerly known as the OPTA LP PTA Balloon Catheter, is used as a delivery system for the PALMAZ GENESIS Transhepatic Biliary Stent. The delivery system utilizes an over-the-wire design and is a catheter with a distal DURALYN™ balloon and a proximal Y-connector. Two radiopaque marker bands aid in stent placement. The injectate lumen (marked "THRU") is used to inject contrast medium via hand injection, and to track the catheter over a guide wire. The inflation lumen (marked "BALLOON") is used to inflate and deflate the balloon. The nominal balloon size is printed on the Y-connector.

A metal introducer tube is included in the packaging.

The PALMAZ GENESIS Transhepatic Biliary stent on OPTA PRO .035" Delivery System is provided sterile (via ethylene oxide) and is intended for single use only.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for the "Cordis PALMAZ® GENESIS™ Transhepatic Biliary Stent on OPTA™ PRO .035" Delivery System". This type of submission is for medical devices and focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel efficacy or performance through extensive clinical trials. Therefore, the information typically requested for AI/device performance studies (like acceptance criteria, sample sizes for test/training sets, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies) is generally not included or applicable in this context.

Here's an breakdown based on the provided text, addressing your questions to the extent possible given the nature of the document:

1. Table of Acceptance Criteria and Reported Device Performance

Feature/TestAcceptance CriteriaReported Device Performance
Overall Safety and EffectivenessSubstantial equivalence to predicate devices (K012090, K012056, K971448, K981407, K991028).Demonstrated via data collected from non-clinical design verification tests and analyses. (Specific performance metrics are not detailed in this summary.)
Intended UseDevice is able to be used for the palliation of malignant neoplasms in the biliary tree.The device is indicated for this use, implying it meets the necessary functional requirements.
Vascular System Use (Limitations)Not established (as per FDA determination).N/A - This is a limitation, not a performance criterion.

Explanation: In a 510(k) submission for a device like this stent and delivery system, the "acceptance criteria" are typically related to meeting design specifications, demonstrating biocompatibility, sterilization efficacy, and functional integrity through non-clinical bench testing. The primary "performance" is its ability to meet the stated intended use and its substantial equivalence to predicate devices. The document explicitly states "Performance standards have not been established for this device by the FDA under section 514 of the Food, Drug and Cosmetic Act."

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not specified. The document states "data collected from non-clinical design verification tests and analyses." This typically refers to bench testing of device components, and the "sample size" would relate to the number of devices or components tested to verify specifications (e.g., stent expansion diameter, balloon integrity, delivery system trackability, etc.). These details are not provided in this summary.
  • Data Provenance: The data is from "non-clinical design verification tests and analyses," implying laboratory and engineering testing. No mention of human clinical data or its geographical origin is made.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This is a medical device 510(k) submission, not an AI performance study requiring expert adjudication for ground truth. Technical specifications and functional tests would be verified against engineering standards and design requirements.

4. Adjudication method for the test set

  • Not Applicable. See point 3. Testing would follow standard engineering and quality control procedures, not expert adjudication in the context of diagnostic performance.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

  • No. An MRMC study is relevant for diagnostic devices, particularly those involving image interpretation with AI assistance. This document describes a physical medical device (stent and delivery system) and its mechanical/functional performance, not a diagnostic tool requiring human reader comparison.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This refers to AI algorithm performance. The device described is a physical medical implant and delivery system, not an algorithm.

7. The type of ground truth used

  • For the non-clinical design verification, the "ground truth" would be established by the device's design specifications and engineering requirements (e.g., material properties, dimensional tolerances, force requirements, fluid flow rates, etc.). These are verified through various physical and chemical tests.

8. The sample size for the training set

  • Not Applicable. This refers to AI/machine learning. The device described does not involve a "training set" in this context.

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

{0}------------------------------------------------

Page 1/2

SEP - 7 2001

Cordis Europa N.V., a Johnson & Johnson company

Attachment 1

Special 510(k) Premarket Notification

KO1259D

Summary of Safety and Effectiveness

Submitter:Bert RoossienManager Regulatory AffairsCordis Europa, N.V.Oosteinde 8NL-9301 LJ RodenThe Netherlands
Tel: +31 - (5050) 22321Fax: +31 - (5050) 22456e-mail: broossie@crdnl.jnj.com
Contact person:Chuck RyanManager, Regulatory AffairsCordis Corporation7 Powder Horn DriveWarren, New Jersey 07059USATel: 908.412.7446Fax: 908.412.3915e-mail: cryan@crdus.jnj.com
Date prepared09 August 2001.
General provisionsTrade name:Cordis PALMAZ® GENESIS™ Transhepatic Biliary Stenton OPTA™ PRO .035" Delivery System.
Common Name:Biliary stent and accessories
Classification Name:21 CFR 876.5010 Biliary Catheter and accessories.
Device ClassificationClass II.
Name of predicate device(s)Cordis PALMAZ® GENESIS™ Transhepatic Biliary Stent (ref. K012090)Cordis PALMAZ® GENESIS™ Transhepatic Biliary Stent on SLALOM .018 Delivery System (ref. K012056)Cordis OPTA ™ PRO (formerly known as OPTA LP) PTA Balloon Catheter (ref. K971448, K981407, K991028)

{1}------------------------------------------------

Special 510(k) Premarket Notification

KD12590

ಬ್ಗ ೭

Cordis Europa N.V.,a Johnson & Johnson companySpecial 510(k) Premarket Notification
PerformancestandardsPerformance standards have not been established for this device by the FDA undersection 514 of the Food, Drug and Cosmetic Act.
DeviceDescriptionThe PALMAZ GENESIS Transhepatic Biliary Stent is a balloon-expandable, laser cutstent made from 316L Stainless Steel tubing. The stent is sold mounted on a Cordispercutaneous transluminal angioplasty (PTA) balloon catheter. The stent is providedin nominal, unexpanded stent lengths from 19 to 79 mm. The stent is designed forexpansion to diameters from 8 to 10 mm, depending on the diameter of theassociated balloon upon which it is mounted.The Cordis OPTA PRO PTA balloon catheter, formerly known as the OPTA LP PTABalloon Catheter, is used as a delivery system for the PALMAZ GENESISTranshepatic Biliary Stent. The delivery system utilizes an over-the-wire design andis a catheter with a distal DURALYN™ balloon and a proximal Y-connector. Tworadiopaque marker bands aid in stent placement. The injectate lumen (marked"THRU") is used to inject contrast medium via hand injection, and to track thecatheter over a guide wire. The inflation lumen (marked "BALLOON") is used toinflate and deflate the balloon. The nominal balloon size is printed on the Y-connector.A metal introducer tube is included in the packaging.The PALMAZ GENESIS Transhepatic Biliary stent on OPTA PRO .035" DeliverySystem is provided sterile (via ethylene oxide) and is intended for single use only.
Intended UseThe Cordis PALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035"Delivery System is indicated for use in the palliation of malignant neoplasms in thebiliary tree.
PerformanceDataThe safety and effectiveness of the Cordis PALMAZ GENESIS Transhepatic BiliaryStent on OPTA PRO .035" Delivery System have been demonstrated via datacollected from non-clinical design verification tests and analyses.
A statement of substantial equivalence to another product is required by 21CFR 807.87 and relates only to whether the present product can be marketedwithout prior reclassification or clinical approval. The present submission istherefore not related to the coverage of any patent and is not to be interpretedas an admission or used as evidence in a patent infringement lawsuit. As thecommissioner of the FDA stated: "A determination of substantial equivalenceunder the Fedral Food, Drug and Cosmetic Act releted to the fact that theproduct can be lawfully marketed without pre-market approval orreclassification. The determination is not intended to have any bearingwhatsoever on the resolution of patent infringement suits." 42 Federal Register42, 50 et seq. (1977).

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Inside the circle is a stylized image of three human profiles facing right, with flowing lines above them that resemble a bird's wings. The logo is black and white.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

7 2001 SEP

Cordis Corporation Cordis Europa, N.V. c/o Mr. Charles J. Ryan, RAC Manager, Regulatory Affairs 7 Powder Horn Drive WARREN NJ 07059

K012590 Re:

Cordis PALMAZ® GENESIS™ Transhepatic Biliary Stent on OPTATM PRO .035" Delivery System Regulatory Class: II 21 CFR 876.5010 Product Code: 78 FGE Dated: August 9, 2001 Received: August 10, 2001

Dear Mr. Ryan:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:

The safety and effectiveness of this device for use in the vascular system have not been established.

Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.

{3}------------------------------------------------

Page 2 - Mr. Charles J. Ryan, RAC

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of R rat 007), advertise (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification if the limitation statement above is added to your labeling, as described.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Philip Phillips

Statland, M.D., Ph.D. Bernard E. Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure .

{4}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K012590

Device Name: Cordis PALMAZ® GENESIS™ Transhepatic Biliary Stent on OPTA™ PRO .035" Delivery System

FDA's Statement of the Indications For Use for device:

The Cordis PALMAZ® GENESIS™ Transhepatic Biliary Stent on OPTA™ PRO .035" Delivery System is indicated for the palliation of malignant neoplasms in the biliary tree.

Nancy C broglon

(Division Sign-Off)
Division of Reproductive, Abdominal,

and Radmiogical Devices 510(k) Number _

OR Prescription Use v (Per 21 CFR 801.109)

Over-The-Counter Use

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.