K Number
K012411
Device Name
HP-ONE
Manufacturer
Date Cleared
2001-09-25

(57 days)

Product Code
Regulation Number
866.3110
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HP-ONE is for in-vitro diagnostic use only. It is intended to detect the urease enzyme of Helicobacter pylori and provide a presumptive diagnosis of H. pylori gastritis.

Device Description

HP-ONE detects the urease enzyme for the presumptive identification of Helicobacter pylori in gastric mucosal biopsies. It is intended for in-vitro diagnostic use only. A gastric mucosal endoscopic biopsy is placed in the HP-ONE tray well and a reagent containing urea, hydrogen peroxide, a pH dye indicator, and an acidic buffer is placed over the biopsy specimen. A bubbling reaction occurs as catalase breaks down the peroxide. If H. pylori is present, the urease in H. pylori converts the urea to ammonia which raises the pH and changes the color of the reagent, indicating a positive test.

AI/ML Overview

This document describes the HP-ONE device, its intended use, and performance characteristics. However, it does not explicitly state "acceptance criteria" as a separate predefined standard. Instead, the "acceptance criteria" are implicitly met by the reported performance characteristics that were considered sufficient for substantial equivalence to a predicate device.

Here's the information requested, organized based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance
Relative SensitivityPerformance comparable to HP-Fast (predicate device). Specific numerical criteria are not stated, but the achieved sensitivity was deemed acceptable for substantial equivalence.92.3%
Relative SpecificityPerformance comparable to HP-Fast (predicate device). Specific numerical criteria are not stated, but the achieved specificity was deemed acceptable for substantial equivalence.100%
Positive Predictive ValueNot explicitly stated as acceptance criteria.100%
Negative Predictive ValueNot explicitly stated as acceptance criteria.91.23%

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 117 patients
  • Data Provenance: The text does not specify the country of origin of the data. It states, "In the clinical study of 117 patients with possible peptic ulcer disease..." This suggests the data was from a clinical study, but whether it was retrospective or prospective is not explicitly stated. Given it's a study for regulatory submission, it is typically a prospective or carefully selected retrospective cohort.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth was established by histological criteria (presence of H. pylori organisms upon staining). The text does not specify the number or qualifications of the experts (e.g., pathologists) who performed the histological assessments.

4. Adjudication Method for the Test Set

The text does not describe an adjudication method for the test set. The ground truth was based on histological diagnosis.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The document explicitly states, "HP-ONE agreed with the positive histological diagnosis in 60 cases (92.3%) and agreed with the negative diagnosis in all 52 (100%) of the cases." This indicates a standalone performance evaluation against a reference standard, not a comparative effectiveness study involving human readers with and without AI assistance.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

Yes. The study presented is a standalone performance evaluation of the HP-ONE device against histological diagnosis as the ground truth. HP-ONE itself is a diagnostic test kit that produces a result; it's not an "algorithm" in the typical sense of AI, but rather a chemical reaction-based diagnostic tool. The performance metrics (92.3% sensitivity, 100% specificity) reflect the device's inherent capability.

7. Type of Ground Truth Used

The ground truth used was histological criteria (presence of H. pylori organisms upon staining).

8. Sample Size for the Training Set

The document does not describe a separate "training set." The clinical study of 117 patients appears to be the primary dataset reported for performance evaluation. For a chemical diagnostic test like HP-ONE, the "training" typically refers to the R&D and optimization process for the reagent formulation, not an algorithmic training process with a distinct dataset.

9. How the Ground Truth for the Training Set Was Established

Since no separate training set is described in the provided text, this question is not applicable. The device's formulation would have been developed and optimized based on scientific principles and internal testing, but without a formally described "training set" with established ground truth in the context of this 510(k) summary.

{0}------------------------------------------------

HP-ONE™ 510k SUMMARY

HP-ONE is a trademark of GI Supply division of Chek-Med Systems Inc. Contact Name: Frank Carter

Address:200 Grandview Ave. Camp Hill, PA 17011
Phone #:717-761-1170
Fax #:717-761-0216

PRODUCT NAME: HP-ONE™ 1.

CLASSIFICATION NAME: Campylobacter pylori test - CODE 83LYR 2.

  • INTENDED USE: 3.
    HP-ONE detects the urease enzyme for the presumptive identification of Helicobacter pylori in gastric mucosal biopsies. It is intended for in-vitro diagnostic use only.

SUMMARY: 3.

Helicobacter pylori has now been shown to be the causative agent in most instances of acute type B gastritis (1). And, it is now known that, in the absence of aspirin, nonsteroidal anti-inflammatory drugs or hyper-secretory states, the bacteria is directly related to peptic ulcer disease such as duodenal and benign gastric ulcers (2). There is epidemiological data linking the presence of the H. pylori with gastric cancer and the lowgrade mucosal associated lymphoid tissue (MALT) lymphoma (3,4). The eradication of H. pylor gastritis by antibiotics has been shown to cure peptic ulcers and prevent recurrence (4,5).

PRINCIPLE: 4.

A gastric mucosal endoscopic biopsy is placed in the HP-ONE tray well and a reagent containing urea, hydrogen peroxide, a pH dye indicator, and an acidic buffer is placed over the biopsy specimen. A bubbling reaction occurs as catalase breaks down the peroxide.

If H. pylori is present, the urease in H. pylori converts the urea to ammonia which raises the pH and changes the color of the reagent, indicating a positive test.

urease (NH2)CO + H2O ------------>NH3 + HCO2 ammonia urea

5. STORAGE/STABILITY:

The HP-ONE reagent is stable and provides accurate results for up to 12 months if refrigerated at 2-8°C. Alternatively, the reagent can be stored at room temperature (18-24°C) but the shelf life will be shortened to 3 months. If stored at room temperature, mark the date of receipt on the label and do not use beyond 3 months. Do not expose to heat or bright light. Do not use the test if the color is not a yellow or yellow-orange color or if the test is past the expiration date. Each HP-ONE reagent bottle contains an expiration date and lot number to permit tracking.

HP-1 SSE rev. 7/01 p.1

{1}------------------------------------------------

  • SUBSTANTIAL EQUIVALENCE AND DIFFERENCES TO PREDICATE DEVICE: The principle of HP-ONE is substantially equivalent to HP-Fast another product marketed by GI-Supply. They share the same operating principal, but HP-ONE has technical differences from HP-Fast. HP-ONE is a liquid rather than an agar gel and also has a lower pH than HP-Fast. HP-One also contains hydrogen peroxide that acts as a preservative and helps to expedite the urease reaction when H. pylori is present.

SUMMARY OF HP-ONE PERFORMANCE CHARACTERISTICS: 7.

In the clinical study of 117 patients with possible peptic ulcer disease, 65 demonstrated positive histological criteria (presence of H. pylori organisms upon staining) for Helicobacter pylori and 52 did not demonstrate the criteria. HP-ONE agreed with the positive histological diagnosis in 60 cases (92.3%) and agreed with the negative diagnosis in all 52 (100%) of the cases. This data provides a relative sensitivity of 92.3% and a relative specificity of 100%.

True +False +
HP-One Positives600
HP-One NegativesFalse -5True -52

HP-One Negatives

Histology Positives = 65 Histology Negatives = 52

HP-One Sensitivity = 92.3% HP-One Specificity = 100%

Positive Predictive Value = 100% Negative Predictive Value = 91.23%

HP-One correctly diagnosed 66.67% of the positives within 5 minutes and 83.3% of the positives by 15 minutes. There were no false positives when diagnosed within the one-hour final reading of the test.

  • MEDICAL DEVICE REPORTS / ADVERSE REACTIONS ENCOUNTERED: 8. To date, no MDR's have been filed and no patient adverse reactions have been encountered.
    HP-1 SSE rev. 7/01 p.2

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, depicted in a minimalist, abstract design.

Public Health Service

SEP 2 5 2001

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Frank Carter President GI Supply, Division of Chek-Med Systems Inc. 200 Grandview Avenue Camp Hill, PA 17011

Re: K012411

Trade/Device Name: HP-ONE Rapid Urea Assay Regulation Number: 21 CFR 866.3110 Regulation Name: Campylobacter Pylori Regulatory Class: I Product Code: LYR Dated: July 27, 2001 Received: July 30, 2001

Dear Mr. Carter:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and v additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Intended Use Certification Section 8.2

510k INDICATION FOR USE STATEMENT

Ko layi ! 510(k) Number (if known): __

Device Name: HP-ONE

Applicant Name: GI Supply division of Chek-Med Systems Inc.

Indication for Use:

HP-ONE is for in-vitro diagnostic use only. It is intended to detect the urease enzyme of Helicobacter pylori and provide a presumptive diagnosis of H. pylori gastritis.

7-26-01

Date:

GI Supply/Chek-Med Signature: c

Frank Carter - CEO

Luedale U. Poole

(Division Sign-Off) Division of Clinical scoratory Devices

510(k) Number K012411

Prescription use

§ 866.3110

Campylobacter fetus serological reagents.(a)
Identification. Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identifyCampylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases.Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.(b)
Classification. Class I (general controls).