K Number
K012306
Date Cleared
2001-09-21

(60 days)

Product Code
Regulation Number
874.3310
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Audioscan® Verifit™ is a hearing aid analysis system that is an electronic reference device used by manufacturers and dispensers of hearing aids intended to calibrate and assess the electroacoustic frequency and sound intensity characteristics emanating from a hearing aid coupled to an ear simulator in a test chamber or a hearing aid when worn by a patient.

Device Description

The Audioscan® Verifit™ is a hearing aid analysis system. The system is integrated with a hearing test chamber physically separated from the display/control unit. Signal processing and analysis is carried out digitally, with the exception of the microphone preamplifiers and output amplifiers. It includes a fluorescent backlit active color display, thermal line printer, and various connectors for peripherals and microphones. It has two stimulus channels and two measurement channels. The hearing aid test chamber has a working space of 8" x 5" x 1.5" (approx.), two 2" x 3" Independent speakers, and induction coils for Telephone Magnetic Field Simulator and a 20 cm diameter test loop. It also includes a battery simulator. The real-ear measurement component has two 2" x 3" ducted ports speakers and a Silicone 1.0mm diameter x 75mm probe microphone tube.

AI/ML Overview

Acceptance Criteria and Device Performance for Audioscan® Verifit™ Model VF-1

This document outlines the acceptance criteria and the reported performance of the Audioscan® Verifit™ Model VF-1, a hearing aid analysis system. The information is derived from the provided 510(k) summary (K012306).

1. Table of Acceptance Criteria and Reported Device Performance

The Audioscan® Verifit™ Model VF-1 demonstrates compliance with established industry standards and its predicate device (Audioscan® Real Ear Measurement System RMS00, K884046/A) across various performance characteristics. The acceptance criteria for this device are implicitly derived from these standards and the performance of the predicate device.

Key Performance Standards Conformance:

  • ANSI S3.46-1997, Methods of Measurement of Real-Ear Performance Characteristics of Hearing Aids
  • ANSI S3.22-1996, Specification of Hearing Aid Characteristics
  • IEC 60601-1, EN60601-1, UL-2601, UL 544 (Safety Standards)
  • Food and Drug Administration radiation safety performance standards

The following table compares key characteristics between the Audioscan® Verifit™ Model VF-1 and its predicate device, Audioscan® Real Ear Measurement System RMS00. The "Acceptance Criteria" are understood to be meeting or exceeding the predicate device's performance, as well as conforming to the cited ANSI, IEC, and UL standards.

CharacteristicPredicate Device (MODEL RMS00)Audioscan® Verifit™ Model VF-1 Performance (Reported)Acceptance Criteria (Implicit)
GENERAL
Overall dimensions15.5" x 13.2" x 6.2"Rear main unit 14.5" x 16" x 6.5"; Test chamber 14" x 14.4" x 4"Device must meet or demonstrate equivalent/improved functionality within reasonable physical configuration changes.
Weight15 lbs16 lbsMinimal increase in weight is acceptable if functionality is improved.
Display typefluorescent backlit blue on whitefluorescent backlit active colorImproved display technology is an enhancement, meeting or exceeding predicate.
Display size6.4" diagonal12.1" diagonalLarger display is an enhancement, meeting or exceeding predicate.
Display pixels (resolution)320 x 200 (VGA)800 x 600 (SVGA)Higher resolution is an enhancement, meeting or exceeding predicate.
Printer typethermal line printerSAME (thermal line printer)Equivalent performance to predicate.
Printer resolution200 dots per inchSAME (200 dots per inch)Equivalent performance to predicate.
Paper width3" (80 mm)SAME (3" (80 mm))Equivalent performance to predicate.
ConnectorsFewer (e.g., 1 external monitor, 1 RS232)More (e.g., mouse, QWERTY keyboard, ethernet, 2 USB, multiple audio)Improved connectivity is an enhancement.
Headphone monitor amplifier1 watt into 16 ohmsSAME (1 watt into 16 ohms)Equivalent performance to predicate.
Power amplifiers1 - 1.7 watts2 - 5 watts eachIncreased power amplifiers are an enhancement.
Stimulus channels12Increased stimulus channels are an enhancement.
Measurement channels2SAME (2)Equivalent performance to predicate.
HEARING AID TEST CHAMBER
Working space5.5" x 3.5" x 2"8" x 5" x 1.5" (approx.)Increased working space is an enhancement.
Speakers1 - 3.5" diameter2 - 2" x 3" IndependentIncreased speakers are an enhancement.
Induction coils1 (TMFS ANSI S3.22-1996, 1 - 20 cm diameter test loop)SAMEEquivalent performance to predicate, conforming to ANSI S3.22-1996.
Battery simulatorper ANSI S3.22 - 1996SAMEConformance to ANSI S3.22-1996.
Frequency range200 to 8000 HzSAMEConformance to established range.
Test stimulitonetone, broad-band noise, recorded speechAdded stimuli are an enhancement.
Test stimulus levels40 to 90 dB SPL in 5 dB stepsSAMEConformance to established levels.
Test stimulus levels (inductive)31.6 mA/m per ANSI S3.22 -1996SAMEEquivalent performance to predicate, conforming to ANSI S3.22-1996.
Test stimulus distortion<2% at 90 dB SPL; <0.5% at 70 dB SPLSAMEMaintaining low distortion, equivalent to predicate.
Test stimulus accuracy at reference microphone (tones 200 - 2000 Hz)± 1.5 dB SPLSAMEConformance to established accuracy.
Test stimulus accuracy at reference microphone (tones 2000 - 8000 Hz)± 2.5 dB SPLSAMEConformance to established accuracy.
Equalization methodpressure methodSAMEEquivalent performance to predicate.
Analysis frequencies per octave12SAMEEquivalent performance to predicate.
Analysis filter bandwidth1/7 octave1/12 octaveImproved (finer) filter bandwidth is an enhancement.
Measurement accuracy at 1 kHz± 1 dBSAMEConformance to established accuracy.
Measurement accuracy re 1 kHz± 1 dB (200-5000 Hz); ± 2.5 dB (5000-8000 Hz)SAMEConformance to established accuracy.
Measurement range30 - 140 dB SPLSAMEConformance to established range.
Harmonic distortion measurement2nd plus 3rd2nd and 3rd or 2nd plus 3rdMore comprehensive measurement options are an enhancement.
Harmonic distortion range200 to 2670 Hz200 to 4000 HzWider range for harmonic distortion measurement is an enhancement.
Harmonic distortion accuracy± 1%SAMEConformance to established accuracy.
Battery drain range0-20 mASAMEConformance to established range.
Battery drain accuracy± .01 mASAMEConformance to established accuracy.
ANSI S3.22-1996 tests availableStandard set (e.g., OSPL90, Full-on Gain)SAME, plus "Coupler SPL - Vertical" and "Magnetic Field - Test Loop Sensitivity"Device must perform existing tests equivalently and offer added features.
Other tests availableStandard set (e.g., coupler SPL vs frequency)SAME, plus "coupler gain vs frequency" and "distortion vs frequency"Device must perform existing tests equivalently and offer added features.
REAL-EAR MEASUREMENT
Speakers1 - 3.5" diameter2 - 2" x 3" ducted portsImproved speaker configuration is an enhancement.
Probe microphone tubeSilicone 1.0mm diameter x 75mmSAMEEquivalent performance to predicate.
Frequency range200 to 8000 HzSAMEConformance to established range.
Test stimulifrequency-modulated tone, tone burst, amplitude-modulated tonefrequency-modulated tone, tone burst, broad-band noise, recorded speechAdded stimuli are an enhancement.
Frequency modulationtriangular ±5% at 36 HzSAMEEquivalent performance to predicate.
Test stimulus levels at reference microphone for tones at 1m40 to 85 dB SPL in 5 dB steps40 to 90 dB SPL in 5 dB stepsWider range for stimulus levels is an enhancement.
Test stimulus accuracy at reference microphone (tones 200 - 2000 Hz)± 1.5 dB SPLSAMEConformance to established accuracy.
Test stimulus accuracy at reference microphone (tones 2000 - 8000 Hz)± 2.5 dB SPLSAMEConformance to established accuracy.
Equalization methodpressure methodSAMEEquivalent performance to predicate.
Analysis frequencies per octave12SAMEEquivalent performance to predicate.
Analysis filter bandwidth1/7 octave1/12 octaveImproved (finer) filter bandwidth is an enhancement.
Measurement accuracy at 1 kHz± 1 dBSAMEConformance to established accuracy.
Measurement accuracy re 1 kHz± 1 dB (200-5000 Hz); ± 2.5 dB (5000-8000 Hz)SAMEConformance to established accuracy.
Measurement range20-135 dB SPL (200-2500 Hz); 30-140 dB SPL (2500-8000 Hz)SAMEConformance to established range.
ANSI S3.46-1997 tests availableStandard set (e.g., Real-Ear Unaided Response)SAMEDevice must perform existing ANSI S3.46-1997 tests equivalently.
Other tests availableStandard set (e.g., Real-ear harmonic distortion)SAMEDevice must perform existing tests equivalently.
Prescriptive fitting methods availableStandard set (e.g., NAL-R, Pogo II)SAMEDevice must incorporate existing fitting methods equivalently.

2. Sample size used for the test set and the data provenance

The document provided (510(k) Summary K012306) does not explicitly detail a clinical test set with a specific sample size for human participants or a defined dataset of hearing aid measurements. Instead, the device's performance is demonstrated through:

  • Conformance to internationally recognized standards: ANSI S3.46-1997, ANSI S3.22-1996, IEC 60601-1, EN60601-1, UL-2601, UL 544, and FDA radiation safety standards. This implies comprehensive testing against these standards, which themselves define test methods and parameters.
  • Substantial Equivalence to a Predicate Device: The primary comparison is against the Audioscan® Real Ear Measurement System RMS00 (K884046/A). This involves demonstrating that the new device performs equivalently or better than the predicate device across all relevant technical specifications. The comparison table above illustrates this.
  • Software Validation: "development, verification, validation, and testing procedures are in accordance with FDA guidances for devices containing software." This refers to internal software testing, not a clinical test set with patient data.

Therefore, it is not possible to specify a sample size, country of origin, or whether the "data" was retrospective or prospective in the context of a clinical test set from this document. The validation relies on engineering testing, compliance with standards, and comparative analysis with a predicate device's established performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Given that the performance assessment relies on conformance to technical standards and comparison with a predicate device, as opposed to a clinical study with human interpretation, the concept of "experts establishing ground truth for a test set" with a specified number and qualifications (e.g., radiologists with 10 years of experience) is not applicable in the context of this device's submission.

The "ground truth" here is essentially the established specifications and accuracy requirements set forth by the ANSI, IEC, and UL standards, and the known performance baseline of the predicate device. The experts involved would be the engineers and quality assurance personnel who performed the testing and verified compliance with these standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

As there is no described clinical "test set" requiring interpretation or consensus among experts, no adjudication method (e.g., 2+1, 3+1) is mentioned or relevant in this 510(k) summary. The evaluation focuses on objective measurements against defined performance criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This device is a "Hearing-Aid Calibrator & Analysis System" designed for the objective measurement and calibration of hearing aids, not a diagnostic AI system assisting human readers. Therefore, an MRMC comparative effectiveness study evaluating human reader improvement with AI assistance is not applicable and was not performed or mentioned in this submission.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The Audioscan® Verifit™ Model VF-1 is inherently a standalone device in terms of its ability to perform measurements and analysis based on its internal algorithms and hardware. Its primary function is to objectively measure electroacoustic characteristics. The "software algorithms are essentially equivalent for both the Audioscan® Verifit™ and predicate devices," and the device performs signal processing and analysis digitally.

Therefore, yes, its performance is de facto standalone in its measurement capabilities. However, it's crucial to understand that this is not an "AI algorithm only" device in the modern sense of a black-box AI standalone performance study. It is a measurement instrument that uses established algorithms to perform its functions. The device's "performance" described is its ability to accurately and consistently perform these measurements against known physical standards, rather than an "algorithm-only" performance compared to human interpretation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the Audioscan® Verifit™ Model VF-1 is based on:

  • Established engineering and electroacoustic standards: Primarily ANSI S3.46-1997 and ANSI S3.22-1996. These standards define how hearing aid characteristics should be measured and what acceptable performance ranges are.
  • Physical calibration references: The device measures physical properties (frequency, sound intensity, harmonic distortion, battery drain) against calibrated references.
  • Performance of the predicate device: The Audioscan® Real Ear Measurement System RMS00 (K884046/A) serves as a benchmark for equivalence.

Therefore, the ground truth is objective measurements against recognized standards and a known predicate device's performance, not expert consensus, pathology, or outcomes data from a clinical trial.

8. The sample size for the training set

The concept of a "training set" in the context of machine learning for an AI-powered device is not applicable here. The Audioscan® Verifit™ is a measurement and calibration instrument based on established scientific principles and algorithms, not a machine learning model that requires a training set of data. Its performance is validated against physical standards and a predicate device.

9. How the ground truth for the training set was established

Since the concept of a "training set" as understood in machine learning is not applicable to this device, the question of how its ground truth was established is also not applicable. The device's foundational principles are rooted in electroacoustic engineering and adherence to international measurement standards.

{0}------------------------------------------------

Edi E

sion incorporated

41 Byron Avenue, Dorchester, Ontario NOL 1G0 Tel: (519) 268-3313 Fax: (519) 268-3256

SEP 2 1 2001

K012306

SECTION E - 510(K) SUMMARY OF SAFETY & EFFECTIVENESS

Audioscan® Verifit™ Model VF-1

Classification Name:Hearing-Aid Calibrator & Analysis System
Common/Usual Name:Hearing Aid Analyzer
Proprietary Name:Audioscan ® Verifit™ Model VF-1
Establishment Registration:8022229
Medical Specialty:Panel 77, Ear, Nose, & Throat
Product Code:ETW
Regulation Number:21 CFR 874.3310
Device Classification:II
Contact Person:William A Cole, President

Indications for Use:

The Audioscan® Verifit™ is a hearing aid analysis system that is an electronic reference device used by manufacturers and dispensers of hearing aids intended to calibrate and assess the electroacoustic frequency and sound intensity characteristics emanating from a hearing aid coupled to an ear simulator in a test chamber or a hearing aid when worn by a patient.

Performance Standards Conformance:

    1. The VF-1 meets or exceeds the equipment requirements of the following standards: American National Standards Institute, Methods of Measurement of Real-Far Performance Characteristics of Hearing Aids, (ANSI S3.46-1997), American National Standards Institute, New York, 1997, and, American National Standards Institute, Specification of Hearing Ald Characteristics, (ANSI S3.22-1996), American National Standards Institute, New York, 1996.
    1. The VF-1 meets or exceeds the applicable requirements of the following standards for safety: IEC 60601-1, EN60601-1, UL-2601 and UL 544.

{1}------------------------------------------------

Etymonic Design Incorporated Applied Microelectronics

  • The VF-1 meets or exceeds the Food and Drug Administration radiation safety 3. performance standards prescribed in the Code of Federal Regulations, Title 21, Chapter I. Subchapter J.

Substantial Equivalence:

Determination of substantial equivalence of the Audioscan® Verifit™ VF-1 instrument to the currently marketed Audioscan® Real Ear Measurement System RMS00, 510(k) No. K884046/A, February 13, 1989, is based on the following safety and effectiveness information:

  • The Audioscan® Verifit™ device is manufactured and delivered completely . assembled to the health care professional using materials and techniques widely used by other manufacturers of devices as with the predicate device.
  • Patient safety is preserved in both the Audioscan® Verifit™ and predicate . devices. There is no direct electrical connection to the patient. The silicone probe microphone assembly that has contact with the patient's ear canal during the testing of the hearing aid, has no exposed metal parts and is classed as an F-Type Applied Part in UL2601-1. Thus, there is no possibility of electrical conductance.
  • Regarding physical configurations, the system is integrated with the hearing test ● chamber physically part of the display/control unit in the predicate device, whereas the Audioscan® Verifit™ hearing aid test chamber is physically separated from the display/control unit.
  • Software algorithms are essentially equivalent for both the Audioscan® Verifit™ t and predicate devices.
  • To establish the safety and effectiveness of the software that controls the . instrument, the development, verification, validation, and testing procedures are in accordance with FDA guidances for devices containing software 12, Quality Assurance procedures are adhered to and test results demonstrate that both system specifications and function requirements are met.

1 Guidance Principles of Software Validation, Draft Guidance Version 1.1, June 9, 1997, Center for Devices and Radiological Health, Food and Drug Administration

2 ODE Guidance for the Content of Premarket Submission for Medical Devices Containing Software, May 29, 1998, Center for Devices and Radiological Health, Food and Drug Administration

{2}------------------------------------------------

Etymonic Design Incorporated Applied Microelectronics

41 Byron Avenue, Dorchester, Ontario NoL. 1G0 Tel: (519) 268-3313 Fax: (519) 268-3256

  • The maximum sound level output of the Audioscan® Verifit™ speakers, if used . as intended, remains below +90 dB SPL throughout the frequency range of 200 to This demonstrates that the Audioscan® Verifit™ unit emits sounds 8000 Hz. well within OSHA permissible exposure limits of 8 hours at 90 dB, The maximum sound level output of the predicate device speakers is 85 dB SPL if used as intended.
  • The signal processing and analysis differ between the Audioscan® Verifit™ and . the predicate device. Signals are processed and analyzed by digitally controlled analog amplifiers, attenuators and filters in the predicate device. In the Audioscan® Verifit™, with the exception of the microphone preamplifiers and output amplifiers, all signal processing and analysis is carried out digitally.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "edi" in a bold, sans-serif font. The letters are composed of small squares, giving them a pixelated appearance. The letters are closely spaced together.

vmon n Incorporated

41 Byron Avenue, Dorchester, Ontanio NOL 1G0

CHARACTERISTICMODEL RM500VERIFIT™ MODELVF-1
GENERAL
Overall dimensions15.5" x 13.2" x 6.2"Rear main unit 14.5" x 16" x 6.5"Test chamber 14" x 14.4" x 4"
Weight15 lbs16 lbs
Display typefluorescent backlit blue onwhitefluorescent backlit active color
Display size6.4" diagonal12.1" diagonal
Display pixels (resolution)320 x 200 (VGA)800 x 600 (SVGA)
Printer typethermal line printerSAME
Printer resolution200 dots per inchSAME
Paper width3" (80 mm)SAME
Connectors1 - external monitor (15HD)1-RS232 (9D)1 - external speaker (RCA)1 - real-ear/coupler mic. (RJ11)1 - mouse (PS2-6 pin)1 - QWERTY keyboard (PS2-6pin)1 - external monitor (15HD)1 - parallel printer (25D)1 - RS232 serial (9D)1 - ethernet (RJ45)2 - USB2 - external speakers (RCA)2 - external speakers (1/4" mono))2 - real-ear mics. (3.5 mm st)1 - RECO transducer (3.5 mm st)1 - monitor headset (3.5 mm st)1 - test chamber ref. mic. (3.5 mm st)1 - coupler microphone (3.5 mm st)1 - battery substitute (3.5 mm st)
Headphone monitor amplifier1 watt into 16 ohmsSAME
Power amplifiers1 - 1.7 watts2 - 5 watts each
Stimulus channels12
Measurement channels2SAME
HEARING AID TEST CHAMBER
Working space5.5" x 3.5" x 2"8" x 5" x 1.5" (approx.)
Speakers1 - 3.5" diameter2 - 2" x 3" Independent
Induction coils1 - Telephone Magnetic FieldSimulator (TMFS ANSI S3.22 -1996)1 - 20 cm diameter test loopSAME
Battery simulatorper ANSI S3.22 - 1996SAME
Frequency range200 to 8000 HzSAME
Test stimulitonetone, broad-band noise, recordedspeech
CHARACTERISTICMODEL RM500VERIFIT™ MODELVF-1
Test stimulus levels40 to 90 dB SPL in 5 dB stepsSAME
Test stimulus levels (inductive)31.6 mA/m per ANSI S3.22 -1996SAME
Test stimulus distortion<2% at 90 dB SPL<0.5% at 70 dB SPLSAME
Test stimulus accuracy at referencemicrophone for tones (200 - 2000 Hz)± 1.5 dB SPLSAME
Test stimulus accuracy at referencemicrophone for tones (2000 -8000 Hz)± 2.5 dB SPLSAME
Equalization methodpressure methodSAME
Analysis frequencies per octave12SAME
Analysis filter bandwidth1/7 octave1/12 octave
Measurement accuracy at 1 kHz± 1 dBSAME
Measurement accuracy re 1 kHz± 1 dB (200 - 5000 Hz)± 2.5 dB (5000 - 8000 Hz)SAME
Measurement range30 - 140 dB SPLSAME
Harmonic distortion measurement2nd plus 3rd2nd and 3rd or 2nd plus 3rd
Harmonic distortion range200 to 2670 Hz200 to 4000 Hz
Harmonic distortion accuracy± 1%SAME
Battery drain range0-20 mASAME
Battery drain accuracy± .01 mASAME
ANSI S3.22-1996 tests available➤ OSPL90➤ Full-on Gain➤ Reference Test Gain➤ Frequency Response➤ Frequency Range➤ Maximum OSPL90➤ Harmonic Distortion➤ Attack & Release time➤ Equivalent Input Noise➤ Input/Output Curves➤ Coupler SPL - TelephoneSimulator➤ Simulated Telecoil➤ Sensitivity➤ Battery drainSAMEWith the added features:➤ Coupler SPL - Vertical➤ Magnetic Field➤ Test Loop Sensitivity
Other tests available➤ coupler SPL vs frequency➤ spectral analysis➤ manual measurement of➤ output, gain and distortionSAMEWith the added features:➤ coupler gain vs frequency➤ distortion vs frequency
REAL-EAR MEASUREMENT
Speakers1 - 3.5" diameter2 - 2" x 3" ducted ports
CHARACTERISTICMODEL RM500VERIFIT™ MODEL
VF-1
Probe microphone tubeSilicone 1.0mm diameter x75mmSAME
Frequency range200 to 8000 HzSAME
Test stimulifrequency-modulated tone, toneburst, amplitude-modulated tonefrequency-modulated tone, toneburst, broad-band noise, recordedspeech
Frequency modulationtriangular ±5% at 36 HzSAME
Test stimulus levels at referencemicrophone for tones at 1 m40 to 85 dB SPL in 5 dB steps40 to 90 dB SPL in 5 dB steps
Test stimulus accuracy at referencemicrophone for tones (200 - 2000 Hz)± 1.5 dB SPLSAME
Test stimulus accuracy at referencemicrophone for tones (2000 -8000 Hz)± 2.5 dB SPLSAME
Equalization methodpressure methodSAME
Analysis frequencies per octave12SAME
Analysis filter bandwidth1/7 octave1/12 octave
Measurement accuracy at 1 kHz± 1 dBSAME
Measurement accuracy re 1 kHz± 1 dB (200 - 5000 Hz)± 2.5 dB (5000 - 8000 Hz)SAME
Measurement range20-135 dB SPL (200 - 2500Hz)30 - 140 dB SPL (2500 - 8000Hz)SAME
ANSI S3,46-1997 tests available> Real-Ear UnaidedResponse> Real-Ear Aided Response> Real-Ear OccludedResponse> Real-Ear Insertion GainSAME
Other tests available> Real-ear harmonicdistortion> Real-ear spectral analysis> manual measurement ofoutput, gain and distortionSAME
Prescriptive fitting methods available> National Acoustics LabsRevised> Pogo II> Berger> Libby> Desired Sensation LevelSAME

Specification Comparison Table

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Etymonic Design Incorporated and Applied Microelectronics. The text is in black and is centered on the page. The words "Etymonic Design Incorporated" are on the top line, and the words "Applied Microelectronics" are on the bottom line. The logo is simple and professional.

.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Etymonic Design Incorporated, Applied Microelectronics. The logo consists of the letters "edi" in a stylized font on the left, followed by the company name in a more standard font on the right. The words "Applied Microelectronics" are written in a smaller font below the company name.

41 Byron Avenue, Dorchester, Ontario NOL 1G0 Tel: (519) 288-3313 Fax: (519) 268-3256

NOTES:

The couplers used in the RM500 hearing aid test chamber are also used in the Verifit™ VF-1. 1)

  1. The probe tubes used with the RM500 real-ear microphone assembly are also used with the VF-1.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or waves.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 1 2001

William A. Cole, President Etymonic Design Incorporated 41 Byron Avenue Dorchester, Ontario N0L 1G0

Re: K012306

Trade Name: Audioscan Verifit™ Regulation Number: 21 CFR 874.3310 Regulation Name: Hearing Aid Calibrator and Analysis System Regulatory Class: Class II Product Code: ETW Dated: July 20, 2001 Received: July 23, 2001

Dear Mr. Cole

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2 - William A. Cole, President

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

A. Ralph Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "edi" in a pixelated font. The letters are formed by small squares, giving it a retro, digital appearance. The font is simple and blocky, making the letters easily readable despite the pixelation.

Etymonic Design Incorporated Applied Microelectronics

41 Byron Avenue, Dorchester, Ontario N0L 1G0 Fax: (519) 268-3256 Tel: (519) 268-3313

SECTION D - INDICATIONS FOR USE STATEMENT

er/ 3 - 4/24/96

Applicant: Etymonic Design Incorporated

510(k) Number (if known):

Device Name: Audioscan® Verifit™

Indications For Use

The Audioscan® Verifit™ is a hearing aid analysis system that is an electronic referenc device used by manufacturers and dispensers of hearing aids intended to calibrate and assess the electroacoustic frequency and sound intensity characteristics emanating from a hearing aid coupled to an ear simulator in a test chamber or a hearing aid when worn by a patient.

Ama H. Kane Ph.D. 9/17/01
Division Sign-Off

Division Sign-Off) Division of Ophthalmic Devices 510(k) Number

Prescription Use
(Per 21 CFR 801.109)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Per 21 CFR 801.109)

(Optional Format 1-2-96)

§ 874.3310 Hearing aid calibrator and analysis system.

(a)
Identification. A hearing aid calibrator and analysis system is an electronic reference device intended to calibrate and assess the electroacoustic frequency and sound intensity characteristics emanating from a hearing aid, master hearing aid, group hearing aid or group auditory trainer. The device consists of an acoustic complex of known cavity volume, a sound level meter, a microphone, oscillators, frequency counters, microphone amplifiers, a distoration analyzer, a chart recorder, and a hearing aid test box.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.