K Number
K011727
Date Cleared
2002-05-02

(332 days)

Product Code
Regulation Number
870.1390
Panel
CV
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Used to create the primary puncture in the interatrial septum when passing an introducer and/or catheter through the septum from the right side of the heart to the left side.

Device Description

The Thomas Medical Transseptal Needle consists of a specifically curved distal portion to accommodate positioning in the cardiac anatomy when used in conjunction with a transseptal introducer. A stopcock is attached to the proximal end of the needle for air aspiration, fluid infusion, blood sampling and pressure monitoring. A pointer is included on the needle shield to show orientation of the curve.

AI/ML Overview

Here's the analysis of the provided text regarding the acceptance criteria and study for the Transseptal Needle:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific TestAcceptance CriteriaReported Device Performance
Mechanical IntegrityHub pull testPassed protocol requirementsAll samples passed
Mechanical IntegrityHub to Stopcock pull testPassed protocol requirementsAll samples passed
Mechanical IntegrityWire to Handle pull test (assuming "Wire" refers to needle)Passed protocol requirementsAll samples passed
Mechanical IntegrityNeedle to Hub torque forcePassed protocol requirementsAll samples passed
Intended UseCreate primary puncture in interatrial septumFunctionally equivalent to predicate devicesSubstantially equivalent to predicate devices

2. Sample Size Used for the Test Set and Data Provenance

The provided text states, "All samples passed the protocol qualification testing requirements." It does not specify an explicit sample size for each test. It also does not provide any information about the data provenance (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The text describes mechanical qualification testing, not a study involving expert assessment of a medical condition or images.

4. Adjudication Method for the Test Set

This information is not applicable as the tests conducted were mechanical qualification tests and did not involve expert adjudication of results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not performed. The submission is for a medical device (a needle) and not an AI or diagnostic imaging device that would typically undergo such a study. The “study” described is qualification testing to ensure the device’s mechanical integrity.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No, a standalone performance study was not done. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the qualification testing was the pre-defined protocol qualification testing requirements. These requirements would have defined acceptable ranges or thresholds for the mechanical properties being tested (e.g., minimum pull force, maximum torque). For the overall device, the "ground truth" for its regulatory approval was its substantial equivalence to predicate devices in terms of intended use and technological characteristics.

8. The Sample Size for the Training Set

This information is not applicable. This submission is for a physical medical device, not a machine learning model, so there is no concept of a "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reasons as #8.

§ 870.1390 Trocar.

(a)
Identification. A trocar is a sharp-pointed instrument used with a cannula for piercing a vessel or chamber to facilitate insertion of the cannula.(b)
Classification. Class II (special controls). Except for trocars that are reprocessed for multiple use, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.