(329 days)
Not Found
Not Found
No
The summary describes standard signal processing algorithms for calculating PR and SpO2 from PPG data and an accelerometer for motion detection. There is no mention of AI, ML, or deep learning in the device description, intended use, or performance studies. The algorithms are described as using a "10 second rolling window," which is a traditional signal processing technique.
No.
The "Intended Use / Indications for Use" states that the device is "not intended to replace traditional methods of monitoring, diagnosis or treatment." It is primarily a monitoring device that provides notifications to prompt assessment, not to actively treat or mitigate a disease or condition.
No
The text explicitly states: "The feature is not intended to replace traditional methods of monitoring, diagnosis or treatment." It is intended to provide notifications and supplement caregivers' decisions, not to diagnose.
No
The device description explicitly states that the device includes three components: the sock hardware, the base station, and the mobile application. The sock hardware contains the PPG sensor and accelerometer, which are physical components essential for data acquisition.
Based on the provided information, the Dream Sock is not an In Vitro Diagnostic (IVD) device.
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information about a person's health.
- Dream Sock's Function: The Dream Sock measures physiological parameters (pulse rate and oxygen saturation) directly from the infant's body using photoplethysmography. It does not analyze samples taken from the body.
- Intended Use: The intended use is for monitoring and providing notifications to caregivers based on these direct measurements, not for analyzing biological specimens for diagnostic purposes.
Therefore, the Dream Sock falls under the category of a medical device, but not specifically an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The Dream Sock analyzes photoplethysmography data to identify instances when the infant's pulse rate (PR) and/or oxygen saturation (SpO2) moves outside a preset range. and provides a notification to the caregiver, prompting them to assess the infant. The Dream Sock also displays the infant's PR and SpO2 values to the caregiver and displays trends in these measured values, and their relationship to the preset ranges, over time. These PR and SpO2 notifications and displays on the Dream Sock are intended for use in infants who are 1 to 18 months of age and between 6 to 30 lbs.
The Dream Sock is intended for over-the-counter (OTC) use only in a home environment. It is not intended to provide notification for every episode of the unexpected occurrences of elevated or depressed PR or a low SpO2 level: rather, the feature is intended to provide a notification only when sufficient data are available for analysis. The notifications and associated data can be used to supplement the decision by caregivers to seek additional guidance for medical care of the infant. The feature is not intended to replace traditional methods of monitoring, diagnosis or treatment.
The Dream Sock is not intended for use with infants previously diagnosed with cardiovascular or respiratory disease or conditions.
Product codes
QYU
Device Description
The Dream Sock uses photoplethysmography (PPG) on the infant's foot to measure the continuous pulse rate (PR) and oxygen saturation (SpO2) in the home environment. The device is intended to monitor healthy infants 1 to 18 months of age and between 6 to 30 lbs. The device includes alarms to alert the caregiver that the pulse rate or SpO2 has gone outside a preset range. The alarm thresholds are pulse rate 220 bpm and the SpO244 weeks and
N/A
0
DE NOVO CLASSIFICATION REQUEST FOR OWLET DREAM SOCK
REGULATORY INFORMATION
FDA identifies this generic type of device as:
Infant pulse rate and oxygen saturation monitor for over-the-counter use. An infant pulse rate and oxygen saturation monitor for over-the-counter use is a device that uses photoplethysmography to measure pulse rate and oxygen saturation in infants. The device may contain alarms that alert the caregiver when vital sign(s) go outside preset threshold(s).
NEW REGULATION NUMBER: 21 CFR 870.2705
CLASSIFICATION: Class II
PRODUCT CODE: QYU
BACKGROUND
DEVICE NAME: Dream Sock
SUBMISSION NUMBER: DEN220091
DATE DE NOVO RECEIVED: December 14, 2022
SPONSOR INFORMATION:
Owlet 3300 N. Ashton Blvd Suite 300 Lehi, UT 84043
INDICATIONS FOR USE
The Dream Sock is indicated as follows:
The Dream Sock analyzes photoplethysmography data to identify instances when the infant's pulse rate (PR) and/or oxygen saturation (SpO2) moves outside a preset range. and provides a notification to the caregiver, prompting them to assess the infant. The Dream Sock also displays the infant's PR and SpO2 values to the caregiver and displays trends in these measured values, and their relationship to the preset ranges, over time. These PR and SpO2 notifications and displays on the Dream Sock are intended for use in infants who are 1 to 18 months of age and between 6 to 30 lbs.
1
The Dream Sock is intended for over-the-counter (OTC) use only in a home environment. It is not intended to provide notification for every episode of the unexpected occurrences of elevated or depressed PR or a low SpO2 level: rather, the feature is intended to provide a notification only when sufficient data are available for analysis. The notifications and associated data can be used to supplement the decision by caregivers to seek additional guidance for medical care of the infant. The feature is not intended to replace traditional methods of monitoring, diagnosis or treatment.
The Dream Sock is not intended for use with infants previously diagnosed with cardiovascular or respiratory disease or conditions.
LIMITATIONS
The Dream Sock does not detect, diagnose or reduce Sudden Infant Death Syndrome (SIDS) or Sudden Unexpected Infant Death (SUID).
The Owlet Dream Sock is not a substitute for adult supervision or safe sleep practices.
The Dream Sock is not intended to diagnose, cure, treat, alleviate, or prevent any disease or condition.
Alternate between feet every 8 hours and also after recharging the Sensor. Check the child's foot often for any signs of irritated skin.
The Dream Sock is not for babies and children with ongoing health conditions.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS. PRECAUTIONS AND CONTRAINDICATIONS.
DEVICE DESCRIPTION
The Dream Sock uses photoplethysmography (PPG) on the infant's foot to measure the continuous pulse rate (PR) and oxygen saturation (SpO2) in the home environment. The device is intended to monitor healthy infants 1 to 18 months of age and between 6 to 30 lbs. The device includes alarms to alert the caregiver that the pulse rate or SpO2 has gone outside a preset range. The alarm thresholds are pulse rate 220 bpm and the SpO210% | PASS |
| Material Mediated
Pyrogenicity | Detection of material
mediated pyrogenicity | ISO 10993-11 | Animals survive and
are healthy via
temperature
monitoring | PASS |
See clinical section for skin observation study that was used to support the safe use of the device in the intended use population.
3
ELECTROMAGNETIC COMPATIBILITY & ELECTRICAL SAFETY
The Dream Sock conforms to the FDA-recognized standards for basic safety and essential performance of Medical Electrical Equipment. These standards include:
- IEC 60601-1-2:2014- Medical Electrical Equipment, Part 1-2: General . requirements for safety- Collateral Standard: Electromagnetic compatibility-Requirements and tests
- IEC 60601-1:2006 + A1:2013- Medical Electrical Equipment, Part 1: General . requirements for basic safety and essential performance
- ASNI C63.27:2017- American National Standard for Evaluation of Wireless . Coexistence
- . AAMI TIR69:2017- Technical Information Report Risk management of radiofrequency wireless coexistence for medical devices and systems.
- IEC 60601-1-11:2015/AMD1:2020- General requirements for basic safety and . essential performance- Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
SOFTWARE
The Dream Sock has a Major Level of Concern (LOC). Appropriate documentation was provided to support the validation of the software for Major LOC in accordance with the FDA Guidance Document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," (issued May 11, 2005).
PERFORMANCE TESTING - BENCH
Test | Purpose | Method | Acceptance Criteria | Results | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Pulse | ||||||||||||
Amplitude | ||||||||||||
(PA) | Verify the | |||||||||||
accuracy of the | ||||||||||||
PA compared | ||||||||||||
against the | ||||||||||||
simulator value | ||||||||||||
& verify PR and | ||||||||||||
SpO2 is accurate | ||||||||||||
at different Pas | 3 devices were | |||||||||||
tested with a | ||||||||||||
simulator and the | ||||||||||||
PR, SpO2, and | ||||||||||||
PA were | ||||||||||||
measured by the | ||||||||||||
DUT | $≤$ (b)% ARMS | |||||||||||
AND | ||||||||||||
As the simulator PA rises | ||||||||||||
from (b)(4)%, the ΔPR≤1bpm | ||||||||||||
& | ||||||||||||
Simulator PA ΔSpO2 (b)%≤PA $≤$ (b)% (b)%≤PA44 weeks and ≥18 months who weigh 6-30 lbs. The device was tested in 35 subjects who are independent from the algorithm training set. The results show that the SpO2 and pulse rate had an average root mean square (Ams) of 2.16% and 3.53 bpm, respectively. |
Image /page/5/Figure/3 description: The image contains two scatter plots comparing measurements from two different devices. The left plot, titled "Figure 11.2 Masimo Radical 97 and Owlet OSS 3.0 PR Agreement", compares pulse rate measurements, with the Owlet OSS 3.0 on the y-axis and the Masimo Radical 97 on the x-axis. A regression line is plotted with the equation y = 1.02 + 0.998x and an R-squared value of 0.95. The right plot, titled "Figure 11.3 Masimo Radical 97 and Owlet OSS 3.0% Oxygen Agreement", compares SpO2 measurements, with the Owlet SpO2 on the y-axis and the Masimo SpO2 on the x-axis, and has a regression line with the equation y = 10.2 + 0.902x and an R-squared value of 0.72.
NICU Study
A prospective, single center, clinical study was conducted to support the alarm functionality. Subjects postmenstrual age