K Number
DEN200072
Device Name
Lumipulse G ß-Amyloid Ratio (1-42/1-40)
Date Cleared
2022-05-04

(530 days)

Product Code
Regulation Number
866.5840
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Lumipulse G B-Amyloid Ratio (1-42/1-40) is an in vitro cerebral spinal fluid (CSF) test that combines the results of Lumipulse G B-Amyloid 1-42 and Lumipulse & B-Amyloid 1-40 assays into a ratio of ß-amyloid 1-42 to ß-amyloid 1-40 concentrations using the LUMIPULSE G 1200 System. The Lumipulse G B-Amyloid Ratio (1-42/1-40) is intended to be used in adult patients, aged 55 years and older, presenting with cognitive impairment who are being evaluated for Alzheimer's disease (AD) and other causes of cognitive decline. A test result ≥ 0.073 is a negative result which is consistent with a negative amyloid positron emission tomography (PET) scan result. A negative result reduces the likelihood that a patient's cognitive impairment is due to AD. A test result ≤ 0.058 is a positive result which is consistent with a positive amyloid PET scan result. A positive result does not establish a diagnosis of AD or other cognitive disorder. A test result between 0.059 and 0.072 is considered as a likely positive result as it is more likely consistent with a positive amyloid PET scan result. A likely positive result does not establish a diagnosis of AD or other cognitive disorders and has increased uncertainty in regard to amyloid PET positivity. The Lumipulse G B-Amyloid Ratio (1-42/1-40) results must be interpreted in conjunction with other patient clinical information . This test is not intended as a screening or stand-alone diagnostic test.
Device Description
The Lumipulse G B-Amyloid Ratio (1-42/1-40) is an in vitro cerebral spinal fluid (CSF) test that calculates the ratio of two analytes, Lumipulse G ß-Amyloid 1-42 and Lumipulse & ß-Amyloid 1-40 assays to generate a numeric value between 0.001 to 1.000. The test system consists of two component assays, Lumipulse & ß-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 assay, running on LUMIPULSE G1200 system, and Lumipulse G B-Amyloid Ratio (1-42/1-40) Calculator Tool to calculate the Lumipulse G B-Amyloid Ratio (1-42/1-40). Lumipulse G B-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 assays are packed individually. Results of individual assays have not been assessed to support the intended use except for determination of the Lumipulse G ß-Amyloid Ratio (1-42/1-40).
More Information

Not Found

No
The device description and intended use indicate a calculation of a ratio from two assay results using a calculator tool, not the use of AI/ML for interpretation or analysis. The performance studies describe the determination of cutoffs based on statistical analysis of clinical data, which is standard practice and does not inherently involve AI/ML.

No
This device is an in vitro diagnostic test intended to aid in the evaluation of Alzheimer's disease by measuring ß-amyloid ratios in CSF, not to treat a condition.

No

The "Intended Use / Indications for Use" section explicitly states: "This test is not intended as a screening or stand-alone diagnostic test."

No

The device description explicitly states that the test system consists of two component assays and the LUMIPULSE G1200 system, which are hardware components used to perform the in vitro diagnostic test. While there is a "Calculator Tool" mentioned, the overall system includes hardware.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use/Indications for Use: The description explicitly states it is an "in vitro cerebral spinal fluid (CSF) test". It is intended to be used on a biological sample (CSF) outside of the body to provide information for the evaluation of Alzheimer's disease and other causes of cognitive decline.
  • Device Description: It is described as an "in vitro cerebral spinal fluid (CSF) test" that measures analytes in CSF.
  • Anatomical Site: It uses Cerebral spinal fluid (CSF), which is a biological sample.

These points clearly align with the definition of an In Vitro Diagnostic device, which is used to examine specimens derived from the human body to provide information for the diagnosis, prevention, or treatment of a disease or condition.

N/A

Intended Use / Indications for Use

The Lumipulse G B-Amyloid Ratio (1-42/1-40) is an in vitro cerebral spinal fluid (CSF) test that combines the results of Lumipulse G B-Amyloid 1-42 and Lumipulse & B-Amyloid 1-40 assays into a ratio of ß-amyloid 1-42 to ß-amyloid 1-40 concentrations using the LUMIPULSE G 1200 System. The Lumipulse G B-Amyloid Ratio (1-42/1-40) is intended to be used in adult patients, aged 55 years and older, presenting with cognitive impairment who are being evaluated for Alzheimer's disease (AD) and other causes of cognitive decline.

A test result ≥ 0.073 is a negative result which is consistent with a negative amyloid positron emission tomography (PET) scan result. A negative result reduces the likelihood that a patient's cognitive impairment is due to AD.

A test result ≤ 0.058 is a positive result which is consistent with a positive amyloid PET scan result. A positive result does not establish a diagnosis of AD or other cognitive disorder.

A test result between 0.059 and 0.072 is considered as a likely positive result as it is more likely consistent with a positive amyloid PET scan result. A likely positive result does not establish a diagnosis of AD or other cognitive disorders and has increased uncertainty in regard to amyloid PET positivity.

The Lumipulse G B-Amyloid Ratio (1-42/1-40) results must be interpreted in conjunction with other patient clinical information.

This test is not intended as a screening or stand-alone diagnostic test.

Product codes (comma separated list FDA assigned to the subject device)

QSE

Device Description

The Lumipulse G B-Amyloid Ratio (1-42/1-40) is an in vitro cerebral spinal fluid (CSF) test that calculates the ratio of two analytes, Lumipulse G ß-Amyloid 1-42 and Lumipulse & ß-Amyloid 1-40 assays to generate a numeric value between 0.001 to 1.000.

The test system consists of two component assays, Lumipulse & ß-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 assay, running on LUMIPULSE G1200 system, and Lumipulse G B-Amyloid Ratio (1-42/1-40) Calculator Tool to calculate the Lumipulse G B-Amyloid Ratio (1-42/1-40). Lumipulse G B-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 assays are packed individually. Results of individual assays have not been assessed to support the intended use except for determination of the Lumipulse G ß-Amyloid Ratio (1-42/1-40).

a) Lumipulse G ß-Amyloid 1-42:
Lumipulse G B-Amyloid 1-42 is used for the quantitative measurement of B-amyloid 1-22 in human CSF in order to calculate the Lumipulse G B-Amyloid Ratio (1-42/1-40). The components of Lumipulse G β-Amyloid 1-42 are as follows:
Antibody Coated Particle Solution: 250 µL, anti-β-amyloid 1.42 monoclonal antibody (mouse)-coated particles and protein stabilizers (bovine) in 50 mM MES buffer. This solution contains gelatin and turns into gel at 15°C or lower. Preservative: 0.1% ProClin 300.
Biotinylated Antibody Solution: 120 µL, biotinylated antibody (mouse), protein (bovine) stabilizers and chemical stabilizers in 50 mM Tris buffer. Preservative: 0.1% ProClin 300.
Enzyme Labeled Streptavidin Solution: 350 µL, alkaline phosphatase (ALP)-labeled streptavidin, protein stabilizers (bovine) and chemical stabilizers in 50 mM MES buffer. Preservative: 0.1% ProClin 300.

Additional materials required but sold separately:
Lumipulse G B-Amyloid 1-42 Calibrators Set: Three Lumipulse G B-Amyloid 1-42 calibrators with at concentration level of 0, 129, and 2.335 pg/mL
Lumipulse ß-Amyloid Controls: Three levels of Lumipulse G ß-Amyloid 1-42 Controls with target concentration of 4,000, 10,000, and 20,000 pg/mL, lyophilized.

b) Lumipulse G B-Amyloid 1-40:
Lumipulse G B-Amyloid 1-40 is used for the quantitative measurement of B-amyloid 1-40 in human CSF in order to calculate the Lumipulse G 3-Amyloid Ratio (1-42/1-40). The components of Lumipulse G B-Amyloid 1-40 are as follows:
Antibody Coated Particle Solution: 150 µL, anti-β-amyloid1-40 monoclonal antibody (mouse)-coated particles and protein stabilizers (bovine) in 50 mM MES buffer. This solution contains gelatin and turns into gel at 15°C or lower. Preservative: 0.1% ProClin 300.
Enzyme Labeled Antibody Solution: 320 µL, alkaline phosphatase (ALP)-labeled anti-β-amyloid antibody (mouse) conjugate, protein stabilizers (bovine) and chemical stabilizers in in 50 mM MES buffer. Preservative: 0.1% ProClin 300

Additional materials required but sold separately:
Lumipulse G 3-Amyloid 1-40 Calibrators Set: three Lumipulse G B-Amyloid 1-40 . calibrators with at concentration level of 0, 500, and 30,000 pg/mL.
Lumipulse ß-Amyloid Controls: three levels of Lumipulse G ß-Amyloid 1-40 . Controls with target concentration of 274. 548, and 1.027 pg/mL, lyophilized.

Both Lumipulse G B-Amyloid 1-42 and Lumipulse G B-Amyloid 1-40 are run on the LUMIPULSE G1200 System. LUMIPULSE G1200 is an instrument platform that can perform automated chemiluminescence immunoassays of specimens using LUMIPULSE G reagents, conducting various processes such as dispensing, agitation, and photometric measurement.

  1. The Lumipulse G B-Amyloid Ratio (1-42/1-40) Calculator Tool: The Lumipulse G B-Amyloid Ratio (1-42/1-40) Calculator Tool is a web-based software to calculate the ratio of B-amyloid 1-40 to B-amyloid 1-40 using results of the Lumipulse G-8-Amyloid 1-42 assay and the Lumipulse G B-Amyloid 1-40 assay in CSF samples. The user is instructed to launch https://www.fujircbio-amyloid-ratio.com/ to perform the result calculation.

  2. Result interpretation:
    The Lumipulse G B-Amyloid Ratio (1-42/1-40) is calculated based on the values from Lumipulse G B-Amyloid 1-42 and Lumipulsc G B-Amyloid 1-40 and generate a numerical ratio in a range of 0.001 to 1.000 as below:
    Ratio ≥ 0.073: Negative, Consistent with a negative amyloid PET scan result. Reduced likelihood that a patient's cognitive impairment is due to AD.
    Ratio ≤ 0.058: Positive, Consistent with a positive amyloid PET scan result. Diagnosis of AD or other cognitive disorder not established.
    0.059 ≤ Ratio ≤ 0.072: Likely Positive, Considered as a likely positive result as it is more likely consistent with a positive amyloid PET scan result. Diagnosis of AD or other cognitive disorders not established and increased uncertainty in regard to amyloid PET positivity.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Cerebral Spinal Fluid (CSF)

Indicated Patient Age Range

Adult patients, aged 55 years and older

Intended User / Care Setting

For prescription use only.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

A clinical study was conducted to evaluate the performance of the Lumipulse G B-Amyloid Ratio (1-42/1-40) as an aid in the assessment of whether a patient presenting with cognitive impairment and being evaluated for AD and other causes of cognitive decline would test positive or negative for amyloid plaques at the time of testing, as measured by an amyloid PET imaging agent. The study included 292 patients aged 55 to 90 years and with sufficient volume of banked CSF samples obtained from the Alzheimer's Disease Neuroimaging Initiative (ADNI) sample bank. A standardized ADNI pre-analytical protocol was used for collection, processing and handling the CSF samples. All study subjects had a clinical diagnosis and had undergone Florbetapir (18F) PET evaluation.

The amyloid PET status of each patient was determined by a minimum of three trained independent readers (two readers and an adjudicator when the first two readers disagree) who were blinded to all other clinical data.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Clinical Performance Study:

  • Study type: Clinical study comparing Lumipulse G B-Amyloid Ratio (1-42/1-40) results to visual amyloid PET read.
  • Sample size: 292 patients
  • Key Results:
    • Agreement with PET Scan:

      • Positive (Ratio ≤ 0.058): 171 PET Positive, 6 PET Negative (Total 177). Predictive Value: 96.6% (171/177). Frequency of Results: 60.6% (177/292).
      • Likely positive (0.059 ≤ Ratio ≤ 0.072): 13 PET Positive, 9 PET Negative (Total 22). Predictive Value: 59.1% (13/22). Frequency of Results: 7.5% (22/292).
      • Negative (Ratio ≥ 0.073): 15 PET Positive, 78 PET Negative (Total 93). Predictive Value: 16.1% (15/93). Frequency of Results: 31.8% (93/292).
    • Sub-group Analysis (Predictive Value for Positive Test Results):

      • Diagnostic group: AD (98.9%), LMCI (97.0%), EMCI (91.5%), SCD (100.0%)
      • Sex: Male (96.1%), Female (97.3%)
      • Age: 55-59 (85.7%), 60-69 (95.3%), 70-79 (97.9%), ≥80 (96.7%)
      • Race (White): 96.4%
    • The data of clinical performance study support that positive Lumipulse G 8-Amyloid Ratio (1-42/1-40) results are consistent with PET scan positive results and negative Lumipulse G B-Amvloid Ratio (1-42/1-40) results are consistent with PET scan negative results.

Analytical Performance:

  • Precision/Reproducibility: Evaluated within-laboratory, lot-to-lot, and site-to-site precision using a seven-member panel. Results presented as SD and %CV for Lumipulse G B-Amyloid 1-42, Lumipulse G B-Amyloid 1-40, and Lumipulse G B-Amyloid Ratio (1-42/1-40). All results met pre-determined acceptance criteria.
  • Linearity: Evaluated for Lumipulse G B-Amyloid 1-42 (37.5 – 2203.5 pg/mL) and Lumipulse G B-Amyloid 1-40 (156.3 – 28450.3 pg/mL). Percent deviation from linearity (DL) was within ± 10% for each dilution level.
  • High-Dose Hook Effect: No high-dose hook effect was observed up to approximately 159,000 pg/mL for B-Amyloid 1-42 and approximately 150,000 pg/mL for B-Amyloid 1-40.
  • Analytical Specificity/Interference: No significant interference was observed (≤±10% difference of test from control) for tested endogenous and exogenous substances.
  • Cross-reactivity: Lumipulse G B-Amyloid 1-42 showed no significant cross-reactivity for most tested cross-reactants except B-Amyloid 1-43. Lumipulse G B-Amyloid 1-40 showed no noticeable cross-reactivity for most tested cross-reactants except B-amyloid 1-40.
  • Detection Limit: LoB, LoD, and LoQ determined for both Lumipulse G B-Amyloid 1-42 and Lumipulse G B-Amyloid 1-40 assays.
    • Lumipulse G B-Amyloid 1-42: LoB = 2.18 pg/mL, LoD = 11.6 pg/mL, LoQ = 38 pg/mL.
    • Lumipulse G B-Amyloid 1-40: LoB = 0.97 pg/mL, LoD = 33.0 pg/mL, LoQ = 158 pg/mL.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

The performance estimates of the Lumipulse G B-Amyloid Ratio (1-42/1-40) is described by PET-positive predictive value (indicated as Predictive Value, PV) and frequency of positive test results of the tested samples (indicated as Frequency of Results, FR) for positive, likely positive and negative test results and it is summarized in the table below.

| | | Visual Amyloid
PET Read | | Predictive
Value %
(n/N)
(95% CI) | Frequency of
Results %
(n/N)
(95% CI) |
|--------------------------------------------------|-----------------------------------------------|----------------------------|-------|--------------------------------------------|------------------------------------------------|
| | | Positive | Total | | |
| Lumipulse G
β-Amyloid
Ratio
(1-42/1-40) | Positive
(Ratio ≤ 0.058) | 171 | 177 | 96.6%
(171/177)
(92.8%; 98.4%) | 60.6%
(177/292)
(54.9%; 66.1%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 13 | 22 | 59.1%
(13/22)
(38.7%; 66.7%) | 7.5%
(22/292)
(5.0%; 11.1%) |
| | Negative
(Ratio ≥ 0.073) | 15 | 93 | 16.1%
(15/93)
(10.0%; 24.9%) | 31.8%
(93/292)
(26.8%; 37.4%) |
| | Total | 199 | 292 | | Prevalence of PET positive = 68.2% |

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

Not Found

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K142895

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

N/A

0

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR Lumipulse G ß-Amyloid Ratio (1-42/1-40) DECISION SUMMARY

I Background Information:

A De Novo Number

DEN200072

B Applicant

Fujirebio Diagnostics, Inc.

C Proprietary and Established Names

Lumipulse G ß-Amyloid Ratio (1-42/1-40)

D Regulatory Information

| Product
Code(s) | Classification | Regulation
Section | Panel |
|--------------------|-----------------------------|-----------------------|-----------------|
| QSE | Class II (special controls) | 21 CFR §866.5840 | Immunology (82) |

II Submission/Device Overview:

A Purpose for Submission:

De Novo request for evaluation of automatic class III designation for Lumipulse & B-Amyloid Ratio (1-42/1-40)

Measurand: B

B-amyloid ratio of two cerebral spinal fluid (CSF) analytes: B-amyloid 1-42 and B-amyloid 1-40

C Type of Test:

Fully automated, chemiluminescent enzyme immunoassays (CLEIA)

1

III Indications for Use:

A Indication(s) for Use:

The Lumipulse G B-Amyloid Ratio (1-42/1-40) is an in vitro cerebral spinal fluid (CSF) test that combines the results of Lumipulse G B-Amyloid 1-42 and Lumipulse & B-Amyloid 1-40 assays into a ratio of ß-amyloid 1-42 to ß-amyloid 1-40 concentrations using the LUMIPULSE G 1200 System. The Lumipulse G B-Amyloid Ratio (1-42/1-40) is intended to be used in adult patients, aged 55 years and older, presenting with cognitive impairment who are being evaluated for Alzheimer's disease (AD) and other causes of cognitive decline.

A test result ≥ 0.073 is a negative result which is consistent with a negative amyloid positron emission tomography (PET) scan result. A negative result reduces the likelihood that a patient's cognitive impairment is due to AD.

A test result ≤ 0.058 is a positive result which is consistent with a positive amyloid PET scan result. A positive result does not establish a diagnosis of AD or other cognitive disorder.

A test result between 0.059 and 0.072 is considered as a likely positive result as it is more likely consistent with a positive amyloid PET scan result. A likely positive result does not establish a diagnosis of AD or other cognitive disorders and has increased uncertainty in regard to amyloid PET positivity.

The Lumipulse G B-Amyloid Ratio (1-42/1-40) results must be interpreted in conjunction with other patient clinical information .

This test is not intended as a screening or stand-alone diagnostic test.

B Special Conditions for Use Statement(s):

For prescription use only For in vitro diagnostic use

C Special Instrument Requirements:

The LUMIPULSE G1200 System (cleared under K142895)

IV Device/System Characteristics:

A Device Description:

    1. Device:
      The Lumipulse G B-Amyloid Ratio (1-42/1-40) is an in vitro cerebral spinal fluid (CSF) test that calculates the ratio of two analytes, Lumipulse G ß-Amyloid 1-42 and Lumipulse & ß-Amyloid 1-40 assays to generate a numeric value between 0.001 to 1.000.

2

The test system consists of two component assays, Lumipulse & ß-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 assay, running on LUMIPULSE G1200 system, and Lumipulse G B-Amyloid Ratio (1-42/1-40) Calculator Tool to calculate the Lumipulse G B-Amyloid Ratio (1-42/1-40). Lumipulse G B-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 assays are packed individually. Results of individual assays have not been assessed to support the intended use except for determination of the Lumipulse G ß-Amyloid Ratio (1-42/1-40).

  • a) Lumipulse G ß-Amyloid 1-42:
    Lumipulse G B-Amyloid 1-42 is used for the quantitative measurement of B-amyloid 1-22 in human CSF in order to calculate the Lumipulse G B-Amyloid Ratio (1-42/1-40). The components of Lumipulse G β-Amyloid 1-42 are as follows:
Lumipulse G β-Amyloid 1-42 (per Immunoreaction Cartridge)
ComponentVolumeContents
Antibody Coated
Particle Solution250 µLanti-β-amyloid 1.42 monoclonal antibody (mouse)-coated
particles and protein stabilizers (bovine) in 50 mM MES
buffer. This solution contains gelatin and turns into gel at
15°C or lower. Preservative: 0.1% ProClin 300.
Biotinylated
Antibody Solution120 µLbiotinylated antibody (mouse), protein (bovine)
stabilizers and chemical stabilizers in 50 mM Tris buffer.
Preservative: 0.1% ProClin 300.
Enzyme Labeled
Streptavidin
Solution350 µLalkaline phosphatase (ALP)-labeled streptavidin, protein
stabilizers (bovine) and chemical stabilizers in 50 mM
MES buffer. Preservative: 0.1% ProClin 300.

Additional materials required but sold separately:

  • · Lumipulse G B-Amyloid 1-42 Calibrators Set: Three Lumipulse G B-Amyloid 1-42 calibrators with at concentration level of 0, 129, and 2.335 pg/mL
  • · Lumipulse ß-Amyloid Controls: Three levels of Lumipulse G ß-Amyloid 1-42 Controls with target concentration of 4,000, 10,000, and 20,000 pg/mL, lyophilized.
  • b) Lumipulse G B-Amyloid 1-40:

Lumipulse G B-Amyloid 1-40 is used for the quantitative measurement of B-amyloid 1-40 in human CSF in order to calculate the Lumipulse G 3-Amyloid Ratio (1-42/1-40). The components of Lumipulse G B-Amyloid 1-40 are as follows:

3

Lumipulse G β -Amyloid 1-40 (per Immunoreaction Cartridge)
ComponentVolumeContents
Antibody Coated
Particle Solution150 µLanti-β-amyloid1-40 monoclonal antibody (mouse)-coated
particles and protein stabilizers (bovine) in 50 mM MES
buffer. This solution contains gelatin and turns into gel at
15°C or lower. Preservative: 0.1% ProClin 300.
Enzyme Labeled
Antibody
Solution320 µLalkaline phosphatase (ALP)-labeled anti-β-amyloid
antibody (mouse) conjugate, protein stabilizers (bovine)
and chemical stabilizers in in 50 mM MES buffer.
Preservative: 0.1% ProClin 300

Additional materials required but sold separately:

  • Lumipulse G 3-Amyloid 1-40 Calibrators Set: three Lumipulse G B-Amyloid 1-40 . calibrators with at concentration level of 0, 500, and 30,000 pg/mL.
  • Lumipulse ß-Amyloid Controls: three levels of Lumipulse G ß-Amyloid 1-40 . Controls with target concentration of 274. 548, and 1.027 pg/mL, lyophilized.

Both Lumipulse G B-Amyloid 1-42 and Lumipulse G B-Amyloid 1-40 are run on the LUMIPULSE G1200 System. LUMIPULSE G1200 is an instrument platform that can perform automated chemiluminescence immunoassays of specimens using LUMIPULSE G reagents, conducting various processes such as dispensing, agitation, and photometric measurement.

    1. The Lumipulse G B-Amyloid Ratio (1-42/1-40) Calculator Tool: The Lumipulse G B-Amyloid Ratio (1-42/1-40) Calculator Tool is a web-based software to calculate the ratio of B-amyloid 1-40 to B-amyloid 1-40 using results of the Lumipulse G-8-Amyloid 1-42 assay and the Lumipulse G B-Amyloid 1-40 assay in CSF samples. The user is instructed to launch https://www.fujircbio-amyloid-ratio.com/ to perform the result calculation.

3. Result interpretation:

The Lumipulse G B-Amyloid Ratio (1-42/1-40) is calculated based on the values from Lumipulse G B-Amyloid 1-42 and Lumipulsc G B-Amyloid 1-40 and generate a numerical ratio in a range of 0.001 to 1.000 as below.

4

| Lumipulse G β-
Amyloid Ratio (1-

42/1-40)Test ResultInterpretationClinical Implication
Ratio ≥ 0.073NegativeConsistent with a
negative amyloid PET
scan result.Reduced likelihood
that a patient's
cognitive impairment
is due to AD
Ratio ≤ 0.058PositiveConsistent with a
positive amyloid PET
scan result.Diagnosis of AD or
other cognitive
disorder not
established
0.059 ≤ Ratio ≤
0.072Likely PositiveConsidered as a likely
positive result as it is
more likely consistent
with a positive amyloid
PET scan result.Diagnosis of AD or
other cognitive
disorders not
established and
increased uncertainty
in regard to amyloid
PET positivity

Principle of Operation B

    1. Specimen collection Refer to the Lumipulse G B-Amyloid Ratio (1-42/1-40) package insert
    1. Lumipulse G B-Amyloid 1-42

Lumipulse G B-Amyloid 1-42 is an assay including a set of immunoassay reagents for the quantitative measurement of ß-amyloid;.42 in human CSF specimens based on CLEIA technology using the LUMIPULSE G System as follows:

    1. First reaction step: B-Amyloid (1-42) calibrator or specimen (50 uL) and biotinylated antibody solution are both added to 'antibody-coated particle solution' compartment in the Immunoreaction Cartridge. B-amyloid 1-42 in specimens or calibrators specifically bind to anti-B-amyloid 1-42 monoclonal antibody (mouse) on the particles, and biotinylated antibody (mouse). Biotinylated antibody-antigen immuno-complexes are formed.
    1. Washing step: The particles are washed and rinsed to remove unbound materials.
    1. Second reaction step: Alkaline phosphatase (ALP)-labeled streptavidin specifically that binds to biotinylated immuno-complexes on the particles are added.
    1. Washing step: The particles are washed and rinsed to remove unbound materials.
    1. Enzyme reaction step: Substrate Solution is added and mixed with the particles. 3-(2 spiroadamantane)-4-methoxy-4-(3"-phosphoryloxy) phenyl-1, 2-dioxetane disodium salt (AMPPD) contained in the Substrate Solution is dephosphorylated by the catalysis of ALP indirectly conjugated to particles.

5

    1. Luminescence measurement step: Luminescence (at a maximum wavelength of 477 mm) is generated by the cleavage reaction of dephosphorylated AMPPD. The luminescent signal reflects the amount of B-amyloid 1-42 present in the sample.
  1. Lumipulse G B-Amvloid 1-40

Lumipulse G B-Amyloid 1-40 is an assay including a set of immunoassay reagents, for the quantitative measurement of B-amyloid 1-40 in human CSF specimens based on CLEIA technology using the LUMIPULSE G System as follows:

    1. First reaction step: B-Amyloid (1-40) calibrator or specimen (40 µL) and biotinylated antibody solution are both added to 'antibody-coated particle solution' compartment in the Immunoreaction Cartridge. B-amyloid 1-40 in specimens or calibrators specifically bind to anti-B-Amyloid 1-40 monoclonal antibody (mouse) on the particles and antigenantibody immunocomplexes are formed.
    1. Washing: The particles are washed and rinsed to remove unbound materials.
    1. Second reaction: Alkaline phosphatase (ALP)-labeled anti-B-amyloid monoclonal antibody (mouse) fragment specifically binds to anti-B-amyloid i.an immune-complexes on the particles and additional immunocomplexes are formed.
    1. Washing: The particles are washed and rinsed to remove unbound materials.
    1. Enzyme reaction: Substrate Solution is added and mixed with the particles. AMPPD contained in the Substrate Solution is dephosphorylated by the catalysis of ALP indirectly conjugated to particles.
    1. Luminescence measurement step: Luminescence (at a maximum wavelength of 477 nm) is generated by the cleavage reaction of dephosphorylated AMPPD. The luminescent signal reflects the amount of B-amyloid 1-40 present in the sample.
    1. Lumipulse G B-Amyloid Ratio (1-42/1-40) Calculator Tool

The Lumipulse G B-Amvloid Ratio (1-42/1-40) Calculator Tool, using input values produced from Lumipulse & B-Amyloid 1-42 and Lumipulse G B-Amyloid 1-40 on the LUMIPULSE G1200 System, is a web-based software available at the following URL ( https://www.fujirebio-amyloid-ratio.com/). The calculator software is validated for use on Microsoft Internet Explorer and Edge, Google Chrome, Mozilla Firefox, and Apple Safari, When invalid values are entered, a 'Invalid Value' message is shown with a description of an error (e.g., "The Lumipulse G B-Amyloid 1-42 assay result value must be less that the Lumipulse G B-Amyloid 1-40 assay result value").

The Lumipulse G ß-Amyloid Ratio (1-42/1-40) is calculated using the following equation and generate a numerical ratio in a range of 0.001 to 1.000 as below.

| Lumipulse G β-Amyloid
Ratio (1-42/1-40) | - | $\frac{Lumipulse\ G\ β-Amyloid\ 1-42\ (results\ in\ pg/mL)}{Lumipulse\ G\ β-Amyloid\ 1-40\ (results\ in\ pg/mL)}$ | - | a numerical value
ranging 0.001 - 1.000 |

-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
  • The Lumipulse G B-Amyloid Ratio (1-42/1-40) Calculator Tool reports the . following: Lumipulse G B-Amyloid Ratio (1-42/1-40) result ≥ 0.073 = Negative
  • Lumipulse G B-Amyloid Ratio (1-42/1-40) result 2200 pg/mL, the value should be set to 2200 pg/mL. The ratio is calculated, and the test results are reported as 'Negative'. If B-amyloid >> 28.450 pg/mL, the ratio is not calculated using the Lumipulse & B-Amvloid Ratio (1-42/1-40) Calculator Tool. The results are manually reported as *ratio undetermined'.

C Instrument Description Information

    1. Instrument Name: The LUMIPULSE G1200 System (cleared under K142895)
    1. Specimen Identification: Refer to K142895
    1. Specimen Sampling and Handling: Refer to K142895
    1. Calibration: Refer to K142895
    1. Quality Control: Refer to K142895

V Standards/Guidance Documents Referenced:

  • . CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures: Approved Guideline - Third Edition
  • . CLSI EP06, 2nd ed .: Evaluation of the Linearity of Quantitative Measurement Procedures Second Edition
  • . CLSI EP07, 3rd ed .: Interference Testing in Clinical Chemistry; - Third Edition
  • . CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline -Second Edition
  • CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical . Laboratory; Approved Guideline Third Edition
  • . CLSI EP37: Supplemental Tables for Interference Testing in Clinical Chemistry: First Edition

7

Performance Characteristics VI

A Analytical Performance:

All results met the pre-determined acceptance criteria.

1. Precision/Reproducibility:

A study was conducted per CLSI guideline EP05-A3 to evaluate the precision including reproducibility of the Lumipulse & B-Amyloid 1-42, Lumipulse & B-Amyloid 1-40, and Lumipulse G B-Amyloid Ratio (1-42/1-40). A seven-member panel (Panel members 1 7) was prepared to achieve target concentrations of B-amyloid 1-42 and B-amyloid 1-40 that cover the measuring ranges of the individual assays. Panel members 1-4 were prepared by pooling native CSF samples and Panel members 5-7 were prepared by spiking recombinant Bamyloid 1.42 and B-amyloid 1.40 proteins into native CSF pools to challenge various ratios of B-amyloid 1-42 to B-amyloid 1-40 proteins. The sample panel was used to evaluate 1) withinlaboratory precision, 2) lot-to-lot precision and 3) site-to-site reproducibility, as described below:

    1. Within-laboratory precision:
      To evaluate the within-laboratory precision, each panel member was tested for 20 days, two runs per day, two replicates per run at a single site, using one instrument and one reagent lot for a total of 80 measurements per sample. The results are summarized in the tables below for each analyte and the ratio of the analytes:

| | | Within-Run | | Between-Run | | Between-Day | | Within-
Laboratory | |
|-----------------|-----------------|------------|-----|-------------|-----|-------------|-----|-----------------------|-----|
| Panel
member | Mean
(pg/mL) | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 129 | 2.4 | 1.9 | 3.8 | 2.9 | 1.4 | 1.1 | 4.7 | 3.6 |
| 2 | 425 | 6.0 | 1.4 | 5.7 | 1.3 | 4.6 | 1.1 | 9.5 | 2.2 |
| 3 | 596 | 7.8 | 1.3 | 13.1 | 2.2 | 0.0 | 0.0 | 15.2 | 2.6 |
| 4 | 700 | 9.0 | 1.3 | 8.3 | 1.2 | 6.2 | 0.9 | 13.7 | 2.0 |
| 5 | 881 | 13.3 | 1.5 | 11.1 | 1.3 | 10.7 | 1.2 | 20.4 | 2.3 |
| 6 | 1,274 | 32.7 | 2.6 | 5.2 | 0.4 | 35.7 | 2.8 | 48.7 | 3.8 |
| 7 | 2,084 | 26.3 | 1.3 | 25.9 | 1.2 | 33.6 | 1.6 | 49.9 | 2.4 |

  • a. Lumipulse G B-Amyloid 1-42

8

| Panel
member | Mean
(pg/mL) | Within-Run | | Between-Run | | Between-Day | | Within-
Laboratory | |
|-----------------|-----------------|------------|-----|-------------|-----|-------------|-----|-----------------------|-----|
| 1 | 238 | 18.1 | 7.6 | 0.0 | 0.0 | 6.2 | 2.6 | 19.1 | 8.0 |
| 2 | 7,700 | 106.1 | 1.4 | 81.8 | 1.1 | 36.1 | 0.5 | 138.7 | 1.8 |
| 3 | 10,620 | 151.0 | 1.4 | 204.9 | 1.9 | 0.0 | 0.0 | 254.5 | 2.4 |
| 4 | 12,455 | 154.7 | 1.2 | 172.3 | 1.4 | 99.0 | 0.8 | 251.9 | 2.0 |
| 5 | 15,515 | 231.4 | 1.5 | 172.8 | 1.1 | 187.3 | 1.2 | 344.2 | 2.2 |
| 6 | 20,850 | 715.2 | 3.4 | 410.2 | 2.0 | 470.2 | 2.3 | 949.1 | 4.6 |
| 7 | 26,545 | 422.8 | 1.6 | 475.9 | 1.8 | 514.6 | 1.9 | 818.5 | 3.1 |

  • b. Lumipulse G B-Amyloid 1-40
  • c. Lumipulse G B-Amyloid Ratio (1-42/1-40)

| Panel
member | Mean
Ratio | Within-Run | | Between-Run | | Between-Day | | Within-
Laboratory | |
|-----------------|---------------|------------|-----|-------------|-----|-------------|-----|-----------------------|-----|
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.542 | 2.5 | 4.7 | 1.2 | 2.3 | 1.4 | 2.5 | 3.1 | 5.8 |
| 2 | 0.055 | 0.1 | 2.0 | 0.0 | 0.7 | 0.1 | 1.0 | 0.1 | 2.4 |
| 3 | 0.056 | 0.1 | 2.0 | 0.1 | 1.0 | 0.1 | 0.9 | 0.1 | 2.4 |
| 4 | 0.056 | 0.1 | 1.9 | 0.0 | 0.7 | 0.0 | 0.7 | 0.1 | 2.1 |
| 5 | 0.057 | 0.1 | 2.4 | 0.0 | 0.0 | 0.0 | 0.5 | 0.1 | 2.4 |
| 6 | 0.061 | 0.2 | 3.7 | 0.0 | 0.0 | 0.1 | 1.2 | 0.2 | 3.9 |
| 7 | 0.079 | 0.2 | 2.0 | 0.1 | 1.9 | 0.1 | 0.9 | 0.2 | 2.9 |

2) Lot-to-lot precision:

To evaluate between-lot precision, the panel members were tested at a single site, using one instrument and three reagent lots which were paired between Lumipulse G B-Amyloid 1-42 and Lumipulse G B-Amyloid 1-40. A paired reagent lot was defined as a set of specific lots of Lumipulse G B-Amyloid 1-42 Immunoreaction Cartridges with Lumipulse G B-Amyloid 1-42 Calibrators and Lumipulse G B-Amyloid 1-40 Immunoreaction Cartridges with Lumipulse & B-Amyloid 1-40 Calibrators. Each panel member was tested for five days, two runs per day, three replicates per run using three lots, for a total of 90 measurements for each sample. The results are summarized in the tables below for each analyte and the ratio of the analytes.

9

| Panel
member | Mean
(pg/mL) | Within-Run | | Between-Run | | Between-Day | | Between-Lot | | Total | |
|-----------------|-----------------|------------|-----|-------------|-----|-------------|-----|-------------|-----|-------|-----|
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 123 | 2.4 | 2.0 | 4.6 | 3.7 | 3.2 | 2.6 | 2.1 | 1.7 | 6.4 | 5.2 |
| 2 | 413 | 5.1 | 1.2 | 13.2 | 3.2 | 7.1 | 1.7 | 8.0 | 1.9 | 17.7 | 4.3 |
| 3 | 569 | 8.7 | 1.5 | 22.9 | 4.0 | 0.0 | 0.0 | 3.9 | 0.7 | 24.8 | 4.4 |
| 4 | 686 | 8.2 | 1.2 | 21.0 | 3.1 | 18.8 | 2.7 | 7.7 | 1.1 | 30.3 | 4.4 |
| 5 | 868 | 8.7 | 1.0 | 30.8 | 3.5 | 4.5 | 0.5 | 13.7 | 1.6 | 35.1 | 4.0 |
| 6 | 1,254 | 16.8 | 1.3 | 33.2 | 2.6 | 18.7 | 1.5 | 18.6 | 1.5 | 45.6 | 3.6 |
| 7 | 2,090 | 17.8 | 0.9 | 39.2 | 1.9 | 15.7 | 0.8 | 37.4 | 1.8 | 59.2 | 2.8 |

  • a. Lumipulse G β-Amyloid 1-42
  • b. Lumipulse G β-Amyloid 1-40

| Panel
member | Mean
(pg/mL) | Within-Run | | Between-Run | | Between-Day | | Between-Lot | | Total | |
|-----------------|-----------------|------------|-----|-------------|-----|-------------|-----|-------------|-----|-------|-----|
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 229 | 8.0 | 3.5 | 4.7 | 2.1 | 3.8 | 1.7 | 6.5 | 2.8 | 11.9 | 5.2 |
| 2 | 7,343 | 108.3 | 1.5 | 249.7 | 3.4 | 0.0 | 0.0 | 138.7 | 1.9 | 305.5 | 4.2 |
| 3 | 9,958 | 150.5 | 1.5 | 284.2 | 2.9 | 0.0 | 0.0 | 0.0 | 0.0 | 321.6 | 3.2 |
| 4 | 11,862 | 198.3 | 1.7 | 412.3 | 3.5 | 195.8 | 1.7 | 104.4 | 0.9 | 508.5 | 4.3 |
| 5 | 14,892 | 207.8 | 1.4 | 489.1 | 3.3 | 0.0 | 0.0 | 111.9 | 0.8 | 543.0 | 3.6 |
| 6 | 19,773 | 265.3 | 1.3 | 427.6 | 2.2 | 321.7 | 1.6 | 0.0 | 0.0 | 597.2 | 3.0 |
| 7 | 25,657 | 345.4 | 1.3 | 372.5 | 1.5 | 251.6 | 1.0 | 0.0 | 0.0 | 566.9 | 2.2 |

10

Within-RunBetween-RunBetween-DayBetween-LotTotal
Panel
memberMean
RatioSD%CVSD%CVSD%CVSD%CVSD%CV
10.5381.83.31.52.81.73.21.73.13.46.3
20.0560.12.00.10.90.12.30.12.20.23.9
30.0570.12.20.11.80.11.40.00.90.23.3
40.0580.12.00.11.80.11.10.11.30.23.2
50.0580.11.50.11.10.11.60.00.30.12.5
60.0630.11.90.11.20.12.10.00.80.23.2
70.0810.11.50.22.40.22.20.11.00.33.6
  • c. Lumipulse G ß-Amyloid Ratio (1-42/1-40)

3) Site-to-site reproducibility

To evaluate site-to-site reproducibility, each panel member was tested for five days, two runs per day, three replicates per run at three sites (one instrument at each site) using one reagent lot, for a total of 90 measurements for each sample. The results are summarized in the tables below for each analyte and the ratio of the analytes:

Within-RunBetween-RunBetween-DayBetween-SiteTotal
Panel
memberMean
(pg/mL)SD%CVSD%CVSD%CVSD%CVSD%CV
11242.42.04.33.53.22.69.67.711.39.1
24185.21.213.93.32.70.614.53.520.95.0
35758.51.521.43.70.00.024.14.233.35.8
46898.61.321.63.115.42.216.42.432.44.7
58779.51.120.82.47.00.833.83.941.44.7
61,25717.61.428.22.20.00.043.93.555.04.4
72,12016.90.833.31.632.81.546.52.268.03.2
  • Lumipulse G B-Amyloid 1-42 a.

11

  • b. Lumipulse G B-Amyloid 1-40

| Panel

member (pg/mL)MeanWithin-RunBetween-RunBetween-DayBetween-SiteTotal
SD%CVSD%CVSD%CVSD%CVSD%CV
12325.92.59.74.218.78.013.05.625.411.0
27,284102.11.4218.43.0222.03.0158.42.2364.05.0
39,946134.01.3358.13.6138.71.4284.22.9496.25.0
411,833175.31.5400.23.4405.83.4313.22.6673.55.7
514,845207.11.4453.73.1263.11.8480.53.2740.85.0
619,558247.61.3495.02.5505.52.6561.92.9936.84.8
725,538258.41.0582.72.3684.22.7828.53.21249.44.9

c. Lumipulse G B-Amyloid Ratio (1-42/1-40)

Within-RunBetween-RunBetween-DayBetween-SiteTotal
Panel
memberMean
RatioSD%CVSD%CVSD%CVSD%CVSD%CV
10.5391.63.02.44.44.38.00.00.05.29.6
20.0570.12.00.11.50.23.00.00.00.23.9
30.0580.11.90.11.60.22.70.00.00.23.6
40.0580.11.90.22.60.22.80.00.00.34.3
50.0590.11.40.11.20.22.90.00.60.23.5
60.0640.12.00.00.70.23.50.00.00.34.1
70.0830.11.40.22.60.33.50.11.60.44.9
  1. Lumipulse G B-Amyloid Ratio (1-42/1-40) Precision Simulation:

The within-laboratory precision and site-tosite reproducibility of Lumipulse & B-Amyloid Ratio (1-42/1-40) was further evaluated by simulating percent positive calls at seven different ratio levels. For a bivariate ratio assay, the precision performance can be different at the same ratio value when the combinations for Lumipulse G B-Amyloid 1-42 and Lumipulse G B-Amyloid 1-40 results are different. Because the assay random measurement error y changes with the analyte concentration (%CV), different combinations for Lumipulse G B-Amyloid 1-42 and Lumipulse G B-Amyloid 1-40 results correspond to different random measurement random errors. In order to evaluate the precision characteristics of Lumipulse G B-Amyloid Ratio (1-42/1-40) numerical values under different combinations of {}-amyloid ;-4} and ß-amyloid ;-40, the variance profile of Lumipulse G B-Amyloid 1-42 and Lumipulse G B-Amyloid 1-40 was considered. The simulated probability of a positive call was estimated for the mean ratio of Panel member 1 7 tested in within-laboratory precision and site-to-site reproducibility above. Random measurement error was considered normally distributed and a generator of random normally distributed numbers was used. The simulated probability of positive call was

12

calculated for different combinations of 9-amyloid 1-42 and 0-amyloid 1-40 and the results met acceptable precision criteria.

2. Linearity:

The linearity of the Lumipulse G B-Amyloid 1-42 and Lumipulse G B-Amyloid 1-40 was evaluated in accordance with the CLSI guideline EP06, 2nd ed.. For each analyte, a high sample pool was diluted with a low sample pool to create a series of 22 dilutions for Bamyloid 1.47 or a series of 21 dilutions for B-amyloid 1-40 that span across the measuring ranges of the respective assays. Each sample dilution was measured in one run with four replicates using one reagent lot of each assay and the mean of the four replicates was calculated for each sample. A weighted regression analysis was performed to determine the predicted values. Percent deviation from linearity (DL) between the observed values and the best linear fit (predicted values) was calculated. The % DL was within ± 10% for each dilution level. The best linear regression fits are summarized in the table below:

AssayRange (pg/mL)Slope (95%CI)Intercept (95% CI)
Lumipulse G
β-Amyloid 1-4237.5 – 2203.51.02
(0.96 1.00)-23.4
(-47.5 0.62)1.00
Lumipulse G
β-Amyloid 1-40156.3 – 28450.31.03
(1.02 – 1.05)13.6
(-167.4 – 194.5)1.00

High-Dose Hook Effect:

The high-dose hook effect was evaluated for the Lumipulse G B-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 on the LUMIPULSE G1200 analyzer. The expected and measured concentrations were compared for the test samples spiked with a high concentration stock of the ß-amyloid 1-42 (159,072 pg/mL) and ß-amyloid 1-40 (168,426 pg/mL). The recovery was 98 103% for the Lumipulse G B-Amyloid 1-42 and 93 108% for the Lumipulse & B-Amyloid. No high-dose hook effect was observed up to approximately 159,000 pg/mL for the Lumipulse G B-Amvloid 1-42 and up to approximately 150,000 pg/mL for the Lumipulse G B-Amyloid 1-40.

3. Analytical Specificity/Interference:

1) Interference:

The effect of potential endogenous and exogenous substances to the Lumipulse G 3-Amyloid Ratio (1-42/1-40) and its component assays, Lumipulse & B-Amyloid 1-42 and Lumipulse G B-Amyloid 1-40, was evaluated in accordance with the CLSI guidelines EP07, 3rd ed, and EP37, Six pooled human CSF samples with various concentrations of B-amyloid 1-42 (303-1,980 pg/mL) and B-amyloid 1-40 (5,997-26,744 pg/mL) were

13

prepared for targeting the ß-Amyloid (1-42/1-40) ratios between 0.050-0.084. Test samples were created by spiking with the potential endogenous and exogenous interfering substances at two different levels. Control samples were spiked only with the appropriate solvent used to create the interfering substances panel. The ß-amyloid 1-42 and ß-amyloid 1-40 in the test samples and control samples were measured in three replicates. The recovery was calculated by comparing measurements of the test and control samples. No significant interference was observed (≤±10% difference of test from control) for the Lumipulse G B-Amyloid 1-42, the Lumipulse G B-Amyloid 1-40, and the Lumipulse G-Amyloid Ratio (1-42/1-40) up to the concentrations of the potential interfering substances tested as shown in the table below:

Endogenous InterferentConcentration
Biotin1300 ng/mL
Conjugated Bilirubin0.5 mg/dL
Free Bilirubin (unconjugated)0.5 mg/dL
Hemoglobin12.5 mg/dL
Human anti-mouse Antibodies, HAMA (IgG)1.5 ng/mL
Human Serum Albumin0.18 g/dL
Immunoglobulin A (IgA)0.018 g/dL
Immunoglobulin G (IgG)0.025 g/dL
Immunoglobulin M (IgM)0.005 g/dL
Rheumatoid Factor15.8 IU/mL
Triglycerides40 mg/dL
Whole Blood2.5% v/v
Exogenous InterferentConcentration
Acetaminophen20 mg/dL
Acetylcysteine15 mg/dL
Acetylsalicylic Acid100 mg/dL
Ampicillin100 mg/dL
Aripiprazole0.45 mg/dL
Ascorbic Acid30 mg/dL
Atorvastatin0.6 mg/dL
Caffeine10.8 mg/dL
Calcium Dobesilate20 mg/dL
Cefoxitin500 mg/dL
Chloramphenicol7.8 mg/dL
Clopidogrel18 mg/dL
Cyclosporine0.5 mg/dL
1Digoxin0.5 mg/dL
Donepezil (Aricept)3 mg/dL
Doxycycline1.8 mg/dL
Escitalopram1.5 mg/dL
Esomeprazole17 mg/dL
Furosemide9 mg/dL

14

Exogenous InterferentConcentration
Galantamine (Reminyl)25 mg/dL
Heparin500 U/dL
Hydrochlorothiazide12 mg/dL
Ibuprofen50 mg/dL
Levodopa2 mg/dL
Lisinopril50 mg/dL
Memantine (Namenda)25 mg/dL
Metformin400 mg/dL
Methyldopa2.25 mg/dL
Metoprolol90 mg/dL
Phenylbutazone32.1 mg/dL
Rifampicin6 mg/dL
Rivastigmine (Exelon)4.5 mg/dL
Simvastatin1.4 mg/dL
Quetiapine (Seroquel)90 mg/dL
Theophylline10 mg/dL
Warfarin7.5 mg/dL

2) Cross-reactivity:

The performance of the Lumipulse G B-Amyloid Ratio (1-42/1-40) in the presence of potential cross-reactive B-amyloid species is based on the effect of these potential crossreactants to the performance of the two component assays, the Lumipulse & S-Amyloid 1-42 and the Lumipulse G ß-Amyloid 1-40. The study was conducted by preparing four pooled human CSF samples with various concentration of B-amyloid 1-47 (349.5 1,808.9 pg/mL) and B-amyloid 1-40 (6,594.3 23,28.0 pg/mL) for challenging the B-amyloid (1-42)1-40, ratios between 0.056-0.077. Test samples were created by spiking with the potential cross-reactive B-amyloid species. Control samples were spiked only with the appropriate solvent used to create the cross-reactive B-amyloid species panel. The B-amyloid 142 and B-amyloid 1-40 in the test samples and control samples were measured in three replicates. The % cross-reactivity of each analyte was calculated by comparing measurements of the test and control samples using the equation below.

% cross-reactivity = {1- [{mean concentration (pg/mL-Test) / (mean concentration (pg/mL-Control)]} x 100

The mean % cross-reactivity values for each cross-reactive B-amyloid species tested at four levels of pooled human CSF samples are summarized for each analyte:

15

| Cross-Reactant | Test Concentration
(pg/mL) | Mean % Cross-Reactivity |
|-----------------|-------------------------------|-------------------------|
| β-amyloid 1-37 | 10,000 | 1.4 |
| β-amyloid 1-38 | 10,000 | 2.0 |
| β-amyloid 1-40 | 10,000 | -0.3 |
| β-amyloid 17-42 | 10,000 | -0.6 |
| β-amyloid 3-42 | 10,000 | 0.5 |
| β-amyloid 1-43 | 5,000 | -49.9 |

  • a. Lumipulse G B-Amyloid 1-42
  • b. Lumipulse G B-Amyloid 1-40

| Cross-Reactant | Test Concentration
(pg/mL) | Mean % Cross-Reactivity
(N=4) |
|-----------------|-------------------------------|----------------------------------|
| β-amyloid 1-37 | 200,000 | -1.3 |
| β-amyloid 1-38 | 200,000 | -2.0 |
| β-amyloid 1-39 | 200,000 | 0.0 |
| β-amyloid 1-42 | 200,000 | -2.0 |
| β-amyloid 11-40 | 200,000 | 6.1 |
| β-amyloid 17-40 | 200,000 | 13.4 |
| β-amyloid 3-40 | 200,000 | 1.6 |
| β-amyloid 1-43 | 200,000 | -0.6 |

The Lumipulse G B-Amyloid 1-42 assay showed no significant cross-reactivity (within ±10% difference of test from Control) up to 10,000 pg/mL for the tested cross-reactants except B-Amyloid 1-43. The Lumipulse G B-Amyloid 1-40 assay showed no noticeable cross-reactivity up to 200,000 pg/mL for the tested cross-reactants except B-amyloid 17.40.

    1. Assay Reportable Range:
      The Lumipulse G B-Amyloid Ratio (1-42/1-40) combines the results of the Lumipulse & B-Amyloid 1-42 assay and the Lumipulse G B-Amyloid 1-40 assay from the same patient CSF specimen into a numerical value in a range of 0.001 to 1.000.

The analytical measuring interval (AMI) for the two component assays are:

  • 38 2,200 pg/mL for the Lumipulse G B-Amyloid 1-42 .
  • 158 28.450 pg/mL for the Lumipulse G B-Amvloid 1-40 .
    1. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
      1. Traceability

The calibrators for use with the Lumipulse G ß-Amyloid 1-42 were prepared gravimetrically under internal quality system and the levels were assigned to the average

16

value of ERM-DA480/IFCC, ERM-DA481/IFCC and ERM-DA482/IFCC. The calibrators for use with Lumipulse G B-Amyloid 1-40 were prepared gravimetrically under internal quality system and are traceable to INNOTEST B-Amyloid 1-40.

    1. Stability
      Stability for the Lumipulse G B-Amyloid Ratio (1-42/1-40) is based on reagent stability for the individual component assays and the ratio. The Lumipulse & B-Amyloid Ratio (1-42/1-40) only uses the assay reagents which are stored under the claimed storage condition and within the claimed expiration date for each assay.

The stability of Lumipulse G B-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 were evaluated based on the following studies:

  • a. Reagent shelf-life:
    To evaluate reagent shelf-life stability, the Lumipulse & B-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 reagents were stored at 2-10°C. Three pooled human CSF samples with varying ß-amyloid 1-42 and ß-amyloid 1-40 concentrations were tested at Day 2, 3, 4, 5, 6, 7, 10, 13, 19, and 25 months after study initiation. The test results obtained at the different time points were compared to the test results at Day 0. Results showed that the Lumipulse & B-Amyloid 1-42 and the Lumipulse & B-Amyloid 1-40 reagents are stable at 2-10℃ for up to 19 months.

  • b. Reagent on-board stability:
    To evaluate reagent on-board stability, the Lumipulse G B-Amyloid 1-42 and the Lumipulse G 3-Amyloid 1-40 reagents were stored on-board the LUMIPULSE G 1200 System for a maximum of 30 days. Three pooled human CSF samples with varying B-amyloid 1-42 and B-amyloid 1-40 concentrations were tested at Day 1, 8, 15, and 31 after study initiation. The test results obtained at the different time points were compared to the test results at Day 1. Results showed that the Lumipulse G B-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 reagents are stable on-board the LUMIPULSE G 1200 System for up to 15 days.

  • c. Transport stability:
    To evaluate reagent transport stability, the Lumipulse C B-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 reagents were stored at two stressed conditions: (i) 25 ± 2℃ for 6 hours and (ii) -5 ± 2℃ for 6 hours. Three pooled human CSF samples with varying B-amyloid 1-42 and B-amyloid 1-40 concentrations were tested. The test results obtained at stressed conditions (i) and (ii) were compared to the test results at normal storage condition (2-10°C). The Lumipulse & B-Amyloid 1-42 and the Lumipulse G B-Amyloid 1-40 reagents are stable at 25 ℃ for up to 6 hours.

  • d. Sample stability:
    CSF samples were evaluated for stability under storage conditions and freeze/thaw cycles.

17

  • Sample storage stability: Seven CSF samples with varying B-amyloid 1-42 and Bi. amyloid 1-40 concentrations were freshly collected to challenge various B-amyloid 11-424 ratios ranged 0.041 to 0.108 . The B-amyloid 1-42 concentration and the Bamyloid 1.40 concentration were measured at Day 0. The CSF samples were stored at four different temperatures: (i) -80 °C (ii) -20 ± 10°C. (iii) 2-10°C, and (iv) 25 ± 2℃ for various storage durations.
  • ii. Sample freeze/thaw stability: Fifteen CSF samples with varying B-amyloid 1-22 and B-amyloid 1-40 concentrations were freshly collected to challenge various Bamyloid 11-42-1-40 ratios ranged 0.034 to 0.110 ). The B-amyloid 1-42 concentration and the B-amyloid 1-40 concentration were measured before the test samples were initially frozen. The CSF samples underwent repeated freeze/thaw cvcles: frozen at -80 ℃ and thawed at room temperature (15-25 ℃).

The CSF sample stability for the Lumipulse G B-Amyloid Ratio (1-42/1-40) were summarized in the table below.

CSF Sample Stability
Storage TemperatureFreeze/Thaw
Cycle
-80 °C-30 – -10°C2 – 8°C23 – 27°C
1 month2 weeks8 days48 hoursUp to 3

6. Detection Limit:

The detection capability for the Lumipulse G B-Amyloid Ratio (1-42/1-40) is based on the detection capability for its component assays, the Lumipulse G ß-Amyloid 1-42 and Lumipulse G B-Amyloid 1-40. The results below the lower limit of the measuring interval of each assay are not imported and do not yield a ratio result.

The detection capabilities for the Lumipulse G ß-Amyloid 1-42 and Lumipulse G ß-Amyloid 1-40 assays were evaluated by Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) studies in accordance with the CLSI guideline EP17-A2. The studies evaluated two lots of Lumipulse G B-Amyloid 1-42 and two lots of Lumipulse G B-Amyloid 1-40 on one LUMIPULSE & 1200 System.

    1. Lumipulse G B-Amyloid 1-42:
      For LoB, a set of four analyte-depleted CSF samples was tested in 10 replicates per run, two runs per day for three days to reach a total of 60 measurements per sample for each reagent lot. The LoB was determined for each lot as 2.18 pg/mL and 1.06 pg/mL. The claimed LoB is the higher value from the two lots as 2.18 pg/mL.

For LoD, a set of seven low level CSF samples with target concentrations of B-amyloid 1. 42 between 10 22 pg/mL was tested in 10 replicates per run, two runs per day for three days to reach a total of 60 measurements per sample for each reagent lot. The LoD was calculated using a precision profile approach. The LoD was determined for each lot as

18

10.6 pg/mL and 11.6 pg/mL. The claimed LoD is the higher value from both lots as 11.6 pg/mL.

The LoQ was estimated based on the lowest value of the test samples at which the variability was below the 20 %CV. The LoO was determined for each lot and the higher LoQ value from both lots was determined. The claimed LoQ was 38 pg/mL at the low limit of the analytical measuring interval of Lumipulse G B-Amyloid 1-42.

    1. Lumipulse G B-Amyloid 1-40:
      For LoB, a set of four analyte-depleted CSF samples was tested in 10 replicates per run, two runs per day for three days to reach a total of 60 measurements per sample for each reagent lot. The LoB was determined for each lot as 0.97 pg/mL and 0.35 pg/mL. The claimed LoB is the higher value from the two lots as 0.97 pg/mL.

For LoD, a set of six low level CSF samples with target concentrations of B-amyloid 1-40 between 25-50 pg/mL was tested in 10 replicates per run. two runs per day for three days to reach a total of 60 measurements per sample for each reagent lot. The LoD was calculated using precision profile. The LoD was determined for each lot as 29.4 pg/mL and 33.0 pg/mL. The claimed LoD is the higher value from both lots as 33.0 pg/mL.

The claimed LoQ was 158 pg/mL (precision was %CV13 | 83.3% | 73.9% | 81.4% | 81.7% | 76.4% | 83.9% | 78.8% |
| | | (15) | (82) | (48) | (85) | (152) | (78) | (230) |
| Visual
Amyloid PET Read | Positive | 38.9% | 42.3% | 55.9% | 90.4% | - | - | 68.2% |
| | | (7) | (47) | (33) | (94) | | | (199) |
| | Negative | 61.1% | 57.7% | 44.1% | 9.6% | - | - | 31.8% |
| | | (11) | (64) | (26) | (10) | | | (93) |

  • including Hawaiian / Pacific Islander

** Mini-Mental State Examination (MMSE)

The above table indicates the majority of the study subjects (80.5%, 235/292) were considered in absence of clinical presentation of dementia based on the Mini-Mental State Examination (MMSE) score between 24 30. The APOE 23e4 and e4e4 genotype, which more often are shown in patients with AD dementia was found in 39% (114/292) and 10.9% (32/292), respectively, of the study population. Most study subjects received >13 years of education (78.8%, 230/292). The diagnostic groups did not differ with respect to age or sex. Patients with AD dementia had lower MMSE scores than those with MCI and SCD. Patients with AD dementia also showed more often a positive PET scan than MCI patients, who in turn had more often a positive PET scan than those with SCD. The prevalence of amyloid positivity among AD was 90.4% (94/104) and it was 55.9% (33/59) and 42.3% (47/11) among those with LMCI and EMCI, respectively.

Results:

    1. Clinical Performance

22

To estimate the clinical performance characteristics of the Lumipulse & B-Amyloid Ratio (1-42/1-40), the test result was compared to the visual amyloid PET read for each patient. The data are summarized in the table below.

Visual Amyloid PET Read
PositiveNegativeTotal
Lumipulse G
β-Amyloid Ratio
(1-42/1-40)Positive (Ratio $\le$ 0.058)1716177
Likely positive
(0.059 $\le$ Ratio $\le$ 0.072)13922
Negative (Ratio $\ge$
0.073)157893
Total19993292

The performance estimates of the Lumipulse G B-Amyloid Ratio (1-42/1-40) is described by PET-positive predictive value (indicated as Predictive Value, PV) and frequency of positive test results of the tested samples (indicated as Frequency of Results, FR) for positive, likely positive and negative test results and it is summarized in the table below.

| | | Visual Amyloid
PET Read | | Predictive
Value %
(n/N)
(95% CI) | Frequency of
Results %
(n/N)
(95% CI) |
|--------------------------------------------------|-----------------------------------------------|----------------------------|-------|--------------------------------------------|------------------------------------------------|
| | | Positive | Total | | |
| Lumipulse G
β-Amyloid
Ratio
(1-42/1-40) | Positive
(Ratio ≤ 0.058) | 171 | 177 | 96.6%
(171/177)
(92.8%; 98.4%) | 60.6%
(177/292)
(54.9%; 66.1%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 13 | 22 | 59.1%
(13/22)
(38.7%; 66.7%) | 7.5%
(22/292)
(5.0%; 11.1%) |
| | Negative
(Ratio ≥ 0.073) | 15 | 93 | 16.1%
(15/93)
(10.0%; 24.9%) | 31.8%
(93/292)
(26.8%; 37.4%) |
| | Total | 199 | 292 | | Prevalence of PET positive = 68.2% |

Confidence intervals were calculated as confidence intervals for binomial proportions using a Wilson score method.

In conclusion, the data of clinical performance study support that positive Lumipulse G 8-Amyloid Ratio (1-42/1-40) results are consistent with PET scan positive results and negative Lumipulse G B-Amvloid Ratio (1-42/1-40) results are consistent with PET scan negative results.

23

2) Sub-group Analysis

a. Diagnostic group

The performance measures of the Lumipulse G (3-Amyloid Ratio (1-42/1-40) stratified by diagnostic groups are summarized in the table below.

| | | Visual Amyloid PET
Read | | Predictive
Value % | Frequency of
Results % |
|-----------------|-----------------------------------------------|----------------------------|----|------------------------------------|-------------------------------------|
| Diagnosis | Test Result | Positive (n) | N | (n/N)
(95% CI) | (n/N)
(95 %CI) |
| AD
(N=104) | Positive
(Ratio ≤ 0.058) | 89 | 90 | 98.9%
(89/90)
(94.0%; 99.8%) | 86.5%
(90/104)
(78.7%; 91.8%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 3 | 4 | 75.0%
(3/4)
(30.1%; 95.4%) | 3.8%
(4/104)
(1.5%; 9.5%) |
| | Negative
(Ratio ≥ 0.073) | 2 | 10 | 20.0%
(2/10)
(5.7%; 51.0%) | 9.6%
(10/104)
(5.3%; 16.8%) |
| LMCI
(N=59) | Positive
(Ratio ≤ 0.058) | 32 | 33 | 97.0
(32/33)
(84.7%; 99.5%) | 55.9%
(33/59)
(43.3%; 67.8%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 4 | 5 | 80.0%
(4/5)
(37.6%; 96.4%) | 8.5%
(5/59)
(3.7%; 18.4%) |
| | Negative
(Ratio ≥ 0.073) | 4 | 21 | 19.0
(4/21)
(7.7%; 40.0%) | 35.6%
(21/59)
(24.6%; 48.3%) |
| EMCI
(N=111) | Positive
(Ratio ≤ 0.058) | 43 | 47 | 91.5%
(43/47)
(80.1%; 96.6%) | 42.3
(47/111)
(33.6%; 51.6%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 5 | 10 | 50.0%
(5/10)
(23.7%; 76.3%) | 9.0%
(10/111)
(5.0%; 15.8%) |
| | Negative
(Ratio ≥ 0.073) | 8 | 54 | 14.8%
(8/54)
(7.7%; 26.6%) | 48.6%
(54/111)
(39.6%; 57.8%) |
| SCD
(N=18) | Positive
(Ratio ≤ 0.058) | 7 | 7 | 100.0%
(7/7)
(64.6%; 100.0%) | 38.9%
(7/18)
(20.3%; 61.4%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 1 | 3 | 33.3%
(1/3)
(6.1% 79.2%) | 16.7%
(3/18)
(5.8%; 39.2%) |
| | Negative
(Ratio ≥ 0.073) | 1 | 8 | 12.5%
(1/8)
(2.2%: 47.1%) | 44.4%
(8/18)
(24.6%: 66.3%) |

24

Across diagnostic groups, the Predictive values (PV) were highest for the positive test results with a decrease for the likely positive test results and lowest in the negative test results within each diagnostic group. The estimate of PV was consistently high for AD (99%), LMCI (97%), EMCI (92%), and SCD (100.0%). for the positive test results. The estimate of PV for the negative test results was 12.5% for SCD, 14.8% for EMCI. 19.0% for LMCI, and 20.0% for AD in each diagnostic group. The % of the likely positive test results within each diagnostic group was 3.8% for AD, 8.5% for LMCl. 9.0% for EMCI. and 16.7% for SCD that comprised 7.5% across all diagnostic groups.

b. Sex

The performance measures of the Lumipulse G B-Amyloid Ratio (1-42/1-40) were also stratified by sex and summarized in the table below.

| | Test Result | Visual Amyloid
PET Read | | Predictive
Value %
(n/N)
(95% CI) | Frequency of
Results %
(n/N)
(95% CI) |
|------------------------------|-----------------------------------------------|----------------------------|-----|--------------------------------------------|------------------------------------------------|
| Sex | | Positive (n) | N | | |
| Male
(N=166) | Positive
(Ratio ≤ 0.058) | 98 | 102 | 96.1%
(98/102)
(90.3%; 98.5%) | 61.4%
(102/166)
(53.9%; 68.5%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 7 | 11 | 63.6%
(7/11)
(35.4%; 84.8%) | 6.6%
(11/166)
(3.7%; 11.5%) |
| | Negative
(Ratio ≥ 0.073) | 12 | 53 | 22.6%
(12/53)
(13.5%; 35.5%) | 31.9%
(53/166)
(25.3%; 39.4%) |
| Female
(N=126) | Positive
(Ratio ≤ 0.058) | 73 | 75 | 97.3%
(73/75)
(90.8%; 99.3%) | 59.5%
(75/126)
(50.8%; 67.7%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 6 | 11 | 54.5%
(6/11)
(28.0%; 78.7%) | 8.7%
(11/126)
(4.9%; 15.0%) |
| | Negative
(Ratio ≥ 0.073) | 3 | 40 | 7.5%
(3/40)
(2.6%; 19.9%) | 31.7%
(40/126)
(24.3%; 40.3%) |
| Age (years) | Result | Visual Amyloid PET
Read | | Predictive
Value % | Frequency of
Results % |
| | | Positive (n) | N | (n/N)
(95% CI) | (n/N)
(95% CI) |
| 55-59
(N=12) | Positive
(Ratio ≤ 0.058) | 6 | 7 | 85.7%
(6/7)
(48.7%; 97.4%) | 58.3%
(7/12)
(32.0%; 80.7%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 0 | 2 | 0.0%
(0/2)
(0.0%; 65.8%) | 16.7%
(2/12)
(4.7%; 44.8%) |
| | Negative
(Ratio ≥ 0.073) | 2 | 3 | 66.7%
(2/3)
(20.8%; 93.9%) | 25.0%
(3/12)
(8.9%; 53.2%) |
| 60-69
(N=85) | Positive
(Ratio ≤ 0.058) | 41 | 43 | 95.3%
(41/43)
(84.5%; 98.7%) | 50.6%
(43/85)
(40.2%; 61.0%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 2 | 5 | 40.0%
(2/5)
(11.8%; 76.9%) | 5.9%
(5/85)
(2.5%; 13.0%) |
| | Negative
(Ratio ≥ 0.073) | 5 | 37 | 13.5
(5/37) (5.9%;
28.0%) | 43.5%
(37/85)
(33.5%; 54.1%) |
| 70-79
(N=146) | Positive
(Ratio ≤ 0.058) | 95 | 97 | 97.9%
(95/97)
(92.8%; 99.4%) | 66.4%
(97/146)
(58.4%; 73.6%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 8 | 11 | 72.7%
(8/11)
(43.4%; 90.3%) | 7.5%
(11/146)
(4.3%; 13.0%) |
| | Negative
(Ratio ≥ 0.073) | 4 | 38 | 10.5%
(4/38)
(4.2%; 24.1%) | 26.0%
(38/146)
(19.6%; 33.7%) |
| ≥80
(N=49) | Positive
(Ratio ≤ 0.058) | 29 | 30 | 96.7%
(29/30)
(83.3%; 99.4%) | 61.2%
(30/49)
(47.2%; 73.6%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 3 | 4 | 75.0%
(3/4)
(30.1%; 95.4%) | 8.2%
(4/49)
(3.2%; 19.2%) |
| | Negative
(Ratio ≥ 0.073) | 4 | 15 | 26.7%
(4/15)
(10.0%; 52.0%) | 30.6%
(15/49)
(19.5%; 44.5%) |
| Race | Result | Visual Amyloid PET
Read | | Predictive
Value % | Frequency of
Results % |
| | | Positive (n) | N | (n/N)
(95% CI) | (n/N)
(95% CI) |
| | | | | 96.4% | 62.4% |
| | Positive
(Ratio ≤ 0.058) | 163 | 169 | (163/169)
(92.5%; 98.4%) | (169/271)
(56.5%; 67.9%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 12 | 20 | 60.0%
(12/20)
(38.7%; 78.1%) | 7.4%
(20/271)
(4.8%; 11.1%) |
| White
(N=271) | Negative
(Ratio ≥ 0.073) | 15 | 82 | 18.3%
(15/82)
(11.4%; 28.0%) | 30.3%
(82/271)
(25.1%; 36.0%) |
| | | | | 100.0% | 42.9% |
| | Positive
(Ratio ≤ 0.058) | 3 | 3 | (3/3)
(43.9%; 100.0%) | (3/7)
(15.8%; 75.0%) |
| Asian
(N=7) | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 0 | 0 | - | 0.0%
(0/7)
(0.0%; 35.4%) |
| | Negative (Ratio ≥
0.073) | 0 | 4 | 0.0%
(0/4)
(0.0%; 49.0%) | 57.1%
(4/7)
(25.0%; 84.2%) |
| | | | | 100.0% | 40.0% |
| | Positive
(Ratio ≤ 0.058) | 2 | 2 | (2/2)
(34.2%; 100.0%) | (2/5)
(11.8%; 76.9%) |
| African
American
(N=5) | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 0 | 0 | - | 0.0%
(0/5)
(0.0%; 43.4%) |
| | Negative
(Ratio ≥ 0.073) | 0 | 3 | 0.0%
(0/3)
(0.0%; 56.1%) | 60.0%
(3/5)
(23.1%; 88.2%) |
| Others*
(N=9) | Positive
(Ratio ≤ 0.058) | 3 | 3 | 100.0%
(3/3)
(43.9%; 100.0%) | 33.3%
(3/9)
(12.1%; 64.6%) |
| | Likely positive
(0.059 ≤ Ratio ≤
0.072) | 1 | 2 | 50.0%
(1/2)
(9.5%; 90.5%) | 22.2%
(2/9)
(6.3%; 54.7%) |
| | Negative
(Ratio ≥ 0.073) | 0 | 4 | 0.0%
(0/4)
(0.0%; 49.0%) | 44.4%
(4/9)
(18.9%; 73.3%) |

The estimates of PV for the positive test results were similar for males (96%) and females (97%). The estimates of PV for the likely positive test results were 63.6% and 54.5% for males and females, respectively. The estimates of PV for the negative test results were 22.6% for males and 7.5% for females. The % of the likely positive test results was similar for males (6.6%) and females (8.7%).

c. Age

The performance measures of the Lumipulse G B-Amyloid Ratio (1-42/1-40) were also stratified by age and summarized in the table below.

25

Across age groups, the Predictive values (PV) were highest for the positive test results with a decrease for the likely positive test results and lowest in the negative test results within each age group. The estimate of PV was 86% for age group 55 59. IV was above 95% for age groups 60 and older. The % of the likely positive test results was highest 16.7% for age group 55-59 and 6-8% for age groups 60 and older.

26

d. Race

The performance measures of the Lumipulse G ß-Amyloid Ratio (1-42/1-40) were also stratified by race and summarized in the table below.

  • including Hawaiian / Pacific Islander

27

Across race groups, the Predictive values (PV) were highest for the positive test results with a decrease for the likely positive test results and lowest in the negative test results within each race group. For White patients, the estimate of PV was 96.4%. 60.0%, and 18.3% for the positive test results, likely positive test results, and negative results, respectively. The % of likely positive results was 7.4% for the White group. The estimates of PV and % of the likely positive results for Asian, African American, and other races had high uncertainty due to limited numbers of patients in the study.

D Clinical Cut-Off:

Refer to Assay Cut-off

Reference Interval / Expected Values: E

  • l) Reference Interval
    A reference interval of the Lumipulse G B-Amyloid Ratio (1-42/1-40) in cogmitively normal subjects aged 55 years and above was determined in accordance with the CLSI guideline EP28-A3c. Cognitively normal subjects were defined as individuals who did not meet clinical criteria for a diagnosis of AD and MCI and had an MMSE of 29 30. A total of 281 human CSF samples including 229 samples from a longitudinal population-based study of normal aging (BIOCARD study) and 52 samples from the real-world clinics at the Washington University Medical Campus. After excluding 69 samples from subjects aged