K Number
DEN180057
Device Name
IB-Stim
Date Cleared
2019-06-07

(225 days)

Product Code
Regulation Number
876.5340
Type
Direct
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IB-Stim is a percutaneous electrical nerve field stimulator (PENFS) system intended to be used in patients 11-18 years of age with functional abdominal pain associated with irritable bowel syndrome (IBS). The IB-Stim is intended to be used for 120 hours per week up to 3 consecutive weeks, through application to branches of Cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination, as an aid in the reduction of pain when combined with other therapies for IBS.

Device Description

The IB-Stim is a device that delivers electrical stimulation through percutaneous electrodes to areas usually innervated by branches of occipital nerves and cranial nerves V, VII, IX and X that are located by transillumination. The device consists of (1) a percutaneous electrical nerve field stimulator (PENFS), (2) a multi-pin wire harness percutaneous electrode array, and (3) a pen light for use in the transillumination technique that aids in positioning of the percutaneous electrodes (Figure 1).

The wire harness percutaneous electrode array consists of 4 leads. The 1-1-1-4 configuration consists of three single-needle leads, and one 4-needle array (Figure 2).

The stimulator is placed behind the ear and the percutaneous electrodes are positioned utilizing the transillumination function of the device (Figure 1). The transillumination technique assists in the visualization of the vasculature of the ear to aid in the placement of the percutaneous electrodes near the nerve branches in the ear.

AI/ML Overview

The provided text describes the regulatory evaluation of the IB-Stim device, a non-implanted nerve stimulator for functional abdominal pain relief associated with Irritable Bowel Syndrome (IBS) in adolescents. The document outlines the studies conducted to establish the device's safety and effectiveness.

It is important to note that the IB-Stim is a medical device performing nerve stimulation, not an AI/ML powered device. Therefore, typical acceptance criteria and study designs for AI/ML devices (e.g., mMRMC studies, stand-alone algorithm performance, ground truth establishment through expert consensus or pathology for image classification, separate training and test sets the way an AI/ML device would need them, etc.) are NOT directly applicable to this device.

The "acceptance criteria" for the IB-Stim are essentially the demonstration of safety and clinical effectiveness as outlined by the FDA's De Novo classification process, largely evidenced through a randomized, double-blind, sham-controlled clinical trial and various non-clinical tests.

Here's an attempt to structure the information according to your request, adapting the concepts to fit a medical device rather than an AI/ML system:


Acceptance Criteria and Device Performance for IB-Stim

The acceptance criteria for the IB-Stim device are based on demonstrating its safety and effectiveness for its intended use, as required by FDA for a De Novo classification. This is primarily achieved through non-clinical testing and a pivotal clinical trial.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria/TestsReported Device Performance/Results
Safety - Non-ClinicalBiocompatibility (for patient-contacting materials)All results demonstrated acceptable performance for cytotoxicity, sensitization, and irritation, leveraging prior evaluations for identical materials in similar devices (DEN170018, K140530).
Sterility (for percutaneous electrode array)Achieved a sterility assurance level of 10^-6 kGy using VDMAX25 method (ISO 11137-2:2007).
Shelf Life (sterility, package integrity, functionality)Validated for a labeled shelf life of (b)(4) using accelerated-aged samples (ISO 11607-1:2009, ASTM F1980-07:2011).
Electromagnetic Compatibility (EMC)Conformed to IEC 60601-1-2.
Electrical SafetyConformed to IEC 60601-1 and IEC 60601-2-10.
Software Verification and ValidationDocumentation provided in K140530; software classified as "Minor" level of concern.
Electrical Bench Performance (verification of pulse train, duty cycle, max operating time)Validated that electrical performance met device specifications (leveraged from K140530).
Safety - ClinicalAbsence of Serious Adverse Events (SAEs)No SAEs recorded in any subject during the clinical study.
Incidence of Non-Serious Adverse Events (AEs)Ten patients reported AEs: Ear discomfort (n=6), Adhesive allergy (n=3), Syncope (n=1). These were considered typical for the device type.
Effectiveness (Clinical Trial)Primary Outcome: Change from baseline to end of Week 3 in worst abdominal pain (IBS subgroup)IB-Stim group had a significantly lower worst pain score (median 5.0, IQR 4.0-7.0) compared to sham group (median 7.0, IQR 5.0-9.0). 59% of IB-Stim vs. 26% of sham patients showed >30% reduction.
Primary Outcome: Change from baseline to end of Week 3 in composite Pain Frequency-Severity-Duration (PFSD) score (IBS subgroup)IB-Stim group had a significantly lower PFSD score (median 7.5, IQR 3.6-14.4) compared to sham group (median 14.4, IQR 4.5-39.2).
Secondary Outcome: Global symptom improvement based on Symptom Response Scale (SRS), Week 3 (IBS subgroup)Median SRS score of 3 (IQR: 2-4) in IB-Stim group vs. 0 (IQR: 0-2) in sham group. 81% of IB-Stim vs. 26% of sham patients reported overall symptom improvement (SRS ≥ 2).
Secondary Outcome: ≥30% improvement in usual abdominal pain, Week 3 (IBS subgroup)52% of IB-Stim vs. 30% of sham patients had >30% reduction in usual abdominal pain.
Risk MitigationImplementation of Special Controls (e.g., biocompatibility standards, electrical safety, labeling warnings)Special controls were deemed sufficient to mitigate identified risks, as evidenced by the granting of the De Novo request.

2. Sample Size and Data Provenance

  • Clinical Trial Test Set Sample Size:
    • Total randomized: 115 subjects (60 active treatment, 55 sham treatment).
    • Completed the trial (primary analysis cohort): 104 subjects (57 active, 47 sham).
    • IBS Subgroup Analysis (post-hoc, for final indication): 51 subjects met Rome III criteria for IBS at baseline. 50 of these completed the study (27 active IB-Stim, 23 sham). This IBS subgroup formed the basis for the specific indication granted.
  • Data Provenance: The study was a randomized, double-blind, sham-controlled clinical trial. It is stated to be published in Lancet Gastroenterol Hepatol in 2017, suggesting it was a prospective clinical trial. The location of the clinical trial (country of origin) is not explicitly stated in the provided text, but the authors listed (K. Kovacic, etc.) typically indicate US-based research centers.

3. Number of Experts and Qualifications for Ground Truth

  • The "ground truth" in this context refers to the patient-reported outcomes of pain and symptom scores (e.g., worst pain, PFSD, SRS). These are direct measures from the participants themselves, not interpretations by experts.
  • The diagnosis of functional abdominal pain, including IBS, was based on established clinical criteria (Rome III criteria for IBS) by the clinicians conducting the study. The number or specific qualifications of these clinicians are not detailed as "experts establishing ground truth" in the way it would be for, e.g., image annotation.

4. Adjudication Method for the Test Set

  • Adjudication methods (e.g., 2+1, 3+1) are typically used for establishing ground truth in AI/ML model training or evaluation where human reviewers provide labels or assessments that need to be reconciled.
  • This was a clinical trial with patient-reported outcomes. While there would be standard operating procedures for data collection and monitoring, there was no "adjudication" in the sense of reconciling disagreeing expert labels. The "ground truth" was the direct patient report and clinical diagnosis.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • An MRMC study evaluates the performance of multiple human readers on multiple cases, often comparing performance with and without AI assistance.
  • No MRMC study was performed as this is a medical device for direct patient treatment, not an AI-based diagnostic tool that assists human readers in interpreting data or images. The study design was a standard randomized controlled trial comparing the device to a sham control.

6. Standalone (Algorithm Only) Performance

  • "Standalone performance" refers to the performance of an algorithm without human-in-the-loop.
  • Not applicable to this device. The IB-Stim is a physical medical device that delivers electrical stimulation to a patient. Its functionality is inherently tied to its application to a patient's body under clinical guidance. There is no "algorithm only" component that could be evaluated for standalone performance.

7. Type of Ground Truth Used

  • The primary ground truth for effectiveness was patient-reported outcome data (pain scores, symptom improvement) collected during the clinical trial.
  • The patient cohort was defined by clinical diagnostic criteria (Rome III for IBS) established by medical professionals.

8. Sample Size for the Training Set

  • Not applicable in the context of AI/ML. The IB-Stim design and functionality are based on medical principles and prior devices, not on machine learning from a "training set." The clinical trial data (the "test set" in your original framework) was used to demonstrate safety and effectiveness for regulatory approval.

9. How Ground Truth for Training Set was Established

  • Not applicable as there is no "training set" for an AI/ML model. The device's design and parameters (e.g., electrical characteristics, duty cycle) were based on existing medical understanding of nerve stimulation and were verified through bench testing. The clinical trial then assessed the device's efficacy in a target patient population.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR IB-STIM

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Non-implanted nerve stimulator for functional abdominal pain relief. A nonimplanted nerve stimulator for functional abdominal pain relief is a device that stimulates nerves remotely from the source of pain with the intent to relieve functional abdominal pain. This generic type of device does not include devices designed to relieve pelvic pain.

NEW REGULATION NUMBER: 21 CFR 876.5340

CLASSIFICATION: Class II

PRODUCT CODE: QHH

BACKGROUND

DEVICE NAME: IB-Stim

SUBMISSION NUMBER: DEN180057

DATE DE NOVO RECEIVED: October 25, 2018

Innovative Health Solutions (IHS), Inc. CONTACT: 829 South Adams St. Versailles. IN 47042

INDICATIONS FOR USE

The IB-Stim is indicated as follows:

The IB-Stim is a percutaneous electrical nerve field stimulator (PENFS) system intended to be used in patients 11-18 years of age with functional abdominal pain associated with irritable bowel syndrome (IBS). The IB-Stim is intended to be used for 120 hours per week up to 3 consecutive weeks, through application to branches of Cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination, as an aid in the reduction of pain when combined with other therapies for IBS.

LIMITATIONS

For prescription use only.

{1}------------------------------------------------

The device is contraindicated for use by patients with cardiac pacemakers, hemophilia, and psoriasis vulgaris.

The device should only be applied to healthy, clean, intact skin.

The device therapy is limited to 120 hours, after which the device should be disposed.

The appliance is splash-proof but not watertight. When showering, the device must not be allowed to come into direct contact with water.

Treatment protocols are for 3 consecutive weeks, and not to exceed 4 weeks

PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.

DEVICE DESCRIPTION

The IB-Stim is a device that delivers electrical stimulation through percutaneous electrodes to areas usually innervated by branches of occipital nerves and cranial nerves V, VII, IX and X that are located by transillumination. The device consists of (1) a percutaneous electrical nerve field stimulator (PENFS), (2) a multi-pin wire harness percutaneous electrode array, and (3) a pen light for use in the transillumination technique that aids in positioning of the percutaneous electrodes (Figure 1).

The wire harness percutaneous electrode array consists of 4 leads. The 1-1-1-4 configuration consists of three single-needle leads, and one 4-needle array (Figure 2).

Image /page/1/Picture/9 description: The image shows two close-up shots of a person's ear with medical devices attached. In the left image, a gloved hand is near the ear, which is glowing with a bright light. A small electronic device is attached behind the ear. The right image shows a similar device attached to the ear, with wires and small pads connected to the skin around the ear.

Figure 1: (Left) Visualization of neurovasculature of the ear using transillumination technique. (Right) Representative image of the placement of the IB-Stim device.

{2}------------------------------------------------

Image /page/2/Picture/0 description: This image shows a medical device with four leads. The leads are thin and gray, and they are connected to a clear, curved device. The device has some electronic components visible inside. The text "FIG. 2" is visible in the bottom right corner of the image.

Figure 2: PENFS System

The stimulator is placed behind the ear and the percutaneous electrodes are positioned utilizing the transillumination function of the device (Figure 1). The transillumination technique assists in the visualization of the vasculature of the ear to aid in the placement of the percutaneous electrodes near the nerve branches in the ear. The system specifications are listed in Table 1.

Device Technology Description
General DeviceCharacteristicsIB-Stim
Power supply1 x 3V battery (Type CR1220 Li)
OutputMax 3.2V @ 1kΩ -10kΩ
Total duration oftreatment5 x 24hr
Weight including battery5g
Dimensions36mm x 16mm x 7mm

Table 1: System Specifications

The following components are part of a convenience kit sold with the IB-Stim:

    1. Tweezers
    1. Transilluminator (pen light)
    1. Alcohol swab
    1. Surgical marker
    1. Adhesive bandages
    1. Foam adhesive
    1. Multi-pin wire harness percutaneous electrode array

The IB-Stim is similar to the NSS-2 BRIDGE device previously granted in DEN170018 for a different intended use. It is also like the Electronic Auricular Device (EAD) electroacupuncture device cleared in K140530, which is the precursor device to both the NSS-2 BRIDGE and IB-Stim device. The IB-Stim duty cycle is identical to the duty cycle cleared in the Electronic Auricular Device (EAD) electroacupuncture device (K140530).

SUMMARY OF NONCLINICAL/BENCH STUDIES

{3}------------------------------------------------

This De Novo request refers to the nonclinical bench testing, biocompatibility, electrical safety, electromagnetic compatibility, shelf life, sterility, and software testing provided in DEN170018 for the NSS-2 BRIDGE device. The DEN170018 used a significant amount of nonclinical testing submitted for the EAD device in K140530.

The sections below summarize testing used from DEN170018 and K140530 to support the safety and effectiveness of this device for its intended use. The technological changes detailed above do not change any results of the prior non-clinical testing. Additional information on these changes and why they do not adversely impact the safety of the device are discussed below.

BIOCOMPATIBILITY/MATERIALS

Table 2, below, summarizes the patient-contacting materials for the IB-Stim device and references the source of any biocompatibility evaluations.

Table 2: List of Patient-Contacting Materials
Patient ContactingDevice ComponentNature of TissueContactDuration of Tissue Contact
TweezersNo skin contactNo skin contact
Transilluminator penlightBehind earLess than 1 minuteMaterials used in the pen light areidentical to the materials used inin the granted NSS-2 BRIDGE(DEN170018).
Kitted Bag containingFixation PlastersOver needle arrayson front surface ofearUp to 120 hoursMaterials used for the fixationplasters are identical to thematerials used in in the grantedNSS-2 BRIDGE (DEN170018)and cleared EAD.
Surgical MarkerOn front surface ofearUp to 120 hoursMaterials used for the surgicalmarker are identical to thematerials used in in the grantedNSS-2 BRIDGE (DEN170018).
Multi-pin wire harnesspercutaneouselectrode arrayImplantedpercutaneously onfront surface of earUp to 120 hoursMaterials used in the needlearrays are identical to thematerials used in the grantedNSS-2 BRIDGE (DEN170018).
Foam AdhesivesBehind earUp to 120 hours

Table 2: List of Patient-Contacting Materials

{4}------------------------------------------------

Foam adhesive material isidentical to the materials used inin the granted NSS-2 BRIDGE(DEN170018).
------------------------------------------------------------------------------------------------------------------

The tweezers, transilluminator (pen light), alcohol swab, and surgical marker are Class I, 510(k)-exempt products packaged as part of the convenience kit.

The fixation plasters and foam adhesive were tested for cytotoxicity, sensitization, and irritation in K140530 per FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. All results demonstrated acceptable performance.

The multi-pin percutaneous electrode array for the IB-Stim is identical to the needle array granted in DEN170018.

STERILITY AND SHELF LIFE

Sterilization validation was conducted on the wire harness percutaneous electrode arrays using the VDMAX25 method according to ISO 11137-2:2007 "Sterilization of health care products -Radiation -- Part 2: Establishing the sterilization dose." The device achieved a sterility assurance kGy. The device fulfilled the requirements of sterility level of 10-6 with (b) (4) according to ISO 11737-2:2009 "Sterilization of medical devices -- Microbiological methods --Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process."

The master packaging for the device is comprised of three parts:

    1. The main tray is a vacuum formed single piece of plastic approximately 12 inches in length and 8 inched wide. The depth of the packaging is ¾ of an inch. There are five vacuum formed compartments that house the internal components in a secure fashion.
    1. The cover for the packaging is clear sheet of film that is heat-sealed to ensure proper closure and protection of vital components.
    1. The sterile sub-pack that houses the wire harness and needle array assembly. It is a 4 ½ x 1-inch vacuum formed enclosure consisting of a top and bottom that is then vacuum sealed inside a plastic medical grade bag.

The performance and stability of the device's packaging system during sterilization, distribution and labeled shelf life of (4) (4) were validated in accordance with ISO 11607-1:2009, using accelerated-aged device samples. Accelerated aging of the device was performed according to ASTM F1980-07:2011. Validation was based on packaging material qualification, according to

{5}------------------------------------------------

ISO 11607-1:2009, and the machine qualifications of the sealing processes, according to ISO 11607-2:2006.

ELECTROMAGNETIC COMPATIBILITY & ELECTRICAL SAFETY

Electrical safety and EMC information is identical to the NSS-2 BRIDGE device granted in DEN170018. In DEN170018, the electrical safety and EMC information was leveraged from information previously provided for the EAD device cleared under K140530 because there were no changes to the electrical components of the device. The only change associated with electrical safety is an alteration to the duty cycle of the IB-Stim to be consistent with the original duty cycle of the EAD device (K140530).

The device conformed to the following electromagnetic compatibility, electrical, and thermal safety standards:

  • IEC 60601-1 Medical Electrical Equipment; Part 1: General Requirements for Basic . Safety and Essential Performance.
  • IEC 60601-1-2 Medical Electrical Equipment: Part 1-2: General Requirements for . Safety - Section 2: Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
  • IEC 60601-2-10 Medical Electrical Equipment; Part 2-10: Particular Requirements for . the Safety of Nerve and Muscle Stimulators.

SOFTWARE

Software verification and validation testing were conducted, and documentation was provided in K140530 as recommended in FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," issued May 11, 2005. The software for this device was considered a "Minor" level of concern.

PERFORMANCE TESTING - BENCH

The electrical bench testing is identical to the NSS-2 BRIDGE device granted in DEN170018. In DEN170018. this electrical bench testing was leveraged from information previously provided for the EAD device cleared under K140530. The following tests were performed under a 1k12 resistance to validate that the electrical performance met the specifications of the device:

  • . Verification of the temporal characteristics and amplitude of the pulse train through the leads and percutaneous electrodes
  • . Verification of the pulse train duty cycle
  • . Verification that the device does not exceed the maximum operating time of 120 hours

{6}------------------------------------------------

SUMMARY OF CLINICAL INFORMATION

Trial Design

A 115-patient, double blind, randomized, sham-controlled trial+ enrolled adolescents 11 to 18 years of age who had abdominal pain-related functional gastrointestinal disorders. During the trial, patients could continue stable doses of medication if there were no dose changes during the four weeks of treatment and two weeks prior to enrolling in the trial. Patients reported any concurrent medication use at each weekly visit.

Patients were provided with an IB-Stim or sham device after baseline measurements and returned to the clinic for a replacement at the end of Weeks 1, 2, and 3 at which time data were recorded. Patients were asked to return to the clinic within 8-12 weeks for long-term follow-up data collection. This follow-up data was ultimately not considered in FDA's decision.

Outcome measures

The study collected the following patient-reported outcome measures at the end of Weeks 1, 2, 3 and at follow-up:

  • Rating of usual, or average, pain experienced in the past week scored from 0 for no pain to 10 for worst pain;
  • Rating of worst pain experienced in the past week scored from 0 for no pain and 10 for ● worst pain:
  • Composite pain frequency-severity-duration (PFSD) score. The PFSD composite score is derived by multiplying the number of days of pain (0-7), the ratings of usual pain (0-10) and worst pain (0-10) over the past week, and then dividing the product by 10 with the highest possible score of 70; and
  • Symptom response scale (SRS) score. Symptoms are recorded as better, worse, or no change based on a 15-point scale across individual domains for both improvement and deterioration of overall symptoms (-7 to -1= worse; 0 = no change; +1 to +7 = better).

Primary outcomes included change from baseline to end of third week in worst abdominal pain and composite PFSD scores.

Secondary outcomes included: 1) the global symptom improvement based on a minimum of +2 in the SRS score and 2) ≥ 30% improvement (defined and analyzed post hoc) in worst and usual abdominal pain at Week 3.

Results

1 Kovacic K, Hainsworth K, Sood M, Chelimsky G, Unteutsch R, Nugent M, Simpson P, Miranda A. Neurostimulation for abdominal pain-related functional gastrointestinal disorders in adolescents: a randomised, double-blind, sham-controlled trial. Lancet Gastroenterol Hepatol. 2017;2:727-737 (https://www.thelancet.com/journals/langas/article/PIIS2468-1253(17)30253-4/fulltext)

{7}------------------------------------------------

The trial enrolled and randomized 60 subjects to receive active treatment and 55 to receive sham treatment for a total of 115 subjects. During the study, 11 subjects either discontinued their participation in the trial or were excluded from the trial. Three patients discontinued in the active treatment group: two patients were excluded as they were diagnosed during the trial with either peptic ulcer or eosinophilic oesophagitis (both are exclusion criteria for the study) and one patient discontinued because they found the device too uncomfortable to continue. Eight patients discontinued from the sham group: three discontinued participation due to a needle phobia or anxiety, two discontinued for aesthetic reasons, one discontinued as they found the device uncomfortable, one discontinued due to an adhesive allergy, and one was excluded as they were diagnosed with eosinophilic oesophagitis (an exclusion criterion for the study). A total of 104 subject completed the trial. The investigator's final analysis was not a true intent-to-treat analysis because it used only the data from the 104 patients who completed the trial (57/60 patients in the active group and 47/55 patients in the sham group).

The primary condition studied was irritable bowel syndrome (IBS) with 28/57 patients in the active group and 23/47 in the sham group. The remaining 29/57 and 24/47 patients in the active and sham groups, respectively, had other functional abdominal pain conditions (e.g., functional dyspepsia, abdominal migraine).

Of the patients who completed the trial, a majority of these patients (N=78) had not previously responded to one or more medications commonly used to treat chronic abdominal pain, whereas 26 patients were treatment naïve.

A blinding assessment was included in the trial after Week 3 for all patients. One half of sham-treated patients believed they were receiving sham treatment compared to one fifth of IB-Stim-treated patients. No blinding assessment analysis was provided for the IBS subgroup.

A pre-specified repeated measures analysis demonstrates a statistically significant difference between groups in all prespecified primary and secondary outcome measures. However, the different etiologies of the patient conditions raises concerns about pooling data from different populations. In this evaluation, the small number of patients with other functional abdominal pain disorders (i.e., functional dyspepsia, abdominal migraine, and functional abdominal pain) and the lack of a prespecified subgroup analysis demonstrating statistical or clinical effectiveness in each studied etiology would not support a general "functional abdominal pain" indication. Therefore, a post hoc sub-group analysis for each of the various disorders studied was analyzed. The study was not sufficiently powered to investigate differences in the subcategories of functional abdominal pain, functional dyspepsia or abdominal migraine. However, the irritable bowel syndrome (IBS) sub-group had a sufficient number of subjects for this analysis and is summarized below.

Post-hoc irritable bowel syndrome (IBS) subgroup analysis requested by FDA

FDA requested this post-hoc analysis to assist in understanding whether the results for patients in the IBS sub-group, who comprised of the majority of the participating subjects in the trial, were consistent with the results reported for the overall cohort studied.

{8}------------------------------------------------

A total of 51 subjects met Rome III criteria for IBS. Thev represent 44% (51/115) of the 115 randomized subjects. Of the 51 subjects with IBS, 50 completed the study and were included in the subgroup analysis (one subject with IBS discontinued participation during the trial, described above). Of these 50 subjects, 23 received the sham device and 27 received the active IB-Stim device. Both groups were comparable in age, gender. BMI, ethnicity, negative diagnostic workup as well as baseline pain (PFSD) and functional disability inventory (FDI) scores (see Table 3). One patient reported taking Tramadol during the study (total of 3 doses). No other medication use during the trial was reported.

The majority of IBS patients in the IB-Stim group (70%; 19/27) and the sham group (87%; 20/23) had not responded to one or more medications commonly used to treat chronic abdominal pain, whereas 30% (8/27) and 13% (3/23) were treatment naïve.

Group CharacteristicsnIB-Stim (n=27)nSham (n=23)p-value
Demographics
Age (years) median (IQR)2715.3 (13.8,16.7)2315.6 (14.2.17.2)(b) (4)
BMI, median (IQR)2722.4 (20.0,29.2)2323.4 (20.9,26.1)(b) (4)
Female2724 (89%)2321 (91%)(b) (4)
Race
Caucasian2722 (82%)2321 (91%)
African American275 (19%)232 (9%)
Diagnostic workup
Normal Screening Labs2727 (100%)2323 (100%)(b) (4)
Normal Abdominal Imaging1919 (100%)1616 (100%)
Normal Endoscopy +/- Colonoscopy1919 (100%)1414 (100%)

Table 3: IBS Group Comparison Data

Assessment in worst pain and PFSD composite score were recorded weekly and compared to sham (weeks 1, 2 and 3). The analysis of change in worst pain and change in composite PFSD scores from baseline included a mixed model using compound symmetry as the covariate structure. Time (week) was a random effect. A Tukey-Kramer was used to adjust for multiple comparisons. The models were run for outcomes in worst pain and PFSD scores, and SRS.

The repeated measures analysis demonstrated group differences in worst pain from baseline to Week 1 and 3 in the IB-Stim group compared to the sham group. Similarly, there was a greater improvement in the composite pain score from baseline to weeks 1, 2 and 3 in the IB-Stim group compared to the sham group. In both worst pain and composite pain scores, the most significant effects were seen at Week 3.

Between group differences in composite pain scores noted after 3 weeks showed that the IB-Stim group had a lower PFSD composite pain score (median 7.5, interquartile range [IQR] 3.6-14.4) compared to the sham group (median 14.4, IQR 4.5-39.2). Similar findings were seen in worst abdominal pain scores: IB-Stim group had a lower worst pain score (median 5.0, IQR 4.0-7.0) compared to the sham group (median 7.0, IQR 5.0-9.0).

{9}------------------------------------------------

Fifty-nine percent (59%; 16/27) of patients receiving the active IB-Stim compared to 26% (6/23) of patients receiving sham showed greater than 30% reduction in worst abdominal pain from baseline to 3 weeks of therapy. Similarly, 52% (14/27) of the IB-Stim group compared to 30% (7/23) of the sham group had greater than 30% reduction in usual abdominal pain after 3 weeks of therapy.

Global Symptoms

The active IB-Stim treatment group reported global symptom improvement after 3 weeks while there was no change in the sham group: median SRS score of 3 (IQR: 2-4) in IB-Stim group versus a median of 0 (IQR: 0-2) in the sham group. At the end of 3 weeks of therapy, 81% (22/27) of the IB-Stim group compared to 26% (6/23) of the sham group reported overall symptom improvement with an SRS score of 2 or more. Similar findings were noted after 2 weeks of therapy where 78% (21/27) of the IB-Stim group compared to 39% (9/23) of the sham group had an SRS score of 2 or more.

Safety analysis

No serious adverse events were recorded in any subject during IB-Stim study. Ten patients reported the following adverse events:

  • Ear discomfort (n=6) -
  • Adhesive allergy (n=3) -
  • -Syncope (n=1)

In addition to the adverse events reported in this study, percutaneous therapies generally have risks of bleeding or infection at the puncture site, skin irritation or pain at the site of application. Additional information about the safety of percutaneous electrical nerve field stimulators (PENFS) is discussed below.

LABELING

Labeling for the IB-Stim resembles labeling previously provided for the NSS-2 BRIDGE and EAD device (DEN170018 and K140530, respectively). The IB-Stim Instructions for Use are consistent with the clinical data and cover all the hazards and other clinically relevant information that may impact use of the device. The labeling includes the following contraindications:

  • Use of cardiac pacemakers because no clinical data is available;
  • Hemophilia; and
  • Psoriasis vulgaris. ●

The labeling also clarifies that an intact skin surface is essential for the use of IB-Stim stimulator. A summary of the clinical data used to support the proposed intended use is provided including a description of the risks and adverse events associated with normal use of the IB-Stim. The labeling provides adequate instructions for use to inform the health care provider in the correct placement and safe use of the IB-Stim including information on the shelf life and technical parameters of the device. Specific instructions for the practical use of the IB-Stim as an aid to IBS therapies was provided:

{10}------------------------------------------------

Patients should be encouraged to make lifestyle modification during treatment. Long-term treatment options with psychological or pharmacological interventions should also be discussed with patient prior to, or at the time of initiating therapy with IB-Stim.

The labeling satisfies the requirements of 21 CFR § 801.109.

RISKS TO HEALTH

The table below identifies the risks to health that may be associated with use of a non-implanted nerve stimulator for functional abdominal pain relief and the measures necessary to mitigate these risks:

Identified Risks to HealthMitigation Measures
Adverse tissue reactionBiocompatibility evaluation
Labeling
Electrical, mechanical, or thermal hazardsleading to user discomfort or injuryElectromagnetic compatibility testing
Electrical, mechanical, and thermal safety testing
Non-clinical performance testing
Software verification, validation and hazardanalysis
Labeling
InfectionSterility testing
Shelf life testing
Labeling

Table 4 - Identified Risks to Health and Mitigation Measures

SPECIAL CONTROLS

In combination with the general controls of the FD&C Act, the non-implanted nerve stimulator for functional abdominal pain relief is subject to the following special controls:

  • (1) The patient-contacting components of the device must be demonstrated to be biocompatible.
  • (2) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed.
  • (3) Electrical performance testing of the device and electrodes must be conducted to validate the specified electrical output and duration of stimulation of the device.
  • (4) Software verification, validation, and hazard analysis must be performed.
  • (5) Sterility testing of the percutaneous components of the device must be performed.

{11}------------------------------------------------

  • (6) Shelf life testing must be performed to demonstrate continued sterility, package integrity, and device functionality over the labeled shelf life.
  • (7) Labeling must include the following:
    • (i) A detailed summary of the device technical parameters;
    • (ii) A warning stating that the device is only for use on clean, intact skin;
    • (iii) Instructions for use, including placement of the device on the patient; and
    • (iv) A shelf life.

BENEFIT-RISK DETERMINATION

The risks of the device are based on data collected in a clinical study described above.

Overall, risks from device use appear to be low. There were no serious adverse events recorded for any subject during the clinical study. There were ten patient reported adverse events of ear discomfort (n=6), adhesive allergy (n=3), and syncope (n=1). These are reported to be typical for this device type.

Although not specifically reported in the study described above, percutaneous therapies generally have risks of bleeding or infection at the puncture site, and skin irritation or pain at the site of application.

The probable benefits of the device are based on data collected in the clinical study as described above.

The overall cohort of the clinical study demonstrate statistically significant results for all prespecified primary and secondary endpoints. However, an indication for use statement can only be supported by the patient population adequately represented in the study, which is reflected only in the IBS subgroup which included 44% (51/115) of patients randomized.

The IBS subgroup analysis shows a decrease in usual pain and worst pain. This reduction was seen in more patients in the active treatment group than those in the sham group. This was also reflected in the responder rates for patients with at least 30% reduction in worst pain and usual pain, as described above. These results demonstrate that there is a clinically meaningful benefit for use of the device in the IBS subpopulation.

This clinically meaningful benefit of reduction in pain for adolescents with IBS treated with this device and other therapies.

The benefits outweigh the probable risks associated with this device, as described above.

PATIENT PERSPECTIVES

This submission did not include specific information on patient perspectives for this device.

{12}------------------------------------------------

BENEFIT/RISK CONCLUSION

In conclusion, given the available information above, for the following indication statement:

The IB-Stim is a percutaneous electrical nerve field stimulator (PENFS) system intended to be used in patients 11-18 years of age with functional abdominal pain associated with irritable bowel syndrome (IBS). The IB-Stim is intended to be used for 120 hours per week up to 3 consecutive weeks, through application to branches of Cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination, as an aid in the reduction of pain when combined with other therapies for IBS.

the device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.

CONCLUSION

The De Novo request for the IB-Stim is granted and the device is classified as follows:

Product Code: QHH Device Type: Non-implanted nerve stimulator for functional abdominal pain relief Regulation Number: 21 CFR 876.5340 Class: II

§ 876.5340 Nonimplanted nerve stimulator for functional abdominal pain relief.

(a)
Identification. A nonimplanted nerve stimulator for functional abdominal pain relief is a device that stimulates nerves remotely from the source of pain with the intent to relieve functional abdominal pain. This generic type of device does not include devices designed to relieve pelvic pain.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed.
(3) Electrical performance testing of the device and electrodes must be conducted to validate the specified electrical output and duration of stimulation of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Sterility testing of the percutaneous components of the device must be performed.
(6) Shelf life testing must be performed to demonstrate continued sterility, package integrity, and device functionality over the labeled shelf life.
(7) Labeling must include the following:
(i) A detailed summary of the device technical parameters;
(ii) A warning stating that the device is only for use on clean, intact skin;
(iii) Instructions for use, including placement of the device on the patient; and
(iv) A shelf life.