K Number
DEN180030
Date Cleared
2018-11-19

(147 days)

Product Code
Regulation Number
862.3460
Type
Direct
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QMS Plazomicin Immunoassay is intended for the quantitative determination of plazomicin in human K2-EDTA plasma on automated clinical chemistry analyzers. The assay results obtained should only be used as an aid in the management of patients with complicated urinary tract infection (cUTI) receiving plazomicin therapy.

The assay should only be used in conjunction with information available from clinical evaluations and other diagnostic procedures.

Device Description

The OMS Plazomicin Immunoassay system is a homogeneous assay utilizing particle agglutination technology and it is based on the competitive binding principle. The assay consists of liquid ready-to-use reagents R1 (anti-plazomicin mouse monoclonal antibody) and R2 (plazomicin-coated microparticles).

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device's performance, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are not explicitly laid out as a numbered list with specific thresholds before the results are presented. Instead, the document describes the types of studies conducted to evaluate performance, and the results of those studies demonstrate that the device meets the "clinically appropriate" standards as determined by the FDA. The special controls outlined in Section P provide a retrospective view of what was deemed acceptable.

Therefore, the table below consolidates the implicit acceptance criteria based on the special controls and the corresponding reported performance.

Performance CharacteristicAcceptance Criteria (Implied from Special Controls & Document Context)Reported Device Performance
PrecisionClinically appropriate precision, including near medical decision points throughout the therapeutic range. Must include a minimum of three samples with different concentrations, including clinical specimens from patients taking plazomicin.Internal Precision: QC materials, spiked plasma samples, and patient plasma pools showed %CVs generally ranging from 2% to 7% for total-run variability. Multi-Laboratory Precision: Overall reproducibility %CVs ranged from 4% to 8% across three sites. (See tables in Section L.1.a)
Linearity/Assay Reportable RangeEvidence of linearity across the claimed measuring range.Supported for 0.8 to 34.0 ug/mL. Regression analysis showed slopes close to 1 (0.98-1.04) and high correlation (r=1.00) for both spiked and patient pooled plasma.
Analytical RecoveryAcceptable recovery throughout the claimed measuring interval.Recovery ranged between 97% and 104% throughout the claimed measuring interval (0.3 to 34.0 ug/mL).
TraceabilityTraceable to reliable reference.Traceable to plazomicin reference calibrators, gravimetrically prepared and value confirmed by LC-MS/MS.
Detection Limit (LoB, LoD, LoQ)Clearly defined and clinically appropriate detection and quantitation limits.LoB: (b)(4) ng/L (specific value redacted) LoD: 0.4 ug/mL LoQ: ≤ 0.8 ug/mL (defined as lowest concentration with interassay precision ≤ 20%CV and bias ≤ 15%)
Analytical Specificity (Endogenous Interference)Free from clinically significant interference from endogenous substances.All tested endogenous substances (Albumin, Bilirubin, Cholesterol, Creatinine, Gamma Globulin, HAMA, Hemoglobin, Rheumatoid Factor, Triglyceride, Uric Acid) resulted in ≤ 10% bias.
Analytical Specificity (Exogenous Interference)Free from clinically significant interference from co-administered medications relevant to cUTI patients and structurally similar compounds.None of the listed concomitant medications or structurally similar compounds, at clinically relevant concentrations, caused ≥ 10% bias in plazomicin measurement.
Method ComparisonClinically appropriate accuracy against a comparator method. Data collected at a minimum of three laboratory sites.Comparison with validated LC-MS/MS method for 134 K2-EDTA plasma samples from patients. Deming: Slope = 1.007, Intercept = 0.72 Passing-Bablok: Slope = 1.039, Intercept = 0.41 Correlation (R): 0.983 (Data across 3 test laboratories, as required).

2. Sample Sizes Used for the Test Set and Data Provenance

  • Test Set (Analytical Studies):

    • Precision (Internal):
      • Quality Control (QC) and Spiked samples: n=80 each (2 replicates/run, 2 runs/day, 20 days).
      • Patient Plasma Pools: n=20 each (2 replicates/run, 2 runs/day, 5 days).
    • Precision (Multi-Laboratory):
      • QC and Spiked samples: n=240 each (n=80 per site x 3 sites).
      • Patient Pools: n=60 each (n=20 per site x 3 sites).
    • Linearity: 9 levels of samples, tested in 5 replicates in a single run.
    • Analytical Recovery: Each sample panel analyzed over 5 days with 4 replicates/day (total 20 measurements per sample).
    • Detection Limit (LoB, LoD, LoQ): Specific sample numbers for LoB/LoD/LoQ are redacted (b)(4). LoB used (b)(4) K2-EDTA samples and LoD used (b)(4) blank plasma K2-EDTA samples.
    • Method Comparison: 134 K2-EDTA plasma samples from patients taking plazomicin.
    • Interference (Endogenous/Exogenous): Not explicitly stated, but implies a sufficient number of spiked samples to test the indicated concentrations.
  • Data Provenance:

    • Country of Origin: Not explicitly stated, but the multi-laboratory precision study and method comparison study imply a multi-site approach, possibly within a single country or across different countries, but this is not detailed.
    • Retrospective or Prospective: Most of the analytical performance studies (precision, linearity, recovery, detection limits, interference) appear to be prospective laboratory studies using prepared or collected samples. The patient samples used for precision and method comparison are collected from patients taking plazomicin, indicating a real-world context for some of the samples. The clinical trial data for plazomicin efficacy/nephrotoxicity (mentioned under "Clinical Studies") was retrospective regarding the device's evaluation, as TDM was not used to adjust dosing during that trial, but the data was later correlated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: This is a quantitative immunoassay, not an imaging device requiring expert reader interpretation. Therefore, the concept of "experts establishing ground truth" in the typical sense (e.g., radiologists reviewing images) does not directly apply.
  • Qualifications of Experts: The ground truth for this device is established through:
    • Reference Methods: The primary ground truth for the device's accuracy and traceability is a validated LC-MS/MS method. Personnel operating and validating LC-MS/MS are highly trained laboratory professionals with expertise in mass spectrometry and analytical chemistry, though their specific qualifications (e.g., years of experience, certifications) are not detailed in this document.
    • Gravimetric Preparation: Plazomicin reference calibrators are gravimetrically prepared, meaning their concentration is determined by precise weighing, which implies a fundamental chemical and metrological expertise.
    • CLSI Guidelines: The studies follow CLSI (Clinical and Laboratory Standards Institute) guidelines, which are established by consensus among experts in laboratory medicine.

4. Adjudication Method for the Test Set

  • Since the ground truth is established by reference methods (LC-MS/MS, gravimetric preparation) and not by subjective expert interpretation, there is no "adjudication method" in the sense of resolving discrepancies between multiple human readers. The analytical methods themselves have defined reconciliation processes, but it's not a multi-reader adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC study was NOT done. This device is a quantitative immunoassay, not an AI-powered diagnostic tool that assists human readers in interpreting complex data like medical images. Therefore, the concept of human readers improving with AI assistance is not applicable here.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Yes, in essence, standalone performance was evaluated. The QMS Plazomicin Immunoassay is an automated laboratory test. Its performance characteristics (precision, linearity, accuracy against LC-MS/MS, detection limits, interference) are all measures of its intrinsic, standalone capability to quantify plazomicin in a sample. There isn't a "human-in-the-loop" component to its analytical function; rather, humans use the results obtained from the automated device.

7. The Type of Ground Truth Used

The ground truth used for evaluating the QMS Plazomicin Immunoassay's performance includes:

  • Reference Measurement Procedure: A validated LC-MS/MS method was used as the comparator for method comparison studies, serving as the "gold standard" for quantifying plazomicin.
  • Gravimetric Preparation: For calibrators and spiked samples, the known concentration derived from gravimetric preparation (weighing out precise amounts of plazomicin) served as the ground truth.
  • Clinical Outcomes Data (Indirectly for clinical utility, not analytical performance): The clinical trial data (retrospectively analyzed post-drug approval) provided evidence for the correlation between plazomicin trough levels and nephrotoxicity, which established the clinical need and medical decision points for the assay, but not the direct analytical accuracy of the device itself.

8. The Sample Size for the Training Set

  • Not Applicable in the traditional machine learning sense. This document describes a traditional in-vitro diagnostic (IVD) device (a chemical immunoassay), not an AI/Machine Learning algorithm that requires a "training set" to learn from data. The device's operational parameters (e.g., reagant concentrations, instrument settings) are determined through development and optimization processes, not by training on a large dataset in the way an AI model would be.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. As this is not an AI/ML device, there is no "training set" or corresponding ground truth establishment process in that context. The "ground truth" for the device's development primarily relies on established principles of analytical chemistry, reagent formulation, and calibration against reference methods like LC-MS/MS and gravimetric standards.

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR QMS Plazomicin Immunoassay DECISION SUMMARY

A. DEN Number:

DEN180030

B. Purpose for Submission:

De Novo request for evaluation of automatic class III designation of the QMS Plazomicin Immunoassay

C. Measurand:

Plazomicin

D. Type of Test:

Quantitative, homogenous particle-enhanced competitive turbidimetric immunoassay

E. Applicant:

Microgenics Corporation

F. Proprietary and Established Names:

QMS Plazomicin Immunoassay

G. Regulatory Information:

    1. Regulation Section:
      21 CFR 862.3460
    1. Classification:
      Class II, Special Controls
    1. Product Code(s):
      QDR
    1. Panel:
      Toxicology (91)

{1}------------------------------------------------

H. Indications for use:

l. Indication(s) for use:

OMS Plazomicin Immunoassay:

The QMS Plazomicin Immunoassay is intended for the quantitative determination of plazomicin in human K2-EDTA plasma on automated clinical chemistry analyzers. The assay results obtained should only be used as an aid in the management of patients with complicated urinary tract infection (cUTI) receiving plazomicin therapy.

The assay should only be used in conjunction with information available from clinical evaluations and other diagnostic procedures.

    1. Special conditions for use statement(s):
      For prescription use only.

3. Special instrument requirements:

Performance is evaluated on the Beckman Coulter AU680 Chemistry Analyzer.

I. Device Description:

The OMS Plazomicin Immunoassay system is a homogeneous assay utilizing particle agglutination technology and it is based on the competitive binding principle. The assay consists of liquid ready-to-use reagents R1 (anti-plazomicin mouse monoclonal antibody) and R2 (plazomicin-coated microparticles).

J. Standard/Guidance Document Referenced:

  • . CLSI EP05-A3 - Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition.
  • CLSI EP06-A Evaluation of the Linearity of Quantitative Measurement Procedures: A . Statistical Approach; Approved Guideline.
  • . CLSI EP07-A2 - Interference Testing In Clinical Chemistry; Approved Guideline -Second Edition.
  • CLSI EP14-A3 Evaluation of Commutability of Processed Samples; Approved ● Guideline - Third Edition.
  • . CLSI EP15-A3 - User Verification of Precision and Estimation of Bias; Approved Guideline - Third Edition.
  • CLSI EP17-A2 - Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition.
  • CLSI EP25-A Evaluation of Stability of In Vitro Diagnostic Reagents; Approved ● Guideline

{2}------------------------------------------------

K. Test Principle:

The QMS Plazomicin Immunoassay is a homogeneous particle-enhanced turbidimetric immunoassay. The assay is based on competition between drug in the sample and drug coated onto microparticles for antibody binding sites of the anti-plazomicin antibody reagent (R1). The plazomicin-coated microparticle reagent (R2) is rapidly agglutinated in the presence of the anti-plazomicin antibody reagent and in the absence of any competing drug in the sample. The assay is based on the rate of absorbance change, which is measured photometrically. When a sample containing plazomicin is added, the agglutination reaction is partially inhibited, slowing down the rate of absorbance change. A concentrationdependent agglutination inhibition curve is obtained, where the maximum rate of agglutination occurs at the lowest sample plazomicin concentration and the lowest agglutination rate occurs at the highest sample plazomicin concentration.

L. Performance Characteristics:

    1. Analytical performance:
    • a. Precision/Reproducibility:

Internal Precision Study

An internal precision study was conducted according to CLSI Document EP05-A3 by measuring trilevel quality control (QC) materials, spiked plasma samples, and patient plasma pools. Each spiked and OC material sample was tested in two replicates per run, two runs per day for 20 days (n=80 per sample). Patient pools were tested for two replicates per run, two runs per day for five days (n=20 per sample). The following results were obtained.

Sample DescriptionNMean(ug/mL)Within-Run SDWithin-Run%CVBetween-Run SDBetween-Run%CVTotal-Run SDTotal-Run%CV
Quality Control 1802.50.12%0.14%0.15%
Quality Control 2808.10.22%0.34%0.34%
Quality Control 38029.61.24%0.83%1.76%
Plasma Patient Pool 1202.30.12%0.14%0.15%
Plasma Patient Pool 2208.00.22%0.22%0.34%
Plasma Patient Pool 32027.30.73%0.93%1.24%
Plasma Spiked 1802.30.13%0.14%0.16%
Plasma Spiked 2807.80.12%0.34%0.56%
Plasma Spiked 38014.80.32%0.85%1.07%
Plasma Spiked 48028.40.93%1.14%1.97%

{3}------------------------------------------------

Multi-Laboratory Precision:

The manufacturer also conducted precision studies at external sites, using tri-level quality control (QC) material, spiked plasma pools, and patient plasma pools. Each spiked and QC material sample was tested in two replicates per run, 2 runs per day for 20 days (n=80 per sample). Patient pools were tested for two replicates per run, 2 runs per day for 5 days (n=20 per sample). The multi-laboratory total repeatability and reproducibility were calculated from combining precision results across three sites. The overall precision study results are shown below:

SampleNGrandMean(ug/mL)TotalRepeatabilityTotalBetween-RunTotalBetween-DayReproducibility
SD%CVSD%CVSD%CVSD%CV
QC12402.50.12%0.14%0.01%0.15%
QC 22408.20.12%0.23%0.11%0.34%
QC 324029.60.93%0.72%0.62%1.24%
Patient Pool 1602.40.15%0.00%0.00%0.27%
Patient Pool 2608.40.44%0.00%0.22%0.67%
Patient Pool 36028.81.04%0.52%0.72%2.48%
Spiked 12402.40.13%0.13%0.12%0.26%
Spiked 22408.10.23%0.23%0.33%0.56%
Spiked 324015.40.64%0.53%0.43%1.17%
Spiked 424029.71.14%0.83%0.62%2.28%

b. Linearity/assay reportable range:

The linearity of the QMS Plazomicin Immunoassay was assessed with two different types of samples; plazomicin-spiked human plasma K2-EDTA, and plazomicin patient pooled plasma K2 EDTA samples. For spiked samples, a high concentration plazomicin stock solution (approximately 1.0 mg/mL) was added to drug-free human K2-EDTA plasma to create a spiked sample with high plazomicin of 34.0 µg/mL. A patient pooled plasma sample was created by pooling patient plasma samples with high plazomicin concentrations to create a sample with concentration of 34.0 µg/mL. The high concentration spiked and pooled patient samples were then diluted with drug-free K2-EDTA plasma to create lower concentrations ranging from 0 ug/mL to 34.0 ug/mL. Nine (9) levels of samples, including concentrations between 0.3 ug/mL and 34.0 µg/mL, were tested in five replicates in a single run.

Regression analysis was performed between the measured mean plazomicin and assigned values for each dilution. The linear regression results from a representative kit lot are shown below:

{4}------------------------------------------------

Plasma K2EDTA (spiked)

SiteSlopeInterceptr
A1.040.321.00
B0.980.361.00
C1.020.341.00

Patient Pooled Plasma

SiteSlopeInterceptr
A0.980.231.00

These results support the claimed measuring range of 0.8 to 34.0 ug/mL, where the lower end of the measuring range is defined by the LoQ.

Analytical Recovery:

Accuracy by recovery was performed following CLSI standard EP15-A3. Two sample panels were prepared by spiking known amounts of plazomicin into negative human K2-EDTA plasma at concentrations throughout the assay range. Each sample was analyzed over a course of five days with four replicates each day for a total of 20 measurements. The mean measured concentration was compared to the gravimetric target concentration for percent recovery. Recovery ranged between 97% and 104% throughout the claimed measuring interval of the device (0.3 to 34.0 ug/mL)

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability:

A six-level kit of QMS Plazomicin Immunoassay Calibrators (A through F) is used to calibrate the assay. A three-level kit of QMS Plazomicin Immunoassay Controls (1 through 3) is used establish control limits. The kits of calibrators and controls are sold separately.

The OMS Plazomicin Immunoassay is traceable to plazomicin reference calibrators which are gravimetrically prepared with plazomicin sulfate in the human serum and value confirmed by LC-MS/MS.

  • d. Detection limit:
    Limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) studies were performed and analyzed as described below.

LoB Test Protocol
To estimate the LoB, (b) (4) K2-EDTA samples were tested in replicates of (b) (4) over (b) days. Three reagent lots and one Beckman Coulter AU680 instrument was used. A total of (b) replicates were obtained on a given (b) (4)

{5}------------------------------------------------

instrument/reagent lot combination. The LoB was determined to be (4) ng/L, using the non-parametric option.

LoD Test Protocol

(b) (4)blank plasma K2-EDTA samples, spiked gravimetrically to (b) (4) of plazomicin, were tested in replicates of five in one run over three days. Three reagent lots and one Beckman Coulter AU680 instrument was used. A total of " replicates were obtained on a given instrument/reagent lot combination. The LoD was determined to be 0.4 ug/mL using the parametric option.

LoQ Test Protocol

The LoQ of the assay was determined based on the guidelines from CLSI standard EP17- A02. The LoO is defined as the lowest concentration which results in interassay precision ≤ 20%CV and bias ≤ 15%. The LoQ was determined to be ≤ 0.8 ug/mL for K2-EDTA plasma samples.

  • e. Analytical specificity:

Endogenous interference:

Interference studies were conducted using CLSI standard protocol EP07-A2 as a guideline. The following endogenous substances, when tested with the QMS Plazomicin Immunoassay at the concentrations indicated, resulted in less than or equal to 10% bias in detecting plazomicin. The results are shown below:

Endogenous SubstancesConcentration
Albumin6 g/dL
Unconjugated Bilirubin30 mg/dL
Conjugated Bilirubin30 mg/dL
Cholesterol500 mg/dL
Creatinine5 mg/dL
Gamma Globulin10 g/dL
Human Anti-Mouse Antibody(HAMA) Type 1&220 ng/mL
Hemoglobin1000 mg/dL
Rheumatoid Factor*1080 IU/mL
Triglyceride1000 mg/dL
Uric Acid30 mg/dL
  • Naturally existing substance

Exogenous interference

Specificity studies were conducted using CLSI standard protocol EP07-A2 as a guideline. Positive and negative interference effects were evaluated for clinically

{6}------------------------------------------------

relevant concentrations of potential cross-reactants that are either concomitant medications or are structurally similar to plazomicin. None of the compounds, when added into the plasma pool containing 2.5 or 30 µg/mL plazomicin at the concentrations listed below, caused greater than or equal to 10% bias in plazomicin measurement. The results are shown below:

Concomitant Medication/ Structurally SimilarCompoundsConcentration ( $µg/mL$ )
Acetyl-Salicylic Acid750
Amikacin200
Amiodarone50
Amlodipine5
Amphotericin20
Ampicillin/ Sulbactam*500
Azithromycin50
Aztreonam750
Carvedilol5
Cefazolin1250
Cefepime500
Ceftaroline fosamil100
Ceftazidime550
Ceftazidime/ Avibactam*300
Ceftolozane/ Tazobactam*300
Ceftriaxone sodium1000
Ciprofloxacin100
Cisatracurium20
Laudanosine (Cisatracurium metabolite)3.5
MQA metabolite (Cisatracurium metabolite)20
Clonidine5
Colistimethate Sodium (Colistin)100
Daptomycin700
Dexmedetomidine5
Diltiazem hydrochloride20
Dobutamine hydrochloride20
Dopamine hydrochloride5
Doripenem100
Enoxaparin200
Epinephrine hydrochloride20
Ertapenem500
Erythromycin100
Esmolol20
Esomeprazole20
Ethacrynic Acid100
Fentanyl2
Concomitant Medication/ Structurally SimilarCompoundsConcentration (µg/mL)
Fluconazole100
Fondaparinux20
Fosfomycin100
Furosemide100
Gentamicin50
Haloperidol20
Hydrocortisone20
Imipenem/ Cilastatin*300
Insulin human regular20
Kanamycin B100
Levetiracetam300
Levofloxacin100
Linezolid100
Lorazepam2
Meropenem600
Metamizole (Dipyrone)100
Metformin100
Methylprednisolone sodium succinate100
Metoclopramide2
Metoprolol20
Metronidazole300
Micafungin100
Midazolam5
Morphine sulfate20
Nafcillin100
N-desethylamiodarone (metabolite of Amiodarone)20
Netilmicin100
Norepinephrine Bitartrate (Noradrenaline)5
Omeprazole20
Pantoprazole20
Paracetamol (acetaminophen)500
Phenylephrine hydrochloride5
Phenytoin (Fosphenytoin IV)20
Phenytoin sodium50
Piperacillin1300
Propofol100
Ramipril5
Ranitidine20
Sisomicin20
Spectinomycin500
Streptomycin300
Tamsulosin5
Teicoplanin200
Concomitant Medication/ Structurally SimilarCompoundsConcentration (µg/mL)
Tedizolid50
Tigecycline20
Tobramycin100
Tramadol20
Vancomycin200
Vasopressin5
Vecuronium bromide20

{7}------------------------------------------------

{8}------------------------------------------------

  • Tested as combination drugs

f. Assay cut-off:

Not applicable.

2. Comparison studies:

  • a. Method comparison study:
    A total of 134 K2-EDTA plasma samples from patients taking plazomicin, within the measuring range of the QMS Plazomicin Immunoassay, were measured by the QMS Plazomicin Immunoassay in singlicate, over the course of 4 to 5 days and across 3 test laboratories. Data from the candidate device were compared with results from a validated LC-MS/MS method using Passing-Bablok regression analysis (similar data were obtained using a second lot of reagents; not shown).

QMS Plazomicin Immunoassay vs. Comparator LC-MS/MS Method

NDemingPassing-BablokCorrelation(R)
SlopeInterceptSlopeIntercept
1341.0070.721.0390.410.983
  • b. Matrix comparison:
    Not applicable. The claimed matrix is K2-EDTA human plasma.

    1. Clinical studies:
      A clinical study was conducted to support approval of Plazomicin that included (b) (4) randomized adults hospitalized with Complicate Urinary Tract Infections (cUTI), in a multinational, double-blind, noninferiority trial comparing plazomicin to (b) (4) No therapeutic drug monitoring (TDM) was used to adjust patient dosing in the drug trial. However, data from the trial were retrospectively correlated to drug concentrations determined by the candidate device to establish the likelihood of an increase in nephrotoxicity associated with plazomicin trough values greater than (b) (4) in patients with renal impairment. As such, TDM is recommended in the plazomicin drug label for

{9}------------------------------------------------

patients with renal impairment. The candidate device was evaluated based on this recommended plasma trough concentration of (b) (4) and determined to have adequate performance characteristics for plazomicin TDM.

    1. Clinical cut-off:
      Not applicable.
    1. Expected values/Reference range:
      Not applicable.

M. Proposed Labeling:

The labeling supports the decision to grant the De Novo request for this device.

N. Identified Risks and Identified Mitigations:

Identified Risks to HealthIdentified Mitigations
Incorrect test resultsGeneral controls and special controls (1) and (2)
Incorrect interpretation of test resultsGeneral controls and special controls (1) and (2)

O. Benefit/Risk Analysis:

Summary of the Assessment of the Benefit For the Proposed Indications for Use

Plazomicin is an intravenous aminoglycoside antibiotic indicated for treatment of complicated Urinary Tract Infections (cUTI) including (b) (4) Due to evidence of increased nephrotoxicity (i.e. creatinine increase of ≥ 0.5 mg/dL from baseline) in the drug trials, plazonicin exposure-response analysis was performed which targeted a trough concentration of (p) (4) to prompt adjustment of the plazomicin dose regimen prior to a second dose. Dosing adjustment of patients over this trough concentration, based on therapeutic drug monitoring (TDM), involves extending dosing intervals by 1.5-fold (i.e., from every 24 hours to every 36 hours or from every 48 hours to every 72 hours) for moderate to severe renal impairment patients (i.e. creatinine clearance <90 and > 15 mL/min). Since the drug label recommends dosing adjustments based on TDM to decrease the incidence of nephrotoxicity, there is a clear clinical need for a plazomicin TDM device.

Summary of the Assessment of Risk For the Proposed Indications for Use

The risk of a falsely elevated plazomicin result is potentially decreased efficacy of the drug due to unnecessary lengthening of the dosing interval. Loss of drug efficacy may present in clinical deterioration leading to a change in antimicrobial regimen (in most cases changing to a broader regimen or possibly a more toxic regimen), clinical deterioration leading to longer

{10}------------------------------------------------

hospitalization and possible clinical complications associated with longer hospitalization, and clinical deterioration leading to significant morbidity or even mortality. Therapeutic drug monitoring (TDM) was not used during the clinical trial for plazomicin. However, based on the observed levels of plazomicin in the majority of subjects in the clinical study who still showed adequate efficacy, it is uncertain that lengthening the dosing interval will result in compromised efficacy in moderately to severely renal impaired subjects given that the drug is primarily excreted through the urine; reductions in efficacy would likely be relative to the degree of bias of the false result.

The risk of a falsely low plazomicin result is a missed indicator to avoid potential nephrotoxicity in complicated UTI patients by increasing the dosing interval, as falsely low results may lead the clinician to erroneously believe that the plazomicin trough is below the threshold level (≤3 ug/mL) associated with an increased incidence of nephrotoxicity. Therefore, a falsely low plazomicin result could lead to a clinician incorrectly maintaining the dosing interval for cUTI patients. According to the exposure response analysis (as described in the drug label), the incidence of nephrotoxicity (defined as serum creatinine increase of 0.5 mg/dL from baseline) was higher in patients with plazomicin trough concentration > 3u/mL (36%, 10/28) compared to patients with plazomicin trough concentration < 3 µg/mL (5%, 11/215). Acute kidney injury is a known risk factor for the aminoglycoside class of drugs. Typically, renal parameters such as serum creatinine and urine output are monitored carefully during aminoglycoside administration, especially in patients with underlying moderate or severe kidney insufficiency. Monitoring serum creatinine and urine output may serve as a mitigation to the potential for nephrotoxicity due to an erroneous assay result. Additionally, although a creatinine increase of 0.5 mg/dL may indicate acute kidney injury, the injury may be reversible after discontinuation of the drug.

Summary of the Assessment of Benefit-Risk For the Proposed Indications for Use

The benefit of the proposed assay does not outweigh the risk.

Summary of the Assessment of Benefit-Risk, considering risk mitigation strategies For the Proposed Indications for Use

Overall, the probable benefits outweigh the probable risks of this device after considering the mitigations provided by the special controls in addition to the general controls.

Patient Perspectives:

Plazomicin is reserved for use in patients with complicated urinary tract infection who have limited or no alternative treatment options. FDA considered the clinical need for this device in this patient population.

P. Conclusion:

The information provided in this de novo submission is sufficient to classify this device into class II under regulation 21 CFR 862.3460. FDA believes that the special controls, in

{11}------------------------------------------------

combination with the general controls provide a reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:

Product Code:QDR
Device Type:Plazomicin test system
Class:II (special controls)
Regulation:862.3460

a) Identification.

A plazomicin test system is a device intended to measure plazomicin in human specimens. Measurements obtained by this device are used in monitoring levels of plazomicin to ensure appropriate therapy in patients with complicated urinary tract infection.

  • b) Classification. Class II (special controls). A plazomicin test system must comply with the following special controls:
    • (1) Design verification and validation must include the following:
      • Precision study data that demonstrates clinically appropriate precision of the (i) plazomicin test system, as determined by FDA. Precision studies must include a minimum of three samples containing different concentrations of plazomicin, including near medical decision points throughout the expected therapeutic range of plazomicin. Samples near the medical decision points must be clinical specimens collected from patients taking plazomicin.
      • (ii) Method comparison data that demonstrates clinically appropriate accuracy of the plazomicin test system, as determined by FDA. Method comparison data must be collected at a minimum of three laboratory sites.
      • (iii) Data from studies appropriate to demonstrate that the device is free from clinically significant interference from co-administered medications that are used in patients with complicated urinary tract infection, as determined by FDA.
    • (2) The device's 809.10 labeling must include a warning statement that reads: "The assay should only be used in conjunction with information available from clinical evaluations and other diagnostic procedures."

§ 862.3460 Plazomicin test system.

(a)
Identification. A plazomicin test system is a device intended to measure plazomicin in human specimens. Measurements obtained by this device are used in monitoring levels of plazomicin to ensure appropriate therapy in patients with complicated urinary tract infection.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Precision study data that demonstrates clinically appropriate precision of the plazomicin test system. Precision studies must include a minimum of three samples containing different concentrations of plazomicin, including near medical decision points throughout the expected therapeutic range of plazomicin. Samples near the medical decision points must be clinical specimens collected from patients taking plazomicin.
(ii) Method comparison data that demonstrates clinically appropriate accuracy of the plazomicin test system, as determined by FDA. Method comparison data must be collected at a minimum of three laboratory sites.
(iii) Data from studies appropriate to demonstrate that the device is free from clinically significant interference from co-administered medications that are used in patients with complicated urinary tract infection, as determined by FDA.
(2) The device's labeling required under § 809.10 of this chapter must include a warning statement that explains: “This assay should only be used in conjunction with information available from clinical evaluations and other diagnostic procedures.”