(424 days)
The COOK Hemospray® Endoscopic Hemostat is used for hemostasis of non-variceal gastrointestinal bleeding.
The Hemospray® Endoscopic Hemostat is a prescription use device consisting of a hemostatic agent and a delivery system. The hemostatic agent is (b) (4) | bentonite powder, a naturallysourced aluminum phyllosilicate clay. The delivery system is an endoscopic accessory used for spraying the powder onto the bleeding surface. The delivery device consists of a 220cm polyethylene application catheter, a handle with a pressurized CO2 cartridge, and a powder chamber containing approximately 20g of the Hemospray® material is propelled through the application catheter by release of CO2 from the cartridge located in the device handle. A trigger valve with an activation button allows the physician to control the amount of powder delivered to the affected area. Each actuation of Hemospray® delivers of CO2 and (b) gram of (b) (4) bentonite. Use of the complete 20-gram canister can produce a volume increase within the lumen of the bowel of 3 liters. In the presence of blood, the (b) (4) bentonite may swell 10-15% in volume.
The delivery system is offered in two configurations (i.e., two different outer diameters of the powder delivery catheter). The HEMO-7 version is used with a 2.8mm endoscope accessory channel, and the HEMO-10 version is used in a 3.7mm endoscope accessory channel.
The Hemospray® Endoscopic Hemostat is designed for hemostasis of non-variceal gastrointestinal bleeding. The acceptance criteria for the device are derived from a combination of non-clinical, animal, and clinical studies, as well as literature reviews and post-market data. The studies aim to demonstrate the device's biocompatibility, sterility, shelf life, and its ability to achieve acute hemostasis while minimizing risks such as re-bleeding, perforation, and other adverse events.
Here's a breakdown of the acceptance criteria and the studies proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Acceptance Criterion | Reported Device Performance (with supporting study reference) |
|---|---|---|
| Biocompatibility | Device materials (powder and catheter) must be non-cytotoxic, non-sensitizing, non-irritant, non-pyrogenic, non-hemolytic, and demonstrate no acute systemic toxicity, genotoxicity risks mitigated, and no local or systemic subchronic toxicity. | Hemospray® Powder: Non-cytotoxic, Non-sensitizer, Non-irritant, No acute systemic toxicity, Non-pyrogenic, Non-hemolytic, Genotoxicity risks mitigated, No local or systemic toxicity (Table 1, GLP Porcine Study Table 6). Delivery Catheter: Non-cytotoxic, Non-sensitizer, Non-irritant (Table 2). |
| Shelf Life/Sterility | Sterilization validation (gamma irradiation), packaging validation (visual inspection, dye leak, burst testing), and device stability (visual inspection, functional testing) must pass. | All sterilization, packaging, and shelf life tests passed (Table 3). |
| MR Compatibility | Safety and compatibility in the MR environment must be evaluated for heating, migration, or image artifact. | "Hemospray® has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration, or image artifact in the MR environment." The safety is unknown, and the labeling includes a warning (page 4). |
| Performance - Bench | Output pressure below specified threshold to mitigate bloodstream entry risk, user can functionally set up and operate the device, handle does not rupture at working pressures, acceptable torque for detach/reattach and force for button press, consistent powder delivery without clogging/intermittence, and cartridge integrity after simulated shipping. Must show no significant pathological alterations or mucosal penetration in ex vivo tissue. | Output pressure: Passed. Functional testing: Passed. Rupture testing: Passed. Human Factors testing: Passed. Powder density: Passed. Cartridge integrity: Passed. Ex vivo tissue trauma: No significant pathological alterations; no mucosal penetration (Table 4). |
| Materials Characterization | Compositional identity and purity, water absorption, moisture content, bioavailable lead/arsenic levels within safe limits, and consistent particle size distribution for each lot. | Compositional identity and purity established (lot-to-lot acceptance criteria set). Free Swell Test and Moisture Content: Passed. Extraction in simulated gastric fluid for lead and arsenic: Passed. Particle Size Characterization: Results demonstrate consistent distribution (Table 5). |
| Performance - Animal | Assess hemostatic ability (effectiveness) and safety (gross and histologic assessment of regional and systemic effects, powder/treatment) in a clinically relevant animal model, including thromboembolic events, local/systemic toxicity, tissue trauma, GI tract obstruction, and bowel distension/perforation. | Acute hemostasis achieved in 6/6 treated animals. One control and two treatment animals had re-bleeding 8-10 days later. Histopathological evaluation showed no important differences between groups. No Hemospray® powder observed in local/regional tissue or lymph nodes, nor embedded in tissue layers (Table 6). The study stated it does not demonstrate safety/effectiveness in large area low-pressure bleeding or assess delayed re-bleeding. |
| Clinical Effectiveness | Achieve acute hemostasis in non-variceal GI bleeding (upper and lower), with acceptable rates of re-bleeding, adverse events, and mortality compared to standard of care. | Pilot Study: Acute hemostasis 95% (19/20 patients). Sustained hemostasis 89.5% (17/19). No treatment-related SAEs. SEAL: Acute hemostasis 100% (89/89). HALT: Initial hemostasis 96.8% (61/63), Successful hemostasis with Hemospray® alone 88.9% (56/63). Persistent bleeding 11.1%. Early recurrent bleed 7.1%. APPROACH: Initial hemostasis 100% (50/50). Successful hemostasis with Hemospray® 98% (49/50). Literature Review: Overall hemostasis 96% (509/532), Endoscopic hemostasis 97% (518/532). (Summarized in Table 25, Clinical Data). |
| Clinical Safety | Low incidence of device-related serious adverse events (SAEs), including perforation, bowel obstruction, aspirations, and device malfunction. Acceptable 30-day mortality. | Pilot Study: No treatment-related or procedure-related SAEs. No bowel obstruction. SEAL: No unanticipated or serious adverse events attributed to Hemospray®. HALT: Device-related SAEs 0% (0/63). 1 intra-operative perforation reported (patient died). APPROACH: Hemospray® related adverse events 0% (0/50). No device-related adverse events. Literature Review: 4 potential device-related adverse events (0.8%) (2 perforations, 2 aspiration pneumonias). Complaint Data: 4 deaths (2 comorbidities, 1 perforation/sepsis, 1 device malfunction). Labeling and design changes implemented to mitigate risks. (Summarized in Table 25 and Table 26). |
2. Sample Sizes and Data Provenance
-
Test Set (Clinical Data):
- Pilot Study: 20 patients. Hong Kong, China. Prospective.
- SEAL Registry: 89 evaluable patients (N=92 enrolled, 3 excluded). Canada, Denmark, England, France, Germany, Italy, Holland. Post-market, observational survey.
- HALT Study: 64 enrolled patients (interim report). EU and Canada. Post-market, single-arm study.
- APPROACH Study: 50 enrolled patients. Canada. Post-market, single-arm, prospective observational cohort study.
- Literature Review: 30 relevant studies, comprising 522 patients. Geographically diverse (Europe, Asia, Canada). Retrospective and prospective studies, case series, case reports, and registry-derived studies.
- Emergency Use: 5 critically ill patients. United States. Pre-market.
- Total Clinical Data (overall summary): 750 patients.
-
Training Set: Not explicitly stated for specific clinical studies, as many were observational or post-market with concurrent data collection. For the algorithm (if applicable, which is not the primary focus for a physical device in this context), this information would be separate. For the manufacturing/design aspects, non-clinical and bench testing would conceptually be "training data" for product development cycles.
- Biocompatibility: GLP porcine gastric bleeding animal study (subchronic toxicity and implantation).
- Performance - Animal: 6 Hemospray® treated animals, 3 sham control (total N=9). Porcine gastric bleeding model.
3. Number of Experts to Establish Ground Truth for Test Set & Qualifications
- For the clinical studies (Pilot, HALT, APPROACH), the ground truth for outcomes like hemostasis, re-bleeding, and adverse events was established by treating physicians (endoscopists/gastroenterologists) at the respective clinical sites. The text doesn't specify the exact number or years of experience for individual experts, but it implies qualified medical professionals for each study. For example, the HALT study had 10 enrollment sites, meaning at least 10 primary investigators or teams. The APPROACH study had 4 clinical sites.
- The SEAL Registry explicitly states that "all physician participants were trained on the use of the Hemospray® kit by Wilson-Cook Medical, Inc. personnel." This implies a level of qualification and standardized training for those assessing outcomes.
- The overall literature review compiled data reported by clinician-authors in their respective publications.
4. Adjudication Method for the Test Set
- The document does not explicitly detail a centralized adjudication method (e.g., 2+1, 3+1 structure with an independent committee) for the clinical studies. For single-center or multi-center, physician-reported outcomes, the assessment of hemostasis, re-bleeding, and adverse events would typically be performed by the local site investigators guided by the study protocol definitions.
- For the SEAL Registry, "No queries or audits to identify and correct missing data were performed," indicating a lack of formal adjudication beyond initial reporting.
- The HALT study mentions endoscopically-confirmed recurrent bleeds, suggesting a visual assessment by the clinician performing the endoscopy.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done. The studies presented are primarily single-arm studies evaluating the performance of Hemospray® alone or in conjunction with other standard therapies.
- The document does compare Hemospray® results to "Standard of Care per ASGE" in Table 20 (APPROACH study results) and generally discusses how Hemospray® "rivaled standard of care (SOC) with 63% initial hemostasis and 38% re-bleed rate" for patients on antithrombotic therapy. However, these are comparisons to historical or published SOC rates, not from a direct head-to-head MRMC study.
- No effect size of human readers improving with AI vs. without AI assistance is provided, as this is a physical medical device, not an AI diagnostic tool.
6. Standalone (Algorithm Only) Performance Study
- Not applicable as the Hemospray® Endoscopic Hemostat is a physical medical device. Performance is inherently tied to its application by a human operator (physician) in a clinical setting.
7. Type of Ground Truth Used
- Clinical Studies (Pilot, HALT, APPROACH, Emergency Use): Clinician assessment and observation (endoscopic visualization) of bleeding cessation, re-bleeding, and adverse events, as defined by study protocols. Outcomes data (e.g., mortality).
- Animal Study: Gross and histologic assessment of regional and systemic effects by veterinary pathologists/researchers.
- Bench Studies: Objective measurements against predefined engineering specifications (e.g., pressure, density, functional checks).
- Biocompatibility: Laboratory testing results based on ISO and ASTM standards.
- Literature Review: Aggregated clinical outcomes reported in peer-reviewed publications.
8. Sample Size for the Training Set
- Animal Study: 6 Hemospray® treated animals, 3 sham control (total N=9). While not a "training set" in the machine learning sense, this GLP porcine study served as crucial in vivo data for validating effectiveness and safety prior to widespread human use.
- Biocompatibility: Various numbers for different tests (e.g., guinea pigs for sensitization, rabbits for pyrogenicity, cell lines for cytotoxicity). The "subchronic toxicity and implantation" was evaluated in the porcine animal study.
- Non-clinical/Bench Studies: Performed on device prototypes and finished devices; specific sample sizes for each bench test are not always detailed but are implicit in the "Passed" results.
- First-in-Human Pilot Study (20 patients): Could be considered early "training" or "feasibility" data for refining clinical protocols and understanding human safety.
9. How the Ground Truth for the Training Set was Established
- Animal Study (Porcine Model): Ground truth for hemostasis was established by direct observation during the procedure and confirmed by follow-up. Safety endpoints (gross and histologic assessment of regional and systemic effects) were established by detailed pathological examination of tissues collected at termination by trained veterinary pathologists.
- Biocompatibility Testing: Ground truth established according to standardized international (ISO, ASTM) and national (USP) testing protocols, which define specific biological responses (e.g., non-cytotoxic, non-irritant) under controlled laboratory conditions.
- Bench Testing: Ground truth established through engineering specifications and predefined pass/fail criteria for physical and functional parameters of the device and its components, measured using calibrated equipment.
{0}------------------------------------------------
DE NOVO CLASSIFICATION REQUEST FOR HEMOSPRAY® ENDOSCOPIC HEMOSTAT
REGULATORY INFORMATION
FDA identifies this generic type of device as:
Hemostatic device for intraluminal gastrointestinal use: A hemostatic device for intraluminal gastrointestinal use is a prescription device that is endoscopically applied to the upper and/or lower gastrointestinal tract and is intended to produce hemostasis via absorption of fluid or by other physical means.
NEW REGULATION NUMBER: 21 CFR 878.4456
CLASSIFICATION: Class II
PRODUCT CODE: OAU
BACKGROUND
DEVICE NAME: Hemospray® Endoscopic Hemostat
SUBMISSION NUMBER: DEN170015
DATE OF DE NOVO: March 9, 2017
CONTACT: Wilson-Cook Medical, Inc. 4900 Bethania Station Road Winston-Salem, NC 27105
INDICATIONS FOR USE
The COOK Hemospray® Endoscopic Hemostat is used for hemostasis of non-variceal gastrointestinal bleeding.
LIMITATIONS
The sale, distribution, and use of the Hemospray® Endoscopic Hemostat is restricted to prescription use in accordance with 21 CFR 801.109. Please refer to the labeling of the device for a complete list of warnings, precautions, and contraindications. Limitations on device use include the following statements in the Instructions for Use:
Contraindication for gastrointestinal endoscopy should also be a contraindication for the use of Hemospray®.
Hemospray® Endoscopic Hemostat is also contraindicated in patients who have
{1}------------------------------------------------
gastrointestinal fistulas, are suspected of having a gastrointestinal perforation, or are at high risk of gastrointestinal perforation during endoscopic treatment.
DEVICE DESCRIPTION
(b)
The Hemospray® Endoscopic Hemostat is a prescription use device consisting of a hemostatic agent and a delivery system. The hemostatic agent is (b) (4) | bentonite powder, a naturallysourced aluminum phyllosilicate clay. The delivery system is an endoscopic accessory used for spraying the powder onto the bleeding surface. The delivery device consists of a 220cm polyethylene application catheter, a handle with a pressurized CO2 cartridge, and a powder chamber containing approximately 20g of the Hemospray® material is propelled through the application catheter by release of CO2 from the cartridge located in the device handle. A trigger valve with an activation button allows the physician to control the amount of powder delivered to the affected area. Each actuation of Hemospray® delivers of CO2 and (b) gram of (b) (4) bentonite. Use of the complete 20-gram canister can produce a volume increase within the lumen of the bowel of 3 liters. In the presence of blood, the (b) (4) bentonite may swell 10-15% in volume.
The delivery system is offered in two configurations (i.e., two different outer diameters of the powder delivery catheter). The HEMO-7 version is used with a 2.8mm endoscope accessory channel, and the HEMO-10 version is used in a 3.7mm endoscope accessory channel.
Image /page/1/Picture/4 description: This image shows a medical device with several labeled parts. The device has a handle, a CO2 primer, a CO2 cartridge, and a regulator. There is also a trigger valve, an on/off valve, a powder canister, and a 7 FR or 10 FR PE delivery catheter.
{2}------------------------------------------------
SUMMARY OF NONCLINICAL/BENCH STUDIES
The sponsor conducted a series of non-clinical performance tests to demonstrate that the Hemospray® Endoscopic Hemostat would perform as anticipated for its intended use population.
BIOCOMPATIBILITY
The patient-contacting materials of the Hemospray® Endoscopic Hemostat include the (b) (4) bentonite powder and the polyethylene delivery catheter. Hemospray® powder is classified as an implant with prolonged tissue contact and indirect contact with blood. The following biocompatibility testing was completed on the Hemospray® powder in accordance with FDA's guidance document entitled "Use of ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process": cytotoxicity, irritation, sensitization, acute systemic toxicity, materials mediated pyrogenicity and hemolysis testing. The subchronic toxicity and implantation endpoints were evaluated in a clinically relevant porcine gastric bleeding animal study. Genotoxicity was evaluated through chemical characterization, toxicological risk assessment and establishment of heavy metal purity specifications. All studies were conducted in compliance with GLP requirements and testing was completed on the final sterilized device. All tests met the acceptance criteria and were supportive of the overall biocompatibility conclusion that the Hemospray® product is expected to be biocompatible when used as indicated. A summary of the biocompatibility testing completed can be found below in Table 1.
The delivery catheter is classified as an externally communicating device with less than 24-hours tissue contact. The following biocompatibility testing was completed on the delivery catheter: cytotoxicity, irritation and sensitization. A summary of the biocompatibility testing completed on the delivery catheter can be found below in Table 2.
| BiocompatibilityEndpoint | Method and Purpose | Result |
|---|---|---|
| Cytotoxicity | ISO 10993-5: MEM Elution Study used to evaluatedevice extracts for cytotoxicity risks. | Non-cytotoxic |
| Sensitization | ISO 10993-10: Guinea Pig MaximizationSensitization Test used to evaluate device extracts fordermal sensitization risks. | Non-sensitizer |
| Irritation | ISO 10993-10: Intracutaneous Irritation Test used toevaluate device extracts for irritation risks. | Non-irritant |
| Acute SystemicToxicity | ISO 10993-11: Acute systemic toxicity study used toevaluate device extracts for systemic toxicity risks. | No acute systemictoxicity |
| Material mediatedpyrogenicity | USP <151>: Rabbit pyrogen test used to evaluatedevice extracts for pyrogenicity risks. | Non-pyrogenic |
| Hemolysis | ASTM: F756: Hemolytic study used to evaluatedevice extracts for hemocompatibility risks. | Non-hemolytic |
| Table 1. Biocompatibility Testing/Assessment Completed on Hemospray ® Powder | |
|---|---|
| -- | ------------------------------------------------------------------------------ |
{3}------------------------------------------------
| Genotoxicity | Genotoxicity risks were evaluated through chemicalcharacterization, toxicological risk assessment andestablishment of heavy metal purity specifications. | Genotoxicity risksmitigated |
|---|---|---|
| SubchronicToxicity andimplantation(Animal study) | GLP porcine gastric bleeding model used toevaluate the Hemospray® product of local andsystemic toxicity risks in a clinically relevant animalmodel. | No local orsystemic toxicity |
Table 2. Biocompatibility Testing/Assessment completed on Hemospray® Delivery Device
| BiocompatibilityEndpoint | Method and Purpose | Result |
|---|---|---|
| Cytotoxicity | ISO 10993-5: MEM Elution Study used to evaluate deviceextracts for cytotoxicity risks. | Non-cytotoxic |
| Sensitization | ISO 10993-10: Guinea Pig Maximization Sensitization Testused to evaluate device extracts for dermal sensitizationrisks. | Non-sensitizer |
| Irritation | ISO 10993-10: Intracutaneous Irritation Test used toevaluate device extracts for irritation risks. | Non-irritant |
SHELF LIFE/STERILITY
Sterilization validation, packaging validation, and shelf life testing completed for the device can be found below in Table 3.
Table 3. Sterilization Validation, Packaging Validation, and Shelf Life Testing Overview for Hemospray® Endoscopic Hemostat
| Test | Purpose | Result |
|---|---|---|
| Sterilization ValidationISO 11137:2006/AAMI TIR 33:2005 | Verify product sterility (gammairradiation) | Passed |
| Packaging ValidationVisual Inspection | Visual assessment of product packagingto ensure no defects | Passed |
| Packaging ValidationDye Leak Test (ASTM F 1929(R2012), (R2015)) | Demonstrate that the packaging is notcompromised | Passed |
| Packaging ValidationBurst Testing (ASTM F 1140-07, R2012) | Assess the strength of the packagingseal | Passed |
| Shelf Life - Device StabilityVisual Inspection | Visual assessment of product to ensurethat it is not compromised | Passed |
| Shelf Life - Device StabilityFunctional Testing | Evaluate the ability of the device tomeet functional requirements after aging | Passed |
MAGNETIC RESONANCE (MR) COMPATIBILITY
Hemospray® has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration, or image artifact in the MR environment.
{4}------------------------------------------------
Hemospray® powder is composed of naturally occurring bentonite clay, which may contain elements containing metals. The safety of Hemospray® in the MR environment is unknown. Scanning a patient who has this device may result in patient injury or image artifact. No MR safety, device-related adverse effects have been observed in clinical use.
PERFORMANCE TESTING - BENCH
A summary of the performance bench testing completed on the Hemospray® Endoscopic Hemostat can be found below in Table 4.
| Test | Purpose | Result |
|---|---|---|
| Outputpressure | Demonstrate that the output pressure at the cathetertip and at the recommended use distance is belowthe specified threshold and therefore mitigates therisk of powder entering the bloodstream | Passed |
| Functionaltesting | Demonstrate that the user can set up the device foruse, insert and remove the device from a compatibleendoscope, adjust the system to reach the targetedsite, detach and reattach catheters during use, pressthe delivery button, and deploy powder to thetargeted site with the Hemospray® device | Passed |
| RuptureTesting | Demonstrate that the Hemospray® handle will notrupture at or below working pressures | Passed |
| HumanFactorsTesting | Demonstrate that the Hemospray® device has anacceptable torque to detach/reattach catheters,torque to open/close valve and force to press button | Passed |
| PowderDensity | Demonstrate that the Hemospray® device candeliver powder to the desired location, that thecatheter will not clog, and that powder will not bedelivered intermittently | Passed |
| CartridgeIntegrity | Demonstrate that the Hemospray® introducercomponent meets functional requirements aftersimulated shipping test | Passed |
| Ex vivotissuetrauma | Evaluate local effects of Hemospray® whendelivery catheter is positioned at recommended usedistance | No significantpathological alterationswere observed.Hemospray® did notpenetrate the mucosallayer |
Table 4. Performance Bench Testing Overview of Hemospray® Endoscopic Hemostat
MATERIALS CHARACTERIZATION
A summary of the materials characterization completed on the Hemospray® Endoscopic Hemostat can be found below in Table 5.
{5}------------------------------------------------
| Test | Purpose | Result |
|---|---|---|
| Mineral phaseidentification andquantification by x-raydiffraction (XRD) on aper-lot basis | Establish the compositionalidentity and purity of Hemospray®powder. Measurements will ensurethe compositional identity andpurity of each lot is withinspecifications. | Compositional identityand purity wereestablished andallowed lot-to-lotacceptance criteria tobe set. |
| Free Swell Test andMoisture Content | Establish that Hemospray®powder's extent of waterabsorption and moisture contentmeet specifications. Measurementswill ensure the swellingcharacteristics and moisturecontent of each lot are withinspecifications. | Passed |
| Extraction in simulatedgastric fluid for lead andarsenic on a per lot basis | Demonstrate that bioavailable leadand arsenic levels are within safelimits for each lot per limitsidentified in the USP monographfor Bentonite. | Passed |
| Particle SizeCharacterization | Characterize the particle sizedistribution of Hemospray®powder to ensure that each lot canbe adequately sprayed. | Results demonstrateconsistent particle sizedistribution across lots. |
Table 5. Materials Characterization Overview of Hemospray® Endoscopic Hemostat
PERFORMANCE TESTING - ANIMAL
The safety and effectiveness of Hemospray® Endoscopic Hemostat was evaluated in a GLP porcine gastric bleeding model. The study is summarized below:
Table 6. Performance Animal Testing Overview of Hemospray® Endoscopic Hemostat
| Test | Purpose | Method | Results |
|---|---|---|---|
| GLPPorcineStudy | Assesshemostaticability(effectiveness)and safety ofthe device | Design: Gastricbleeding model.Animals: 6Hemospray®treated, 3 shamcontrol.Duration: 10-dayfollow-up followedby termination | Acute hemostasis was achieved in all animals inthe treatment group (6/6). One control animal(1/3) and two (2/6) treatment group animalsshowed re-bleeding 8 to 10 days post-operatively.The re-bleeding observed were most likelyassociated with limitations of the chosen surgicalprocedure. Histopathological evaluation of thedevice application site did not reveal anyimportant histological differences betweentreatment group and control group animals.No evidence of Hemospray® powder was |
{6}------------------------------------------------
| Endpoints: grossand histologicassessment ofregional andsystemic effects ofpowder/treatment | observed in any local or regional tissue sectionsof the gastric surgical site or the regional lymphnodes. No Hemospray® powder particles wereobserved to have been embedded in any tissuelayers in any of the three histological sectionscollected at and around the treatment site. | |
|---|---|---|
| -- | ---------------------------------------------------------------------------------------------------------------- | ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Results of the animal study suggest that the Hemospray® Endoscopic Hemostat is effective at achieving initial hemostasis in a worst-case gastric bleeding model (Forrest 1a and 1b peptic ulcer tvpe bleeding). This study does not demonstrate safety/effectiveness in large area low-pressure bleeding where the probability of embolization may be more likely nor did it assess the potential rate of delayed re-bleeding.
SUMMARY OF CLINICAL INFORMATION
Wilson-Cook Medical, Inc. has sponsored a pre-market pilot study, a survey and 2 postmarket clinical studies of Hemospray® to date, all performed outside the United States (OUS). The pilot study was a "first-in-human" feasibility study to determine performance and safety issues associated with the endoscopic use of Hemospray® powder as a hemostatic material to treat bleeding peptic ulcers. This study was conducted in Hong Kong, China from 2009 to 2010.
The Survey to Evaluate the Application of Hemospray® in the Luminal Tract (SEAL), was conducted in Canada, Denmark, England, France, Germany, Italy and Holland.
The first post-market study. Hemostasis of Active GI Luminal Tract Bleeding (HALT) trial. was conducted in the European Union (EU). It was intended to study the safety and effectiveness of Hemospray® in treating upper gastrointestinal bleeds (UGIB), specifically Forrest 1a and 1b actively bleeding ulcers. Peptic ulcers are the most common cause of UGIB and are usually classified as spurting/pulsatile (1a) or oozing (1b) according to the Forrest scale. They account for up to 50% of peptic ulcer bleeds. The study began in 2011 with an intent to enroll 80 patients; 64 patients have been enrolled to date. The HALT study continues to enroll patients in the EU and Canada.
Finally, the second post-market study, APPROACH (A Prospective Observational Cohort Study of Hemospray® for Lower Gastrointestinal Hemorrhage), was conducted in Canada and designed to collect safety and performance data on non-variceal lower gastrointestinal bleeding (NVLGIB). APPROACH has recently completed enrolling its entire complement of 50 patients at 4 Canadian sites. The results of all the studies are presented below.
PILOT STUDY
This pilot clinical trial was designed as a single-center, early feasibility study of Hemospray® powder for the treatment of bleeding peptic ulcers. The aim of the study was to investigate the safety and effectiveness of Hemospray® in achieving acute hemostasis in peptic ulcer bleeding in subjects diagnosed with Forrest 1a or 1b bleeding peptic ulcers. The protocol for this prospective trial was approved by the ethics committee of the faculty of medicine of the Chinese University
{7}------------------------------------------------
of Hong Kong. China and conducted in accordance with the Declaration of Helsinki. International Conference on Harmonization, Good Clinical Practices, and other regulations in force in Hong Kong. Written informed consent was obtained from all patients were to undergo upper GI endoscopy within approximately 24 hours of hospital admission. Each patient was allowed up to 2 applications of Hemospray® (potentially using multiple canisters during each application) during the index procedure, not exceeding a total of 150 g. Recurrent bleeding was monitored post-procedure and at a second-look endoscopy (72 hours posttreatment) was conducted on all patients. Patients were contacted by phone at 30 days to monitor the occurrence of any serious adverse events and mortality.
The safety and performance outcome measures of this study were:
Primary safety endpoint:
-
The incidence of procedural and treatment-related serious adverse events for subjects . treated with Hemospray®.
Primary effectiveness endpoint: -
Proportion of subjects with acute (procedural) hemostasis. .
-
Rate of recurrent bleeding within 72 hours of the application of Hemospray®
Secondary Outcome Measures
- Severe adverse events and mortality within 30 days of the study procedure .
- . Surgery required for failure to achieve hemostasis within 72 hours
Rate of hemostasis, rate of recurrent bleeding, need for surgical intervention, treatment-related and procedure-related serious adverse events and major adverse events (e.g., arterial embolization, allergic reaction, bowel obstruction) were also assessed.
Twenty patients were enrolled in the study. An interim analysis was conducted on data collected from the first 10 patients. Since no major adverse events related to the Hemospray® powder or delivery catheters were reported. 10 additional patients were enrolled. The study consisted of 18 males and 2 females. Mean age was 60.2 years (range 37-85 y). The first patient was enrolled in the study on November 17, 2009 and the last patient was enrolled on August 27, 2010. The final follow-up visit was on September 27, 2010.
All 20 patients included in this study presented with melena at hospital admission, 7 of whom also had concurrent hematemesis. Nineteen patients were diagnosed with a Forrest 1b ulcer and 1 patient had Forrest 1a bleeding. Most ulcers were in the duodenum 14 (70%) and the remaining 6 (30%) were in the stomach. A single application of Hemospray® was used in 17 (85%) patients, and two applications were used in 3 (15%) patients. The number of syringes of Hemospray® used for each patient ranged between 1 and 7, with most patients (13/20, 65%) receiving only 1 syringe (20g). Of the remaining 7 patients, 5 received 2 syringes, one patient received 4 syringes, and one patient received 7 syringes.
{8}------------------------------------------------
Results:
Endoscopic application of Hemospray® onto the ulcer bleeding site was successfully performed in all patients. Acute hemostasis was achieved in 19 of 20 patients (95%). There were no treatment-related or procedure-related serious adverse events. All patients were stable after the procedure and recovered uneventfully. The one patient in which acute hemostasis was not achieved was treated with 7 syringes of Hemospray® or 140 grams in one application. This patient was then treated according to institutional standard of care (endoscopic epinephrine injection and hemostasis clips) in 3 additional unsuccessful attempts to achieve endoscopic hemostasis and was referred for angiography during which a pseudoaneurysm was identified in the artery feeding the ulcer site. Ultimately, TAE (trans-arterial embolization) of the feeding artery was required to stop bleeding in this patient. The patient was contacted at 30 days and reported no serious adverse events.
Of the 19 patients in which acute hemostasis was achieved, sustained hemostasis as defined by the study was observed in 17 (89.5%) patients through 72 hours. Two patients met the study definition of recurrent bleeding or re-bleeding within 72 hours. However, in both patients, no active bleeding was observed at the treated lesion sites at the planned second-look endoscopy. Both patients reported no adverse event or serious adverse event at 30-day follow-up.
Safety outcome:
The Hemospray® powder was found to have been eliminated from the stomach and duodenum in all patients at the second-look endoscopy, conducted on Day 2 for one patient and Day 3 for all others. None of the patients, including the patient who received 140 grams or 7 syringes reported any adverse issues with passage of the powder through the gastrointestinal tract (e.g., no bowel obstruction). All patients were stable after the procedure and discharged uneventfully. There was no mortality reported during the 30-day follow-up. No treatment-related or procedurerelated serious adverse events and no major adverse events (e.g., embolization, bowel obstruction, or allergic reaction) were reported. While there were no reports of serious adverse events related to either the treatment or the procedure, 5 cases of non-device related serious adverse events were reported. These included 4 cases of prolonged hospitalization: 2 due to upper respiratory tract infection, 1 due to altered liver function and psychotic depression, and 1 patient who required more time to recover and regain stable blood values. The fifth event was the diagnosis of adenocarcinoma that was initially diagnosed as a gastric ulcer during the Hemospray® treatment procedure. In this last patient, a large gastric ulcer oozing from the base was found during endoscopy. There were some features suggestive of malignancy and biopsy was performed. The bleeding was successfully controlled by Hemospray®, and no Hemospray® powder was observed at the 72-hour endoscopy at the treatment site. Histology subsequently confirmed malignancy, and this patient was referred to a medical oncologist for further management. The only observations of relevance were clogging of the Hemospray® application catheter, but in each case a new catheter was used and the procedure completed.
SURVEY TO EVALUATE THE APPLICATION OF HEMOSPRAY IN THE LUMINAL TRACT (SEAL)
Wilson-Cook Medical, Inc. initiated the post-market Survey to Evaluate the Application of Hemospray® in the Luminal Tract (SEAL) to collect limited, non-protected health information
{9}------------------------------------------------
and data on the Wilson-Cook Medical, Inc. Hemospray® Endoscopic Hemostat as part of a limited marketing release OUS. The information was used to develop educational materials for physicians new to this technology. Hemospray® used endoscopically by gastroenterologists provided effective acute hemostasis in the most common types of gastrointestinal (GI) bleeding, including Forrest 1a and 1b bleeding peptic ulcers.
As a requirement of participation in the SEAL registry, all physician participants were trained on the use of the Hemospray® kit by Wilson-Cook Medical, Inc. personnel. Training included discussion of hemostasis techniques, appropriate intended use and product labeling in the countries where the evaluation took place, and hands-on experience with the Hemospray® device. Each participating physician group was assigned a unique sign-on to access an electronic data capture form hosted on a secure internet portal. All cases entered into the database were numbered sequentially in the order of entry. No queries or audits to identify and correct missing data were performed. Summaries generated from the collected information were, therefore, constrained by the scope of the information collected, and are limited to counts of cases, common location of bleeding sites, severity of the bleeding treated, and other methods to achieve hemostasis if used in conjunction with Hemospray®. Complaints and adverse events were to be reported directly to the manufacturer. Ninety-two patients treated in Canada, Denmark, England, France, Germany, Italy and Holland were registered in this survey. Two patients had no data entered in the electronic database by the site other than age and sex. In a third patient, the CO2 cartridge in the Hemospray® delivery system was found to be defective so the device was not used and the patient was not treated with Hemospray®. These 3 patients were excluded from analysis. Therefore, evaluable data from 89 patients were available for analysis.
Eight-four percent (84%) of patients were over the age of 50 (n = 75), and 71% (n = 63) were male. Acute hemostasis was achieved in all (100%) cases. Approximately 98% of the bleeding sites were in the upper GI tract and were identified in the duodenum (n = 38), stomach (n = 29), and esophagus (n = 20). Peptic ulcers constituted most cause of GI bleeding treated with Hemospray® (n = 40; 45%). Of the 40 patients with peptic ulcers, 16 were Forrest 1a bleeds (spurting or pulsatile), 19 were Forrest Class 1b (oozing), 3 were Forrest 2b (adherent clot), and 2 patients were unclassified. In one Forrest 2b case, a procedural video confirmed the clot was removed prior to treatment with Hemospray®. It is not known if the other 2 cases were treated in a similar fashion. Other conditions treated with Hemospray® included: bleeding after endoscopic mucosal resection or dissection (n = 9; 10%); diffuse bleeding from gastric malignancy (n = 8; 9%); Mallory-Weiss tears (n = 6; 7%); and upper GI post-polypectomy bleeding (n = 5; 6%); Hemospray® was used successfully in the colon for acute hemostasis in 2 cases.
Fifty-one (51) patients (57%) were successfully treated with Hemospray® alone, and the remaining 38 patients (43%) were successfully treated with Hemospray® as an adjunct to other standard of care methods. Hemospray® was identified as the final hemostasis treatment in 24 of 38 cases where more than one other method was used. In 5 patients, Hemospray® was used as the initial treatment and was followed by alternate methods including epinephrine injection, argon-plasma coagulation, bi-polar electrocautery, and hemoclips. In 9 patients, the order of use was unclear. Hemospray® was used as a single modality in 51 patients. Of the 51 patients treated with Hemospray® alone, 11 were diagnosed with Forrest 1a bleeds, 38 with Forrest 1b, and 2 with Forrest 2b. Acute hemostasis was achieved with Hemospray® alone in all 51 of
{10}------------------------------------------------
these patients. The remaining 38 patients were treated with Hemospray® and at least one other modality; 16 of these with Forrest 1a, 19 with Forrest 1b, 1 with Forrest 2b, and 2 were unclassified. Acute hemostasis was achieved in all 38. There were no unanticipated adverse events or serious adverse events attributed to the Hemospray® Endoscopic Hemostat. Holster I.L. et. al., published a study of patients in the SEAL survey with upper gastrointestinal bleeding who were on antithrombotic therapy which included mostly patients on antiplatelet agents and a few on anticoagulation therapy. Patients had 63% initial hemostasis with a 38% rebleeding rate following Hemospray® treatment.
HEMOSTASIS OF ACTIVE GI LUMINAL TRACT BLEEDING (HALT) STUDY
The HALT trial was designed to study the safety and effectiveness of Hemospray® in treating upper GI bleeds (UGIB), specifically Forrest 1a and 1b actively bleeding ulcers. The HALT trial continues to enroll patients in Canada and Europe with a final goal of 80 enrollments. An interim report of the study findings at 64 enrollments is presented as part of this submission. The data lock was instituted on August 24, 2016. The HALT study was initiated with Version 1 of the Hemospray® device (55 psi CO2 canisters), but all sites were converted to the Version 2 device (37 psi CO2 canisters) effective January 2013. There are 10 enrollment sites.
| %(n/N) | |
|---|---|
| Hypertension | 47.6% (30/63a) |
| Diabetes | 27.0% (17/63a) |
| Patient already hospitalized for another illnesswhen GI bleeding occurred | 10.9% (7b/64) |
| Previous treatment(s) for peptic ulcer bleedingc | 22.2% (14/63a) |
| Hemostasis clips | 35.7% (5/14) |
| Injection | 42.9% (6/14) |
| Thermal probe | 28.6% (4/14) |
| Upper GI surgery | 0.0% (0/14) |
| Other Endoscopic/surgical treatment | 7.1% (1/14) |
| Transfusion | 64.3% (9/14) |
| IV administration of PPI | 64.3% (9/14) |
| Oral administration of PPI | 21.4% (3/14) |
| Other Medical treatment | 0.0% (0/14) |
| Current symptoms of peptic ulcer bleedingd | 98.4% (62/63a) |
| Hematemesis | 22.6% (14/62) |
| Hematochezia | 12.9% (8/62) |
| Melena | 91.9% (57/62) |
Table 7: Patient Comorbidities
a Data were outstanding at time of data lock for one patient.
{11}------------------------------------------------
b Prior hospitalization for chest, abdominal, and loin pain (1); fall(s) (2); syncope related to GI bleed (1); low
hemoglobin, chest pain, shortness of breath (1); prior GI bleed (1); loose stools and generally unwell (1).
6 Patients may have more than one treatment on prior admissions for peptic ulcer bleeding.
d Patients may have more than one current symptom of peptic ulcer bleeding.
| Lesion size (diameter)a,b | |
|---|---|
| < 1 cm | 32.3% (20/62) |
| 1-2 cm | 54.8% (34/62) |
| > 2 cm | 12.9% (8/62) |
| Lesion locationb | |
| Stomach | 25.4% (16/63) |
| Duodenum | 74.6% (47/63) |
| Forrest classificationb | |
| 1a (spurting) | 15.9% (10/63) |
| 1b (oozing) | 84.1% (53/63) |
Table 8. Baseline Bleed Site Characteristics
ª One patient did not have an ulcer diameter measurement.
b Data were outstanding at time of data lock for one patient.
Primary Endpoint
The primary effectiveness endpoint is the proportion of patients with further bleeds within 72 hours of the index procedure. Further bleed within 72 hours is defined as persistent bleeding (bleeding at the conclusion of the index endoscopy: 7 patients did not achieve initial hemostasis with Hemospray®) or recurrent bleeding (visualization of bleeding at the treated study lesion [assessed only in patients with successful initial hemostasis]) within 72 hours.
Table 9. Primary Endpoint
| % (n/N) | |
|---|---|
| Patients with further bleeds within 72 hours | 17.2% (11/63a) |
| Persistent bleeding after Hemospray application(s) | 11.1% (7/63a) |
| Endoscopically-confirmed recurrent bleed within 72 hours | 7.1% (4/56b) |
a Data were outstanding at time of data lock for one patient.
b Recurrent bleed was only assessed in patients with successful initial hemostasis (i.e., patients without persistent bleeding).
To further investigate if demographics or baseline characteristics affected the primary endpoint, persistent bleeding, or early recurrent bleed, a stepwise logistic regression was performed using sex, ulcer location, ulcer diameter, Forrest score, and hypertension as variables (criterion was p<0.3 for inclusion in the model; criterion for removal was p>0.35). No variables were found to significantly affect the primary endpoint of persistent bleeding; however, stepwise logistical regression indicated that patients with Forrest 1a bleeds are more likely to have an early recurrent bleed than patients with Forrest 1b bleeds. Further, patients with Forrest 1a bleeds are
{12}------------------------------------------------
more likely to have any recurrent bleed (early or late) than patients with Forrest 1b bleeds based on stepwise logistical regression.
Secondary Endpoints
Secondary endpoints include: proportion of patients with hemostasis at the conclusion of the index procedure (i.e., initial hemostasis), clinical success (i.e., initial hemostasis and no SAE within 72 hours of the index procedure), early recurrent bleed (i.e., recurrent bleeding within 72 hours of the application of Hemospray®), late recurrent bleed (i.e., recurrent bleed occurring 72 hours-30 days), incidence of serious adverse GI events within 30 days of the application of Hemospray®, incidence of serious adverse events within 30 days of the application of Hemospray® and incidence of mortality at 30 days.
| Number | Percent | |
|---|---|---|
| Patients achieving hemostasis during index endoscopy | 61 | 96.8 |
| Successful hemostasis achieved after Hemospray® application | 56 | 88.9 |
| Re-bleeding within 72 hours after successful initial hemostasis | 4 | 7.1b |
| 30-day all-cause mortality | 2a | 3.2 |
| Patients with device-related Serious Adverse Events within 30 days of index procedure | 0 | 0.0 |
Table 10. HALT Study Secondary Endpoints
4First patient cause of death was liver failure. Second patient cause of death was pneumonia.
b Re-bleeding was only assessed in patients with successful initial hemostasis (n=56).
Of the twelve patients with serious adverse events, one patient had an intra-operative perforation reported. This patient died two days post-procedure due to liver failure. Ten patients had clinical signs and symptoms of re-bleeds (seven with confirmed recurrent bleeding at the study lesion site, one which endoscopically confirmed that the initial study lesion was not involved, and two which were not considered to be a recurrent bleed at the study lesion site by the physician and therefore did not have repeat endoscopy performed). Of note, clinical signs and symptoms of rebleed are a common adverse event in patients recently treated for peptic ulcer bleeds. One patient had no signs of re-bleeding, however, developed pneumonia that resulted in death.
| Table 11. Adverse Events by Organ System | |||||
|---|---|---|---|---|---|
| ------------------------------------------ | -- | -- | -- | -- | -- |
| Organ System | 0-3 Days (n) | 4-7 Days (n) | 8-30 Days (n) |
|---|---|---|---|
| Gastrointestinal | 15 | 5 | 5 |
| Cardiovascular | 0 | 0 | 1 |
| Pulmonary | 0 | 0 | 0 |
| Miscellaneous | 4 | 2 | 4 |
| Table 12. Gastrointestinal Events | |
|---|---|
| -- | ----------------------------------- |
| Event | 0-3 Days (n) | 4-7 Days (n) | 8-30 Days (n) |
|---|---|---|---|
| ------- | -------------- | -------------- | --------------- |
{13}------------------------------------------------
| Clinical signs and symptoms of re-bleed | 11 | 5 | 4 |
|---|---|---|---|
| Other | 2 | 0 | 1 |
| Perforation | 2 | 0 | 0 |
Specifically, 11 events of clinical signs and symptoms of re-bleeding in 10 patients and 2 perforations were considered serious adverse events. Of the reports of re-bleeding, two were not considered recurrent bleeds because hemostasis had not been achieved with Hemospray® during the initial procedure. No treatment or repeat endoscopy was performed for either patient and neither patient had further reports of clinical signs and symptoms of re-bleeding throughout the follow-up period. In three additional cases that were not considered recurrent bleeds, correspondence between Wilson-Cook Medical, Inc. and the study physician indicated that the clinical signs and symptoms of re-bleeding were attributable to the original bleed, rather than recurrent bleeding; therefore, additional endoscopy was not performed. None of the signs or symptoms were treated (although one patient was instructed to take an iron supplement), and no further reports of clinical signs or symptoms of re-bleeding were submitted throughout the remainder of the study follow-up. In one case that was not considered recurrent bleeding, it was reported that re-bleed from the treated site was not suspected. One individual with hypotension and melena after normalization of stool color was not re-scoped as the study site reported that the patient was not well enough to undergo repeat endoscopy and the patient's melena had resolved on the same day. The remaining patient with clinical signs and symptoms of re-bleeding did have recurrent bleed with endoscopic confirmation of study lesion involvement, which was treated with epinephrine injection, hemostasis clips, and transfusion.
Initial hemostasis was achieved using Hemospray® as a single-modality treatment method in 88.9% (56/63) of cases. Initial hemostasis was defined as patients with hemostasis at the conclusion of the index procedure, where 'index procedure' is considered to be the application of Hemospray® and a 5-minute observation period. Patients that did not achieve initial hemostasis with Hemospray® were treated with hemostasis clips (1 patient), injection (with epinephrine) and hemostasis clips (2 patients), injection and argon beam therapy (1 patient), injection and thermal probe (1 patient), conversion to surgical repair (1 patient is described further in Table 10, HALT Summary Study Data), and proton pump inhibitor (PPI) therapy (1 patient). Subsequent events of perforation and death were reported for 1 patient.
Fifty-five patients completed their 30-day follow-up; eight patients exited the study prior to completion of 30-day follow-up. Of these eight, one patient died 2 days after the procedure, and one patient died 18 days after the procedure. Both deaths occurred after surgical intervention, one for bowel perforation and one for re-bleeding, resulting in death from postoperative liver failure and pneumonia, respectively. Six other patients were lost to follow up prior to completing the study follow-up schedule.
A PROSPECTIVE OBSERVATIONAL COHORT STUDY OF HEMOSPRAY™ FOR LOWER GASTROINTESTINAL HEMORRHAGE (APPROACH)
The second study is the APPROACH study in Canada. APPROACH is designed to collect safety and performance data on NVLGIB. Although the rate of occurrence and mortality in LGIB is lower than UGIB, bleeding in the lower GI tract can become clinically significant and/or
{14}------------------------------------------------
exacerbate existing co-morbidities. This prospective, single-arm, post-market study collected data on the safety and performance of the Hemospray® device when used as a treatment to achieve hemostasis of non-variceal lower GI bleeding. The study enrolled 50 patients at 4 clinical sites in Canada. Patient enrollment and data collection were completed in October and November 2016, respectively. The final site monitoring visit was completed in February 2017. Results are presented below.
| Number | Percent | |
|---|---|---|
| Number of sites | 4 | na |
| Patients enrolled | 50 | na |
| Patients achieving hemostasis during index endoscopy | 50 | 100.0 |
| Successful hemostasis achieved using Hemospray® | 49a | 98.0 |
| Patients with recurrent bleed as defined by the study protocol | 5 | 10.0 |
| 30-day all-cause mortality | 1b | 2.0 |
| Patients with device-related adverse events within 30 days ofthe index procedure | 0 | 0.0 |
Table 13. APPROACH Study Summary Data
a In one patient, hemostasis was not achieved with the initial, single modality use of Hemospray . Hemostasis clips were used also in order to achieve hemostasis.
b Includes 1 patient who died within 30 days. Cause of death: gastrointestinal bleeding (site unclear) secondary to pre-existing idiopathic thrombocytopenia.
| Table 14. Patient Demographics | |
|---|---|
| -------------------------------- | -- |
| Demographic Characteristic | Reported (n=50) |
|---|---|
| Age (years) | $64.6 \pm 12.5$ |
| Mean $\pm$ SD (number of patients, age range) | (50, 37 - 85) |
| Sex | |
| Male | 72.0% (36/50) |
| Female | 28.0% (14/50) |
Table 15. Pre-procedure Co-Morbid Conditions
| Co-morbid Conditions | Percent of Patients (n/N) |
|---|---|
| Patient already hospitalized for another illnesswhen lower GI bleeding occurred | 8.0% (4/50) |
| Previous episodes of GI bleeding | 56.0% (28/50) |
| Current symptoms of lower GI bleeding | 52.0% (26/50) a |
| Hematochezia | 24.0% (12/50) |
| Melena | 8.0% (4/50) |
{15}------------------------------------------------
| Chronic anemia | 6.0% (3/50) |
|---|---|
| Acute anemia | 6.0% (3/50) |
| Positive fecal occult blood test | 18% (9/50) |
a Patients may have more than one current symptom of lower GI bleeding.
| Hemospray® Use | Percent of Patients(n/N) | |
|---|---|---|
| Number of bleed sites treated per patient | 1 | 96.0% (48/50) |
| 2 | 4.0% (2/50) | |
| Index procedural use of Hemospray® | Initial intervention(Hemospray® asmonotherapy) | 26.0% (13/50) |
| Supplementalintervention(Hemospray® inconjunction with anotherintervention) | 40.0% (20/50) | |
| Rescue intervention(Hemospray® after failureof prior intervention) | 34.0% (17/50) | |
| Was Hemospray® applied to bleed site asintended? | Yes | 100% (50/50) |
Table 16. Hemospray® Use during Study Procedure
The Primary Endpoint is Hemospray -related adverse events occurring within 30 days of the index procedure. Hemospray® related adverse events include, but are not limited to, powder impaction in colon or embolization.
| Table 17. Primary Endpoint | - 2017 - 10 10000 | 1986 - 12 |
|---|---|---|
| ---------------------------- | ------------------- | ----------- |
| Hemospray® Related Adverse Eventswithin 30 Days of the Index Procedure | Percent ofPatients (n/N) |
|---|---|
| Yes | 0.0% (0/50) |
| No | 100% (50/50) |
Secondary Endpoints:
- · Hemostasis following application of Hemospray®
- · Recurrent bleeding within 30 days of the index procedure
- · Mortality within 30 days of the index procedure
{16}------------------------------------------------
| Endpoints | Percent of Patients (n/N) | |
|---|---|---|
| Hemostasis following application of Hemospray® | Yes | 98.0% (49/50) |
| No | 2.0% (1/50) | |
| Recurrent BleedingWithin 30 Days of Index Procedure | Yes | 10.0% (5/50) |
| No | 90.0% (45/50) | |
| Mortality Within 30 Days of Index Procedure | Yes | 2.0% (1/50) |
| No | 98.0% (49/50) |
Table 18. Secondary Endpoints
Hemostasis was achieved after the index procedural use of Hemospray® as an initial, supplemental, or rescue intervention treatment method in 97.3% of cases. Hemostasis, defined as the absence of persistent bleeding at the conclusion of the index colonoscopy, was achieved in all
but one patient who had a visibly oozing or spurting bleed.
| Percent of Data AvailableEndpoint Since Previous Time Period | ||||||
|---|---|---|---|---|---|---|
| Period | Patients Eligible for Follow-Upa | Telephone Follow-up | Death | Lost to Follow-up | Withdrawal | Not Eligible for Next Follow-up Visit |
| Procedure | 50 | N/A | 0 | 1 | 0 | 0 |
| 14-day | 49 | 98.0%(48/49) | 0 | 0 | 0 | 0 |
| 30-day | 49 | 98.0%(48/49) | 1 | N/A | N/A | N/A |
Table 19. Clinical Follow-up Data
a Number of patients eligible = previous number eligible - (death + LTF + withdrawal + not eligible for next follow-up visit).
Table 20. Comparison of APPROACH Study Results with American Society of Gastrointestinal Endoscopy (ASGE) Standard of Care
| Endpoint | APPROACH StudyResults | Standard of Care per ASGE |
|---|---|---|
| 30-day Mortality Rate | 2.0% | 2.0-4.0% |
| Hemostasis AchievedDuring IndexEndoscopy | 100.0% | 50-100.0% |
{17}------------------------------------------------
| Re-bleeding Rate | 10.0% | 18-22.0% |
|---|---|---|
| ------------------ | ------- | ---------- |
Serious Adverse Events
Of the six patients with serious adverse events, one patient died from gastrointestinal bleeding secondary to pre-existing chronic idiopathic thrombocytopenia 27 days after the procedure. Five cases of clinical signs and symptoms of lower GI bleeding were reported in four patients. Two of these cases were confirmed to have a recurrent bleed at the study lesion site, one case was conservatively assumed to be a recurrent bleed at the study lesion site as visual confirmation was not attained, and three cases were considered not to be recurrent bleeds. Two additional postoperative serious adverse events were reported in which clinical signs and symptoms of lower GI bleeding were not noted.
| Category | 0-14 Days (N) | 15-30 Days (N) |
|---|---|---|
| Gastrointestinal | 10 | 1 |
| Pulmonary | 0 | 0 |
| Cardiovascular | 0 | 0 |
| Miscellaneous | 4 | 1 |
Table 21. Patient Morbidity
Forty-eight patients completed their final follow-up. One patient exited the study prior to completion of the 14-day and 30-day follow-ups. One patient died 27 days after the procedure, as noted above. No other patients were lost to follow up prior to completing the study follow-up schedule.
Literature Summary of Recent Clinical Hemospray® Use
An extensive search was performed to identify the existing clinical literature on the recent use of Hemospray®. The search terms "Hemospray" or "TC-325" were used in PubMed and Google Scholar. Only publications in clinical journals were included in the analysis. Non-clinical studies, review articles, conference abstracts, and publications not in English were excluded. Publications that appeared to have overlapping patients with another study were excluded. Treatments of contraindicated bleeding types (e.g., varices or fistulas) were excluded from the analysis. Thirty (30) relevant studies were identified comprising the treatment of 522 patients with Hemospray®. Overall, the hemostasis rate after the use of Hemospray® was 96% and the endoscopically achieved hemostasis rate was 97%. The 30-day gastrointestinal bleeding-related mortality rate was 3%. Four potential device-related adverse events (0.8%) were identified in the literature (2 reports of perforation and 2 reports of aspiration pneumonia); however, none of these events were reported to be directly attributed to the use of Hemospray® in the publications.
{18}------------------------------------------------
| Number | Percent | |
|---|---|---|
| Number of studies | 30 | n/a |
| Patients treated | 522 | n/a |
| Hemospray® procedures | 532 | n/a |
| Successful hemostasis achieved using Hemospray® | 509 | 96% |
| Patients achieving hemostasis during index endoscopy | 518 | 97% |
| Re-bleeding within 7 days after successful initial hemostasis | 97 | 19% |
| Re-bleeding 8-30 days after successful initial hemostasis | 20 | 4% |
| 30-day all-cause mortality | 80 | 11% |
| 30-day gastrointestinal bleeding mortality | 23 | 3% |
| Potential device-related adverse eventsa | 4 | 0.8% |
Table 22. Hemospray® Clinical Literature Summary
a Potential device-related adverse events were 2 perforations and 2 cases of aspiration pneumonia. None of these adverse events could be directly attributed to the use of Hemospray®.
The literature reports consisted of the following levels of evidence:
- . Level II: 12 Registry Derived Studies
- Level IV: The remainder of the studies consist of level IV evidence single arm ● retrospective or prospective studies, case series and case reports.
For a total of 30 studies in which there were 532 Hemospray® applications, resulting in 96% hemostasis rate and an overall re-bleeding rate of 23%. There were 2 instances of bowel perforation which may have been attributed to the use of Hemospray® and 2 aspiration pneumonias which could not be conclusively attributed to Hemospray®. There were no reports of bowel powder impaction or thromboembolic events.
{19}------------------------------------------------
| Study | Patients(n) | HemosprayApplications | HemostasisafterHemosprayApplication | EndoscopicHemostasis | HemosprayHemostasisRate | EndoscopicHemostasisRate | Rebleedingat riskpatients | RebleedingEvents (7d) | RebleedingRate (7d) | RebleedingEvents(8-30d) | RebleedingRate(8-30d) |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Sung 2011 | 19 | 20 | 19 | 19 | 95% | 95% | 19 | 2 | 11% | 0 | 0% |
| Chen 2012 | 5 | 6 | 5 | 5 | 83% | 83% | 5 | 1 | 20% | 0 | 0% |
| Granata 2013 | 1 | 1 | 1 | 1 | 100% | 100% | 1 | 0 | 0% | 0 | 0% |
| Holster 2013 | 15 | 15 | 12 | 14 | 80% | 93% | 15 | 5 | 33% | 0 | 0% |
| Leblanc 2013 | 12 | 12 | 12 | 12 | 100% | 100% | 12 | 0 | 0% | 0 | 0% |
| Smith 2013 | 4 | 4 | 4 | 4 | 100% | 100% | 4 | 0 | 0% | 0 | 0% |
| Smith 2013 (SEAL) | 62 | 63 | 56 | 61 | 89% | 97% | 61 | 8 | 13% | 0 | 0% |
| Soulellis 2013 | 4 | 4 | 4 | 4 | 100% | 100% | 4 | 0 | 0% | 0 | 0% |
| Appleby 2014 | 1 | 1 | 1 | 1 | 100% | 100% | 1 | 0 | 0% | 0 | 0% |
| Chen 2014 | 67 | 67 | 66 | 67 | 99% | 100% | 63 | 6 | 10% | 9 | 14% |
| Curcio 2014 | 1 | 1 | 1 | 1 | 100% | 100% | 1 | 0 | 0% | 0 | 0% |
| Dietrich 2014 | 2 | 2 | 2 | 100% | 100% | 2 | 1 | 50% | 0 | 0% | |
| Holster 2014 | 9 | 9 | 9 | 9 | 100% | 100% | 9 | 2 | 22% | 0 | 0% |
| Kratt Endoscopy 2014 | 1 | 2 | 2 | 2 | 100% | 100% | 2 | 0 | 0% | 0 | 0% |
| Masci 2014 | 13 | 13 | 13 | 13 | 100% | 100% | 13 | 2 | 15% | 0 | 0% |
| Selvapatt 2014 | 1 | 1 | 1 | 1 | 100% | 100% | 1 | 0 | 0% | 0 | 0% |
| Sulz 2014 | 15 | 15 | 14 | 15 | 93% | 100% | 14 | 2 | 14% | 0 | 0% |
| Tarantino 2014 | 1 | 1 | 1 | 1 | 100% | 100% | 1 | 0 | 0% | 0 | 0% |
| Yau 2014 | 19 | 19 | 18 | 18 | 95% | 95% | 18 | 7 | 39% | 0 | 0% |
| Zimmer 2014 | 1 | 1 | 1 | 1 | 100% | 100% | 1 | 0 | 0% | 0 | 0% |
| Chen Endoscopy 2015 | 1 | 1 | 1 | 1 | 100% | 100% | 1 | 0 | 0% | 0 | 0% |
| Granata Int J Colorectal 2015 | 4 | 5 | 5 | 5 | 100% | 100% | 5 | 1 | 20% | 0 | 0% |
| Granata Endoscopy 2015 | 1 | 1 | 1 | 1 | 100% | 100% | 1 | 0 | 0% | 0 | 0% |
| Ivekovic 2015 | 1 | 1 | 1 | 1 | 100% | 100% | 1 | 0 | 0% | 0 | 0% |
| Giles 2016 | 36 | 36 | 36 | 36 | 100% | 100% | 36 | 4 | 11% | 0 | 0% |
| Gubler 2016 | 5 | 7 | 7 | 7 | 100% | 100% | 7 | 2 | 29% | 0 | 0% |
| Haddara 2016 | 202 | 202 | 195 | 195 | 97% | 97% | 191 | 51 | 27% | 11 | 6% |
| Sinha 2016 | 20 | 20 | 19 | 19 | 95% | 95% | 19 | 3 | 16% | 0 | 0% |
| Venezia 2016 | 1 | 1 | 1 | 1 | 100% | 100% | 1 | 0 | 0% | 0 | 0% |
| Xavier 2016 | 1 | 1 | 1 | 1 | 100% | 100% | 1 | 0 | 0% | 0 | 0% |
| Total | 522 | 532 | 509 | 518 | 96% | 97% | 510 | 97 | 19% | 20 | 4% |
Table 23. Summary of Clinical Literature on the use of Hemospray®
Successful hemostasis defined as hemostasis achieved during procedure where Hemospray resoue, or intila followed by additional modalities.
{20}------------------------------------------------
| Study | AdverseEvents(n) | UpperGI | LowerGI | Lesion Types (See Table # for identifiers) | 30-dayAll-causemortality(n) | 30-day GIbleedingmortality(n) | PrimaryTherapy(n) | SecondaryTherapy(n) |
|---|---|---|---|---|---|---|---|---|
| Sung 2011 | 0 | 20 | 0 | PU (20) | 0 | 0 | 20 | 0 |
| Chen 2012 | 0 | 5 | 0 | Tumor (5) | 1 | 1 | 5 | 0 |
| Granata 2013 | 0 | 0 | 1 | OU (1) | 0 | 0 | 0 | 1 |
| Holster 2013 | 0 | 15 | 0 | PU (9), Tumor (2), Div (1), DL (1), TR (1) | 0 | 0 | 6 | 11 |
| Leblanc 2013 | 0 | 12 | 0 | TR (11), post-EBS (1) | 0 | 0 | 8 | 4 |
| Smith 2013 | P | 3 | 1 | PH (4) | 1 | 1 | 3 | 1 |
| Smith 2013 (SEAL) | P, AP | 62 | 0 | PU (30), TR (9), Tumor (6), DL (1), Post-EBS (1), Div (1), Other (5) | 4 | 0 | 55 | 8 |
| Soulellis 2013 | 0 | 0 | 4 | TR (2), DL (1), RP (1) | 0 | 0 | 0 | 4 |
| Appleby 2014 | 0 | 1 | 0 | post-EBS (1) | 0 | 0 | 1 | 0 |
| Chen 2014 | 0 | 50 | 17 | PU (13), OU (4), Tumor (21), DL (2), RP (1), AVM (3) | 4 | 1 | ||
| Curcio 2014 | 0 | 0 | 1 | TR (1) | 0 | 0 | 1 | 0 |
| Dietrich 2014 | 0 | 1 | 0 | Div | 0 | 0 | 1 | 1 |
| Holster 2014 | 0 | 0 | 9 | OU (3), Div (1), Tumor (1), RP (1), TR (4) | 0 | 0 | 6 | 3 |
| Kratt Endoscopy 2014 | 0 | 0 | 2 | OU (2) | 0 | 0 | 2 | 0 |
| Masci 2014 | 0 | 13 | 0 | PU (13) | 0 | 0 | 7 | 6 |
| Selvapatt 2014 | 0 | 1 | 0 | Tumor (1) | 0 | 0 | 1 | 0 |
| Sulz 2014 | 0 | 14 | 1 | PU (3), OU (3), post-EBS (2), TR (1), PH (1), DL (1), Tumor (2), Other (2) | 0 | 0 | 2 | 13 |
| Tarantino 2014 | 0 | 1 | 0 | Other (1) | 0 | 0 | 0 | 1 |
| Yau 2014 | 0 | 19 | 0 | PU (13), AVM (1), TR (1), DL (1), Other (2) | 4 | 1 | 3 | 16 |
| Zimmer 2014 | 0 | 1 | 0 | PU (1) | 0 | 0 | 1 | 0 |
| Chen Endoscopy 2015 | 0 | 1 | 0 | Tumor (1) | 0 | 0 | 1 | 0 |
| Granata Int J Colorectal 2015 | 0 | 0 | 5 | OU (5) | 0 | 0 | 0 | 5 |
| Granata Endoscopy 2015 | 0 | 1 | 0 | Other (1) | 0 | 0 | 0 | 1 |
| Ivekovic 2015 | 0 | 0 | 1 | TR (1) | 0 | 0 | 0 | 1 |
| Giles 2016 | 0 | 36 | 0 | PU (25), TR (6), Tumor (2), DL (1), PH (1), Other (1) | 9 | 5 | 9 | 27 |
| Gubler 2016 | 0 | 1 | 4 | OU (5) | 0 | 0 | 6 | 1 |
| Haddara 2016 | 0 | 202 | 0 | PU (72), Tumor (61), TR (25), post-EBS (7), DL (3), PH (7), OU (1), Other (64) | 30 | 7 | 94 | 108 |
| Sinha 2016 | AP | 20 | 0 | PU (20) | 3 | 1 | 0 | 20 |
| Venezia 2016 | 0 | 1 | 0 | Tumor (1) | 0 | 0 | 0 | 1 |
| Xavier 2016 | 0 | 1 | 0 | Tumor (1) | 0 | 0 | 0 | 1 |
| Total | 4 | 481 | 46 | 56 (11%) | 17 (3.0%) | 232 | 234 |
Table 24. Summary of Clinical Literature on the use of Hemospray® (Continued)
Potential Adverse Events reviewed for are: perforation (P), aspiration pneumonia (AP), embolization, and bowel impaction.
Primary Therapy = Monotherapy or first-line as Hemospray followed by additional modalities.
Secondary Therapy = Adjunct therapy or Resoue therapy, including from previous treatment of same lesion during procedure.
| Lesion Identifier | Description |
|---|---|
| PU | Peptic Ulcer |
| Tumor | Tumor |
| TR | Tissue Resection (e.g. endoscopic mucosal resection, endoscopic submucosal dissection, polypectomy, and ampullary resection) |
| post-EBS | Endoscopic biliary sphincterotomy (aka ERCP) |
| DL | Dieulafoy's lesion |
| OU | other Ulcer (e.g. anastomotic ulcer, ulcerative colitis, post-polypectomy ulcer, post-EMR ulcer, post-Barrett's Ablation Ulcer, post-banding) |
| PH | portal hypertensive gastropathy |
| RP | radiation proctitis |
| Div | Diverticulum |
| AVM | arteriovenous malformation (includes angiodysplasia) |
| Other | Includes Esophagitis, Gastritis, Mallory-Weiss tear, Gastric antral vascular ectasia, pseudocyst, infection, ischemic colitis, unknown causes |
Wilson-Cook Medical, Inc. has sponsored 3 clinical studies and 1 survey of the use of Hemospray® to date, and has also conducted a literature search of relevant studies comprising the treatment of 522 patients with Hemospray®. The first and only premarket study was a pilot "firstin-human" feasibility study completed on 20 patients in Hong Kong, China who presented with upper gastrointestinal bleeding. The other two clinical studies and the survey were also
{21}------------------------------------------------
conducted outside the U.S. as post-market studies. Wilson-Cook Medical, Inc. initiated the postmarket Survey to Evaluate the Application of Hemospray® in the Luminal Tract (SEAL) to collect limited, non-protected health information and data on the Wilson-Cook Medical, Inc. Hemospray® Endoscopic Hemostat as part of a limited marketing release. The survey was conducted in Canada, Denmark, England, France, Germany, Italy and Holland and currently stores data on over 100 patients, primarily with upper gastrointestinal bleeding. The post-market studies include the Hemostasis of Active GI Luminal Tract Bleeding (HALT) study, which was a European and Canadian single-arm study recruiting only patients with upper gastrointestinal bleeding, and A Prospective Observational Cohort Study of Hemospray® for Lower Gastrointestinal Hemorrhage (APPROACH) study, a Canadian single-arm study evaluating the use of Hemospray® in lower gastrointestinal bleeding. Published clinical literature information was cited for use as historical control data for evaluation of each of these single-armed studies.
Acute (procedural) hemostasis was achieved in 19 of 20 patients (95%) in the pilot study. There were no treatment-related or procedure-related serious adverse events. The HALT study continues to enroll patients in Europe and Canada, and the APPROACH study reports on 50 patients, demonstrating 100% initial hemostasis with use of Hemospray® in this group of patients with diverse lower gastrointestinal etiologies of bleeding. This endoscopic hemostat is approved for use in upper gastrointestinal nonvariceal bleeding in Europe and both upper and lower gastrointestinal nonvariceal bleeding in Canada. The post-market experience reported in medical literature from Europe, Canada, and Asia all report favorable results with the use of Hemospray® in the treatment of nonvariceal gastrointestinal bleeding. 92% of the literature reports are on the treatment of upper gastrointestinal bleeding with Hemospray®, while 8% of the reports include only lower gastrointestinal bleeding cases.
Finally, the sponsor reports on 5 cases of Hemospray® application under humanitarian device exemption emergency use in critically ill patients who were very high-risk surgical patients. These were patients from one center in the United States. Hemospray® was used after all other modalities of endoscopic treatment failed. Hemospray® stopped the upper gastrointestinal bleeding in all these patients; none died of recurrent bleeding and only one died from progression of lymphoma during the time interval reported.
The clinical data are summarized in the following table:
{22}------------------------------------------------
| Study | N | Hemostasis onIndexEndoscopy (%) | Re-bleedRate (%) | 30-dayMortality(%) | BowelPerforation(%) | PowderImpaction(%) | ThromboembolicEvent (%) |
|---|---|---|---|---|---|---|---|
| Pilot Study | 20 | 95 | 10 | 0 | 0 | 0 | 0 |
| SEALRegistry | 89 | 100 | 19 | 5.6 | 3.4 | 0 | 0 |
| HALT Study | 64 | 97 | 20 | 3.2 | 3.1 | 0 | 0 |
| APPROACHStudy | 50 | 100 | 10 | 2 | 0 | 0 | 0 |
| LiteratureHemospray®Studies | 522 | 97.4 | 22 | 10.7 | 0.4 | 0 | 0 |
| EmergencyUse | 5 | 100 | 0 | 20 | 0 | 0 | 0 |
| Total | 750 | 97.8 | 20.2 | 11.6 | 0.9 | 0 | 0 |
Table 25: Summary of Hemospray® Clinical Data
The emergency pre-market use of Hemospray® was accomplished in the treatment of 5 patients, all with intractable upper gastrointestinal bleeding requiring multiple transfusions and very highrisk, precluding other interventions, including surgery. All these patients attained successful hemostasis with Hemospray ; one died from the progression of their underlying malignancy.
Holster I.L. et. al., published a study which reported on patients with upper gastrointestinal bleeding and antithrombotic therapy and demonstrated 63% initial hemostasis with a 38% rebleeding rate. These results are similar or better than those of traditional endoscopic modalities of treating gastrointestinal bleeding in this subgroup of patients.
The sponsor also reports on outside the United States Complaint Data: 280 total reports for over 50.000 units sold. Only 10% of these complaints required medical intervention. There were 4 deaths reported in the complaint data. Two deaths were related to patient co-morbidities and 2 may have been related to the device - one perforation, sepsis, and death, possibly from bowel over-inflation, and one death from exsanguination due to device malfunction. The latter resulted in change in device design to mitigate for the risk of malfunction. The table below summarizes the complaint data, stratified by Risk Severity Score (RSS) 3:
{23}------------------------------------------------
Table 26: Risk Severity Score
| RSS Value: | 2 | 3 | 4 | 5 | 6 | 7 | 10 |
|---|---|---|---|---|---|---|---|
| Number ofComplaintReports | 1 | 246 | 3 | 16 | 4 | 6 | 4 |
| Descriptionof Harm | Labeldamage | Lowimpact,loss of allor part ofdevicefunction;nuisanceto patientor enduser | Negligibleharm; harmnotrequiringmedicalintervention | Minorharm; harmrequiringmedicalintervention | Moderateharm; harmrequiringmedicalintervention | Significantharm; harmresulting inhospitalization,major | Criticalharm;death. |
a Risk Severity Score (RSS)
RSS 10: 4 deaths - 2 comorbidities, 1 perforation sepsis, 1 device malfunction resulting in exsanguination
RSS 7: 5 re-bleed, 1 perforation requiring surgery
RSS 6: less significant re-bleed. 1 requiring surgery
RSS 5: re-bleed, stricture, failed hemostasis
Conclusions:
- . Hemospray® has been used as primary treatment of non-variceal gastrointestinal bleeding, rescue therapy, adjunctive therapy, bridging therapy, and, in some cases, prophylactic therapy after polyp excision or mucosectomy. Both upper gastrointestinal and lower gastrointestinal non-variceal bleeding have been treated with Hemospray® outside the United States for the last seven years with few reports of device-related adverse events and possible device-related adverse events. In all these clinical applications of Hemospray® initial hemostasis was achieved in over 90% of patients.
- . The HALT and APPROACH studies were prospective, post-market, single-armed studies that demonstrated the effectiveness of Hemospray® in treating both upper and lower gastrointestinal bleeding.
- . Complaint data resulted in label warnings and device manufacturing changes to limit the risk of bowel perforation and device malfunction.
- In Patients on antithrombotic therapy (ATT), Hemospray® rivaled standard of care (SOC) . with 63% initial hemostasis and 38% re-bleed rate.
Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
{24}------------------------------------------------
LABELING
Labeling provided for the Hemospray® Endoscopic Hemostat includes Instructions for Use and packaging labels. The labeling satisfies the requirements of 21 CFR § 801.109 Prescription devices, and includes information regarding specifications, instructions for use, contraindications, warnings, and precautions, as well as a prescription statement.
Important components of the labeling include:
- Contraindications, warnings, cautions, or limitations needed for safe use of the device, . including
- Contraindications specific to the primary endoscopic procedure to be performed in gaining access to desired target site
- Specific contraindication for patients who have gastrointestinal fistulas, are suspected of having a gastrointestinal perforation, or are at high risk of gastrointestinal perforation during endoscopic treatment.
- A caution that patients with gastrointestinal bleeding that are on antithrombotic medication may be at an increased risk of re-bleeding
- A caution that use of more than (3) Hemospray® devices per patient may result in impaction in colon and is not recommended
- A caution that use of Hemospray® in the presence of bowel obstruction and/or an anastomosis may pose a risk of injury due to over-distention.
- A caution that Hemospray® may occlude ducts and orifices which communicate with the main bowel lumen. Use caution when using Hemospray® in the vicinity of these orifices.
- A warning that although not seen in clinical practice, there is a risk of aspiration of Hemospray® powder resulting in respiratory complications. It is prudent to restrict the use of Hemospray® to 5 cm below the upper esophageal sphincter (UES).
- A statement that use of an end-viewing endoscope is recommended
- Information on how the device operates and directions for use ●
- . A shelf life
- A prescription statement as required by 21 CFR 801.109
- Information regarding magnetic resonance compatibility.
{25}------------------------------------------------
RISKS TO HEALTH
Table 27 identifies the risks to health that may be associated with use of a hemostatic device for intra-luminal gastrointestinal internal use and the measures necessary to mitigate these risks.
| Identified Risk | Mitigation Measures |
|---|---|
| Bleeding- Inability to achieve hemostasis- Recurrence of bleeding | In vivo performance testingNon-clinical performance testingLabeling |
| Infection | Sterilization validationShelf life testingLabeling |
| Adverse tissue reaction | In vivo performance testingNon-clinical performance testingBiocompatibility evaluationLabeling |
| Obstruction of GI tract | In vivo performance testingLabeling |
| GI distension or perforation | In vivo performance testingLabeling |
| Vascular obstruction- Ischemia- Emboli formation | In vivo performance testingNon-clinical performance testingLabeling |
| Tissue trauma | In vivo performance testingNon-clinical performance testingLabeling |
| Improper device use | In vivo performance testingLabeling |
Table 27: Identified Risks to Health and Mitigation Measures
SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the hemostatic device for intraluminal gastrointestinal internal use is subject to the following special controls:
-
- The device must be demonstrated to be biocompatible.
-
- Performance data must support the sterility and pyrogenicity of the device.
-
- Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
-
- In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The testing must evaluate the following:
- a) The ability to deliver the hemostatic material to the bleeding site;
{26}------------------------------------------------
- b) The ability to achieve hemostasis in a clinically relevant model of gastrointestinal bleeding: and
- c) Safety endpoints, including thromboembolic events, local and systemic toxicity, tissue trauma, gastrointestinal tract obstruction, and bowel distension and perforation.
-
- Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
- a) Materials characterization of all components must demonstrate the device meets established specifications, which must include compositional identity and purity, characterization of impurities, physical characteristics, and reactivity with fluids,
- b) Performance testing must demonstrate the mechanical integrity and functionality of the system used to deliver the device and demonstrate the device meets established specifications, including output pressure for propellant-based systems.
-
- Labeling must include:
- a) Information identifying and explaining how to use the device and its components; and
- b) A shelf life.
BENEFIT/RISK DETERMINATION
The benefits and risks of the device are based on the data collected in clinical studies as well as on the outside U.S. post-market complaint data. These benefits and risks are summarized below:
Benefits
-
- Ease of application; no need for direct targeting, direct pressure or en face positioning of the endoscope to the bleeding site or touching of tissue. The device can be applied tangentially or at any angle with good effect. It may be less injurious to a bleeding fresh bowel anastomosis and can be applied in areas that are difficult to reach anatomically. Such areas include the proximal lesser curvature of the stomach and the back wall of the duodenum. Hemospray® can be effective where bleeding is originating at an ulcer base that is fibrotic and on which a clip cannot be easily applied.
-
- Rapid application with minimal need for anaesthesia that is easily repeatable.
-
- Can be applied to a wide area to cover diffuse or multifocal bleeding.
- Can be applied by gastroenterologists and general surgeons in smaller hospitals to control 4. initial bleeding as a bridge to more definitive treatment in tertiary care hospitals with greater clinical resources
{27}------------------------------------------------
-
- Can be used as a novel non-contact rescue therapy when other traditional therapies have failed in a patient who would otherwise require a high-risk operation.
-
- Can be used therapeutically as well as prophylactically to treat recently active bleeding lesions of the GI tract or after procedures with high risk of delayed bleeding, such as endoscopic submucosal dissection or endoscopic mucosal resection.
-
- Can be used in coagulopathic patients who are bleeding from a GI lesion where the coagulopathy cannot be easily reversed since Hemospray® powder, which is both cohesive and adhesive when exposed to blood, provides a mechanical barrier to bleeding.
Risks
-
- Literature-supported risks of Hemospray®: Re-bleeding 20%, bowel perforation and continued bleeding requiring surgical intervention 1%, 30-day mortality 20%.
-
- Hemospray® can cause bowel obstruction or obstruction of ducts and GI structures emptying into the main GI tract. Anatomic and physiologic gastrointestinal abnormalities can contribute to this problem.
-
- Over-inflation of bowel may result in perforation. Patients with gastrointestinal fistulas or compromised bowel may be at greatest risk. Each canister of Hemospray® could potentially add over 3 litres of additional volume to the GI tract comprised primarily of CO2 and to a lesser extent Hemospray® powder.
-
- Embolization of CO2 gas or Hemospray® to distant sites, resulting in adverse cardiopulmonary events or migration of (b) (4) | bentonite into nearby feeding vessels causing more diffuse or uncontrolled thrombosis and local ischemic events.
-
- Local toxicity to endothelial cells or macrophages causing thrombosis of blood vessels and affect healing.
-
- Allergic reaction, systemic toxicity from leached components of the device.
-
- The device is an ion exchange resin and a bulk laxative which may affect absorption of electrolytes, drugs, vitamins, and nutrients from the GI Tract.
-
- When used for proximal esophageal bleeding, the risk of patient aspiration of the (b) (4) bentonite may be present especially in an unprotected airway.
-
- Mechanical malfunction of the CO2 delivery mechanism resulting from obstruction of the endoscopic channel or catheter or CO2 canister malfunction may result in delayed treatment of uncontrolled gastrointestinal bleeding. The resulting risk of explosion with harm to patient or health personnel is unknown.
-
- Hemospray® cannot be easily applied through a side viewing duodenoscope and may require exchange to a forward viewing gastroscope. The resulting treatment delay may be detrimental to the patient.
{28}------------------------------------------------
Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
Benefit/Risk Conclusion
Traditional endoscopic treatment of gastrointestinal bleeding includes: injection of vasoconstrictors, directed by endoscopic ultrasound, thermos-probe or bipolar cautery, argon plasma coagulator (APC), Over the Scope Clip (OTSC), endo-suturing, bands, detachable snares, covered stents, radiation therapy, angiography with angio-embolization and surgery. Blood vessels feeding the gastrointestinal bleeding site can now be identified with endoscopic ultrasound and directly occluded with cyanoacrylic glues (off-label) and coils. These modalities are effective in greater than 80% of cases, with a re-bleed rate in up to 10% of cases. However, they require specific targeting of the bleeding lesion, specialized expertise and, in some cases, are invasive. All these modalities require touching of the tissue with potential collateral injury. This disadvantage may be most appreciated in the patient with a fresh gastrointestinal anastomosis who can least afford injury to that surgical site during an effort to achieve hemostasis. Many of these treatment modalities are time consuming and difficult to accomplish further compromising a hemodynamically unstable patient. Hemospray® offers a less technically demanding, more rapid means of achieving hemostasis. The risks of bowel obstruction, allergic reaction, and embolization were not realized in all the clinical evidence submitted. There were episodes of device malfunction that resulted in delayed treatment as well as bowel overdistension, which resulted in pain and in some cases bowel perforation. The perforation rates in all cases did not exceed that of traditional endoscopic treatment of gastrointestinal bleeding. The early and late re-bleeding rates, especially for malignant and large friable lesions, were significantly less than that noted with traditional treatment. The lesions that are more likely to rebleed or continue bleeding may still benefit from the application of Hemospray® to stop the bleeding acutely, allowing for patient stabilization and bridging to more definitive therapy.
Given the available information summarized above, the data support the conclusion that Hemospray® is an endoscopic hemostat with a very low risk profile compared to traditional endoscopic hemostatic procedures, with a greater than 95% immediate hemostasis in most cases of upper and lower non-variceal gastrointestinal bleeding. While Hemospray® has a similar early and late re-bleeding rate when compared to many traditional techniques of endoscopic hemostasis, it has been shown to decrease the need for salvage surgical intervention and angiographic embolization. Because direct targeting and specialized skill are not necessary with the use of Hemospray®, it is especially useful in controlling gastrointestinal bleeding in difficultto-reach anatomic locations and for controlling diffuse bleeding in large lesions, such as tumors, and large ulcers. The most significant advantage of Hemospray® is its ability to convert a bleeding, hemodynamically unstable patient to one in which bleeding is controlled, the patient resuscitated and bridged to a more elective definitive treatment of the target bleeding site. The literature and studies provided on Hemospray® use in Europe. Asia, and Canada report no episodes of embolization of the device components, major thrombosis outside the target bleeding site, bowel obstruction, or allergic reaction. There have been reports of transient episodes of pain and rare perforation events, both attributed to over-distension of bowel with application of Hemospray®. This problem may be mitigated with labeling and instruction to partially de
{29}------------------------------------------------
sufflate the bowel prior to device application. Additionally, if CO2 is used as the insufflating gas rather than air, the more rapid absorption of CO2 may mitigate the risk of persistent bowel distension.
The device provides benefits, and the risks can be mitigated using general controls and the identified special controls.
CONCLUSION
The De Novo request for the Hemospray® Endoscopic Hemostat is granted and the device is classified under the following:
Product Code: QAU Device Type: Hemostatic device for intraluminal gastrointestinal use Class: II Regulation: 21 CFR 878.4456
{30}------------------------------------------------
REFERENCES
Alzoubaidi D, Magee C, Gulati S, Haji A, Coron E, Morris AJ, Bergman J, Weusten BL, Goetz M, Kiesslich R, Lovat L. Mo1093 Outcomes From an International Multicentre Registry of Patients With Gastrointestinal Bleeding Undergoing Endoscopic Treatment With Hemospray. Gastrointestinal Endoscopy. 2017 May 1;85(5):AB424.
Arena M, Masci E, Eusebi LH, Iabichino G, Mangiavillano B, Viaggi P, Morandi E, Fanti L, Granata A, Traina M, Testoni PA. Hemospray for treatment of acute bleeding due to upper gastrointestinal tumours. Digestive and Liver Disease. 2017 May 1;49(5):514-7.
Balachandar S, Eaton JK. Turbulent dispersed multiphase flow. Annual Review of Fluid Mechanics. 2010;42:111-33.
Barkun AN, Adam V, Lu Y, Chen YI, Martel M. Using Hemospray Improves the Costeffectiveness Ratio in the Management of Upper Gastrointestinal Nonvariceal Bleeding. Journal of clinical gastroenterology. 2018 Jan 1:52(1):36-44.
Beye B, Barret M, Alatawi A, Beuvon F, Nicco C, Pratico CA, Chereau C, Chaussade S, Batteux F, Prat F. Topical hemostatic powder promotes reepithelialization and reduces scar formation after extensive esophageal mucosal resection. Diseases of the Esophagus. 2016 Sep 1:29(6):520-7.
Carrier B, Wang L, Vandamme M, Pelleng RJ, Bornert M, Tanguy A, Van Damme H. ESEM study of the humidity-induced swelling of clay film. Langmuir. 2013 Oct 2;29(41):12823-33.
Champion JA, Katare YK, Mitragotri S. Particle shape: a new design parameter for micro-and nanoscale drug delivery carriers. Journal of Controlled Release. 2007 Aug 16;121(1-2):3-9.
Chen YI, Barkun A, Nolan S. Hemostatic powder TC-325 in the management of upper and lower gastrointestinal bleeding: a two-year experience at a single institution. Endoscopy. 2015 Feb;47(02):167-71.
Eisenhour DD, Brown RK. Bentonite and its impact on modern life. Elements. 2009 Apr 1;5(2):83-8.
Faita F, Cori L, Bianchi F, Andreassi MG. Arsenic-induced genotoxicity and genetic susceptibility to arsenic-related pathologies. International journal of environmental research and public health. 2013 Apr 12;10(4):1527-46.
Friedrich F, Schild D, Weidler PG, Schäfer T. The Clay Minerals Society Workshop Lectures Series. 2016:21
Gerlach T. Grayson JK. Pichakron KO. Sena MJ. DeMartini SD. Clark BZ. Estep JS. Zierold D. Preliminary study of the effects of smectite granules (WoundStat) on vascular repair and wound
{31}------------------------------------------------
healing in a swine survival model. Journal of Trauma and Acute Care Surgery. 2010 Nov 1:69(5):1203-9.
Giday SA, Kim Y, Krishnamurty DM, Ducharme R, Liang DB, Shin EJ, Dray X, Hutcheon D, Moskowitz K. Donatelli G. Rueben D. Long-term randomized controlled trial of a novel nanopowder hemostatic agent (TC-325) for control of severe arterial upper gastrointestinal bleeding in a porcine model. Endoscopy. 2011 Apr:43(04):296-9.
Grim RE, Guven N. Bentonites: geology, mineralogy, properties and uses. Elsevier; 2011 Sep 22.
Gubler C, Metzler JM, Turina M. Hemospray treatment for bleeding intestinal anastomoses in the early postoperative period: a novel non-operative approach. Techniques in coloproctology. 2016 Jul 1;20(7):495-8.
Haddara S, Jacques J, Lecleire S, Branche J, Leblanc S, Le Baleur Y, Privat J, Heyries L, Bichard P. Granval P. Chaput U. A novel hemostatic powder for upper gastrointestinal bleeding: a multicenter study (the "GRAPHE" registry). Endoscopy. 2016 Dec;48(12):1084-95.
Hagel AF, Albrecht H, Nägel A, Vitali F, Vetter M, Dauth C, Neurath MF, Raithel M. The application of hemospray in gastrointestinal bleeding during emergency endoscopy. Gastroenterology research and practice. 2017:2017.
Harper MS, Carpenter C, Klocke DJ, Carlson G, Davis T, Delaney B. E. coli lipopolysaccharide: acute oral toxicity study in mice. Food and chemical toxicology. 2011 Aug 1;49(8):1770-2.
Holster IL, Kuipers EJ, Tiwa ET. Hemospray in the treatment of upper gastrointestinal hemorrhage in patients on antithrombotic therapy. Endoscopy. 2013 Jan;45(01):63-6.
Ibrahim M. El-Mikkawy A. Abdalla H. Mostafa I. Devière J. Management of acute variceal bleeding using hemostatic powder. United European gastroenterology journal. 2015 Jun;3(3):277-83.
Koedrith P, Seo YR. Advances in carcinogenic metal toxicity and potential molecular markers. International journal of molecular sciences. 2011 Dec 20;12(12):9576-95.
Leblanc S. Vienne A, Dhooge M, Coriat R, Chaussade S, Prat F. Early experience with a novel hemostatic powder used to treat upper GI bleeding related to malignancies or after therapeutic interventions (with videos). Gastrointestinal endoscopy. 2013 Jul 1:78(1):169-75.
Li Y, Li H, Xiao L, Zhou L, Shentu J, Zhang X, Fan J. Hemostatic efficiency and wound healing properties of natural zeolite granules in a lethal rabbit model of complex groin injury. Materials. 2012 Dec 3;5(12):2586-96.
Moore DM, Reynolds RC. X-ray Diffraction and the Identification and Analysis of Clay Minerals. New York: Oxford university press: 1989 Oct.
{32}------------------------------------------------
Moosavi S, Chen YI, Barkun AN. TC-325 application leading to transient obstruction of a postsphincterotomy biliary orifice. Endoscopy. 2013 Dec;45(S 02):E130.
Morgan DJ, Eslinger EV, Eberl DD, Karlinger MR. Chemistry of illite/smectite and end-member illite. Clays and Clay Minerals. 1986;34(4):368-78.
Morris AJ, Smith LA, Stanley A, Bergman JJ, Kiesslich R, Hoffman A, Tjwa ET, Kuipers EJ, Von Holstein CS, Oberg S, Brullet E. 284 Hemospray for non-variceal upper gastrointestinal bleeding: results of the seal dataset (survey to evaluate the application of Hemospray in the luminal tract). Gastrointestinal Endoscopy. 2012 Apr 1;75(4):AB133-4.
Smith LA, et. al, Hemospray for Non-Variceal Upper Gastrointestinal Bleeding: Results of the SEAL Dataset (Survey to Evaluate the Application of Hemospray in the Luminal Tract), GUT, Vol 61 Suppl 2, July 2012, A61.
Sung JJ, Luo D, Wu JC, Ching JY, Chan FK, Lau JY, Mack S, Ducharme R, Okolo P, Canto M, Kalloo A. Early clinical experience of the safety and effectiveness of Hemospray in achieving hemostasis in patients with acute peptic ulcer bleeding. Endoscopy. 2011 Apr:43(04):291-5.
Talibudeen O. Goulding KW. Charge heterogeneity in smectites. Clays and Clay Minerals. 1983 Feb 1:31(1):37-42.
Ufer K, Stanjek H, Roth G, Dohrmann R, Kleeberg R, Kaufhold S. Quantitative phase analysis of bentonites by the Rietveld method. Clays and Clay Minerals. 2008 Apr 1;56(2):272-82.
United States Pharmacopeia 40. National Formulary 35 (USP), Monograph: Bentonite (2017)
Yau AH, Ou G, Galorport C, Amar J, Bressler B, Donnellan F, Ko HH, Lam E, Enns RA. Safety and efficacy of Hemospray® in upper gastrointestinal bleeding. Canadian Journal of Gastroenterology and Hepatology. 2014;28(2):72-6.
§ 878.4456 Hemostatic device for intraluminal gastrointestinal use.
(a)
Identification. A hemostatic device for intraluminal gastrointestinal use is a prescription device that is endoscopically applied to the upper and/or lower gastrointestinal tract and is intended to produce hemostasis via absorption of fluid or by other physical means.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Performance data must support the sterility and pyrogenicity of the device.
(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(4) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The testing must evaluate the following:
(i) The ability to deliver the hemostatic material to the bleeding site;
(ii) The ability to achieve hemostasis in a clinically relevant model of gastrointestinal bleeding; and
(iii) Safety endpoints, including thromboembolic events, local and systemic toxicity, tissue trauma, gastrointestinal tract obstruction, and bowel distension and perforation.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
(i) Materials characterization of all components must demonstrate the device meets established specifications, which must include compositional identity and purity, characterization of impurities, physical characteristics, and reactivity with fluids.
(ii) Performance testing must demonstrate the mechanical integrity and functionality of the system used to deliver the device and demonstrate the device meets established specifications, including output pressure for propellant-based systems.
(6) Labeling must include:
(i) Information identifying and explaining how to use the device and its components; and
(ii) A shelf life.