K Number
DEN170010
Date Cleared
2018-04-20

(423 days)

Product Code
Regulation Number
862.3590
Type
Direct
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of carisoprodol metabolite (meprobamate) in human urine at a cutoff value of 100 ng/mL when calibrated against meprobamate. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for verification by a confirmatory method such as GC/MS, LC/MS or permitting laboratories to establish quality control procedures.

The assay provides only a preliminary analytical result. A more specific alternative chemical confirmatory method (e.g., gas or liquid chromatography and mass spectrometry) must be used to obtain a confirmed analytical result. Clinical consideration and professional judgment must be exercised with any drug of abuse test, particularly when the preliminary test result is positive.

Device Description

The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is a kit comprised of two reagents (antibody/substrate reagent R1 and enzyme-drug conjugate reagent R2), which are bottled separately but sold together within the kit. The Ri solution contains mouse monoclonal anti-meprobamate antibody, glucose-6- phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains meprobamate-labeled glucose-6-phosphate dehydrogenase (G6PDH) in buffer with sodium azide (0.09 %) as a preservative.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay, based on the provided document:

Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from study results and regulatory requirements)Reported Device Performance
Accuracy (Qualitative Mode)High agreement with confirmatory method (LC/MS or GC/MS) for positive and negative samples, particularly around the cutoff.Qualitative Analysis Agreement: - Positive Agreement: 98.5% - Negative Agreement: 96.6% *2 false positives in "Near Cutoff Negative" (92 and 98 ng/mL, both identified as Negative by GC/MS/LC/MS) *1 false negative in "Near Cutoff Positive" (103 ng/mL, identified as Positive by GC/MS/LC/MS)
Accuracy (Semi-Quantitative Mode)High agreement with confirmatory method (LC/MS or GC/MS) for positive and negative samples, particularly around the cutoff.Semi-Quantitative Analysis Agreement: - Positive Agreement: 98.5% - Negative Agreement: 98.3% *1 false positive in "Near Cutoff Negative" (98 ng/mL, identified as Negative by GC/MS/LC/MS) *1 false negative in "Near Cutoff Positive" (103 ng/mL, identified as Positive by GC/MS/LC/MS)
Reproducibility/Precision (Qualitative Mode)Consistent results at specific concentrations, especially near the cutoff.At cutoff (100 ng/mL, GC/MS confirmed 94.9 ng/mL): 40 Negative / 48 Positive out of 88 determinations. At -25% Cutoff (75 ng/mL, GC/MS confirmed 76.8 ng/mL): 88 Negative / 0 Positive. At +25% Cutoff (125 ng/mL, GC/MS confirmed 122.3 ng/mL): 0 Negative / 88 Positive.
Reproducibility/Precision (Semi-Qualitative Mode)Consistent results at specific concentrations, especially near the cutoff.At cutoff (100 ng/mL, GC/MS confirmed 94.9 ng/mL): 60 Negative / 28 Positive out of 88 determinations. At -25% Cutoff (75 ng/mL, GC/MS confirmed 76.8 ng/mL): 88 Negative / 0 Positive. At +25% Cutoff (125 ng/mL, GC/MS confirmed 122.3 ng/mL): 0 Negative / 88 Positive.
Linearity/Reportable RangeAcceptable recovery across a range of concentrations.Mean Recovery: 66.6% to 109.2% for expected concentrations from 10 ng/mL to 400 ng/mL. (Note: 10 ng/mL showed lower recovery, but higher concentrations were generally good).
Analytical Specificity (Endogenous Compounds)No interference from common endogenous compounds at specified concentrations.No positive or negative interference observed for the listed endogenous compounds when meprobamate was at -25% or +25% of the cutoff.
Analytical Specificity (Urine Preservatives)No significant interference from common urine preservatives.Sodium azide and sodium fluoride: No interference. Boric acid: Caused false negative results at +25% cutoff (125 ng/mL) and up to +125% cutoff (225 ng/mL). Labeling mitigation required.
Analytical Specificity (pH)No interference across a physiological pH range.No positive or negative interference observed at urine pH values ranging from 3 to 11.
Analytical Specificity (Specific Gravity)No interference across a physiological specific gravity range.No positive or negative interference observed at urine specific gravities ranging from 1.002 to 1.029.
Cross-Reactivity (Structurally Related Compounds)Acceptable cross-reactivity profile.Carisoprodol (90.9%), Felbamate (25.0%), Meprobamate-N-Glucuronide (0.5%), Hydroxymeprobamate (0.2%). Other listed compounds showed <0.1% cross-reactivity.
Interference (Structurally Unrelated Compounds)No interference from common structurally unrelated drugs.No positive or negative interference observed from a wide range of common drugs/substances (e.g., Acetaminophen, Codeine, Ibuprofen) at high concentrations (e.g., 100,000 ng/mL) when meprobamate was at -25% or +25% of the cutoff.
Clinical Validity of CutoffMeprobamate levels measurable for a relevant post-dose period.Pharmacokinetic studies of carisoprodol and meprobamate in plasma, and meprobamate elimination in urine, indicated levels above 100 ng/mL are present for "up to a few days" post-single dose.

Study Details

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set:

    • Precision Study: 88 determinations per concentration level for
      each of 8 concentrations (a total of 704 determinations for qualitative and 704 for semi-quantitative).
    • Method Comparison Study: 127 clinical urine samples.
    • Linearity Study: Each sample run in 10 replicates across 11 concentrations.
    • Analytical Specificity (Endogenous, Preservatives, pH, Specific Gravity, Unrelated Compounds): Samples tested in duplicate.
    • Cross-reactivity (Structurally Related Compounds): Not explicitly stated, but typically involves multiple runs.
  • Data Provenance: Retrospective clinical urine samples "from subjects that were (b) (4) taking prescribed or nonprescribed Carisoprodol." The origin country is not explicitly stated, but given this is an FDA submission, it's likely primarily US-based or from regions with comparable medical practices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • The ground truth for the test set was established by Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS). These are established analytical methods for drug confirmation. There is no mention of "experts" in the sense of medical professionals (e.g., radiologists) establishing ground truth, as this is an in vitro diagnostic device. The "experts" in this context would be the technicians/scientists operating and interpreting the GC/MS/LC/MS results, who are implied to be qualified in their field based on the use of these validated methods.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • There was no human adjudication process described. The reference method (GC/MS or LC/MS) provided the definitive "ground truth" for each sample.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, this was not a multi-reader multi-case (MRMC) comparative effectiveness study. This device is an automated enzyme immunoassay for chemical analysis, not an imaging-based AI system that would typically involve human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, this was a standalone (algorithm only) performance study. The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is an automated system designed to provide a preliminary analytical result without human interpretation of the assay itself. Human involvement comes in confirming positive results with alternative chemical methods and clinical judgment, as stated in the intended use.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The ground truth used was confirmatory analytical methods, specifically Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS). These methods are considered the gold standard for drug quantification and identification in urine.

8. The sample size for the training set

  • The document does not explicitly describe a separate "training set" for the device. As an enzyme immunoassay, the "training" (calibration) involves using known calibrator concentrations (0, 50, 100, 200, and 400 ng/mL meprobamate, as mentioned in the linearity section) to establish the standard curve against which unknown samples are measured. There isn't an "algorithm" in the conventional AI sense that undergoes training with a large dataset. The performance characteristics described are related to the analytical validation of the assay itself.

9. How the ground truth for the training set was established

  • As there isn't a traditional "training set" in the AI sense, the concept of ground truth for it doesn't directly apply. The "ground truth" for the calibrators used in the assay system would be their known, precisely manufactured concentrations of meprobamate, traceable to a commercially available meprobamate source material with 99% analytical purity (as mentioned in section L.1.c).

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay

DECISION SUMMARY

A. DEN Number:

DEN170010

B. Purpose for Submission:

De Novo request for evaluation of automatic class III designation for the LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay

C. Measurand:

Meprobamate

D. Type of Test:

Homogenous Enzyme Immunoassay, Qualitative and Semi-quantitative

E. Applicant:

Lin-Zhi International, Inc.

F. Proprietary and Established Names:

LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay

G. Regulatory Information:

    1. Regulation: 21 CFR 862.3590
    1. Classification: Class II
    1. Product code: QBK
    1. Panel: Toxicology (91)

H. Intended Use:

    1. Indications for use:
      The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is intended for

{1}------------------------------------------------

the qualitative and semi-quantitative determination of carisoprodol metabolite (meprobamate) in human urine at a cutoff value of 100 ng/mL when calibrated against meprobamate. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for verification by a confirmatory method such as GC/MS, LC/MS or permitting laboratories to establish quality control procedures.

The assay provides only a preliminary analytical result. A more specific alternative chemical confirmatory method (e.g., gas or liquid chromatography and mass spectrometry) must be used to obtain a confirmed analytical result. Clinical consideration and professional judgment must be exercised with any drug of abuse test, particularly when the preliminary test result is positive.

    1. Special conditions for use statement(s):
    • For in vitro diagnostic use only.
    • For prescription use only. ●
    1. Special instrument requirements:

The assay was validated on the Beckman AU400e automated clinical chemistry analyzer. Clinical chemistry analyzers capable of maintaining a constant temperature, pipetting samples, mixing reagents, measuring enzyme rates at 340 nm and timing the reaction accurately can be used to perform this assay.

I. Device Description:

The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is a kit comprised of two reagents (antibody/substrate reagent R1 and enzyme-drug conjugate reagent R2), which are bottled separately but sold together within the kit. The Ri solution contains mouse monoclonal anti-meprobamate antibody, glucose-6- phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains meprobamate-labeled glucose-6-phosphate dehydrogenase (G6PDH) in buffer with sodium azide (0.09 %) as a preservative.

J. Standard/Guidance Documents Referenced:

CLSI EP07-A2: Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition.

{2}------------------------------------------------

K. Test Principle:

The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is based on competition between meprobamate in the sample and the enzyme glucose-6-phosphate dehydrogenase (G6PDH) labeled with meprobamate for a fixed amount of antibody in the reagent. G6PDH enzyme activity decreases upon binding to the antibody, and thus meprobamate concentration in the sample is proportional to enzyme activity. In the absence of meprobamate in the sample, meprobamate-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. When meprobamate is present in the sample competes with drug-labeled G6PDH for binding to antibody; the unbound meprobamatelabeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.

L. Performance Characteristics (if/when applicable):

The following performance characteristics were obtained on the Beckman AU400e automated clinical chemistry analyzer.

1. Analytical performance:

a. Reproducibility/Precision

A precision study was performed over 22 days, with 2 runs per day in duplicate (n = 88) using pooled negative urine samples spiked with meprobamate to concentrations of 25, 50, 75, 100, 125, 150, 175, and 200 ng/mL. All concentrations for precision studies were confirmed by Gas Chromatography/Mass Spectrometry (GC/MS) testing. The results from samples assaved using the qualitative and semi-quantitative modes are summarized below.

Targetmeprobamateconcentration(ng/mL)GC/MSmeprobamateconcentration(ng/mL)% ofCutoff# ofDeterminationsResult
0notdetermined-100%8888 Neg / 0Pos
2524.7-75%8888 Neg / 0Pos
5051.7-50%8888 Neg / 0Pos
7576.8-25%8888 Neg / 0Pos
10094.9Cutoff8840 Neg / 48Pos

Qualitative Analysis

{3}------------------------------------------------

Targetmeprobamateconcentration(ng/mL)GC/MSmeprobamateconcentration(ng/mL)% ofCutoff# ofDeterminationsResult
125122.3+25%880 Neg / 88Pos
150149.4+50%880 Neg / 88Pos
ા તર176.8+75%880 Neg / 88Pos
200211.0+100%880 Neg / 88Pos
Semi-Quantitative Analysis
Sampleconcentration(ng/mL)GC/MSmeprobamateconcentration(ng/mL)% of Cutoff# ofDeterminationsResult
0notdetermined-100%8888 Neg / 0Pos
2524.7-75%8888 Neg / 0Pos
5051.7-50%8888 Neg / 0Pos
7576.8-25%8888 Neg / 0Pos
10094.9Cutoff8860 Neg / 28Pos
125122.3+25%880 Neg / 88Pos
150149.4+50%880 Neg / 88Pos
175176.8+75%880 Neg / 88Pos
200211.0+100%880 Neg / 88Pos

b. Linearity/assay reportable range:

A recovery study was performed by spiking a pool of negative urine with a high dose of meprobamate and generating serial dilutions to achieve the following concentrations: 10, 40, 80, 120, 160, 200, 240, 280, 320, 360, and 400 ng/mL meprobamate. Each sample was run in replicates of 10 in semi-quantitative mode with a calibration curve established with five meprobamate calibrators (0, 50, 100, 200, and 400 ng/mL). Percent recovery was calculated using the mean concentration of the 10 replicates relative to the expected concentration as shown below.

{4}------------------------------------------------

ExpectedConcentration(ng/mL)Mean ObservedConcentration(ng/mL)Mean Recovery(%)Range ofRecovery (%)
02.2N/AN/A
106.766.638.0 - 87.0
4039.999.793.8 - 104.8
8081.8102.298.0 - 108.4
120123.3102.899.9 - 107.3
160163.7102.398.3 - 107.5
200195.397.795.4 - 99.7
240251.6104.895.3 - 110.5
280305.7109.2102.0 - 114.3
320348.9109.0105.5 - 111.9
360386.6107.4102.3 - 111.8
400412.5103.1100.3 - 105.4

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is traceable to a commercially available meprobamate source material having 99% analytical purity as determined by LC/MS.

Specimen stability information was provided to support the following specimen storage conditions: storage at 2-8℃ for up to 1 week and at -20℃ for up to 17 months.

  • d. Detection limit
    Not applicable.

  • e. Analytical specificity:
    Endogenous Compounds: Potential interference from endogenous compounds was evaluated in the qualitative and semi-quantitative modes by spiking these compounds (at the test concentrations shown in the table below) into urine containing either 75 ng/mL or 125 ng/mL meprobamate (±25% of the assay cutoff). Samples were tested in duplicate. The results, which were the same for the qualitative and semiquantitative modes, are shown below.

{5}------------------------------------------------

EndogenousConcentration-25% Cutoff+25% Cutoff
CompoundTested (mg/dL)(75 ng/mL)(125 ng/mL)
Acetone1000NegativePositive
Ascorbic acid1500NegativePositive
Beta-hydroxybutyricacid sodium salt100NegativePositive
Bilirubin2NegativePositive
Calcium chloridedihydrate300 (saturatedsolution)NegativePositive
Citric acid800NegativePositive
Creatinine500NegativePositive
Ethanol1000NegativePositive
Galactose10NegativePositive
γ-Globulin500NegativePositive
Glucose3000NegativePositive
Hemoglobin300NegativePositive
Human SerumAlbumin500NegativePositive
Oxalic Acid100NegativePositive
Potassiumchloride6000NegativePositive
Riboflavin0.3NegativePositive
Urea6000NegativePositive
Uric acidmonosodium salt10NegativePositive
Sodium Chloride6000NegativePositive
Sodium phosphatedibasic salt300NegativePositive

Urine sample preservatives: To evaluate potential interference from common urine sample preservatives, device performance in qualitative and semi-quantitative modes was tested by spiking sodium azide (1% w/v), sodium fluoride (1% w/v), or boric acid (1% w/v) into negative urine to ± 25% of the 100 ng/mL meprobamate cutoff (75 ng/mL and 125 ng/mL). Samples were tested in duplicate. Boric acid was found to cause false negative results at +25% of the 100 ng/mL cutoff (125 ng/mL) and up to +125% of the 100 ng/mL cutoff (225 ng/mL) in both the qualitative and semiquantitative modes. The following statement is provided in the limitations section of the labeling: Boric Acid at 1% w/v may cause false negative results. Boric Acid is not recommended as a preservative for urine.

pH: To evaluate potential interference from the pH of urine, device performance in the qualitative and semi-quantitative modes was tested using a range of urine pH values (3, 4, 5, 6, 7, 8, 9, 10, and 11). Test samples were prepared in negative urine, which was spiked to ±25% of the 100 ng/mL meprobamate cutoff (75 ng/mL and 125 ng/mL), and tested in duplicate. No positive or negative interference was observed at

{6}------------------------------------------------

urine pH values ranging from 3 to 11 for the qualitative and semi-quantitative modes.

Specific Gravity: To evaluate potential interference from the specific gravity of urine. device performance in the qualitative and semi-quantitative modes was tested using a range of urine specific gravities (1.002, 1.003, 1.006, 1.007, 1.008, 1.012, 1.014, 1.015, 1.019, 1.025, and 1.029). These 12 negative urine samples were spiked with meprobamate to ±25% of the 100 ng/mL meprobamate cutoff (75 ng/mL and 125 ng/mL), and tested in duplicate. No positive or negative interference was observed at urine specific gravities ranging from 1.002 to 1.029 for the qualitative and semi-quantitative modes.

Cross-reactivity of structurally related compounds: The cross-reactivity of various structurally related drugs was evaluated in qualitative and semi-quantitative modes by spiking each substance into negative urine. The following table shows the quantity of each potential cross-reacting compound that produced assay reactivity equivalent to the 100 ng/mL meprobamate cutoff, or the maximum concentration tested (concentration tested), and calculated cross-reactivity for each compound (% crossreactivity). The results were the same for the qualitative and semi-quantitative modes.

CompoundConcentration Tested (ng/mL)% Cross-Reactivity
Carisoprodol11090.9
Darunavir200,000<0.1
Efavirenz200,000<0.1
Felbamate40025.0
Hydroxymeprobamate65,0000.2
Meprobamate100100.0
Meprobamate-N-Glucuronide20,0000.5
Methocarbamol200,000<0.1
Mitocycin C200,000<0.1
Neostigmine bromide200,000<0.1
Retigabine200,000<0.1
Ritonavir100,000<0.1
Rivastigmine tartrate200,000<0.1
Zafirlukast200,000<0.1

Interference by structurally unrelated compounds: The potential for positive or negative interference by various structurally unrelated compounds was evaluated by spiking these compounds (at the test concentrations shown in the table below) into urine containing either 75 ng/mL or 125 ng/mL meprobamate (±25% of the assay cutoff). Samples were tested in duplicate. The results, which are the same for the qualitative and semi-quantitative modes, are shown below.

{7}------------------------------------------------

-25%+25%
CutoffCutoff
Concentration(75(125
CompoundTested (ng/mL)ng/mL)ng/mL)
Acetaminophen100,000NegativePositive
6-Acetylmorphine10,000NegativePositive
Acetylsalicylic Acid100,000NegativePositive
Albuterol(Salbutamol)100,000NegativePositive
Amitriptyline100,000NegativePositive
d-Amphetamine100,000NegativePositive
Benzoylecgonine100,000NegativePositive
Buprenorphine15,000NegativePositive
Bupropion100,000NegativePositive
Caffeine100,000NegativePositive
Carbamazepine100,000NegativePositive
Cetirizine20,000NegativePositive
Chlorpheniramine100,000NegativePositive
Chlorpromazine100,000NegativePositive
Clomipramine100,000NegativePositive
Codeine100,000NegativePositive
Cyclobenzaprine100,000NegativePositive
Desipramine100,000NegativePositive
Diphenhydramine100,000NegativePositive
Ephedrine100,000NegativePositive
Fentanyl10,000NegativePositive
Fluoxetine100,000NegativePositive
Fluphenazine100,000NegativePositive
Hydrocodone100,000NegativePositive
Hydromorphone100,000NegativePositive
Ibuprofen100,000NegativePositive
Imipramine100,000NegativePositive
Lidocaine100,000NegativePositive
Loratadine100,000NegativePositive
Maprotiline30,000NegativePositive
MDA (3,4-methylenedioxyamphetamine)100,000NegativePositive
MDEA100,000NegativePositive
MDMA (3,4-methylenedioxymethamphetamine)100,000NegativePositive
Meperidine100,000NegativePositive
Methadone100,000NegativePositive
d-Methamphetamine100,000NegativePositive
Methapyrilene100,000NegativePositive
Methaqualone100,000NegativePositive
-25%Cutoff(75ng/mL)+25%Cutoff(125ng/mL)
CompoundConcentrationTested (ng/mL)
Metronidazole100,000NegativePositive
Morphine100,000NegativePositive
Nicotine100,000NegativePositive
Nortriptyline100,000NegativePositive
Oxazepam100,000NegativePositive
Oxycodone100,000NegativePositive
Oxymorphone100,000NegativePositive
PCP (phencyclidine)10,000NegativePositive
Pentazocine20,000NegativePositive
Phenobarbital100,000NegativePositive
d-Propoxyphene100,000NegativePositive
Propranaolol100,000NegativePositive
Ranitidine100,000NegativePositive
Sertraline100,000NegativePositive
THC-COOH (11-Nor-Delta-9-THC-9-carboxylic acid)1,000NegativePositive
Thioridazine100,000NegativePositive
Tramadol100,000NegativePositive
Valproic Acid100,000NegativePositive

{8}------------------------------------------------

f. Assay cut-off:

Characterization of how the device performs analytically around the claimed cutoff concentrations of 100 ng/mL is described in the precision section, M.1.a. above.

2. Comparison studies:

a. Method comparison:

A method comparison study was performed using 127 unaltered clinical urine samples from subjects that were (b) (4) taking prescribed or nonprescribed Carisoprodol. Each sample was run in singlicate using the LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay on the AU400e automated clinical analyzer and the result was compared to that obtained by LC/MS or GC/MS. Samples with meprobamate concentrations < 100 ng/mL by LC/MS or GC/MS were defined as "negative." Samples with meprobamate concentrations > 100 ng/mL by LC/MS or GC/MS were defined as "positive." The results are summarized below.

{9}------------------------------------------------

Qualitative analysis

CandidatedeviceresultsNegative(Drug-free orless than 50%of the cutoffconcentration)Near CutoffNegative(Between50% belowthe cutoff andthe cutoffconcentration)Near CutoffPositive(Between thecutoff and50% abovethe cutoffconcentration)High Positive(Greater than50% abovethe cutoffconcentration)%Agreement
Positive02*115698.5
Negative43141**096.6
Sample#LC/MS or GC/MSMeprobamate (ng/mL)LC/MS orGC/MS resultCandidate deviceresult
58*92NegativePositive
59*98NegativePositive
60**103PositiveNegative

Semi-Quantitative analysis

CandidatedeviceresultsNegative(Drug-free orless than 50%of the cutoffconcentration)Near CutoffNegative(Between50% belowthe cutoff andthe cutoffconcentration)Near CutoffPositive(Between thecutoff and50% abovethe cutoffconcentration)High Positive(Greater than50% abovethe cutoffconcentration)%Agreement
Positive01*115698.5
Negative43151**098.3
Sample#LC/MS Meprobamate(ng/mL)LC/MSresultCandidate deviceresult
59*98NegativePositive
60**103PositiveNegative
  • b. Matrix comparison:
    Not applicable. Urine is the only claimed matrix for the candidate device.

    1. Clinical studies:
    • a. Clinical Sensitivity: Not applicable.
    • b. Clinical Specificity: Not applicable.

{10}------------------------------------------------

  • c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.
    1. Clinical cut-off:

The assay has a clinical cut-off of 100 ng/mL meprobamate in human urine. To support the clinical validity of this cut-off, the sponsor submitted information from pharmacokinetic studies of carisoprodol and meprobamate in plasma following single dose administration and detailed information about meprobamate elimination in urine as a first-order elimination. This information indicates that meprobamate levels above the cutoff of this assay (100 ng/mL) are present in the urine for up to a few days following a single dose of drug. The information provided supported the clinical validity of the claimed cutoff of this device.

    1. Expected values/Reference range:
      Not applicable.

M. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Not applicable.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and the special controls for this device type.

O. Patient Perspectives:

This submission did not include specific information on patient perspectives for this device. In general, patients benefit from reliable tests to screen for the presence of drugs of abuse.

P. Identified Risks to Health and Required Mitigations

Identified Risks to HealthRequired Mitigations
Clinical action based on incorrect test results(false positive results, false negative results)may lead to inappropriate clinical decisionmaking.Special controls (1), (2), and (3)
Incorrect understanding of the device,including the results, may lead toinappropriate clinical decision making.Special controls (2) and (3)

{11}------------------------------------------------

Q. Benefit/Risk Analysis

Summary
Summary ofthe Benefit(s)There are benefits related to the detection of the presence of the carisoprodol metabolite ormeprobamate drug in human urine. While the test is not indicated for acute clinicalmanagement of patients it may have value in settings of drug treatment programs,including decreasing the risk of potentially dangerous medication interactions. Forexample, identifying the presence of meprobamate may help prevent drug interactions withother sedating drugs that could potentiate the sedating effect. The test also has benefit foruse in substance abuse programs to monitor the illicit use of carisoprodol or meprobamatewhich have abuse potential and are categorized as Schedule IV controlled substances bythe Drug Enforcement Administration (DEA).
Summary ofthe Risk(s)The risk of the test is limited to false negative or false positive test results and the incorrectunderstanding of how to interpret the results.
A false positive result could result in the misidentification of a patient as having takencarisoprodol or meprobamate; however, device labelling states that positive results must beconfirmed with a more specific analytical testing method; and when performed thisconfirmatory testing mitigates the risk of false positive results.
A false negative result could increase the risk of drug interactions if other drugs whichcontribute to or potentiate the side effects are prescribed. However, the analyticalperformance of the test appears mitigates the risk of false negative results. Negativeagreement among the 127 samples tested was 96.6% and the one false negative result inthe study showed a level of meprobamate (established by a confirmatory analyticalmethod) that was very near the cut-off.
Misinterpretation of a positive test as indicative of toxicity, when the test is used in anacute care setting, could delay appropriate treatment if the result leads to confusion in thediagnosis of another acute condition that is present. This risk is mitigated by labelling ofthe device which specifically limits against the use of the test in being used for diagnosingdrug intoxication or for the purposes of determining appropriate therapy.
Summary ofOtherFactorsThe studies conducted included assessment of the accuracy, specificity, and precisionaround the cutoff of the device.
ConclusionsDo the probable benefits outweigh the probable risks?
Given the device's indications for use, required general controls and special controls established for this device,the probable benefits outweigh the probable risks.

{12}------------------------------------------------

R. Conclusion:

The information provided in this de novo submission is sufficient to classify this device into class II under regulation 21 CFR 862.3590. FDA believes that the stated special controls, and applicable general controls, including design controls, provide reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:

Product Code:QBK
Device Type:Meprobamate test system
Class:II (special controls)
Regulation:21 CFR 862.3590

(a) Identification. A meprobamate test system is a device intended to measure meprobamate in human specimens. Measurements obtained by this device are used to detect the presence of meprobamate to diagnose the use or overdose of meprobamate or structurally-related drug compounds (e.g., prodrugs).

(b) Classification. Class II (special controls). The special controls for this device are:

    1. Design verification and validation must include:
  • Robust data demonstrating the accuracy of the device when used in the intended (i) specimen matrix. The accuracy data must include a comparison between the meprobamate test system results and meprobamate results that are measured on an FDA-accepted measurement method that is specific and accurate (e.g., gas or liquid chromatography combined with tandem mass spectrometry).
  • (ii) Robust analytical data demonstrating the performance characteristics of the device, including, but not limited to, specificity, cross-reactivity to relevant endogenous and exogenous substances, and the reproducibility of analyte detection around the cutoff(s).
    1. The intended use of the device must not include an indication for use in monitoring therapeutic drug concentrations or informing dosing adjustment decisions.
    1. Your 21 CFR 809.10 labeling must include the following:
  • (i) If indicated for use as a screening test to identify preliminary results for further confirmation, the intended use must state "This assay provides only a preliminary analytical result. A more specific alternative chemical confirmatory method (e.g., gas or liquid chromatography and mass spectrometry) must be used to obtain a confirmed analytical result. Clinical consideration and professional judgment must be exercised with any drug of abuse test, particularly when the preliminary test result is positive."
  • (ii) A limiting statement that reads as follows: "This test should not be used to monitor therapeutic drug concentrations or to inform dosing adjustment decisions."

§ 862.3590 Meprobamate test system.

(a)
Identification. A meprobamate test system is a device intended to measure meprobamate in human specimens. Measurements obtained by this device are used to detect the presence of meprobamate to diagnose the use or overdose of meprobamate or structurally-related drug compounds (e.g., prodrugs).(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) Robust data demonstrating the accuracy of the device when used in the intended specimen matrix. The accuracy data must include a comparison between the meprobamate test system results and meprobamate results that are measured on an FDA-accepted measurement method that is specific and accurate (e.g., gas or liquid chromatography combined with tandem mass spectrometry).
(ii) Robust analytical data demonstrating the performance characteristics of the device, including, but not limited to, specificity, cross-reactivity to relevant endogenous and exogenous substances, and the reproducibility of analyte detection around the cutoff(s).
(2) The intended use of the device must not include an indication for use in monitoring therapeutic drug concentrations or informing dosing adjustment decisions.
(3) Your 21 CFR 809.10 labeling must include the following:
(i) If indicated for use as a screening test to identify preliminary results for further confirmation, the intended use must state “This assay provides only a preliminary analytical result. A more specific alternative chemical confirmatory method (e.g., gas or liquid chromatography and mass spectrometry) must be used to obtain a confirmed analytical result. Clinical consideration and professional judgment must be exercised with any drug of abuse test, particularly when the preliminary test result is positive.”
(ii) A limiting statement that reads as follows: “This test should not be used to monitor therapeutic drug concentrations or to inform dosing adjustment decisions.”