K Number
DEN140001
Device Name
ZINC TRANSPORTER 8 ANTIBODY (ZNT8AB) ELISA ASSAY KIT
Date Cleared
2014-08-20

(65 days)

Product Code
Regulation Number
866.5670
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay is for the semi-quantitative determination of autoantibodies to Zinc Transporter 8 (ZnT8) in human serum. The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay may be useful as an aid in the diagnosis of Type 1 diabetes mellitus (autoimmune mediated diabetes). The ZnT8Ab assay is not to be used alone and is to be used in conjunction with other clinical and laboratory findings.
Device Description
The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay contains the following: strip wells coated with ZnT8 (96 wells in total) and supplied as 12 strips of 8 wells in a frame and sealed in a foil bag with desiccant; ready-to-use 4 levels of calibrators (10, 20, 75, and 500 U/mL, 5x0.7 mL each); one each ready-to-use positive I, positive II and negative control serum (1x0.7 mL); Zinc T8 Biotin (lyophilized) 3x5.5 mL (reconstituted); ready-to-use reconstitution buffer for Zinc T8 Biotin 2x15 mL (colored red); Streptavidin Peroxidase (SA-POD dilute before use) 1x0.7 mL; ready-to-use diluent for SA-POD 1x15 mL; ready-to-use peroxidase substrate (TMB) 1x15 mL; concentrated wash solution (dilute with deionized water before use) 1x125 mL and ready-to-use stop solution 1x12 mL.
More Information

Not Found

Not Found

No
The summary describes a standard ELISA assay kit for detecting autoantibodies and does not mention any AI or ML components in its description, intended use, or performance evaluation.

No
The device is an in vitro diagnostic assay used to aid in the diagnosis of Type 1 diabetes, not to treat or prevent it.

Yes

The device aids in the diagnosis of Type 1 diabetes mellitus by determining autoantibodies in human serum, which is a diagnostic purpose.

No

The device description clearly lists multiple physical components, including strip wells, calibrators, control serum, and various reagents, indicating it is a laboratory assay kit, not a software-only device.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use explicitly states it is for the "semi-quantitative determination of autoantibodies to Zinc Transporter 8 (ZnT8) in human serum." This involves testing a sample taken from the human body (serum) in vitro (outside the body) to provide information about a medical condition (aid in the diagnosis of Type 1 diabetes mellitus).
  • Device Description: The description details a kit containing reagents and components designed for performing a laboratory test (ELISA assay) on a biological sample.
  • Intended User / Care Setting: It specifies "laboratory professionals in a clinical laboratory setting," which is the typical environment for performing IVD tests.

All these elements align with the definition of an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay is for the semi-quantitative determination of autoantibodies to Zinc Transporter 8 (ZnT8) in human serum. The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay may be useful as an aid in the diagnosis of Type 1 diabetes mellitus (autoimmune mediated diabetes). The ZnT8Ab assay is not to be used alone and is to be used in conjunction with other clinical and laboratory findings.

Product codes

PHF

Device Description

The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay contains the following: strip wells coated with ZnT8 (96 wells in total) and supplied as 12 strips of 8 wells in a frame and sealed in a foil bag with desiccant; ready-to-use 4 levels of calibrators (10, 20, 75, and 500 U/mL, 5x0.7 mL each); one each ready-to-use positive I, positive II and negative control serum (1x0.7 mL); Zinc T8 Biotin (lyophilized) 3x5.5 mL (reconstituted); ready-to-use reconstitution buffer for Zinc T8 Biotin 2x15 mL (colored red); Streptavidin Peroxidase (SA-POD dilute before use) 1x0.7 mL; ready-to-use diluent for SA-POD 1x15 mL; ready-to-use peroxidase substrate (TMB) 1x15 mL; concentrated wash solution (dilute with deionized water before use) 1x125 mL and ready-to-use stop solution 1x12 mL.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

The clinical study included subjects 11-20, 21-25, 26-37, and 38+ years old.

Intended User / Care Setting

For prescription use only in accordance with 21 CFR 801.109. The device is for use by laboratory professionals in a clinical laboratory setting.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

A clinical performance study evaluated 569 clinically defined patient samples: 323 patients with Type 1 diabetes mellitus ("T1D" or "T1 DM") and 246 patients with other diseases (60 Type 2 diabetes mellitus and 186 other diseases including Graves' Disease, Myasthenia Gravis, Rheumatoid Arthritis, Addison's Disease, Neuromyelitis Optica, Hashimoto's Disease, Celiac Disease, Systemic Lupus Erythematosus, Testicular Cancer, Metabolic syndrome, Kidney Disease, and Urinary Tract Infection).

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Precision/Reproducibility:

  • Intra-assay: Tested nine serum specimens (ranges from 6.9-471.5 U/mL) with 20-25 replicates each. %CVs ranged from 1.7-5.6%.
  • Inter-assay: Tested eleven samples (ranges from 5.8-465.1 U/mL) over 10-20 days. %CVs ranged from 2.9-17.7%.
  • Lot-to-lot reproducibility: Ten kit lots tested over 96 weeks using two control samples (48.4 and 156.3 U/mL). %CV for lot-to-lot reproducibility ranged from 7.6-9.2%.
  • Lab to Lab reproducibility: Thirty-five samples (0 to 1299 U/mL) assayed at two different laboratories. Correlation R = 0.984 (y = 0.9199x +0.3183).

Linearity/assay reportable range:

  • Three positive samples diluted (10 dilutions from neat to 1:500). Linear for concentrations of 7.0 to 485 U/mL.
  • No hook effect observed up to 1980 U/mL.

Detection limit:

  • LoB: 2.1 U/mL (determined by testing negative control 60 times).
  • LoD: 5.6 U/mL (determined by testing 100 healthy blood donor samples).

Analytical specificity:

  • Endogenous Interference: No significant interference observed in 10 specimens spiked with hemoglobin (500 mg/dL), bilirubin (20 mg/dL), and Intralipids (1000 and 3000 mg/dL).
  • Cross-reactivity: Tested with 246 sera from other autoimmune diseases and conditions. 241/246 specimens were negative. Five positive sera: 1 Graves' Disease (15.6 U/mL), 2 Addison Disease (48.1 U/mL and 23.5 U/mL), 1 Type 2 Diabetes (>500 U/mL), and 1 Kidney Disease (65 U/mL).

Assay cut-off:

  • Determined by testing 397 US healthy blood donors. 394/397 (99%) were negative for ZnT8Ab. The cut-off is >= 15 U/mL for positive.

Clinical studies:

  • Clinical Sensitivity and Specificity: Evaluated on 569 clinically defined patient samples (323 T1D, 60 T2D, and 186 other diseases).
    • Sensitivity: 68 % (220/323) (95% C.I.: 63-73%)
    • Specificity: 98 % (241/246) (95% C.I.: 95-99%)
    • Overall Agreement: 81% (461/569) (95% C.I.: 76-85 %)
    • Positivity rate in T1D population decreased with age:
      • 11-20 years: 77% (166/215)
      • 21-25 years: 67% (18/27)
      • 26-37 years: 45% (33/73)
      • 38+ years: 38% (3/8)

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Sensitivity: 68 % (220/323) (95% C.I.: 63-73%)
Specificity: 98 % (241/246) (95% C.I.: 95-99%)
Overall Agreement: 81% (461/569) (95% C.I.: 76-85 %)

Predicate Device(s)

Not Found

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 866.5670 Zinc transporter 8 autoantibody immunological test system.

(a)
Identification. A zinc transporter 8 autoantibody immunological test system is a device that consists of reagents used to measure, by immunochemical techniques, the autoantibodies in human serum samples that react with Zinc Transporter 8 (ZnT8). The measurements aid in the diagnosis of Type 1 diabetes mellitus (autoimmune mediated diabetes) in conjunction with other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) A detailed description of the device that includes:
(A) A detailed description of all components in the test system, including a description of the assay components in the kit and all required ancillary reagents;
(B) A detailed description of instrumentation and equipment, and illustrations or photographs of non-standard equipment or methods if applicable;
(C) Detailed documentation of the device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software where applicable;
(D) A detailed description of appropriate internal and external quality controls that are recommended or provided. The description must identify those control elements that are incorporated into the recommended testing procedures;
(E) Detailed specifications for sample collection, processing, and storage;
(F) A detailed description of methodology and assay procedure; and
(G) Detailed specification of the criteria for test results interpretation and reporting.
(ii) Information that demonstrates the performance characteristics of the device, including:
(A) Device precision/reproducibility data generated from within-run, between-run, between-day, between-lot, between-operator, between-instruments, between-site, and total precision for multiple nonconsecutive days as applicable. A well characterized panel of patient samples or pools from the intended use population that covers the device measuring range must be used;
(B) Device linearity data generated from patient samples covering the assay measuring range if applicable;
(C) Information on traceability to a reference material and description of value assignment of calibrators and controls if applicable;
(D) Device analytical sensitivity data, including limit of blank, limit of detection and limit of quantitation if applicable;
(E) Device analytical specificity data, including interference by endogenous and exogenous substances, as well as cross-reactivity with samples derived from patients with other autoimmune diseases or conditions;
(F) Device instrument carryover data when applicable;
(G) Device stability data including real-time stability under various storage times and temperatures;
(H) Specimen stability data, including stability under various storage times, temperatures, freeze-thaw, and transport conditions where appropriate;
(I) Method comparison data generated by comparison of the results obtained with the device to those obtained with a legally marketed predicate device with similar indication of use. Patient samples from the intended use population covering the device measuring range must be used;
(J) Specimen matrix comparison data if more than one specimen type or anticoagulant can be tested with the device. Samples used for comparison must be from patient samples covering the device measuring range;
(K) A description of how the assay cut-off (the medical decision point between positive and negative) was established and validated as well as supporting data;
(L) Clinical performance must be established by comparing data generated by testing samples from the intended use population and the differential diagnosis groups with the device to the clinical diagnostic standard. The diagnosis of Type 1 diabetes mellitus must be based on clinical history, physical examination, and laboratory tests, such as one or more pancreatic or insulin autoantibody test. Because the intended use population for Type 1 diabetes mellitus includes subjects less than 18 years old, samples from representative numbers of these subjects must be included. Representative numbers of samples from all age strata must also be included. The differential diagnosis groups must include, but not be limited to the following: Type 2 diabetes mellitus; metabolic syndrome; latent autoimmune diabetes in adults; other autoimmune diseases such as celiac disease (without a concomitant diagnosis of Type 1 diabetes mellitus), systemic lupus erythematosus, rheumatoid arthritis, and Hashimoto's thyroiditis; infection; renal disease; and testicular cancer. Diseases for the differential groups must be based on established diagnostic criteria and clinical evaluation. For all samples, the diagnostic clinical criteria and the demographic information must be collected and provided. The clinical validation results must demonstrate clinical sensitivity and clinical specificity for the test values based on the presence or absence of Type 1 diabetes mellitus. The data must be summarized in tabular format comparing the interpretation of results to the disease status; and
(M) Expected/reference values generated by testing an adequate number of samples from apparently healthy normal individuals.
(iii) Identification of risk mitigation elements used by the device, including description of all additional procedures, methods, and practices incorporated into the directions for use that mitigate risks associated with testing.
(2) Your 21 CFR 809.10(a) compliant label and 21 CFR 809.10(b) compliant labeling must include warnings relevant to the assay including:
(i) A warning statement that reads, “The device is for use by laboratory professionals in a clinical laboratory setting”;
(ii) A warning statement that reads, “The test is not a stand-alone test but an adjunct to other clinical information. A diagnosis of Type 1 diabetes mellitus should not be made on a single test result. The clinical symptoms, results on physical examination, and laboratory tests (
e.g., serological tests), when appropriate, should always be taken into account when considering the diagnosis of Type 1 diabetes mellitus and Type 2 diabetes mellitus”;(iii) A warning statement that reads, “Absence of Zinc T8 autoantibody does not rule out a diagnosis of Type 1 diabetes mellitus”; and
(iv) A warning statement that reads, “The assay has not been demonstrated to be effective for monitoring the stage of disease or its response to treatment.”
(3) Your 21 CFR 809.10(b) compliant labeling must include a description of the protocol and performance studies performed in accordance with paragraph (b)(1)(ii) of this section and a summary of the results.

0

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR KRONUS Zinc Transporter 8 Autoantibody ELISA Assay DECISION SUMMARY

A. DEN:

DEN140001

B. Purpose for Submission:

De Novo request for evaluation of automatic class III designation for the KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay

C. Measurand:

Zinc Transporter 8 Autoantibody (ZnT8Ab)

D. Type of Test:

The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay is for the semiquantitative determination of autoantibodies to Zinc Transporter 8 (ZnT8) in human serum. The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay may be useful as an aid in the diagnosis of Type 1 diabetes mellitus (autoimmune mediated diabetes).

E. Applicant:

KRONUS Market Development Associates, INC.

F. Proprietary and Established Names:

KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay

G. Regulatory Information:

    1. Regulation section:
      21 CFR § 866.5670
    1. Classification:
      Class II
    1. Product code:
      PHF

1

4. Panel:

82-Immunology

H. Intended Use:

    1. Intended use(s):
      The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay is for the semi-quantitative determination of autoantibodies to Zinc Transporter 8 (ZnT8) in human serum. The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay may be useful as an aid in the diagnosis of Type 1 diabetes mellitus (autoimmune mediated diabetes). The ZnT8Ab assay is not to be used alone and is to be used in conjunction with other clinical and laboratory findings.

2. Indication(s) for use:

Same as Intended use.

    1. Special conditions for use statement(s):
      For prescription use only in accordance with 21 CFR 801.109.
    1. Special instrument requirements:
      ELISA Plate Reader suitable for 96-well format and capable of measuring at 450 nm, and ELISA plate shaker capable of 500 shakes/ minute.

I. Device Description:

The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay contains the following: strip wells coated with ZnT8 (96 wells in total) and supplied as 12 strips of 8 wells in a frame and sealed in a foil bag with desiccant; ready-to-use 4 levels of calibrators (10, 20, 75, and 500 U/mL, 5x0.7 mL each); one each ready-to-use positive I, positive II and negative control serum (1x0.7 mL); Zinc T8 Biotin (lyophilized) 3x5.5 mL (reconstituted); ready-to-use reconstitution buffer for Zinc T8 Biotin 2x15 mL (colored red); Streptavidin Peroxidase (SA-POD dilute before use) 1x0.7 mL; ready-to-use diluent for SA-POD 1x15 mL; ready-to-use peroxidase substrate (TMB) 1x15 mL; concentrated wash solution (dilute with deionized water before use) 1x125 mL and ready-to-use stop solution 1x12 mL.

J. Standard/Guidance Document Referenced (if applicable):

Not applicable

K. Test Principle:

2

The KRONUS Zinc Transporter 8 Autoantibody (ZnT8Ab) ELISA Assay depends on the ability of ZnT8 autoantibodies to act divalently and form a bridge between ZnT8 coated on the ELISA plate wells and liquid phase ZnT8- biotin. The ZnT8-biotin bound is then quantitated by addition of streptavidin peroxidase and a colorgenic substrate (TMB) with reading of the final absorbances at 450 nm. The absorbance of each vial is directly proportional to the amount of antibody present. Calibrator values are plotted on linear graph paper and the antibody concentrations of the controls and patient specimens are interpolated from the curve.

L. Performance Characteristics (if/when applicable):

1. Analytical performance:

  • Precision/Reproducibility: a.

Intra-assay

The intra-assay precision was determined by testing nine serum specimens which included one close to the assay cut-off (15.4 U/mL); three low (6.9, 7.8, and 10.2 U/mL); one moderate (25.9 U/mL), and four high (66.3, 222.2, 441.5 and 471.5 U/mL). Twenty five (20-25) replicates of each sample were assayed in a single assay on a single day. Results showed %CVs ranged from 1.7-5.6% (see table below).

Sample123456789
n202020252520202020
Mean (U/mL)6.97.810.215.425.966.3222.2441.5471.4
SD (U/mL)0.20.40.40.51.21.711.87.613.5
% CV3.35.64.23.34.52.55.31.72.9

Inter-assay

The inter-assay precision was determined by testing eleven samples: five serum specimens once a day for twenty days, two serum specimens once a day for 18 days and four sera for once a day for 10 days. The serum specimens consisted of two specimens with anti-ZnT8Ab concentration of 16.1 and 16.3 U/mL close to the assay 15 U/mL cut-off; four low levels (5.8, 7.2, 10.9 and 14.2 U/mL); one moderate level (28.7 U/mL) and three high levels (73.7, 250.3. 446.5 and 465.1 U/mL). The %CVs ranged from 2.9-17.7% (see table below).

Sample1234567891011
n1818101020101020202020
Mean (U/mL)5.87.210.916.128.773.7250.314.216.3446.5465.1
SD (U/mL)0.90.90.81.62.93.444.20.50.514.716.9
% CV16.012.37.69.810.24.717.73.52.93.33.6

Lot-to-lot reproducibility:

3

Ten kit lots were tested using two control samples (48.4 and 156.3 U/mL) over a period of 96 weeks. The %CV for lot-to-lot reproducibility ranged from 7.6-9.2%.

Lab to Lab reproducibility:

Thirty-five samples (ranging from 0 to 1299 U/mL) and kit controls were assayed at two different laboratories. The correlation between laboratories was R = 0.984 (y = 0.9199x +0.3183).

  • b. Linearity/assay reportable range:
    Three positive samples were diluted (10 dilutions ranged from neat to 1:500). The observed values were graphed against the calculated values and a linear regression was performed. Results are summarized below:

| Sample | Dilution range
(U/mL) | Slope
(95%CI) | Y - Intercept
(95% CI) | R2 |
|--------|--------------------------|------------------|---------------------------|--------|
| A | 75.4 - 485.3 | 0.9539 | -59.55 | 0.9499 |
| B | 7.0 - 57.8 | 0.9556 | -0.5562 | 0.9783 |

The assay is linear for concentrations of 7.0 to 485 U/mL. The Package Insert Calculation Section states: "For samples that result in values greater than the 500 U/mL (greater than the highest calibrator) KRONUS recommends reporting the value as "greater than 500 U/mL"

High dose hook effect:

No hook effect was observed for ZnT8Ab concentration up to estimated 1980 U/mL.

  • Traceability, Stability, Expected values (controls, calibrators, or methods): C.
    There is no recognized reference standard available. The ZnT8 calibrator and controls (positive and negative) are prepared in-house and arbitrary units are assigned during the development process.

Stability

A real-time stability study was performed on three lots of ZnT8 Ab ELISA kit using 2 positive controls, a negative control and 11 samples (5 positive and 6 negative samples). Data support a shelf life of 9 months.

  • d. Detection limit:
    The limit of blank (LoB) was determined by sequentially testing the negative control included with the kit sixty (60) times. A calibration curve of absorbance450mm VS. concentration was constructed. The mean and SD for the absorbance450mm of the negative control were calculated and the mean + 2 SDs was read off the calibration curve to give a U/mL value. The LoB was computed to be 2.1 U/mL.

4

The limit of detection (LoD) was determined by sequentially testing twenty five (25) replicates each of four healthy blood donor samples for a total of 100 samples. The LoD determination used a non-parametric statistical calculation. The 95th percentile corresponds to the 95.5 ordered observation of results placed in ascending order which yields a LoD estimate of 5.6 U/mL.

  • e. Analytical specificity:

Endogenous Interference

No significant interference was observed in 10 specimens (with ZnT8Ab levels ranging from 2.1 - 356 U/mL) spiked with hemoglobin (at 500 mg/dL), bilirubin (at 20 mg/dL) and Intralipids (at 1000 and 3000 mg/dL). The Package Insert 'Specimen Collection and Handling Section' states: "Sera should be clear and grossly lipemic or hemolyzed samples should not be used".

Cross-reactivity:

The KRONUS ZnT8Ab ELISA Assay Kit was tested with 246 sera from other autoimmune diseases and conditions. These samples comprised of 24 Graves' Disease, 9 Myasthenia gravis, 26 Rheumatoid arthritis, 23 Addison Disease, 3 Neuromyelitis optica, 24 Hashimoto's Disease; 60 Type 2 Diabetes mellitus; 10 Celiac Disease; 9 Systemic Lupus Erythematosus; 6 Testicular Cancer; 37 Metabolic syndrome; 10 Kidney Disease and 5 Urinary Tract Infection.

Two hundred forty one (241) of 246 specimens were negative with the KRONUS ZnT8Ab ELISA Assay Kit. The five positive sera were as follows: 1 Graves' Disease at 15.6 U/mL; 2 Addison Disease at 48.1 U/mL and 23.5 U/mL; 1 Type 2 Diabetes at >500 U/mL and 1 Kidney Disease at 65 U/mL.

| Patient Group | No. of Samples Positive for
ZnT8Ab | % |
|------------------------------|---------------------------------------|----|
| Graves' Disease | 1/24 | 4 |
| Hashimoto's Thyroiditis | 0/24 | 0 |
| Addison's Disease | 2/23 | 9 |
| Myasthenia Gravis | 0/9 | 0 |
| Neuromyelitis Optica | 0/3 | 0 |
| Type 2 Diabetes | 1/60 | 2 |
| Rheumatoid Arthritis | 0/26 | 0 |
| Celiac Disease | 0/10 | 0 |
| Systemic Lupus Erythematosus | 0/9 | 0 |
| Testicular Cancer | 0/6 | 0 |
| Metabolic Syndrome | 0/37 | 0 |
| Kidney Disease | 1/10 | 10 |
| Urinary Tract Infection | 0/5 | 0 |

5

f. Assay cut-off:

The assay cut-off (greater than or equal to 15 U/mL is positive) for the KRONUS ZnT8Ab Assay Kit was determined by testing specimens from 397 US healthy blood donors (291 Black and 6 Caucasian males as well as 98 Black, 1 Hispanic and 1 Asian females). Ninety-nine percent (394/397) were negative for ZnT8Ab. The ZnT8Ab concentrations in the three positive samples were 19, 41 and 45 U/mL. The mean result was 1.90 U/mL with a SD of 3.84.

2. Comparison studies:

  • a. Method comparison with predicate device:
    Refer to Clinical studies.

  • b. Matrix comparison:
    Not applicable

    1. Clinical studies:
    • Clinical Sensitivity and Specificity: a.

The clinical performance was evaluated on 569 clinically defined patient samples. These samples had the following diagnosis: 323 patients with Type 1 diabetes mellitus ("T1D" or "T1 DM"), 60 Type 2 diabetes mellitus ("T2 DM") and 186 other diseases (24 Graves' Disease, 9 Myasthenia Gravis, 26 Rheumatoid Arthritis, 23 Addison's Disease, 3 Neuromyelitis Optica, 24 Hashimoto's Disease; 10 Celiac Disease; 9 Systemic Lupus Erythematosus; 6 Testicular Cancer; 37 Metabolic syndrome; 10 Kidney Disease and 5 Urinary Tract Infection). The clinical sensitivity, specificity and overall agreement are summarized below:

Diagnosis
Positive
(TID)Negative
(Non-Target Diseases)Totals
ZnT8Ab
ELISA
Assay KitPositive2205225
Negative103241344
Total323246569

Sensitivity: 68 % (220/323) (95% C.I.: 63-73%) 98 % (241/246) (95% C.I.: 95-99%) Specificity: Overall Agreement: 81% (461/569) (95% C.I.: 76-85 %)

The rate of positivity was found to decrease with increasing age for the T1D population. Positivity rates based on age strata in the T1D population is shown below.

6

Agen# PositivePositive (%)Negative (%)
11-202151667723
21-2527186733
26-3773334555
38+833862
Total3232206832
  • c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.
    1. Clinical cut-off:

Refer to assay cut-off.

    1. Expected values/Reference range:
      The expected value from 394 of the total 397 healthy individual blood donors was 15 U/mL values).

M. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 801, 21 CFR Part 809, 21 CFR 801.109, and the special controls.

N. Identified Risks and Required Mitigations:

Identified RisksRequired Mitigations
Inaccurate test results that provide false positive or false negative results can lead to improper patient management.Special controls (1), (2), and (3)
Failure to correctly interpret test results can lead to false positive or false negative resultsSpecial controls (1) (iii), (2), and (3)

7

O. Benefit/Risk Analysis:

Summary
Summary of
the Benefit(s)Patients with Type 1 diabetes mellitus (“T1D” or “T1DM”) often have antibodies to
pancreatic islet cell antigens. This assay represents the first to target antibodies to the
zinc transporter, expressed in pancreatic islet cells. The assay can help in the diagnosis
of T1D patients. In the clinical study, the assay demonstrated a clinical sensitivity of
68% and specificity of 98%. The study identified 11 of 323 T1D patients (3.4%) who
were positive only for ZnT8Ab, and negative for other existing pancreatic
autoantibodies, such as insulin antibodies (IAA), insulinoma-associated protein 2
autoantibodies (IA-2A), and glutamic acid decarboxylase autoantibodies (GADA).
This highlights the clinical utility of adding ZnT8 to the existing islet cell
autoantibodies to increase the capture of patients who potentially have T1D.
Summary of
the Risk(s)The risks are related to the consequences of decisions made based on false negative
and false positive results due to inaccurate test results, or failure to correctly
interpret test results. The combination of general and special controls would
mitigate these risks. A false negative result could lead to a missed or delayed
diagnosis of T1D, and a false positive result could lead to additional, unnecessary
evaluation and testing. The false negative risk of 32% found in the clinical study is
sufficiently mitigated by language in the Intended Use statement in the package
insert (labeling) that the assay is not to be used alone and is to be used in
conjunction with other clinical and laboratory findings. The false positive risk of
2% is extremely low.
Conclusions
Do the
probable
benefits
outweigh the
probable risks?The probable benefit of aiding in the increased detection of T1D patients
outweighs the probable risk of a false negative assay result in 32% of patients with
T1D and a false positive result in 2% of patients without T1D. The combination of
general controls and special controls, including product labeling (package insert),
adequately mitigates the risks posed by use of the assay.

The addition of ZnT8Ab to the existing islet cell autoantibodies increases the
capture of patients who potentially have T1D and outweighs risks of inaccurate test
results, which are adequately described and sufficiently mitigated by statements in
the Intended Use and Limitations sections of the package insert (product labeling). |

P. Conclusion:

The information provided in this de novo submission is sufficient to classify this device into class II under regulation 21 CFR 866.5670 with special controls. FDA believes that special controls, along with the applicable general controls, provide reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:

Product Code: PHF Zinc Transporter 8 Autoantibody immunological test system Device Type:

8

Class: II (special controls) Regulation: 21 CFR 866.5670

i.

Identification. A Zinc Transporter 8 Autoantibody immunological test system is a device that consists of reagents used to measure, by immunochemical techniques, the autoantibodies in human serum samples that react with Zinc Transporter 8 (ZnT8). The measurements aid in the diagnosis of Type 1 diabetes mellitus (autoimmune mediated diabetes) in conjunction with other clinical and laboratory findings.

  • Classification. Class II (special controls). Zinc Transporter 8 Autoantibody immunological (a) test system must comply with the following special controls:
      1. Premarket notification submissions must include the following information:
      • A detailed description of the device that includes:
        • A ) A detailed description of all components in the test system, including a description of the assay components in the kit and all required ancillary reagents.
        • B) A detailed description of instrumentation and equipment, and illustrations or photographs of non-standard equipment or methods if applicable.
        • C) Detailed documentation of the device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software where applicable.
        • D) A detailed description of appropriate internal and external quality controls that are recommended or provided. The description must identify those control elements that are incorporated into the recommended testing procedures.
        • E) Detailed specifications for sample collection, processing and storage.
        • F) A detailed description of methodology and assay procedure.
        • G) Detailed specification of the criteria for test results interpretation and reporting.
      • Information that demonstrates the performance characteristics of the ii. device, including:
        • A) Device precision/reproducibility data generated from within-run, between-run, between-day, between-lot, between-operator, betweeninstruments, between-site, and total precision for multiple nonconsecutive days as applicable. A well characterized panel of patient samples or pools from the intended use population that covers the device measuring range must be used.
        • B) Device linearity data generated from patient samples covering the assay measuring range if applicable.
        • C') Information on traceability to a reference material and description of value assignment of calibrators and controls if applicable.
        • D) Device analytical sensitivity data, including limit of blank, limit of detection and limit of quantitation if applicable.

9

  • E) Device analytical specificity data, including interference by endogenous and exogenous substances, as well as cross-reactivity with samples derived from patients with other autoimmune diseases or conditions.
  • F) Device instrument carryover data when applicable.
  • G) Device stability data including real-time stability under various storage times and temperatures.
  • H) Specimen stability data, including stability under various storage times, temperatures, freeze-thaw and transport conditions where appropriate.
  • I) Method comparison data generated by comparison of the results obtained with the device to those obtained with a legally marketed predicate device with similar indication of use. Patient samples from the intended use population covering the device measuring range must be used.
  • J) Specimen matrix comparison data if more than one specimen type or anticoagulant can be tested with the device. Samples used for comparison must be from patient samples covering the device measuring range.
  • K ) A description of how the assay cut-off (the medical decision point between positive and negative) was established and validated as well as supporting data.
  • L) Clinical performance must be established by comparing data generated by testing samples from the intended use population and the differential diagnosis groups with the device to the clinical diagnostic standard. The diagnosis of Type 1 diabetes mellitus must be based on clinical history, physical examination, and laboratory tests, such as one or more pancreatic or insulin autoantibody test. Because the intended use population for Type 1 diabetes mellitus includes subjects less than 18 years old, samples from representative numbers of these subjects must be included. Representative numbers of samples from all age strata must be also be included. The differential diagnosis groups must include, but not be limited to, the following: Type 2 diabetes mellitus; metabolic syndrome; latent autoimmune diabetes in adults; other autoimmune diseases such as celiac disease (without a concomitant diagnosis of Type 1 diabetes mellitus), systemic lupus erythematosus, rheumatoid arthritis, and Hashimoto's thyroiditis; infection; renal disease; and testicular cancer. Diseases for the differential groups must be based on established diagnostic criteria and clinical evaluation. For all samples, the diagnostic clinical criteria and the demographic information must be collected and provided. The clinical validation results must demonstrate clinical sensitivity and clinical specificity for the test values based on the presence or absence of Type 1 diabetes mellitus. The data must be summarized in tabular format comparing the interpretation of results to the disease status.

10

  • M) Expected/ reference values generated by testing an adequate number of samples from apparently healthy normal individuals.
  • Identification of risk mitigation elements used by the device, including iii description of all additional procedures, methods, and practices incorporated into the directions for use that mitigate risks associated with testing.
    1. Your 21 CFR 809.10(a) compliant label and 21 CFR 809.10(b) compliant labeling must include warnings relevant to the assay including:
    • i. A warning statement that reads "The device is for use by laboratory professionals in a clinical laboratory setting."
    • ii. A warning statement that reads "The test is not a stand-alone test but an adjunct to other clinical information. A diagnosis of Type 1 diabetes mellitus should not be made on a single test result. The clinical symptoms, results on physical examination, and laboratory tests (e.g., serological tests), when appropriate, should always be taken into account when considering the diagnosis of Type 1 diabetes mellitus and Type 2 diabetes mellitus."
    • A warning statement that reads "Absence of Zinc T8 autoantibody does not rule iii. out a diagnosis of Type 1 diabetes mellitus."
    • iv. A warning statement that reads "The assay has not been demonstrated to be effective for monitoring the stage of disease or its response to treatment."
    1. Your 21 CFR 809.10(b) compliant labeling must include a description of the protocol and performance studies performed in accordance with special control (1)(ii) and a summary of the results.