(387 days)
HeartFlow FFRcT is a post-processing software for the clinical quantitative and qualitative analysis of previously acquired Computed Tomography (CT) DICOM data for clinically stable symptomatic patients with coronary artery disease. It provides FFRcT, a mathematically derived quantity, computed from simulated pressure, velocity and blood flow information obtained from a 3D computer model generated from static coronary CT images. FFRcT analysis is intended to support the functional evaluation of coronary artery disease.
The results of this analysis are provided to support qualified clinicians to aid in the evaluation and assessment of coronary arteries. The results of HeartFlow FFR-c are intended to be used by qualified clinicians in conjunction with the patient's clinical history, symptoms, and other diagnostic tests, as well as the clinician's professional judgment.
The device is only for prescription use.
FFR v1.4 is post-processing image analysis software developed for the clinical quantitative and qualitative analysis of previously physician-acquired DICOM-compliant cardiac CT images and data, to assess the anatomy and function of the coronary arteries. The software displays the resulting coronary anatomy combined with functional information using graphics and text, including a computed and derived quantification of blood flow. termed FFR - to aid the clinician in the assessment of coronary artery disease.
The HeartFlow FFR cr software is housed at Heart Flow, Inc. The health care provider electronically sends the patient's CT scan data to HeartFlow. Inc. where a 3D computer model of the coronary arteries is developed and simulates blood flow in the models using computational fluid dynamics. A resulting report is electronically sent to the physician with the estimated fractional flow reserve (FFR) values (called FFRct values) displayed as color images of the patient's heart (Figure 1) and an associated color interpretation table (Table 1) indicating the error associated with each measurement range in the HFNXT clinical study.
Here's a breakdown of the acceptance criteria and the study detailing the device's performance:
Acceptance Criteria and Device Performance (HFNXT Study)
Acceptance Criteria (Target Rate) | Reported Device Performance (HFNXT Study - per-vessel) | Met? |
---|---|---|
Sensitivity: Lower 95% CI > 65% | 83.5% (Lower 95% CI: 75.3%) | MET |
Specificity: Lower 95% CI > 55% | 85.8% (Lower 95% CI: 81.5%) | MET |
Note: The document states that the prespecified target goals identified by the sponsor for sensitivity and specificity were 65%, respectively. However, it also clarifies that these target goals were not agreed upon by the FDA. The table reflects the FDA's acceptance based on the reported lower one-sided 95% confidence bounds being "significantly above the pre-specified target goals of 65% and 55%, respectively, and were considered acceptable." This implies the FDA may have had slightly different internal thresholds but accepted these results.
Additional Per-Subject Diagnostic Performance (FFR_CT vs. Site-Read cCTA with FFR ≤ 0.80 as reference standard):
Performance Metric | FFRCT ≤ 0.80 (Estimate % (95% Wilson CI)) | Site-Read cCTA > 50% (Estimate % (95% Wilson CI)) |
---|---|---|
Diagnostic Accuracy | 81.1% (95.8%-85.4%) | 52.8% (46.6%-58.8%) |
Sensitivity | 86.3% (77.0%-92.1%) | 93.8% (86.2%-97.3%) |
Specificity | 78.7% (72.1%-84.2%) | 33.9% (27.3%-41.2%) |
PPV | 65.1% (55.6%-73.5%) | 39.5% (32.8%-46.6%) |
NPV | 92.6% (87.2%-95.8%) | 92.2% (83.0%-96.6%) |
Study Details for HeartFlowNXT (HFNXT)
-
Sample Size used for the Test Set and Data Provenance:
- Sample Size: 254 adult subjects comprise the intention-to-diagnose (ITD) population. In total, 484 vessels were analyzed for direct comparison of invasive FFR and FFR_CT.
- Data Provenance: Prospective, multicenter, nonrandomized study conducted at 11 sites in 8 countries across Canada, Europe, and Asia. Data was collected from September 2012 to August 2013.
-
Number of Experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document states that "All invasive FFR data was reviewed by an independent FFR/QCA core laboratory." The number of experts or their specific qualifications (e.g., years of experience, subspecialty) are not explicitly detailed in the provided text.
-
Adjudication Method for the test set:
- The document implies a core lab review for invasive FFR data, suggesting a centralized and likely standardized method, but a specific "adjudication method" like 2+1 or 3+1 for discrepancies is not explicitly described for the final ground truth determination. The FFR_CT and cCTA data were interpreted in a blinded fashion, but the process for resolving disagreements or establishing the definitive FFR ground truth isn't detailed.
-
If a multi-reader, multi-case (MRMC) comparative effectiveness study was done:
- The HFNXT study compared the diagnostic performance of FFR_CT against cCTA alone. While it assesses the effectiveness of FFR_CT, it doesn't explicitly describe itself as a typical MRMC study designed to show how much human readers improve with AI vs. without AI assistance.
- Effect Size with AI vs. without AI assistance: The study showed "Per-subject FFR_CT specificity compared to site-read cCTA demonstrated superior diagnostic ability (p
§ 870.1415 Coronary vascular physiologic simulation software device.
(a)
Identification. A coronary vascular physiologic simulation software device is a prescription device that provides simulated functional assessment of blood flow in the coronary vascular system using data extracted from medical device imaging to solve algorithms and yield simulated metrics of physiologic information (e.g., blood flow, coronary flow reserve, fractional flow reserve, myocardial perfusion). A coronary vascular physiologic simulation software device is intended to generate results for use and review by a qualified clinician.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Adequate software verification and validation based on comprehensive hazard analysis, with identification of appropriate mitigations, must be performed, including:
(i) Full characterization of the technical parameters of the software, including:
(A) Any proprietary algorithm(s) used to model the vascular anatomy; and
(B) Adequate description of the expected impact of all applicable image acquisition hardware features and characteristics on performance and any associated minimum specifications;
(ii) Adequate consideration of privacy and security issues in the system design; and
(iii) Adequate mitigation of the impact of failure of any subsystem components (
e.g., signal detection and analysis, data storage, system communications and cybersecurity) with respect to incorrect patient reports and operator failures.(2) Adequate non-clinical performance testing must be provided to demonstrate the validity of computational modeling methods for flow measurement; and
(3) Clinical data supporting the proposed intended use must be provided, including the following:
(i) Output measure(s) must be compared to a clinically acceptable method and must adequately represent the simulated measure(s) the device provides in an accurate and reproducible manner;
(ii) Clinical utility of the device measurement accuracy must be demonstrated by comparison to that of other available diagnostic tests (
e.g., from literature analysis);(iii) Statistical performance of the device within clinical risk strata (
e.g., age, relevant comorbidities, disease stability) must be reported;(iv) The dataset must be adequately representative of the intended use population for the device (
e.g., patients, range of vessel sizes, imaging device models). Any selection criteria or limitations of the samples must be fully described and justified;(v) Statistical methods must consider the predefined endpoints:
(A) Estimates of probabilities of incorrect results must be provided for each endpoint,
(B) Where multiple samples from the same patient are used, statistical analysis must not assume statistical independence without adequate justification, and
(C) The report must provide appropriate confidence intervals for each performance metric;
(vi) Sensitivity and specificity must be characterized across the range of available measurements;
(vii) Agreement of the simulated measure(s) with clinically acceptable measure(s) must be assessed across the full range of measurements;
(viii) Comparison of the measurement performance must be provided across the range of intended image acquisition hardware; and
(ix) If the device uses a cutoff threshold or operates across a spectrum of disease, it must be established prior to validation, and it must be justified as to how it was determined and clinically validated;
(4) Adequate validation must be performed and controls implemented to characterize and ensure consistency (
i.e., repeatability and reproducibility) of measurement outputs:(i) Acceptable incoming image quality control measures and the resulting image rejection rate for the clinical data must be specified, and
(ii) Data must be provided within the clinical validation study or using equivalent datasets demonstrating the consistency (
i.e., repeatability and reproducibility) of the output that is representative of the range of data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment;(A) Testing must be performed using multiple operators meeting planned qualification criteria and using the procedure that will be implemented in the production use of the device, and
(B) The factors (
e.g., medical imaging dataset, operator) must be identified regarding which were held constant and which were varied during the evaluation, and a description must be provided for the computations and statistical analyses used to evaluate the data;(5) Human factors evaluation and validation must be provided to demonstrate adequate performance of the user interface to allow for users to accurately measure intended parameters, particularly where parameter settings that have impact on measurements require significant user intervention; and
(6) Device labeling must be provided that adequately describes the following:
(i) The device's intended use, including the type of imaging data used, what the device measures and outputs to the user, whether the measure is qualitative or quantitative, the clinical indications for which it is to be used, and the specific population for which the device use is intended;
(ii) Appropriate warnings specifying the intended patient population, identifying anatomy and image acquisition factors that may impact measurement results, and providing cautionary guidance for interpretation of the provided measurements;
(iii) Key assumptions made in the calculation and determination of simulated measurements;
(iv) The measurement performance of the device for all presented parameters, with appropriate confidence intervals, and the supporting evidence for this performance. Per-vessel clinical performance, including where applicable localized performance according to vessel and segment, must be included as well as a characterization of the measurement error across the expected range of measurement for key parameters based on the clinical data;
(v) A detailed description of the patients studied in the clinical validation (
e.g., age, gender, race or ethnicity, clinical stability, current treatment regimen) as well as procedural details of the clinical study (e.g., scanner representation, calcium scores, use of beta-blockers or nitrates); and(vi) Where significant human interface is necessary for accurate analysis, adequately detailed description of the analysis procedure using the device and any data features that could affect accuracy of results.