(680 days)
The BrainPort V100 is an oral electronic vision aid that provides electro-tactile stimulation to aid profoundly blind patients in orientation, mobility, and object recognition as an adjunctive device to other assistive methods such as the white cane or a guide dog.
The BrainPort V100 is an electronic assistive aid that translates images of objects captured by a digital camera into electro-tactile signals that are presented to the user's tongue. With training, users are able to use the electrotactile signals to perceive shape, size, location and motion of objects. The BrainPort V100 is intended to augment, rather than replace, other assistive technology such as the white cane or guide dog. The BrainPort V100 is not used to diagnose or treat the underlying condition that leads to the user's visual impairment. The BrainPort V100 is intended for prescription use only and for single patient use. The BrainPort V 100 consists of three components: the headset, the controller (also known as the handset), and the battery charger.
Acceptance Criteria and Device Performance for BrainPort V100
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Safety | Acceptable rate (< 10%) of clinically significant device-related adverse events. | 100% (57/57) success rate based on completers, with a one-sided 97.5% confidence interval of 93.7%, meeting the primary safety endpoint. Most adverse events were non-serious and did not require action or cessation of use. |
| Effectiveness | Subjects perform at a rate of 50% or better on the object recognition test. | 92.2% (52/57) success rate based on completers, with a one-sided 97.5% confidence interval of 81.7%, meeting the primary effectiveness endpoint. Worst-case analysis for ITT cohort: 69.3% (52/75) responder subjects, lower limit of 97.5% CI was 57.6% (exceeds 50%). |
| Training Effectiveness | Training program sufficiently addresses usability and validates the effectiveness of training methods and the "train-the-trainer" process. | Training program was utilized and validated. End-user success in quantifiable skills (FrACT test, Object recognition, O&M Hallway signs test) validated the training manual and train-the-trainer process. |
| Biocompatibility | Patient-contacting components (IOD, cable & headset sleeve) must be demonstrated to be biocompatible. | Biological risk assessment and applicable tests based on ISO10993-1 were conducted. Patient-contacting materials are well-characterized in oral electro-stimulating applications, and assessment deemed adequate. |
| Shelf Life/Sterility | Non-sterile, single-user reusable device. No specified shelf life. Cleaning instructions for IOD. | The device is non-sterile and reusable. No specified shelf life. IOD cleaning instructions with 70% isopropyl alcohol are included and tested for material compatibility. |
| Electrical Safety | Conformance with IEC 60601-1:2005 (3rd edition) and IEC 60601-1-2:2007. | Device tested per IEC 60601-1:2005 and IEC 60601-1-2:2007, conforming and passing applicable sub-clauses. |
| EMC | Conformance with IEC 60601-1-2:2007 criteria and withstand higher levels for home-use (ESD, magnetic field, conducted RF, radiated RF). | Device tested per IEC 60601-1-2:2007, including higher home-use levels (+/- 8kV contact, +/-15kV air ESD, 30 A/m magnetic field, 6 V r.m.s. conducted RF, 10 V/m radiated RF). All results demonstrated acceptable performance. |
| Battery Reliability | Sufficiently able to detect low battery capacity and alert the user. Lasts for a reasonable duration and charge cycles. | Device automatically announces "low battery" when capacity is low. Fully charged battery provides ~3 hours of use and is expected to last >500 charge cycles. |
| Software Verification | Software verification, validation, and hazard analysis performed according to FDA guidance for moderate level-of-concern software. | Software V&V included information on hazards, requirements, development, traceability, and anomalies per FDA guidance. (Note: data exchange is unencrypted, but no PII is stored). |
| Moisture Ingress | Characterize internal moisture penetration, ensuring no penetration to electronic components. | Testing showed very limited moisture ingress and no penetration to the electronic component area after 690 hours of continuous immersion (simulating 6.5 years of use). IOD is encapsulated in epoxy. Device is IP20 (controller/headset) and IPX4 (IOD). |
| IOD Durability | Durability of IOD and electrodes under continuous stimulation. | IOD electrodes continuously stimulated for 8 days (simulating 38 months of normal use). No tarnishing or pitting observed under visual or 120x optical magnification. |
| MR Compatibility | If not MR compatible, clearly label and warn users. | Device is "MR Unsafe" and labeling clearly states it has not been evaluated, should be removed before entering an MR environment, and that safety is unknown due to potential heating, migration, or artifact. |
Study Proving Device Acceptance Criteria
The BrainPort V100's acceptance criteria were primarily proven through a one-year clinical study, augmented by additional adverse event data collected outside the United States, and supported by various bench tests.
Clinical Study Elements:
-
Sample Size used for the Test Set and Data Provenance:
- Test Set Sample Size: 75 enrolled subjects (74 completed training, 57 completers for primary endpoints).
- Data Provenance: Prospective, single-arm, open-label clinical study conducted at 7 sites in the U.S. and Canada.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:
- The document does not explicitly state the number of experts used to establish a ground truth in the traditional sense of consensus reading for diagnostic imaging.
- Instead, the "ground truth" for effectiveness in object recognition and mobility was established by the subjects' observed performance on standardized tasks (e.g., correctly identifying and touching objects, ambulating to signs).
- The study involved a test administrator for the Object Recognition test and a similar setup for the Mobility Test, who would objectively observe and record subject performance. While their specific qualifications are not detailed beyond "test administrator," they would be trained in conducting the assessments.
- Oral health exams were conducted by unspecified healthcare professionals as part of quarterly follow-up.
-
Adjudication Method for the Test Set:
- Not applicable in the typical sense for image or clinical data interpretation. The effectiveness endpoints (object recognition, mobility) were based on direct observation and recording of subject performance against defined criteria (e.g., touching the correct object, reaching the correct sign). Adverse events were determined by the investigator.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This device is an assistive technology, not an AI-powered diagnostic tool. The goal was to demonstrate if the device itself, when used by profoundly blind individuals, improved their ability in orientation, mobility, and object recognition, rather than evaluating human reader performance with or without AI assistance. The study assessed the users' absolute performance with the device after training, not their improvement over a baseline without the device (though they are profoundly blind, implying severe functional limitation without assistive aids).
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No, an algorithm-only standalone performance was not done. The BrainPort V100 is designed as a human-in-the-loop assistive device. Its function is to convert visual information into tactile signals for human interpretation via the tongue. Therefore, evaluating the algorithm's performance without a human user is not relevant to its intended use or clinical effectiveness.
-
The Type of Ground Truth Used:
- Behavioral Performance and Clinical Observation: For effectiveness, the ground truth was the observed, objective behavioral performance of the profoundly blind subjects on defined tasks (e.g., correctly identifying an object, successfully navigating to a sign). For safety, the ground truth was clinical assessment of adverse events by investigators and oral health examinations, alongside subject-reported adverse events.
-
The Sample Size for the Training Set:
- The document describes a "training phase" for the clinical study subjects, where 74 out of 75 enrolled subjects completed 2-3 days (10 hours) of supervised training. This initial training segment is part of preparing the subjects to use the device for the study, rather than training an algorithm.
- The document does not specify a "training set" sample size for the device's internal algorithms (e.g., for camera processing or tactile signal generation). The device's core functionality (translating luminance into electrotactile signals) appears to be based on pre-programmed logic rather than machine learning models that require a distinct training set.
-
How the Ground Truth for the Training Set Was Established:
- Given that the device's algorithms do not appear to be machine learning-based with a "training set" in the traditional sense, this question is not directly applicable.
- The "ground truth" for the human training (i.e., for the users and trainers) was established through best practices in blindness rehabilitation and education. User training protocols were developed by Wicab and validated in the study. Trainers had relevant experience (CLVS, COMS, TVI) and were further trained by Wicab according to specific procedures. The effectiveness of this human training was validated by end-user success in the objective performance tasks.
{0}------------------------------------------------
DE NOVO CLASSIFICATION REQUEST FOR BRAINPORT V100
REGULATORY INFORMATION
FDA identifies this generic type of device as:
Oral electronic vision aid. An oral electronic vision aid is a battery-powered prescription device that contains an electrode stimulation array to generate electro-tactile stimulation patterns that are derived from digital object images captured by a camera. It is intended to aid profoundly blind patients in orientation, mobility, and object recognition as an adjunctive device to other assistive methods such as the white cane or a guide dog.
NEW REGULATION NUMBER: 886.5905
CLASSIFICATION: CLASS II
PRODUCT CODE: PIC
BACKGROUND
DEVICE NAME: BRAINPORT V100
SUBMISSION NUMBER DEN130039
DATE OF DE NOVO: August 7, 2013
- CONTACT: Wicab Inc. 8313 Greenway Blvd. Suite 100 Middleton, WI 53562 608-829-4500
REQUESTER'S RECOMMENDED CLASSIFICATION: CLASS II
INDICATIONS FOR USE
The BrainPort V100 is an oral electronic vision aid that provides electro-tactile stimulation to aid profoundly blind patients in orientation, mobility, and object recognition as an adjunctive device to other assistive methods such as the white cane or a guide dog.
LIMITATIONS
BrainPort V100 is indicated for prescription use only.
Candidates for the device should be profoundly blind.
Contraindications:
- . Any neurological condition that causes impaired sensitivity to the tongue or loss of consciousness
{1}------------------------------------------------
Warnings:
- . Limited data are available on the long-term effects of electrical stimulation of the tongue. Long-term effects (beyond one year) have not been evaluated in clinical trials.
- The BrainPort V100 should be not used in patients experiencing numbness or lack of feeling of the tongue, or in patients with a history of injury to the tongue resulting in impaired sensation or use of the tongue.
- The BrainPort V100 should only be used after the user has completed training. Do not give the device to untrained individuals for use.
- Limited data are available on stimulation sensitivity for individuals with oral conditions such as oral ulcerations, herpes simplex, oral thrush, and geographic tongue. If use of the device causes discomfort, discontinue use.
- . The BrainPort V100 has not been thoroughly evaluated in the presence of dental implants. The safety of dental implants in BrainPort users is unknown. The use of this device potentially may cause heating of dental implants; chronic use of this device potentially may result in loosening and failure of dental implants.
- The BrainPort V100 is not used to diagnose or treat the underlying condition that leads to ● the user's visual impairment.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS. PRECAUTIONS AND CONTRAINDICATIONS.
DEVICE DESCRIPTION
The BrainPort V100 is an electronic assistive aid that translates images of objects captured by a digital camera into electro-tactile signals that are presented to the user's tongue. With training, users are able to use the electrotactile signals to perceive shape, size, location and motion of objects. The BrainPort V100 is intended to augment, rather than replace, other assistive technology such as the white cane or guide dog. The BrainPort V100 is not used to diagnose or treat the underlying condition that leads to the user's visual impairment. The BrainPort V100 is intended for prescription use only and for single patient use
The BrainPort V 100 consists of three components: the headset, the controller (also known as the handset), and the battery charger as depicted in Figure 1. The headset is typically worn on the user's head and provides the image input and output functions of the device. The controller is generally handheld and provides the processing and power functions of the device. The battery charger is a commercial off-the-shelf unit used to charge the removable battery.
Image /page/1/Figure/11 description: The image shows a person wearing a headset and glasses with a device in their mouth. The headset is connected to the glasses and has a wire that runs down to the device in the person's mouth. The person is also holding a controller in their hand. The controller has several buttons and a screen.
Figure 1: Artist Rendition of BrainPort V100 components: Controller and Headset
{2}------------------------------------------------
Headset
A digital video camera is mounted on a pair of sunglasses at the nose bridge. The field of view of the camera is user-controlled and varies from narrow to wide angle views. There are two cables permanently attached to the left ear piece: the IOD assembly and the headset cable.
Intra Oral Device (IOD)
The IOD (tongue electrode array) contains electrodes that act as "pixels" for the tongue. The flat side with the electrodes should be in contact with the front top surface of the tongue. The user should close their lips around the thin stem, maximizing tongue contact with the electrodes. The electrodes provide stimulation to the tongue based on information received from the camera. The electrodes provide electro-tactile information that varies based on the size, location and relative luminance of objects within a visual scene. There is one cable exiting the thin stem of the IOD that is permanently attached to the ear piece of the headset. Figure 2 below presents the IOD sub-assembly of the BrainPort V100.
Image /page/2/Picture/4 description: The image shows a polished finished IOD subassembly. The subassembly appears to be a tool with a rectangular head and a handle. The head is light-colored, possibly white or beige, and has a grid-like pattern on its surface. The handle is dark, likely brown or black, and is attached to the head.
Figure 2 - IOD Subassembly of the BrainPort V100
Controller
The Controller contains the battery as well as the user control features for the BrainPort V100. An antenna provides wireless connection capabilities. The Controller is generally handheld. A belt clip is provided for hands free operation.
Battery & Battery Charger
Rechargeable Lithium-Polymer batteries and a battery charger with factory instructions are included with each BrainPort V100. All batteries included with the system should be fully charged at 3.7VDC prior to initial use.
The general specifications for the BrainPort V100 are listed in Table 1:
{3}------------------------------------------------
| Physical | Length 13.3 cmwidth 5.6 cmheight 3.5 cm (excluding belt clip)weight <175 g (including battery)Cable length to headset 106 cm |
|---|---|
| Power | Internally powered by a lithium polymerrechargeable battery (3.7V, 2260mAh,8.4Wh) |
| Output Waveforms | MonophasicCapacitive coupling |
| Pulse Frequency | 200 Hz |
| Pulse Width | 25 μs |
| Surface Area | 0.46 mm2 (per electrode) |
| Voltage | 0 to 1.414 V rms (per electrode)0 to 14.14 V rms (device) |
| Current | 0 to 0.518 mA rms (per electrode)0 to 20.7 mA rms (device) |
| Energy/pulse | 10.35μJ |
Table 1- Relevant Product Specifications
TRAINING
The training philosophy is to facilitate the user's self-discovery of linking BrainPort V100 derived perception with the user's lifetime of multi-sensory experiences. The trainer provides activities where the user explores a situation in his typical manner (i.e., by reaching, touching, listening, etc.). The user then observes how that same situation is represented by the BrainPort V100.
To ensure proper use of the device, BrainPort V100 professionals and users are required to undergo training. Potential trainers of users of the BrainPort V100 should have relevant experience, such as experience working with the blind or visually impaired as well as professional credentials, such as certification as a Certified Low Vision Specialists (CLVS), Certified Orientation and Mobility Specialist (COMS) or Teachers of the Visually Impaired (TVI). All potential trainers will be further trained by Wicab according to Wicab procedures and only those who have successfully completed the training will be considered qualified to train users of the BrainPort V100. As part of the training program, a specialist from Wicab provides a demonstration to the trainer of typical training session(s), covering all tasks that are typically completed over several days of end user training. First, the specialist operates the BrainPort V 100 and focuses the trainer on the vRemote output to demonstrate how to monitor a user's performance and provide appropriate feedback and instruction to the user.
Second, the specialist supervises the trainer performing training with a new user (or blind-folded simulation) as outlined in the training procedures. During this hands-on phase, the specialist interjects and provides suggestions for improved training or user performance. This exercise is to ensure that the trainer can provide suitable training with the BrainPort V100.
Lastly, a Trainer and Patient Manual (printed and auditory format) of the BrainPort V100 is provided. The user profile is summarized in Table 2:
{4}------------------------------------------------
| Table 2 - User Profile | ||||
|---|---|---|---|---|
| ------------------------ | -- | -- | -- | -- |
| User | For use by individuals who are profoundly blind. |
|---|---|
| Education | Completed traditional blindness rehabilitation: white cane, and/or guidedog, and rehabilitation in activities of daily living |
| Knowledge | Minimum: ability to understand verbal training instructionsReading and comprehension at 10th grade level |
| Experience | All users must participate in a minimum of 10 hours of supervisedtraining per Wicab's training protocol prior to unsupervised use of thedevice |
| PermissibleImpairments | Diagnosis of no light perception or light perceptionBlindness may be acquired or congenitalAbsence of oral sensory impairments |
vRemote Feature
There are two ways to use the BrainPort V100: in standalone mode or in Wi-Fi mode to connect to Wicab's proprietary application software vRemote (Wicab Reference Number: SW- 000014). The software package (vRemote) receives data from a BrainPort V100 device and displays it on a computer. The optional Wi-Fi communications capability provides one way only transfer from the V100 to the PC screen. It provides information to the trainer about stimulation intensity, zoom, exposure, tilt, and inversion settings of the BrainPort V100 device. There is no patient information contained in Wi-Fi communications. The user will be advised that if the vRemote is used by the user's companion, it should not be used by the companion to provide any instructions to the user, but is simply for viewing.
The hardware and software of the BrainPort V100 device were designed and developed to ensure accurate and reliable wireless transmission of device data. The personal computer used to run the vRemote software must have an operating system of Windows XP or higher (.NET 4.0 or higher), and must be able to establish an 'ad-hoc' Wi-Fi connection (minimum 802.11 a/b) with the BrainPort V100 device.
SUMMARY OF NONCLINICAL/BENCH STUDIES
A. BIOCOMPATIBILITY/MATERIALS
The patient-contacting components of the BrainPort V100 are the IOD and cable & headset sleeve, which are identified in Table 3.
| Component | Material |
|---|---|
| Intra-oral Device (IOD)(oral contact) | b(4) |
| Cable & Headset Sleeve(skin contact) | b(4) |
Table 3 - Materials of Patient Contact
{5}------------------------------------------------
The applicant conducted a biological risk assessment and applicable test on these materials based on ISO10993-1 (Biological Evaluation of Medical Devices). Additionally, the patient contacting materials in BrainPort V100 are well characterized in oral electro- stimulating application and an assessment was deemed adequate to ensure sufficient biocompatibility.
B. SHELF LIFE/STERILITY
The BrainPort V100 is a non-sterile, single user reusable device.
There is no specified shelf life for the device, which is acceptable based upon the nature of the device components. However, because the IOD does come in contact with the patient's lips and tongue, and is further intended for re-use, the Instructions for Use include cleaning. It recommended to clean the IOD once per week using 70% isopropyl alcohol (rubbing alcohol), which has been tested and shown to be compatible with the materials used in the IOD. To avoid damage to the IOD, it should not be exposed to any bleach agents.
C. ELECTRICAL SAFETY, ELECTROMAGNETIC COMPATIBILITY (EMC) AND GENERAL ELECTRICAL TESTING
1. ELECTRICAL SAFETY
The device was tested per IEC 60601-1: 2005 (3rd edition) and IEC 60601-1-2: 2007. The device is in conformance with the standards and passed applicable sub-clauses.
2. ELECTROMAGNETIC COMPATIBILITY (EMC)
Testing was performed to address EMC concerns for this device, which is intended to be used in hospital and home-use environments. The device was tested and evaluated per the IEC 60601-1-2:2007 criteria and the device was tested at higher levels for home-use: specifically electrostatic discharge (ESD) at +/- 8 kV contact discharge, +/-15 kV air discharge, power frequency magnetic field at 30 A/m, Conducted RF Immunity: 6 V r.m.s. and radiated RF at 10 V/m. All results demonstrated acceptable performance.
EMC labeling, as necessary to claim compliance with IEC 60601-1-2:2007, has been included in the Trainer and User Manuals. This includes tests at higher test levels than the recommended hospital environment.
3. BATTERY TESTING
Battery testing was conducted to show the device is sufficiently able to detect when the battery capacity is low. In a low battery condition, the device may shutdown automatically leaving the patient without feedback. Thus, when the battery level decreases to the designated low level, the device will repeatedly announce "low battery." This announcement acts as a reminder that a replacement battery should be inserted in order to continue using the device.
The battery supplied with the BrainPort V100 is a rechargeable Lithium-Polymer battery. When fully charged, the battery provides approximately three hours of use and is expected to last through more than 500 charge cycles.
{6}------------------------------------------------
D. MAGNETIC RESONANCE (MR) COMPATIBILITY
The BrainPort V100 has not been tested for MR compatibility. Therefore, it is not appropriate for this device to be in or near forces encountered in MR imaging suites. Accordingly, the following has been included as part of the labeling:
MR Unsafe. The BrainPort V100 has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating migration, or image artifact in the MR environment. The safety of BrainPort V100 in the MR environment is unknown. The BrainPort V100 should be removed before entering an MR scan room or having an MR scan. Scanning patient who has this device could result in patient injury.
E. SOFTWARE
The BrainPort V100 acquires data from a digital camera and transforms the data to a 400 byte array (an IOD image). Each byte in the IOD image corresponds to an electrode on the IOD. Each value in the array is used to set to a voltage level related to the luminance data in the corresponding camera image: a dark region in the image will result in a low value and a bright region will result in a higher value. Using a progressive scanning technique, the IOD image is presented to electrodes.
Cyber and Information Security Risks
- No patient identifiable information is requested or stored, nor is there a means for the . user to enter any such information on the V100 or vRemote (application on PC).
- There is a data exchange between the V100 and vRemote to establish an 'ad hoc' . connection between the devices and for the V100 to send status information (including camera and IOD images) to vRemote.
- The data exchange is unencrypted. .
Software Verification and Validation
Per the FDA software guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005), the software is considered as a moderate level-of-concern. Among other information, the software documentation included information relating to hazards, software requirements, software development, traceability and potential unresolved anomalies.
F. PERFORMANCE TESTING - BENCH
1. MOISTURE INGRESS TESTING
Testing was conducted to characterize the Internal Moisture Penetration of the device. b(4)
It should
be noted that the duration of this test exceeded the observed device usage life of even the most active users by a safety factor of nearly 10. For this test, there were 30 active days of ~23hr of testing per day (690 hrs.). The most active users from the sponsor's clinical study on average had a usage of 530 min per month; 106 hours per
{7}------------------------------------------------
year. Also. 690 hr / 106hr = 6.5 years of use simulated which is more than three times the proposed device service life of 2 years.
Since continuous immersion in a saliva-like environment showed very limited moisture ingress and no penetration to the electronic component area, the opportunity for dendrite growth is limited. In addition, the entire IOD is encapsulated in epoxy. including the electrical components that comprise the active circuitry. However, if there was a short within the IOD electrical components for any reason, including dendrite growth, there would be no anticipated impact on the basic safety of the device.
The BrainPort V100 is classified as IP20. The controller and headset must be kept dry. The device is protected against solid foreign objects over 12 mm. The IOD assembly is IPX4 rated (protection against splashing water).
2. IOD DURABILITY TEST
Testing was performed to determine the IOD durability. IOD with b(4)
. The IOD electrodes were continuously stimulated for 8 days (equivalent to simulating 38 months of normal use). No tarnishing or pitting was observed under visual inspection or 120x optical magnification on any of the electrodes.
SUMMARY OF CLINICAL INFORMATION
The applicant conducted a one-year clinical study to demonstrate the safety and effectiveness of the BrainPort V100 device in assisting blind and visually impaired individuals in object recognition and orientation and mobility activities. This study was augmented by additional adverse event data collected outside the United States.
Study Design
The BrainPort V100 device was evaluated in a single arm, open label clinical study of 75 enrolled blind/profoundly blind subjects (visual acuity of light perception or worse) at 7 sites in the U.S. and Canada. Subjects underwent 2-3 days (10 hours) of training followed by in-home use over the span of 12 months.
The duration of each subject's participation was intended to be approximately 12 months. Total study duration was targeted to be approximately 18 months, or a sufficient time to enroll and evaluate 75 subjects with 12 months of device use.
The primary safety objective was to demonstrate an acceptable rate (< 10%) of clinically significant device-related adverse events, which required an observed event-free rate of approximately 97%. In addition, all adverse events were collected and summarized according to seriousness, event type, relationship to device and study procedures as determined by the investigator, severity, action taken, outcome, and whether the event lead to subject withdrawal from study participation.
The applicant conducted the following study assessments:
{8}------------------------------------------------
-
- Object Recognition test: Four objects were placed in a row on a table draped in black cloth. Subjects were asked to reach out and touch the object requested by the test administrator, without touching any other object. This was repeated for 20 trials. The objects were: softball, coffee mug, plastic banana and a highlighter marker.
-
- Mobility Test: Four signs were placed in a 15 foot (4.57 meter) hallway. Subjects were asked to find and ambulate to one requested target sign, indicating their response by touching the found sign or within 5 inches (12.7 centimeters) around the perimeter of the designated sign. Signs included: Men's Room, Women's Room, Stairs and Danger.
-
- Oral Health Exam: Study participants were required to undergo quarterly follow-up evaluations during the year of participation, visiting the study site to receive an oral health exam and to complete assessments of threshold intensity and effectiveness.
-
- Minimum Threshold Test: A minimum perception threshold on the tongue was recorded using the BrainPort V100 device to assess changes in threshold sensitivity over time. A 4x4 square (16 electrodes) was presented to the user across three trials. Users were instructed to increase intensity to a point where they first feel the stimulation level and the intensity level output will be recorded. Three trials were completed and recorded.
-
- Adverse Event Reports: All user-initiated adverse event reports, monthly telephone query adverse event reports and adverse event reports resulting from the oral exam were collected.
Study Results
A total of 75 subjects were enrolled, with 74 completing the required device training (Training Phase) prior to entering the "At-Home" Phase. A total of 18 study participants either withdrew consent, or were withdrawn from participation by the principal investigator, prior to completion of the study. The results are provided for both the intent to treat (ITT) and the 'completer' cohorts below.
An internal-control device log was available that was generated by the device. Despite the lower usage reported by subjects in the home logs (i.e., the home training log reported by the subjects), the majority of study subjects achieved or exceeded the minimum device usage target at each quarter and over the 12 month study period based on the device log of the BrainPort V100 usage.
Safety Evaluation: For the primary safety endpoint, the success rate is 100% (57/57) based on the completers and its one-sided 97.5% confidence interval is 93.7%, and therefore the study met the primary safety endpoint.
Adverse Events: The majority of the subjects who reported any adverse events did not require any action and device use continued unchanged. Of the subjects who reported adverse events, 18 subjects had decreased device use or discontinued device use (mostly temporarily), with only two of these subjects reporting events that were considered possibly or probably related to the device. Forty-seven (47) subjects had a total of 137 adverse events during the 12 month investigation. Of these 137 adverse events, 121 were non-serious events. Twenty-eight (28) of these 121 non-serious events were "device-related" and five (5) of those 28 events were "definitely device-related" and constituted changes in taste which were not persistent.
{9}------------------------------------------------
In addition, supplemental safety data (e.g., adverse events) were provided for Wicab's Balance device, which uses the same IOD unit as the one used in the BrainPort V100 device. Of the 148 users from which data were collected, there were no reports of serious device-related adverse events. Three subjects reported non-serious adverse events considered possibly or probably device-related that led to withdrawal or treatment/medical intervention. In addition, Wicab's Balance device has also been used by 617 users outside of the U.S., including consumers and subjects from additional studies that occurred prior to the safety evaluation for the BrainPort V 100.
Effectiveness Evaluation: The primary effectiveness endpoint was considered to be met if subjects performed at a rate of 50% or better on the object recognition test. The results showed a success rate was 92.2% (52/57) and its one-sided 97.5% confidence interval was 81.7% for the complete cohort, and therefore the study met the primary effectiveness endpoint. In the worstcase analysis for the ITT cohort, the proportion of responder subjects was 69.3% (52/75) and the lower limit of the one-sided 97.5% confidence interval was 57.6%, which exceeds the performance goal of 50%.
In addition, the training program utilized and validated for this study sufficiently addresses the usability of the device for its intended population. It addressed all aspects of the device and the specific environments in which it is intended to be used. Effectiveness would not have been able to be demonstrated without this thorough training program.
Usability: Usability: (i.e., ensuring that the end user can properly use the device) has been assessed by the following:
- . The applicant validated the BrainPort V100 device training methods as described in the Training Manual and evaluated the effectiveness of the trainer process as outlined in the "train-the-trainer" course syllabus.
- Trainers are professionals who have both credentials and practical experience working with the blind, such as: Certified Low Vision Specialists (CLVS), Certified Orientation and Mobility Specialist (COMS) and Teachers of the Visually Impaired (TVI).
- As part of the usability assessment, three quantifiable skills were tested to measure . the validity of the BrainPort V100 device Training Manual techniques: 1) Freiberg Visual Acuity Test (FrACT test); 2) Object recognition (Table- top test); and 3) Orientation and Mobility (Hallway signs test).
- End user success was used to validate the effectiveness of the both the BrainPort V100 device Training Manual techniques and the train-the-trainer process. The training procedures used and validated by the applicant are thorough, task specific, and representative of the various types of environmental obstacles that a blind person may encounter in navigation both at home and in public places.
LABELING
The BrainPort V100 Trainer and User manuals are consistent with the clinical data and sufficiently cover the hazards and other clinically relevant information that may impact use of
{10}------------------------------------------------
the device. The labeling is sufficient and satisfies the requirements of 21 CFR §801.109 Prescription devices. The User manual is available in both print and audio formats.
The labeling includes the following information:
- . Contraindications, warnings, precautions, and limitations needed for safe use of the device.
- . A summary of the clinical performance testing, including adverse events and complications.
- Appropriate instructions for training and use of device.
- Product specifications and technical references specifications, including EMC ● compatibility and wireless technology information
- The potential risks associated with use of the device.
In addition, the labeling includes the following warnings that are specific to the risks associated with the use of IOD (tongue electrode array):
- . Long-term use. Limited data are available on the long-term effects of electrical stimulation of the tongue. Long-term effects (beyond one year) have not been evaluated in clinical trials.
- . Limited data is available on use of the device exceeding an average of between 250 and 400 minutes per BrainPort V100 User Manual 8 month, and a maximum of 1550 minutes per month. Wicab recommends that you tailor your use of the device to be within these time limits since long-term effects (beyond one year) exceeding this usage have not been evaluated in clinical trials.
- Oral Health. Limited data is available on stimulation sensitivity for individuals with oral ● conditions such as oral ulcerations, herpes simplex, oral thrush, and geographic tongue. If use of the device causes discomfort, discontinue use.
- . Oral Health. Individuals with high, narrow palatal vaults should discontinue use of the device if use causes discomfort.
- Oral Health. Individuals with maxillary or mandibular tori that interfere with the IOD placement such that full contact with the tongue is prevented should seek additional training to gain the most benefit of the device.
- . Dental Appliances (orthodontic appliances, removable partial or full dentures, lingual amalgam alloy restoration, metal crowns, etc.) Electrical stimulation of the tongue when metal appliances/surfaces are present may change results and/or cause unintended stimulation. If the stimulation causes discomfort, remove the IOD from your mouth and discontinue use of the device. If you notice a change in your dental device or appliance (warmth or looseness), discontinue the use of the device and contact your dentist.
- . Dental Implants. The BrainPort device has not been thoroughly evaluated in the presence of dental implants. The safety of dental implants in BrainPort users is unknown. The use of this device potentially may cause heating of dental implants; chronic use of this device potentially may result in loosening and failure of dental implants.
RISKS TO HEALTH
Table 1 identifies the risks to health that may be associated with use of an oral electronic vision aid and the measures necessary to mitigate these risks.
{11}------------------------------------------------
| Identified Risk | Mitigation Method |
|---|---|
| Irritation, Discomfort or Adverse EventsInvolving the Mouth, Tongue, or Gums | Clinical TestingLabeling |
| Adverse Tissue Reaction | Biocompatibility TestingLabeling |
| Unit (Hardware) Malfunction, FunctionalReliability | Non-Clinical Performance TestingClinical TestingLabeling |
| Software Malfunction | Software Verification, Validation, andHazard Analysis |
| Use Error | Clinical TestingHealthcare Professional TrainingPatient TrainingLabeling |
| Interference with Other Devices | Electromagnetic Compatibility (EMC)and Electromagnetic Interference(EMI) TestingWireless Coexistence TestingLabeling |
| Electrical Shock | Electrical Safety TestingLabeling |
Table 1 - Identified Risks to Health and Mitigation Measures
SPECIAL CONTROLS
In combination with the general controls of the Food, Drug& Cosmetic Act, the oral electronic vision aid is subject to the following special controls:
-
- Clinical performance testing must demonstrate an acceptable adverse event profile, including adverse events involving the mouth, tongue, and gums and demonstrate the effect of the stimulation to provide clinically meaningful outcomes. The clinical performance testing must also investigate the anticipated conditions of use, including potential use-error, intended environment of use, and duration of use.
-
- Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including simulated moisture ingress, device durability, and battery reliability.
-
- Software verification, validation and hazard analysis must be performed.
-
- Analysis/testing must validate electromagnetic compatibility (EMC).
- Analysis/testing must validate electrical safety. 5.
-
- Analysis/testing must assess and validate wireless coexistence concerns. Any elements of the device that contact the patient must be demonstrated to be biocompatible.
-
- Training must include elements to ensure that the healthcare provider and user can identify the safe environments for device use, use all safety features of the device, and operate the device in the intended environment of use.
{12}------------------------------------------------
-
- Labeling for the trainer and user must include a summary of the clinical testing including, adverse events encountered under use conditions, summary of study outcomes and endpoints; and information pertinent to use of the device including the conditions under which the device was studied (e.g., level of supervision or assistance, and environment of use).
BENEFIT/RISK DETERMINATION
The risks of the device are described above in the Summary of Clinical Information.
The risks of the device are based on non-clinical data as well as data collected from clinical studies described above.
Potential risks arising from the use of the BrainPort V100 include falls, allergic reaction to the materials in the device, tongue irritation from the electrodes or excessive stimulation and continuous electrical stimulation at a specific locus on the tongue along with the tissue damage that may occur (thus the perception of taste may change). Additionally, there is unknown potential risk associated with long-term use (e.g., damage to existing dental appliances worn and to dental implants due to potential long-term heating)
The probable benefits of the device are also based on data collected in clinical studies described above. The clinical information has demonstrated the effectiveness of BrainPort V100 in assisting profoundly blind (blind and visually impaired) individuals in object recognition and orientation and mobility activities. The BrainPort device demonstrates potential utility in assisting severely visually impaired and blind individuals in recognition of crude geometric and alphabetic shapes and characters, and as a potential augmentation tool to mobility and way finding tasks.
Additional factors considered in determining probable risks and benefits for the BrainPort V100 include: (1) the patient population is profoundly blind. The disease severity is not addressed by any form of currently-available treatment method, other than utilization of alternative assistive methods (such as a white cane or a guide dog). Therefore, tolerance of risk for benefit is not an issue. (2) Long-term safety and effectiveness data for the BrainPort V100 device are not available.
Given the available information above, the data support the conclusion that for aid in orientation, mobility, and object recognition, the probable benefits outweigh the probable risks for the BrainPort V100.
CONCLUSION
The de novo for the Brainport V100 is granted and the device is classified under the following:
Product Code: PIC Device Type: Oral electronic vision aid Class: II Regulation: 21 CFR 886.5905
§ 886.5905 Oral electronic vision aid.
(a)
Identification. An oral electronic vision aid is a battery-powered prescription device that contains an electrode stimulation array to generate electrotactile stimulation patterns that are derived from digital object images captured by a camera. It is intended to aid profoundly blind patients in orientation, mobility, and object recognition as an adjunctive device to other assistive methods such as a white cane or a guide dog.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate an acceptable adverse event profile, including adverse events involving the mouth, tongue, and gums and demonstrate the effect of the stimulation to provide clinically meaningful outcomes. The clinical performance testing must also investigate the anticipated conditions of use, including potential use error, intended environment of use, and duration of use.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including simulated moisture ingress, device durability, and battery reliability.
(3) Software verification, validation, and hazard analysis must be performed.
(4) Analysis/testing must validate electromagnetic compatibility.
(5) Analysis/testing must validate electrical safety.
(6) Analysis/testing must assess and validate wireless coexistence concerns.
(7) Any elements of the device that contact the patient must be demonstrated to be biocompatible.
(8) Training must include elements to ensure that the healthcare provider and user can identify the safe environments for device use, use all safety features of the device, and operate the device in the intended environment of use.
(9) Labeling for the trainer and user must include a summary of the clinical testing including adverse events encountered under use conditions, summary of study outcomes and endpoints, and information pertinent to use of the device including the conditions under which the device was studied (
e.g., level of supervision or assistance, and environment of use).