(609 days)
The DUET™ EDMS is indicated for temporary draining and monitoring of cerebrospinal fluid (CSF) flow from the lumbar subarachnoid space in:
-
- Patients undergoing open descending thoracic aortic aneurysm (open TAA) or open descending thoraco-abdominal aortic aneurysm (open TAAA) repair surgery.
-
- Patients post TAA/TAAA repair that become symptomatic with neurological deficit such as paraplegia.
The Medtronic DUET™ External Drainage and Monitoring System (EDMS) that is the subject of this De Novo request is designed to drain and monitor cerebrospinal fluid (CSF) from the lumbar subarachnoid space.
The DUET™ EDMS consists of the following: a green-striped patient connection line (pressure tubing) with an inner diameter of 0.075 ± 0.005 inches, an outer diameter of 0.124 ± 0.003 inches, and a total length of 60 inches (9), a patient line stopcock (10), a main system stopcock (8) that may be attached at two locations on the main panel, a drip chamber (4) with a drip chamber stopcock (5), a rotatable pressure scale (3), three latex-free needleless injection/CSF sampling sites (Figure 1b (IS-5), (IS-10) and (IS-11)) and a removable drainage bag (7) with approximate volumetric graduations and a hydrophobic microbial barrier air vent. There is a pole mount clamp (6) and a cord (12) with a cord lock (13) to enable independent suspension of the system, or to provide additional security when using the pole clamp as identified in Figure 1 below. It should be noted that the numbers in parentheses correspond with the numbers in Figure 1.
The DUET™ EDMS are not long-term implants but are intended for limited external drainage of CSF. The drainage flow of CSF into the DUETTM EDMS is uni-directional and gravity-driven; there is no recirculation of the CSF. During use, an external lumbar catheter inserted into the lumbar subarachnoid space is connected to the DUET™ EDMS patient connection line. The CSF drains through the catheter, into the patient connection line and into the graduated drip chamber. CSF collects in the drip chamber, exits the bottom of the chamber via another connection line and is collected in a drainage bag. In the event that the patient may require administration of fluid directly into the lumbar subarachnoid space or CSF sampling is required, the DUET™ EDMS features injection/CSF sampling ports integrated into the patient connection line. The DUET™ EDMS is completely disposable. The DUET™ EDMS is recommended for use with the Clear-Site™ Laser Level (cleared under K984053) that is provided separately.
The provided document describes the Medtronic Neurosurgery DUET™ External Drainage and Monitoring System (EDMS), a device intended for temporary draining and monitoring of cerebrospinal fluid (CSF) flow. The document does not describe an AI/ML powered device, and therefore does not contain information on acceptance criteria for AI performance metrics (like sensitivity, specificity, AUC), nor does it have a study comparing human readers with and without AI assistance (MRMC study).
Instead, the document focuses on the safety and performance of a medical device through biocompatibility, shelf-life, and performance bench testing, along with a review of existing clinical literature to support its intended use.
Here's a breakdown of the requested information based on the provided text, adapted to the nature of the device:
1. A table of acceptance criteria and the reported device performance
The document provides extensive tables for biocompatibility and shelf-life testing, and a list of performance testing - bench. Since there are many individual tests within these categories, I will present a summary table derived from them. Each of these tests had specific acceptance criteria and the document explicitly states that the device "passed" or "meets requirements" for all of them.
| Category | Acceptance Criteria (Summary) | Reported Device Performance (Summary) |
|---|---|---|
| Biocompatibility | - Cytotoxicity: Meets ISO 10993-1 requirements (Non-Cytotoxic). - Sensitization: Meets ISO 10993-1 requirements (Non-Sensitizer). - Intracutaneous Reactivity: Meets ISO 10993-1 requirements (Non-Irritant). - Acute Systemic Toxicity: Meets ISO 10993-1 requirements (No Acute Systemic Toxicity, Biocompatible Materials). | All tests passed: Non-Cytotoxic, Non-Sensitizer, Non-Irritant, No Acute Systemic Toxicity, Biocompatible Materials. |
| Shelf-Life/Sterility | - Functional and Package Integrity Testing: Maintain secure connections (e.g., MSS assembly, clamp, drip chamber, junctions) under specified torque/load. - Vent Integrity: Withstand 150 mm Hg fluid pressure; allow fluid withdrawal without compromising mechanical integrity; allow drainage of blood/Ringer's solution; provide CSF flow with minimal resistance. - Leakage: Bonds to withstand 200 mmHg air with <= 0.5cc/min leakage; no leak at 1 psi for drainage bag seal weld; drainage bag to withstand inversion with 700mL fluid with <= 2mL leakage after 1 minute. - Flow Initiation: Initiate flow by 15 cm water. - Sterility: Sterility Assurance Level (SAL) of 10⁻⁶; Endotoxin limit of 2.15 EU/device per USP <161>; non-pyrogenic. - Packaging: Secure, undamaged, and meet specifications after simulated shipping. | All shelf-life testing, including functional and package integrity, passed acceptance criteria after 3 years and one month of real-time aging and EO sterilization. Device achieved SAL of 10⁻⁶, met endotoxin limits, and was non-pyrogenic. Packaging integrity was maintained. |
| Performance Testing | - Dimensional measurements within specifications. - Mechanical Strength: MSS assembly, clamp, cord, drip chamber attachment, junctions, bottom cap bond to withstand specified forces/torques. - Vent Functionality: Ensure proper fluid pressure withstand, blood solution drainage, and minimal CSF flow resistance. - Leakage: UV-cure bonds and drainage bag to show no leakage under specified conditions. - Flow Initiation Pressure within limits. - Drip Chamber: Verify fluid weight and proper volume. - Hydrophobic Microbial Barrier Vent: Demonstrate 99.9% Bacterial Filtration Efficiency (BFE). | All "Performance Testing - Bench" 항목 had specific test methods and the document states the DUET™ EDMS "was tested and passed" all of them. Specific quantitative results are not provided for each test in the summary. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
-
Bench Testing (Biocompatibility, Shelf-Life, Performance): The sample sizes for each specific test (e.g., number of devices, number of material samples) are not explicitly stated in the provided text. The data provenance is internal testing conducted by Medtronic (the manufacturer).
-
Clinical Literature:
- "Coselli et al." (2002): 145 subjects. Prospective, randomized study.
- "Estrera et al." (2001): 148 subjects. Retrospective study.
- "Estrera et al." (2005): 238 subjects in adjunct group, 62 subjects in "other" groups (12 CSF drainage alone, 34 DAP alone, 16 neither). Retrospective study.
- "Safi et al." (2003): 1004 subjects (741 in adjunct group, 263 in non-adjunct group). Retrospective study.
- "Svensson et al." (1998): 17 subjects in treatment group, 16 subjects in control group. Randomized, prospective study.
The countries of origin for these clinical studies are not explicitly stated, but are likely in countries with advanced surgical capabilities (e.g., USA, Europe), based on the typical publication venues for such medical research.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This question is not applicable in the context of this device and study. The "ground truth" for the bench testing is established by engineering standards and measurement accuracy, not by human expert assessment. For the clinical literature review, the "ground truth" refers to patient outcomes and diagnoses reported by medical professionals in those studies, determined by standard clinical practice in those contexts. The number and qualifications of experts involved in establishing those clinical outcomes are not detailed in this document.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. The bench tests are objective measurements against defined standards, not subjective assessments requiring adjudication. The clinical studies cited were clinical trials or retrospective reviews, not studies involving adjudication of a test set for this specific device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device, and no MRMC study was performed or is relevant to its evaluation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML device. The device's performance is standalone in the sense that it functions mechanically and physically as designed, but it always requires a human in the loop for its operation, monitoring, and interpretation of its output (e.g., CSF pressure readings).
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Bench Testing (Biocompatibility, Shelf-Life, Performance): The "ground truth" is based on pre-defined engineering specifications, international/national standards (e.g., ISO, ASTM, USP), and objective measurements (e.g., torque, load, pressure, leakage volume, dimensional measurements).
- Clinical Literature: The "ground truth" for the effectiveness section is patient outcomes data (e.g., incidence of spinal cord injury, neurological deficit rates) collected in the cited clinical studies, as diagnosed and managed by the clinical teams in those studies.
8. The sample size for the training set
Not applicable. This is not an AI/ML device, so there is no "training set."
9. How the ground truth for the training set was established
Not applicable. There is no training set for this device.
{0}------------------------------------------------
DE NOVO CLASSIFICATION REQUEST FOR MEDTRONIC NEUROSURGERY DUET™ EXTERNAL DRAINAGE AND MONITORING SYSTEM (EDMS)
REGULATORY INFORMATION
FDA identifies this generic type of device as:
Cerebrospinal fluid shunt system. A cerebrospinal fluid shunt system is a prescription device used to monitor and divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of preventing spinal cord ischemia or injury during procedures that require reduction in central nervous system pressure. A cerebrospinal fluid shunt system may include catheters, valved catheters, valves, connectors, and pressure monitors intended to facilitate use of the shunt or evaluation of a patient with a shunt.
NEW REGULATION NUMBER: 21 CFR 882.5560
CLASSIFICATION: Class II
PRODUCT CODE: PCB
BACKGROUND
DEVICE NAME: Medtronic Neurosurgery DUETTM External Drainage and Monitoring System (EDMS)
DE NOVO REQUEST: DEN120017
DATE OF DE NOVO REQUEST: December 21, 2012
Medtronic Neurosurgery REQUESTER CONTACT: 125 Cremona Drive Goleta, CA 93117
INDICATIONS FOR USE
The DUET™ EDMS is indicated for temporary draining and monitoring of cerebrospinal fluid (CSF) flow from the lumbar subarachnoid space in:
-
- Patients undergoing open descending thoracic aortic aneurysm (open TAA) or open descending thoraco-abdominal aortic aneurysm (open TAAA) repair surgery.
-
- Patients post TAA/TAAA repair that become symptomatic with neurological deficit such as paraplegia.
{1}------------------------------------------------
LIMITATIONS
For prescription use only.
The use of a DUET™ EDMS lumbar catheter for drainage and monitoring of cerebrospinal fluid from the lumbar subarachnoid space is contraindicated in a patient with:
- . non-communicating hydrocephalus;
- large intracranial mass lesions, tumors, hematomas, or cysts;
- infections in the area surrounding the lumbar puncture which includes the skin. o subcutaneous tissue, bone and the epidural space; or
- demonstrated blockage of cerebrospinal fluid to the subarachnoid space due to 0 trauma, hematoma, fracture or tumor.
The use of a DUET™ EDMS requires 24-hour-a-day availability of trained personnel to supervise monitoring and drainage.
Literature suggests a maximum CSF drainage duration of 3 days for aneurysm repair patients who do not exhibit symptoms of neurological deficit, with longer durations for those exhibiting symptoms; however, drainage duration should be at the medical discretion of the physician and based on the institution's protocol.
Warnings & Precautions:
- It is possible that the puncture of the ventricle or the opening of the dura will ● result in an intracranial hemorrhage.
- o It is possible that if too much CSF is removed from the ventricles, either during a drainage procedure or when the ventricle is first punctured, the ventricle may collapse and occlude the catheter.
- . It is possible that the monitoring system may give a false pressure reading either due to a pressure line becoming clogged or kinked or from an air bubble lodged in the system. An incorrect pressure reading may lead to the wrong therapy being given to the patient. The irrigation of the catheter or the performance of a Volume Pressure Relationship (VPR) study may induce pressure waves in the patient. For this reason, irrigation or VPR studies should be done only by, or on the order of, a physician.
- o In order to minimize the possibility of infection, meningitis or ventriculitis, several steps should be observed. First, the injection sites should always be cleaned with alcohol and the alcohol allowed to dry before a needle is inserted into them. Second, sterile technique should be observed in setting up the system and in the placement of the catheter. Third, subgaleal tunneling of the ventricular catheter should be approximately one to two inches.
- . Leakage from the system, which can result from damaged system components or improper use of handling, can potentially result in over-drainage, the need to replace the drainage system and/or other complications to the patient.
- o In order to ensure against ventricular collapse and the possible consequences of tentorial herniation, always perform a drainage maneuver against a positive
{2}------------------------------------------------
pressure head on the order of 20 cm H2O or 15 mmHg. In addition, when the ventricle or lumbar subarachnoid space is first punctured during the insertion of the catheter, care should be taken so as little CSF as possible is lost.
- o Whenever irrigation of the catheter or the performance of the VPR is decided upon, great care must be used so that pressure waves are not initiated. Only a small volume of saline should ever be injected into the ventricular system, and this only done by, or on the order of, a physician. In general, in monitoring intracranial pressure, one should always be aware of the waveform on the oscilloscope. If the waveform begins to dampen out. it is important that the entire monitoring system be examined. Ensure that the line to the patient is not kinked and that all air bubbles or blood or other debris are removed from the system. Ascertain that the transducer is on the same level as the patient's ventricular system to ensure the proper reference level in the manometer tube for use in calibration procedures. Pressure monitoring with the manometer may result in over-drainage of the ventricles.
- Improper vigilance or improper drainage system setup can lead to over- or under-. drainage and potentially serious injury to the patient. Intracranial and lumbar pressure monitoring has been associated with intracranial infection, meningitis and ventriculitis. This hazard has been quoted at less than 1% to more than 5%. The risk of infection is probably influenced both by the number of times a system is opened and by the duration of the monitoring. Prolonged steroid therapy can also increase the risk of infection.
DEVICE DESCRIPTION
The Medtronic DUET™ External Drainage and Monitoring System (EDMS) that is the subject of this De Novo request is designed to drain and monitor cerebrospinal fluid (CSF) from the lumbar subarachnoid space.
The DUET™ EDMS consists of the following: a green-striped patient connection line (pressure tubing) with an inner diameter of 0.075 ± 0.005 inches, an outer diameter of 0.124 ± 0.003 inches, and a total length of 60 inches (9), a patient line stopcock (10), a main system stopcock (8) that may be attached at two locations on the main panel, a drip chamber (4) with a drip chamber stopcock (5), a rotatable pressure scale (3), three latex-free needleless injection/CSF sampling sites (Figure 1b (IS-5), (IS-10) and (IS-11)) and a removable drainage bag (7) with approximate volumetric graduations and a hydrophobic microbial barrier air vent. There is a pole mount clamp (6) and a cord (12) with a cord lock (13) to enable independent suspension of the system, or to provide additional security when using the pole clamp as identified in Figure 1 below. It should be noted that the numbers in parentheses correspond with the numbers in Figure 1.
{3}------------------------------------------------
Image /page/3/Figure/0 description: The image shows two medical devices labeled 1a and 1b. Both devices have a clear plastic container with volume measurements on the side. The device in 1a has labels pointing to different parts, including a dial (1), a measurement scale (3), a valve (5), a collection bag (8), and a tube (9).
Figure 1: Medtronic DUET EDMS (a) and Location of Injection/CSF Sampling Sites (b)
The DUET™ EDMS are not long-term implants but are intended for limited external drainage of CSF. The drainage flow of CSF into the DUETTM EDMS is uni-directional and gravity-driven; there is no recirculation of the CSF. During use, an external lumbar catheter inserted into the lumbar subarachnoid space is connected to the DUET™ EDMS patient connection line. The CSF drains through the catheter, into the patient connection line and into the graduated drip chamber. CSF collects in the drip chamber, exits the bottom of the chamber via another connection line and is collected in a drainage bag. In the event that the patient may require administration of fluid directly into the lumbar subarachnoid space or CSF sampling is required, the DUET™ EDMS features injection/CSF sampling ports integrated into the patient connection line. The DUET™ EDMS is completely disposable. The DUET™ EDMS is recommended for use with the Clear-Site™ Laser Level (cleared under K984053) that is provided separately.
SUMMARY OF NONCLINICAL/BENCH STUDIES
The non-clinical/bench studies conducted on the DUET™ EDMS to demonstrate the safety and effectiveness of the device are summarized in the sections below. The technological characteristics of the DUET™ EDMS for the subject De Novo is identical to the system cleared in K984053.
{4}------------------------------------------------
BIOCOMPATIBILITY/MATERIALS
The components of the patient lines in the DUET™ EDMS were tested for biocompatibility because these patient lines have the potential for contact with CSF that could be re-introduced to the patient in the case of retrograde flow. The patient lines are classified as external communicating devices of limited contact duration (< 24 hours). The biocompatibility tests conducted on the patient lines of the DUET™ EDMS are shown in Table 1.
| Test | Purpose | Acceptance Criteria | Results |
|---|---|---|---|
| Cytotoxicity (MEMElution) | Determine the lysis of cells(cell death), the inhibitionof cell growth, and othereffects on cells caused bythe device, materials and/orother extracts. | Meets requirements inISO 10993-1:2009/(R)2013 | Non-Cytotoxic |
| Sensitization (Saline,Sesame Oil (SO)) | Estimate the potential forcontact sensitization of thedevice, materials and/orother extracts. | Meets requirements inISO 10993-1:2009/(R)2013 | Non-Sensitizer |
| IntracutaneousReactivity (Saline,SO) | Evaluate the local dermalirritant or toxic effects ofleachables extracted fromthe test article followingintracutaneous injection inrabbits. | Meets requirements inISO 10993-1:2009/(R)2013 | Non-Irritant |
| Acute SystemicToxicity | Estimate the potentialharmful effects of eithersingle or multipleexposures, during a periodof less than 24 hours, todevices, materials and/orextracts. | Meets requirements inISO 10993-1:2009/(R)2013 | No Acute SystemicToxicity,BiocompatibleMaterials |
TABLE 1: DUET EDMS™ PATIENT LINE BIOCOMPATIBILITY TESTING
SHELF LIFE/STERILITY
The DUET™ EDMS is labeled with a 3 year shelf-life. Table 2 contains a summary of the shelf-life testing conducted on the DUET™ EDMS and its packaging after 3 years and one month of real-time aging and ethylene oxide sterilization to validate the 3 year expiration date. All shelf-life testing, which included both functional and package integrity testing, passed the corresponding acceptance criteria.
{5}------------------------------------------------
TABLE 2: DUET™ EDMS SHELF-LIFE TESTING
| DUET System |
|---|
| Main System Stopcock (MSS) assembly to attachment arm:MSS assembly must not be displaced or dislodged with 5 in-lbs torqueapplied to arm of stopcock. |
| Main System Stopcock (MSS) assembly to attachment arm: MSSassembly must not be displaced or dislodged with a 5 lb load applieddownward onto the core of the stopcock arm. |
| Clamp to maintain secure attachment of device to 0.75"-1.25" diameterpole with 5 pound load in the downward axial direction for a minimum of 1minute with 3 in-lbs torque applied to the pole clamp thumb screw. |
| Drip chamber/bag subassembly to panel: Subassembly to remainstationary with 3.5 pound load applied in the axial direction with 3 in-lbtorque applied to thumbscrew used to secure subassembly to panel. |
| Attachment of junctions - tubes to luers: Must withstand minimum 5 poundload in the axial direction. |
| Attachment of junctions - stopcock to drip assembly: Must withstandminimum 5 pound load in the axial direction. |
| Stopcock/bottom cap junction: The Stopcock/Bottom cap Bond shallwithstand 3.1 in-lbs torque when tested per TM-488-QA. |
| Drip assembly and drainage bag vent integrity: Vents must withstand 150mm Hg fluid pressure. |
| Drip assembly vent integrity: With Main System Stopcock closed relativeto the drip assembly, vent must allow withdrawal of 30 mL distilled waterfrom the bottom of the drip chamber in 15 seconds without compromisingthe mechanical integrity of the vent. |
| Leakage of UV-cure bonds (Patient line & Drip Chamber subassembly):bonds to withstand 200 mmHg of air with no more than 0.5cc/min ofleakage. |
| Cord to I.V. pole:Cord and cord lock to maintain secure hanging of system with 5 poundload in the downward axial direction for a minimum of 1 minute. |
| Drip assembly and drainage bag vent integrity: Vents must withstand 150mm Hg fluid pressure.Tensile strength of drainage bag inlet port: 10 lbs. minimum in axialdirection. |
| Disposable drainage bag seal weld: No leak at 1 psi. |
| Flow initiation pressure of drainage bag (includes anti-reflux valve): unitsmust initiate flow by 15 cm water. |
| Leakage of drainage bag: bag must withstand being inverted with 700 mLof fluid with no more than 2 mL of leakage after one minute. |
| Drip assembly vent: with vent wet with a mixture of 10% blood and 90%lactated Ringer's solution for 30 minutes at 50 mmHg fluid pressure, ventmust: |
| • allow drainage of 30 mL of blood and Ringer's solution mixture within 3minutes• provide CSF flow through system of no more resistance than +2.2 cmwater at an inflow rate of 20 mL/hour after drainage of blood and |
| Ringer's solution mixture |
| PackagingShipping - Product packaging and contained product must be secure, |
| undamaged, and must meet all specifications after simulated shippingconditions per ISTA-2A and simulated shelf life conditions.(Dye Penetration Test and Bubble Leak) |
| Shipping - Product inspection method: visual |
{6}------------------------------------------------
The package integrity testing for the DUET EDMS conforms to the following FDA recognized consensus standards:
- . AAMI ANSI ISO 11607-1:2006/(R)2010 – Packaging for Terminally Sterilized Medical Devices – Part 1: Requirements for Materials, Sterile Barrier Systems, and Packaging Systems
- AAMI ANSI ISO 11607-2:2006/(R)2010 Packaging for Terminally Sterilized . Medical Devices - Part 2: Validation Requirements for Forming, Sealing and Sealing Processes Standards
- ASTM F1886 - Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- . ASTM F2096 – Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test)
- . ASTM F88/F88M-09 – Standard Test Method for Seal Strength of Flexible Barrier Material
The DUET™ EDMS is sterilized by 100% ethylene oxide (EO) with a validated Sterility Assurance Level (SAL) of 10℃. The device is indicated for single use only and disposable. The sterilization validation for the DUET™ EDMS is in compliance with ISO 11135-1:2007 and ISO 10993-7:2008 with no deviations. In addition, endotoxin testing using the limulus amebocyte lysate (LAL) method was conducted on the fluid path of the DUET™ EDMS to meet the endotoxin limit of 2.15 EU/device per USP <161>.
Material-mediated pyrogenicity testing was also conducted on the DUET™ EDMS to detect material-mediated pyrogenic reactions of extracts of the device and/or materials. The testing meets the requirements in ISO 10993-1: 2009/(R)2013. The results show that the extracts of the device and/or its materials are non-pyrogenic.
PERFORMANCE TESTING - BENCH
The DUET™ EDMS was tested and passed the following performance (bench) tests listed in Table 3.
| Test Name | Test Method |
|---|---|
| Dimensional | Measure the dimensions of the scale label vertical and horizontalalignment, pressure scale lengths, patient line tubing inner diameter(ID) and length, cord length, drainage path ID (from bottom of dripchamber), and stopcock flow path diameter. |
| Drip Chamber Graduations |
TABLE 3: DUET™ EDMS PERFORMANCE TESTS
{7}------------------------------------------------
| Main System Stopcock (MSS)Assembly Torque Applied to Arm ofStopcock | Record the peak torque at which the MSS assembly fails or detachesfrom the panel. |
|---|---|
| MSS Assembly Load Applied to Coreof Stopcock Arm | Record peak load at which MSS assembly failed. |
| Clamp to I.V. Pole AttachmentStrength | Ensure secure attachment of clamp to I.V. pole. |
| Cord to I.V. Pole Attachment Strength | Ensure cord and cord lock maintains secure hanging of system. |
| Drip Chamber to Back PanelAttachment Strength | Ensure secure attachment of drip chamber/bag subassembly to panel. |
| Strength of Attached Junctions (i.e.,Tubing to Luer) | Ensure secure attachment of junctions of tubes to luer. |
| Bottom Cap to Stopcock JunctionTorque | Ensure secure bond of stopcock/bottom cap. |
| Drip Assembly and Drainage BagVent Integrity | Ensure that drip assembly and drainage bag vent can withstandappropriate fluid pressures. |
| Tensile Strength of Drainage BagInlet Port | Evaluate the tensile strength of the drainage bag inlet port to failure. |
| Drainage Bag Seal Weld | Ensure there are no leaks in the drainage bag. |
| Flow Initiation Pressure | Record pressure at which flow initiates, for each drainage bag. |
| Drip Assembly Vent Test (Exposureof Vent to Blood Solution) | Ensure that the drip assembly vent allows drainage of blood andprovide CSF flow through system with minimal resistance. |
| Drip Assembly Vent Integrity | Test the drip assembly vent to withdraw fluid without compromising itsmechanical integrity. |
| Leakage of UV-Cure Bonds | Record any leakage from the UV-cure bonds between the patient lineand drip chamber subassembly. |
| Leakage of Drainage Bag | The drainage bag must withstand being inverted without leaking. |
| Drip Chamber Volume | Verify fluid weight in the drip chamber. |
| Attachment of I.V. Pole and Positionof Adjustable Drip Chamber | Visually verify that clamping thumbscrews (and cord locks) have notslipped from initial positions (using visual marks to identify anyslippage). |
| Leakage of UV Cure Bonds | The UV-cure bonds between the patient line and drip chambersubassembly should withstand air pressure without causing leaks. |
| Attachment of Junctions | Junctions must be able to withstand minimum of 5 pound load in theaxial direction. |
| Bottom Cap to Stopcock JunctionTorque | Test the torque of the stopcock/bottom cap bond. |
| Hydrophobic Microbial Barrier Venton the Drainage Bag | The supplier for the material used as the drainage bag vents conductedmicrobial barrier testing to demonstrate a 99.9% Bacterial FiltrationEfficiency (BFE) |
SUMMARY OF CLINICAL INFORMATION
Although there was no formal clinical study conducted using the DUET™ EDMS for the expanded indication of temporary draining and monitoring of CSF flow from the lumbar subarachnoid space in patients undergoing open descending TAA/TAAA repair surgeries and patients post TAA/TAAA repair that become symptomatic with neurological deficit such as paraplegia. FDA believes that there is sufficient clinical data to support the expanded indication for this device. Given the worldwide clinical experience, the data and information provided support use of the device for the expanded indication as long as it is performed with a clear understanding of the risks associated with the device and clinical procedure. In describing the apparatus set-up, the CSF drains used in the clinical literature had similar characteristics (i.e., gravitational based pole-mounted apparatus with a
{8}------------------------------------------------
drainage bag) and operating principles as the subject device. The following is a summary of the clinical literature used to support the De Novo Indications for Use for the DUET™ EDMS in general:
-
- Coselli et al. "Cerebrospinal Fluid Drainage Reduces Paraplegia after Thoracoabdominal Aortic Aneurysm Repair: Results of a Randomized Clinical Trial" (J. Vascular Surgery 2002; 35; p. 631-639). In this prospective, randomized study of 145 subjects, the effect of CSF drainage on the incidence of spinal cord injury (SCI) was evaluated in subjects undergoing surgical repair of Type I or II aortic aneurysms. In this study, additional concomitant methods were used for spinal cord protection; however, administration of CSF drainage was the only variable between the two treatment groups. During the procedure, CSF was drained freely with gravity if the CSF pressure exceeded 10 mmHg. The drain was removed two days post-operatively if the subject did not experience SCI, and was maintained bevond two days if SCI occurred. The authors reported a significant difference (p = 0.03) in the SCI rate, which was 2.7% in the CSF drainage group compared to 12.2% in the control group. This difference represents an 80% reduction in the occurrence of SCI and the authors concluded that CSF drainage is beneficial during the repair of Type I and II aortic aneurysms.
-
- Estrera et al. "Descending Thoracic Aortic Aneurysm: Surgical Approach and Treatment using the Adjuncts Cerebrospinal Fluid Drainage and Distal Aortic Perfusion" (Ann. Thorac. Surg. 2001; 72; p. 481-486). A retrospective study was conducted to evaluate the concomitant use of Distal Aortic Perfusion (DAP) and CSF drainage in the prevention of neurological deficit during 148 non-emergent repairs of descending thoracic aortic aneurysms. The authors reported an overall neurological deficit rate of 2.7%.
-
- Estrera et al. "Descending Thoracic Aortic Aneurysm Repair: 12-Year Experience using Distal Aortic Perfusion and Cerebrospinal Fluid Drainage" (Ann. Thorac. Surg. 2005; 80; p. 1290-1296). The authors presented the results of their 12-year experience comparing the rate of neurological deficit in 238 subjects who underwent aortic aneurysm repair in which DAP and CSF drainage were both administered (adjunct group) to 62 subjects with the use of CSF drainage alone (12 subjects). DAP alone (34 subjects). or neither adjunct was used (16 subjects). The results reported a neurological deficit rate of 1.3% for the adjunct group compared to 6.5% for the non-adjunct group (p < 0.02).
-
- Safi et al. "Distal Aortic Perfusion and Cerebrospinal Fluid Drainage for Thoracoabdominal and Descending Thoracic Aortic Repair: Ten Years of Organ Protection" (Ann. Surg. 2003: 238; p. 372-380). The authors retrospectively examined the long term results of the combined administration of DAP. CSF drainage, and moderate hypothermia (adjunct group) compared to subjects in which no adjuncts were used for repairs of descending thoracic and TAAA. A total of 1004 subjects were evaluated with 741 subjects in the adjunct group (73.8%) and 263 subjects in the non-adjunct group (26.2%). Within the adjunct group, intraoperative CSF pressure was maintained at 10 mmHg and the mean arterial pressure between 90-100 mmHz. The CSF drain was employed for 3 days postoperatively and was continued for an additional 72 hours if neurological complications occurred. Results from this retrospective study demonstrated a neurological deficit rate of 2.4% for the adjunct group compared to a rate of 6.8% in the non-adjunct group (p < 0.0009). Use of the adjuncts prevented neurological
{9}------------------------------------------------
deficit in 1 in 20 cases for all patients, and 1 in 5 for Type II TAAA. The authors concluded that the study results suggest a multimodal approach including CSF drainage intra- and postoperatively can protect the spinal cord and reduce the risk of neurological complications.
-
- Svensson et al. "Reduction of Neurologic Injury after High-Risk Thoracoabdominal Aortic Operation" (Ann. Thorac. Surg. 1998; 66; p. 132-138). The authors conducted a randomized, prospective study comparing the use of CSF drainage and Intrathecal papaverine (a vasodilator) in subjects undergoing high-risk TAAA repairs (treatment group) compared to no adjuncts (control group). Seventeen (17) subjects were enrolled in the treatment group and 16 subjects were enrolled in the control group. Within the treatment group, while the aorta was cross-clamped, CSF drained freely by gravity and once the clamp was removed, CSF drainage was discontinued unless CSF pressure exceeded 7 - 10 cm H2O. Neurological deficit occurred in 2 subjects (11.8%) in the treatment group compared to 7 subjects (43.8%) in the control group (p = 0.0392). The authors concluded that for high-risk TAAA repairs, the combined CSF drainage and Intrathecal papaverine approach significantly reduced the incidence and severity of neurologic injury and that active cooling may further reduce the risk. The Institutional Review Board (IRB) terminated the study (for ethical reasons) with one-third of the target number of enrolled subjects, and an interim analysis of safety and effectiveness demonstrated a significant difference between the treatment and control groups. Consequently, early termination reduced the power of the study (α = 0.1) and increased the possibility of statistical error. In conclusion, despite the reduced power of the study, CSF drainage and Intrathecal papaverine reduced the occurrence of SCI in high-risk patients.
The provided clinical literature demonstrates reasonable assurance of safety and effectiveness in the use of a lumbar drainage system with the characteristics of the DUET™ EDMS for temporary CSF drainage to prevent neurological deficit and reduce spinal cord perfusion pressure and ischemia in patients undergoing open TAA/TAAA repair surgeries.
- Svensson et al. "Reduction of Neurologic Injury after High-Risk Thoracoabdominal Aortic Operation" (Ann. Thorac. Surg. 1998; 66; p. 132-138). The authors conducted a randomized, prospective study comparing the use of CSF drainage and Intrathecal papaverine (a vasodilator) in subjects undergoing high-risk TAAA repairs (treatment group) compared to no adjuncts (control group). Seventeen (17) subjects were enrolled in the treatment group and 16 subjects were enrolled in the control group. Within the treatment group, while the aorta was cross-clamped, CSF drained freely by gravity and once the clamp was removed, CSF drainage was discontinued unless CSF pressure exceeded 7 - 10 cm H2O. Neurological deficit occurred in 2 subjects (11.8%) in the treatment group compared to 7 subjects (43.8%) in the control group (p = 0.0392). The authors concluded that for high-risk TAAA repairs, the combined CSF drainage and Intrathecal papaverine approach significantly reduced the incidence and severity of neurologic injury and that active cooling may further reduce the risk. The Institutional Review Board (IRB) terminated the study (for ethical reasons) with one-third of the target number of enrolled subjects, and an interim analysis of safety and effectiveness demonstrated a significant difference between the treatment and control groups. Consequently, early termination reduced the power of the study (α = 0.1) and increased the possibility of statistical error. In conclusion, despite the reduced power of the study, CSF drainage and Intrathecal papaverine reduced the occurrence of SCI in high-risk patients.
LABELING
The Instructions for Use for the DUET™ EDMS (for the De Novo indications) are consistent with the clinical data and cover all the hazards and other clinically relevant information that may impact use of the device. The labeling for the DUET™ EDMS is sufficient and satisfies the requirements of 21 CFR § 801.109 Prescription Devices. The labeling for the DUET™ EDMS includes:
-
- The indicated use population and environment.
-
- Contraindications with respect to patients who should not receive a lumbar drain.
-
- Requirement with respect to 24-hour-a-day availability of trained personnel to supervise monitoring and drainage.
-
- Detailed instructions on proper device set-up, positioning and monitoring.
-
- Appropriate warnings and precautions to inform the user of the potential serious hazards and special care associated with the use of the device.
-
- A detailed summary of the device- and procedure-related complications or adverse events.
-
- Instructions which state that the device is not to be reused, reprocessed, or resterilized when open but unused.
-
- Cleaning instructions for the injection sites.
{10}------------------------------------------------
-
- Recommendation for use with the Clear-Site™ Laser Level (cleared under K984053) that is provided separately.
RISKS TO HEALTH
Table 4 below identifies the risks to health that may be associated with use of a cerebrospinal fluid shunt system and the measures necessary to mitigate these risks.
| Identified Risk | Mitigation Measure |
|---|---|
| Pyrogenicity/adverse tissue reaction | Biocompatibility evaluationPyrogenicity testingLabelingShelf life testingSterilization validation |
| Infection (including meningitis) | LabelingSterilization validationPackage integrity testing |
| Cerebrospinal fluid (CSF) leakage | LabelingNon-clinical performance testing |
| Over- & under-drainage• Spinal headache with and without CSF leakage• Intracranial hemorrhage• Hematoma (e.g., spinal, subdural)• Paraplegia• Foreign body obstruction | LabelingNon-clinical performance testing |
| Procedural/use errors | Labeling |
Table 4: Identified Risks to Health and Mitigation Measures
SPECIAL CONTROLS:
In combination with the general controls of the FD&C Act, the cerebrospinal fluid shunt system is subject to the following special controls:
-
- The device description must include a detailed summary of the device technical parameters, including design configuration, dimensions, engineering drawings, and a list of all components with identification of their materials of construction.
-
- The patient-contacting components of the device must be demonstrated to be biocompatible.
-
- Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
- a. Simulated use testing must be conducted to characterize fluid flow and resistance to leakage.
- b. Mechanical integrity testing of all connections must be conducted.
{11}------------------------------------------------
-
- Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the specified shelf life.
-
- Performance data must demonstrate the sterility and pyrogenicity of patient-contacting components of the device.
-
- The labeling must include:
- a. Contraindications with respect to patients who should not receive a lumbar drain;
- b. A warning that the device should have 24-hour-a-day availability of trained personnel to supervise monitoring and drainage;
- Instructions on proper device set-up, positioning and monitoring; c.
- d. Warnings and precautions to inform the user of serious hazards and special care associated with the use of the device;
- e. A statement that the device is not to be reused, reprocessed, or resterilized when open but unused; and
- f. Cleaning instructions for the injection sites.
BENEFIT/RISK DETERMINATION
The risks of the device are based on data as evidenced in the clinical literature. The risks associated with lumbar CSF drainage using the subject device include excessive drainage of CSF that can lead to cranial subdural hematoma, infection/meningitis, CSF leakage with or without spinal headache, and drainage fluid that is not clear or not flowing indicating problems with coagulopathy or a malfunction of the drain (e.g., catheter fracture or occlusion). In the clinical literature titled, "Complications of Spinal Fluid Drainage in Thoracoabdominal Aortic Aneurysm Repair: A Report of 486 Patients Treated from 1987 to 2008" (Wynn et al. Journal of Vascular Surgery 2009. 49; p. 29-34), the results revealed a 5% rate of bloody spinal fluid and 2.9% rate of intracranial blood without neurological deficit. In addition, it was reported that neurological deficits and mortality due to CSF drainage occurred in 1% and 0.6% of patients, respectively.
The probable benefits of the device are also based on data collected in the clinical literature as described above. The benefit of the device includes the reduction of paraplegia after open thoracic aortic aneurysm (TAA) or open thoraco-abdominal aortic aneurysm (TAAA) repair surgery with CSF drainage. In the published clinical literature from Coselli et al., a prospective, randomized study of 145 subjects determined that the risk of spinal cord injury (SCI) was 12.2% in the control group compared to 2.7% in the lumbar CSF drainage group for the surgical repair of open Type I/II aortic aneurysms. In patients who achieve a probable benefit from using this device, the effect would be the prevention of SCI.
Additional factors to be considered in determining probable risks and benefits for the DUET™ EDMS include: The effectiveness data were primarily based on one randomized prospective study of 145 subjects in the Coselli et al. publication. There were several other retrospective studies published in the clinical literature including a retrospective study from Estrera et al. who used distal aortic perfusion (DAP) concurrent with CSF drainage in the prevention of neurological deficit during 148 non-emergent repairs of descending thoracic aortic aneurysms.
{12}------------------------------------------------
The results showed that the rate of neurological deficit in 238 subjects was 1.3% for the adjunct group in which DAP was used concurrently with CSF drainage compared to 6.5% for the control group (i.e., only CSF drainage, only DAP, and neither CSF drainage or DAP). In a separate retrospective study from Safi et al., the long term results of the combined administration of DAP, CSF drainage, and moderate hypothermia (adjunct group) were compared to subjects in which no adjuncts were used for repairs of descending thoracic and thoracoabdominal aortic aneurysms. A total of 1004 subjects were evaluated, and the results showed a neurological deficit rate of 2.4% for the adjunct group compared to a rate of 6.8% in the non-adjunct group. Svensson et al. conducted a randomized prospective study comparing the use of CSF drainage and intrathecal papaverine (a vasodilator) in subjects undergoing high-risk TAAA repairs (17 subjects) in comparison with no adjuncts (16 subjects). The results demonstrated that neurological deficit occurred in 11.8% of subjects in the treatment group compared to 43.8% in the control group. Based on the data in the clinical literature, the benefit of using the subject device is most effective in patients undergoing open TAA and open TAAA repairs. Because SCIs can be difficult to treat, are chronic conditions, and there are few good interventions to help these patients, the use of CSF drainage has the potential to lower the risk of SCI and improve the quality of life in patients undergoing open TAA and open TAAA repair surgeries. Currently, CSF drainage is the most common treatment for open TAA and open TAAA repair surgeries and involves the off-label use of lumbar drain systems. Intrathecal agents may be an alternative to CSF drainage.
In conclusion, given the available information, the data support that the probable benefits outweigh the probable risks of temporary draining and monitoring CSF flow from the lumbar subarachnoid space in patients undergoing open descending TAA or open descending TAAA repair surgery or in post TAA/TAAA patients who are symptomatic for neurological deficit such as paraplegia. The Medtronic Neurosurgery DUET™ EDMS, as a lumbar drainage system, provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
CONCLUSION
The De Novo request for the Medtronic Neurosurgery DUET™ External Drainage and Monitoring System (EDMS) is granted and the device is classified under the following:
Product Code: PCB Device Type: Cerebrospinal fluid shunt system Class: II Regulation: 21 CFR 882.5560
§ 882.5560 Cerebrospinal fluid shunt system.
(a)
Identification. A cerebrospinal fluid shunt system is a prescription device used to monitor and divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of preventing spinal cord ischemia or injury during procedures that require reduction in central nervous system pressure. A cerebrospinal fluid shunt system may include catheters, valved catheters, valves, connectors, and pressure monitors intended to facilitate use of the shunt or evaluation of a patient with a shunt.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The device description must include a detailed summary of the device technical parameters, including design configuration, dimensions, engineering drawings, and a list of all components with identification of their materials of construction.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Simulated use testing must be conducted to characterize fluid flow and resistance to leakage; and
(ii) Mechanical integrity testing of all connections must be conducted.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the specified shelf life.
(5) Performance data must demonstrate the sterility and pyrogenicity of patient-contacting components of the device.
(6) The labeling must include:
(i) Contraindications with respect to patients who should not receive a lumbar drain;
(ii) A warning that the device should have 24-hour-a-day availability of trained personnel to supervise monitoring and drainage;
(iii) Instructions on proper device setup, positioning, and monitoring;
(iv) Warnings and precautions to inform the user of serious hazards and special care associated with the use of the device;
(v) A statement that the device is not to be reused, reprocessed, or resterilized when open but unused; and
(vi) Cleaning instructions for the injection sites.