K Number
K994390

Validate with FDA (Live)

Manufacturer
Date Cleared
2000-10-26

(303 days)

Product Code
Regulation Number
876.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

POLYFLUX R is indicated for use in hemodialysis for the treatment of chronic or acute renal failure. The choice of the filter is the responsibility of the physician. Special attention must be paid in connection with pediatric use. CAUTION! If POLYFLUX R is reused, the procedure and disinfection specified in the RENATRON INSTRUCTION MANUAL must be followed. The POLYFLUX R may be reprocessed for reuse on the same patient.

Device Description

The Gambro Polyflux 17R and 21R, Capillary Dialyzers/Filters labeled for multiple use (reuse) are identical in construction in function to Gambro Polyflux 17S and 21S Hemodialyzers / Hemofilters labeled for single use which are currently marketed in the United States and have been previously cleared by the FDA under 510(k) Notification K981414. Only the product designation "R" has been changed from "S" to designate that they are labeled for multiple use (reuse). These devices are intended for use in hemodialysis for the treatment of acute and chronic renal failure and for certain types of intoxications for both single when reprocessed for reuse for a maximum of 15 reprocessing reuse cycles on the same patient. If reprocessing and reuse is practiced, it is recommended that the reuse be done under the conditions as existed in the in vitro and confirmatory clinical studies undertaken by Gambro and presented in the labeling for this device. It may also be used in cases of acute fluid overload for the removal of plasma water. The membrane used in these devices is polyethersulfone (PES) which is identical to the membrane utilized in the Gambro Polyflux 17S and 21S Hemodialyzers / Hemofitters for labeled for single use which have been previously approved for marketing in the United States under a 510K Notification (K982414). Blood enters a blood inlet port where it is distributed to polyethersulfone hollow fibers. Each hollow fiber has an inner diameter of approximately 215 microns (wet hollow fiber internal diameter) and a wall thickness of 50 microns. The number of polyethersulfone hollow fibers in each hemodialyzer / hemofilter is 10,000 for the Polyflux 17R and 12,500 for the 21R. These dialyzers have an effective membrane length of 250 mm. The effective membrane surface area is 1.7 square meters for the 17R and 2.1 square meters for the 21R. The housing and end caps of these hemodialyzers / hemofiters are made of polycarbonate. The fibers used in the Gambro Polyflux 17R and 21R are identical in design and materials to the previously approved Gambro Polyflux 17S and 21S Hemodialyzers / Hemofilters for labeled for single use (K982414). The patient's blood traverses the inside of the hollow fibers and exits the device via a blood exit port. By means of a hydrostatic pressure or transmembrane pressure which is created by a combination of positive and negative pressures across the polyethersulfone membrane, plasma water along with certain lower molecular weight solutes of plasma water pass through the membrane and into the dialysate or filtrate compartment of the device. Removal of uremic toxins and waste products are removed from the patient's blood in this device by means of both diffusion and convection through polyethersulfone membrane and into the counter current flowing dialysis solution during hemodialysis. The dialysate exits the devices via a dialysate outlet port.

AI/ML Overview

This document describes the 510(k) Notification for the Gambro Polyflux 17R & 21R Capillary Dialyzers/Filters, labeled for multiple use. This submission seeks clearance for the multiple-use labeling of these devices, which are identical in construction and function to the previously cleared single-use Gambro Polyflux 17S & 21S Hemodialyzers/Hemofilters (K981414).

1. Table of Acceptance Criteria and Reported Device Performance:

The document explicitly states that the proposed devices (Polyflux 17R & 21R for multiple use) are identical in construction and function to the predicate devices (Polyflux 17S & 21S for single use), with the only difference being the labeling for multiple use. Therefore, the acceptance criteria are implicitly linked to demonstrating that the reprocessing for reuse does not compromise the device's original performance, safety, and effectiveness.

Acceptance Criteria (Implicit from FDA Guidance for Hemodialyzer Reuse Labeling)Reported Device Performance (Summary from Submission)
Device performance after reprocessing for reuse remains equivalent to its performance as a single-use device or to existing products in commercial distribution."Testing performed on the Gambro Polyflux 17R and 21R Capillary Dialyzers/Filters indicates that they are safe, effective, and perform as well as the predicate devices, when used in accordance with the instructions for use." "With respect to performance, these hemodialyzers/filters perform in a manner substantially equivalent to existing products in commercial distribution in the United States." This implies that the performance after reuse cycles met the expected performance standards. The submission also states: "The POLYFLUX R may be reprocessed for reuse on the same patient... If reprocessing and reuse is practiced, it is recommended that the reuse be done under the conditions as existed in the in vitro and confirmatory clinical studies undertaken by Gambro and presented in the labeling for this device." This suggests that specific performance metrics (e.g., clearance rates, ultrafiltration rates, fiber integrity) were maintained through up to 15 reprocessing reuse cycles.
Safety of the device is maintained after reprocessing for reuse, including biocompatibility and integrity of materials."Testing performed on the Gambro Polyflux 17R and 21R Capillary Dialyzers/Filters indicates that they are safe..." This implies that tests related to material integrity, absence of harmful leachates, and potential for contamination after reprocessing were successfully conducted.
Effectiveness of the device is maintained after reprocessing for reuse for its intended use (hemodialysis for treatment of renal failure and certain intoxications)."Testing performed on the Gambro Polyflux 17R and 21R Capillary Dialyzers/Filters indicates that they are... effective..." This implies that the device continues to adequately remove uremic toxins and waste products via diffusion and convection after reprocessing.
Reprocessing instructions are clear, validated, and result in safe and effective reuse."In vitro and in vivo performance data and directions for reuse have been included in the labeling." "CAUTION! If POLYFLUX R is reused, the procedure and disinfection specified in the RENATRON INSTRUCTION MANUAL must be followed." This indicates that a validated reprocessing procedure was established and is to be followed to ensure the device continues to meet the safety and effectiveness criteria.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Sample Size for Test Set: The document simply states that "Clinical data was collected according to the FDA Guidance for Hemodialyzer Reuse Labeling," and refers to "confirmatory clinical studies undertaken by Gambro." However, no specific sample sizes (number of devices or patients) for the clinical test set are provided within the excerpt.
  • Data Provenance: The country of origin of the data is not explicitly stated. The nature of the studies ("confirmatory clinical studies undertaken by Gambro") suggests they were prospective studies, as they were conducted to support the multiple-use claim.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

This information is not provided in the given excerpts. For medical devices like dialyzers, ground truth often relates to measurable clinical and laboratory parameters (e.g., solute clearance, ultrafiltration rates, patient outcomes), rather than interpretation by a panel of experts in the same way an imaging study would.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not applicable/provided for this type of device and study. Adjudication methods like 2+1 or 3+1 are typically used in imaging studies where subjective interpretation is involved. For a hemodialyzer, the performance is assessed through objective measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This excerpt describes a 510(k) submission for a medical device (hemodialyzer) for multiple use, not an AI or imaging diagnostic device. Therefore, an MRMC study or assessment of human reader improvement with AI assistance is entirely irrelevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a hardware medical device, not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

Based on the nature of the device (hemodialyzer), the ground truth for performance and safety would have been established through a combination of:

  • In vitro performance criteria: Objective measurements of device function (e.g., solvent removal, solute clearance, fiber integrity, pressure drop, biocompatibility measures like membrane integrity after reprocessing challenges).
  • Clinical outcomes data: Patient safety data and clinical effectiveness parameters (e.g., maintaining patient electrolyte balance, fluid removal, absence of adverse events related to reprocessing) over the course of multiple reuse cycles. The document refers to "in vitro and confirmatory clinical studies."

8. The sample size for the training set:

Not applicable. As this is a medical device, not an AI algorithm, there is no "training set" in the computational learning sense. The "training" for the device's design and manufacturing would come from extensive engineering design, material science, and manufacturing process validation, which are not detailed in this excerpt.

9. How the ground truth for the training set was established:

Not applicable. As there is no "training set" for an AI algorithm, this question does not apply. The "ground truth" for the device's initial design and subsequent multiple-use validation is based on established engineering principles, regulatory standards (like the FDA Guidance for Hemodialyzer Reuse Labeling), and clinical guidelines for hemodialysis treatment.

{0}------------------------------------------------

OCT 26 2000510(k) Notification Supplement Gambro Polyflux 17R, & 21 R Labeled for Multiple Use July 17th, 2000 510K(k) SUMMARY
----------------------------------------------------------------------------------------------------------------------------------------------------------------------
SUBMITTER:Gambro Healthcare 1185 Oak Street Lakewood, CO 80215 (303) 231-4436
DATE PREPARED:July 17th, 2000
DEVICE NAME:Gambro Polyflux 17R & 21R Capillary Dialyzers/Filters Labeled for Multiple Use
CLASSIFICATION NAMES:High Permeability Hemodialyzer / Hemofilters
PREDICATE DEVICE:Gambro Polyflux 17S & Gambro 21S Hemodialyzers/Filters Labeled for Single Use

Kgq4390

p.113

Device Description:

Gambro Polyflux 17R & 21 R Capillary Dialyzers/Filters Labeled for Multiple Use

The Gambro Polyflux 17R and 21R, Capillary Dialyzers/Filters labeled for multiple use (reuse) are identical in construction in function to Gambro Polyflux 17S and 21S Hemodialyzers / Hemofilters labeled for single use which are currently marketed in the United States and have been previously cleared by the FDA under 510(k) Notification K981414. Only the product designation "R" has been changed from "S" to designate that they are labeled for multiple use (reuse).

These devices are intended for use in hemodialysis for the treatment of acute and chronic renal failure and for certain types of intoxications for both single when reprocessed for reuse for a maximum of 15 reprocessing reuse cycles on the same patient. If reprocessing and reuse is practiced, it is recommended that the reuse be done under the conditions as existed in the in vitro and confirmatory clinical studies undertaken by Gambro and presented in the labeling for this device. It may also be used in cases of acute fluid overload for the removal of plasma water. The membrane used in these devices is polyethersulfone (PES) which is identical to the membrane utilized in the Gambro Polyflux 17S and 21S Hemodialyzers / Hemofitters for labeled for single use which have been previously approved for marketing in the United States under a 510K Notification (K982414). A copy of this approval letter is included in Appendix I of this Notification.

Blood enters a blood inlet port where it is distributed to polyethersulfone hollow fibers. Each hollow fiber has an inner diameter of approximately 215 microns (wet hollow fiber internal diameter) and a wall thickness of 50 microns. The number of polyethersulfone hollow fibers in each hemodialyzer / hemofilter is 10,000 for the Polyflux 17R and 12,500 for the 21R. These dialyzers have an effective membrane length of 250 mm. The effective membrane surface area is 1.7 square meters for the 17R and 2.1 square meters for the

000263

{1}------------------------------------------------

K994390

p.213

510K Notification Gambro Polyflux 17R Labeled for Multiple Use December 2nd , 1999

21R. The housing and end caps of these hemodialyzers / hemofiters are made of polycarbonate. The fibers used in the Gambro Polyflux 17R and 21R are identical in design and materials to the previously approved Gambro Polyflux 17S and 21S Hemodialyzers / Hemofilters for labeled for single use (K982414). The patient's blood traverses the inside of the hollow fibers and exits the device via a blood exit port.

By means of a hydrostatic pressure or transmembrane pressure which is created by a combination of positive and negative pressures across the polyethersulfone membrane, plasma water along with certain lower molecular weight solutes of plasma water pass through the membrane and into the dialysate or filtrate compartment of the device. Removal of uremic toxins and waste products are removed from the patient's blood in this device by means of both diffusion and convection through polyethersulfone membrane and into the counter current flowing dialysis solution during hemodialysis. The dialysate exits the devices via a dialysate outlet port.

Predicate Devices:

The Gambro Polyflux 17R and 21R Capillary Dialyzers / Filters labeled for multiple use (reuse) is identical in design and construction to the currently marketed Gambro Polyflux 17S and 21S, Hemodialyzers / Filters, respectively, labeled for single use which have been approved for marketing / sale in the United States under a 510K Notification (K982414). Both the predicate and the proposed devices, incorporate an identical polyethersulfone membrane and other blood contact materials. They are the same device with the exception of the labeling. The intended use for the proposed and predicate devices is also the same; hemodialyzer/ filter.

PREDICATE DEVICE

DEVICE NAMESGambro Polyflux 17S & 21S, Hemodialyzers / Filterslabeled for Single use
INTENDED USEHemodialyzer/Filter
510K NUMBERK982414
APPROVAL DATE3/26/99

With respect to performance, these hemodialyzers/filters perform in a manner substantially We therefore consider the predicate devices substantially equivalent to each other. equivalent to existing products in commercial distribution in the United States

Intended Use:

POLYFLUX R is indicated for use in hemodialysis for the treatment of chronic or acute renal failure. The choice of the filter is the responsibility of the physician. Special attention must be paid in connection with pediatric use.

CAUTION! If POLYFLUX R is reused, the procedure and disinfection specified in the RENATRON INSTRUCTION MANUAL must be followed.

000264

{2}------------------------------------------------

p. 3/3

The POLYFLUX R may be reprocessed for reuse on the same patient.

This indication statement is essentially the same as the indication statement for the predicate device.

Technological Characteristics:

Comparing the proposed devices to the predicate devices, they are identical with the exception that one is labeled for single use (Gambro Polyflux 17S & 21S Hemodialyzer/Filters and the other is labeled for multiple use (reuse) (Gambro Polyflux 17R & 21R Hemodialyzer/Filter. Both the proposed and predicate devices use the same polyethersulfone, hollow fiber membrane. Both the proposed and predicate devices use polycarbonate for the housing and header material and polyurethane for the membrane potting material.

Summary of Non-Clinical Tests:

In vitro data was collected according to the FDA Guidance for Hemodialyzer Reuse Labeling.

Clinical Test Results:

Clinical data was collected according to the FDA Guidance for Hemodialyzer Reuse Labeling.

Conclusions:

Testing performed on the Gambro Polyflux 17R and 21R Capillary Dialyzers/Filters indicates that they are safe, effective, and perform as well as the predicate devices, when used in accordance with the instructions for use. In vitro and in vivo performance data and directions for reuse have been included in the labeling.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 6 2000

Jeffrey R. Shideman, Ph.D. Gambro Healthcare 225 Union Boulevard Suite 600 Lakewood, CO 80228

Re: K994390

Multiple Use Labeling for Gambro Polyflux 17R and 21R Hemodialyzers Dated: July 17, 2000 Received: July 28, 2000 Regulatory Class: II 21 CFR §876.5860/Procode: 78 MSF

Dear Dr. Shideman:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have we nave tevice is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate comment for the more in the mactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the may, uculone, market the do not of the gisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject If your device is dassinod (oo acon of any regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Fride requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (2) CFR Part 1 raches requirements, to sol tour mare Quand the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation regulatory action. In addition, FDA may publish further announcements to comply with the Federal Register. Please note: this response to your premarket notification submission does not comeching you device in allo reast to sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding I fills letter will and you to obgally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro If you don't devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers-Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Daniel G. Schultz, M.D.

Captain, USPHS Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure (s)

{4}------------------------------------------------

510 (k) NUMBER (IF KNOWN): K994390

DEVICE NAME: Gambro Polyflux 17R & 21R

INDICATIONS FOR USE:

Indications for Use Statement

POLYFLUX R is indicated for use in hemodialysis for the treatment of chronic or acute renal failure. The I OLITLOA It is thatea you use in not the physician. Special attention must be paid in connection with pediatric use.

CAUTION! If POLYFLUX R is reused, the procedure and disinfection specified in the RENATRON INSTRUCTION MANUAL must be followed.

The POLYFLUX R may be reprocessed for reuse on the same patient.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter-Use (Optional Format 1-2-96)

David A. Berman

(Division Sign-Off) Division of Reproductive, Abdominal. EN and Radiological De 510(k) Number

000204

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”