K Number
K994069

Validate with FDA (Live)

Date Cleared
2000-06-02

(183 days)

Product Code
Regulation Number
892.1570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Diagnostic ultrasound imaging or fluid flow analysis or the human body as follows: the urinary tract.

Device Description

The Olympus UM-S30-25R Ultrasonic Probe has been designed to be used with an Olympus Endoscopic Ultrasound center, Probe Driving Unit and an endoscope for intraluminal ultrasonic imaging of the urinary tract.

The probe is connected to the endoscopic ultrasound center through the probe driving unit. The probe is attached to the endoscope which allows the user to use probe. The probe is inserted into the patient through a channel of the endoscope. A probe driving unit controls the rotation of the transducer.

UM-S30-25R Ultrasound Probe produces a B-mode scans using the de-aerated water immersion method and balloon method. The probeoffers 360 degree mechanical /radial scanning of the tissue under observation. The outer diameter of the insertion tube is 2.4mm and the length is 2050mm.

AI/ML Overview

The provided text is a 510(k) summary for the Olympus UM-S30-25R Ultrasonic Probe and associated equipment. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study to prove acceptance criteria were met for a new AI/software-based medical device.

Therefore, many of the requested elements are not applicable to the information contained within this 510(k) summary. This document describes the device, its intended use, and argues for substantial equivalence based on design, materials, and compliance with standards, not performance against specific clinical acceptance criteria through a dedicated study.

Here's an breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

  • Not applicable / Not provided. This 510(k) summary does not define specific performance acceptance criteria (e.g., sensitivity, specificity, accuracy thresholds) for the device. Instead, it relies on demonstrating that the device meets safety and effectiveness standards through design compliance and material equivalence to a predicate device.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable / Not provided. The document does not describe a performance study with a test set of data. The non-clinical tests mentioned are related to design compliance (IEC 60601-1, IEC 60601-1-2) and material equivalence, not performance on a dataset.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable / Not provided. As no test set performance study is described, there's no mention of experts establishing ground truth for such a study.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable / Not provided. No test set performance study is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable / Not provided. This document concerns a physical ultrasonic probe and its associated hardware, not an AI or software component that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not applicable / Not provided. This device is an ultrasonic probe, a hardware component, which requires human operation. It is not an algorithm that can perform standalone.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable / Not provided. No ground truth for device performance is described in a study context.

8. The sample size for the training set

  • Not applicable / Not provided. The device is a hardware probe, not a machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

  • Not applicable / Not provided. Not a machine learning device.

Summary of the study/evidence presented in the document:

The provided document describes non-clinical tests focused on design compliance and material equivalence to demonstrate substantial equivalence to a legally marketed predicate device (K982610 for the probe and associated units, K982734 for the balloon sheath).

  • Design Tests:
    • Endoscopic Ultrasound Center EU-M30 and Probe Driving Unit MH-240 were tested for compliance with IEC 60601-1 and the Revised 510(k) Diagnostics ultrasound Guidance for 1993, 1995.
    • Ultrasonic Probe UM-S30-25R was tested for electromagnetic compatibility (EMC) with IEC 60601-1-2.
    • Results: The devices "meet the requirements" of these standards.
  • Materials Test:
    • The patient contacting materials are stated to be "identical to predicate device."

Conclusion: The summary concludes that, "When compared to the predicate device, the Olympus UM-S30-25R Ultrasonic probe and associated ancillary equipment do not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness." This statement forms the basis for the FDA's finding of substantial equivalence.

{0}------------------------------------------------

K994069
Page 1 of 3

510(k) SUMMARY OLYMPUS UM-S30-25R ULTRASONIC PROBE AND ASSOCIATED ANCILLARY EQUIPMENT FOR URINARY TRACT USE

A. Submitter's Name, Address, Phone and Fax Numbers

1. Manufacturer of the subject devices

Name & Address of manufacturer:

Registration No .: Address, Phone and Fax Numbers: of R&D Department, Endoscope Division

Olympus Optical Co., Ltd. 2-3-1 Shinjyuku Monolis Nishishinjyuku Shinjuku-ku, Tokyo, Japan 8010047 2951 Ishikawa-Cho, Hachioji-shi, Tokyo 192-8507 Japan TEL 0426-42-5101 FAX 0426-46-2786

B. Name of Contact Person

Name: Address, Phone and Fax Numbers: Ms. Laura Storms-Tyler Olympus America Inc. Endoscope Division Two Corporate Center Drive Melville, New York 11747-3157 TEL: (516) 844-5474 FAX: (516) 844-5416

C. Device Name, Common Name, Classification Name and Predicate Devices

Device Name:Ultrasonic Probe UM-S30-25REndoscopic Ultrasound Center EU-M30Probe Driving Unit MH-240
Common Name:Ultrasonic Probe
Classification Name:21 CFR892.1570 Diagnostic Ultrasonic transducer

{1}------------------------------------------------

K994069.
page 2 of 3

Predicate Device:

MODELNAME510(k)#
UM-2RUltrasonic ProbeK982610
UM-3RUltrasonic ProbeK982610
EU-M30Endoscopic Ultrasound CenterK982610
MH-240Probe Driving UnitK982610
MH-246Balloon SheathK982734 (for UM-2R/3R)

D. Description of the Device(s)

The Olympus UM-S30-25R Ultrasonic Probe has been designed to be used with an Olympus Endoscopic Ultrasound center, Probe Driving Unit and an endoscope for intraluminal ultrasonic imaging of the urinary tract.

The probe is connected to the endoscopic ultrasound center through the probe driving unit. The probe is attached to the endoscope which allows the user to use probe. The probe is inserted into the patient through a channel of the endoscope. A probe driving unit controls the rotation of the transducer.

UM-S30-25R Ultrasound Probe produces a B-mode scans using the de-aerated water immersion method and balloon method. The probeoffers 360 degree mechanical /radial scanning of the tissue under observation. The outer diameter of the insertion tube is 2.4mm and the length is 2050mm.

E. Intended Use of the Device(s)

The Olympus UM-S30-25R Ultrasonic Probe has been designed to be used with an Olympus Endoscopic Ultrasound center, Probe Driving Unit and an endoscope for intraluminal ultrasonic imaging of the urinary tract.

F. Summary of the Technological Characteristics of the Device compared to the Predicate Device(s)

Ultrasonic Frequency

The ultrasonic frequency is changed to 30MHz.

{2}------------------------------------------------

K994069
page 3 of 3

G. Summary including a Brief Discussion of Non-clinical Tests and How their Results support Determination of Substantial Equivalence

1. Design

Endoscopic Ultrasound Center EU-M30 and Probe Driving Unit MH-240 have been designed, manufactured and tested in compliance with IEC 60601-1 and Revised 510(k) Diagnostics ultrasound Guidance for 1993,1995. It meets the requirements of IEC 60601-1 and Revised 510(k) Diagnostics ultrasound Guidance for 1993,1995.

Ultrasonic Probe UM-S30-25R has been tested to verify electromagnetic compatibility (EMC) with IEC 60601-1-2. It meets the requirements of IEC 6060-1-1-2.

2. Materials

The patient contacting materials are identical to predicate device.

H. Summary including Conclusions drawn from Non-clinical Tests

When compared to the predicate device, the Olympus UM-S30-25R Ultrasonic probe and associated ancillary equipment do not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of an eagle or bird-like figure, composed of three curved lines that suggest feathers or wings.

Public Health Service

JUN - 2 2000

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Laura Storms-Tyler Director, Regulatory Affairs and Quality Assurance Olympus America, Inc. Two Corporate Center Drive Melville, NY 11747-3157

Re: K994069

UM-S30-25R Ultrasonic Probe Regulatory Class: II Product Code: 90 ITX/21 CFR 892.1570 Dated: March 23, 2000 Received: March 29, 2000

Dear Ms. Storms-Tyler:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the UM-S30-25R Ultrasonic Probe intended for use with the Olympus EU-M30 Endoscopic Ultrasound Center as described in your premarket notification.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 – Ms. Laura Storms-Tyler

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and rr you desire specific as reviews and the see contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers on your responsive manaber (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

If you have any questions regarding the content of this letter, please contact Mr. Rodrigo C. Perez at (301) 594-1212.

Sincerely yours,

Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosures

{5}------------------------------------------------

Appendix F

Diagnostic Ultrasound Indications for Use Form OLYMPUS EU-M30 ENDOSCOPIC ULTRASOUND CENTER (ancillary system) Intended Use: Diagnostic ultrasound imaging or fluid flow analysis or the human body as follows:

Mode of Operation
Clinical ApplicationABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
Ophthalmic
Fetal
Abdominal
Intraoperative (Specify)
Intraoperative Neurological
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Cardiac
Transesophageal
Transrectal
Transvaginal
Transurethral
Intravascular
Peripheral Vascular
LaparoscopicMusculo-skeletalConventional
Musculo-skeletal Superficial
Other (Specify)

N = new indication: P = previously cleared by EDA: E = added under Appendix FN = new indication; P = previously cleared by FDA; E = added under Appendix E Additional Comments:

the urinary tract.

(PLEASE DO NOT WRITE BELOW THE LINE-CONTIINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evalution (ODE)
---------------------------------------------------------

Prescription Use (Per 21 CFR 801.109)

F-3

(Division Sign-Off)

Division of Reproductive, Abdominal, ENT, and Radiological Devi

510(k) Number K94406

{6}------------------------------------------------

Appendix F

Diagnostic Ultrasound Indications for Use Form OLYMPUS UM-S30-25R ULTRASONIC PROBE (Subject Device) Intended Use: Diagnostic ultrasound imaging or fluid flow analysis or the human body as follows:

Intended Use. Diagnostic ultrasound imaging or fluid flow analysis of the human Body as follows:

Mode of Operation
Clinical ApplicationABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
Ophthalmic
Fetal
Abdominal
Intraoperative (Specify)
Intraoperative Neurological
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Cardiac
Transesophageal
Transrectal
Transvaginal
TransurethralN
Intravascular
Peripheral Vascular
LaparoscopicMusculo-skeletalConventional
Musculo-skeletal Superficial
Other (Specify)N

N = new indication; P = previously cleared by FDA; E = added under Appendix E Additional Comments: the urinary tract.

(PLEASE DO NOT WRITE BELOW THE LINE CONTIN

(PLEASE DO NOT WRITE BELOW THE LINE-CONTIINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evalution (ODE)

Prescription Use (Per 21 CFR 801.109)

F-3

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radialasia

Radiological Devices
510(k) Number K994069

§ 892.1570 Diagnostic ultrasonic transducer.

(a)
Identification. A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.(b)
Classification. Class II.