(223 days)
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Neonatal Cephalic, Intravascular, Peripheral Vascular, Adult Cephalic
SYSTEM LEGEND TC 22
This document is an FDA 510(k) clearance letter dated June 29, 2000, for the Nicolet EME Legend TC22 ultrasound system and associated transducers. It outlines the intended uses of the device and confirms its substantial equivalence to legally marketed predicate devices.
The document does not contain information about specific acceptance criteria, a study proving the device meets those criteria, sample sizes for test or training sets, data provenance, expert qualifications, adjudication methods, multi-reader multi-case studies, standalone algorithm performance, or how ground truth was established. This type of detailed performance data is typically found in the 510(k) summary or premarket notification submission itself, not in the clearance letter.
The tables presented in the document are "Diagnostic Ultrasound Indications for Use Forms," which specify the clinical applications and modes of operation for which each transducer is cleared. They are not performance reports.
Therefore, I cannot provide the requested information from the provided text.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle. The text is in all capital letters and is arranged to follow the curve of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 9 2000
David W. Wagner Director Quality Assurance and Regulatory Affairs Nicolet Vascular. 6355 Joyce Drive Golden, Colorado 80403
K993939 Re: Nicolet EME Legend TC22 Regulatory class: II 21 CFR 892.1556/Procode: 90 IYN 21 CFR 892.1570/Procode: 90 ITX Dated: March 30, 2000 Received: April 4, 2000
Dear Mr. Wagner:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls rowisions of the Act. The general controls provisions of the Act include requirements for annual registration, provisions of the research gractice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Nicolet EME Legend TC22, as described in your premarket notification:
Transducer Model Number
2 MHz 10 mm Pulsed Wave 2 MHz 15 mm Pulsed Wave 4 MHz Pulsed Wave 4 MHz Continuous Wave 8 MHz Pulsed Wave 8 MHz Continuous Wave
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{1}------------------------------------------------
Page -2- Mr. Wagner
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Paul M. Gammell, Ph.D. at (301) 594-1212.
Sincerely yours,
Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K 993939
SYSTEM LEGEND TC 22 Tiran's CRANINL FADULT CEPHALIC
Appendix F
Poriphern Vascular Diagnstic Dev
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined - (specify) | Other (specify) | ||
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intraoperative (specify) | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (specify) | ||||||||||||
| Neonatal Cephalic | ◼ | |||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intravascular | ◼ | ◼ | ||||||||||
| Peripheral Vascular | ◼ | ◼ | ||||||||||
| Laparoscopic | ||||||||||||
| Musculo-skeletalConventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other (specify) |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
| Additional Comments: | SYSTEM UTILIZES 2EA 2MHZ PULSED WAVE PROBES, 1 EA 4MHZ PULSED AND CW PROBE AND 1 EA 8MHZ PULSED AND CW PROBE. |
|---|---|
| ---------------------- | --------------------------------------------------------------------------------------------------------------- |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Cadomanal. ONT
and Radiological Devices
| 510(k) Number | K993939 |
|---|---|
| --------------- | --------- |
{3}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined(specify) | Other(specify) |
| Ophthalmic | ||||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intraoperative (specify) | ||||||||||
| Intraoperative Neurological | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | ||||||||||
| Laparoscopic | ||||||||||
| Musculo-skeletalConventional | ||||||||||
| Musculo-skeletal Superficial | ||||||||||
| Other (specify) | ||||||||||
| N= new indication; P= previously cleared by FDA; E= added under Appendix | ||||||||||
| Additional Comments: | E- 2MHZ 10MM PULSED WAVE |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Prescription Use (Per 21 CFR 801.109)
F-3 Division of Reproductive, Abdominal, ENT,
and Radiological Devices
F-3
510(k) Number J
{4}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined(specify) | Other(specify) | |
| Ophthalmic | |||||||||||
| Fetal | |||||||||||
| Abdominal | |||||||||||
| Intraoperative (specify) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Pediatric | |||||||||||
| Small Organ (specify) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transurethral | |||||||||||
| Intravascular | |||||||||||
| Peripheral Vascular | |||||||||||
| Laparoscopic | |||||||||||
| Musculo-skeletalConventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other (specify) | |||||||||||
| N= new indication; P= previously cleared by FDA; E= added under AppendixAdditional Comments:E_______________________________________________________________________________________________________________________________________________________ | 2MHZ | ||||||||||
| (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)Concurrence of CDRH, Office of Device Evaluation (ODE) | |||||||||||
| Prescription Use (Per 21 CFR 801.109)ﺎ | (Division Sign-Ja)and Radiological Devices | Division of Reproductive, Abdominal, ENT |
510(k) Number _
{5}------------------------------------------------
PROBE FOR LEGEND TC 22 SYSTEM
(TRANSDUCER) Diagnostic Ultrasound
Appendix F
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined(specify) | Other(specify) |
| Ophthalmic | ||||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intraoperative (specify) | ||||||||||
| Intraoperative Neurological | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | ||||||||||
| Laparoscopic | ||||||||||
| Musculo-skeletalConventional | ||||||||||
| Musculo-skeletal Superficial | ||||||||||
| Other (specify) | ||||||||||
| N= new indication; P= previously cleared by FDA; E= added under Appendix EAdditional Comments: | E- 4 MHZ PULSED WAVE | |||||||||
| Prescription Use (Per 21 CFR 801.109) | (Division Sign-Off)Division of Reproductive, Abdominal, ENT,and Radiological Devices |
510(k) Number _
199
{6}------------------------------------------------
Appendix F ドゥツアクラク
PROBE FOR LEGEND TC-22 SYSTEM
(TRANSSICE)
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | ||||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intraoperative (specify) | ||||||||||
| Intraoperative Neurological | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | X | |||||||||
| Laparoscopic | ||||||||||
| Musculo-skeletalConventional | ||||||||||
| Musculo-skeletal Superficial | ||||||||||
| Other (specify) |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: E- 4 MHZ CONTINUOUS WAVE
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
F-3
(Division Sign-off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
and Radiological Devices
510(k) Number .
{7}------------------------------------------------
PROBE FOR LEGEND TC22 SYSTEM
(TRANSDUCER)
Diagnostic Ultrasound Inc.
Appendix F
ドゥッスリ39
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | ||||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intraoperative (specify) | ||||||||||
| Intraoperative Neurological | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | ||||||||||
| Laparoscopic | ||||||||||
| Musculo-skeletal Conventional | ||||||||||
| Musculo-skeletal Superficial | ||||||||||
| Other (specify) |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: E-8 MHZ PULSED WAVE
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
F-3
Prescription Use (Per 21 CFR 801.109)
510(k) Number _
99
39
{8}------------------------------------------------
PLEASE FOR LEGEND TC-22 SYSTEM
Appendix F ドゥッショタ
(THINSDIXER
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined(specify) | Other(specify) |
| Ophthalmic | ||||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intraoperative (specify) | ||||||||||
| Intraoperative Neurological | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | ||||||||||
| Laparoscopic | ||||||||||
| Musculo-skeletalConventional | ||||||||||
| Musculo-skeletal Superficial | ||||||||||
| Other (specify) |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: E - 8 MHZ CONTINUOUS WAVE
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
Prescription Use (Per 21 CFR 801.109)
F-3
510(k) Number ________________________________________________________________________________________________________________________________________________________________
N/A