ACON ONE STEP PREGNANCY TEST STRIP (URINE)
K993203 · ACON Laboratories, Inc. · JHI · Dec 1, 1999 · Clinical Chemistry
Device Facts
| Record ID | K993203 |
| Device Name | ACON ONE STEP PREGNANCY TEST STRIP (URINE) |
| Applicant | ACON Laboratories, Inc. |
| Product Code | JHI · Clinical Chemistry |
| Decision Date | Dec 1, 1999 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1155 |
| Device Class | Class 2 |
Intended Use
ACON™ One Step Pregnancy Test Strip (Urine) is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid for the early detection of pregnancy. This test is for professional use.
Device Story
Rapid chromatographic immunoassay; detects hCG in urine samples. Principle: capillary action moves urine along membrane; monoclonal/polyclonal antibodies react with hCG to form colored lines. Output: visual colored line in test area indicates positive result; control line indicates procedural validity. Used in clinical settings by healthcare professionals. Detects hCG concentrations ≥25 mIU/ml. Standardized to WHO Third International Standard. Benefits: provides rapid, qualitative pregnancy status confirmation to assist clinical decision-making.
Clinical Evidence
Multi-center clinical evaluation comparing ACON™ One Step Pregnancy Test Strip to a commercially available urine membrane test. Study included 150 urine specimens. Results showed 100% concordance between the subject device and the comparator.
Technological Characteristics
Rapid chromatographic immunoassay; utilizes monoclonal and polyclonal antibodies. Detects hCG at concentrations ≥25 mIU/ml. Standardized to WHO Third International Standard. No cross-reactivity with hLH (300 mIU/ml), hFSH (1000 mIU/ml), or hTSH (1000 μIU/ml). Form factor: test strip.
Indications for Use
Indicated for professional use as an aid in the early detection of pregnancy by qualitatively detecting human chorionic gonadotropin (hCG) in urine.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Related Devices
- K993317 — ACON HCG ONE STEP PREGNANCY TEST DEVICE (URINE) · ACON Laboratories, Inc. · Dec 1, 1999
- K022680 — AZOG, INC. HCG ONE-STEP URINE PREGNANCY TEST STRIP (DIPSTICK) · Azog, Inc. · Nov 1, 2002
- K062361 — INNOVACON HCG UKTRA TEST DEVICE · Innovacon, Inc. · Oct 23, 2006
- K022681 — AZOG, INC. HCG ONE-STEP URINE PREGNANCY TEST (DEVICE OR CASSETTE) · Azog, Inc. · Nov 1, 2002
- K041946 — ACON SPECTRUM URINE/SERUM PREGNANCY TEST DEVICE · ACON Laboratories, Inc. · Aug 16, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
K993203
DEC - 1 1996
## "Summary of Safety & Effectiveness"
ACON™ One Step Pregnancy Test Strip (Urine) is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid for the early detection of pregnancy. The test utilizes a combination of monoclonal and polyclonal antibodies to selectively detect elevated levels of hCG. The assay is conducted by immersing the test strip in the urine sample and observing for the formation of colored lines. The specimen migrates via capillary action along the membrane to react with the colored conjugate. Positive specimens react with the specific antibody-hCG-colored conjugate and form a colored line in the Test Area of the membrane. Absence of this colored line suggests a negative result. To serve as a procedural control, a colored line in the Control Area will always appear regardless of the presence or absence of hCG.
A multi-center clinical evaluation was conducted comparing the results obtained using ACON™ One Step Pregnancy Test Strip (Urine) and another commercially available urine membrane test. The study included 150 urine specimens tested with both assays. The following results were found:
| | Positive<br>Results | Negative<br>Results |
|--------------------------------------------------|---------------------|---------------------|
| ACON™ One Step<br>Commercially<br>Available Test | 78 | 72 |
| | 78 | 72 |
ACON™ One Step Pregnancy Test Strip (Urine) showed a 100% concordance with the other commercially available test.
ACON™ One Step Pregnancy Test Strip (Urine) detects hCG concentrations of 25 mIU/ml and greater. The test has been standardized to the World Health Organization Third International Standard. The addition of hLH (300mUl/ml, hFSH (1000 mIU/ml), and hTSH (1000μIUml) to negative and positive serum and urine specimens showed no cross-reactivity.
Ara C. R. York
Nora C.R. ACON Laboratories, Inc. 11175 Flintkote, Avenue, Suite F San Diego, CA 92121 UDA
9/23/99
**Date**
Premarket Notification 510(k) Number
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or birds in flight, arranged in a row.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC - 1 1999
Ms. Nora C.R. York Regulatory Affairs Manager Acon Laboratories, Inc. 11175 Flintkote Avenue Suite F San Diego. California 92121
K993203 Re:
> Trade Name: Acon™ One Step Pregnancy Test Strip (Urine) Regulatory Class: II Product Code: JHI Dated: September 23, 1999 Received: September 24, 1999
Dear Ms. York:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
## Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrlv/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications For Use
Registration 510(k)Number:
Device Name: ACON™ One Step Pregnancy Test Strip (Urine)
"Indications For Use" - ACON™ One Step Pregnancy Test Strip (Urine) is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid for the early detection of pregnancy. This test is for professional use.
(Please do not write below this point)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(per 21 CFR 801.109)
Over-The-Counter Use Or