INNOVACON HCG UKTRA TEST DEVICE
K062361 · Innovacon, Inc. · JHI · Oct 23, 2006 · Clinical Chemistry
Device Facts
| Record ID | K062361 |
| Device Name | INNOVACON HCG UKTRA TEST DEVICE |
| Applicant | Innovacon, Inc. |
| Product Code | JHI · Clinical Chemistry |
| Decision Date | Oct 23, 2006 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1155 |
| Device Class | Class 2 |
Intended Use
The Innovacon™ hCG Ultra Test Device is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine to aid in the early detection of pregnancy. It is for healthcare professionals only.
Device Story
Rapid chromatographic immunoassay; detects hCG in serum or urine samples. Operates via capillary action; specimen migrates along membrane to react with mouse monoclonal and goat polyclonal anti-hCG antibody conjugates. Positive result indicated by colored line at test region; procedural control line confirms proper test execution. Used by healthcare professionals in clinical settings. Provides qualitative visual output; aids in early pregnancy diagnosis. Standardized to WHO Fourth International Standard (NIBSC 75/589). Sensitivity: 10 mIU/mL (serum), 20 mIU/mL (urine).
Clinical Evidence
Clinical study conducted at three U.S. sites by healthcare professionals. Demonstrated performance equivalency to the Fisher Sure-Vue Serum/Urine hCG-STAT. Study confirmed high reproducibility and precision. Additional laboratory studies verified robustness regarding specificity, interference substances, urinary pH, specific gravity, dose hook effect, time flexibility, and sample volume.
Technological Characteristics
Rapid chromatographic immunoassay; lateral flow membrane. Uses mouse monoclonal and goat polyclonal anti-hCG antibodies. Sensitivity: 10 mIU/mL (serum), 20 mIU/mL (urine). Standardized to WHO Fourth International Standard (NIBSC 75/589). No cross-reactivity with FSH, LH, or TSH. Visual readout.
Indications for Use
Indicated for healthcare professionals to qualitatively detect human chorionic gonadotropin (hCG) in serum or urine for early pregnancy detection.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
- Fisher Sure-Vue Serum/Urine HCG-STAT
Related Devices
- K993203 — ACON ONE STEP PREGNANCY TEST STRIP (URINE) · ACON Laboratories, Inc. · Dec 1, 1999
- K993317 — ACON HCG ONE STEP PREGNANCY TEST DEVICE (URINE) · ACON Laboratories, Inc. · Dec 1, 1999
- K041946 — ACON SPECTRUM URINE/SERUM PREGNANCY TEST DEVICE · ACON Laboratories, Inc. · Aug 16, 2004
- K022681 — AZOG, INC. HCG ONE-STEP URINE PREGNANCY TEST (DEVICE OR CASSETTE) · Azog, Inc. · Nov 1, 2002
- K955707 — ONESTEP URINE/SERUM HCG MODULE PREGNANCY TEST KIT · Excel Scientific, Inc. · May 2, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
K062361
### Conclusion:
OCT 2 3 2006
Applying 2 to 6 drops of samples to the Innovacon hCG Ultra Test Devices yielded correct results as read at 3 or 5 and 10 minutes after sample application. However, Innovacon recommends using 3 drops of the urine or serum sample to obtain the best test result, as prescribed in our package insert.
# XV. 510(k) Summary
#### Device Names:
Innovacon™ hCG Ultra Test Device
#### Common Name:
Pregnancy Test Kit, Professionals
# Medical Specialty:
Clinical Chemistry
#### Intended Use:
The Innovacon™ hCG Ultra Test Device is a rapid chromatographic immunoassay for the qualitative detection of human Chorionic Gonadotropin (hCG) in serum or urine to aid in the early detection of pregnancy. It is for professional in-vitro diagnostic use only
### Device Description:
The test utilizes a combination of antibodies including mouse monoclonal antihCG antibodies and goat polyclonal anti-hCG antibodies to selectively detect elevated levels of hCG. The assay is conducted by adding urine or serum specimens to the specimen well of the test device and observing the formation of colored lines. The specimen migrates via capillary action along the membrane to react with the colored conjugate.
Positive specimens react with the specific colored antibody conjugates and form a colored line at the test line region of the membrane. Absence of this colored line suggests a negative result. To serve as a procedural control, a colored line will always appear at the control line region if the test has been performed properly.
The Innovacon™ hCG Ultra Test Device is a rapid test that qualitatively detects hCG at the sensitivity of 10 mIU/mL in serum and 20 mIU/mL in urine. The cutoff concentration of this test has been standardized to the World Health Organization Fourth International Standard for Chorionic Gonadotropin (NIBSC Code: 75/589). At the level of claimed sensitivity, the hCG Ultra Test Device
{1}------------------------------------------------
shows no cross-reactivity interference from the structurally related glycoprotein hormones FSH, LH and TSH at extra high physiological levels.
# Clinical Studies:
A clinical study was conducted in three sites in the U.S. by healthcare professionals with varying educational backgrounds and laboratory experience and demonstrated performance equivalency between the Fisher Sure-Vue Serum/Urine hCG-STAT test and the Innovacon hCG Ultra Test Device by professionals. The POL study demonstrated high degree of reproducibility and precision of the Innovacon hCG Ultra Test Device
The POL sites also found the Innovacon hCG Ultra Test Device very easy to use, and that they have had no trouble understanding the labeling, reading the instructions, or interpreting the results.
## Additional Laboratory Studies:
Additional laboratory study results on performance including specificity, interference substances, urinary pH, urinary specific gravity, dose hook effect, time flexibility, and sample volume flexibility studies are also included in this submission. These results indicate that the Innovacon hCG Ultra Test Device is robust and will give accurate results under many adverse conditions.
# Substantial Equivalency on Performance:
The overall performance data indicate that the Innovacon hCG Ultra Test Device is safe, effective and substantially equivalent to the Fisher Sure-Vue Serum/Urine HCG-STAT legally sold in the U. S. market.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with flowing lines, positioned to the right of a circular arrangement of text. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" and is arranged around the upper half of the circle.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Edward Tung, Ph.D. Vice President, Regulatory Affairs Innovacon Inc 4106 Sorrento Valley Blvd. San Diego, CA 92121
OCT 2 3 2006
Re: k062361 Trade/Device Name: Innovacon™ hCG Ultra Test Device Regulation Number: 21 CFR 862.1155 Regulation Name: Human chorionic gonadotropin (HCG) test system Regulatory Class: Class II Product Code: JHI Dated: August 11, 2006 Received: August 14, 2006
Dear Dr. Tung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Alberto Garcia
Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications For Use V.
510(k) Number (if known): K062361
Device Name:
Innovacon™ hCG Ultra Test Device
"Indications for Use":
Innovacon™ hCG Ultra Test Device is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine to aid in the early detection of pregnancy. It is for healthcare professionals only.
Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Device
Evaluation and Safety Evaluation and Safety
K062361
Page 1 of