K Number
K993190

Validate with FDA (Live)

Date Cleared
2000-02-10

(140 days)

Product Code
Regulation Number
886.4370
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SIS, Ltd., ACCM microkeratome is a precision-manufactured instrument indicated for lamellar resection of the cornea.

Device Description

The SIS, Ltd ACCM Microkeratome is a precision-manufactured instrument designed for cutting a precise corneal disc of preselected thickness and diameter. The design, material and operating principle are very similar to those of the predicate devices.

AI/ML Overview

The provided text does not contain detailed acceptance criteria, specific performance metrics, or a study design for a device. It primarily focuses on the submission of a 510(k) for a microkeratome and its substantial equivalence to predicate devices. Therefore, I cannot populate the table or answer the questions related to acceptance criteria and study details.

Here's a breakdown of what the document does include:

  • Device Name: Surgical Instrument Services (SIS), Ltd. Advanced Computer Controlled Microkeratome (ACCM)
  • Intended Use: Lamellar resection of the cornea.
  • Predicate Devices:
    • Bausch & Lomb Surgical: Hansatome Microkeratome (K972808)
    • Bausch & Lomb Surgical: Advance Corneal Shaper - ACS (K913697)
  • Technological Comparison (Table 1): This table outlines various technical characteristics of the SIS-ACCM and its predicate devices, demonstrating similarity in operating principle, intended use, blade material, and general features. This comparison is used to establish substantial equivalence.
  • Nonclinical Tests: The summary states that "The ACCM microkeratome has been designed and tested to applicable safety standards. Parameters related to lamellar resection were validated through extensive preclinical testing. The ACCM microkeratome does not raise any new issues of safety, effectiveness, or performance of the product." However, specific details of these tests, their results, or acceptance criteria are not provided.
  • FDA Clearance: The document includes the FDA's letter granting 510(k) clearance based on substantial equivalence, permitting the device to be marketed.

Therefore, the requested information regarding acceptance criteria and a study proving the device meets those criteria cannot be extracted from this document. The document's purpose is to demonstrate substantial equivalence, not to present a detailed performance validation study with specific acceptance criteria and results.

{0}------------------------------------------------

FEB 1 0 2000

K993190

SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter's name, address, telephone number, contact person, and date summary 1. prepared:

  • Surgical Instrument Services (SIS), Ltd. a. P. O. Box CH-2501 Biel Switzerland Phone: 41 (32) 332-91 61 41 (32) 332-91 62 Fax:
  • b. Contact Person: Frank Ziemer Managing Director, New Product Development
  • Date Summary Prepared: January 28, 2000 c.

Name of device, including trade name and classification name: 2.

  • Trade/Proprietary Name: Keratome a.
  • Classification Name: Keratome b.

Identification of the predicate device or legally marketed device or devices to which 3. substantial equivalence is being claimed:

Company:Bausch & Lomb Surgical
Device:Hansatome Microkeratome
510(k) :K972808
Date Cleared:October 24, 1997
Company:Bausch & Lomb Surgical
Device:Advance Corneal Shaper - ACS
510(k) :K913697
Date Cleared:November 5, 1991

{1}------------------------------------------------

4. A description of the device that is the subject of the 510(k), including explanation of how the device functions, basic scientific concepts, significant physical and performance characteristics (design, material, physical properties):

The SIS, Ltd ACCM Microkeratome is a precision-manufactured instrument designed for cutting a precise corneal disc of preselected thickness and diameter. The design, material and operating principle are very similar to those of the predicate devices.

ട. Statement of intended use:

The SIS, Ltd. ACCM Microkeratome is indicated for lamellar resection of the cornea.

{2}------------------------------------------------

Statement of how the technological characteristics of the device compare to those 6. of the predicate or legally marketed device.

Surgical Instrument Systems Advanced Computer Controlled Microkeratome

TABLE 1 TECHNOLOGICAL COMPARISON

CHARACTERISTICSSIS-ACCMPredicate DeviceAutomatic Corneal ShaperPredicate DeviceHansatome Microkeratome
Intended UseLamellar Corneal ResectionsLamellar Corneal ResectionsLamellar Corneal Resections
Operating PrincipleElectrically driven oscillating bladeElectrically driven oscillating bladeElectrically driven oscillating blade
Suction RingInterchangeable heads with fixedthickness platesOne head with interchangeablethickness platesInterchangeable heads with fixedthickness plates
Blade Drive SourceElectric Motor 6-9V DC(Reusable)Electric Motor 12V DC (Reusable)Electric Motor 6-9V DC(Reusable)
Thickness ControlFixed thickness plate in blade holder;two blade holders availableThickness PlatesThickness Plates
Blade SpeedBlade Oscillation 8000 RPM.Blade Oscillation 7,500 RPMBlade Oscillation >7,500 RPM
Blade Angle25°25°Unknown
Blade MaterialLow carbon stainless steel -Disposable, Single-UseLow carbon stainless steel -Disposable, Single-UseLow carbon stainless steel -Disposable, Single Use
Blade Sterilization MethodGamma irradiationGamma irradiationEtO
Flap Diameter8.5 mm and 9.5 mm9.0 mm10 mm
Flap Thickness Options160µm and 180 µm160µm and 180 µm160µm and 180 µm
Hinge LocationNasal or superiorNasalSuperior
Keratome MechanismDual rectilinear guide rails which aresymmetrical.Dual linear guideways with singlelinear gear trackSingle arcuate gear rack with temporalpivot pin
Console Details
Electrical110V or 215V AC110V or 215V AC110V or 215V AC
Vacuum PumpAC PoweredAC PoweredAC Powered
Blade Height VerificationClinically measured with microscope.Quality control check in house concerning mountingClinically measured with microscope.Clinically measured with microscope.
Foot ControlsDC PoweredDC PoweredDC Powered

{3}------------------------------------------------

7. Brief summary of nonclinical tests and results:

The ACCM microkeratome has been designed and tested to applicable safety standards. Parameters related to lamellar resection were validated through extensive preclinical testing. The ACCM microkeratome does not raise any new issues of safety, effectiveness, or performance of the product.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure with three flowing lines representing hair or movement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 0 2000

Ms. Judy Gordon Regulatory Consultant to S.I.S. ClingReg Consulting Services, Inc. 18732 Saginaw Irvine, CA 92612

Re: K993190 Trade Name: Advanced Computer Controlled Microkeratome (ACCM) Regulatory Class: I Product Code: 86 HNO Regulation: 886.4370 (Keratome) Dated: January 13, 2000 Received: January 14, 2000

Dear Ms. Gordon:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .

{5}------------------------------------------------

Page 2 - Ms. Judy Gordon

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

A. Ralph Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page of

K993190 NUMBER (IF KNOWN) : 510 (k)

Advanced Computer Controlled Microkeratome - ACCM DEVICE NAME :

INDICATIONS FOR USE:

The SIS, Ltd., ACCM microkeratome is a precision-manufactured instrument indicated for lamellar resection of the cornea.

Evenette T. Beens
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K993190

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter-Use (Optional Format 1-2-9

§ 886.4370 Keratome.

(a)
Identification. A keratome is an AC-powered or battery-powered device intended to shave tissue from sections of the cornea for a lamellar (partial thickness) transplant.(b)
Classification. Class I.