Intended Use

The Beckman Coulter OP 300 Low and High Urine Calibrators, in conjunction with SYNCHRON Reagents, are intended for use on SYNCHRON Systems for the calibration of Opiate 300 ng (Part No. 475024) enzyme immunoassays.

Device Description

The SYNCHRON® Systems OP 300 Low and High Urine Calibrators are intended for use on SYNCHRON Systems for the calibration of Opiate 300 ng enzyme immunoassays. This product contains a 5.0 mL bottle of the Low Urine Calibrator and a 5.0 mL bottle of the High Urine Calibrator. The storage temperature for the calibrators is +2°C to +8°C.

The OP Low and High Urine Calibrators are ready-to-use human, urine-based. liquid calibrators. They are derived by adding known quantities (traceable to Gas Chromatography/Mass Spectrometry) of morphine (for opiate) to human urine to achieve each drug analyte concentration.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the SYNCHRON® Systems OP 300 Low and High Urine Calibrators. This is a submission for an in vitro diagnostic (IVD) device, specifically calibrators used with enzyme immunoassays for detecting opiates in urine. The nature of this device means the "acceptance criteria" and "study" are focused on analytical performance and comparability to a predicate device, rather than a clinical study evaluating diagnostic accuracy from an AI algorithm.

Here's a breakdown of the information requested, tailored to the context of this IVD calibrator submission:


1. Table of Acceptance Criteria and Reported Device Performance

For an IVD calibrator, the acceptance criteria typically revolve around demonstrating that the new calibrator performs equivalently to a legally marketed predicate device, ensuring similar assay calibration and resulting in comparable analytical performance of the associated assay. The document doesn't explicitly state a "table of acceptance criteria" with numerical targets for performance metrics in the way an AI diagnostic would. Instead, the core of the submission is demonstrating substantial equivalence to the predicate device.

The "performance" in this context refers to the ability of the calibrators to effectively calibrate the Opiate 300 ng enzyme immunoassays on SYNCHRON Systems, yielding results comparable to those obtained using the predicate calibrators.

Based on the provided text, the primary "acceptance criterion" is:

  • Substantial Equivalence: The device must demonstrate substantial equivalence to a legally marketed predicate device (SYNCHRON® Systems DAT Low and High Urine Calibrators II, K983747) for its intended use.

The "reported device performance" is implicitly established by the FDA's decision:

  • FDA Finding: The FDA reviewed the 510(k) and determined the device is substantially equivalent to the predicate device for the stated indications for use. This implies the submitted data supported this finding.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a "test set" in the context of human-subject data for general diagnostic performance. For calibrators, performance is typically evaluated through analytical studies, such as:

  • Method Comparison Studies: Comparing results obtained using the new calibrator against the predicate calibrator.
  • Accuracy/Trueness Studies: Assessing how closely the calibrated assay measures known concentrations.
  • Precision Studies: Evaluating the reproducibility of results.

While the document mentions "adding known quantities (traceable to Gas Chromatography/Mass Spectrometry) of morphine (for opiate) to human urine to achieve each drug analyte concentration," this describes how the calibrators themselves are manufactured and their traceability, not a "test set" sample size for evaluating an algorithm.

Data Provenance: The calibrators are "human, urine-based" liquids. The origin of the human urine used for formulating the calibrators (e.g., country of origin) and whether the data for the equivalency studies were retrospective or prospective are not specified in the provided text.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to this type of device submission. Calibrators are analytical tools; their "ground truth" is established through metrological traceability (e.g., to GC/MS for concentration). There are no human experts "establishing ground truth" for diagnostic performance in the way a radiologist would for an image.

4. Adjudication Method for the Test Set

This information is not applicable for a calibrator submission. Adjudication methods like 2+1 or 3+1 refer to agreement among human readers for diagnostic tasks, which is not relevant here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a MRMC comparative effectiveness study was not done. This type of study assesses how AI assistance impacts human reader performance on diagnostic tasks, which is not relevant for a calibrator.

6. If a Standalone (Algorithm Only Without Human-in-the Loop Performance) Was Done

Not Applicable. This device is a calibrator, an essential component for an in vitro assay system. It does not perform diagnostic tasks independently or as an algorithm. Its "performance" is its ability to correctly calibrate the associated immunoassay.

7. The Type of Ground Truth Used

For the calibration values within the device, the ground truth is established through:

  • Metrological Traceability: Known quantities of morphine, traceable to an accepted reference method like Gas Chromatography/Mass Spectrometry (GC/MS), were added to human urine to achieve specific drug analyte concentrations. This serves as the "ground truth" for the calibrator's assigned values.
  • The "clinical significance" section reinforces this: "They are derived by adding known quantities (traceable to Gas Chromatography/Mass Spectrometry) of morphine (for opiate) to human urine to achieve each drug analyte concentration."

8. The Sample Size for the Training Set

This information is not applicable to this type of device. Calibrators are not "trained" in the way AI algorithms are. Their formulation and target values are determined by established analytical chemistry principles and traceability to reference methods.

9. How the Ground Truth for the Training Set Was Established

Not Applicable. As there is no "training set" in the AI sense, there is no ground truth established for it. The ground truth for the calibrator's values is established as described in point 7 (metrological traceability to GC/MS).


In summary: The provided document pertains to an IVD calibrator seeking substantial equivalence to a predicate device. The information requested aligns with assessing AI diagnostic devices, which is a different category. Therefore, many of the questions are not directly applicable to a calibrator submission. The critical information for this device is its analytical performance and traceability to reference standards, which underpinned the FDA’s substantial equivalence determination.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the text 'K993022' in a handwritten style. The text appears to be a code or identifier, with a combination of letters and numbers. The characters are bold and slightly slanted, giving the impression of quick handwriting. The background is plain white, which makes the text stand out.

Image /page/0/Picture/2 description: The image shows the logo for Beckman Coulter. The logo consists of a stylized, abstract symbol on the left, resembling a rounded shape with two curved lines running through it. To the right of the symbol, the words "BECKMAN" are stacked above the word "COULTER" in a bold, sans-serif font. The logo is black and white.

Summary of Safety and Effectiveness SYNCHRON® Systems OP 300 Low and High Urine Calibrators

1.0 Submitted By:

Gail Lefebvre Associate Regulatory Specialist, Product Submissions Beckman Coulter, Inc. 200 S. Kraemer Blvd., W-104 Brea, California 92822-8000 Telephone: (714) 993-8503 FAX: (714) 961-4123

2.0 Date Submitted:

September 7, 1999

3.0 Device Name(s):

3.1 Proprietary Names

SYNCHRON® Systems OP 300 Low and High Urine Calibrators

3.2 Classification Name

Clinical Toxicology Calibrator (21 CFR § 862.3200)

4.0 Predicate Device(s):

SYNCHRON SystemsReagentPredicateManufacturerDocketNumber
SYNCHRON® SystemsOP 300 Low and HighUrine CalibratorsSYNCHRON®Systems DAT Lowand High UrineCalibrators IIBeckman Coulter, Inc.K983747

5.0 Description:

The SYNCHRON® Systems OP 300 Low and High Urine Calibrators are intended for use on SYNCHRON Systems for the calibration of Opiate 300 ng enzyme immunoassays. This product contains a 5.0 mL bottle of the Low Urine Calibrator and a 5.0 mL bottle of the High Urine Calibrator. The storage temperature for the calibrators is +2°C to +8°C.

Mailing Address: 200 S. Kraemer Boulevard P.O. Box 8000 Brea, CA 92822-8000

Telephone: (714) 993-5321 Facsimile: (714) 961-4165 Internet: www.beckmancoulter.com

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, overlaid on three human profiles. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

NOV - 9 1999

Ms. Gail Lefebvre Associate Regulatory Affairs Specialist Beckman Coulter, Inc. 200 S. Kraemer Blvd., M/S W-104 Brea. California 92822-8000

  • Re: K993022
    Trade Name: SYNCHRON® Systems OP 300 Low and High Urine Calibrators Regulatory Class: II Product Code: DLJ Dated: September 7, 1999 Received: September 9, 1999

Dear Ms. Lefebvre:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page of of

510(k) Number (if known): K993022

Device Name: SYNCHRON® Systems OP 300 Low and High Urine Calibrators

Indications for Use:

The Beckman Coulter OP 300 Low and High Urine Calibrators, in conjunction with SYNCHRON Reagents, are intended for use on SYNCHRON Systems for the calibration of Opiate 300 ng (Part No. 475024) enzyme immunoassays.

Clinical Significance:

The OP Low and High Urine Calibrators are ready-to-use human, urine-based. liquid calibrators. They are derived by adding known quantities (traceable to Gas Chromatography/Mass Spectrometry) of morphine (for opiate) to human urine to achieve each drug analyte concentration.

Jean Cooper
(Division Sign-Off)
Division of Clinical Labora
ces
510(k) Number K993022

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(per 21 CFR 801.109)

OR

Over-the-Counter Use Optional Format 1-2-96

§ 862.3200 Clinical toxicology calibrator.

(a)
Identification. A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.