(263 days)
Brachytherapy planning, utilizing a 3D ultrasound imaging cube, with Sonoaraphic Planning of Oncology Treatments (SPOT) is intended for use with brachytherapy procedures, i.e. intercavitary, interstitial, intraluminal, involving manual or remote afterloading radioactive sources. The software program provides the physician with anatomical and dosimetric information to determine the positioning and loading of the radioactive sources prior to insertion. The software also provides the treatment time and dose distribution for the specified loading. From this information the patient can be treated with radioactive sources.
Sonographic Planning of Oncology Treatments (SPOT) as described in this submission is a software package which utilizes the acquisition of 2D ultrasound images and 3D reconstruction of the ultrasound images for treatment planning of brachytherapy procedures. This software runs on a Windows NT workstation.
Nucletron Sonographic Planning of Oncology Treatments (SPOT) software initially acquires 2D ultrasound images and generates 3D ultrasound image processing for definition of the treatment volume. These images are utilized for reconstructing the target volume of the brachytherapy implant, needles placement, defining the location of the radioactive sources within the implant, identifying a reference point, i.e. dose points. anatomical points. Once the treatment volume is identified and the implant is reconstructed the prescription dose is defined and the software program calculates treatment data, including dose distributions and treatment/implantation time.
The brachytherapy treatment planning session allows the physician to evaluate the implant prior to insertion of the radioactive sources in order to determine the most optimal dose distribution within the treatment volume. Once the physician approves the treatment plan the implant is loaded with radioactive sources manually or via the Nucletron remote afterloading equipment. The Sonographic Planning of Oncology Treatments (SPOT) software can export the treatment data to the Nucletron remote afterloading system and the PLATO Brachytherapy Planning System via a network. The Sonographic Planning of Oncology Treatments (SPOT) software does not control the treatment unit it strictly exports treatment times and related information. The Nucletron remote afterloading system and the clinical staff verify this data input prior to treatment.
Nucletron' Sonographic Planning of Oncology Treatments (SPOT) software includes remote afterloading radioactive source brachytherapy treatment planning and manual loaded radioactive source brachytherapy treatment planning. The program provides a variety of plan evaluation tools to assist in generating the most optimal dose distribution, i.e. dose volume histograms, dose verification points, dose profiles, etc.
The provided document, a 510(k) summary for the Nucletron Sonographic Planning of Oncology Treatments (SPOT) device, does not contain the information requested regarding acceptance criteria and a study proving those criteria are met.
The document primarily focuses on establishing substantial equivalence to predicate devices, describing the device, its intended use, and technological considerations. It is a regulatory submission for premarket notification, not a report detailing performance studies against specific acceptance criteria.
Therefore, I cannot populate the table or answer the specific questions about acceptance criteria, study design, sample sizes, ground truth establishment, or multi-reader multi-case studies based on the provided text. The document states:
- d. Description: "Sonographic Planning of Oncology Treatments (SPOT) as described in this submission is a software package which utilizes the acquisition of 2D ultrasound images and 3D reconstruction of the ultrasound images for treatment planning of brachytherapy procedures."
- f. Summary of technological considerations: "The Sonographic Planning of Oncology Treatments (SPOT) software is substantially equivalent to the predicate devices. It combines the functionality of the defined predicate devices into one user interface for brachytherapy treatment planning."
These statements indicate a descriptive explanation and a claim of substantial equivalence, but they do not outline performance metrics or studies to validate them against acceptance criteria.
{0}------------------------------------------------
VA- L 1
Premarket Notification Nucletron Sonographic Planning of Oncology Treatments (SPOT) Date: July 1, 1999
Image /page/0/Picture/2 description: The image shows the word "Nucletron" in a bold, sans-serif font. To the left of the word is a square logo. Inside the square is a black circle with two white circles on either side of it. The logo and the word "Nucletron" are both black.
NUCLETRON B.V. Waardgelder 1 3905 TH Veenendaal P.O.Box 930 3900 AX Veenendaal The Netherlands Phone +31 318 557133 Fax +31 318 550485
Department of Health and Human Services Center of Devices and Radiological Health Office of Device Evaluation Pre-Market Notification section
510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
as required by section 807.92(c)
Submitter of 510(k) a.
| Company name: | Nucletron Corporation |
|---|---|
| Registration # | 1121753 |
| Address: | 7080 Columbia Gateway Drive |
| Columbia, MD 21046-2133 | |
| Contact Person: | Ralph E. Shuping |
| Regulatory Affairs Manager | |
| Phone:. | 410-312-4100 |
| Fax: | 410-312-4197 |
| Name: |
b. Device Name:
Trade/Proprietary Name: Common/Usual Name: Classification Name:
Sonographic Planning of Oncology Treatments (SPOT) Brachytherapy planning system Accessory to remote afterloader 21 CFR 892.5700 Class II.
Legally Marketed Predicate Device(s) C.
Our device is substantially equivalent to the legally marketed predicate devices cited in the table below.
| Manufacturer | Device | 510(k) # |
|---|---|---|
| Nucletron BV | PLATO Brachytherapy Planning System | K915562 |
| Nucletron BV | Nucletron Planning System | K911612 |
| LIS, Inc | LIS 6000A | K961403 |
{1}------------------------------------------------
Premarket Notification Nucletron Sonographic Planning of Oncology Treatments (SPOT) Date: July 1, 1999
d. Description
Sonographic Planning of Oncology Treatments (SPOT) as described in this submission is a software package which utilizes the acquisition of 2D ultrasound images and 3D reconstruction of the ultrasound images for treatment planning of brachytherapy procedures. This software runs on a Windows NT workstation.
Nucletron Sonographic Planning of Oncology Treatments (SPOT) software initially acquires 2D ultrasound images and generates 3D ultrasound image processing for definition of the treatment volume. These images are utilized for reconstructing the target volume of the brachytherapy implant, needles placement, defining the location of the radioactive sources within the implant, identifying a reference point, i.e. dose points. anatomical points. Once the treatment volume is identified and the implant is reconstructed the prescription dose is defined and the software program calculates treatment data, including dose distributions and treatment/implantation time.
The brachytherapy treatment planning session allows the physician to evaluate the implant prior to insertion of the radioactive sources in order to determine the most optimal dose distribution within the treatment volume. Once the physician approves the treatment plan the implant is loaded with radioactive sources manually or via the Nucletron remote afterloading equipment. The Sonographic Planning of Oncology Treatments (SPOT) software can export the treatment data to the Nucletron remote afterloading system and the PLATO Brachytherapy Planning System via a network. The Sonographic Planning of Oncology Treatments (SPOT) software does not control the treatment unit it strictly exports treatment times and related information. The Nucletron remote afterloading system and the clinical staff verify this data input prior to treatment.
Nucletron' Sonographic Planning of Oncology Treatments (SPOT) software includes remote afterloading radioactive source brachytherapy treatment planning and manual loaded radioactive source brachytherapy treatment planning. The program provides a variety of plan evaluation tools to assist in generating the most optimal dose distribution, i.e. dose volume histograms, dose verification points, dose profiles, etc.
Intended use e.
Brachytherapy planning, utilizing a 3D ultrasound imaging cube, with Sonoaraphic Planning of Oncology Treatments (SPOT) is intended for use with brachytherapy procedures, i.e. intercavitary, interstitial, intraluminal, involving manual or remote afterloading radioactive sources. The software program provides the physician with anatomical and dosimetric information to determine the positioning and loading of the radioactive sources prior to insertion. The software also provides the treatment time and dose distribution for the specified loading. From this information the patient can be treated with radioactive sources.
{2}------------------------------------------------
Premarket Notification Nucletron Sonographic Planning of Oncology Treatments (SPOT) Date: July 1, 1999
With remote afterloading brachytherapy treatment planning the Sonographic Planning of Oncology Treatments (SPOT) program exports the treatment data, i.e. treatment times to the remote afterloader. With manual brachytherapy treatment planning the program exports treatment data regarding, source loading, total dose in permanent implantation and dose distributions for patient records. With both types of software the dose grid is exported and can be evaluated in conjunction with the Nucletron PLATO treatment planning software modules.
Summary of technological considerations f.
The Sonographic Planning of Oncology Treatments (SPOT) software is substantially equivalent to the predicate devices. It combines the functionality of the defined predicate devices into one user interface for brachytherapy treatment planning.
J. Kimble-
Name J. Kindlein Title Product Manager Nucletron BV Veenendaal The Netherlands
July 1, 1999
Date
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight, with three stylized human profiles incorporated into the design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 7 2000
Ralph Shuping, Sc.D. Regulatory Affairs Manager Nucletron Corporation 7080 Columbia Gateway Drive Columbia, MD 21046
Re:
K992303 Sonographic Planning or Oncology Treatments (SPOT) Dated: January 18, 2000 Received: January 18, 2000 Regulatory class: II 21 CFR 892.5700/Procode: 90 MUJ
Dear Dr. Shuping:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in Interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Aproval), it may be subject to such additional controls. Existing major requirecting vour device can be found in the Code of Federal Requirations. Title 21. Parts 800 to 895. A substantially equivation assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-5597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Daniel G. Schultz, M.D.
Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
Statement of indications for use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
Brachytherapy planning, utilizing 3D Ultrasound Imaging, Sonographic Planning of Oncology Treatments (SPOT) is intended for use with Brachytherapy procedures, i.e. intercavitary, interstitial, intraluminal, involving manual or remote afterloading radioactive sources. The software program provides the physician with anatomical and dosimetric information to determine the positioning and loading of the radioactive sources prior to insertion. From this information the patient can be treated with radioactive sources.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | |
|---|---|
| --------------------------------------- | -- |
OR
| Over the Counter Use | |
|---|---|
| ---------------------- | -- |
(Optional Format 1-2-96)David A. Segner
Division Sign Off
(Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological Devic 510(k) Number
N/A