K Number
K992256
Date Cleared
2000-01-11

(189 days)

Product Code
Regulation Number
868.5690
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is for use as a Positive Expiratory Pressure (PEP) device and Inspiratory Muscle Trainer (IMT) in one device, for use by adults and children. The device is for single patient use, not for use on multiple patients.

Device Description

The breathing trainer includes an inner chamber, an outer chamber, a breathing tube, and a perforated bottom cap to the inner chamber. Four teaspoons of water (20 ml.) are placed in the outer chamber and the patient breaths in and out of the breathing tube. The training effect is caused by the added pressure of the water. US Patent Number 5,755,640.

AI/ML Overview

The Frolov's Respiration Training Device, FRTD-01, is a breathing training device intended for use as a Positive Expiratory Pressure (PEP) device and Inspiratory Muscle Trainer (IMT).

1. Acceptance Criteria and Reported Device Performance:

The provided document does not explicitly state quantitative acceptance criteria for the Frolov's Respiration Training Device, FRTD-01. Instead, the submission relies on demonstrating substantial equivalence to a legally marketed predicate device, the Mini-Ciser (K974848).

Therefore, the reported device performance is implicitly that it is "as safe and effective as the predicate device," meaning it functions comparably in its intended use as a PEP and IMT device. Without specific performance metrics for the predicate device within this document, we cannot provide quantitative values for FRTD-01's performance.

Acceptance CriteriaReported Device Performance
Safety"as safe as the predicate device"
Effectiveness"as effective as the predicate device"
Design/Function"similar in design and function to the Mini-Ciser, K974848"

2. Sample size used for the test set and the data provenance:

The document mentions "bench and user testing" but does not specify the sample size for either the bench testing or the user testing. The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the supplied documentation. The study appears to focus on comparative effectiveness to a predicate device rather than establishing a new "ground truth" for the a medical condition using expert review.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not provided in the supplied documentation.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable as the device is a mechanical breathing trainer and not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This is not applicable as the device is a mechanical breathing trainer and does not involve an algorithm. The "user testing" would inherently involve human users.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

Given the nature of the device (a breathing trainer) and the lack of specific clinical trial data, the "ground truth" for its effectiveness and safety would likely be established through:

  • Bench testing: Verifying the physical and functional specifications of the device (e.g., pressure output, air resistance).
  • User testing: Assessing patient comfort, ease of use, and potentially physiological responses consistent with PEP and IMT, as compared to the predicate device.
  • Comparison to predicate device: The primary "truth" is that its performance matches that of the legally marketed predicate device, which would have had its own "ground truth" established during its clearance process.

8. The sample size for the training set:

This information is not provided. It's important to note that for a mechanical device, the concept of a "training set" in the context of machine learning is not directly applicable. If "training set" refers to data used to refine the device's design, that is not detailed.

9. How the ground truth for the training set was established:

Not applicable in the context of device design and development as described in this 510(k) summary. The "ground truth" for guiding the design process would be engineering specifications, existing scientific understanding of respiratory physiology, and performance data from the predicate device.

{0}------------------------------------------------

K992256

EXHIBIT 2 Intercontinental New Technologies, Inc. 1389 Bristol Lane Buffalo Grove, IL 60089 Tel. 847-955-0311 Fax 847-955-0311 Contact: Aleksey Shlyakhta, President June 30, 1999

510(k) Summary of Safety and Effectiveness

  • a) Identification of the Device: Proprietary-Trade Name: Frolov's Respiration Training Device, FRTD-01 Classification Name: 73BWF Common/Usual Name: Breathing training device
  • b) Equivalent legally marketed devices This product is similar in design and function to the Mini-Ciser, K974848
  • c) Indications for Use (intended use) The device is for use as a Positive Expiratory Pressure (PEP) device and Inspiratory Muscle Trainer (IMT) in one device.
  • d) Description of the Device: The breathing trainer includes an inner chamber, an outer chamber, a breathing tube, and a perforated bottom cap to the inner chamber. Four teaspoons of water (20 ml.) are placed in the outer chamber and the patient breaths in and out of the breathing tube. The training effect is caused by the added pressure of the water. US Patent Number 5,755,640.
  • e) Safety and Effectiveness, comparison to predicate device. The results of bench and user testing indicates that the new device is as safe and effective as the predicate device.
  • f) Conclusion. After analyzing both bench and user testing data, it is the conclusion of Intercontinental New Technologies, Inc. that the Frolov's Respiration Training Device, FRTD-01 is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the border of the circle. Inside the circle is a stylized image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 11 2000

Mr. Daniel Kamm Intercontinental New Technologies, Inc. c/o Kamm & Associates P.O. Box 7007 Deerfield, IL 60015

Re: K992256 Frolov's Respiration Training Device, FRTD-01 Regulatory Class: II (two) Product Code: 73 BWF October 13, 1999 Dated: October 14, 1999 Received:

Dear Mr. Kamm:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Mr. Daniel Kamm

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Joanne A. Wintershaven

Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

j) Indications for Use

510(k) Number K992256

Device Name: Frolov's Respiration Training Device, FRTD-01

Indications for Use: The device is for use as a Positive Expiratory Pressure (PEP) device and Inspiratory Muscle Trainer (IMT) in one device, for use by adults and children. The device is for single patient use, not for use on multiple patients.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

Over the Counter Use (Per 21 CFR 801.109)

J.A. Westman

OR

Division Sign-Off) ivision of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number

§ 868.5690 Incentive spirometer.

(a)
Identification. An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.(b)
Classification. Class II (performance standards).