ETI-EA-G ASSAY

K992191 · DiaSorin, Inc. · LSE · Jul 12, 1999 · Microbiology

Device Facts

Record IDK992191
Device NameETI-EA-G ASSAY
ApplicantDiaSorin, Inc.
Product CodeLSE · Microbiology
Decision DateJul 12, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3235
Device ClassClass 1
AttributesPediatric, 3rd-Party Reviewed

Intended Use

The DiaSorin ETI-EA-G kit uses Enzyme Linked Immunosorbent Assay (ELISA) technology for the qualitative and/or semi-quantitative detection of IgG antibodies to the EBV Early Antigen Diffuse. The assay is designed for human serum. The presence of EA antibodies is used as an aid in the diagnosis of EBV associated infectious mononucleosis when used in conjunction with other EBV serologies in pediatric, adult, transplant donor and transplant recipient populations. When evaluating properly paired sera, the results of these assays are used to demonstrate seroconversion or significant change in antibody titer as evidence of recent infection. Both specimens should be tested simultaneously.

Device Story

DiaSorin ETI-EA-G is an ELISA-based in vitro diagnostic kit for detecting IgG antibodies to Epstein-Barr Virus (EBV) Early Antigen Diffuse (EA(D)). Input is human serum; assay utilizes polystyrene microtiter wells coated with recombinant EA(D) antigen. Patient serum is incubated; captured IgG antibodies are detected via goat anti-human IgG conjugated to horseradish peroxidase. Addition of tetramethylbenzidine chromogen produces a color reaction measured by photometer at 450 nm. Enzyme activity is directly proportional to specific anti-EA(D) IgG concentration. Used in clinical laboratories by trained personnel. Results assist clinicians in diagnosing EBV-associated infectious mononucleosis and identifying recent infection via paired sera analysis. Benefits include standardized, objective serological assessment compared to traditional immunofluorescence assays.

Clinical Evidence

Clinical trials evaluated performance across primary, seronegative, reactivated, and past infection populations (adults and pediatrics). Sensitivity ranged from 36.4% to 100% depending on disease state; specificity ranged from 33.3% to 100%. Agreement with predicate IFA was high in acute/reactivated states. Discrepancies in 'Past Infection' populations were explained by the specific detection of Diffuse vs. Restricted antigens. Reproducibility studies (3 sites, 6 samples) showed total %CV ranging from 6.64% to 21.63%.

Technological Characteristics

ELISA-based immunoassay. Components: Polystyrene microtiter wells coated with recombinant EA(D) antigen; affinity-purified polyclonal goat anti-human IgG (Fc) conjugated to horseradish peroxidase; tetramethylbenzidine chromogen. Detection via 450 nm photometer. Manual or automated plate processing. No software algorithm; results calculated via photometric absorbance.

Indications for Use

Indicated for qualitative/semi-quantitative detection of IgG antibodies to EBV Early Antigen Diffuse in human serum. Used as an aid in diagnosing EBV-associated infectious mononucleosis in pediatric, adult, transplant donor, and transplant recipient populations. Used with other EBV serologies to demonstrate seroconversion or titer changes indicating recent infection.

Regulatory Classification

Identification

Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUL 12 1560 # K992191 ## 510 (k) SUMMARY SUBMITTED BY: Judith J. Smith DiaSorin, Inc. 9175 Guilford Rd. Suite 100 Columbia, MD 21046 NAME OF DEVICES: Trade Name: DiaSorin ETI-EA-G Epstein-Barr Viral Early Antigen Diffuse EA (D) IaG ELISA Immunoassay for the detection of IgG Common Names/Descriptions: antibodies to the Epstein-Barr Viral (EBV) Early Antigen Diffuse antigen Classification Names: EBV Serology Test PREDICATE DEVICES: DiaSorin EBV EA (D) IgG Clin-ELISA™ Gull Laboratories EBV EA IgG IFA DEVICE DESCRIPTION: INTENDED USE: The DiaSorin ETI-EA-G kit uses Enzyme Linked Immunosorbent Assay (ELISA) technology for the qualitative and/or semi-quantitative detection of IgG antibodies to the EBV Early Antigen Diffuse. The assay is designed for human serum. The presence of EA antibodies is used as an aid in the diagnosis of EBV associated infectious mononucleosis when used in conjunction with other EBV serologies in pediatric, adult, transplant donor and transplant recipient populations. When evaluating properly paired sera, the results of these assays are used to demonstrate seroconversion or significant change in antibody titer as evidence of recent infection. Both specimens should be tested simultaneously. KIT DESCRIPTION: The method for the determination of specific anti-EA (D) IgG utilizes the enzyme-linked immunosorbent assay (ELISA) technique. Polystyrene microtiter wells are coated with recombinant EA (D) antigen. Diluted patient serum is incubated with the purified antigen bound to the solid surface of a microtiter well. The EA (D) IgG antibodies present in a patient's serum will be captured by the solid phase. After washing, affinity purified polyclonal goat anti-human IgG (Fc) antibodies coniugated to horseradish peroxidase are added to the well. After this incubation, chromogen containing tetramethylbenzidine is added. Enzyme action on the chromogen results in a color reaction. The color can be detected with a photometer at a wavelength of 450 nm. The measured enzyme activity is directly proportional to the concentration of specific anti-EA (D) IgG bound to the solid phase. {1}------------------------------------------------ #### PERFORMANCE DATA: Comparative Clinical Trials Clinical trials were conducted at 1 clinical laboratory to evaluate the performance of ETI-EA-G in detecting IgG antibodies to EBV EA (D). Assay performance was compared to the Gull Laboratories, Inc. EA IgG IFA test. Patients from the disease states defined below were tested. The screening population is a group of samples from patients suspected of disease. The transplant recipients were patients who had received, or were awaiting, solid organ or bone marrow transplants. Patient samples excluded if they failed to fit a recognized EBV marker pattern determined by EBV VCA (IgG, IgM) and EBNA testing. Results are shown below. | Expected Pattern | + | |---------------------|----------------| | Sensitivity | 45/46 = 97.8% | | (95% CI) | (88.5 - 99.9%) | | Specificity | N/A | | (95% CI) | | | Agreement | 45/46 = 97.8% | | Prevalence ELISA | 45/46 = 97.8% | | Prevalence IFA | 46/46 = 100% | | Expected Prevalence | 80-100% | Primary Disease State: Adult Population Primary Disease State; Pediatric Population | Expected Pattern | + | |---------------------|----------------| | Sensitivity | 26/29 = 89.7% | | (95% CI) | (72.6 – 97.8%) | | Specificity | N/A | | (95% CI) | | | Agreement | 26/29 = 89.7% | | Prevalence ELISA | 26/29 = 89.7% | | Prevalence IFA | 29/29 = 100% | | Expected Prevalence | 80-100% | {2}------------------------------------------------ # Seronegative; Adult Population | Expected Pattern | - | |---------------------|-----------------| | Sensitivity | N/A | | (95% CI) | | | Specificity | 7/7 = 100.0% | | (95% CI) | (59.0 - 100.0%) | | Agreement | 7/7 = 100.0% | | Prevalence ELISA | 0/7 = 0.0% | | Prevalence IFA | 0/7 = 0.0% | | Expected Prevalence | 0% | : : ## Seronegative; Pediatric Population | Expected Pattern | - | |---------------------|----------------| | Sensitivity | N/A | | (95% CI) | | | Specificity | 48/50 = 96.0% | | (95% CI) | (86.3 - 99.5%) | | Agreement | 48/50 = 96.0% | | Prevalence ELISA | 2/50 = 4.0% | | Prevalence IFA | 0/50 = 0.0% | | Expected Prevalence | 0% | ## Reactivated; Adult Population | Expected Pattern | + | |---------------------|---------------| | Sensitivity | 30/31 = 96.8% | | (95% CI) | (83.3-99.9%) | | Specificity | N/A | | (95% Cl) | N/A | | Agreement | 30/31 = 96.8% | | Prevalence ELISA | 30/31 = 96.8% | | Prevalence IFA | 31/31 = 100% | | Expected Prevalence | 90-100% | {3}------------------------------------------------ Reactivated; Pediatric Population | Expected Pattern | + | |---------------------|-----------------| | Sensitivity | 4/4 = 100.0% | | (95% CI) | (39.8 - 100.0%) | | Specificity | N/A | | (95% CI) | | | Agreement | 4/4 = 100.0% | | Prevalence ELISA | 4/4 = 100.0% | | Prevalence IFA | 4/4 = 100.0% | | Expected Prevalence | 90-100% | . . Past Infection; Adult Population : | Expected Pattern | -/+ | |---------------------|----------------| | Sensitivity | 12/33 = 36.4% | | (95% CI) | (20.4 - 54.9%) | | Specificity | 3/4= 75.0% | | (95% CI) | (19.4-99.4) | | Agreement | 15/37 = 40.5% | | Prevalence ELISA | 13/37 = 32.4% | | Prevalence IFA | 33/37 = 89.2% | | Expected Prevalence | 10-40% | Past Infection; Pediatric Population | Expected Pattern | -/+ | |---------------------|--------------| | Sensitivity | 2/3 = 66.7% | | (95% CI) | (9.0 - 99.9) | | Specificity | N/A | | (95% CI) | | | Agreement | 2/3 = 66.7% | | Prevalence ELISA | 2/3 = 66.7% | | Prevalence IFA | 3/3 = 100.0% | | Expected Prevalence | 10-40% | {4}------------------------------------------------ | ELISA Result | EA (D) IgG | |------------------|----------------| | Sensitivity | 10/21 = 47.6% | | (95% CI) | (25.7 - 70.2%) | | Specificity | 1/3 = 33.3% | | (95% CI) | (0.8% - 90.6%) | | Agreement | 11/24 = 45.8% | | Prevalence ELISA | 12/24 = 50.0% | | Prevalence IFA | 21/24 = 87.5% | Transplant Recipient patients: Adult Population Transplant Recipient patients; Pediatric Population | ELISA Result | EA (D) IgG | |-------------------------|-------------------------------| | Sensitivity<br>(95% CI) | N/A | | Specificity<br>(95% CI) | 6/7 = 85.7%<br>(42.1 - 99.6%) | | Agreement | 6/8 = 75.0% | | Prevalence ELISA | 1/8 = 12.5% | | Prevalence IFA | 1/8 = 12.5% | The ETI-EA-G ELISA demonstrates good sensitivity and specificity in Primary. Seronegative and Reactivated populations, The prevalence rates were within the expected range for each of these populations. In the Past Infection population the prevalence rate for ETI-EA-G was consistent with published rates of 10-40%, while IFA rates were 90%. This appears to reflect the fact that the IFA detects both the Restricted and the Diffuse EA antigen, while ETI-EA-G detects only the Diffuse antigen. The Restricted antigen is only occasionally seen in acute IM cases, and then only late in the acute phase. Following IM, however, the Restricted antigen persists in a substantial number of patients for at least 10-104 months. In light of this, it is predictable that the Gull IFA would detect Early Antigen at a much higher rate in a Past Infection population than would the ETI-EA-G ELISA. The same situation appears to account for the results seen in the Transplant Recipients who included a high percentage of samples with EBV markers consistent with Past Infection. The results support the Transplant claim in the Intended Use. Within each of the defined disease states, results were similar across the adult and pediatric populations, supporting the separate adult and pediatric claims in the Intended Use for ETI-EA-G. Reproducibility: Reproducibility studies were performed at 3 sites using one lot of ETI-EA-G reagents. Assay reproducibility was determined using 6 samples that spanned {5}------------------------------------------------ the range of the assay. Samples were tested in triplicate once a day for 3 days. Combined results are summarized below. | | | Within-run | Between day | Between site | Total | |----------|-----------|------------|-------------|--------------|-------| | Sample | Mean (AU) | %CV | %CV | %CV | %CV | | Negative | 7.6 | 12.14 | 22.02 | 14.77 | 21.63 | | Negative | 9.2 | 14.24 | 24.19 | 20.01 | 19.61 | | low pos | 29.4 | 7.57 | 4.97 | 11.37 | 8.18 | | low pos | 26.0 | 13.47 | 12.32 | 9.59 | 15.88 | | mid pos | 74.5 | 9.71 | 2.68 | 3.36 | 10.12 | | high pos | 122.4 | 6.74 | 3.06 | 0.98 | 6.64 | Reproducibility for ETI-EA-G based on Arbitrary Units (AU) � {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure or a bird in flight, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1.2 1999 DiaSorin, Inc. c/o Ms. Carole Stamp TÜV Product Service, Inc. 1775 Old Highway 8 NW Suite 104 New Brighton, MN 55112-1891 Re: K992191 Trade Name: DiaSorin ETI-EA-G Regulatory Class: I Product Code: LSE Dated: June 25, 1999 Received: June 28, 1999 Dear Ms. Stamp: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {7}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ## INDICATIONS FOR USE 510(k) Number (if known): Not known Device Name: DiaSorin ETI-EA-G Indications For Use: The DiaSorin ETI-EA-G kit uses Enzyme Linked Immunosorbent Assay (ELISA) technology for the qualitative and/or semi-quantitative detection of IgG antibodies to the EBV Early Antigen Diffuse. The assay is designed for human serum. The presence of EA antibodies is used as an aid in the diagnosis of EBV associated infectious mononucleosis when used in conjunction with other EBV serologies in pediatric, adult, transplant donor and transplant recipient populations. When evaluating properly paired sera, the results of these assays are used to demonstrate seroconversion or significant change in antibody titer as evidence of recent infection. Both specimens should be tested simultaneously. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Woody Dubois ical I aboratory Devices 510(k) Number Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%