K Number
K991961

Validate with FDA (Live)

Device Name
3M LVR SYSTEM
Manufacturer
Date Cleared
1999-08-09

(60 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is indicated for:

  • Base/liner under direct restorations .
  • Pit and fissure sealant .
  • Veneer Cementation .
  • Undercut blockout ●
  • Restoration of minimal, shallow, Class V preparations ●
  • Restoration of minimally invasive cavity preparations, including air abrasion and tunnel . preparations
  • . Repair of small defects in composite and ceramic indirect restorations including veneers, inlays, onlays, crowns and bridges
  • . Repair of small marginal defects in direct restorations
  • Repair of acrylic temporary marginal defects .
Device Description

3M™ LVR System is a single part, light curing restorative composite. This device, as well as the predicate devices, are based on methacrylate resin chemistry incorporating an inorganic filler.

AI/ML Overview

The provided text describes the 3M™ LVR System, a light-curing restorative composite, and its 510(k) submission for market clearance. However, it does not include the detailed information needed to fill out all sections of your request regarding acceptance criteria and a study proving the device meets those criteria, especially in the context of an AI/human-in-the-loop system.

The submission is for a dental material, not an AI-powered diagnostic or assistive device. Therefore, many of your requested fields, such as "multi reader multi case (MRMC) comparative effectiveness study," "human readers improve with AI vs without AI assistance," "standalone (i.e. algorithm only without human-in-the-loop performance)," and specifics about training and test sets common in AI studies, are not applicable to this type of device.

The "study" mentioned for this device is a series of bench tests comparing its physical and chemical properties to predicate devices and adherence to an ISO standard.

Here's an attempt to extract and present the relevant information based on the provided text, acknowledging the limitations for an AI-focused analysis:

Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance

The document states that the 3M™ LVR System "meets ISO 4049-1988 'Dentistry - Resin-based filling materials'" and that bench tests were conducted. The acceptance criteria are implicitly those defined by the ISO standard and the performance of the predicate devices. The reported device performance is that it met these criteria.

Acceptance Criteria CategorySpecific Criteria (Implicitly from ISO 4049-1988 and Predicate Device Performance)Reported Device Performance
Physical/Chemical PropertiesWatts shrinkage (comparable or better than predicates)Met (comparable results to predicate devices)
Wear resistance (comparable or better than predicates)Met (comparable results to predicate devices)
Compressive strength (comparable or better than predicates)Met (comparable results to predicate devices)
Diametral tensile strength (comparable or better than predicates)Met (comparable results to predicate devices)
Film thickness (comparable or better than predicates)Met (comparable results to predicate devices)
Thermal Expansion Coefficient (comparable or better than predicates)Met (comparable results to predicate devices)
Flow (comparable or better than predicates)Met (comparable results to predicate devices)
Water sorption (comparable or better than predicates)Met (comparable results to predicate devices)
International StandardCompliance with ISO 4049-1988 "Dentistry - Resin-based filling materials"Device "meets ISO 4049-1988"
Safety and EffectivenessSafe, effective, and performs as well or better than predicate devicesConcluded to be "safe, effective and performs as well or better than the predicate devices"

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not specified for each individual bench test. Bench tests typically involve a defined number of samples per test, but the exact count is not provided in this summary.
  • Data Provenance: The text does not specify the country of origin for the data. The tests were presumably conducted internally by 3M Dental Products Laboratory. The data is retrospective in the sense that the tests were completed before the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. For a dental material, "ground truth" is established by laboratory measurement equipment and adherence to defined physical and chemical standards (e.g., ISO methods) rather than expert consensus on diagnostic images.

4. Adjudication method for the test set

Not applicable. Adjudication methods (like 2+1, 3+1) are relevant for human-interpreted data, such as medical image diagnostics, not for laboratory bench tests of material properties.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a dental restorative material, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical product, not an algorithm.

7. The type of ground truth used

The ground truth used for the "study" (bench tests) is based on objective measurements of material properties (e.g., shrinkage, strength, wear) according to established laboratory protocols and adherence to an international standard (ISO 4049-1988). The performance of predicate devices also serves as a comparative "ground truth" for demonstrating substantial equivalence.

8. The sample size for the training set

Not applicable. This device is not an AI algorithm that requires a "training set." The materials themselves are developed through R&D, not trained on data in the AI sense.

9. How the ground truth for the training set was established

Not applicable. See point 8.

{0}------------------------------------------------

AUG. 2. 1999 1:37PM

AUG -- 9 1999

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1980 and 21 CFR $807.92.

Name:Lael J. Pickett
Regulatory Affairs Specialist
Address:3M Dental Products Laboratory
3M Center, Building 260-2B-12
St. Paul, MN 55144-1000
Telephone:612-733-3594
Fax:612-736-0990
Trade Name:3M™ LVR System
Common Names:Tooth shade resin material, light cure restorative
Resin cement, luting composite, luting cement
Dental sealant, pit and fissure sealant.
Classification Name:21 CFR § 872.3690 Tooth Shade Resin Material
21 CFR § 872.3275 Dental cement, Class II.
21 CFR § 872.3765 Pit and fissure sealant, Class II
Predicate Devices:DYRACT® FLOW, DENTSPLY Caulk
AELITEFLO™, Bisco Dental Products
Revolution Flowable Paste Product, Kerr Corp.
TETRIC® FLOW Flowable Ceromer, Ivoclar Vivadent

3M™ LVR System is a single part, light curing restorative composite. This device, as well as the predicate devices, are based on methacrylate resin chemistry incorporating an inorganic filler.

The 3M™ LVR System can be used as a: baselliner under direct restorations, pit and fissure sealant, veneer cement, undercut blockout, restorative for minimal - shallow Class V preparations, restorative for minimally invasive cavity proparations (including air abrasion and tunnel preparations), repair of small defects in composite and ceramic indirect restorations (including veneers, inlays, crowns and bridges), repair of small marginal defects in direct restorations and for repair of acrylic temporary marginal defects. This product can be used for both anterior and posterior restorations and functions as a permanent veneer cement.

3M™ LVR System and predicate devices have similar technological characteriatios as indicated by their methacrylate resin chemistry. This is further validated by the comparative results of the bench tests conducted. These tests include Watts shrinkage, wear resistance, compressive strength, diametral tensile strength, film thickness, Thermal Expansion Coefficient, flow and water sorption.

The 3M™ LVR System also meets ISO 4049-1988 "Dentistry - Resin-based filling materials".

Based on the conclusions drawn from the safety analysis conducted for this device and the results of the bench testing, 3M™ LVR System II Plus is safe, effective and performs as well or better than the predicate devices mentioned above.

3M™ LVR System K991961 August 2, 1999

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG - 9 1999

Mr. Lael J. Pickett Regulatory Affairs Specialist 3M Dental Products Laboratory 3M Center, Building 260-2B-09 St. Paul, Minnesota 55144-1000

K991961 Re: 3MTM LVR SYSTEM Trade Name: Requlatory Class: II Product Code: EBF Dated: June 2, 1999 Received: June 10, 1999

Dear Mr. Pickett:

We have reviewed your Section 510(k) notification of intent to we have reviewed your below and we have determined the market the device relies suivalent (for the indications for use device is substancially equilies (marketed in interstate commerce scated in the Cholorals, is enactment date of the Medical Device prior to hay 20, 1970) one base been reclassified in accordance Amendicites) or oo of the Federal Food, Drug, and Cosmetic Act You may, therefore, market the device, subject to the (Act). (Acc). Tod may) choresons of the Act. The general controls general concrors provisions e requirements for annual registration, proving of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be (Spectur controls) Stronal controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Mr. Pickett

This letter will allow you to begin marketing your device as Inis lected will arrow you ou begdescribed in your stow, promise of your device to a legally marketed of Subscuncial equivalisin a classification for your device and thus, permits your device to proceed to the market. enab, Delmire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in xitro reguration (ar orn), please contact the Office of Compliance at (1801) 594-4692. Additionally, for questions on the promotion and (301) 394-4022. Addressmarly) ase contact the Office of Compliance Also, please note the regulation entitled, at (301) 594-4639. "Misbranding by reference to premarket notification" (21 CFR Other general information on your responsibilities under 807.97) . . our. The ocher general from the Rivision of Small Manufacturers ene may be obsolisfree number (800) 638-2041 or (301) 443-6597 Assistance ac Its Coll Irec namby (voo) (voo) (voo) cdrh/dsmamain.html").

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known): K991961

Device Name: 3M™ LVR System_

Indications For Use: This device is indicated for:

  • Base/liner under direct restorations .
  • Pit and fissure sealant .
  • Veneer Cementation .
  • Undercut blockout ●
  • Restoration of minimal, shallow, Class V preparations ●
  • Restoration of minimally invasive cavity preparations, including air abrasion and tunnel . preparations
  • . Repair of small defects in composite and ceramic indirect restorations including veneers, inlays, onlays, crowns and bridges
  • . Repair of small marginal defects in direct restorations
  • Repair of acrylic temporary marginal defects .

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use(Per 21 CFR 801.109)OROver-The-Counter Use
August 2, 1999(Division Sign-Off)Division of Dental, Infection Control,and General Hospital DevicesPage x
510(k) NumberK99961

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.