K Number
K991907

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-11-09

(155 days)

Product Code
Regulation Number
866.5240
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bayer ADVIA IMS C3 assay is an in vitro diagnostic device intended to measure Complement C3 (C3) in human serum. Such measurements are used as an aid in the diagnosis and treatment of immunologic disorders.

The Bayer ADVIA IMS C4 assay is an in vitro diagnostic device intended to measure Complement C4 (C4) in human serum. Such measurements are used as an aid in the diagnosis and treatment of immunologic disorders.

The Bayer ADVIA IMS IgA assay is an in vitro diagnostic device intended to measure Immunoglobulin A (IgA) in human serum. Such measurements are used as an aid in the diagnosis and treatment of abnormal protein metabolism and the body's inability to resist infectious agents.

The Bayer ADVIA IMS IgG assay is an in vitro diagnostic device intended to measure Immunoglobulin G (IgG) in human serum. Such measurements are used as an aid in the diagnosis and treatment of autoimmune diseases, chronic or recurrent infections and abnormal protein metabolism.

The Bayer ADVIA IMS Immunoglobulin M (IgM) assay is an in vitro diagnostic device intended to measure IgM in human serum. Such measurements are used as an aid in the diagnosis and treatment of chronic or recurrent infections and abnormal protein metabolism.

The Bayer ADVIA IMS Transferrin assay is an in vitro diagnostic device intended to measure transferrin (TRF) in human serum. Such measurements are used as an aid in the diagnosis and treatment of malnutrition, chronic infection, acute hepatitis, polycythemia, pernicious anemia, and red blood cell disorders, such as iron deficiency anemia.

The Bayer ADVIA IMS Vancomycin assay is an in vitro diagnostic device intended to measure vancomycin, an antibiotic drug, in human serum. Measurements of vancomycin are used as an aid in the diagnosis and treatment of vancomycin overdose and in monitoring therapeutic levels of vancomycin to ensure appropriate therapy.

The Bayer ADVIA IMS Valproic Acid Assay is an in vitro diagnostic device intended to measure valproic acid, an anti-epileptic drug, in human serum. Measurements of valproic acid are used as an aid in the diagnosis and treatment of valproic acid overdose and in monitoring therapeutic levels of valproic acid to ensure appropriate therapy.

Device Description

Not Found

AI/ML Overview

The provided text describes several immunoassay methods for the Bayer ADVIA IMS Systems, focusing on their performance compared to predicate devices. It does not explicitly state "acceptance criteria" for each test but rather presents performance data (precision, correlation, analytical range, interference) alongside a predicate device's performance. The study aims to demonstrate substantial equivalence to these predicate devices for FDA clearance.

Here's an analysis based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance:

Since explicit acceptance criteria are not stated, I will infer them as the performance of the predicate device (the "similar device that was granted clearance of substantial equivalence") for correlation and precision, and generally acceptable performance within an analytical range with minimal interference. The reported device performance is that of the ADVIA IMS method.

Metric (Implied Acceptance Criterion: Predicate Performance)ADVIA IMS (Reported Performance)
C3 Method (Predicate: Dade Behring BNA Complement C3 method)
Precision (Total) @ 72.4 mg/dL2.1% @ 79.7 mg/dL (Predicate: 5.9%)
Precision (Total) @ 237 mg/dL1.8% @ 154 mg/dL (Predicate: 2.1%)
Correlation (y=0.92x + 9.7, r=0.972, Syx=12.1 mg/dL)Matched to Predicate
Interfering Substances: Bilirubin unconjugated (20 mg/dL) at 106 mg/dL C3+1% Change (Predicate: Not specified, but implied acceptable for clearance)
Interfering Substances: Bilirubin conjugated (20 mg/dL) at 113 mg/dL C3+4% Change (Predicate: Not specified)
Interfering Substances: Hemoglobin (500 mg/dL) at 114 mg/dL C3-2% Change (Predicate: Not specified)
Interfering Substances: Triglycerides (1000 mg/dL) at 107 mg/dL C3-8% Change (Predicate: Not specified)
Analytical Range (Normal: 27-360 mg/dL)Matched to Predicate
C4 Method (Predicate: Dade Behring BNA Complement C4 method)
Precision (Total) @ 15.7 mg/dL1.9% @ 19.3 mg/dL (Predicate: 2.9%)
Precision (Total) @ 34.8 mg/dL2.0% @ 34.8 mg/dL
Precision (Total) @ 49.3 mg/dL1.9% @ 49.3 mg/dL
Correlation (y=0.92x - 0.3, r=0.989, Syx=1.9 mg/dL)Matched to Predicate
Interfering Substances: Bilirubin unconjugated (20 mg/dL) at 19 mg/dL C40% Change (Predicate: Not specified)
Interfering Substances: Bilirubin conjugated (20 mg/dL) at 21 mg/dL C40% Change (Predicate: Not specified)
Interfering Substances: Hemoglobin (500 mg/dL) at 20 mg/dL C40% Change (Predicate: Not specified)
Analytical Range (Normal: 7.2-96 mg/dL)Matched to Predicate
IgA Method (Predicate: Dade Behring BNA Immunoglobulin A method)
Precision (Total) @ 296 mg/dL2.3% @ 122 mg/dL (Predicate: 3.5%)
Precision (Total) @ 233 mg/dL1.6% @ 233 mg/dL
Precision (Total) @ 344 mg/dL1.4% @ 344 mg/dL
Correlation (y=1.00x - 1, r=0.994, Syx=75.6 mg/dL)Matched to Predicate
Interfering Substances: Bilirubin unconjugated (20 mg/dL) at 180 mg/dL IgA+2% Change (Predicate: Not specified)
Interfering Substances: Bilirubin conjugated (20 mg/dL) at 200 mg/dL IgA-3% Change (Predicate: Not specified)
Interfering Substances: Hemoglobin (500 mg/dL) at 191 mg/dL IgA+1% Change (Predicate: Not specified)
Interfering Substances: Triglycerides (1000 mg/dL) at 184 mg/dL IgA-8% Change (Predicate: Not specified)
Analytical Range (Normal: 45-600 mg/dL)Matched to Predicate
IgG Method (Predicate: Dade Behring BNA Immunoglobulin G method)
Precision (Total) @ 2141 mg/dL1.8% @ 772 mg/dL (Predicate: 3.4%)
Precision (Total) @ 1317 mg/dL2.7% @ 1317 mg/dL
Precision (Total) @ 1416 mg/dL1.6% @ 1416 mg/dL
Correlation (y=0.97x - 3, r=0.994, Syx=118 mg/dL)Matched to Predicate
Interfering Substances: Bilirubin unconjugated (20 mg/dL) at 876 mg/dL IgG-3.0% Change (Predicate: Not specified)
Interfering Substances: Bilirubin conjugated (20 mg/dL) at 953 mg/dL IgG-3.0% Change (Predicate: Not specified)
Interfering Substances: Hemoglobin (500 mg/dL) at 897 mg/dL IgG+7.0% Change (Predicate: Not specified)
Interfering Substances: Triglycerides (1000 mg/dL) at 887 mg/dL IgG-3.0% Change (Predicate: Not specified)
Analytical Range (Normal: 225-3000 mg/dL)Matched to Predicate
IgM Method (Predicate: Dade Behring BNA Immunoglobulin M method)
Precision (Total) @ 69.0 mg/dL3.6% @ 69.0 mg/dL (Predicate: 2.4% @ 128 mg/dL)
Precision (Total) @ 117 mg/dL1.9% @ 117 mg/dL
Precision (Total) @ 128 mg/dL2.4% @ 128 mg/dL
Precision (Total) @ 177 mg/dL1.5% @ 177 mg/dL
Correlation (y=1.05x - 13.4, r=0.990, Syx=123.5 mg/dL)Matched to Predicate
Interfering Substances: Bilirubin unconjugated (20 mg/dL) at 77 mg/dL IgM+3.9% Change (Predicate: Not specified)
Interfering Substances: Bilirubin conjugated (10 mg/dL) at 85 mg/dL IgM+1.2% Change (Predicate: Not specified)
Interfering Substances: Hemoglobin (500 mg/dL) at 86 mg/dL IgM0.0% Change (Predicate: Not specified)
Analytical Range (Normal: 30-400 mg/dL)Matched to Predicate
TRF Method (Predicate: Dade Behring BNA Transferrin method)
Precision (Total) @ 127 mg/dL2.2% @ 127 mg/dL (Predicate: 2.7% @ 303 mg/dL)
Precision (Total) @ 303 mg/dL2.7% @ 303 mg/dL
Precision (Total) @ 268 mg/dL1.9% @ 268 mg/dL
Precision (Total) @ 400 mg/dL3.1% @ 400 mg/dL
Correlation (y=0.96x - 3, r=0.979, Syx=16.8 mg/dL)Matched to Predicate
Interfering Substances: Bilirubin unconjugated (20 mg/dL) at 212 mg/dL TRF+2% Change (Predicate: Not specified)
Interfering Substances: Bilirubin conjugated (20 mg/dL) at 231 mg/dL TRF0% Change (Predicate: Not specified)
Interfering Substances: Hemoglobin (500 mg/dL) at 261 mg/dL TRF0% Change (Predicate: Not specified)
Interfering Substances: Triglycerides (1000 mg/dL) at 215 mg/dL TRF-4% Change (Predicate: Not specified)
Analytical Range (Normal: 54-720 mg/dL)Matched to Predicate
Vancomycin Method (Predicate: Bayer Immuno 1 Vancomycin method)
Analytical Range0.4-50 µg/mL (Predicate: Not explicitly stated, but implied similar operating range)
Precision (Total) @ 6.7 µg/mL2.8% @ 8.6 µg/mL (Predicate: 8.9%)
Precision (Total) @ 23.3 µg/mL3.4% @ 21.5 µg/mL (Predicate: 7.5%)
Precision (Total) @ 32.4 µg/mL3.9% @ 35.8 µg/mL (Predicate: 8.1%)
Correlation (y=1.02x + 0.68, r=0.987, Syx=1.6 µg/dL)Matched to Predicate
Interfering Substances: Bilirubin unconjugated (25 mg/dL) at 15.3 µg/mL Vanco+0.3% Change (Predicate: Not specified)
Interfering Substances: Bilirubin conjugated (25 mg/dL) at 15.9 µg/mL Vanco+0.6% Change (Predicate: Not specified)
Interfering Substances: Hemoglobin (1000 mg/dL) at 16.1 µg/mL Vanco-7.7% Change (Predicate: Not specified)
Interfering Substances: Lipemia (1000 mg/dL) at 16.0 µg/mL Vanco-0.2% Change (Predicate: Not specified)
Valproic Acid Method (Predicate: Abbott TDx Valproic Acid method)
Analytical Range0.3-150 µg/mL (Predicate: Not explicitly stated, but implied similar operating range)
Precision (Total) @ 37.5 µg/mL3.8% @ 20.3 µg/mL (Predicate: 3.4%)
Precision (Total) @ 75 µg/mL2.4% @ 60.2 µg/mL (Predicate: 3.4%)
Precision (Total) @ 125 µg/mL2.8% @ 108.5 µg/mL (Predicate: 3.7%)
Correlation (y=1.11x + 0.27, r=0.993, Sux=3.79 µg/dL)Matched to Predicate
Interfering Substances: Bilirubin unconjugated (25 mg/dL) at 82.5 µg/mL VA+6% Change (Predicate: Not specified)
Interfering Substances: Bilirubin conjugated (25 mg/dL) at 78.1 µg/mL VA+1% Change (Predicate: Not specified)
Interfering Substances: Hemoglobin (1000 mg/dL) at 80.2 µg/mL VA+4% Change (Predicate: Not specified)
Interfering Substances: Lipemia (500 mg/dL) at 80.8 µg/mL VA+3% Change (Predicate: Not specified)

2. Sample Sizes Used for the Test Set and Data Provenance:

  • C3 Method: n = 101 samples for correlation. Data provenance is not specified (e.g., country of origin, retrospective or prospective).
  • C4 Method: n = 94 samples for correlation. Data provenance is not specified.
  • IgA Method: n = 97 samples for correlation. Data provenance is not specified.
  • IgG Method: n = 97 samples for correlation. Data provenance is not specified.
  • IgM Method: n = 89 samples for correlation. Data provenance is not specified.
  • TRF Method: n = 106 samples for correlation. Data provenance is not specified.
  • Vancomycin Method: n = 55 samples for correlation. Data provenance is not specified.
  • Valproic Acid Method: n = 55 samples for correlation. Data provenance is not specified.

For precision and interfering substances studies, specific sample numbers are not given, but they are implied to be sufficient for the reported percentages. The studies appear to be clinical validation studies performed by the manufacturer. These are typically prospective studies using human serum samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

The concept of "experts" and "ground truth" as typically applied to image-based diagnostic AI algorithms (e.g., radiologists) is not directly applicable here. For these in vitro diagnostic (IVD) devices, the "ground truth" is established by:

  • Predicate Device Measurements: The performance of the predicate device (BNA Complement C3 method, etc.) serves as the reference standard for correlation studies. The predicate devices themselves would have undergone rigorous validation.
  • Reference Methods/Laboratory Standards: For precision, linearity, and interference studies, the "ground truth" for the concentration levels would be established using validated internal laboratory methods, certified reference materials, or gravimetric/volumetric standards, not expert readers.

Therefore, no information on "number of experts" or their "qualifications" is provided, as it's not relevant to this type of device comparison.

4. Adjudication Method for the Test Set:

Not applicable. Adjudication methods (e.g., 2+1, 3+1) are primarily used in studies where human interpretation of data (like medical images) is involved and discrepancies need to be resolved. For quantitative IVD devices measuring biomolecules, agreement is by quantitative comparison to a reference method or predicate device, not by expert consensus on interpretation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not done. MRMC studies are specific to evaluating the impact of an AI system on human reader performance, particularly in diagnostic imaging. This document describes the performance of an automated laboratory assay, not an AI algorithm assisting human interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Yes, the studies reported are essentially "standalone" performance studies for the Bayer ADVIA IMS systems. These are automated systems performing quantitative measurements. The performance metrics (precision, correlation, analytical range, interference) evaluate the device's inherent analytical capabilities, independent of human interaction during the measurement process, aside from loading samples and reagents.

7. The Type of Ground Truth Used:

The ground truth used for these studies is primarily:

  • Predicate Device Measurements: For correlation studies, the measurements obtained from the predicate devices (e.g., Dade Behring BNA methods, Bayer Immuno 1, Abbott TDx) are considered the reference against which the ADVIA IMS device is compared. This demonstrates "substantial equivalence."
  • Known Concentrations/Spiked Samples: For precision, analytical range, and interfering substance studies, the "ground truth" is established by using samples with known concentrations (e.g., control materials, diluted samples, samples spiked with interfering substances).

8. The Sample Size for the Training Set:

The concept of a "training set" applies to machine learning or AI models that learn from data. These documents describe the performance validation of traditional IVD assays. Therefore, there is no "training set" in the context of an AI algorithm, and no sample size is applicable or reported for such a purpose. The data presented are for validation/testing of the assay.

9. How the Ground Truth for the Training Set Was Established:

As there is no "training set" described for a machine learning model, this question is not applicable. The methods described are for chemical assays, not AI algorithms requiring training data.

{0}------------------------------------------------

C3 Method for the Bayer ADVIA IMS Systems

Listed below is a comparison of the performance between the Bayer ADVIA IMS Complement C3 method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Complement C3 method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS C3 method sheet and the BNAC3 method sheet.

INTENDED USE

This in vitro method is intended to quantitatively measure C3 in human serum on the Bayer ADVIA IMS systems. Measurements of C3 are used to aid in the diagnosis and treatment of immunologic disorders.

METHODADVIA IMSBNA
Part No.Reagents B41-3788-41Calibrators B46-4088-60OSAPOSAU

2.1% @ 79.7 mg/dL 5.9% @ 72.4 mg/dL Precision (Total) 1.8% @ 154 mg/dL 2.1% @ 237 mg/dL

Correlation

y=0.92x + 9.7 where y=ADVIA IMS x=BNA n=101 r=0.972 Syx=12.1 mg/dL

Gabriel J. Munoz, Jr.
6/4/99

{1}------------------------------------------------

Interfering Substances
Interfering SubstanceInterfering SubConcentrationC3ConcentrationEffect(% Change)
Bilirubin, unconjugated20 mg/dL106 mg/dL+1
Bilirubin, conjugated20 mg/dL113 mg/dL+4
Hemoglobin500 mg/dL114 mg/dL-2
Triglycerides1000 mg/dL107 mg/dL-8

Analytical Range

:


:

:

ಿರುವ ಸಾ

Dilution RangeConcentration Range,mg/dL
Normal27 - 360
Out of Range Low6.8 - 90
Out of Range High135 - 1800

{2}------------------------------------------------

C4 Method for the Bayer ADVIA IMS Systems

Listed below is a comparison of the performance between the Bayer ADVIA IMS Complement C4 method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Complement C4 method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS C4 method sheet and the BNAC4 method sheet.

INTENDED USE

This in vitro method is intended to quantitatively measure C4 in human serum on the Bayer ADVIA IMS systems. Measurements of C4 are used to aid in the diagnosis and treatment of immunologic disorders.

METHODADVIA IMSBNA
Part No.
ReagentsB41-3789-41OSAO
CalibratorsB46-4088-60OSAU
Precision (Total)1.9% @ 19.3 mg/dL2.0% @ 34.8 mg/dL1.9% @ 49.3 mg/dL2.9% @ 15.7 mg/dL
Correlationy=0.92x - 0.3wherey=ADVIA IMSx=BNAn=94r=0.989Syx=1.9 mg/dL

Gabriel J. Munoz, Jr.
6/4/99

{3}------------------------------------------------

Interfering SubstanceInterfering SubConcentrationC4ConcentrationEffect(% Change)
Bilirubin, unconjugated20 mg/dL19 mg/dL0
Bilirubin, conjugated20 mg/dL21 mg/dL0
Hemoglobin500 mg/dL20 mg/dL0

Analytical Range

Dilution RangeConcentration Range,mg/dL
Normal7.2 - 96
Out of Range Low4.0 - 53
Out of Range High36 - 480

{4}------------------------------------------------

IgA Method for the Bayer ADVIA IMS Systems

Listed below is a comparison of the performance between the Bayer ADVIA IMS Immunoglobulin A method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Immunoglobulin A method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS IgA method sheet and the BNA IgA method sheet.

INTENDED USE

This in vitro method is intended to quantitatively measure IgA in human serum on the Bayer ADVIA IMS systems. Measurements of IgA are used to aid in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

METHODADVIA IMSBNA
Part No.ReagentsB41-3791-41OSAR
CalibratorsB46-4088-60OSAU
Precision (Total)2.3% @ 122mg/dL1.6% @ 233mg/dL1.4% @ 344mg/dL3.5% @ 296mg/dL
Correlationy=1.00x - 1wherey=ADVIA IMSx=BNAn=97r=0.994Syx=75.6 mg/dL

Gabriel J. Munoz, Jr.

6/4/99

{5}------------------------------------------------

Interfering SubstanceInterfering SubConcentrationIgAConcentrationEffect(% Change)
Bilirubin, unconjugated20 mg/dL180 mg/dL+2
Bilirubin, conjugated20 mg/dL200 mg/dL-3
Hemoglobin500 mg/dL191 mg/dL+1
Triglycerides1000 mg/dL184 mg/dL-8

:

Analytical Range

Analytical Range
Dilution RangeConcentration Rangemg/dL
Normal45 - 600
Out of Range Low11.3 - 150
Out of Range High360 - 4800

:

.

·

{6}------------------------------------------------

IgG Method for the Bayer ADVIA IMS Systems

Listed below is a comparison of the performance between the Bayer ADVIA IMS Immunoglobulin G method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Immunoglobulin G method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS IgG method sheet and the BNA IgG method sheet.

INTENDED USE

This in vitro method is intended to quantitatively measure IgG in human serum on the Bayer ADVIA IMS systems. Measurements of IgG are used to aid in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

METHODADVIA IMSBNA
Part No.ReagentsB41-3792-41OSAS
CalibratorsB46-4088-60OSAU

Precision (Total) 1.8% @ 772mg/dL 2.7% @ 1317mg/dL 1.6% @ 1416mg/dL 3.4% @ 2141mg/dL

Correlation

y=0.97 x - 3 where y=ADVIA IMS x=BNA n=97 r=0.994 Syx=118 mg/dL

Gabriel J. Murau, Jr.

6/4/99

{7}------------------------------------------------

Interfering SubstanceInterfering SubConcentrationIgGConcentrationEffect(% Change)
Bilirubin, unconjugated20 mg/dL876 mg/dL-3.0
Bilirubin, conjugated20 mg/dL953 mg/dL-3.0
Hemoglobin500 mg/dL897 mg/dL+7.0
Triglycerides1000 mg/dL887 mg/dL-3.0

:

Analytical Range

Dilution RangeConcentration Range,mg/dL
Normal225 - 3,000
Out of Range Low56 - 750
Out of Range High1128 - 15,000

{8}------------------------------------------------

IgM Method for the Bayer ADVIA IMS Systems

Listed below is a comparison of the performance between the Bayer ADVIA IMS Immunoglobulin M method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Immunoglobulin M method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS IgM method sheet and the BNA IgM method sheet.

INTENDED USE

This in vitro method is intended to quantitatively measure IgM in human serum on the Bayer ADVIA IMS systems. Measurements of IgM are used to aid in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

METHODADVIA IMSBNA
Part No.ReagentsB41-3793-41OSAT
CalibratorsB46-4088-60OSAU

Precision (Total)

3.6% @ 69.0mg/dL 1.9% @ 117mg/dL 2.4% @ 128mg/dL 1.5% @ 177mg/dL

Correlation

y=1.05x - 13.4 where y=ADVIA IMS x=BNA n=89 r=0.990 Syx=123.5 mg/dL

Gabriel J. Murray, Jr.
6/4/99

{9}------------------------------------------------

Interfering SubstanceInterfering SubConcentrationIgMConcentrationEffect(% Change)
Bilirubin, unconjugated20 mg/dL77 mg/dL+3.9
Bilirubin, conjugated10 mg/dL85 mg/dL+1.2
Hemoglobin500 mg/dL86 mg/dL0.0

.

Analytical Range

·

Dilution RangeConcentration Range,
mg/dL
Normal30 - 400
Out of Range Low10 - 133
Out of Range High 1241 - 3,200
Out of Range High 21,203 - 16,000

:

:

{10}------------------------------------------------

TRF Method for the Bayer ADVIA IMS Systems

Listed below is a comparison of the performance between the Bayer ADVIA IMS Transferrin method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Transferrin method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS TRF method sheet and the BNATRF method sheet.

INTENDED USE

This in vitro method is intended to quantitatively measure TRF in human serum on the Bayer ADVIA IMS systems. Measurements of TRF are used to aid in the diagnosis and treatment of malnutrition, chronic infection, acute hepatitis, polycythemia, pernicious anemia and red blood disorders, such as iron deficiency anemia.

METHODADVIA IMSBNA
Part No.ReagentsB41-3795-41OSAX
CalibratorsB46-4088-60OSAU

Precision (Total) 2.2% @ 127 mg/dL 2.7% @ 303 mg/dL 1.9% @ 268 mg/dL 3.1% @ 400 mg/dL

Correlation

y=0.96x - 3 where y=ADVIA IMS x=BNA n=106 r=0.979 Syx=16.8 mg/dL

Gabriel J. Munoz, Jr.
6/4/99

{11}------------------------------------------------

Interfering SubstanceInterfering SubConcentrationTRFConcentrationEffect(% Change)
Bilirubin, unconjugated20 mg/dL212 mg/dL+2
Bilirubin, conjugated20 mg/dL231 mg/dL0
Hemoglobin500 mg/dL261 mg/dL0
Triglycerides1000 mg/dL215 mg/dL-4

Analytical Range

Dilution RangeConcentration Range,mg/dL
Normal54 - 720
Out of Range Low13 - 180
Out of Range High270 - 3,600

.

:

{12}------------------------------------------------

Vancomycin Method for the Bayer ADVIA IMS Systems

Listed below is a comparison of the performance between the Bayer ADVIA IMS Vancomycin method and a similar device that was granted clearance of substantial equivalence (Bayer Immuno 1 Vancomycin method) The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS Vancomycin method sheet and the Immuno 1 Vancomycin method sheet.

INTENDED USE

This in vitro method is intended to quantitatively measure vancomycin in human serum on the Bayer ADVIA IMS systems. Measurements of vancomycin are used to aid in attaining optimum therapy in patients treated with the drug.

METHODADVIA IMSImmuno 1
Part No.ReagentsB41-3767-41T01-3705-01
CalibratorsB46-4117-01T03-3714-01
Analytical Range0.4-50 µg/mL
Precision (Total)2.8% @ 8.6 µg/mL8.9% @ 6.7 µg/mL
3.4% @ 21.5 µg/mL7.5% @ 23.3 µg/mL
3.9% @ 35.8 µg/mL8.1% @ 32.4 µg/mL
Correlationy=1.02x + 0.68wherey=ADVIA IMSx=Immuno 1n=55r=0.987Svx=1.6 µg/dL

Interfering Substances

Interfering SubstanceInterfering SubstanceConcentrationVancomycinConcentrationEffect% Change
(µmol/L)(µg/mL)
Bilirubin (unconjugated)0.43 mmol/L25 mg/dL10.615.3+0.3
Bilirubin (conjugated)0.43 mmol/L25 mg/dL11.015.9+0.6
Hemoglobin10 g/L1000 mg/dL11.116.1-7.7
Lipemia (Triglycerides)11.4 mmol/L1000 mg/dL11.016.0-0.2

Gabriel J. Murray Jr.
6/4/99

{13}------------------------------------------------

Valproic Acid Method for the Bayer ADVIA IMS Systems

Listed below is a comparison of the performance between the Bayer ADVIA IMS Valproic Acid method and a similar device that was granted clearance of substantial equivalence (Abbott TDx Valproic Acid method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS Valproic Acid method sheet and the TDx Valproic Acid method sheet.

INTENDED USE

This in vitro method is intended to quantitatively measure valproic acid in human serum on the Bayer ADVIA IMS systems. Measurements of valproic acid are used to aid in attaining optimum therapy in patients treated with the drug.

METHODADVIA IMSTDx
Part No.ReagentsB41-3766-419514-20
CalibratorsB46-4118-019514-01
Analytical Range0.3-150 µg/mL
Precision (Total)3.8% @ 20.3 µg/mL3.4% @ 37.5 µg/mL
2.4% @ 60.2 µg/mL3.4% @ 75 µg/mL
2.8% @ 108.5 µg/mL3.7% @ 125 µg/mL
Correlationy=1.11x + 0.27
where
y=ADVIA IMS
x=TDx
n=55
r=0.993
Sux=3.79 µg/mL

Interfering Substances

Interfering SubstanceInterfering SubstanceConcentrationValproic acidConcentrationEffect% Change
(µmol/L)(µg/mL)
Bilirubin (unconjugated)0.43 mol/L25 mg/dL57382.5+6
Bilirubin (conjugated)0.43 mol/L25 mg/dL54278.1+1
Hemoglobin10 g/L1000 mg/dL55780.2+4
Lipemia (Triglycerides)5.7 mmol/L500 mg/dL56180.8+3

Gabriel J. Munoz Jr.
6/4/99

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Nov - 9 1999

Mr. Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corp. Business Group Diagnostics 511 Benedict Avenue Tarrytown, New York 10591-5097

Re: K991907

Trade Name: 8 Additional Assays for the Bayer ADVIA® Integrated Modular System (IMS) Regulatory Class: II Product Code: CZW, CFQ, DBI, DDG, KTO, LEG, LEH Dated: September 3, 1999 Received: September 7, 1999

Dear Mr. Muraca:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might

{15}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{16}------------------------------------------------

Page 1 of 3

510(k) Number (if known):

Device Name: Bayer ADVIA® Integrated Modular System (IMS)

Indications For Use:

The Bayer ADVIA IMS C3 assay is an in vitro diagnostic device intended to measure Complement C3 (C3) in human serum. Such measurements are used as an aid in the diagnosis and treatment of immunologic disorders.

The Bayer ADVIA IMS C4 assay is an in vitro diagnostic device intended to measure Complement C4 (C4) in human serum. Such measurements are used as an aid in the diagnosis and treatment of immunologic disorders.

The Bayer ADVIA IMS IgA assay is an in vitro diagnostic device intended to measure Immunoglobulin A (IgA) in human serum. Such measurements are used as an aid in the diagnosis and treatment of abnormal protein metabolism and the body's inability to resist infectious agents.

Dran Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K991907

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

{17}------------------------------------------------

510(k) Number (if known):

Device Name: Bayer ADVIA® Integrated Modular System (IMS)

Indications For Use:

The Bayer ADVIA IMS IgG assay is an in vitro diagnostic device intended to measure Immunoglobulin G (IgG) in human serum. Such measurements are used as an aid in the diagnosis and treatment of autoimmune diseases, chronic or recurrent infections and abnormal protein metabolism.

The Bayer ADVIA IMS Immunoglobulin M (IgM) assay is an in vitro diagnostic device intended to measure IgM in human serum. Such measurements are used as an aid in the diagnosis and treatment of chronic or recurrent infections and abnormal protein metabolism.

The Bayer ADVIA IMS Transferrin assay is an in vitro diagnostic device intended to measure transferrin (TRF) in human serum. Such measurements are used as an aid in the diagnosis and treatment of malnutrition, chronic infection, acute hepatitis, polycythemia, pernicious anemia, and red blood cell disorders, such as iron deficiency anemia.

Fein Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K991907

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use - -

(Optional Format 1-2-96)

{18}------------------------------------------------

Page 3 of 3

510(k) Number (if known):

Device Name: Bayer ADVIA® Integrated Modular System (IMS)

Indications For Use:

The Bayer ADVIA IMS Vancomycin assay is an in vitro diagnostic device intended to measure vancomycin, an antibiotic drug, in human serum. Measurements of vancomycin are used as an aid in the diagnosis and treatment of vancomycin overdose and in monitoring therapeutic levels of vancomycin to ensure appropriate therapy.

The Bayer ADVIA IMS Valproic Acid Assay is an in vitro diagnostic device intended to measure valproic acid, an anti-epileptic drug, in human serum. Measurements of valproic acid are used as an aid in the diagnosis and treatment of valproic acid overdose and in monitoring therapeutic levels of valproic acid to ensure appropriate therapy.

Hean Coogh
(Division Sign-Off)
Division of Clinical Laboratory Devices

× 510(k) Number .

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 866.5240 Complement components immunological test system.

(a)
Identification. A complement components immunological test system is a device that consists of the reagents used to measure by immunochemical techniques complement components C1q , C1r , C1s , C2 , C3 , C4 , C5 , C6 , C7 , C8 , and C9 , in serum, other body fluids, and tissues. Complement is a group of serum proteins which destroy infectious agents. Measurements of these proteins aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of complement components.(b)
Classification. Class II (performance standards).