K Number
K991734

Validate with FDA (Live)

Date Cleared
1999-06-09

(19 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HERMES OR Control Center is indicated for use with Stryker Endoscopy 882 Camera, Quantum 5000 Light Source, Stryker SES Shaver, W.O.M. 20L Insufflator, W.O.M. 2.0L Arthroscopy Pump. and the Stryker Total Performance System (endoscopy port only). It can be used in general laparoscopy, nasopharyngoscopy, ear endoscopy, and sinuscopy where a laparoscope/endoscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic symph node dissection, laparoscopically assisted hysterectomy, laparoscopic anterior spinal fusion, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization in indicated and examination of the evacuated chamber during performance of valve replacement. The users of the HERMES ORCC are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

Device Description

The HERMES Operating Room Control Center is a computer-driven system whose basic function is offer the additional option for surgeon selection of attachment device parameter settings utilizing voice control.

The intent of the HERMES OR Control Center is to allow for simplified and more direct control of medical device settings by the physician, thereby eliminating the necessity of using the various interfaces existing on the Stryker Endoscopy 882 Camera, Quantum 5000 Light Source, Stryker SE5 Shaver, WOM 20L Insufflator, WOM 2.0L Arthroscopy Pump and the Stryker Total Performance System (endoscopy port only) in the Operating setting, or relying upon verbal communications between the surgeon and other personnel in the operation room in order to adjust surgical equipment.

AI/ML Overview

The HERMES™ Operating Room Control Center is a computer-driven system that allows surgeons to control medical device settings using voice commands.

Here's an analysis based on the provided document:

Acceptance Criteria and Device Performance

This 510(k) summary does not include a table of specific acceptance criteria with corresponding device performance metrics (e.g., accuracy, precision, sensitivity, specificity). Instead, it primarily focuses on demonstrating that the HERMES™ Operating Room Control Center conforms to established safety and electrical standards.

The document lists the following standards to which the device was designed and tested:

Acceptance Criterion (Standard Compliance)Reported Device Performance (Compliance Statement)
IEC 601-1 Second Edition 1990 Medical Electrical EquipmentDevice was designed and tested to this standard.
IEC 601-1 Second Amendment 1 1991 Medical Electrical EquipmentDevice was designed and tested to this standard.
IEC 601-2-18 First Edition 1990 Medical Electrical EquipmentDevice was designed and tested to this standard.
UL 2601-1Device was designed and tested to this standard.
Conducted & Radiated Emission EN55022/A1: 1995Device was designed and tested to this standard.
Immunity Tests EN61000-4-2: 1995Device was designed and tested to this standard.
Immunity Tests EN61000-4-3: 1995Device was designed and tested to this standard.
Immunity Tests EN50140:1994Device was designed and tested to this standard.
Immunity Tests EN61000-4-4:1995Device was designed and tested to this standard.
Immunity Tests EN61000-4-5:1995Device was designed and tested to this standard.
Immunity Tests EN61000-4-6:1995Device was designed and tested to this standard.
CAN/CSA-C22.2 NO. 601Device was designed and tested to this standard.

Study Details:

The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed clinical study with performance metrics in the way one might expect for diagnostic AI or imaging devices.

  1. Sample size for the test set and data provenance:
    The document does not specify a "test set" in the context of clinical performance data (e.g., patient data, image dataset). The testing refers to compliance with electrical and safety standards. Therefore, information on sample size and data provenance for a test set in the traditional sense is not applicable or provided.

  2. Number of experts used to establish the ground truth for the test set and qualifications:
    Not applicable, as the provided document focuses on compliance with engineering and safety standards, not on clinical performance or diagnostic accuracy that would require expert-established ground truth.

  3. Adjudication method for the test set:
    Not applicable, for the reasons stated above.

  4. Multi-reader multi-case (MRMC) comparative effectiveness study:
    No, an MRMC comparative effectiveness study was not done or reported in this 510(k) summary. This device is an operating room control system for device settings, not a diagnostic or AI-assisted interpretation tool that would typically involve MRMC studies.

  5. Standalone (algorithm only without human-in-the-loop performance) study:
    While the system operates on voice control, the "standalone" performance isn't described in terms of a clinical outcome or diagnostic accuracy. Its performance is measured by its ability to reliably and correctly interpret voice commands to adjust device parameters according to the surgeon's intent, and its compliance with safety standards. The document doesn't provide specific data on voice recognition accuracy or response time for device control.

  6. Type of ground truth used:
    The "ground truth" for this device's performance testing revolved around engineering standards and functional requirements. For example, whether it correctly changed a setting when a specified voice command was given, or if it met electromagnetic compatibility limits. It is not expert consensus, pathology, or outcomes data in the clinical sense.

  7. Sample size for the training set:
    This information is not provided. If a voice recognition component involves machine learning, there would typically be a training set, but this detail is absent from the summary.

  8. How the ground truth for the training set was established:
    Not provided. This detail would only be relevant if the device incorporated advanced AI/ML for voice recognition requiring specific training data and ground truth labeling, which is not elaborated upon in this 1999 document.

{0}------------------------------------------------

×991734 Premarket Notification 510(k) Safety and Effectiveness Su 10.

HERMES™ Operating Room Control Center 510(k) Summary

Computer Motion, Inc. is submitting the following safety and effectiveness summary.

    1. Submitter Information:
      Computer Motion 130-B Cremona Drive Goleta, CA 93117

Contact: David U. Thomas, Regulatory Affairs Specialist Phone: (805) 968-9600 Ext. 214 FAX (805) 685-9277 Prepared: May 20, 1999

    1. Name of Device:
      Proprietary Name: HERMESTM Operating Room Control Center Common Name is HERMESTM Classification Name: Laparoscope for Use in General and Plastic Surgery, Regulation Number 876.1500, Class II.
    1. Substantially equivalent to SE 5 Shaver and HERMES™ (510(k)'s K974771 and K990691, respectively)
    1. The HERMES Operating Room Control Center is a computer-driven system whose basic function is offer the additional option for surgeon selection of attachment device parameter settings utilizing voice control.

The intent of the HERMES OR Control Center is to allow for simplified and more direct control of medical device settings by the physician, thereby eliminating the necessity of using the various interfaces existing on the Stryker Endoscopy 882 Camera, Quantum 5000 Light Source, Stryker SE5 Shaver, WOM 20L Insufflator, WOM 2.0L Arthroscopy Pump and the Stryker Total Performance System (endoscopy port only) in the Operating setting, or relying upon verbal communications between the surgeon and other personnel in the operation room in order to adjust surgical equipment.

The HERMES OR Control Center is indicated for use with Stryker Endoscopy 882 Camera, Quantum 5000 Light Source, Stryker SE5 Shaver, WOM 20L Insufflator, WOM 2.0L Arthroscopy Pump and the Stryker Total Performance System (endoscopy port only). It can be used in general laparoscopy, nasopharyngoscopy, ear endoscopy, and sinuscopy where a laparoscope/endoscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectorny, laparoscopic hernia repair, laparoscopic appendectorny, laparoscopic pelvic symph node dissection, laparoscopically assisted hysterectomy, laparoscopic & thorascopic anterior spinal fusion, decompression fixation,

{1}------------------------------------------------

wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal weage resources, iang over for coronary artery bypass, coronary artery bypass grafting where maninal y artery diosenian indicated and examination of the evacuated cardiac chamber during performance of valve replacement. The users of the HERMES ORCC are general surgeons, performation of varies repeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

The HERMES ORCC was designed and tested to the following standards:

IEC 601-1 Second Edition 1990 International Standard for Medical Electrical Equipment IEC 601-1 Second Domn 1 1991 International Standard for Medical Electrical Equipment IEC 601-2-18 First Edition 1990 International Standard for Medical Electrical Equipment UL 2601-1

Conducted & Radiated Emission EN55022/A1: 1995

Immunity Tests EN61000-4-2: 1995; EN61000-4-3: 1995; EN50140:1994; EN61000-4-4:1995; EN61000-4-5:1995; EN61000-4-6:1995.

CAN/CSA-C22.2 NO. 601.

{2}------------------------------------------------

Public Health Service

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest wings and a body. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

0 1999 JUN

Mr. David U. Thomas Regulatory Affairs Specialist Computer Motion, Inc. 130 Cremona Drive, Suite B Goleta, California 93117

K991734
Trade Name: HERMES™ Operating Room Control Center
Trade Name: HERMESTM Operating Room Control Center Re: Regulatory Class: II Product Code: GCJ Dated: May 20, 1999 Received: May 21, 1999

Dear Mr. Thomas:

Deal WIT Tribution of the Stock) notification of intent to market the device referenced.
The have reviews in the device is substantially equivalent (for the indications for We have reviewed your Section 510(k) notification of themsel and contracted to the indications for
above and we have determined the devices marketed in interstate commerce pr above and we have determined the devices marketed in interstate commerce prior to use stated in the enclosure) to devices marketed in merisiale comments, or to devices that
May 28, 1976, the enactment date with the novisions of the Federal Food, Drug, and May 28, 1976, the enactment date of the Meeting of the Federal Food, Drug, and have been reclassified in accordance with the provisions of the general controls
Cosmetic Act (Act). You may, therefore, market the device, subject to the equirements for Cosmetic Act (Act). You may, therefore, market the aevice, shoped to enquirements for
Cosmetic Act (Act). You may, the no, good manufacturing provisions for
provisions of the Act. The general controls provisions of the research of the office of
annual registration, listing direction, and gdulteration, annual registration, fisting of 2014-02-2
prohibitions against misbranding and adulteration.

.
If your device is a more to subscribed into either class II (Special Controls) or class III
n and controller and control and intribut to such additional controls. Existing If your device is classified (see above) into ender class in (oposto). Existing major
(Premarket Approval), it may be subject to such additional controls, Existing major (Premarket Approval), it may be subject to such adultions Regulations, Title 21,
regulations affecting your device can be found in the Code of Federal Regulance with the regulations affecting your device can be found in the Counces compliance with the
Parts 800 to 895. A substantially equivalent as set forth in the Quality System Parts 800 to 895. A substantally equivalent determination in the Quality System
current Good Manufacturing Practice requirement, as set forth in the Quality System current Good Manufacturing Practice requirencin, as soon (21 CFR Part 820) and that,
Regulation (QS) for Medical Devices: General regulation (FDA) will verif Regulation (QS) for Mcdical Devices: Geheral Ingualibin (FDA) will verify
through periodic (QS) inspections, the Food and Drug Administration (FDA) werify through periodic (QS) inspections, the roomply with the GMP regulation may result in regulatory
such assumptions. Failure to comply with the GMP regulation may result in the such assumptions. Failure to comply with the United Concerning your device in the
action. In addition, FDA may publish further announcements concerning your commission does action. In addition, FDA may publish turner announcements ontification submission does
Federal Register. Please note: this response to your premarks notification on for Federal Register. Please note: this response or your promants of the Act for not affect any obligation you might have under seculous 531 urough on 100 million was or regulations.

{3}------------------------------------------------

Page 2 -- Mr. David U. Thomas

This letter will allow you to begin marketing your device as described in your 510(k) This letter will anow you to begin maiketing your device of your device and thus, ne premarket notification. The FDA midnig of subtainter equal device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFFP Part 801 If you desire specific advice for your donostic devices), please contact the Office of
and additionally 809.10 for in vitro diagnostic devices the promotion and and additionally 809.10 for in Vitto ditignally, for questions on the promotion and Compliance at (301) 394-4.997. Aduntonatify 16th-office of Compliance at (301) 594-4639.
advertising of your device, please contact the Office of Compliance to premarket Also, please note the regulation entitled, "Misbranding by reference to premarket Also, please note the regulation entities, "Information on your responsibilities under the notification" (21 CFR 807.97). This general Manufactures Assistance at its toll-free Act may be obtain-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K99173<

Page 1 of 1

510(k) Number (if known): 991734

Device Name: HERMES™ with Stryker Total Performance System (endoscopy port only)

Indications For Use:

The HERMES OR Control Center is indicated for use with Stryker Endoscopy 882 Camera, Quantum 5000 Light Source, Stryker SES Shaver, W.O.M. 20L Insufflator, W.O.M. 2.0L Arthroscopy Pump. and the Stryker Total Performance System (endoscopy port only). It can be used in general laparoscopy, nasopharyngoscopy, ear endoscopy, and sinuscopy where a laparoscope/endoscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic symph node dissection, laparoscopically assisted hysterectomy, laparoscopic anterior spinal fusion, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization in indicated and examination of the evacuated chamber during performance of valve replacement. The users of the HERMES ORCC are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER P. NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K991734

Prescription Use X
(Per 21 CFR 801.109)
OR Over-The-Counter Us
(Optional Form

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.