K Number
K991628

Validate with FDA (Live)

Date Cleared
1999-08-09

(90 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The modified LX-20 Laser System Family is a general surgical instrument used to vaporize, incise, excise, ablate, or photo coagulate soft tissue in surgical procedures in the listed surgical specialties.

The system may be effectively utilized for those applications recognized as within the accepted indications for use of C0, laser devices performed within the listed specialties and which require power densities as calculated by power and spot size referenced in the power density charts in the operator's manual.

Indications for Use - Gynecology
Representative examples: Condyloma, excision and vaporization, cervical, vulvar, & perineal; vaginal intraepithelial neoplasia ablation (VAIN); vulvar intraepithelial neoplasia ablation (VIN); herpes vaporization; urethral caruncle vaporization; I&D Bartholin's & nubothian cysts.

Indications for Use - Laparoscopic Surgery
Vaporization, incision, excision, ablation, or photo coagulation of soft tissue in endoscopic and laparoscopic surgery including GYN laparoscopy where delivery of energy by the hollow fiber may be more convenient than delivery of energy by a handpiece. Representative examples: endometriosis ablation; excision of adhesions; salpingotomy; metroplasty,

Indications for Use - Dermatology
Representative examples: Tattoo removal; port wine hemangioma removal; rhinophyma reduction; telangiectasia removal; keloid reduction; wart removal; basal & squamous cell carcinoma removal; blepharoplasty; xanthalasma removal; removal of neurofibromas, hemangiomas, nevi, and tricoepitheliomas; dermabrasion such as for lentigos, keratoses, actinic keratosis & cheilitis.

Indications for Use - Dental/Oral Surgery
Representative examples: Gingivectomy; frenum release; removal of soft tissue, cysts, and tumors.

Indications for Use - Orthopedic
Representative examples: Menisectomy; chondromalacia ablation; partial synovectomy; lateral release; PMMA removal.

Indications for Use - General Surgery
Representative examples: Hemorthoid removal; skin tag vaporization; pilodidal cyst removal and repair: debridement of decubitus ulcers and statis ulcers; mastectomy, breast biopsy, reduction mammoplasty; cytoreduction for metastatic disease; many dermatological procedures.

Indications for Use - Otorhinolaryngology (ENT)
Representative examples: Subglottic stenosis vaporization; tonsillectomy; removal of vocal cord papillomas, nodules, and polyps; lymphangioma removal; pulmonary bronchial and tracheal lesion removal: turbinectomy; removal of leukoplakia of larynx; ablation of choanal atresia. Myringotomy/tympanostomy; Laser Assisted Uvulopalatoplasty (LAUP).

Indications for Use - Podiatry
Representative examples: Plantar wart vaporization; fungal nail treatment; partial and complete matrixectomy; porokeratoma ablation; Morton's neruoma removal; ingrown nail treatment.

Device Description

The Luxar modified LX-20 Surgical Laser System produces laser (infrared) energy at a wavelength of 10.6 microns which is directed to soft tissue through a knuckled beam delivery assembly, articulated arm or articulated arm plus hollow fiber waveguide, handpieces, and sterile delivery tips.

AI/ML Overview

The provided text describes a 510(k) summary for the Luxar modified LX-20 C0, Surgical Laser System. This document focuses on demonstrating substantial equivalence to previously cleared devices rather than presenting new clinical study data to prove the device meets acceptance criteria.

Key takeaway: The submission states, "There are no new indications for use beyond those cleared for the predicate devices, and no clinical data were presented." This directly indicates that a study proving the device meets acceptance criteria, as typically understood for new performance claims, was not conducted or presented in this 510(k) submission. Therefore, I cannot provide information on acceptance criteria or a study proving they were met.

However, I can extract information related to the device's intended use and the basis for its clearance:


1. Table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not explicitly stated as this is a substantial equivalence submission based on predicate devices, not a de novo clearance requiring new performance data. The "acceptance criteria" effectively are that the device performs equivalently to the predicate devices for the stated uses.
  • Reported Device Performance: The document states that "Basic system operation and performance are unchanged from the cleared LX-20 Laser System Family." The modified LX-20 shares the same RF excited lasing tube, main chassis, power supply, safety systems, microprocessor controller, and much of the same electronics. The primary modification is the addition of a Sharplan articulated arm, which the submitter believes "do not raise any new issues of safety and effectiveness."

The following information cannot be provided as no new clinical data was presented or a study described in this 510(k) submission:

  • 2. Sample sized used for the test set and the data provenance
  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
  • 4. Adjudication method for the test set
  • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
  • 7. The type of ground truth used
  • 8. The sample size for the training set
  • 9. How the ground truth for the training set was established

Summary of Device Information Presented:

  • Device Name: Luxar modified LX-20 C0, Surgical Laser System
  • Intended Use: Vaporization, incision, excision, ablation, or photocoagulation of soft tissue in various surgical specialties (Gynecology, Laparoscopic Surgery, Dermatology, Dental and Oral surgery, Orthopedic, General Surgery, Otorhinolaryngology, Podiatry).
  • Basis for Clearance: Substantial equivalence to the Luxar LX-20 C0, Surgical Laser System Family (K960475) and Sharplan 20C C02 Laser (K963229 and K935563).
  • Key Modification: Addition of the Sharplan articulated arm.
  • Performance Claim: "Basic system operation and performance are unchanged from the cleared LX-20 Laser System Family." No new clinical data was presented to support new performance claims or new indications for use.

{0}------------------------------------------------

AUG - 9 1999

991628

13.0 510(K) SUMMARY

Page 1 of 2

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

  • Submitter's name, address, telephone, fax, & contact person 1.
    ESC Medical Systems Yokneam Industrial Park, PO Box 240, Yokneam 20692, Israel Phone: 972-4-959-9000 Fax: 972-4-959-9050 Zvi Ladin, Vice President Clinical Applications and Regulatory Affairs

  • Date summary prepared: 2. January 15, 1999

  • Product trade or proprietary name: 3. Luxar modified LX-20 C0, Surgical Laser System

    1. Product common name: C0, Laser

న్. Product classification name

21C.F.R. § 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology

21C.F.R. § 874.4500 Ear, nose and throat microsurgical carbon dioxide laser

21C.F.R. § 884.4550 Gynecologic surgical laser

  • Legally marketed predicate device used for equivalency: 6.
    Luxar LX-20 C0, Surgical Laser System Family (K960475) and Sharplan 20C C02 Laser (K963229 and K935563).

    1. Description:
      The Luxar modified LX-20 Surgical Laser System produces laser (infrared) energy at a wavelength of 10.6 microns which is directed to soft tissue through a knuckled beam delivery assembly, articulated arm or articulated arm plus hollow fiber waveguide, handpieces, and sterile delivery tips.

{1}------------------------------------------------

Statement of intended use: 8.

The intended use of the modified LX-20 is identical to that of the LX-20 Laser System Family and the predicate Sharplan 20C, namely the vaporization, incision, excision, ablation, or photocoagulation of soft tissue in the listed surgical specialties of:

  • · Gynecology
  • · Laparoscopic Surgery Including GYN Laparoscopy
  • · Dermatology
  • · Dental and Oral surgery
  • · Orthopedic
  • · General Surgery
  • Otorhinolaryngology
  • · Podiatry

The Modified LX-20 Laser has harmonized indications for use that are identical to the cleared LX-20 Laser System Family combined with the equivalent Sharplan 20C laser. There are no new indications for use beyond those cleared for the predicate devices, and no clinical data were presented.

9. Technological characteristics:

Basic system operation and performance are unchanged from the cleared LX-20 Laser System Family. The modified LX-20 and the cleared LX-20 Laser System Family share the same RF excited lasing tube, main chassis, power supply, safety systems, microprocessor controller, and much of the same electronics. The modified LX-20 laser system maintains the same 3 base consoles established pursuant to the predicate K960475 - namely the LX-20SI, the LX-20SP, the LX-20LP, and adds the fourth, the LX-20SA. The technological characteristics of the modified LX-20 that are different from the predicate LX-20 Laser System Family are as follows:

The modified unit adds the Sharplan articulated arm from the comparable Sharplan CO, laser. The articulated arm, used unchanged, allows the attachment of the Sharplan series of laser accessories. .

ESC believes that the minor differences in the delivery capabilities of the modified LX-20 and the LX-20 Laser System Family do not raise any new issues of safety and effectiveness.

Advisory: This information was prepared for the sole purpose of compliance with the Safe Medical Devices Act of 1990. It does not imply that the procedures described herein can be performed with the equipment described without substantial risk of personal iniury or death to patients or operators due to operator error or in high risk procedures.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles or faces stacked on top of each other, with flowing lines suggesting movement or water.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG - 9 1999

Mr. Zvi Ladin Vice President for Clinical Applications and Regulatory Affairs ESC Medical Systems Ltd. Yokneam Industrial Park P.O. Box 240 Yokneam, 20692 Isreal

Re: K991628

Trade Name: Luxar Modified LX-20 CO2 Family of Surgical Lasers and Accessories Regulatory Class: II Product Code: GEX Dated: May 9, 1999 Received: May 11, 1999

Dear Mr. Ladin:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Zvi Ladin

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.

Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use - Dermatology

Representative examples: Tattoo removal; port wine hemangioma removal; rhinophyma reduction; telangiectasia removal; keloid reduction; wart removal; basal & squamous cell carcinoma removal; blepharoplasty; xanthalasma removal; removal of neurofibromas, hemangiomas, nevi, and tricoepitheliomas; dermabrasion such as for lentigos, keratoses, actinic keratosis & cheilitis.

Indications for Use - Dental/Oral Surgery

Representative examples: Gingivectomy; frenum release; removal of soft tissue, cysts, and tumors.

Indications for Use - Orthopedic

Representative examples: Menisectomy; chondromalacia ablation; partial synovectomy; lateral release; PMMA removal.

Indications for Use - General Surgery

Representative examples: Hemorthoid removal; skin tag vaporization; pilodidal cyst removal and repair: debridement of decubitus ulcers and statis ulcers; mastectomy, breast biopsy, reduction mammoplasty; cytoreduction for metastatic disease; many dermatological procedures.

Indications for Use - Otorhinolaryngology (ENT)

Representative examples: Subglottic stenosis vaporization; tonsillectomy; removal of vocal cord papillomas, nodules, and polyps; lymphangioma removal; pulmonary bronchial and tracheal lesion removal: turbinectomy; removal of leukoplakia of larynx; ablation of choanal atresia. Myringotomy/tympanostomy; Laser Assisted Uvulopalatoplasty (LAUP).

Indications for Use - Podiatry

Representative examples: Plantar wart vaporization; fungal nail treatment; partial and complete matrixectomy; porokeratoma ablation; Morton's neruoma removal; ingrown nail treatment.

Acollafo

991678

Prescription Use
(Per 21 CFR 801.109)

Page 11

Page 11

{5}------------------------------------------------

510(k) Number (if known): K9916 28

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use:

Intended Use

The modified LX-20 Laser System Family is a general surgical instrument used to vaporize, incise, excise, ablate, or photo coagulate soft tissue in surgical procedures in the listed surgical specialties.

The system may be effectively utilized for those applications recognized as within the accepted indications for use of C0, laser devices performed within the listed specialties and which require power densities as calculated by power and spot size referenced in the power density charts in the operator's manual.

Indications for Use - Gynecology

Representative examples: Condyloma, excision and vaporization, cervical, vulvar, & perineal; vaginal intraepithelial neoplasia ablation (VAIN); vulvar intraepithelial neoplasia ablation (VIN); herpes vaporization; urethral caruncle vaporization; I&D Bartholin's & nubothian cysts.

Indications for Use - Laparoscopic Surgery

Vaporization, incision, excision, ablation, or photo coagulation of soft tissue in endoscopic and laparoscopic surgery including GYN laparoscopy where delivery of energy by the hollow fiber may be more convenient than delivery of energy by a handpiece. Representative examples: endometriosis ablation; excision of adhesions; salpingotomy; metroplasty,

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Sign-Off)
of General Restorative Devices
(k) Number
K991628
Prescription UseOROver-The-Counter Use
(Per 21 CFR 801.109)(Optional Format 1-2-96)

Page 10

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.