K Number
K991487

Validate with FDA (Live)

Date Cleared
1999-10-08

(163 days)

Product Code
Regulation Number
880.6885
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CIDEX® OPA Solution is a high level disinfectant for reprocessing heat sensitive medical devices, and when used according to the Directions for Use.

High Level Disinfectant: CIDEX® OPA Solution is a high level disinfectant when used or reused, according to the Directions for Use, at or above its Minimum Effective Concentration (MEC) as determined by CIDEX® OPA Solution Test Strips, at 20° C (68° F) with an immersion time of at least 12 minutes for a reuse period not to exceed 14 days.

Minimum Effective Concentration (MEC): 0.3%.

Device Description

CIDEX® OPA Solution is formulated to contain 0.55% w/v of ortho-phthalaldehyde. The resultant solution contains a corrosion inhibitor, chelating agents, and a dye in a phosphate buffer. ortho-Phthalaldehyde is chemically related to glutaraldehyde in that they are both aldehydes. The mechanism of action of ortho-phthalaldehyde is postulated to be similar to glutaraldehyde and is based on powerful binding of the aldehyde to the outer cell wall of the organism.

AI/ML Overview

Based on the provided text, here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:

Device Name: CIDEX® OPA Solution (0.55% ortho-phthalaldehyde Solution)
Intended Use: High level disinfectant for reprocessing heat-sensitive medical devices.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Defined by Intended Use/Claims)Reported Device Performance and Test Methods
High-level Disinfection Efficacy: The solution must function as a high-level disinfectant at or above its Minimum Effective Concentration (MEC) of 0.3%, at 20°C (68°F), with an immersion time of at least 12 minutes, for a reuse period not exceeding 14 days. This implies:Efficacy Testing: CIDEX® OPA Solution was tested using a "standard array of microbiology tests for germicidal efficacy."
* Bactericidal Efficacy: Effective at killing bacteria.* Bactericidal: Demonstrated (implicitly by the general efficacy claim and the Mycobacterium terrae test).
* Fungicidal Efficacy: Effective at killing fungi.* Fungicidal: Demonstrated.
* Tuberculocidal Efficacy: Effective at killing Mycobacterium (e.g., M. terrae), which is a key indicator for high-level disinfection.* Tuberculocidal: Demonstrated. Specifically tested in "Simulated Use" against Mycobacterium terrae in artificial soil at MEC of 0.3% and 20°C, and found effective.
* Virucidal Efficacy: Effective at inactivating viruses.* Virucidal: Demonstrated.
* Sporicidal Efficacy: While high-level disinfectants are not expected to kill all spores, some level of sporicidal activity is often evaluated. The document specifically mentions a sporicidal claim.* Sporicidal: Demonstrated as a sporicidal agent as defined by the AOAC Sporicidal Activity Test with an exposure time of at least 32 hours at 20°C. Note: The 32 hours for sporicidal activity is significantly longer than the 12-minute high-level disinfection time, indicating sporicidal claim is separate or for specific conditions.
Minimum Effective Concentration (MEC): The lowest concentration at which the disinfectant is effective, stated as 0.3%.Throughout the reported efficacy tests (e.g., Simulated Use, general efficacy), the solution was tested at its "MEC concentration of 0.3% conditions" and found effective.
Biocompatibility: The product (active ingredient and formulated product) must be safe for human use and product residues on reprocessed devices must not cause toxic effects.* Biocompatibility (product itself): Evaluated through a panel of toxicologic tests: acute oral and dermal toxicity, skin sensitization, genetic toxicity (in vitro and in vivo), subchronic oral toxicity, and developmental toxicity tests. "All animal toxicity data indicate that the product is at least as safe for human use as the predicate device."
* Biocompatibility (residues): "Biocompatibility of product residues testing was conducted." Results indicated "levels well below those, which cause toxic effects in animals."
Material Compatibility: The solution should not significantly damage medical device materials after multiple disinfection cycles."Evaluated for its effect on materials commonly used to fabricate medical devices." "Multiple disinfection soaks over extended periods of time resulted in minimal effect on the test articles." Effects were "similar to those seen with the predicate device."
Stability: The product must remain stable for its claimed shelf life."Tested and found stable for 24 months at 15-30℃."

2. Sample Size Used for the Test Set and Data Provenance

  • Simulated Use (Tuberculocidal): Specific sample size for this test is not explicitly stated, but it involved efficacy against Mycobacterium terrae in artificial soil.
  • In Use Testing: One hundred endoscopes were studied.
  • Data Provenance: The document does not specify the country of origin of the data. The studies appear to be prospective in nature, designed to test the performance of the CIDEX® OPA Solution under controlled and simulated-use conditions.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • The document describes laboratory efficacy tests and clinical "in use" testing. For these types of studies involving disinfection and sterilization, "ground truth" is typically established by standardized laboratory protocols and microbiological cultures, rather than expert medical consensus in the diagnostic sense.
  • The "In Use Testing" on 100 endoscopes would have involved microbiologists or laboratory personnel responsible for culturing and determining the presence/absence of microorganisms, and potentially clinicians handling the endoscopes. No specific number or qualifications for "experts" establishing ground truth are provided in the context of consensus interpretation, as this is not an image-based or qualitative diagnostic assessment. The "ground truth" is typically objective microbiological growth or lack thereof.

4. Adjudication Method for the Test Set

  • Adjudication methods like "2+1" or "3+1" are relevant for subjective assessments (e.g., interpreting medical images).
  • For the types of tests described (microbiological efficacy, biocompatibility, material compatibility), adjudication by consensus of multiple human readers is not typically involved. Results are usually objective measurements (e.g., colony counts, toxicity levels, material degradation) performed according to established protocols. Therefore, the adjudication method would effectively be none in the traditional sense, relying instead on the established scientific reproducibility of the tests.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, an MRMC comparative effectiveness study was not done.
  • MRMC studies are typically used to evaluate the diagnostic performance of human readers, with and without the assistance of AI, usually in the context of medical imaging or pathology. This submission is for a chemical disinfectant, which is evaluated based on objective laboratory and in-use efficacy against microorganisms, not human diagnostic interpretation.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • This question is not applicable to this device. CIDEX® OPA Solution is a chemical disinfectant, not an algorithm or an AI software device. Its performance is inherent to its chemical properties and formulation.

7. The Type of Ground Truth Used

The "ground truth" for the studies described is primarily:

  • Microbiological Culture Results: For efficacy testing (sporicidal, bactericidal, fungicidal, tuberculocidal, virucidal, and in-use testing), the ground truth is the presence or absence of viable microorganisms after exposure to the disinfectant, as determined by standard microbiological culture and enumeration techniques.
  • Toxicological Endpoints: For biocompatibility, the ground truth is based on observed biological responses in in vitro and in vivo animal models (e.g., acute toxicity, sensitization, genetic toxicity, developmental toxicity).
  • Material Integrity Assessments: For material compatibility, the ground truth is based on visual inspection, mechanical property testing, or other methods to assess degradation or changes to medical device materials.

8. The Sample Size for the Training Set

  • This question is not applicable as CIDEX® OPA Solution is a chemical disinfectant, not a machine learning or AI-based device that requires a "training set" in the computational sense. The formulation of the disinfectant is based on chemical principles and experimental optimization, not on a machine learning algorithm.

9. How the Ground Truth for the Training Set Was Established

  • This question is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Page I of 4

K991487

0CT 8 1999

510(K) SUMMARY ·N.

1. 510(k) Summary of Safety and Effectiveness

ContactJohn E. Friend, R.A.C.Vice President, Worldwide Regulatory Affairs(949)789-3885 (Telephone)(949) 789-3900 (Fax)Advanced Sterilization ProductsDivision of Ethicon Inc.33 Technology DriveIrvine, Ca 92618
DateOctober 6, 1999
StatementInformation supporting claims of substantial equivalence, as definedunder the Federal Food, Drug and Cosmetic Act, respecting safety andeffectiveness is summarized below.For the convenience of the Reviewer, this summary is formatted inaccordance with the Agency's final rule “...510(k) Summaries and510(k) Statements...” (21 CFR §807) and can be used to provide asubstantial equivalence summary to anyone requesting it from theAgency.
Device nameClassification: Liquid chemical germicideTrade Name: CIDEX® OPA SolutionProprietary Name: 0.55% ortho-phthalaldehyde Solution
Legally marketed deviceCIDEX® OPA Solution claims equivalence to the following legallymarketed medical devices: CIDEX® Activated Dialdehyde Solution andPeract 20.

{1}------------------------------------------------

Device

Pugg 2 J

CIDEX® OPA Solution is formulated to contain 0.55% w/v of orthophthalaldehyde. The resultant solution contains a corrosion inhibitor, description chelating agents, and a dye in a phosphate buffer, ortho-Phthalaldehyde is chemically related to glutaraldehyde in that they are both aldehydes. The mechanism of action of ortho-phthalaldehyde is postulated to be similar to glutaraldehyde and is based on powerful binding of the aldehyde to the outer cell wall of the organism.

CIDEX® OPA Solution is a high level disinfectant for reprocessing heat Intended use sensitive medical devices, for which sterilization is not suitable, and when used according to the Directions for Use.

High level disinfectant: CIDEX® OPA Solution is a high level disinfectant when used or reused, according to the Directions for Use, at or above its Minimum Effective Concentration (MEC) as determined by the CIDEX® OPA Solution Test Strips, at 20℃ (68 F) with an immersion time of at least 12 minutes for a reuse period not to exceed 14 days.

Minimum Effective Concentration (MEC): 0.3%

Page 234 10/6/1999

{2}------------------------------------------------

K991487

page 3 of

  • CIDEX® OPA Solution was tested using the standard array of Efficacy Testing microbiology tests for germicidal efficacy. Tests demonstrated sporicidal, bactericidal, fungicidal, tuberculocidal and virucidal efficacy of CIDEX® OPA Solution. The CIDEX® OPA Solution used within the test protocols represented end of shelf life, reuse stressed, and MEC concentration of 0.3% conditions.
    Sporicidal: CIDEX® OPA Solution is a sporicidal agent as defined by the AOAC Sporicidal Activity Test with an exposure time of at least 32 hours at 20°C.

Simulated Use: The reused and diluted CIDEX® OPA Solution at an MEC of 0.3% concentration at 20℃ is effective against Mycobacterium terrae in artificial soil.

In Use Testing: One hundred endoscopes used for bronchoscopy, gastroscopy, and colonoscopy were studied to determine the efficiency of CIDEX® OPA Solution for high level disinfection. The results support the claim of high level disinfectant for CIDEX® OPA Solution.

CIDEX® OPA Solution was evaluated for biocompatibility. The active Biocompatibility ingredient and the formulated product were subjected to a panel of toxicologic tests, including acute oral and dermal toxicity, skin sensitization, genetic toxicity, in vitro and in vivo systems, subchronic oral toxicity and developmental toxicity tests. All animal toxicity data indicate that the product is at least as safe for human use as the predicate device.

Biocompatibility of product residues testing was conducted. Results indicate that CIDEX® OPA Solution residuals absorbed onto materials commonly used in reprocessed medical devices are at levels well below those, which cause toxic effects in animals.

Page 235 10/6/1999

{3}------------------------------------------------

page 4 J 4

CIDEX® OPA Solution was evaluated for its effect on materials Material commonly used to fabricate medical devices. Multiple disinfection compatibility soaks over extended periods of time resulted in minimal effect on the test articles. The effects seen were similar to those seen with the predicate device. CIDEX® OPA Solution was tested and found stable for 24 months at Stability 15-30℃. The data presented and the equivalence demonstrated to the predicate Conclusion device support the claim of substantial equivalency for CIDEX® OPA Solution. CIDEX® OPA Solution is safe and effective as a high level disinfectant when used as labeled for reprocessing heat sensitive AOAC Sporicidal Test results showed sporicidal medical devices. activity.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human profile, composed of three curved lines.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

8 1999 OCT

John E. Friend, M.Sc., R.Ph., R.A.C. Vice President, Worldwide Regulatory Affairs Advanced Sterilization Products® A Johnson & Johnson Company Requlatory Affairs Department Division of Ethicon, Inc. 33 Technoloqy Drive Irvine, CA 92618

Re: K991487

CIDEX® OPA Solution Trade Name: Requlatory Class: Unclassified Product Code: MED September 28, 1999 Dated: September 29, 1999 Received:

Dear Mr. Friend:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in

{5}------------------------------------------------

Page 2 - Mr. Friend

the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

ADVANCED STERILIZATION PRODUCTS®

a Johnson Johnson company

REGULATORY AFFAIRS DEPARTMENT

Indications for Use

510(k) Number: K991487

Device Name: CIDEX® OPA Solution (0.55% ortho-phthalaldehyde solution)

Indications For Use:

CIDEX® OPA Solution is a high level disinfectant for reprocessing heat sensitive medical devices, and when used according to the Directions for Use.

High Level Disinfectant: CIDEX® OPA Solution is a high level disinfectant when used or reused, according to the Directions for Use, at or above its Minimum Effective Concentration (MEC) as determined by CIDEX® OPA Solution Test Strips, at 20° C (68° F) with an immersion time of at least 12 minutes for a reuse period not to exceed 14 days.

Minimum Effective Concentration (MEC): 0.3%.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-the-Counter Use X

(Optional Format 1-2-96)

Olia S. Lin

(Division Sign-Off) Division of Dental, Infection Co and General Hospita 5 1 O(k) Number

§ 880.6885 Liquid chemical sterilants/high level disinfectants.

(a)
Identification. A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.(b)
Classification. Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.