K Number
K991397

Validate with FDA (Live)

Date Cleared
2000-02-11

(296 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

the SEIN SE-30 TENS device is indicated for the symptomatic relief and management of chronic (long-term) intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.

Device Description

Proprietary Name : SE-30 TENS Common or Usual Name : Transcutaneous Electrical Nerve Stimulator Classification Name : Stimulator, Nerve, Transcutaneous (Pain Relief)
The SEIN SE-30 TENS device is equivalent to the Graham-Field, Inc. Micro TENS Plus transcutaneous electrical nerve stimulator.
Each product is used for the symptomatic relief of and management of long term intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic pain.
One output channel, Asymmetric bi-phasic & phasic rectangular shape with zero net direct current, Pulse Width 105-450 us programmed insert protected circuit, Pulse Rate 1-200 pulse/sec programmed, Power Source two alkaline AAA batteries 1.5 volt, Maximum Voltage Load 200 ohm 26 V, 500 ohm 40 V, 1 Kohm 52 V, open 90 V, Maximum Output charge/pulse (500 ohm load for safety) 58.5 uc/pulse, Maximum Output average current (500 ohm load for safety) 3.375 mA, Size 70(W) x 79(H) x 17(D) mm, Weight 80 grams, Recharge Battery Option N, Battery Check N, Water Use Option N, Discriminate Polality Automatically Change, Gel Pad Adhesive Type, Treatment 15 min. (no daily limit), MODES AVAILABLE Mode 1: [Modulation] 1. Pulse Rate : 1-100 pps 2. Pulse Width: 220-450 usec, Mode 2... [Burst] 1. Pulse Rate: 16 pulse/burst. 4b/sec 16 pulse/burst. 7b/sec 63-125 pps 2. Pulse Width : 105-450 usec, Mode 3... [Modulation] 1. Pulse Rate : 1-5 pps 2. Pulse Width : 220 usec, Mode 4... [Burst] 1. Pulse Rate: 220 pulse/burst, 1b/sec 90 pulse/burst. 1b/sec 2. Pulse Width: 220-450 usec

AI/ML Overview

Here's an analysis of the provided text regarding the SEIN SE-30 TENS device, framed by the requested acceptance criteria and study information.

It's important to note that the provided document is a 510(k) summary for a Transcutaneous Electrical Nerve Stimulator (TENS) device. For such devices, "acceptance criteria" and "study" typically refer to demonstrating substantial equivalence to a predicate device, as well as adherence to recognized performance standards. This is different from the type of clinical performance study often associated with AI/software devices. The acceptance criteria here are about demonstrating that the new device is as safe and effective as a legally marketed predicate device.


Acceptance Criteria and Study for SEIN SE-30 TENS Device

The SEIN SE-30 TENS device demonstrated substantial equivalence to a predicate device (Graham-Field, Inc. Micro TENS Plus) and compliance with the ANSI/AAMI NS4-1985 standard. The "acceptance criteria" are implied by these comparisons and the requirements of the 510(k) process for demonstrating substantial equivalence.

1. Table of Acceptance Criteria and Reported Device Performance

For a TENS device seeking 510(k) clearance, the "acceptance criteria" are generally met by demonstrating that the device is as safe and effective as a legally marketed predicate device and complies with relevant performance standards. The reported device performance is primarily a comparison of technical specifications.

Acceptance Criteria (Implied)SEIN SE-30 TENS Reported PerformanceSatisfies?
Substantial Equivalence:
- Equivalent intended useSymptomatic relief and management of chronic pain, post-surgical/post-traumatic pain. (Identical to predicate)Yes
- Similar technological characteristics (where differences don't raise new questions of safety/effectiveness)See "Comparison of General Specifications" below; differences evaluated by FDA.Yes
Compliance with Recognized Standards:
- ANSI/AAMI NS4-1985 (Electromyographs and Evoked Potential Instruments)Device was tested and found to meet the requirements of ANSI/AAMI NS4-1985.Yes

Comparison of General Specifications between SEIN SE-30 TENS and GRAHAM-FIELD Micro TENS Plus (Predicate Device)

ITEMSEIN SE-30 TENSGRAHAM-FIELD Micro TENS Plus
Output ChannelsOne output channelTwo with individual controls
Pulse ShapeAsymmetric bi-phasic & phasic rectangular shape with zero net direct currentAsymmetric bi-phasic & phasic rectangular shape with zero net direct current
Pulse Width105-450 us programmed50-200 us variable
Pulse Rate1-200 pulse/sec programmed1-200 pulse/sec variable
Power Source2 alkaline AAA batteries 1.5 volt9 volt alkaline or NiCd bat. (rechargeable option)
Max. Voltage (200 ohm)26 V17 V
Max. Voltage (500 ohm)40 V32 V
Max. Voltage (1 Kohm)52 V45 V
Max. Voltage (open)90 V90 V
Max. Output charge/pulse (500 ohm load)58.5 uc/pulse14.3 uc/pulse
Max. Output average current (500 ohm load)3.375 mA2.3 mA
Size70(W) x 79(H) x 17(D) mm90(W) x 59(H) x 25(D) mm
Weight80 grams142 grams
Recharge Battery OptionNY
Battery CheckNY
Water Use OptionNY
Discriminate PolarityAutomatically ChangeRequired
Gel PadAdhesive TypeAdhesive Type
Modes AvailableMode 1: Modulation (1-100 pps, 220-450 usec) Mode 2: Burst (16 p/burst, 4/7 b/sec, 63-125 pps, 105-450 usec) Mode 3: Modulation (1-5 pps, 220 usec) Mode 4: Burst (220/90 p/burst, 1 b/sec, 220-450 usec)Constant Mode (1-200 pps, 50-200 usec) Burst Mode (10 p/burst, 2 b/sec, 50-200 usec) Modulation Mode (1-200 pps, 50-200 usec)

2. Sample Size Used for the Test Set and Data Provenance

For TENS devices seeking 510(k) clearance, there isn't typically a "test set" in the sense of a dataset for an AI algorithm. Instead, the "test" involved comparing the new device against a predicate device's specifications and performance against a recognized standard (ANSI/AAMI NS4-1985).

  • Sample Size for Test Set: Not applicable in the context of a dataset for an AI algorithm. The "test" involved a single physical SEIN SE-30 TENS device being evaluated against specifications and standards.
  • Data Provenance: The data is primarily derived from engineering specifications, design documents, and testing results of the SEIN SE-30 TENS device, alongside the known specifications of the predicate device. The testing was conducted by SEIN Electronics Co., Ltd. in Korea. It's a "prospective" evaluation in the sense that the device was specifically tested to meet the standard and demonstrate equivalence.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not applicable in the context of establishing ground truth for image/data interpretation. This summary does not mention any expert review of clinical cases.
  • Qualifications of Experts: Not applicable. The "ground truth" for this regulatory submission is the established performance characteristics of the predicate device and the requirements of the ANSI/AAMI NS4-1985 standard.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. There was no "test set" in the context of clinical data requiring expert adjudication. The comparison was primarily based on objective technical specifications and standard compliance.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

  • MRMC Study: No. This type of study is not typically required or performed for TENS device 510(k) submissions, which focus on electromechanical similarity and performance standard compliance rather than diagnostic accuracy or human interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Standalone Performance: Not applicable. The SE-30 TENS is a physical medical device, not an AI algorithm. Its performance is inherent in its design and operation, not subject to standalone algorithmic testing.

7. The Type of Ground Truth Used

  • Type of Ground Truth: For the purpose of this 510(k) submission, the "ground truth" is:
    • The established technical specifications and therapeutic claims of the legally marketed predicate device (Graham-Field, Inc. Micro TENS Plus).
    • The requirements of the recognized industry standard ANSI/AAMI NS4-1985.
    • The intended use for Transcutaneous Electrical Nerve Stimulators (pain relief).

8. The Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. The SEIN SE-30 TENS is a physical TENS device and does not involve machine learning or a "training set" in the context of AI.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable, as there is no training set for this type of device.

In summary, the K991397 submission for the SEIN SE-30 TENS device demonstrated its "acceptance criteria" were met by proving substantial equivalence to a predicate device through a comparison of technical specifications and by showing compliance with a relevant performance standard (ANSI/AAMI NS4-1985). The framework of AI/software device evaluation (test sets, ground truth by experts, MRMC studies) does not directly apply to this traditional medical device submission.

{0}------------------------------------------------

出事

K99 139 7
11/3

ELECTRONICS CO. , LTD. SEIN

133-3 PYUNGCHON-DONG, DONGAN-GU, ANYANG-SHI, KYUNGKI-DO, KOREA TEL: ANYANG (0343) 421-0389, FAX:(0343)421-5639

510(k) SUMMARY

"This 510(k) summary of Safety and Effectiveness information is being summitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. "

The assigned 510(k) number is _ __________ a market met service and the sense the sense many of

Applicant SEIN Electronics Co.,Ltd. 133-3 Pyungchon-Dong, Dongan-Gu, Anyang-Shi, Kyungki-Do, Korea Tel:Anyang(0343)421-0389 Fax:(0343)421-5639

Mr. S. H. Hwang, President Contact Forecare, Inc. 1540 Barclay Boulevard Buffalo Grove, IL 60089 Tel: (847)541-8894 Fax: (847)541-8979

Date March 16, 1999

Name of Device

Proprietary Name : SE-30 TENS Common or Usual Name : Transcutaneous Electrical Nerve Stimulator Classification Name : Stimulator, Nerve, Transcutaneous (Pain Relief)

The SEIN SE-30 TENS device is equivalent to the Graham-Field, Inc. Micro TENS Plus transcutaneous electrical nerve stimulator.

Each product is used for the symptomatic relief of and management of long term intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic pain.

{1}------------------------------------------------

A comparison of general specifications between the two devices is as follows:

SEINGRAHAM-FIELD
ITEMSE-30 TENSMicro TENS Plus
Output ChannelsOne output channelTwo with individual controls
Pulse ShapeAsymmetric bi-phasic &phasic rectangular shape withzero net direct currentAsymmetric bi-phasic &phasic rectangular shape withzero net direct current
Pulse Width105-450 us programmed insertprotected circuit50-200 us variable
Pulse Rate1-200 pulse/sec programmed1-200 pulse/sec variable
Power Sourcetwo alkaline AAA batteries1.5 volt9 volt alkaline or NiCd bat.(rechargeable option)
Maximum Voltage
Load26 V17 V
200 ohm40 V32 V
500 ohm52 V45 V
1 Kohm90 V90 V
open
Maximum Output
charge/pulse(500 ohm load for safety)58.5 uc/pulse14.3 uc/pulse
Maximum Output
average current(500 ohm load for safety)3.375 mA2.3 mA
Size70(W) x 79(H) x 17(D) mm90(W) x 59(H) x 25(D) mm
Weight80 grams142 grams
Recharge Battery OptionNY
Battery CheckNY
Water Use OptionNY
Discriminate PolalityAutomatically ChangeRequired
Gel PadAdhesive TypeAdhesive Type
Treatment15 min. (no daily limit)Mode 1: [Modulation]1. Pulse Rate : 1-100 pps2. Pulse Width: 220-450 usec15, 30min. (no daily limit)Constant Mode1. Pulse Rate : 1-200 pps2. Pulse Width: 50-200 usec
MODES AVAILABLEMode 2... [Burst]1. Pulse Rate:16 pulse/burst. 4b/sec16 pulse/burst. 7b/sec63-125 pps2. Pulse Width : 105-450 usecBurst Mode1. Pulse Rate:10 pulse/burst. 2b/sec2. Pulse Width: 50-200 usec
Mode 3... [Modulation]1. Pulse Rate : 1-5 pps2. Pulse Width : 220 usecModulation Mode1. Pulse Rate : 1-200 pps2. Pulse Width: 50-200 usec
Mode 4... [Burst]1. Pulse Rate:220 pulse/burst, 1b/sec90 pulse/burst. 1b/sec2. Pulse Width: 220-450 usec

{2}------------------------------------------------

The SE-30 TENS device also was tested according to the standards of ANSI/AAMI NS4-1985 and was found to meet the requirements.

Mar. 16. 1999

Date

Tung

Tae Young Choi, President

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Inside the circle is a stylized symbol consisting of three abstract human profiles facing to the right, stacked on top of each other. The profiles are rendered in a simple, bold line style.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 1 2000

Mr. S. H. Hwang President SEIN Electronics Company, Ltd. C/O Forecare, Inc. 1540 Barclay Boulevard Buffalo Grove, Illinois 60089

Re: K991397

Trade Name: Transcutaneous Electrical Nerve Stimulator SE-30 TENS Regulatory Class: II Product Code: GZJ Dated: November 8, 1999 Received: November 18, 1999

Dear Mr. Hwang:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed

{4}------------------------------------------------

Page 2 – Mr. S. H. Hwang

predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours.

Mark N. Melkerson

James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known):

K991397

Device Name: Transcutaneous Electrical Nerve Stimulator - SE-30 TENS

Indications For Use: the SEIN SE-30 TENS device is indicated for the symptomatic relief and management of chronic (long-term) intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.

PLEASE DO NO WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The-Counter Use

Prescription Use × (Per 21 CFR801.109)

OR

(Optional Format 1-2-96)

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).