K Number
K991371

Validate with FDA (Live)

Device Name
CARESIDE AMMONIA
Manufacturer
Date Cleared
1999-08-03

(105 days)

Product Code
Regulation Number
862.1065
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use with the CARESIDE Analyzer™ to measure ammonia from anti-coagulated whole blood or plasma specimens to aid in the diagnosis and treatment of patients with severe liver disorders such as cirrhosis, hepatitis, and Reye's syndrome.

Device Description

CARESIDE™ Ammonia cartridges are used with the CARESIDE Analyzer™ to measure ammonia in anti-coagulated whole blood or plasma specimens. The CARESIDE™ Ammonia cartridge, a single use disposable in vitro diagnostic test cartridge, delivers a measured volume of plasma to a dry film to initiate the measurement of ammonia The film cartridge (patent pending) contains all reagents necessary to measure ammonia.

Each CARESIDE™ Ammonia cartridge consists of an ammonia-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the specimen into the cartridge Sample Well, closes the lid and inserts the cartridge into the CARESIDE Analyzer™.

Once loaded, the CARESIDE Analyzer™ scans the cartridge barcode, brings the cartridge and the contained specimen to 37°C, and spins the cartridge to move the sample from the Sample Well into the cartridge channels and chambers. 8.5 microliters of sample remains in the metering passage. Any excess sample flows into the Overflow Well.

The 8.5 microliters of sample is automatically dispensed onto the multi-layer reagent film. The spreading layer distributes the sample evenly on the film before the sample moves through the reaction, porous, and detection layers where a blue-green dye forms in the presence of ammonia. The color intensity of the resulting blue-green dye, as measured by the amount of reflected light at 590 nanometers, directly relates to the ammonia concentration of the specimen.

As the cartridges spin, a photodiode measures reflectance of light emitted by a wavelength-specific light emitting diode (LED) over a fixed time period. The analyzer uses the reflectance measurements and the lot-specific standard curve to calculate ammonia concentration.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the CARESIDE™ Ammonia device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

It's important to note that the provided text does not explicitly state pre-defined acceptance criteria in terms of thresholds for satisfactory performance. Instead, it presents the comparative performance characteristics of the CARESIDE™ Ammonia device alongside the predicate device (Vitros NH3 DT Slides) and implicitly argues for substantial equivalence based on these comparisons.

Therefore, the "acceptance criteria" listed below are inferred from the demonstrated performance and the context of establishing substantial equivalence to a legally marketed predicate device. The underlying acceptance is likely that the device performs as well as or better than the predicate device for key analytical metrics.

Acceptance Criterion (Inferred)CARESIDE™ Ammonia Reported PerformancePredicate Device (Vitros NH3 DT Slides) Reported Performance
Detection Limit7 µmol/L1 µmol/L
Reportable Range7 to 350 µmol/L1 to 500 µmol/L
Accuracy (Recovery)Mean recovery 99%Not provided
Precision (Total CV)11% at 109 µmol/L10% at 100 µmol/L
Method ComparisonCARESIDE™ = 0.98 (RAICHEM) + 3.9 µmol/L, r = 0.99Not provided
LinearitySlope and correlation coefficient within acceptable limitsNot provided
InterferenceNo significant interference observed at tested concentrations (Ascorbic Acid, Total Protein, Hemoglobin, Triglycerides, Urea)No reported interference

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document does not explicitly state the sample size used for the performance comparison studies (accuracy, precision, method comparison, linearity, interference).
  • Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • This information is not provided in the text. As an in vitro diagnostic device for quantitative measurement, the "ground truth" would typically come from a reference method of known accuracy rather than expert interpretation of images or clinical findings.

4. Adjudication Method for the Test Set

  • This information is not applicable for this type of analytical device performance study, which relies on quantitative measurements and comparison to a reference method, rather than subjective expert adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a MRMC comparative effectiveness study was not done. This type of study is relevant for imaging devices or diagnostic tests that involve human interpretation of results. The CARESIDE™ Ammonia device is an automated, quantitative in vitro diagnostic, so human "readers" are not directly involved in interpreting the result to the same extent as in, for example, radiology.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Yes, a standalone performance study was done. The entire "Comparative Performance Characteristics" section describes the performance of the CARESIDE™ Ammonia device (algorithm/system only) in generating quantitative results. The device measures ammonia automatically and provides a µmol/L NH3 value. There is no indication of human intervention or interpretation required during the test process itself.

7. Type of Ground Truth Used

  • The ground truth for the performance studies was established using a reference method, specifically the Glutamate dehydrogenase method. This is explicitly stated as the "Reference Method." For the method comparison study, it was compared against a "RAICHEM" method, suggesting another established analytical method.

8. Sample Size for the Training Set

  • This information is not provided in the text. The document describes a specific analytical method and its performance, but not the development or "training" of an algorithm in the sense of machine learning, which would typically involve a separate training set. The "standard curve" is mentioned as lot-specific, which is a form of calibration rather than algorithm training with a large dataset.

9. How the Ground Truth for the Training Set Was Established

  • This information is not provided as a "training set" in the context of machine learning algorithms is not discussed. For traditional analytical devices like this, calibration (which uses a set of known standards to establish the relationship between signal and concentration) is the equivalent of "training," and the ground truth for this would be the precisely known concentrations of the calibrator materials. The text notes "# Calibration information bar-coded on each cartridge. Calibration information may change with each lot." This implies known calibrator values are used for establishing the standard curve.

{0}------------------------------------------------

CARESIDE, Inc. Page 10

510(K) SUMMARY: CARESIDE™ AMMONIA SAFETY V. AND EFFECTIVENESS

I. Applicant Information

  • Applicant Name A.
  • Applicant/Manufacturer Address B.
  • C. Telephone Number
  • Contact Person D.
  • E. FAX Number
  • F. e-Mail Address
  • Date 510(k) Summary prepared G.

II. Device Information

  • A. Device Name (Trade)
  • B. Device Name (Classification)
  • C. Device Classification

CARESIDE, Inc.

6100 Bristol Parkway Culver City, CA 90230 310-338-6767 Kenneth B. Asarch, Pharm.D., Ph.D. 310-338-6789 AsarchK@CARESIDE.com April 16, 1999

CARESIDE™ Ammonia

Ammonia test system Clinical chemistry panel Ammonia test system Regulation Number: 21 CFR 862.1065 Regulatory Class 1 Classification Number: 75JID None applicable

  • D. Special controls and performance standards

III. Substantial Equivalence Claim

  • A. General equivalency claim
    The ability to monitor analyte-specific biochemical reactions in dry film and other formats is widely recognized and has gained widespread acceptance for use in chemistry assays.

Ammonia in vitro diagnostic products, in both dry film and other formats, are already on the U.S. market. Ammonia products include those that use bromophenol blue as an ammonia indicator.

B. Specific equivalency claim

This CARESIDE™ Ammonia test is substantially equivalent in intended use and clinical performance to the currently marketed Vitros slides for the quantitative measurement of ammonia on the Vitros DT 60 II system. Both are based on the principle of dry film, both use the bromphenol blue ammonia indicator, and both are read by reflectance photometry.

Name of Predicate Device: Johnson and Johnson's (formerly Eastman Kodak, Inc.) Vitros NH3 Slides for Johnson and Johnson's Vitros DT 60 II system (formerly Eastman Kodak's DT 60 II). Predicate Device 510K number: K912844/A Product Code: 75JID

{1}------------------------------------------------

IV. Device Description

CARESIDE™ Ammonia cartridges are used with the CARESIDE Analyzer™ to measure ammonia in anti-coagulated whole blood or plasma specimens. The CARESIDE™ Ammonia cartridge, a single use disposable in vitro diagnostic test cartridge, delivers a measured volume of plasma to a dry film to initiate the measurement of ammonia The film cartridge (patent pending) contains all reagents necessary to measure ammonia.

Explanation of Device Function A. .

Each CARESIDE™ Ammonia cartridge consists of an ammonia-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the specimen into the cartridge Sample Well, closes the lid and inserts the cartridge into the CARESIDE Analyzer™.

Once loaded, the CARESIDE Analyzer™ scans the cartridge barcode, brings the cartridge and the contained specimen to 37°C, and spins the cartridge to move the sample from the Sample Well into the cartridge channels and chambers. 8.5 microliters of sample remains in the metering passage. Any excess sample flows into the Overflow Well.

The 8.5 microliters of sample is automatically dispensed onto the multi-layer reagent film. The spreading layer distributes the sample evenly on the film before the sample moves through the reaction, porous, and detection layers where a blue-green dye forms in the presence of ammonia. The color intensity of the resulting blue-green dye, as measured by the amount of reflected light at 590 nanometers, directly relates to the ammonia concentration of the specimen.

Test Reaction Sequence:

Bromophenol blue (yellow) + NH3 -- > Blue-green dye

As the cartridges spin, a photodiode measures reflectance of light emitted by a wavelength-specific light emitting diode (LED) over a fixed time period. The analyzer uses the reflectance measurements and the lot-specific standard curve to calculate ammonia concentration.

  • B. Test Summary
    The majority of ammonia in the body is generated in the gastrointestinal tract through the action of bacterial enzymes on dietary protein in the colon and from the catabolism of amino acids. Ammonia is detoxified in the liver by conversion to urea. Elevated ammonia levels are caused by liver dysfunction from such causes as viral hepatitis, cirrhosis, or Reye's syndrome, and inherited deficiencies of enzymes involved in the conversion of ammonia to urea. Excess ammonia exerts toxic effects on the central nervous system,

V. Intended Use

11 23

  • A. Intended Use
    The CARESIDE™ Ammonia cartridge is intended for in vitro diagnostic use in conjunction with the CARESIDE Analyzer™ to quantitatively measure ammonia in anticoagulated whole blood or plasma.

  • B. Indications for Use
    This product is indicated for use in the diagnosis and treatment of patients with severe liver disorders such as cirrhosis, hepatitis, and Reye's syndrome.

{2}------------------------------------------------

t

VI. Technological Characteristics

  • Similarities A.
CARESIDE™ AmmoniaVitros NH3 DT Slides
Intended UsePrimarily to aid in thediagnosis and treatment ofpatients with severe liverdisorders such as cirrhosis,hepatitis, and Reye'ssyndrome.Primarily to aid in thediagnosis of hepatic coma andReye's syndrome, as anindicator during the finalstages of terminal cirrhosis.
IndicationsFor in vitro diagnostic use.For professional laboratory:not for point of care orphysician office laboratoryuse.For in vitro diagnostic use.
MeasurementQuantitativeSame
Method PrincipleDry film, bromophenol bluedye.Dry film, bromophenol bluedye.
MaterialsBromophenol blueBromophenol blue
DetectorReflectance (590 nm)Reflectance (605 nm)
Test time4 minute warm-up (on-board)plus 3 minute test time.15 minute warm-up (off-line)plus 5 minutes test time.
Reference MethodGlutamate dehydrogenaseGlutamate dehydrogenase
Sample TypeAnti-coagulated whole bloodor plasma.Heparinized plasma
Specimen volume8.5 µl test volume(85 ± 15 µl applied volume)10 µl
CalibrationCalibration information bar-coded on each cartridge.Calibration information maychange with each lot.Run Vitros DT II calibratorswhenever a new slide lot isused or when necessary.
Quality Control2 levelsSame
Reporting Unitsµmol/L NH3Same
Reaction Temp.37 °CSame

B. Differences

Qaran

CARESIDE™ AmmoniaVitros NH3 DT Slides
SpecimenProcessingNot requiredRequired
AccuratepipettingNot requiredRequired
Reagent pre-warmingNot requiredRequired

{3}------------------------------------------------

C. Comparative Performance Characteristics

CARESIDE™ AmmoniaVitros NH3 DT Slides
Detection limit7 µmol/L1 µmol/L
Reportable range7 to 350 µmol/L1 to 500 µmol/L
AccuracyMean recovery 99%Not provided
PrecisionTotal CV, 109 µmol/L, 11%Total CV, 100 µmol/L, 10%
MethodcomparisonCARESIDE™ = 0.98 (RAICHEM) + 3.9 µmol/L, r = 0.99
LinearityLinearity yielded slope andcorrelation coefficient withinacceptable limits.Not provided
InterferenceNo significant interferenceobserved at testedconcentration of interferent:Ascorbic Acid, ........20 mg/dLTotal Protein, ........12 g/dLHemoglobin ........500 mg/dLTriglycerides ........3000 mg/dLUrea (BUN).........40 mg/dLNo reported interference

D. Conclusion

·

..............................................................................................................................................................................

The nonclinical and clinical data provided demonstrate that the CARESIDE™ Ammonia product is as safe, effective, and performs as well as or better than the legally marketed predicate device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three curved lines that suggest the head and wings.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AUG - 3 1999

Kenneth B. Asarch, Ph.D. VP Quality Systems and Regulatory Affairs CARESIDE, Inc. 6100 Bristol Parkway Culver City, California 90230

Re: K991371

Trade Name: CARESIDE™ AMMONIA Regulatory Class: I reserved Product Code: JID Dated: June 15, 1999 Received: June 16, 1999

Dear Dr. Asarch:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Gutman

Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

CARESIDE, Inc. Page 15

VII. INDICATIONS FOR USE

510(k) Number:

Device Name:

CARESIDE™ Ammonia

Indications for use:

For in vitro diagnostic use with the CARESIDE Analyzer™ to measure ammonia from anti-coagulated whole blood or plasma specimens to aid in the diagnosis and treatment of patients with severe liver disorders such as cirrhosis, hepatitis, and Reye's syndrome.

Dean Cooper
(Division Sign-Off)

Division of Clinical Laboratory Devices 510/k) Number

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use (Optional Format 1-2-96)

§ 862.1065 Ammonia test system.

(a)
Identification. An ammonia test system is a device intended to measure ammonia levels in blood, serum, and plasma, Ammonia measurements are used in the diagnosis and treatment of severe liver disorders, such as cirrhosis, hepatitis, and Reye's syndrome.(b)
Classification. Class I.