K Number
K991062

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-09-21

(174 days)

Product Code
Regulation Number
866.3830
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QualiCode™ B. burgdorferi IgM Western Blot Kit is an in vitro qualitative assay for the detection of human IgM antibodies reactive with Borrelia burgdorferi antigens present on a membrane strip. The QualiCode™ B. burgdorferi IgM Western Blot Kit is intended for supplemental testing of human serum specimens which yield positive or equivocal results on B. burgdorferi ELISA or IFA screening assays. QualiCode test results can provide additional, more specific evidence of infection with B. burgdorferi which may be useful in the diagnosis of Lyme disease.

The QualiCode™ B. burgdorferi IgM Western Blot Kit can be used to test human sera for IgM antibodies during the acute phase of disease, within one month of infection, in accordance with CDC/ASTPHLD guidelines. During this period, patients may not yet have developed a detectable IgG response. Beyond one month from infection, an IgG Western Blot test is recommended. Patients should not be tested solely for IgM antibodies at this stage due to the higher probability of false-positive results. Symptomatic patients showing IgM but not IgG antibodies at an early stage in infection should be re-tested for IgG antibodies 2-4 weeks later.

Device Description

The device is a Western Blot assay. Proteins and other antigenic components of the Borrelia spirochete are fractionated by polyacrylamide gel electrophoresis in the presence of sodium dodecyl sulfate. The separated proteins are electrophoretically transferred from the gel to nitrocellulose membranes. which are subsequently blocked to minimize non-specific binding and cut into strips. These nitrocellulose strips with resolved Borrelia burgdorferi antigens are then reacted with diluted serum and controls (positive and negative sera of defined reactivity).

During the incubation period, human antibodies specific to the B, burgdorferi antigens, if present in the sample or control, will bind to the antigen to which they have affinity. Unbound serum and non-specific antibodies are washed from the strip. Detection of bound IgM antibodies is accomplished by reacting and incubating the strips with a solution containing anti-human IgM antibodies conjugated with alkaline phosphatase. Unbound conjugate antibodies are removed by washing. The qualitative assessment of the detected IgM antibodies is then accomplished by the reaction of the alkaline phosphatase with a chemical substrate, which is cleaved into a colored, insoluble product that can be visualized. The determination of the reactivity of each unknown specimen is accomplished by comparison of the identified, visualized bands to the Band Identifying and Band Intensity Controls.

AI/ML Overview

Here's an analysis of the provided text to extract the requested information about device acceptance criteria and studies:

Immunetics QualiCode™ B. burgdorferi IgM Western Blot Kit (K991062)

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria (e.g., "The device must achieve X% sensitivity and Y% specificity"). Instead, it presents the performance characteristics found during the clinical trials. We can infer the expected performance based on these reported values.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
SpecificityHigh specificity across normal and cross-reactive populationsNormal - Overall: 96% (95% CI: 93-97%)
Normal - Endemic Region: 95% (95% CI: 92-97%)
Normal - Non-Endemic Region: 96% (95% CI: 92-98%)
Cross-Reactive Panel: 87% (95% CI: 88-96%)
SensitivityHigh sensitivity particularly in the early acute phase (<1 month)Draw Time <1 month: 91% (95% CI: 82-96%)
Declining sensitivity for later stagesDraw Time Unknown: 74% (95% CI: 60-85%)
Draw Time 1-2 months: 83% (95% CI: 61-94%)
Draw Time 3-12 months: 63% (95% CI: 35-82%)
Draw Time >12 months: 47% (95% CI: 25-72%)
Inter-Lot ReproducibilityHigh agreement between different kit lots80% Interpretation Agreement
Inter-Run ReproducibilityHigh agreement between different runs85% Interpretation Agreement (overall, averaged over 3 sites)
Inter-Reader ReproducibilityHigh agreement between different readers93% Interpretation Agreement (overall, averaged over 3 sites)

2. Sample Sizes and Data Provenance for the Test Set

  • Specificity Test Set:

    • Normal Population: n=430 (279 from Lyme disease endemic regions, 151 from non-endemic regions).
    • Cross-Reactive Panel: n=172 (samples from patients with diseases other than Lyme disease that may be cross-reactive).
    • Total Specificity Samples: 602
    • Provenance: Not explicitly stated by country, but regional (endemic/non-endemic) and from clinical trial sites. The data is retrospective, as these were "well-characterized, archived Lyme Disease specimens" and other pre-existing panels.
  • Sensitivity Test Set:

    • Lyme Disease Panel: n=186 (samples from patients with a clinical diagnosis of Lyme disease).
    • Provenance: Not explicitly stated by country, but from clinical trial sites. The data is retrospective, consisting of "well-characterized, archived Lyme Disease specimens." These samples were drawn at different times after onset of disease.
  • Reproducibility Test Set:

    • Panel: n=20 specimens (including positive, weakly reactive, and negative sera).
    • Provenance: Not specified, but likely internal or from the clinical trial sites.

3. Number of Experts and their Qualifications for Ground Truth

The document does not explicitly state the number of experts or their specific qualifications for establishing the ground truth for the test set. However, it does refer to:

  • "Clinical diagnosis of Lyme Disease in accordance with the CDC case definition, i.e. based on the presence of EM (erythema migrans) or the presentation of late Lyme clinical manifestations (e.g., arthritic, cardiac, or neurological symptoms)." This implies diagnosis by clinicians.
  • "Infection was confirmed by culture of Borrelia from biopsies in many examples."
  • "Specimens selected for the trial were required to have tested positive or indeterminate on a B. burgdorferi screening assay, typically an ELISA, in accordance with the two-tier testing protocol recommended by CDC/ASTPHLD." This indicates that initial characterization involved standard serological testing overseen by labs or clinicians.
  • The "Lyme Disease Serum Panel" from the Centers for Disease Control (CDC) was also used, which would have its own established ground truth.

For the reproducibility studies, "two independent readers" were used for inter-reader reproducibility, but their qualifications are not specified.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method (like 2+1 or 3+1) for establishing the ground truth of the test set specifically for this study. Instead, the samples were "well-characterized" based on clinical diagnosis, culture confirmation, and pre-existing serological testing (two-tier protocol) under CDC/ASTPHLD guidelines. The CDC Lyme Disease Serum Panel also likely has its ground truth externally defined and accepted.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported. The study focuses on the standalone performance of the device and its reproducibility characteristics (inter-lot, inter-run, and inter-reader agreement), not on comparing human readers with and without AI assistance.

6. Standalone Performance Study (Algorithm Only)

Yes, a standalone performance study was done. The entire "Summary of Performance" section describes the device's performance (sensitivity, specificity, reproducibility) on various sample panels without human interpretation assistance for the core performance metrics. The result interpretations from the device were compared against the established ground truth of the samples.

7. Type of Ground Truth Used for the Test Set

The ground truth for the test set was a combination of:

  • Clinical Diagnosis: Based on CDC case definition (erythema migrans or late Lyme clinical manifestations).
  • Pathology/Culture Confirmation: Infection confirmed by culture of Borrelia from biopsies in many examples.
  • Established Serological Testing: Initial positive or indeterminate results on B. burgdorferi screening assays (ELISA/IFA) following CDC/ASTPHLD two-tier testing protocol.
  • Reference Panels: Utilization of the Centers for Disease Control Lyme Disease Serum Panel, which comes with its own established reference status.

8. Sample Size for the Training Set

The document does not provide information on a specific training set or its sample size. The description of "clinical trials" and "characterized samples" refers to the general methodology for assessing the device's performance, but there is no explicit mention of data used to "train" the assay itself in the context of machine learning or AI. This is a traditional in-vitro diagnostic (IVD) device, not an AI/ML-based device.

9. How Ground Truth for the Training Set Was Established

Since there's no mention of a dedicated training set in the context of an AI/ML algorithm, this question is not applicable. The "training" of a traditional IVD device like this is typically through its design, reagent selection, and optimization during development, rather than a data-driven training set used for machine learning. The clinical performance data presented here serves to validate the final device's performance against established ground truths.

{0}------------------------------------------------

510(k) Safety and Effectiveness Summary

Submitted by:

Immunetics, Inc. 63 Rogers Street Cambridge, MA 02142

Contact Person:

Andrew E. Levin, Ph.D. Scientific Director

(617) 492 - 5416

Date of Preparation:

September 9, 1999

1. Name and Address of Owner/Operator and Manufacturer

Immunetics, Inc. 63 Rogers Street Cambridge, MA 02142

2. Product Name

Trade Name: QualiCode™ B. burgdorferi IgM Western Blot Kit

Common Name: B. burgdorferi IgM Western Blot Kit

3. Claim of Substantial Equivalence

The characterized samples used for the establishment of Substantial Equivalence were obtained from patients with a clinical diagnosis of Lyme Disease in accordance with the CDC case definition, i.e. based on the presence of EM (erythema migrans) or the presentation of late Lyme clinical manifestations (e.g., arthritic, cardiac, or neurological symptoms). Infection was confirmed by culture of Borrelia from biopsies in many examples.

Each of the clinical trial sites provided specimens that were well characterized by the site using Lyme-specific serological analyses, including EIA and/or Western Blot testing. In particular, specimens selected for the trial were required to have tested positive or indeterminate on a B. burgdorferi screening assay, typically an ELISA, in accordance with the two-tier testing protocol recommended by CDC/ASTPHLD (CDC (1995) "Recommendations for test performance and interpretation from the Second

{1}------------------------------------------------

National Conference on Serologic Diagnosis of Lyme Disease", Morbid. Mort. Weekly Rep. 44:590-591.).

Substantial equivalence of this device is based on the assessment of performance of the device in these clinical trials in which the well-characterized, archived Lyme Disease specimens, the Centers for Disease Control Lyme Disease Serum Panel, normal donor specimens (from endemic and non-endemic regions), and samples from diverse disease conditions were analyzed.

4. Description

The device is a Western Blot assay. Proteins and other antigenic components of the Borrelia spirochete are fractionated by polyacrylamide gel electrophoresis in the presence of sodium dodecyl sulfate. The separated proteins are electrophoretically transferred from the gel to nitrocellulose membranes. which are subsequently blocked to minimize non-specific binding and cut into strips. These nitrocellulose strips with resolved Borrelia burgdorferi antigens are then reacted with diluted serum and controls (positive and negative sera of defined reactivity).

During the incubation period, human antibodies specific to the B, burgdorferi antigens, if present in the sample or control, will bind to the antigen to which they have affinity. Unbound serum and non-specific antibodies are washed from the strip. Detection of bound IgM antibodies is accomplished by reacting and incubating the strips with a solution containing anti-human IgM antibodies conjugated with alkaline phosphatase. Unbound conjugate antibodies are removed by washing. The qualitative assessment of the detected IgM antibodies is then accomplished by the reaction of the alkaline phosphatase with a chemical substrate, which is cleaved into a colored, insoluble product that can be visualized. The determination of the reactivity of each unknown specimen is accomplished by comparison of the identified, visualized bands to the Band Identifying and Band Intensity Controls.

5. Intended use

The QualiCode™ B. burgdorferi IgM Western Blot Kit is an in vitro qualitative assay for the detection of human IgM antibodies reactive with Borrelia burgdorferi antigens present on a membrane strip. The QualiCode™ B. burgdorferi IgM Western Blot Kit is intended for supplemental testing of human serum specimens which yield positive or equivocal results on B. burgdorferi ELISA or IFA screening assays. QualiCode test results can provide additional, more specific evidence of infection with B. burgdorferi which may be useful in the diagnosis of Lyme disease.

The QualiCode™ B. burgdorferi IgM Western Blot Kit can be used to test human sera for IgM antibodies during the acute phase of disease, within one month of infection, in accordance with CDC/ASTPHLD guidelines. During this period, patients may not yet have developed a detectable IgG response. Beyond one month from infection, an IgG Western Blot test is recommended. Patients should not be tested solely for IgM

{2}------------------------------------------------

antibodies at this stage due to the higher probability of false-positive results. Symptomatic patients showing IgM but not IgG antibodies at an early stage in infection should be re-tested for IgG antibodies 2-4 weeks later.

Summary of Performance 6.

From a summary of the clinical trial data, the following performance characteristics are described:

Expected values

Three (3) investigational sites, including Immunetics and two (2) independent off-site investigators, assayed samples from the following patient populations:

    1. Normal population (n=430) comprised of samples from Lyme disease endemic (n=279) and non-endemic (n=151) regions.
    1. Cross Reactive Panel (n=172) comprised of samples from patients with diseases other than Lyme disease that may be cross - reactive.
    1. Lyme Disease Panel (n=186) comprised of samples from patients with a clinical diagnosis of Lyme disease, based on presence of erythema migrans or one or more symptoms of late Lyme disease, and which tested positive or equivocal on ELISA or other screening assays for B. burgdorferi antibodies.

Specificity

Specificity of the device was determined from analysis of results of testing normal donor specimens from endemic and non-endemic regions and potentially cross reactive disease specimens (602 total samples). The specificity values derived from testing each population are shown in the table below in the "% Negative" column:

Sample Typen% Positive% Negative95% CI
Normal - endemic region2795%95%92-97 %
Normal - non-endemic region1514%96%92-98 %
Normal - overall4304%96%93-97 %
Cross - Reactive Panel17213%87%88-96%

Sensitivity

Sensitivity of the device was determined by testing a total of 186 well-characterized sera from patients with Lyme disease, which had been drawn at different times after onset of disease. Sensitivity for each time after onset category is shown in the table below in the "% Positive" column:

{3}------------------------------------------------

Sensitivity of the Immunetics, Inc. QualiCode™ B. burgdorferi IgM Western Blot Kit vs. Time after Onset

Draw Time (months)n% Positive% Negative95% CI
Unknown5474%26%60-85 %
<17691%9%82-96 %
1-22383%17%61-94 %
3-121663%37%35-82 %
>121747%53%25-72 %

Reproducibility

Inter-lot reproducibility was determined by assaying a panel of 20 specimens, including positive, weakly reactive and negative sera, on three lots of the Immunetics QualiCode™ kit. There was 80% agreement between interpretations from the three kit lots. The reproducibility of scoring of individual bands between the three lots varied from 80% to 100%, averaging 88%.

Inter-run reproducibility was determined by assaying the 20 specimen panel twice, on separate days, at each of three sites. There was 85% agreement overall between interpretations from the two runs averaged over all three sites. The reproducibility of individual band scoring between the two runs averaged over all three sites varied between 75% and 93%, with an average of 85% over all three bands.

Inter-reader reproducibility was determined by assaying the 20 specimen panel twice, on separate days, at each of three sites, with each strip interpreted by two indevendent readers. There was 93% agreement overall between interpretations from the two readers averaged over the three sites. Band scoring reproducibility varied between 97% and 98% averaged over the three sites, with an average of 97% over all three bands.

Studyn% Interpretation Agreement
Inter-lot2080
Inter-run2085
Inter-reader2093

7. Conclusions

Based on the clinical performance, this device has been shown to be safe and effective for the intended use in the qualitative detection of human Immunoglobulin M (IgM) antibodies in serum or plasma to Borrelia burgdorferi antigens, and as a supplemental, more specific, test to aid in the diagnosis of infection or exposure to Borrelia burgdorferi, the causative agent of Lyme disease.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

SEP 2 1 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Andrew E. Levin. Ph.D. Scientific Director Immunetics, Inc. 63 Rogers Street Cambridge, Massachusetts 02142

Re: K991062

Trade Name: QualiCode™ B. burgdorferi IgM Western Blot Kit Regulatory Class: II Product Code: LSR Dated: July 8, 1999 Received: July 9, 1999

Dear Dr. Levin:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known):

Device Name: QualiCode B. burgdorferi IgM Western Blot Kit

Indications For Use:

The Immunetics QualiCode" B. burgdorferi IgM Western Blot Kit is intended for use in testing human serum samples which have demonstrated positive or eguivocal responses using EIA or IFA test procedures to provide supportive evidence of infection with Borrelia burgdorferi.

The QualiCode B. burgdorferi IgM Western Blot Kit can be used to test human sera for IqM antibodies during the acute phase of disease, within one month of infection, in accordance with CDC/ASTPHLD During this period, patients may not yet have developed quidelines. a detectable IgG response. Beyond one month from infection, an IgG Western Blot test is recommended. Patients should not be tested solely for IgM antibodies at this stage due to the higher probability of false positive results. Symptomatic patients showing IgM but not IqG antibodies at an early stage in infection should be re-tested for IgG antibodies 2-4 weeks later.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Woody Dubres

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional format 1-2-96)

§ 866.3830

Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).