K Number
K990852

Validate with FDA (Live)

Date Cleared
1999-06-11

(88 days)

Product Code
Regulation Number
876.4300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CooperSurgical IRC is indicated for the treatment of hemorrhoids, especially bleeding, non-prolapsed internal hemorrhoids, as a means of hemostasis in the treatment of hemorrhoids by routine excision, or to coagulate external thrombotic hemorrhoids that have been removed ("collectively referred to as hemorrhoids") through the coagulative necrosis of the mucosa proximal to the base of the hemorrhoid.

Device Description

The Cooper IRC delivers short pulses of visible and infrared light through a small contact tip applicator that is applied to the tissue. This light causes thermal coagulation that results in tissue necrosis. The user sets the pulse duration control knob depending on the depth of the tissue necrosis required. The depth of coagulation is directly related to the pulse length delivered to a given area of tissue. The area(s) to be treated may be locally anesthetized at the physician's discretion.

The CooperSurgical IRC consists of the following main components: (1) a console unit; (2) a hand piece; (3) a removable light guide; and (4) a disposable contact tip.

AI/ML Overview

This document is a 510(k) summary for CooperSurgical, Inc.'s Infrared Coagulator (IRC), seeking clearance for the treatment of hemorrhoids. It details the device's intended use, principles of operation, technical characteristics, and the basis for its substantial equivalence to previously cleared devices.

Here's an analysis of the provided information, specifically addressing the questions about acceptance criteria and supporting studies:

  1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state specific acceptance criteria in terms of numerical performance metrics (e.g., sensitivity, specificity, accuracy, or a specific success rate for treatment). Instead, the basis for clearance is substantial equivalence to predicate devices. This means the device is deemed safe and effective because it is essentially the same as a device already on the market.

    Acceptance CriteriaReported Device Performance
    Implicit Criterion: Substantial Equivalence to Predicate Devices for the Treatment of Hemorrhoids. (This implies it must perform similarly in safety and effectiveness as the predicate devices for the same indication.)The CooperSurgical IRC is "essentially the same device" as Redfield Corporation's Infrared Coagulator already cleared for hemorrhoid treatment. It is also "identical" (with the exception of four features which are not specified as relevant to hemorrhoid treatment or the PVC tips) to CooperSurgical's own IRC cleared for genital condylomas and general warts. The safety and effectiveness for hemorrhoid treatment are based on the FDA's prior clearance of these predicate devices and "CooperSurgical's testing of the PVC contact tips."
    Implicit Criterion: The device's technical characteristics (console unit, hand piece, removable light guide, disposable contact tip) must be comparable to the predicate devices and suitable for its intended use.The technical characteristics are described as: "The CooperSurgical IRC consists of the following main components: (1) a console unit; (2) a hand piece; (3) a removable light guide; and (4) a disposable contact tip. The CooperSurgical IRC is essentially the same device as Redfield Corporation's Infrared Coagulator that has already been cleared by FDA for the treatment of hemorrhoids. With the exception of four features, the CooperSurgical IRC is identical to the CooperSurgical IRC that is cleared for the treatment of genital condylomas and general warts." The functioning principle (delivery of short pulses of visible and infrared light, causing thermal coagulation and tissue necrosis) is also described, aligning with known IRC technology. The user can set pulse duration based on desired depth of tissue necrosis.
    Implicit Criterion: The PVC contact tips used with the device must be safe and effective."CooperSurgical's testing of the PVC contact tips" was performed. (No details on the nature or results of this testing are provided in the summary.)
  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not report any clinical study data from a "test set" in the traditional sense (i.e., a clinical trial comparing the device's performance against a reference standard or control group). The basis for clearance is substantial equivalence. Therefore, there is no sample size, data provenance, or information on retrospective/prospective studies mentioned that specifically tested this version of the CooperSurgical IRC for hemorrhoids. The "testing of the PVC contact tips" is mentioned, but no details regarding sample size or nature of testing are provided.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. As noted above, there is no disclosed "test set" and thus no mention of experts establishing ground truth for performance metrics.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No "test set" or clinical performance study for this specific device for this indication is described.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an Infrared Coagulator, a treatment device, not an AI-assisted diagnostic or interpretation device. Therefore, no MRMC study or AI assistance is relevant.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a manual medical instrument, not an algorithm.

  7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    Not applicable. The clearance is based on substantial equivalence, not on a de novo clinical trial with explicit ground truth determination for performance evaluation. The "ground truth" for the predicate devices' clearance would have established their safety and effectiveness, but those details are not provided here.

  8. The sample size for the training set

    Not applicable. This device is not an AI/ML-based device that requires a training set.

  9. How the ground truth for the training set was established

    Not applicable. This device is not an AI/ML-based device.

{0}------------------------------------------------

JUN 11 1999

510(K) SUMMARY FOR COOPERSURGICAL, INC.'S INFRARED COAGULATOR

K990852

Submitter's Name, Address, Telephone Number, and Contact Person

CooperSurgical, Inc. 15 Forest Parkway Shelton, Connecticut 06484

Contact:Deb PekarManager, Regulatory Affairs, CooperSurgical, Inc.
Phone:(203) 929-6321
Facsimile:(203) 925-0135

Date Prepared: March 15, 1999

Name of the Device

CooperSurgical InfraRed Coagulator

Common or Usual Name

Infrared Coagulator

Predicate Devices

  • (1) Redfield Corporation's Redfield Infrared Coagulator for the treatment of hemorrhoids, especially bleeding, non-prolapsed internal hemorrhoids, as a means of hemostasis in the treatment of hemorrhoids by routine excision, or to coagulate external thrombotic hemorrhoids that have been removed; and
  • (2) CooperSurgical's InfraRed Coagulator for the treatment of genital condylomas (condyloma accuminata) and general warts.

Intended Use

The CooperSurgical IRC is indicated for the treatment of hemorrhoids, especially bleeding, non-prolapsed internal hemorrhoids, as a means of hemostasis in the treatment of hemorrhoids by routine excision, or to coagulate external thrombotic hemorrhoids that have been removed ("collectively referred to as hemorrhoids") through the coagulative necrosis of the mucosa proximal to the base of the hemorrhoid.

{1}------------------------------------------------

Principles of Operation

The Cooper IRC delivers short pulses of visible and infrared light through a small contact tip applicator that is applied to the tissue. This light causes thermal coagulation that results in tissue necrosis. The user sets the pulse duration control knob depending on the depth of the tissue necrosis required. The depth of coagulation is directly related to the pulse length delivered to a given area of tissue. The area(s) to be treated may be locally anesthetized at the physician's discretion.

Page 2 of 2

K990852

Contact photocoagulation requires direct contact of the entire optical window of the disposable contact tip. The physician applies light mechanical pressure to the optical window against the tissue to be treated before depressing the activation trigger of the hand piece, waiting at least five seconds between exposures. To achieve coagulation, the user depresses the activation trigger. A built-in digital timer deactivates the lamp automatically according to the pulse duration setting preselected by the physician.

Technical Characteristics

The CooperSurgical IRC consists of the following main components: (1) a console unit; (2) a hand piece; (3) a removable light guide; and (4) a disposable contact tip. The CooperSurgical IRC is essentially the same device as Redfield Corporation's Infrared Coagulator that has already been cleared by FDA for the treatment of hemorrhoids. With the exception of four features, the CooperSurgical IRC is identical to the CooperSurgical IRC that is cleared for the treatment of genital condylomas and general warts.

Summary of the Basis for the Finding of Substantial Equivalence

The safety and effectiveness of the CooperSurgical IRC for the treatment of hemorrhoids is based on FDA's clearance of: (1) Redfield's Infrared Coagulator for the treatment of hemorrhoids; (2) CooperSurgical's InfraRed Coagulator for the treatment of genital condylomas and general warts; and (3) CooperSurgical's testing of the PVC contact tips.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three curved lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 1 1 1999

CooperSurgical, Inc. c/o Mr. Jonathan S. Kahan Hogan & Hartson, L.L.P. 555 Thirteenth Street, N.W. Washington, D.C. 20004

Re: K990852 CooperSurgical InfraRed Coaqulator Dated: March 15, 1999 Received: March 15, 1999 Requiatory Class: II 21 CFR §876.4300/ Procode : 78 KNS

Dear Mr. Kahan:

We have reviewed vour Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regult in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compiance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours.

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: CooperSurgical InfraRed Coagulator

Indications for Use:

The CooperSurgical InfraRed Coagulator is indicated for the treatment of hemorrhoids, especially bleeding, non-prolapsed internal hemorrhoids, as a means of hemostasis in the treatment of hemorrhoids by routine excision, or to coagulate external thrombotic hemorrhoids that have been removed through coagulative necrosis of the mucosa proximal to the base of the hemorrhoid.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use | (Per 21 C.F.R. 801.109)

(Division Sign-Off)

and Radiological Dev 510(k) Number.

Division of Reproductive, Abdominal, B

OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).