K Number
K990379

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-03-04

(24 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cannulas and accessories are to be used with a Dexide seal package and spike for creating a point of entry for laparoscopic instruments into the abdominal cavity and maintaining pneumoperitoneum.

MultAport Cannulas:
The MultAport reusable cannula is used with an AccuPort Seal or Orbital Reducer Seal and Trocar Spike for creating a point of entry for laparoscopic instruments into the abdominal cavity.

AccuPort 5-12mm Seal:
The AccuPort Seal is a single-used during laparoscopy with MultAport Cannulas to maintain pneumoperitoneum while accommodating instruments with diameters ranging from 5mm to 12mm.

Orbital Reducer Seal:
The Orbital Reducer is a single used during laparoscopy with MultAport Cannulas to maintain pneumoperitoneum while accommodating instruments of equal and smaller size.

Blunt Spike:
The reusable Blunt Trocar spike is used with the MultAport Cannulas and AccuPort or Orbital Reducer Seal for creating a point of entry for laparoscopic instruments into the abdominal cavity.

Sealing Cone:
The Sealing Cone is a reusable product designed for use during open laparoscopy procedures. Used in conjunction with the MultAport Cannula and Blunt Spike, the Sealing Cone provides a means to maintain pneumoperitoneum.

Shaft Seal (replacement part for Sealing Cone):
The Sealing Cone is a reusable product designed for used during open laparoscopy procedures. Used in conjunction with the MultAport Cannula and Blunt Spike, the Sealing Cone provides a means to maintain pneumoperitoneum. The shaft seal is designed for use when sealing the cone to the cannula.

Device Description

The "candidate device" is a reusable, stainless steel Cannula anc disposable, single-use reduction seal system that threads into the Cannula. The reducer system is composed of silicon rubber, thermoplastic rubber, and plastics. The reusable Sealing Cone and its Seal accommodate the cannula to assist the surgical procedure.

AI/ML Overview

This document is a 510(k) summary for the MultAport™ Cannula, Reducer and Accessories. It focuses on the substantial equivalence of this device to a previously cleared predicate device (K981941) rather than presenting a de novo study with detailed acceptance criteria and performance data in the typical sense for an AI/ML medical device.

Therefore, the requested information points (1-9) regarding acceptance criteria and a study proving performance as one might expect for a novel AI/ML diagnostic or prognostic device are not present in this document.

This 510(k) submission establishes equivalence based on:

  • Identical Intended Use and Operating Principle: Both the candidate and predicate device are used for creating an entry point for laparoscopic instruments and maintaining pneumoperitoneum.
  • Similar Basic Design Principles and Materials: The devices share fundamental design and material compositions.
  • Same Shelf Life, Similar Packaging, and Identical Sterilization Instructions: These aspects are also comparable.

Because this is a 510(k) for a physical medical device (cannula, reducer, accessories) claiming substantial equivalence to an existing device, it does not involve AI/ML components, clinical studies for diagnostic accuracy, or performance metrics like sensitivity, specificity, AUC, or the need for ground truth establishment.

In summary, none of the requested information points (1-9) can be extracted from the provided text because the document is a 510(k) for a conventional medical device seeking clearance through substantial equivalence, not a performance study of a novel device (especially not an AI/ML device).

{0}------------------------------------------------

3/4/99

Attachment 3

990379

SUMMARY OF SAFETY AND EFFECTIVENESS

SPECIAL 510(k) SUMMARY

1)Submitter:Dexide, Inc.7509 Flagstone DriveFort Worth, TX 76118-6995Phone No.: (817) 589-1454
Contact Person:John CorzineVice President, Research and Development
Date Prepared:February 1, 1999
2)Name of Device:MultAport™ Cannula, Reducer and Accessories
Common Name:Cannula, Reducer Seal, and Spike
Classification Name:Laparoscopic, General & Plastic Surgery
3)Predicate Device:Dexide Inc., MultAport™ Cannula, Reducer and Accessories,K981941
4)Description of Device:The "candidate device" is a reusable, stainless steel Cannula ancdisposable, single-use reduction seal system that threads into theCannula. The reducer system is composed of silicon rubber,thermoplastic rubber, and plastics. The reusable Sealing Coneand its Seal accommodate the cannula to assist the surgicalprocedure.
5)Intended Use:The Cannulas and accessories are to be used with a Dexide sealpackage and spike for creating a point of entry for laparoscopicinstruments into the abdominal cavity and maintainingpneumoperitoneum.
  • The candidate device has comparable technological characteristics to the predicate device in () that it has the same intended use and operating principle. The same basic design principles and incorporates the same basic materials. The same shelf life, similar packaging and the instructions for sterilization are identical.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three curved lines representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR - 4 1999

Mr. John Corzine Vice President, Research and Development Dexide, Inc. 7509 Flagstone Drive Fort Worth, Texas 76118

Re: K990379

Trade Name: MultAport™ Cannula, Reducer and Accessories Regulatory Class: II Product Code: GCJ Dated: February 1. 1999 Received: February 8, 1999

Dear Mr. Corzine:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Mr. John Corzine

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours.

Stipt Quvelo

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known): K 990379

Device Name: MultAport™ Cannula, Reducer and Accessories

Indications For Use:

MultAport Cannulas:

The MultAport reusable cannula is used with an AccuPort Seal or Orbital Reducer Seal and Trocar Spike for creating a point of entry for laparoscopic instruments into the abdominal cavity.

AccuPort 5-12mm Seal:

The AccuPort Seal is a single-used during laparoscopy with MultAport Cannulas to maintain pneumoperitoneum while accommodating instruments with diameters ranging from 5mm to 12mm.

Orbital Reducer Seal:

The Orbital Reducer is a single used during laparoscopy with MultAport Cannulas to maintain pneumoperitoneum while accommodating instruments of equal and smaller size.

Blunt Spike:

The reusable Blunt Trocar spike is used with the MultAport Cannulas and AccuPort or Orbital Reducer Seal for creating a point of entry for laparoscopic instruments into the abdominal cavity.

Sealing Cone:

The Sealing Cone is a reusable product designed for use during open laparoscopy procedures. Used in conjunction with the MultAport Cannula and Blunt Spike, the Sealing Cone provides a means to maintain pneumoperitoneum.

Shaft Seal (replacement part for Sealing Cone):

The Sealing Cone is a reusable product designed for used during open laparoscopy procedures. Used in conjunction with the MultAport Cannula and Blunt Spike, the Sealing Cone provides a means to maintain pneumoperitoneum. The shaft seal is designed for use when sealing the cone to the cannula.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General Restorative Devices
510(k) NumberK990379
Prescription Use X(Per 21 CFR 801.109)OROver-The-Counter Use ____
(Optional Format 1-2-96)

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.